Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K162603
    Manufacturer
    Date Cleared
    2017-05-26

    (249 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K140188, K151644, K153225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The noninvasive Masimo O3 Regional Oximeter System and accessories are intended for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥ 40 kg and on pediatrics > 5 kg and

    Device Description

    The Masimo Regional Oximetry System (O3 System) includes the O3 Sensors and the O3 Module. The O3 System measures hemoglobin under the sensor, allowing clinicians to continuously and accurately determine the absolute and trend measurements of regional blood oxygenation saturation in the tissue (rSO2). The O3 Sensors includes optical components that collect physiological signals. The O3 Module includes Masimo technology for processing those signals which resulted in regional oximetry (rSO2) measurements. In turn, these measurements are displayed on the Host/Backboard device.

    The O3 Sensor is a single-patient use adhesive sensor and is supplied non-sterile. The O3 Sensor, comprising of an emitter and two detectors, is applied to the patient's forehead at one end. The other end of the sensor is connected to a patient cable which in turn is connected to the 03 Module. Up to two 03 Sensors can be connected to each 03 Module and both sensors can be connected to a patient.

    The O3 Module includes Near InfraRed Spectroscopy (NIRS) technology. The O3 Sensor uses multiple wavelengths in the range of near infrared wavelengths to measure light absorption in the tissue. The O3 Sensors and O3 Module make up the O3 System for the monitoring of absolute regional hemoglobin oxygen saturation of blood (rSO2) under the sensors. The O3 System does not have its own power. The O3 Module is powered by connecting to a Host/Backboard Device such as the Root Monitoring System (Root in turn is powered by either AC power or internal rechargeable batteries.

    The O3 System provides the following key measurements:

    • Regional Oxygenation (rSO2): Regional tissue oxygenation level in the deep tissue local to the sensor site, including cerebral tissue
    • Delta Baseline (Abase): Relative difference in rSO2 with respect to baseline rSO2
    • Area Under the Limit (AUL index): Index that quantifies the duration (amount of time the patient stays below rSO2 low alarm limit) and depth (refers to the gap between the patient's rSO2 level and the rSO2 low alarm limit) of patient's stay below the user-defined rSO2 low alarm limit (LAL)
    • Delta SpO2 (ΔSpO2): The difference between SpO2 and rSO2. The source of SpO2 is from peripheral SpO2 measurement (using pulse oximeter).
    AI/ML Overview

    The provided document describes the Masimo O3 Regional Oximeter System, specifically focusing on the O3 Pediatric Sensor, and its equivalence to predicate devices (K160526 and K082327). The document outlines non-clinical and clinical testing performed to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for accuracy are presented as ARMS (Accuracy Root Mean Square) values.

    FeatureAcceptance Criteria (Specification)Reported Device Performance (O3 Pediatric Sensor)
    rSO2 Measurement Accuracy
    Adults ≥ 40kg (using O3 Large Sensor - Predicate)Absolute ARMS, 4% for SavO2 of 45%-85%N/A (this is predicate performance)
    Trending ARMS, 3% for SavO2 of 45%-85%N/A (this is predicate performance)
    Pediatrics ≥ 5 kg and
    Ask a Question

    Ask a specific question about this device

    K Number
    K160526
    Manufacturer
    Date Cleared
    2016-06-09

    (105 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K140188

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The noninvasive Masimo O3 Regional Oximeter System and accessories are intended for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (tSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥ 40 kg in healthcare environments.

    Device Description

    The Masimo Regional Oximetry System (O3 System) includes the O3 Sensors and the O3 Module. The O3 System measures hemoglobin under the sensor, allowing clinicians to continuously and accurately determine the absolute and trend measurements of regional blood oxygenation saturation in the tissue (rSO2). The O3 Sensors includes optical components that collect physiological signals. The O3 Module includes Masimo technology for processing those signals which resulted in regional oximetry (rSO2) measurements. In turn, these measurements are displayed on the Host/Backboard device.

    The O3 Sensor is a single-patient use adhesive sensor and is supplied non-sterile. The O3 Sensor, comprising of an emitter and two detectors, is applied to the patient's forehead at one end. The other end of the sensor is connected to a patient cable which in turn connected to the 03 Module. Up to two 03 Sensors can be connected to each O3 Module and both sensors can be connected to a patient.

    The O3 Module includes Near InfraRed Spectroscopy (NIRS) technology. The O3 Sensor uses multiple wavelengths in the range of near infrared wavelengths to measure light absorption in the tissue. The O3 Sensors and O3 Module make up the O3 System for the monitoring of absolute regional hemoglobin oxygen saturation of blood (rSO2) under the sensors. The O3 System does not have its own power. The O3 Module is powered by connecting to a Host/Backboard Device such as the Root Monitoring System (Root). Root in turn is powered by either AC power or internal rechargeable batteries.

    The O3 System provides the following key measurements:

    • Regional Oxygenation (rSO2): Regional tissue oxygenation level in the deep o tissue local to the sensor site, including cerebral tissue
    • o Delta Baseline (Abase): Relative difference in rSO2 with respect to baseline rSO2
    • Area Under the Limit (AUL index): Index that quantifies the duration (amount of ● time the patient stays below rSO2 low alarm limit) and depth (refers to the gap between the patient's rSO2 level and the rSO2 low alarm limit) of patient's stay below the user-defined rSO2 low alarm limit (LAL)
    • o Delta SpO2 (dSpO2): The difference between SpO2 and rSO2. The source of SpO2 is from peripheral SpO2 measurement (using pulse oximeter).
    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Masimo O3 Regional Oximeter System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    FeatureAcceptance Criteria (Specification)Reported Device Performance
    rSO2 (trending) Accuracy≤ 3% (RMS) for 45% to 85% SavO2Meets requirements (specific RMS value not provided, but stated to be within ≤3%)
    rSO2 (absolute) Accuracy≤ 4% (RMS) for 45% to 85% SavO2Meets requirements (specific RMS value not provided, but stated to be within ≤4%)

    Note on "SavO2": The document mentions "SavO2" in the context of the clinical study, stating accuracy against "30% arterial and 70% jugular venous blood oxygen saturations". This implies that SavO2 is a composite measure or target for regional oxygen saturation, often calculated based on arterial and venous contributions. The acceptance criteria for rSO2 accuracy are directly related to this measured range.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: "healthy adult male and female subjects with light to dark skin pigmentation." The specific number of subjects is not provided in the document.
    • Data Provenance: The study was a prospective clinical study. The country of origin is not explicitly stated, but given the FDA submission, it can be inferred to be from a region where clinical trials meet FDA standards, likely the USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: This information is not applicable as the ground truth was established by direct physiological measurements, not expert review of images or data.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. The ground truth was established by direct physiological measurements using a laboratory CO-Oximeter.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is a measurement device, not an imaging interpretation aid for human readers.

    6. Standalone Performance Study

    • Standalone Performance: Yes, the clinical study explicitly evaluates the device's accuracy ("trending and absolute rSO2 accuracies were determined") against a laboratory standard, indicating a standalone performance study without human intervention in the primary measurement.

    7. Type of Ground Truth Used

    • Ground Truth Type: Physiological measurements: "30% arterial and 70% jugular venous blood oxygen saturations, measured by a laboratory CO-Oximeter." This is a direct, objective physiological measurement.

    8. Sample Size for the Training Set

    • Sample Size: The document does not specify a separate training set or its sample size. For simple measurement devices like an oximeter, "training" often refers to the device's inherent calibration and algorithm development, which might not involve a distinct, large-scale "training set" in the same way machine learning models for image analysis do. The clinical study described appears to be the primary validation set.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training: As no distinct training set is described, this information is not provided in the document. It's likely that the device's algorithms were developed and refined through engineering and calibration processes against established physiological principles and potentially internal studies, rather than a publicly reported "training set" with ground truth established through a formal process analogous to a clinical trial.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1