K Number
K172890

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-01-26

(126 days)

Product Code
Regulation Number
882.1400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SedLine Sedation Monitor is intended to monitor the state of the data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentanil, Desflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.

Device Description

SedLine® Sedation Monitor is a patient-connected, 4-channel processed Electroencephalograph (EEG) monitor designed specifically for intraoperative or intensive care use. It displays electrode status, EEG waveforms, Density Spectral Array (DSA), and Patient State Index (PSI), EMG Index, Suppression Ratio (SR) and Artifact (ARTF). The operator controls the unit using menus and dedicated buttons to select various display options. The system consists of 4 major components: Root, SedLine Module, SedLine Patient Cable, and SedLine Sensor.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Key Takeaway: This 510(k) submission is for a modification to an existing device (SedLine Sedation Monitor), specifically a modified Patient State Index (PSI) algorithm and an optional additional DSA display. The primary goal is to demonstrate substantial equivalence to the predicate device.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a "table of acceptance criteria" with corresponding "reported device performance" in the typical format of a clinical trial results table that would define specific metrics and thresholds for success for the new PSI algorithm. Instead, the non-clinical testing section lists various standards and guidances it aimed to satisfy, and the clinical testing section describes how the new algorithm was compared to the predicate.

However, based on the context of a 510(k) submission for a device modification, the implicit acceptance criteria would revolve around demonstrating that the modified device performs at least as well as, or is substantially equivalent to, the predicate device. The general performance specifications for the SedLine Sedation Monitor are provided in Table 5.2.

Implicit Acceptance Criteria and Reported Performance (derived from document):

Acceptance Criteria CategorySpecific Areas (Implicit)Reported Performance/Outcome (from text)
Non-Clinical Performance (Safety & Functionality)Electrical Safety (IEC 60601-1)Satisfied all requirements and performance specifications.
EMC (IEC 60601-1-2)Satisfied all requirements and performance specifications.
Alarm Testing (IEC 60601-1-8)Satisfied all requirements and performance specifications.
Usability (FDA Human Factors & Usability Draft Guidance)Satisfied all requirements and performance specifications.
Software Verification (FDA Software Guidance)Validated DSA by comparing multi-taper DSA against the predicate's Hanning DSA for known input signals, testing: dynamic range, frequency range, spectral edge frequency, and high-contrast feature. Satisfied all requirements and performance specifications.
Mechanical Testing (EN 60601-2-26)Satisfied all requirements and performance specifications.
Environmental Testing (EN 60601-2-26)Satisfied all requirements and performance specifications.
Clinical Performance (Equivalence of new PSI algorithm)Comparison of subject PSI algorithm to predicate PSI algorithm"The subject PSi algorithm was compared to the predicate PSi algorithm." (Implies the comparison satisfied the criteria for substantial equivalence, though specific statistical equivalence metrics are not detailed in this summary). The overall conclusion is that clinical testing "demonstrates that the subject device... is substantially equivalent to its predicate."

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size for Test Set: 100 surgical patients
  • Data Provenance:
    • Country of Origin: Not explicitly stated, but the submission is to the FDA (USA), and typically, studies cited in such submissions are either US-based or explicitly noted if international.
    • Retrospective or Prospective: Retrospective analysis of clinical data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not specify the number or qualifications of experts used to establish ground truth for the test set. For this type of device (Sedation Monitor), the "ground truth" for the PSI algorithm's performance is typically related to the administered anesthetic drug doses and observed physiological responses, rather than expert interpretation of EEG waveforms alone for classifying sedation depth. The study states "Clinical data used for the analysis includes continuous EEG, anesthetic drug dose information, and other physiological vital signs such as mean arterial blood pressure and heart rate," which would serve as the reference against which the PSI algorithm's output is compared.


4. Adjudication Method for the Test Set

The document does not mention any adjudication method for the test set. Given it's a retrospective analysis of clinical data including objective measurements (EEG, drug doses, vital signs), an adjudication process involving multiple human readers for "ground truth" might not have been applied in the same way as, for example, in an imaging study.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

No, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed or described. This study focused on the performance of the modified algorithm itself in comparison to its predicate, using retrospective clinical data. The SedLine Sedation Monitor is an monitoring device that provides a computed index (PSI) for clinicians to interpret, it's not a diagnostic AI intended to assist human interpretation of complex images in an MRMC setting.


6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, a standalone performance evaluation of the algorithm was done. The clinical study describes a "retrospective analysis of the clinical data" where "The subject PSi algorithm was compared to the predicate PSi algorithm" using collected physiological and drug administration data. This directly assesses the algorithm's output (PSI) based on its input (EEG signals, etc.) without human intervention in its calculation or interpretation to determine its output.


7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The ground truth for evaluating the PSI algorithm's relation to anesthetic depth would be based on:

  • Anesthetic drug dose information: The type and amount of anesthetic agents administered.
  • Physiological vital signs: Mean arterial blood pressure and heart rate.
  • Continuous EEG data: The raw EEG signals from which the PSI is derived, allowing for comparison of the algorithm's output against the expected EEG changes under anesthesia.

Combined, these elements serve as the reference for the "effect of anesthetic agents" that the PSI is designed to reflect. It's not a single "expert consensus" or "pathology" but rather a composite of objective clinical data related to the patient's state of anesthesia.


8. The Sample Size for the Training Set

The document does not provide the sample size for the training set. It only mentions the "validation" of the subject algorithm through retrospective analysis of clinical data in 100 surgical patients. This 100-patient dataset appears to be the test/validation set for assessing the modified algorithm, not necessarily a training set. Given that this is a modification of an existing algorithm, the original algorithm would have been developed and trained using prior data, but details about that are not included here.


9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for the training set (if a separate training set was used for the modified algorithm) was established. It only describes the data used for the validation comparison of the subject algorithm against the predicate. For the original development of the PSI, ground truth would typically involve correlating EEG patterns with known states of consciousness/sedation induced by controlled anesthetic administration, likely established by expert assessment (e.g., Riker Sedation-Agitation Scale, Observer's Assessment of Alertness/Sedation Scale) and objective physiological markers during prospective studies. However, these details are absent for this specific submission's context.

{0}------------------------------------------------

January 26, 2018

Masimo Corporation Matthew Tiacharoen Regulatory Affairs Specialist II, Regulatory Affairs 52 Discovery Irvine, California 92618

Re: K172890

Trade/Device Name: SedLine Sedation Monitor Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLW, GXY, OLT, OMC, ORT Dated: December 22, 2017 Received: December 26, 2017

Dear Matthew Tiacharoen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tara A. Ryan -
S

for

Digitally signed by Tara A. Ryan -S
DN: c=US, o=U.S. Government, ou=HHS,
ou=FDA, ou=People, cn=Tara A. Ryan -S,
0.9.2342.19200300.100.1.1=1300030749
Date: 2018.01.26 14:48:19 -05'00'

Tina Kiang, Ph.D.
Acting Director
Division of Anesthesiology,
General Hospital, Respiratory,
Infection Control, and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172890

Device Name SedLine Sedation Monitor

Indications for Use (Describe)

The SedLine Sedation Monitor is intended to monitor the state of the data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentanil, Desflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.

Submitter and Address ofManufacturing Facility:Masimo Corporation52 DiscoveryIrvine, CA 92618Phone: (949) 297-7982FAX: (949) 297-7199
Date:January 26, 2018
Contact:Matthew TiacharoenRegulatory Specialist II, Regulatory Affairs
Trade Name:SedLine Sedation Monitor
Common Name:Brain Function Monitor
Classification Regulation/Product Code:21 CFR 882.1400, Class II/OLW
Additional Product Codes:GXY,OLT, OMC, ORT
EstablishmentRegistration Number:2031172
Reason for PremarketNotification:Device modification
Predicate Device:K140188 - Masimo Root Monitoring System
Performance StandardsNo performance standards for the above device have beenpromulgated pursuant to Section 514 of the Food and DrugAdministration Modernization Act of 1997 (FDAMA)

5.1 Device Description

SedLine® Sedation Monitor is a patient-connected, 4-channel processed Electroencephalograph (EEG) monitor designed specifically for intraoperative or intensive care use. It displays electrode status, EEG waveforms, Density Spectral Array (DSA), and Patient State Index (PSI), EMG Index, Suppression Ratio (SR) and Artifact (ARTF).

The operator controls the unit using menus and dedicated buttons to select various display options. The system consists of 4 major components: Root, SedLine Module, SedLine Patient Cable, and SedLine Sensor.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in bold, black letters. To the right of the logo is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.

Note: For clarification, the purpose of this submission is to demonstrate that the subject device, SedLine® Sedation Monitor, is the substantially equivalent to the predicate device's SedLine Module.

5.2 Significant Physical and Performance Characteristics of the Device

The device specifications are shown below for the general functions of the subject device.

TABLE 5.2 SedLine® Sedation Monitor Specifications
FEATURESPECIFICATION
Display
Display Range
PSI0 to 100
EMG0 to 100%
SR0 to 100%
ARTF0 to 100%
DSA Amplitude (Leftand Right-60 to 40 dB
SEFL/SEFR0-30Hz
DSA Asymmetry-100% to +100%
Electrode Impedance0 to 65 KOhms
Resolution
PSI1
EMG1%
SR2%
ARTF1%
DSA Amplitude (Leftand Right≤1db
SEFL/SEFR1 Hz
DSA Asymmetry1%
Electrode Impedance1 KOhm
General
Visual/audible alarmHost/Backboard Device (Masimo Root Monitoring System) isIEC60601-1-8 compliant perK140188
Storage/recordingMasimo Root Monitoring System has trend/data storage perK140188
Electrical
AC PowerHost/Backboard Device (Masimo Root Monitoring System)provides AC power per K140188
Rechargeable batteryHost/Backboard Device (Masimo Root Monitoring System)provides internal battery power per K140188
Interface
SedLine ModuleConnectionMOC-9 interface with Host/Backboard device (Masimo RootMonitoring System, per K140188)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.

TABLE 5.2 SedLine® Sedation Monitor Specifications
FEATURESPECIFICATION
Mechanical
Module: Dimensions1 3/10 in (3.3 cm) x 4 in (10.2 cm) x .8 in (2.0 cm)
Environmental
Operating Conditions
Temperature+41°F to +104°F (+5°C to +40°C)
Humidity15% to 95%, non-condensing
Storage Conditions
Temperature-40°F to +158°F (-40°C to +70°C)
Humidity15–95%, non-condensing
Pressure500 to 1060 mbar

5.3 Intended Use

The SedLine® Sedation Monitor is indicated for use in the operating room (OR), intensive care unit (ICU), and clinical research laboratory. It is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The computed values are displayed on a Host/Backboard monitor such as the Masimo Root Monitoring System (K140188).

5.4 Indications For Use

The SedLine Sedation Monitor is indicated for use in the operating room (OR), intensive care unit (ICU), and clinical research laboratory. It is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals for adult patients (18 years of age and older). The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents include: Alfentanil, Desflurane, Fentanyl, Isoflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane.

Note: The Indication for Use is the same as the SedLine Sedation Monitor, cleared under K140188, with the exception of identification of anesthetic agents used for validation and the intended use population age.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red checkmark symbol to the left of the word. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".

5.5 Technological Characteristics

5.5.1 Principle of Operation

The Patient State Index (PSI) formula was constructed based upon multivariate combinations of quantitative electroencephalogram (OEEG) variables found to be sensitive to changes in the level of anesthesia but insensitive to the specific substances producing such changes. The PSI is the result of a complex computation that combines weighted quantitative values reflecting many dimensions of brain electrical activity, such as: (1) changes in power in various EEG frequency bands, (2) changes in symmetry and synchronization between critical brain regions, and (3) the inhibition of regions of the frontal cortex.

The PSI is computed from continuously monitored changes in the OEEG during surgery, using statistical analysis to estimate the likelihood that the patient is anesthetized. The SedLine performs these computations automatically on the continuously recorded EEG after automatic removal of data contaminated with artifact from physiological and environmental signals. The computed PSI is periodically updated, displayed in numeric form, and presented in a color-coded trend graphic for monitoring the effect of certain anesthetics on the state of the brain.

5.5.2 Mechanism of Action for Achieving the Intended Effect

The SedLine EEG Sensor is noninvasively applied to the patient on one end. The other end of the SedLine EEG Sensor connects to the SedLine Module. In turn, the SedLine Module connects to a Host/Backboard device. The SedLine EEG Sensor collects patient EEG signals which are processed by the SedLine Module and displayed on the Host/Backboard device.

5.5.3 Summary of Technological Characteristics of Subject Device Compared to Predicate Device

The subject device, SedLine® Sedation Monitor, and the predicate device. Masimo Root Monitoring System (K140188), have the following key similarities:

  • Both devices have the same parameters to measure patient sedation state;
  • Both devices display their calculated values on a Host/Backboard device . (Masimo Root Monitoring System);
  • Both devices have the same indications for use;
  • Both devices have the same intended use; ●
  • Both devices have the same measurement site; ●

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Masimo Corporation. On the left is a red circle with a white line going through it. To the right of the logo is the word "Masimo" in black, and to the right of that is the text "MASIMO CORPORATION 52 Discovery Irvine, CA 92618" in gray.

  • Both devices include EEG sensors and patient cables.
    The subject device and the predicate device. Root Monitoring System (K140188). differ only in the SedLine software. The differences are:

  • . An optional additional DSA display is now available that uses a multi-taper window EEG to display the spectral waveform;

  • Modified Patient State Index (PSI) algorithm.

5.5.4 Non-clinical Testing

See below for the non-clinical testing that was completed.

  • Electrical safety testing per IEC 60601-1
  • EMC testing per IEC 60601-1-2 ●
  • Alarm testing per IEC60601-1-8 ●
  • Usability testing per FDA Human Factors and Usability Draft Guidance
  • Software verification per FDA Software Guidance ●
  • Mechanical Testing per EN 60601-2-26 ●
  • Environmental testing per EN 60601-2-26

The results demonstrate that all requirements and performance specifications were satisfied, and that the subject device is substantially equivalent to the predicate device.

The multi-taper DSA evaluation was conducted as part of the "Software verification per FDA Software Guidance." The tests validated DSA by comparing multi-taper DSA against the predicate's Hanning DSA, both computed for a known input signal. The following functionality were tested for multi-taper DSA:

  • Dynamic range of the power spectrum,
  • Frequency range,
  • Spectral Edge Frequency, and
  • High-contrast feature of the multi-taper DSA against the predicate's Hanning DSA.

5.5.5 Clinical Testing

Validation of the subject algorithm was performed by retrospective analysis of the clinical data in 100 surgical patients (age: 18-77 years, male/female: 26/74). The

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red symbol on the left, followed by the word "Masimo" in black. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray. The logo is clean and modern, with a focus on the company name and location.

subject PSi algorithm was compared to the predicate PSi algorithm. The clinical data was collected at the operating room from three different hospitals. Clinical data used for the analysis includes continuous EEG, anesthetic drug dose information, and other physiological vital signs such as mean arterial blood pressure and heart rate.

5.5.6 Conclusion

The clinical and non-clinical testing provided in this 510(k) submission demonstrates that the subject device, SedLine® Sedation Monitor, is substantially equivalent to its predicate.

§ 882.1400 Electroencephalograph.

(a)
Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.(b)
Classification. Class II (performance standards).