K Number
K172468
Manufacturer
Date Cleared
2017-12-06

(113 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 091 Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasulature.

Device Description

The InNeuroCo, Inc. 091 Long Sheath is a variable stiffness catheter that has a catheter shaft reinforced with a stainless steel double coil. It has a radiopaque Platinum/lridium marker band on the distal end. The distal 10 cm of the 091 Long Sheath Catheter has a hydrophilic coating. The catheter has a nominal outer diameter of 0.109 inches and a nominal inner diameter of 0.091 inches. It is available in three working lengths: 70 cm, and 90 cm. The 091 Long Sheath has a PTFE-lined lumen. The 091 Long Sheath is inserted at a vascular access point to provide access to the target site and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst Hemostasis Valve and a Vessel Dilator. The 091 Long Sheath is supplied sterile, non-pyrogenic, and intended for single use only.

AI/ML Overview

The provided text is a 510(k) summary for the InNeuroCo 091 Long Sheath peripheral, coronary, and neuro vasculature catheter. It outlines the device's characteristics and its comparison to a predicate device, along with the results of performance testing to demonstrate substantial equivalence.

Based on the provided document, here's a description of the acceptance criteria and the study proving the device meets these criteria:

This document describes the validation of a medical device, the InNeuroCo 091 Long Sheath, for regulatory approval. It is not an AI/ML device, and therefore several of the requested sections (e.g., number of experts, adjudication methods, MRMC study, standalone performance, training set details) are not applicable as they relate to the validation of AI/ML algorithms, not traditional medical devices like catheters.


Acceptance Criteria and Device Performance Study for the InNeuroCo 091 Long Sheath

The InNeuroCo 091 Long Sheath is a percutaneous catheter intended for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. The study conducted was a series of non-clinical (bench and animal) tests to demonstrate substantial equivalence to a legally marketed predicate device (Stryker AXS Infinity LS, K152876).

1. Table of Acceptance Criteria and Reported Device Performance

TestTest Method SummaryAcceptance CriteriaReported Device Performance (Conclusions)
Biocompatibility-Material Mediated Pyrogen TestingCompleted per ISO 10993-11The test article extracts must not cause a febrile reaction greater than 0.5°C in any individual subject.Test results for the InNeuroCo Super Distal Access (K161262) and the Zenith Catheter (K171672) were leveraged as materials and manufacturing processes are equivalent. Met acceptance criteria.
Biocompatibility-Cytotoxicity MEM ElutionCompleted per ISO 10993-5The cultures treated with the test article must not have a reactivity grade greater than 2.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
Biocompatibility-Hemolysis ASTM Method, extract human bloodCompleted per ISO 10993-4The hemolytic index above the negative control article must be less than 5%.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
Biocompatibility-Hemolysis, ASTM method, direct contact (human blood)Completed per ISO 10993-4The hemolytic index above the negative control article must be less than 5%.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
Biocompatibility-Unactivated Partial Thromboplastin TimeCompleted per ISO 10993-4There must be no statistical decrease between the UPTT of plasma exposed to the test article and to the negative or untreated control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
Biocompatibility-Complement ActivationCompleted per ISO 10993-4There must be no statistical increase between either the C3a or SC5b-9 concentrations in plasma exposed to the test article as compared to the negative and untreated controls.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
Biocompatibility-Dog ThrombogenicityCompleted per ISO 10993-4The test articles must receive a thrombus formation score less than or equal to that of the control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
Biocompatibility-Maximization SensitizationCompleted per ISO 10993-10The test article must elicit a positive response in less than 10% of the test animals.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
Biocompatibility-Intracutaneous Toxicity/ReactivityCompleted per ISO 10993-10The test article extracts must not induce a significantly greater biological reaction than the control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
Biocompatibility-Acute Systemic Toxicity TestCompleted per ISO 10993-11The test article extracts must not induce a significantly greater biological reaction than the control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
Animal Testing-Angiographic EvaluationProduct evaluated within a porcine model for clinically significant injury.No units tested can cause vessel injury.Test article met the acceptance criteria for Animal Testing.
Bench Testing - Chemical CompatibilityCatheter exposed to chemicals readily available in a clinical setting.Chemicals have no negative effect on the catheter or accessories.Leveraged results from InNeuroCo Super Distal Access (K161262). Met acceptance criteria.
Bench Testing - Visual InspectionCompleted per ISO 10555-1Catheter shall appear free from damage, including a rounded tip and smooth transition points.091 Long Sheath test samples met the acceptance criteria.
Bench Testing - DimensionsCompleted per ISO 10555-1Test samples should be within existing dimensional specifications.091 Long Sheath test samples met the acceptance criteria. Leveraged results for Hemostasis Valve from InNeuroCo Super Distal Access (K161262), which also met the criteria.
Bench Testing - Hub compatibilityCompleted per ISO 594-1 and ISO 594-2Hub shall meet existing Luer specifications.Leveraged results for the 091 Long Sheath and Dilator from InNeuroCo Intermediate Catheter (K152202) and for the Hemostasis Valve from Super Distal Access (K161262). All tested samples met the acceptance criteria.
Bench Testing - Leak - AirCompleted per ISO 10555-1Test samples should be within existing Air – Leak specifications.091 Long Sheath test samples met the acceptance criteria.
Bench Testing - Leak - LiquidCompleted per ISO 10555-1Test samples should be within existing Leak - Liquid specifications.091 Long Sheath test samples met the acceptance criteria.
Bench Testing - ParticulatesCompleted per USP 788Test samples should be within existing Particulate specifications.091 Long Sheath test samples met the acceptance criteria.
Bench Testing - Simulated UseLS underwent simulated use testing by a physician in a benchtop model.Test samples must meet predetermined user needs.091 Long Sheath test samples met the acceptance criteria.
Bench Testing - Catheter BurstCompleted per ISO 10555-1Test sample burst pressures must meet or exceed existing minimum burst pressure specification.091 Long Sheath test samples met the acceptance criteria. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bench Testing - TensileCompleted per ISO 10555-1Test sample ultimate tensile strength must meet or exceed existing tensile strength specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bench Testing - CorrosionCompleted per ISO 10555-1Test samples shall exhibit no evidence of corrosion.Leveraged results from InNeuroCo Intermediate Catheter (K152202). All samples met the acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bench Testing - Packaging - Dye LeakCompleted per ASTM F1929-12Test sample shall not exhibit any visual leaks or channels.091 Long Sheath test samples met the acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bench Testing - Packaging - PeelCompleted per ASTM F88-09Test sample tensile strength must meet or exceed existing tensile strength specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bench Testing - Kink ResistanceSamples are subjected to different diameters until kink is observed.Test sample kink resistance must meet or exceed existing Kink Resistance specifications.091 Long Sheath test samples met the acceptance criteria. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bench Testing - TorqueConditioned samples are torqued to failure.Test sample torque results must meet or exceed existing torque specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bench Testing - PTFE Liner inspectionLS was challenged to demonstrate liner adherence.Test sample liner adhesion must meet or exceed existing PTFE Liner inspection specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bench Testing - Hydrophilic Coating IntegrityConditioned samples were repeatedly exposed to friction to demonstrate that the hydrophilic coating is not affected.Test sample results must meet or exceed existing Hydrophilic Coating Integrity specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bench Testing - Labeling LegibilityLabel is legible after printing.Test samples shall demonstrate text legibility.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bench Testing - BarcodeBarcode is readable with a standard barcode reader.Test samples shall demonstrate readily readable barcodes.091 Long Sheath test samples met the acceptance criteria. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
SterilizationISO 11135 and AAMI TIR 28Sterilization load shall pose an equal or lesser challenge to sterilize than existing sterile product loads.091 Long Sheath sterilization load met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Bacterial EndotoxinAAMIST72Test samples must meet the bacterial endotoxin acceptance criteria.The 091 Long Sheath samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Shelf LifeASTM F1980Aged test samples must meet or exceed existing specifications.091 Long Sheath test samples met the acceptance criteria. -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The exact sample sizes for each specific test are not explicitly enumerated in the provided summary table. The statements typically indicate "Test samples" or "All samples," implying that a sufficient number were tested according to the referenced ISO or ASTM standards.
  • Data Provenance: The data is a mix of:
    • Direct testing: For many bench tests (e.g., Visual Inspection, Leaks, Particulates, Simulated Use, Catheter Burst, Tensile, Kink Resistance, Torque, PTFE Liner inspection, Hydrophilic Coating Integrity, Labeling Legibility, Barcode, Sterilization, Bacterial Endotoxin, Shelf Life). These tests were conducted on the 091 Long Sheath.
    • Leveraged data: For all biocompatibility tests, Chemical Compatibility, Hub compatibility, and Corrosion, data was leveraged from previously approved InNeuroCo devices (Super Distal Access K161262, Zenith Catheter K171672, and Intermediate Catheter K152202). This leveraging is justified by stating that the materials and manufacturing processes are equivalent.
    • Animal Testing: Product evaluated within a porcine model for angiographic evaluation.
  • Retrospective/Prospective: All described testing appears to be prospective (new tests conducted for this device and leveraged results from other devices), not retrospective analysis of existing clinical data.
  • Country of Origin: Not specified in the document, but given the FDA 510(k) submission, it would be expected to follow US or internationally recognized standards. The company address is in Florida, USA.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

  • Not Applicable. This is a traditional medical device (catheter) validation, not an AI/ML algorithm. Ground truth for device performance is established through standardized physical, chemical, biological, and animal testing, not human expert interpretation of data.
  • For "Simulated Use" testing, it mentions "by a physician," indicating one or more physicians were involved in that specific assessment, but their number and specific qualifications are not detailed beyond "physician."

4. Adjudication Method for the Test Set

  • Not Applicable. As this is not an AI/ML study, there is no need for adjudication of conflicting expert opinions on a dataset. Test results are objective measurements against defined acceptance criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No. This type of study is specifically designed for evaluating the impact of AI on human reader performance, which is not relevant for the validation of a physical medical device like a catheter.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not Applicable. There is no AI algorithm being validated in this submission. The "device" is a physical catheter.

7. The Type of Ground Truth Used

  • The "ground truth" for this device's performance is established by:
    • Standardized Test Methods: Adherence to established international (ISO, ASTM, USP) and industry standards for material properties, mechanical performance, biocompatibility, and sterility.
    • Physical Measurements: Objective measurements of dimensions, burst pressure, tensile strength, etc.
    • Chemical/Biological Assays: Results from biocompatibility tests (e.g., cytotoxicity, hemolysis, pyrogenicity, sensitization) and bacterial endotoxin tests.
    • Functional Demonstrations: Successful performance in simulated use environments and in an animal model (porcine model for angiographic evaluation, showing no vessel injury).
    • Predetermined Specifications: Comparison against established product specifications and the performance of the predicate device.

8. The Sample Size for the Training Set

  • Not Applicable. This is not an AI/ML device, so there is no "training set."

9. How the Ground Truth for the Training Set was Established

  • Not Applicable. As there is no AI/ML training set, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 6, 2017

InNeuroCo, Inc. Marianne Grunwaldt Director, Quality Assurance and Regulatory Affairs 4635 NW 103rd Avenue Sunrise, Florida 33351

Re: K172468

Trade/Device Name: 091 Long Sheath Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY, DTL, GCC Dated: November 3, 2017 Received: November 6, 2017

Dear Marianne Grunwaldt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Carlos L. Pena -S

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172468

Device Name 091 Long Sheath

Indications for Use (Describe)

The 091 Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasulature.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Submitter's Name and Address

InNeuroCo, Inc. 4635 NW 103rd Avenue Sunrise, Florida 33351 Telephone: 1-954-742-5988 Facsimile: 1-954-742-5989

Contact Information

Marianne Grunwaldt Director, Quality Assurance & Regulatory Affairs InNeuroCo, Inc 4635 NW 103rd Avenue Sunrise, Florida 33351 Telephone: 1-954-703-5271 Facsimile: 1-954-742-5989 E-Mail: Marianne@InNeuroCo.com

Date Prepared

December 4, 2017

Device Trade or Proprietary Name

091 Long Sheath

Device Common or Classification Name:

Catheter, Percutaneous, 21CFR870.1250, Class II Hemostasis Valve, 21CFR870.4290, Class II Vessel Dilator, 21 CFR 870.4200, Class I

{4}------------------------------------------------

Product Code:

DQY (Catheter) DTL (Hemostasis Valve) GCC (Vessel Dilator)

Identification of the Legally Marketed Devices to which Equivalence is Being Claimed

Name of Predicate DeviceName of Manufacturer510(k) Number
AXS Infinity LSStrykerK152876

Device Description

The InNeuroCo, Inc. 091 Long Sheath is a variable stiffness catheter that has a catheter shaft reinforced with a stainless steel double coil. It has a radiopaque Platinum/lridium marker band on the distal end. The distal 10 cm of the 091 Long Sheath Catheter has a hydrophilic coating. The catheter has a nominal outer diameter of 0.109 inches and a nominal inner diameter of 0.091 inches. It is available in three working lengths: 70 cm, and 90 cm. The 091 Long Sheath has a PTFE-lined lumen. The 091 Long Sheath is inserted at a vascular access point to provide access to the target site and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst Hemostasis Valve and a Vessel Dilator. The 091 Long Sheath is supplied sterile, non-pyrogenic, and intended for single use only.

Indications for Use

The 091 Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

{5}------------------------------------------------

Comparison to Predicate Device

StrykerAXS Infinity LSInNeuroCo 091 LongSheath
510(k) NumberK152876K172468
Classification21CFR870.1250, Class II21CFR870.1250, Class II
Product CodeDQYSame
Review PanelCardiovascularSame
Indications For UseThe AXS Infinity LS isindicated for theintroduction ofinterventional devicesinto the peripheral,coronary, and neurovasculature.The 091 Long Sheath isindicated for theintroduction ofinterventional devicesinto the peripheral,coronary, and neurovasculature.
ComponentsSuppliedSheath, Vessel Dilator,Hemostasis ValveSame
Catheter ShaftMaterialPolyether Block Amide,(PEBAX) andChronoflexSame
Inner LinerPTFESame
Hub MaterialPolycarbonateSame
Strain ReliefPolyolefinSame
Catheter ShaftReinforcementStainless Steel BraidStainless Steel Double-Coil
Lubricious CoatingHarland HydrophilicCoatingSame
Radiopaque MarkerBandPlatinum/ IridiumSame
StrykerAXS Infinity LSInNeuroCo 091 Long Sheath
PackagingTyvek/Nylon Pouch, polyethylene support tube, packaging card, SBS cartonSame
SterilizationEthylene OxideSame
PyrogenicityNonpyrogenicSame
Working Lengths70, 80, 90 cmSame
Internal Diameter0.088 inches0.091 inches
Outer Diameter0.109 inchesSame
Vessel DilatorAid in sheath introduction during procedureSame
Hemostasis ValveMinimizes blood lossSame
Luer Tapered HubYesSame

{6}------------------------------------------------

091 Long Sheath Performance Testing

TestTest MethodSummaryAcceptance CriteriaConclusionsPredicateDevice Testing(Yes/Unknown)
Biocompatibility-MaterialMediated PyrogenTestingcompleted perISO 10993-11The test articleextracts must notcause a febrilereaction greaterthan 0.5°C in anyindividual subject.Test results for theInNeuroCo SuperDistal Access(K161262) and theZenith Catheter(K171672) wereleveraged for the091 Long Sheathas the materialsand manufacturingYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicateDevice Testing(Yes/Unknown)
Biocompatibility-Cytotoxicity MEM ElutionTestingcompleted perISO 10993-5The culturestreated with the testarticle must nothave a reactivitygrade greater than2.Test results for theInNeuroCo SuperDistal Access(K161262) and theZenith Catheter(K171672) wereleveraged for the091 Long Sheathas the materialsand manufacturingprocesses areequivalent. Testarticles met theacceptancecriteria forCytotoxicity MEMElution.Yes
Biocompatibility-Hemolysis ASTMMethod, extract humanbloodTestingcompleted perISO 10993-4The hemolyticindex above thenegative controlarticle must be lessthan 5%.Test results for theInNeuroCo SuperDistal Access(K161262) and theZenith Catheter(K171672) wereleveraged for the091 Long Sheathas the materialsand manufacturingprocesses areequivalent. Testarticles met theacceptancecriteria forHemolysis ASTMYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicateDevice Testing(Yes/Unknown)
Biocompatibility-Hemolysis, ASTMmethod, direct contact(human blood)Testingcompleted perISO 10993-4The hemolyticindex above thenegative controlarticle must be lessthan 5%.Method, extracthuman blood.Test results for theInNeuroCo SuperDistal Access(K161262) and theZenith Catheter(K171672) wereleveraged for the091 Long Sheathas the materialsand manufacturingprocesses areequivalent. Testarticles met theacceptancecriteria forHemolysis, ASTMmethod, directcontact (humanblood).Yes
Biocompatibility-Unactivated PartialThromboplastin TimeTestingcompleted perISO 10993-4There must be nostatistical decreasebetween the UPTTof plasma exposedto the test articleand to the negativeor untreatedcontrol.Test results for theInNeuroCo SuperDistal Access(K161262) and theZenith Catheter(K171672) wereleveraged for the091 Long Sheathas the materialsand manufacturingprocesses areequivalent. Testarticle met theacceptancecriteria forUnactivatedPartialThromboplastinTime.Yes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicateDevice Testing(Yes/Unknown)
Biocompatibility-Complement ActivationTestingcompleted perISO 10993-4There must be nostatistical increasebetween either theC3a or SC5b-9concentrations inplasma exposed tothe test article ascompared to thenegative anduntreated controls.Test results for theInNeuroCo SuperDistal Access(K161262) and theZenith Catheter(K171672) wereleveraged for the091 Long Sheathas the materialsand manufacturingprocesses areequivalent. Testarticle met theacceptancecriteria forComplementActivation.Yes
Biocompatibility-DogThrombogenicityTestingcompleted perISO 10993-4The test articlesmust receive athrombus formationscore less than orequal to that of thecontrol.Test results for theInNeuroCo SuperDistal Access(K161262) and theZenith Catheter(K171672) wereleveraged for the091 Long Sheathas the materialsand manufacturingprocesses areequivalent. Testarticle met theacceptancecriteria for DogThromboresistance.Yes
Biocompatibility-MaximizationSensitizationTestingcompleted perISO 10993-10The test articlemust elicit apositive responsein less than 10% ofthe test animals.Test results for theInNeuroCo SuperDistal Access(K161262) and theZenith Catheter(K171672) wereYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicateDevice Testing(Yes/Unknown)
leveraged for the091 Long Sheathas the materialsand manufacturingprocesses areequivalent. Testarticle met theacceptancecriteria forMaximizationSensitization.
Biocompatibility-IntracutaneousToxicity/ReactivityTestingcompleted perISO 10993-10The test articleextracts must notinduce asignificantly greaterbiological reactionthan the control.Test results for theInNeuroCo SuperDistal Access(K161262) and theZenith Catheter(K171672) wereleveraged for the091 Long Sheathas the materialsand manufacturingprocesses areequivalent. Testarticle met theacceptancecriteria forIntracutaneousToxicity/Reactivity.Yes
Biocompatibility-AcuteSystemic Toxicity TestTestingcompleted perISO 10993-11The test articleextracts must notinduce asignificantly greaterbiological reactionthan the control.Test results for theInNeuroCo SuperDistal Access(K161262) and theZenith Catheter(K171672) wereleveraged for the091 Long Sheathas the materialsand manufacturingprocesses areequivalent. TestYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicateDevice Testing(Yes/Unknown)
Animal Testing-Angiographic EvaluationProductevaluated withina porcine modelfor clinicallysignificantinjury.No units tested cancause vessel injury.Test article metthe acceptancecriteria for AnimalTesting-Yes
Bench Testing -Chemical CompatibilityCatheterexposed tochemicalsreadily availablein a clinicalsetting.Chemicals have nonegative effect onthe catheter oraccessories.Test results for theInNeuroCo SuperDistal Access(K161262) wereleveraged for the091 Long Sheathas the materialsand manufacturingprocesses areequivalent. Testarticles met theacceptancecriteria forChemicalCompatibilityYes
Bench Testing - VisualInspectionTestingcompleted perISO 10555-1Catheter shallappear free fromdamage, includinga rounded tip andsmooth transitionpoints.091 Long Sheathtest samples metthe acceptancecriteria for VisualInspection.Yes
Bench Testing -DimensionsTestingcompleted perISO 10555-1Test samplesshould be withinexistingdimensionalspecifications.091 Long Sheathtest samples metthe acceptancecriterial fordimensions. Testresults for theYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicateDevice Testing(Yes/Unknown)
InNeuroCo SuperDistal Access(K161262) wereleveraged for theHemostasis Valve,which met thedimensionalacceptancecriteria.
Bench Testing - HubcompatibilityTestingcompleted perISO 594-1 andISO 594-2Hub shall meetexisting Luerspecifications.Test results for theInNeuroCoIntermediateCatheter(K152202) wereleveraged for the091 Long Sheathand the Dilator asthe materials andmanufacturingprocesses areequivalent. Thetest results for theSuper DistalAccess (K161262)were leveraged forthe 091 LongSheathHemostasis Valve.All tested samplesmet theacceptancecriteria for Hubcompatibility.Yes
Bench Testing - Leak -AirTestingcompleted perISO 10555-1Test samplesshould be withinexisting Air – Leakspecifications.091 Long Sheathtest samples metthe acceptancecriteria for Leak -Air.Yes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicateDevice Testing(Yes/Unknown)
Bench Testing - Leak -LiquidTestingcompleted perISO 10555-1Test samplesshould be withinexisting Leak -Liquidspecifications.091 Long Sheathtest samples metthe acceptancecriteria for Leak -Liquid.Yes
Bench Testing -ParticulatesTestingcompleted perUSP 788Test samplesshould be withinexisting Particulatespecifications.091 Long Sheathtest samples metthe acceptancecriteria forParticulates.Yes
Bench Testing -Simulated UseLS underwentsimulated usetesting by aphysician in abenchtop modelTest samples mustmeetpredetermined userneeds091 Long Sheathtest samples metthe acceptancecriteria forSimulated Use -Bench.Yes
Bench Testing - CatheterBurstTestingcompleted perISO 10555-1Test sample burstpressures mustmeet or exceedexisting minimumburst pressurespecification.091 Long Sheathtest samples metthe acceptancecriteria forCatheter Burst.Yes
Bench Testing - TensileTestingcompleted perISO 10555-1Test sampleultimate tensilestrength must meetor exceed existingtensile strengthspecifications.091 Long Sheathtest samples metthe acceptancecriteria.Yes
Bench Testing -CorrosionTestingcompleted perISO 10555-1Test samples shallexhibit no evidenceof corrosion.Test results for theInNeuroCoIntermediateCatheter(K152202) wereleveraged for the091 Long Sheathas the materialsand manufacturingprocesses areYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicateDevice Testing(Yes/Unknown)
Bench Testing -Packaging - Dye LeakTestingcompleted perASTM F1929-12Test sample shallnot exhibit anyvisual leaks orchannelsequivalent. Allsamples met theacceptancecriteria.091 Long Sheathtest samples metthe acceptancecriteria forPackaging - DyeLeak.Yes
Bench Testing -Packaging - PeelTestingcompleted perASTM F88-09Test sample tensilestrength must meetor exceed existingtensile strengthspecifications.091 Long Sheathtest samples metthe acceptancecriteria forPackaging - Peel.Yes
Bench Testing - KinkResistanceSamples aresubjected todifferentdiameters untilkink isobserved.Test sample kinkresistance mustmeet or exceedexisting KinkResistancespecifications.091 Long Sheathtest samples metthe acceptancecriteria for KinkResistance.Yes
Bench Testing - TorqueConditionedsamples aretorqued tofailureTest sample torqueresults must meetor exceed existingtorquespecifications.091 Long Sheathtest samples metthe acceptancecriteria for Torque.Yes
Bench Testing - PTFELiner inspectionLS waschallenged todemonstrateliner adherence.Test sample lineradhesion mustmeet or exceedexisting PTFE Linerinspectionspecifications.091 Long Sheathtest samples metthe acceptancecriteria for PTFELiner inspection.Yes
Bench Testing -Hydrophilic CoatingIntegrityConditionedsamples wererepeatedlyexposed tofriction toTest sample resultsmust meet orexceed existingHydrophilic Coating091 Long Sheathtest samples metthe acceptancecriteria forYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicateDevice Testing(Yes/Unknown)
demonstratethat thehydrophiliccoating is notaffected.Integrityspecifications.HydrophilicCoating Integrity.
Bench Testing - LabelingLegibilityLabel is legibleafter printing.Test samples shalldemonstrate textlegibility.091 Long Sheathtest samples metthe acceptancecriteria for labelinglegibility.Yes
Bench Testing - BarcodeBarcode isreadable with astandardbarcode reader.Test samples shalldemonstratereadily readablebarcodes091 Long Sheathtest samples metthe acceptancecriteria for barcodetesting.Yes
SterilizationISO 11135 andAAMI TIR 28Sterilization loadshall pose an equalor lesser challengeto sterilize thanexisting sterileproduct loads.091 Long Sheathsterilization loadmet theacceptancecriteria forsterilization.Yes
Bacterial EndotoxinAAMIST72Test samples mustmeet the bacterialendotoxinacceptance criteriaThe 091 LongSheath samplesmet the acceptancecriteria for bacterialendotoxins.Yes
Shelf LifeASTM F1980Aged test samplesmust meet orexceed existingspecifications091 Long Sheathtest samples metthe acceptancecriteria for shelflife.Yes

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

Summary of Non-Clinical Data

The conclusions drawn from the physical, mechanical, and performance testing of the subject 091 Long Sheath along with the data leveraged from other commercially available products demonstrate that the product is Substantially Equivalent to the legally marketed predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).