K Number
K160743
Device Name
Philips Ingenuity CT
Date Cleared
2016-08-08

(144 days)

Product Code
Regulation Number
892.1750
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Ingenuity CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The Ingenuity CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature, for further information,
Device Description
The Philips Ingenuity CT consists of three system configurations, the Philips Ingenuity CT, the Philips Ingenuity Core and the Philips Ingenuity Core128. These systems are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of Jung nodules that may represent cancer*. The main components (detection system, the reconstruction algorithm, and the x-ray system) that are used in the Philips Ingenuity CT have the same fundamental design characteristics and are based on comparable technologies as the predicate. The main system modules and functionalities are: - 1. Gantry. The Gantry consists of 4 main internal units: - a. Stator a fixed mechanical frame that carries HW and SW. - b. Rotor A rotating circular stiff frame that is mounted in and supported by the stator. - c. X-Ray Tube (XRT) and Generator fixed to the Rotor frame. - d. Data Measurement System (DMS) a detectors array, fixed to the Rotor frame. - 2. Patient Support (Couch) carries the patient in and out through the Gantry bore synchronized with the scan. - 3. Console A two part subsystem containing a Host computer and display that is the primary user interface and the Common Image Reconstruction System (CIRS) - a dedicated powerful image reconstruction computer. In addition to the above components and the software operating them, each system includes a workstation hardware and software for data acquisition, display, manipulation, storage and filming as well as post-processing into views other than the original axial images. Patient supports (positioning aids) are used to position the patient.
More Information

No
The document describes standard CT technology and image reconstruction techniques (FBP and iDose4, which is an iterative reconstruction technique, not necessarily AI/ML). There is no mention of AI, ML, or related terms like deep learning or neural networks in the device description or performance studies.

No
Explanation: The device is described as an X-ray Computed Tomography system intended for diagnostic imaging and screening, not for therapeutic intervention.

Yes

The "Intended Use / Indications for Use" section explicitly states, "These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*." This directly indicates its use for diagnosis.

No

The device description explicitly lists multiple hardware components including a Gantry (with stator, rotor, X-Ray Tube, Generator, and Data Measurement System), Patient Support (Couch), and Console (with Host computer and Common Image Reconstruction System). It also mentions a workstation hardware.

Based on the provided text, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: The Ingenuity CT is an X-ray system that produces images of the body using computer reconstruction of X-ray transmission data. It is used for diagnostic imaging and screening of the body itself, not for analyzing samples taken from the body.
  • Intended Use: The intended use clearly states it's for producing images of the head and body for diagnostic imaging and lung cancer screening. This involves directly imaging the patient's internal structures.

Therefore, the Ingenuity CT falls under the category of medical imaging devices, not In Vitro Diagnostics.

N/A

Intended Use / Indications for Use

The Ingenuity CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The Ingenuity CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.

These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature, for further information.

Product codes

JAK

Device Description

The Philips Ingenuity CT consists of three system configurations, the Philips Ingenuity CT, the Philips Ingenuity Core and the Philips Ingenuity Core128. These systems are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of Jung nodules that may represent cancer*.

The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature, for further information.

The Philips Ingenuity CT is available in three system configurations, Ingenuity CT, Ingenuity Core, Device Description: and Ingenuity Core128

The main components (detection system, the reconstruction algorithm, and the x-ray system) that are used in the Philips Ingenuity CT have the same fundamental design characteristics and are based on comparable technologies as the predicate.

The main system modules and functionalities are:

  1. Gantry. The Gantry consists of 4 main internal units:
    a. Stator a fixed mechanical frame that carries HW and SW.
    b. Rotor A rotating circular stiff frame that is mounted in and supported by the stator.
    c. X-Ray Tube (XRT) and Generator fixed to the Rotor frame.
    d. Data Measurement System (DMS) a detectors array, fixed to the Rotor frame.
  2. Patient Support (Couch) carries the patient in and out through the Gantry bore synchronized with the scan.
  3. Console A two part subsystem containing a Host computer and display that is the primary user interface and the Common Image Reconstruction System (CIRS) - a dedicated powerful image reconstruction computer.

In addition to the above components and the software operating them, each system includes a workstation hardware and software for data acquisition, display, manipulation, storage and filming as well as post-processing into views other than the original axial images. Patient supports (positioning aids) are used to position the patient.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Computed Tomography X-Ray

Anatomical Site

head, whole body, cardiac and vascular

Indicated Patient Age Range

patients of all ages

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

An image evaluation was performed to compare images of the brain, chest, abdomen and pelvis/peripheral orthopedic body areas that were reconstructed using either FBP or an iterative reconstruction technique (iDose4). Images were compared using a 5 point Likert scale by a qualified radiologist to determine if the images were of diagnostic quality. Results indicated that there was no difference in image quality observed between the images reconstructed with iDose4 compared to FBP reconstruction, with images reconstructed using iDose4 scoring higher in most cases. Therefore there was no impact on diagnostic performance and images using iDose4 which is provided on the Phillips Ingenuity CT are of diagnostic quality and equivalent to FBP reconstructed images provided on the predicate system.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K033326

Reference Device(s)

K131773, K033357, K060937, K123599, K133674, K123576, K033677, K042293, K091195

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.

0

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a stylized emblem. The emblem is a graphic representation of a human figure, with three overlapping profiles suggesting community and support. The overall design is simple and professional, reflecting the department's role in public health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 8, 2016

Philips Medical Systems (Cleveland), Inc. % Ms. Susan Quick Regulatory Affairs Specialist 595 Miner Road CLEVELAND OH 44143

Re: K160743

Trade/Device Name: Philips Ingenuity CT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: July 18, 2016 Received: July 20, 2016

Dear Ms. Quick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

1

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

2

Image /page/2/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.

Philips Ingenuity CT 510(k) Submission

Section 4

Indications For Use Statement

3

Indications for Use

510(k) Number (if known)

K160743

Device Name Philips Ingenuity CT

Indications for Use (Describe)

The Ingenuity CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The Ingenuity CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.

These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature, for further information,

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

4

Image /page/4/Picture/0 description: The image contains the word "PHILIPS" in all capital letters. The text is written in a bold, sans-serif font. The color of the text is blue, and the background is white.

Philips Ingenuity CT 510(k) Submission

Section 5

510(k) Summary

Rev. 3

5

Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.

510(k) Summary of Safety and Effectiveness

[As required by 21 CFR 807.92(c)]

| Applicant's Name:
Address: | Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland, OH
44143 USA |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Contact Person:
Title:
Address: | Susan Quick
Regulatory Affairs Specialist
595 Miner Road
Cleveland, OH
44143 USA |
| Telephone number:
Fax number:
E-mail: | +1 440 483-3284
+1 440 483-4918
susan.quick@philips.com |
| 510(k) Summary Date of
Preparation: | 29-February-2016 |
| Device Trade Name: | Philips Ingenuity CT |
| Common or Usual Name: | Computed Tomography X-ray system |
| Classification
Name:
Regulation:
Class:
Product Code:
Panel: | Computed Tomography X-ray system
21 CFR 892.1750
II
JAK
Radiology |
| Primary Predicate device | K033326 - Philips Plus CT Scanner |

6

Indications for Use:

The Philips Ingenuity CT consists of three system configurations, the Philips Ingenuity CT, the Philips Ingenuity Core and the Philips Ingenuity Core128. These systems are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of Jung nodules that may represent cancer*.

The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature, for further information.

The Philips Ingenuity CT is available in three system configurations, Ingenuity CT, Ingenuity Core, Device Description: and Ingenuity Core128

The main components (detection system, the reconstruction algorithm, and the x-ray system) that are used in the Philips Ingenuity CT have the same fundamental design characteristics and are based on comparable technologies as the predicate.

The main system modules and functionalities are:

    1. Gantry. The Gantry consists of 4 main internal units:
    • a. Stator a fixed mechanical frame that carries HW and SW.
    • b. Rotor A rotating circular stiff frame that is mounted in and supported by the stator.
    • c. X-Ray Tube (XRT) and Generator fixed to the Rotor frame.
    • d. Data Measurement System (DMS) a detectors array, fixed to the Rotor frame.
    1. Patient Support (Couch) carries the patient in and out through the Gantry bore synchronized with the scan.
    1. Console A two part subsystem containing a Host computer and display that is the primary user interface and the Common Image Reconstruction System (CIRS) - a dedicated powerful image reconstruction computer.

In addition to the above components and the software operating them, each system includes a workstation hardware and software for data acquisition, display, manipulation, storage and filming as well as post-processing into views other than the original axial images. Patient supports (positioning aids) are used to position the patient.

Device Modifications

This 510(k) addresses the following minor changes (documented internally via a letter to file) that were implemented in the primary predicate device, Philips Plus CT System, cleared by FDA via 510(k) number K033326, (Oct. 29, 2003). Table 5-1 provides a listing of the minor changes that were implemented since clearance of the predicate device, Philips Plus CT System.

7

PHILIPS

Table 5-1
Table of minor changes that were implemented to the primary predicate device, Philips Plus CT
System
Design featureDescription
Dell 670 host to Dell 690 hostEnd of life components, no change in functionality
New CIRS computers
Generator, kV Settings80KW and four kV settings (adding 100kV): the functionality is the same as on
the 60KW generator and the addition of the 100 kV provides another setting
that may be useful in standard scanning.
Generator TubeMRC Ice Tube (880) Identical tube technology, the two tubes share the same
anode, cathode, bearing, anode drive, and electrical interfaces, but differ in
housing.
Detector40 to 64 slices - The move to 64 slices was the addition of physical detectors
increasing the total number of slices to 64.
64 to 128 slices - With a reconstruction process and reduction in ("windmill")
artifacts, 128 slices can be achieved from the 64 detectors.
The NanoPanels of the predicate and the Ingenuity Core are the same, the
NanoPanel Elite is a revision of the NanoPanel with slight performance
improvements but still has the same product specifications.
Gantry Covers and
ControlsGantry Covers & Controls, CT Control Box: new cover design for gantry
control/display panels, simplified CT control box for cost reduction.
Load and Unload Foot
PedalsLoad and Unload foot pedals allow the operator to move the patient couch to
the load or unload position using a foot pedal thus improving patient handling
efficiency by freeing the operator's hands to prepare, restrain, or release the
patient. The functionality was available in the predicate device through the
Gantry panel buttons.
DMS CoverRedesigned DMS cover for improved airflow and ventilation
Interventional Couch
ControlPhilips' interventional couch control, Panning Handle & Joystick, improves
operational efficiency during CT-guided interventional procedures through
tableside control of longitudinal movements for patient positioning. The
functionality was available in the predicate device through the Gantry panel
buttons.
IQ Improvement - Head
Centering ApplicationThe Head Centering Application helps to improve the brain image quality by
giving the user the option to change the bed vertical position (height) after
planning on Surview without forcing the user to re-plan the scan.
FDOM (Dose
Modulation)Dose Modulation function has been cleared on the Brilliance iCT submission
K131773. This function modulates the tube current based on patient body
symmetry change. FDOM incorporates the features of Z-DOM and D-DOM and
does not introduce new hazards, but only supports a mode which combines
angular and longitudinal modulation.
Low Dose Ring Artifact
ReductionLow Dose Ring Artifact Reduction is not a software feature or a software
algorithm. This occurs when scanning at a low dose and/or scanning large
attenuation objects. This was improved by making a change to the data
collection process in the Detector Module System (DMS) which was achieved
with the NanoPanel Elite detector. There is no new technology being
introduced.
Pulmonary GatingIs similar to the pulmonary gating of the AcQSim Multislice CT,
K033357. Retrospective gating (tagging) was added to enhance the ease of use
without significantly changing the device's safety and effectiveness.
CCT (Axial Step &
Shoot)The Step&Shoot (APEX) option is identical to the Axial Cardiac feature cleared
in "Brilliance Volume" CT system K060937.
Table 5-1
Table of minor changes that were implemented to the primary predicate device, Philips Plus CT System
Design featureDescription
SyncRightSyncRight has been cleared on the Ingenuity Digital PET/CT submission
K123599. SyncRight is used in conjunction with the SAS to improve clinical
workflow of contrast enhanced scanning. The SyncRight feature is supported by
the communication between the injector and the CT scanner, to improve clinical
workflow of contrast enhanced scanning.
Dose ManagementThe iDose4 feature has been cleared most recently on the Brilliance iCT
submission K131773. iDose4 is an iterative reconstruction technique that
improves image quality through artifact prevention and increased spatial
resolution at low dose. It is comparable to the present Adaptive Filtering
techniques and edge enhancing features used on the Brilliance 64.
iPatientiPatient has been most recently cleared on the IQon Spectral CT submission
K133674. iPatient was originally called "results driven scanning", this software
platform allows the operator to plan the required reconstruction results prior to
the scan, without the need to go through the 2D images and re-recons. It
enables the system to provide scan acquisition coverage based on the planned
results. It does not affect image reconstruction or clinical algorithms, but only
improves and simplifies the clinical workflow.
Reconstruction During
ReadyReconstruction During Ready has been cleared on the IQon Spectral CT
submission K133674. This feature enables image construction of previous
scans when the system is in "ready-for-scan' mode. The feature does not affect
image reconstruction algorithms, but only reconstruction timing.
IMRThe IMR reconstruction feature has been cleared under K123576 for any Philips
CT system with the stated minimum operational parameters. This is a
purchasable option. The IMR (K123576) reconstruction feature is intended as
an alternative to standard reconstruction methods (filtered back projection) for
the reconstruction of CT scanner data to produce diagnostic images. The IMR
reconstruction feature is designed to reduce image noise, increase high-
contrast spatial resolution, and improve low contrast detectability. IMR is
designed to reduce dose required for diagnostic CT imaging. Image quality
improvements and dose reduction depend on the clinical task, patient size,
anatomical location, and clinical practice.
iBrain/head filtersBrain/head reconstruction filters have been cleared on the IQon Spectral CT
submission K133674. They work to improve Low Contrast Detectability by
improving the Contrast to Noise Ratio (CNR) in brain images. The filters
increase the contrast of white matter vs. gray matter resulting in better white
matter/gray matter separation in brain scans.
Orthopedic Metal Artifact
Reduction (O-MAR)O-MAR stands for orthopedic metal artifact reduction. This post processing
capability reduces metal induced artifacts and is directed for large orthopedics
metals that cause photon starvation of the rays that pass through the metal
object. Comparable to the MX 16-slice feature, K091195.
Advanced Brain
PerfusionThis application was cleared under Brain Perfusion Option K033677 for use on
the Philips Brilliance Workspace of a CT system. This Philips Brilliance
Workspace is currently being used on the Philips Ingenuity CT.
Advanced Vessel
Analysis (AVA)
Comprehensive Cardiac
Analysis (CCA)These features were cleared under Brilliance CT, Private Practice CV
Configuration CT Scanner K042293. The Philips Ingenuity CT and the Brilliance
CT, Private Practice have the same CT functionality and both are using the
same predicate device, the Philips Plus CT K033326. This feature functions the
same on the Philips Ingenuity CT as it does on the Brilliance CT, Private
Practice.

8

PHILIPS

9

Image /page/9/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.

Substantial Primary Predicate Device: Philips Plus CT System Equivalence: Manufacturer: Philips Healthcare Predicate Device k#: K033326

The design, intended use and technology provided with the proposed Philips Ingenuity CT is identical to the predicate device Philips Plus CT System.

Substantial Equivalence Comparison Table 5-2
Primary
Predicate Device
Philips Plus CT
Scanner
(K033326)Proposed Philips
Ingenuity CT
(Ingenuity CT,
Ingenuity Core,
Ingenuity Core128)Comments
Design/Fundamental Scientific Technology
ApplicationHead/BodyHead/BodyIdentical
Scan RegimeContinuous RotationContinuous RotationIdentical
No. of Slices40Ingenuity CT
Configurations:
Ingenuity CT - 128
Ingenuity Core - 64
Ingenuity Core128 - 128The move to 64 slices was the
addition of physical detectors
increasing the total number of
slices to 64.
With a reconstruction process
and reduction in ("windmill")
artifacts, 128 slices can be
achieved from the 64 detectors.
Neither of these changes affect
safety and effectiveness.
Scan ModesSurview
Axial Scan
Helical ScanSurview
Axial Scan
Helical ScanIdentical
Minimum Scan Time0.42 sec for 360°
rotation0.42 sec for 360°
rotationIdentical
Image (Spatial
Resolution)24 lp/cm/max.24 lp/cm/max.Identical
Image Noise0.27% at 120 kV,
250 mAs, 10 mm
slice thickness0.27% at 120 kV, 250
mAs, 10 mm slice
thicknessIdentical
Image MatrixUp to 1024 x 1024Up to 1024 x 1024Identical
Display1024 x 12801024 x 1280Identical
Host InfrastructureWindows XPWindows 7Identical, Windows based
computer running iPatient
system software, does not affect
safety and effectiveness.
Focal SpotDynamic Focal SpotDynamic Focal Spot
in X axisIdentical
Power (kW Output)60 kW80 kWGenerator is FDA class I,
510(k) exempt, no change in
functionality, physically the
same. This change does not
affect safety and effectiveness.
Substantial Equivalence Comparison Table 5-2
Primary Predicate
Device
Philips Plus CT
Scanner
(K033326)Proposed Philips
Ingenuity CT
(Ingenuity CT,
Ingenuity Core,
128
Ingenuity Core)Comments
Design/Fundamental Scientific Technology
mA Range30-500mA20-665mAThe impact of increasing the
tube power is an extended
mA range, difference in
range does not affect safety
and effectiveness.
kV Settings80, 120, 14080, 100, 120, 140Provides another setting that
may be useful in standard
scanning, does not affect
safety and effectiveness.
Tube TypeMRC 800MRC Ice Tube (880)Identical tube technology, the
two tubes share the same
anode, cathode, bearing,
anode drive, and electrical
interfaces, but differ in
housing, does not affect
safety and effectiveness.
Scan Field of ViewUp to 500 mmUp to 500 mmIdentical
Detector TypeSingle layer ceramic
scintillator plus a
photodiodeSingle layer ceramic
scintillator plus a
photodiodeIdentical
Gantry Tilt$\pm 30^\circ$$\pm 30^\circ$Identical
Gantry Rotation Speed143 RPM143 RPMIdentical
Bore Size700 mm700 mmIdentical

10

Image /page/10/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters against a white background. The font is sans-serif and the letters are evenly spaced. The word is horizontally oriented and centered in the image.

Summary of Non-Clinical Testing:

This 510(k) premarket notification contains technical documentation, which demonstrates that the proposed Philips Ingenuity CT is substantially equivalent to the predicate device, Philips Plus CT System in terms of safety and effectiveness. Testing was performed on the proposed Philips Ingenuity CT according to the following international and FDA recognized consensus standards and FDA guidance documents:

  • · IEC 60601-1:2006 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2:2007, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
  • IEC 60601-1-3 Ed 2.0:2008 Medical electrical equipment Part 1-3: General requirements for basic safety - Collateral standard: Radiation protection in diagnostic X-ray equipment
  • · IEC 60601-1-6:2010 Medical electrical equipment -- Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • · IEC 60601-2-44:2009 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment
  • IEC 62304:2006 First Edition Medical device software Software life cycle processes
  • ISO 14971 Medical devices Application of risk management to medical devices (Ed. 2.0, 2007)
  • Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 11, 2005, document number 337).

11

Image /page/11/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.

The completed bench test protocols included the above IEC standard requirements in addition to other requirements found in the main system level software requirements of the System Requirements Specification and the Subsystem Requirement Specifications as well as the identified hazard mitigations from the Safety Risk Management Report. The traceability between the requirements, the hazard mitigations and the test protocols are described in the Traceability Matrix. The Traceability Matrix also shows the overall test results per requirement and per hazard mitigation. The results of the functional and non-functional regression tests as well as the user interface verification are provided in the Traceability Matrix.

Some of the bench tests included patient support/gantry positioning repeatability and accuracy. laser alignment accuracy, CT image quality metrics testing including CT number accuracy and uniformity within each slice for body and head. MTF. noise reduction performance of iDose*, slice thickness, slice sensitivity profiles and radiation metrics. Sample phantom images were provided to show the performance of the system in presence of implants, also when using a uniform phantom and a phantom with embedded test objects.

Software Documentation for a Moderate Level of Concern, per the FDA quidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.

  • All verification tests were executed and passed the specified requirements, which validate that the Conclusion: subject device, Philips Ingenuity CT is substantially equivalent to the predicate device, Philips Plus CT Scanner, with regards to system performance.

Summary

of Clinical The completed validation test plan identified the tasks, deliverables, methodology, requirements Testing: and the resources for validation of the intended use and meets customer needs.

The validation testing covered clinical validation, serviceability validation, manufacturing validation and validation by analysis. The clinical validation covered requirements related to clinical workflows and features. The serviceability validation covered requirements related to upgrade, installation, servicing and troubleshooting of the system. The manufacturing validation covered requirements related to operations and manufacturing and the validation by analysis covered all other requirements that affect the end user that were not covered by the above testing.

The validation test plan was executed as planned and acceptance criteria met for each requirement. All defects were managed and closed. All validation tests demonstrate the safety and effectiveness of the Philips Inqenuity CT in its performance as a CT system. The results of this validation testing are available in the Final Validation Report.

An image evaluation was performed to compare images of the brain, chest, abdomen and pelvis/peripheral orthopedic body areas that were reconstructed using either FBP or an iterative reconstruction technique (iDose4). Images were compared using a 5 point Likert scale by a qualified radiologist to determine if the images were of diagnostic quality. Results indicated that there was no difference in image quality observed between the images reconstructed with iDose4 compared to FBP reconstruction, with images reconstructed using iDose4 scoring higher in most cases. Therefore there was no impact on diagnostic performance and images using iDose4 which is provided on the Phillips Ingenuity CT are of diagnostic quality and equivalent to FBP reconstructed images provided on the predicate system.

The proposed Philips Ingenuity CT can be used as defined in its clinical workflow and intended use.

12

Image /page/12/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.

Overall

Based on the conformance to standards, development under Philips Medical System's quality Conclusion: system, the successful verification testing, additional engineering testing, and the clinical evaluation, Philips Medical Systems believes that the Philips Ingenuity CT is substantially equivalent to the predicate device Philips Plus CT Scanner (K033326). The proposed Philips Ingenuity CT is as safe and effective as the currently marketed and predicate device, Philips Plus CT Scanner, without raising any new safety and/or effectiveness concerns.