(178 days)
The Nova Stat Profile pHOx Ultra Analyzer with CO-Oximeter is intended for in vitro diagnostic use by health care professionals and for point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), total Hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), Deoxyhemoglobin (HHb), and total bilirubin (tBil) in heparinized whole blood; Na+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma. Total Bilirubin (tBil) was not evaluated on neonatal samples.
The Nova Stat Profile pHOx Ultra Analyzer without CO-Oximeter is intended for in vitro diagnostic use by health care professionals and/or point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Hemoglobin (Hb) in heparinized whole blood; Na+, K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma.
The intended use of the Nova STP pHOx Ultra Calibrator Cartridge is for the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Hemoglobin (Hb) in heparinized whole blood; Na+, K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma.
The intended use of the Nova Stat Profile pHOx Ultra Analyzer CO-Oximeter Calibrator Cartridge with Bilirubin and Deproteinizing Solution is for the quantitative determination of total Hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), Deoxyhemoglobin (HHb), and total bilirubin (tBil) in human blood using the Nova Stat Profile pHOx Ultra Analyzer System with CO-Oximeter.
Nova Stat Profile pHOx Ultra Analyzer CO-Oximeter Controls and Autocartridge QC are intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Nova Stat Profile pHOx Ultra Analyzer.
As in the Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System (K061830) predicate device, the Nova Stat Profile pHOx Ultra Analyzer System combines Blood Gas/pH, Chemistry, bilirubin and CO-Oximetry testing into one Point-of-Care Analyzer. This device is analyte configurable by the end user, based on tests needed.
As with the predicate, this device is microprocessor-based and incorporates:
- traditional electrode technology to measure blood pH, pCO2, pO2 .
- Nova Biomedical proprietary optical reflectance technology for the measurement of oxygen . saturation
- ion selective electrode technology to measure blood sodium, chloride, ionized . calcium, ionized magnesium
- enzyme/amperometric technology for glucose, urea nitrogen, lactate and creatinine . measurements
- conductivity/Na+ correction for hematocrit .
- multi-wavelength reflectance/conductivity correction for hemoglobin. .
Calibration standards with dissolved gases are provided in sealed pouches eliminating the need for users to calibrate the blood qas electrodes using external compressed gas cylinders. Quality control materials are available as external ampules and as internal auto-cartridge quality control packs. Sampling, calibration and quality control are fully automated.
Nova will market the Nova Stat Profile pHOx Ultra Analyzer System in two configurations. The proposed Nova Stat Profile pHOx Ultra Analyzer System with CO-Ox module (Catalog #42013) will be offered with all the parameters listed above. A second configuration will be offered, called the Nova Stat Profile pHOx Ultra Analyzer System without the CO-Ox module (Catalog #42014). This configuration will not have the capability to measure Oxyhemoglobin (O₂Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), and Reduced Hemoglobin (HHb) or bilirubin.
This 510(k) summary (K110648) describes the Nova Stat Profile pHOx Ultra Analyzer System as substantially equivalent to the predicate device, the Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System (K061830). The submission primarily focuses on demonstrating equivalence due to component obsolescence, rather than presenting a study to prove new performance criteria.
Therefore, the acceptance criteria are implicitly those established for the predicate device, and the "study" is a comparison and verification that the changes to the Nova Stat Profile pHOx Ultra Analyzer System did not affect the performance and maintain equivalence.
Here's the information broken down based on the provided text, recognizing the nature of this 510(k) (substantial equivalence based on component updates):
1. Table of Acceptance Criteria and Reported Device Performance
(Note: Since this is a substantial equivalence claim based on component changes, explicit acceptance criteria values for each analyte are not provided in this document. Instead, the "acceptance criteria" for the new device is to perform equivalently to the predicate. The reported device performance is that it met this equivalence.)
| Characteristic | Acceptance Criteria (based on predicate performance) | Reported Device Performance (Nova Stat Profile pHOx Ultra) |
|---|---|---|
| Analyte Measurement | Performance equivalent to predicate for all analytes (pH, PCO2, PO2, SO2%, Hct, tHb, O2Hb, COHb, MetHb, HHb, tBil, Na+, K+, Cl-, Ca++, Mg++, Glu, Lac, BUN, Creat) | "the results of software validation and performance verification testing confirmed that the modifications made to the hardware and software of the Nova Stat Profile pHOx Ultra Analyzer System did not affect the safety, efficacy or performance of the system is substantially equivalent to the predicate device." |
| User Interface | Functionality equivalent to predicate | "Functionality is equivalent to the predicate." |
| Accuracy | Equivalent to predicate | Confirmed through performance verification testing |
| Precision | Equivalent to predicate | Confirmed through performance verification testing |
| Reproducibility | Equivalent to predicate | Confirmed through performance verification testing |
| Software Validation | All software requirements met | Successfully validated |
| Performance Verification | All performance specifications met against predicate | Successfully verified |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size or data provenance (country of origin, retrospective/prospective) for the "performance verification testing." Given the nature of a 510(k) for component changes, this testing would typically involve a series of internal laboratory tests comparing the updated device's output to the predicate's output across a range of relevant samples (e.g., blood controls, patient samples).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This type of 510(k) submission, focused on component equivalence for an in vitro diagnostic device, typically relies on established reference methods or calibrated controls as the "ground truth" rather than expert consensus on individual cases. The document does not specify the use of "experts" to establish ground truth for a test set in the traditional sense of clinical imaging or diagnostic interpretation.
4. Adjudication Method for the Test Set
Not applicable. This submission does not describe a process involving adjudication of test results, as it is focused on technical equivalence to a predicate device rather than human interpretation or complex diagnostic assessment.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This is an IVD device for quantitative measurements, not an AI-assisted diagnostic imaging device. An MRMC study is not relevant to this submission.
6. Standalone Performance Study
Yes, in essence, a "standalone" performance verification for the new device was implicitly conducted against the predicate device. The submission states: "The results of software validation and performance verification testing confirmed that the modifications made to the hardware and software of the Nova Stat Profile pHOx Ultra Analyzer System did not affect the safety, efficacy or performance of the system is substantially equivalent to the predicate device." This indicates that the device was tested to ensure its standalone performance matches that of the predicate.
7. Type of Ground Truth Used
The ground truth for this device would be established through a combination of:
- Traceability to Reference Methods: Calibrators and controls used for the device would be traceable to recognized reference measurement procedures or certified reference materials for each analyte.
- Comparison to Predicate Device: The primary "ground truth" in this context is the established performance of the legally marketed predicate device (Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System). The new device's performance was verified to be equivalent.
8. Sample Size for the Training Set
Not applicable. This is not a machine learning or AI device that relies on a "training set" to learn. It is a measurement device with established electrochemical and optical principles.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" for this type of IVD device.
{0}------------------------------------------------
AUG 2 9 2011
510(k) Summary
| 510(K) Owner: | Nova Biomedical Corporation |
|---|---|
| Registration Number: | 1219029 |
| Address: | 200 Prospect St.Waltham, MA 02454 |
| Phone: | 781-894-0800 |
| Fax Number: | 784-891-4806 |
| Contact Person: | Paul W. MacDonald |
| Date Prepared: | 08/24/2011 |
Proprietary Name: Nova Stat Profile pHOx Ultra Analyzer System
Common or Usual Name: Blood Gas/Electrolyte/Metabolite/CO-Oximetry Analyzer
Classification Name: Multiple
| Classification Names: | Class No. | Reg. No. | Class |
|---|---|---|---|
| Blood Gases and Blood pH system | 75CHL | 862.1120 | II |
| Sodium Test System | 75JGS | 862.1665 | II |
| Potassium Test System | 75CEM | 862.1600 | II |
| Calcium Test System | 75JFP | 862.1145 | II |
| Chloride Test System | 75CGZ | 862.1170 | II |
| Creatinine Test System | 75CGL | 862.1225 | II |
| Magnesium Test System | 75CFA | 862.1495 | I |
| Glucose Test System | 75CGA | 862.1345 | II |
| Lactic Acid Test System | 75KHP | 862.1450 | I |
| Urea Nitrogen System | 75CDS | 862.1770 | II |
| Oxygen Saturation (SO2%) | 91JKS | 862.3220 | I |
| Hematocrit Measuring Device | 81JPI | 864.6400 | II |
| Hemoglobin System | 81GKR | 864.5620 | II |
| Oxyhemoglobin | 91JKS | 862.3220 | I |
| Carboxyhemoglobin | 81GHS | 864.7425 | II |
| Methemoglobin | 81JBJ | 864.7360 | II |
| Hemoglobin Standards and Controls | 81JPK | 864.8625 | II |
| Hemoglobin Calibrator | 81KRZ | 864.8165 | II |
| Calibrators | 75JIX | 862.1150 | II |
| Quality Control Materials | 81JJS | 862.1660 | I |
Product Codes: CHL, JGS, CEM, JFP, CGZ, CGL, CFA, CGA,KHP,CDS, JKS, GHS, JPI, GKR, JKS, JBJ, JPK,KRZ,JIX, JJS
Predicate Device: Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System (K061830)
{1}------------------------------------------------
Device Description:
As in the Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System (K061830) predicate device, the Nova Stat Profile pHOx Ultra Analyzer System combines Blood Gas/pH, Chemistry, bilirubin and CO-Oximetry testing into one Point-of-Care Analyzer. This device is analyte configurable by the end user, based on tests needed.
As with the predicate, this device is microprocessor-based and incorporates:
- traditional electrode technology to measure blood pH, pCO2, pO2 .
- Nova Biomedical proprietary optical reflectance technology for the measurement of oxygen . saturation
- ion selective electrode technology to measure blood sodium, chloride, ionized . calcium, ionized magnesium
- enzyme/amperometric technology for glucose, urea nitrogen, lactate and creatinine . measurements
- conductivity/Na+ correction for hematocrit .
- multi-wavelength reflectance/conductivity correction for hemoglobin. .
Calibration standards with dissolved gases are provided in sealed pouches eliminating the need for users to calibrate the blood qas electrodes using external compressed gas cylinders. Quality control materials are available as external ampules and as internal auto-cartridge quality control packs. Sampling, calibration and quality control are fully automated.
Nova will market the Nova Stat Profile pHOx Ultra Analyzer System in two configurations. The proposed Nova Stat Profile pHOx Ultra Analyzer System with CO-Ox module (Catalog #42013) will be offered with all the parameters listed above. A second configuration will be offered, called the Nova Stat Profile pHOx Ultra Analyzer System without the CO-Ox module (Catalog #42014). This configuration will not have the capability to measure Oxyhemoglobin (O₂Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), and Reduced Hemoglobin (HHb) or bilirubin.
Summary of the Technological Characteristics:
The Nova Stat Profile pHOx Ultra Analyzer System is identical to the previously cleared for market Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System in intended use. Modifications were made necessary due to the obsolescence and availability of certain components and hardware, which required minor modifications to the device software to support these changes. The software algorithms used in performing the analysis of samples are unchanged from the predicate device. Minor modifications have been made to simplify the User Interface. The same reagents, calibrators, and controls are used for all of the analytes. The comparison table below compares the Nova Stat Profile pHOx Ultra Analyzer System to the predicate Nova Stat Profile Critical Care Xpress (CCX), Model 1+ Analyzer System.
The results of software validation and performance verification testing confirmed that the modifications made to the hardware and software of the Nova Stat Profile pHOx Ultra Analyzer System did not affect the safety, efficacy or performance of the system is substantially equivalent to the predicate device.
{2}------------------------------------------------
Comparison Table
| Characteristic | PredicateStat Profile Critical Care Xpress with COOXAnalyzer System | ProposedStat Profile pHOx Ultra with COOX AnalyzerSystem |
|---|---|---|
| Intended Use | The Stat Profile Critical Care Xpress (STPCCX) Analyzer with CO-Oximeter is intendedfor in vitro diagnostic use by health careprofessionals and for point-of-care usage in thequantitative determination of pH, PCO2, PO2,SO2%, Hematocrit (Hct), total Hemoglobin(tHb), Oxyhemoglobin (O2Hb),Carboxyhemoglobin (COHb), Methemoglobin(MetHb), Deoxyhemoglobin (HHb), and totalbilirubin (tBil) in heparinized whole blood; Na+,K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate(Lac), BUN (Urea), and Creatinine (Creat) inheparinized whole blood, serum, or plasma.Total Bilirubin (tBil) was not evaluated onneonatalThe STP CCX Analyzer without CO-Oximeteris intended for in vitro diagnostic use by healthcare professionals and/or point-of-care usagein the quantitative determination of pH, PCO2,PO2, SO2%, Hematocrit (Hct), Hemoglobin(Hb) in heparinized whole blood; Na+, K+, Cl-,Ca++, Mg++, Glucose (Glu), Lactate (Lac),BUN (Urea), and Creatinine (Creat) inheparinized whole blood, serum, or plasmasamples.The intended use of the Nova STP CCX CO-Oximeter Calibrator Cartridge with Bilirubin andDeproteinizing Solution is for the quantitativedetermination of total Hemoglobin (tHb),Oxyhemoglobin (O2Hb), Carboxyhemoglobin(COHb), Methemoglobin (MetHb),Deoxyhemoglobin (HHb), and total bilirubin(tBil) in human blood using the STP CCXAnalyzer with CO-Oximeter.Nova STP CCX CO-Oximeter Controls andAutocartridge QC are intended for in vitrodiagnostic use by healthcare professionals formonitoring the performance of Nova STP CCXAnalyzers. | The Stat Profile pHOx Ultra Analyzer with CO-Oximeter is intended for in vitro diagnostic useby health care professionals and for point-of-care usage in the quantitative determination ofpH, PCO2, PO2, SO2%, Hematocrit (Hct), totalHemoglobin (tHb), Oxyhemoglobin (O2Hb),Carboxyhemoglobin (COHb), Methemoglobin(MetHb), Deoxyhemoglobin (HHb), and totalbilirubin (tBil) in heparinized whole blood; Na+,K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate(Lac), BUN (Urea), and Creatinine (Creat) inheparinized whole blood, serum, or plasma.Total Bilirubin (tBil) was not evaluated onneonatal samples.The Stat Profile pHOx Ultra Analyzer withoutCO-Oximeter is intended for in vitro diagnosticuse by health care professionals and/or point-of-care usage in the quantitative determinationof pH, PCO2, PO2, SO2%, Hematocrit (Hct),Hemoglobin (Hb) in heparinized whole blood;Na+, K+, Cl-, Ca++, Mg++, Glucose (Glu),Lactate (Lac), BUN (Urea), and Creatinine(Creat) in heparinized whole blood, serum, orplasma.The intended use of the Nova STP pHOx UltraCalibrator Cartridge is for the quantitativedetermination of pH, PCO2, PO2, SO2%,Hematocrit (Hct), Hemoglobin (Hb) inheparinized whole blood; Na+, K+, Cl-, Ca++,Mg++, Glucose (Glu), Lactate (Lac), BUN(Urea), and Creatinine (Creat) in heparinizedwhole blood, serum, or plasma.The intended use of the Nova STP pHOxUltra/CCX CO-Oximeter Calibrator Cartridgewith Bilirubin and Deproteinizing Solution is forthe quantitative determination of totalHemoglobin (tHb), Oxyhemoglobin (O2Hb),Carboxyhemoglobin (COHb), Methemoglobin(MetHb), Deoxyhemoglobin (HHb), and totalbilirubin (tBil) in human blood using STP pHOxUltra/CCX Analyzers with CO-OximeterNova STP pHOx Ultra/CCX ABG and CO-Oximeter Controls and Autocartridge QC areintended for in vitro diagnostic use byhealthcare professionals for monitoring theperformance of Nova STP pHOx Ultra/CCXAnalyzers. |
| Analyzer | Nova Stat Profile Critical Care Xpress (CCX),Model 1+ Analyzer System with CO-OX -37413Nova Stat Profile Critical Care Xpress (CCX),Model 1+ Analyzer System - 35942 | Nova Stat Profile pHOx Ultra Analyzer Systemwith CO-OX - 42013Nova Stat Profile pHOx Ultra Analyzer System- 42014 |
| Consumables | ||
| Calibrators | ABG, SO2, CO-Ox | Same |
| Controls | ABG Gas, ABG Chem, CO-Ox, External LinStandards | Same |
| Sensors | PH, PCO2, PO2, SO2, Na+, K+, Cl-, Ca++,Mg++, Glu, Lac, Creat, BUN, REF | Same |
| Membranes | Glu, Lac, Creat, BUN, PCO2, PO2 | Same |
| Characteristic | PredicateStat Profile Critical Care Xpress with COOXAnalyzer System | ProposedStat Profile pHOx Ultra with COOX AnalyzerSystem |
| Pump Tubing | ABG, CO-Ox | Same |
| Acceptable Samples | Syringe, Capillary | Same |
| Sample Volumes | ||
| ABG Only | 150μL | Same |
| CO-Ox Only | 100μL | Same |
| Combined Analysis | 210µL | Same |
| Micro SampleVolumes | ||
| Full | 120µL | Same |
| ABG | 60µL | Same |
| ABG Plus | 70µL | Same |
| ABG Nano | 70µL | Same |
| Sensor Slope Limits | ||
| pH | 8.7 - 11.6 | Same |
| PCO2 | 7.9 - 12.6 | Same |
| PO2 | 15.0 - (-1.6) | Same |
| SO2 | 8.0 - 18.0 | Same |
| Hct | 14.0 - 45.0 | Same |
| Hb | 40.0 - 98.0 | Same |
| Na+ | 8.8 - 11.0 | Same |
| K+ | 8.3 - 12.1 | Same |
| Cl- | 7.7 - 13.2 | Same |
| Ca++ | 8.3 - 12.6 | Same |
| Mg++ | 9.5 - 15.0 | Same |
| Glu | 5.0 - 60.0 | Same |
| Lac | 6.0 - 100.0 | Same |
| Creat | 3.0 - 25.0 | Same |
| BUN | 9.3 - 13.5 | Same |
| Principles ofMeasurement | ||
| pH | Hydrogen ion-selective glass sensor | Same |
| PCO2 | Severinghaus-type sensor | Same |
| PO2 | Polarographic Clark-type sensor | Same |
| SO2 | Reflectance photometry (fiber optics) | Same |
| Hct | Impedance sensor | Same |
| Hb | Impedance sensor | Same |
| Na+ | Sodium ion-selective glass sensor | Same |
| K+ | Potassium ion-selective sensor | Same |
| Cl- | Chloride ion-selective sensor | Same |
| Ca++ | Calcium ion-selective sensor | Same |
| Mg++ | Magnesium ion-selective sensor | Same |
| Glu | Enzymatic sensor | Same |
| Lac | Enzymatic sensor | Same |
| Creat | Enzymatic sensor | Same |
| BUN | Enzymatic sensor | Same |
| THb | Spectrophotometric | Same |
| O2Hb | Spectrophotometric | Same |
| COHb | Spectrophotometric | Same |
| MetHb | Spectrophotometric | Same |
| TBil | Spectrophotometric | Same |
| Design | ||
| Touch Screen | 10.4" LCD, 640x480 pixel, Resistive Touch | 12.1" LCD, 1024x768 pixel, Resistive Touch |
| Menu | Fully configurable test menu based on abovesensors | Test menu has not changed. |
| Printer | 4.25" Roll, Thermal Transfer | 2" Roll, Thermal Transfer |
| Characteristic | Predicate | Proposed |
| Chassis | Stat Profile Critical Care Xpress with COOX Analyzer System | Stat Profile PHOx Ultra with COOX Analyzer System |
| Molded GF PC Base w/ Aluminum Welded Frame and Molded ABS/PC Cover. | Aluminum Riveted and Welded Base and Frame w/ Aluminum Cover | |
| ABG System | ||
| ABG Sensor Module | Combined Flow-through/Flow-by w/ optical SO2 | Same |
| ABG Probe/S-Line | Stainless Probe w/ Polyurethane and Saran flexible tubing, In-Line Air detector | No Polyurethane used, shorter stainless steel portion dictated use of less gas permeable material. |
| ABG Pump | Peristaltic Pump w/ Pressure Plate, TPE Tubing (Pharmed BPT) | Same |
| Reagent Valves | Ceramic Shear Valves | Same |
| Other Valves | Pinch Style, Extruded TPE and Molded Silicone Tubing | Same |
| CO-Ox System | ||
| CO-Ox Probe/S-Line | Stainless Probe w/ Polyurethane and Saran flexible tubing, In-Line Air detector | No Polyurethane used, shorter stainless steel portion dictated use of less gas permeable material. |
| CO-Ox Pump | Peristaltic Pump w/o Pressure Plate, Molded Silicone Tubing (Q7-4840) | Peristaltic Pump w/ Pressure Plate, Extruded TPE Tubing (Pharmed BPT) |
| Reagent Valves | Ceramic Shear Valves | Same |
| Spectrophotometer | Broad Spectrum Spectrometer | Same |
| Lamp House | Krypton Calibration Lamp / LED Light Source | Same |
| Cuvette Slide | Machined aluminum pre-heated slide w/ acrylic | Cast aluminum pre-heated slide w/ acrylic and |
| Assembly | and sapphire cuvette assembly | sapphire cuvette assembly |
| Analog Board | precision low level analog front end w/amperometric and potentiometric amplifiers, air detector circuitry and temperature control circuitry | Same |
| Deproteinizing | "bleach type" agent can be scheduled to run | Same process, but spectrophotometer |
| Algorithm | through COOX branch for regulardeproteinization | compares cuvette absorbance before and after sample, and schedules immediate deproteinization if difference is detected. |
| CoHb algorithm | Multi-wavelength Linear RegressionChemometrics | Same |
| tBil algorithm | Multi-wavelength Linear RegressionChemometrics | Same |
| Hemolyzer | Chemically Lysed with mechanical mixing | Same |
| Software/HWArchitecture | ||
| Graphical User | CCX user interface | Interface updated and tested as part of software validation. Functionality is equivalent to the predicate. |
| Interface (GUI) | ||
| Storage media | IDE Hard Drive | SATA Hard Drive |
| Operating System | Windows 2000 Embedded | Windows 7 Embedded |
| Host Computer | ibase IB880 | Axiomtek 84831 |
, •
{3}------------------------------------------------
{4}------------------------------------------------
:
・
:
.
.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features an eagle-like bird symbol on the right side. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the left side of the seal. The text is in all capital letters.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Silver Spring, MD 20993
Food and Drug Administration
10903 New Hampshire Avenue
Nova Biomedical Corporation c/o Paul W. MacDonald 200 Prospect St. Waltham, MA 02454
AUG 2 9 2011
Re: K110648
Trade/Device Name: Nova Stat Profile pHOx Ultra Analyzer System with CO-OX Regulation Number: 21 CFR 862.1120 Regulation Name: Blood gases (pCO2, pO2) and blood pH test system Regulatory Class: Class II Product Code: CHL, JGS, CEM, CGZ, JFP, CFA, NGD, CDS, CGL, GGZ, GLY, GKK, GHS, GJC, JJB, CGA, BXX, JIX Dated: 2 August 2011 Received: 4 August 2011
Dear Dr. MacDonald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{6}------------------------------------------------
Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation (27 CFK Pat(1) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding of reference to prematics of prematics under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillace.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
C.C.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
: 1441
Enclosure
{7}------------------------------------------------
Indications for Use Form
| 510(k) Number (if known): | K110648 |
|---|---|
| --------------------------- | --------- |
Nova Stat Profile pHOx Ultra Analyzer System Device Name:
Indications for Use:
The Nova Stat Profile pHOx Ultra Analyzer with CO-Oximeter is intended for in vitro diagnostic use by health care professionals and for point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), total Hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), Deoxyhemoglobin (HHb), and total bilirubin (tBil) in heparinized whole blood; Na+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma. Total Bilirubin (tBil) was not evaluated on neonatal samples.
The Nova Stat Profile pHOx Ultra Analyzer without CO-Oximeter is intended for in vitro diagnostic use by health care professionals and/or point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Hemoglobin (Hb) in heparinized whole blood; Na+, K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma.
The intended use of the Nova STP pHOx Ultra Calibrator Cartridge is for the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Hemoglobin (Hb) in heparinized whole blood; Na+, K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Qus
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K110648
Page 1 of 2
{8}------------------------------------------------
Indications for Use Form
510(k) Number (if known): K110648
Nova Stat Profile pHOx Ultra Analyzer System Device Name:
Indications for Use:
The intended use of the Nova Stat Profile pHOx Ultra Analyzer CO-Oximeter Calibrator Cartridge with Bilirubin and Deproteinizing Solution is for the quantitative determination of total Hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), Deoxyhemoglobin (HHb), and total bilirubin (tBil) in human blood using the Nova Stat Profile pHOx Ultra Analyzer System with CO-Oximeter.
Nova Stat Profile pHOx Ultra Analyzer CO-Oximeter Controls and Autocartridge QC are intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Nova Stat Profile pHOx Ultra Analyzer.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K110648
Page 2 of 2
§ 862.1120 Blood gases (P
CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.