(215 days)
No
The summary describes a standard blood gas analyzer that measures and calculates parameters based on established physiological formulas. There is no mention of AI or ML in the device description, intended use, or performance studies.
No
This device is an analyzer system used for diagnostic purposes (measuring pH, PCO2, PO2 in blood) to aid in the diagnosis and treatment of acid-base disturbances, but it does not directly provide therapy or therapeutic effect.
Yes
The "Intended Use / Indications for Use" section explicitly states that the measurements obtained from this device are "used in the diagnosis and treatment of life-threatening acid base disturbances." This directly indicates a diagnostic purpose.
No
The device description explicitly states it consists of hardware components such as the analyzer, sensor cartridges, and thermal paper for an onboard printer.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The device is indicated for the "quantitative determination of pH, Partial Pressure of Carbon Dioxide, and Partial Pressure of Oxygen in heparinized arterial and venous whole blood." This involves testing biological samples (whole blood) in vitro (outside the body).
- Purpose: The measurements are used "in the diagnosis and treatment of life-threatening acid base disturbances." This clearly indicates a diagnostic purpose.
- Device Description: It analyzes blood samples using sensor cartridges to measure specific analytes.
- Sample Types: It accepts "lithium heparinized arterial and venous whole blood," which are biological specimens.
- Measured Parameters: It measures pH, pCO2, and pO2, which are clinical parameters determined from the analysis of the blood sample.
All these points align with the definition of an In Vitro Diagnostic device, which is used to examine specimens taken from the human body to provide information for diagnostic, monitoring, or compatibility purposes.
N/A
Intended Use / Indications for Use
The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings for quantitative determination of pH, Partial Pressure of Carbon Dioxide, and Partial Pressure of Oxygen in heparinized arterial and venous whole blood.
pH, pCO2, pO2 Measurements are used in the diagnosis and treatment of life-threatening acid base disturbances.
Product codes (comma separated list FDA assigned to the subject device)
CHL
Device Description
The Stat Profile Prime Plus Analyzer System is designed to be a low cost, low maintenance analyzer for the hospital laboratory setting. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).
The system architecture and user interface for this proposed device is based on the previously cleared Stat Profile Prime CCS Analyzer System (K131703). The primary predicate for this proposed device is the Stat Profile pHOx Ultra Analyzer System (K110648).
The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed.
Primary Sensor Card Port:
There are two options for the primary sensor card:
- Primary Sensor Card 1 shall enable and report the following listed analytes: oPO2, PCO2, pH
- Primary Sensor Card 2 shall enable and report the following listed analytes: o PO2, PCO2, pH
Similar to the primary predicate device, the Stat Profile Prime Plus Analyzer is a blood gas/cooximetry/electrolyte/chemistry and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control will be used, as well as an on-board Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly at all times.
The Stat Profile Prime Plus Analyzer accepts samples from syringes, and small cups. The minimum sample size for analysis is 135 uL.
Sample collection, preparation and application to the analyzer are the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel.
Stat Profile Prime Plus Analyzer System Components:
The Stat Profile Prime Plus Analyzer System is comprised of the following components.
- Stat Profile Prime Plus Analyzer System
- Primary Sensor Cartridge
- Auxiliary Sensor Cartridge
- Stat Profile Prime Plus Auto-Cartridge Quality Control Pack
- Stat Profile Prime Plus Calibrator Cartridge
- Stat Profile Prime Plus External Ampuled Control
- IFU/Labeling
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
healthcare professionals in clinical laboratory settings
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance testing was completed to demonstrate that the Stat Profile Prime Plus Analyzer is substantially equivalent in performance, safety and efficacy to the Stat Profile pHOx Ultra Analyzer System. The performance testing included:
Method Comparison Studies:
A study was performed to compare the Stat Profile Prime Plus to the Nova Stat Profile pHOx Ultra analyzer to assess the equivalence of the analyzers in the measurement of pH, PCO2, and PO2 in heparinized whole blood in a clinical laboratory setting.
The blood comparison data for pH, PCO2, and PO2 for the Stat Profile Prime Plus analyzers meet the acceptance criteria.
Precision/Reproducibility - Within Run and Run to Run Studies:
Within Run and Run to Run precision was evaluated by replication studies performed on three Stat Profile Prime Plus analyzers.
Within Run Precision testing consisted of one run of each of the following sample types and levels was performed, 20 replicates per run:
- Stat Profile Prime Plus Internal Controls: Levels 1- 3
- Stat Profile Prime Plus Ampuled Controls: Levels 1-3
- Two whole bloods, sampled from syringes
To assess Run to Run Precision for whole blood, triplicate analyses were performed on a single whole blood sample in ten separate runs during a single day. The systems were recalibrated before each triplicate run.
The precision data for all parameters meet the within run imprecision specifications for the Stat Profile Prime Plus analyzers.
The precision data for all parameters meet the between analyzer run to run imprecision specifications for the Stat Profile Prime Plus analyzers.
Linearity Testing:
The study assessed the linearity of all parameters to establish and/or verify the Analytical Measurement Range (AMR) for the Stat Profile Prime Plus Analyzer on whole blood. The linearity was measured using method comparison. The evaluation of the linear range included lower and upper limits of the AMR and various medical decision limits.
All Stat Profile Prime Plus results were compared to the reference analyzer and/or the product specifications.
The linearity comparison data for all parameters for the Stat Profile Prime Plus analyzers shows good correlation and linearity to the reference analyzers across the claimed measurement range for all parameters and met the acceptance criteria.
Specificity / Interference Testing:
The purpose of this study was to identify substances that may interfere with the Stat Profile Prime Plus sensors. If interference was identified, a dose response study was performed to determine the concentration where the interfering substance may alter results.
This Interference Testing study was performed using whole blood collected in lithium heparin vacutainers. The possible interfering substances were tested at two analyte concentrations.
Many substances were screened as potential interferents through analysis on the Prime Plus analyzers. Substance screening was completed according to the CLSI EP7-A2 guideline. There were no interfering substances related to pH, pCO2, or pO2 identified in the study.
Conclusion:
The results of software validation and performance verification testing confirmed that the Stat Profile Prime Plus Analyzer is safe and effective for its intended purpose and that the Stat Profile Prime Plus Analyzer System is substantially equivalent to that of the predicate Stat Profile pHOx Ultra Analyzer System (K110648).
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1120 Blood gases (P
CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 17, 2018
Nova Biomedical Corporation Cesidio Tempesta Sr. Regulatory Affairs Specialist 200 Prospect St. Waltham, MA 02454
Re: K173797
Trade/Device Name: Stat Profile Prime Plus Analyzer System Regulation Number: 21 CFR 862.1120 Regulation Name: Blood gases (PCO2, PO2) and blood pH test system Regulatory Class: Class II Product Code: CHL Dated: June 8, 2018 Received: June 11, 2018
Dear Cesidio Tempesta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
1
Page 2 - Cesidio Tempesta
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known)
Device Name
Stat Profile® Prime Plus Analyzer System
Indications for Use (Describe)
The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings for quantitative determination of pH, Partial Pressure of Carbon Dioxide, and Partial Pressure of Oxygen in heparinized arterial and venous whole blood.
pH, pCO2, pO2 | Measurements are used in the diagnosis and treatment of life-threatening acid base disturbances. |
---|---|
--------------- | -------------------------------------------------------------------------------------------------- |
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
---|---|
--------------------------------------------------------------------------------------------------------------------------------- | -------------------------------------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary K173797
510(K) Owner: | Nova Biomedical Corporation |
---|---|
Registration Number: | 1219029 |
Address: | 200 Prospect St. |
Waltham, MA 02454 | |
Phone: | 781-894-0800 |
Fax Number: | 784-891-4806 |
Contact Person: | Cesidio Tempesta, Regulatory Affairs Specialist |
Date Prepared: | July 17, 2018 |
Proprietary Name: Stat Profile Prime Plus Analyzer System
Common or Usual Name: Blood Gas Analyzer
Classification Name: Blood Gases and Blood pH system
Product Code: CHL
Predicate Device: K110648 - Stat Profile pHOx Ultra Analyzer System
Device Description:
The Stat Profile Prime Plus Analyzer System is designed to be a low cost, low maintenance analyzer for the hospital laboratory setting. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).
The system architecture and user interface for this proposed device is based on the previously cleared Stat Profile Prime CCS Analyzer System (K131703). The primary predicate for this proposed device is the Stat Profile pHOx Ultra Analyzer System (K110648).
The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed.
Primary Sensor Card Port:
There are two options for the primary sensor card:
- Primary Sensor Card 1 shall enable and report the following listed analytes: o PO2, PCO2, pH
- . Primary Sensor Card 2 shall enable and report the following listed analytes: o PO2, PCO2, pH
Similar to the primary predicate device, the Stat Profile Prime Plus Analyzer is a blood gas/cooximetry/electrolyte/chemistry and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control will be used, as well as an on-board Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly at all times.
4
The Stat Profile Prime Plus Analyzer accepts samples from syringes, and small cups. The minimum sample size for analysis is 135 uL.
Sample collection, preparation and application to the analyzer are the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel.
Stat Profile Prime Plus Analyzer System Components:
The Stat Profile Prime Plus Analyzer System is comprised of the following components.
- . Stat Profile Prime Plus Analyzer System
- Primary Sensor Cartridge
- Auxiliary Sensor Cartridge ●
- Stat Profile Prime Plus Auto-Cartridge Quality Control Pack .
- Stat Profile Prime Plus Calibrator Cartridge ●
- Stat Profile Prime Plus External Ampuled Control
- . IFU/Labeling
Sample Types:
The Stat Profile Prime Plus Analyzer System accepts lithium heparinized arterial and venous whole blood.
Measured Parameters:
The Stat Profile Prime Plus Analyzer measures:
- . pH
- Partial Pressure of Carbon Dioxide (pCO2)
- Partial Pressure of Oxygen (pO2)
Calculated Parameters:
The following parameters are calculated by the Prime Plus Analyzer based on results of the directly measured parameters.
- Base Excess of Extracellular Fluid (BE-ecf)
- Base Excess of the Blood (BE-b)
- Standard Bicarbonate Concentration (SBC) ●
- Bicarbonate level (HCO3) ●
- Oxygen Content (O2Ct) .
- P50 ●
- Respiratory Index (RI)
- PO2 to FIO2 Ratio (PO2/FIO2) ●
- Normalized Calcium (nCa) ●
- Normalized Magnesium (nMq)
- Anion Gap (Gap) ●
- nCA to nMg Ratio (nCa/nMg) ●
- Arterial Oxygen Content (CaO3) .
Intended Use:
The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings for quantitative determination of pH, Partial Pressure of Carbon Dioxide, and Partial Pressure of Oxygen in heparinized arterial and venous whole blood.
pH, pCO2, pO2 | Measurements are used in the diagnosis and treatment of life-threatening acid base disturbances. |
---|---|
--------------- | -------------------------------------------------------------------------------------------------- |
5
Summary of the Technological Characteristics:
The Stat Profile Prime Plus Analyzer is substantially equivalent to the previously cleared for market Stat Profile pHOx Ultra Analyzer System (K110648) in intended use. It uses the same sensor technology and measurement algorithms, and the formulations of the internal and external controls and the calibration cartridge are the same for the tested parameters. The External Control solutions for use with the Stat Profile Prime Plus Analyzer are substantially equivalent to those cleared for use with the predicate Stat Profile pHOx Ultra Analyzer System (K110648). See Table 1 for comparison of the predicate and proposed devices.
Principles of Measurement:
pH:
pH is measured using a hydrogen ion selective glass membrane. One side of the glass is in contact with a solution of constant pH. The other side is in contact with a solution of unknown pH. A change in potential develops which is proportional to the pH difference of these solutions. This change in potential is measured against a reference electrode of constant potential. The magnitude of the potential difference is a measure, then, of the pH of the unknown solution.
pCO2:
PCO2 is measured with a modified pH sensor. Carbon dioxide in the unknown solution makes contact with a gas permeable membrane mounted on a combination measuring/ reference electrode. CO2 diffuses across the membrane into a thin layer of electrolyte solution in response to partial pressure difference. This solution then becomes equilibrated with the external gas pressure. CO2 in the solution becomes hydrated producing carbonic acid, which results in a change in hydrogen ion activity.
pO2:
PO2 is measured amperometrically by the generation of a current at the sensor surface. As oxygen diffuses through a gas permeable membrane, the oxygen molecules are reduced at the cathode, consuming 4 electrons for every molecule of oxygen reduced. This flow of electrons is then measured by the sensor and is directly proportional to the partial pressure of oxygen.
Summary of Performance Testing:
Performance testing was completed to demonstrate that the Stat Profile Prime Plus Analyzer is substantially equivalent in performance, safety and efficacy to the Stat Profile pHOx Ultra Analyzer System. The performance testing included:
Method Comparison Studies:
A study was performed to compare the Stat Profile Prime Plus to the Nova Stat Profile pHOx Ultra analyzer to assess the equivalence of the analyzers in the measurement of pH, PCO2, and PO2 in heparinized whole blood in a clinical laboratory setting.
The blood comparison data for pH, PCO2, and PO2 for the Stat Profile Prime Plus analyzers meet the acceptance criteria.
Precision/Reproducibility - Within Run and Run to Run Studies:
Within Run and Run to Run precision was evaluated by replication studies performed on three Stat Profile Prime Plus analyzers.
Within Run Precision testing consisted of one run of each of the following sample types and levels was performed, 20 replicates per run:
- . Stat Profile Prime Plus Internal Controls: Levels 1- 3
- . Stat Profile Prime Plus Ampuled Controls: Levels 1-3
- . Two whole bloods, sampled from syringes
To assess Run to Run Precision for whole blood, triplicate analyses were performed on a single whole blood sample in ten separate runs during a single day. The systems were recalibrated before each triplicate run.
6
The precision data for all parameters meet the within run imprecision specifications for the Stat Profile Prime Plus analyzers.
The precision data for all parameters meet the between analyzer run to run imprecision specifications for the Stat Profile Prime Plus analyzers.
Linearity Testing:
The study assessed the linearity of all parameters to establish and/or verify the Analytical Measurement Range (AMR) for the Stat Profile Prime Plus Analyzer on whole blood. The linearity was measured using method comparison. The evaluation of the linear range included lower and upper limits of the AMR and various medical decision limits.
All Stat Profile Prime Plus results were compared to the reference analyzer and/or the product specifications.
The linearity comparison data for all parameters for the Stat Profile Prime Plus analyzers shows good correlation and linearity to the reference analyzers across the claimed measurement range for all parameters and met the acceptance criteria.
Specificity / Interference Testing:
The purpose of this study was to identify substances that may interfere with the Stat Profile Prime Plus sensors. If interference was identified, a dose response study was performed to determine the concentration where the interfering substance may alter results.
This Interference Testing study was performed using whole blood collected in lithium heparin vacutainers. The possible interfering substances were tested at two analyte concentrations.
Many substances were screened as potential interferents through analysis on the Prime Plus analyzers. Substance screening was completed according to the CLSI EP7-A2 guideline. There were no interfering substances related to pH, pCO2, or pO2 identified in the study.
Conclusion:
The results of software validation and performance verification testing confirmed that the Stat Profile Prime Plus Analyzer is safe and effective for its intended purpose and that the Stat Profile Prime Plus Analyzer System is substantially equivalent to that of the predicate Stat Profile pHOx Ultra Analyzer System (K110648).
7
Table 1: Comparison of Predicate and Proposed devices
Characteristic | Predicate: K110648 - Stat Profile pHOx Ultra Analyzer | Proposed: Stat Profile Prime Plus Analyzer |
---|---|---|
Indication For Use | The Stat Profile pHOx Ultra Analyzer without CO-Oximeter is intended for in | |
vitro diagnostic use by health care professionals and/or point-of-care usage in the | ||
quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), | ||
Hemoglobin (Hb) in heparinized whole blood; Na+, K'-, Cl-, Ca++, Mg++, Glucose | ||
(Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole | ||
blood, serum, or plasma. |
The Stat Profile pHOx Ultra Analyzer with CO-Oximeter is intended for in vitro
diagnostic use by health care professionals and for point-of-care usage in the
quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), total
Hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb),
Methemoglobin (MetHb), Deoxyhemoglobin (HHb), and total bilirubin (tBil) in
heparinized whole blood; Na+, K'-, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac),
BUN (Urea), and Creatinine (Creat) in Heparinized whole blood, serum, or
plasma. Total Bilirubin (tBil) was not evaluated on neonatal samples. | The Stat Profile Prime Plus Analyzer System is indicated for use by
healthcare professionals in clinical laboratory settings for quantitative
determination of pH, Partial Pressure of Carbon Dioxide, and Partial
Pressure of Oxygen in heparinized arterial and venous whole blood. |
| Acceptable Samples | | |
| Sample Types | Sodium or lithium heparinized whole blood, serum, or plasma samples from
syringes, open tubes, small cups, and capillary tubes. | Lithium heparin whole blood from syringes, open tubes, small cups, and
capillary tubes. |
| Sample Volumes | 60-200µL (dependent on panel selected) | 135µL |
| Measurement Range | | |
| pH | 6.500 - 8.000 | 6.500 to 8.000 (pH units) |
| PCO2 | 3.0 - 200 mmHg | 3.0 to 200.0 mmHg |
| PO2 | 0 - 800 mmHg | 5.0 to 765 mmHg |
| Principles of
Measurement | | |
| pH | Hydrogen ion-selective glass sensor | Same |
| PCO2 | Severinghaus-type sensor | Same |
| PO2 | Polarographic Clark-type sensor | Same |
| | | |
| Touch Screen | 12.1" LCD, 1024x768 pixel, Resistive Touch | 10.1" WXGA 1280 x 800 color touch screen |
| Menu | Fully configurable test menu based on available sensors | Same |
| Bar Code Scanner | Internal Integrated 1D/2D | Same |
| Printer | 2" Roll, Thermal Transfer | Same |
| Pump | Peristaltic Pump w/ Pressure Plate, TPE Tubing (Pharmed BPT) | Same |
| Analog Board | Precision low level analog front end w/ amperometric and potentiometric
amplifiers, air detector circuitry and temperature control circuitry | Same |