(357 days)
No
The summary describes a standard CT system with spectral capabilities for material analysis and image reconstruction. There is no mention of AI or ML in the intended use, device description, or performance studies. The focus is on the hardware and spectral data processing.
No
Explanation: The device is explicitly described as a "diagnostic tool for the visualization and analysis of anatomical and pathological structures," not for treating conditions.
Yes
The "Intended Use / Indications for Use" section explicitly states, "This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures."
No
The device description explicitly details multiple hardware components including a gantry, patient couch, operator console, x-ray tube, and detection system, in addition to the software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The Philips IQon Spectral CT is a medical imaging device that uses X-rays to create cross-sectional images of the body. It analyzes the interaction of X-rays with the body's tissues in vivo (within the living body), not in vitro (outside the body).
- Intended Use: The intended use clearly states it's for producing images of the body and providing information about the chemical composition of body materials and/or contrast agents within the body.
- Device Description: The description details a system that interacts directly with the patient (gantry, patient couch, X-ray tube, detector system).
While the device provides information about the chemical composition of body materials, this is done through the analysis of X-ray attenuation within the body, not through testing samples taken from the body. Therefore, it falls under the category of medical imaging devices, not IVDs.
N/A
Intended Use / Indications for Use
The Philips IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment parts, and accessories.
The IQon Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectramay be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.
This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.
Product codes (comma separated list FDA assigned to the subject device)
JAK
Device Description
The Philips IQon Spectral CT is a whole-body computed tomography (CT) X-Ray System featuring a continuously rotating x-ray tube and detectors gantry and multi-slice capability. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body taken at different angles and planes. This device also includes signal analysis and display equipment; patient and equipment supports; components; and accessories. The Philips IQon Spectral CT includes the detector array previously described in K131773 "Modified Brilliance iCT".
The IQon Spectral CT consists of three main components – a scanner system that includes a rotating gantry, a movable patient couch, and an operator console for control and image reconstruction; a Spectral Reconstruction System; and a Spectral CT Viewer. On the gantry, the main active components are the x-ray high voltage (HV) power supply, the x-ray tube, and the detection system.
In addition to the above components and the software operating them, the system includes workstation hardware and software for data acquisition; and image display, manipulation, storage, and filming, as well as post-processing into views other than the original axial images. Patient supports (positioning aids) are used to position the patient.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Computed Tomography X-Ray
Anatomical Site
Body
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Trained healthcare professional
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Summary of Non-Clinical Testing:
The following tests demonstrated that the IQon Spectral CT system continues to conform to IEC 61223-3-5:2004:
CT Number, Uniformity, Noise and Tomographic Section Thickness Measurements
CTDI Dose Measurements
Air Dose Measurements
Spatial Resolution Measurements
Low Contrast Detectability Measurements
Acceptance and Constancy Test
Performance testing demonstrates following spectral capabilities:
Monoenergetic Images - keV and HU stability
Monoenergetic Images – CT linearity at 70 keV
Iodine Quantification and Water-No-Iodine
Iodine Map Imaging
Calcium-No-Iodine images, and Iodine-No-Calcium images
Calcium-No-Uric-Acid images, and Uric-Acid-No-Calcium images
Virtual Non-Contrast (VNC) images
Effective Atomic Number
Beam Harding Artifact Reduction
Summary of Clinical Testing:
Clinical images were collected and analyzed, to ensure that images constructed by the IQon Spectral CT system meet user needs. This evaluation demonstrated that spectral images were useful for the visualization and analysis of anatomical and pathological structures.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Philips Brilliance iCT (K131773)
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing design element below them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Philips Medical Systems (Cleveland), Inc. % Ms. Catherine M. Connell Regulatory Affairs Specialist 595 Miner Road CLEVELAND OH 44143
November 21, 2014
Re: K133674
Trade/Device Name: Philips IOon Spectral CT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: October 23, 2014 Received: October 24, 2014
Dear Ms. Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K133674
Device Name Philips IQon Spectral CT
Indications for Use (Describe)
The Philips IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment parts, and accessories.
The IQon Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectramay be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.
This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.
Type of Use (Select one or both, as applicable) | |
---|---|
☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
Section 5: 510(k) Summary [As required by 21 CFR 807.92(c)]
| Applicant's Name:
Address: | Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland, OH 44143
USA |
|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Telephone number: | +1 440 483 3000 |
| Contact Person:
Title:
Address: | Catherine M. Connell
Regulatory Affairs Specialist
595 Miner Road
Cleveland, OH 44143
USA |
| Telephone number:
Fax number:
E-mail: | +1 440 483 5581
+1 440 483 4918
catherine.connell@philips.com |
| 510(k) Summary
Date of Preparation: | 23-October-2014 |
| Device Trade Name: | Philips IQon Spectral CT |
| Common or Usual Name: | Computed tomography x-ray system |
| Classification
Name:
Regulation:
Class:
Product Code:
Panel: | Computed tomography x-ray system
21 CFR 892.1750
II
JAK
Radiology |
| Predicate devices | Philips Brilliance iCT (K131773) |
4
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters against a white background. The font is sans-serif and the letters are evenly spaced. The overall impression is clean and modern, reflecting the brand's identity. The color contrast makes the word stand out.
The Philips IQon Spectral CT is a whole-body computed Device Description: tomography (CT) X-Ray System featuring a continuously rotating x-ray tube and detectors gantry and multi-slice capability. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body taken at different angles and planes. This device also includes signal analysis and display equipment; patient and equipment supports; components; and accessories. The Philips IQon Spectral CT includes the detector array previously described in K131773 "Modified Brilliance iCT".
The IQon Spectral CT consists of three main components – a scanner system that includes a rotating gantry, a movable patient couch, and an operator console for control and image reconstruction; a Spectral Reconstruction System; and a Spectral CT Viewer. On the gantry, the main active components are the x-ray high voltage (HV) power supply, the x-ray tube, and the detection system.
In addition to the above components and the software operating them, the system includes workstation hardware and software for data acquisition; and image display, manipulation, storage, and filming, as well as post-processing into views other than the original axial images. Patient supports (positioning aids) are used to position the patient.
5
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters against a white background. The font is sans-serif and the letters are evenly spaced. The overall impression is clean and modern, reflecting the brand's identity. The color contrast makes the word stand out.
The Philips IQon Spectral CT is a Computed Tomography X-Ray Intended Use: System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
The IQon Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a lowerenergy detected x-ray spectrum. The two spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.
This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.
6
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill the majority of the frame. The background is plain white, which makes the blue letters stand out.
Substantial Equivalence: The Philips IQon Spectral CT has an intended use and technology that are similar to the predicate device.
As described above, the Philips IQon Spectral CT includes the hardware configuration previously described in K131773 "Modified Brilliance iCT", updated system software, a Spectral Reconstruction System; and a Spectral CT Viewer application.
The IQon Spectral CT system is able to generate conventional (combined) CT images (equivalent to a CT image series from a non-spectral detector CT system) as described in K131773. The scanner software has been updated to a more recent version of software that provides for an easier user workflow (iPatient).
To address the addition of spectral CT imaging capability, a reference device has been identified. Because both devices are able to acquire two CT spectra, each is able to perform various types of spectral analysis, such as:
- Monoenergetic images
- . Materials Basis/Density Pairs, such as
- o 1/ H20
- o I / Ca
- o Ca / Uric Acid
- I Effective Atomic Number
- 트 Material Separation/Differentiation
- I Attenuation Curves
- Density Measurements/Visualization
- Reduction of Beam Hardening
- Reduction of Calcium Blooming
The Summaries of Non-Clinical Testing and Clinical Testing, presented below, provide a brief outline of the testing data that has been presented to demonstrate substantial equivalence.
7
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is plain white, which makes the blue letters stand out.
| Summary of Non-Clinical
Testing: | The following tests demonstrated that the IQon Spectral CT system
continues to conform to IEC 61223-3-5:2004:
CT Number, Uniformity, Noise and Tomographic Section
Thickness Measurements CTDI Dose Measurements Air Dose Measurements Spatial Resolution Measurements Low Contrast Detectability Measurements Acceptance and Constancy Test Performance testing demonstrates following spectral capabilities: Monoenergetic Images - keV and HU stability Monoenergetic Images – CT linearity at 70 keV Iodine Quantification and Water-No-Iodine Iodine Map Imaging Calcium-No-Iodine images, and Iodine-No-Calcium images Calcium-No-Uric-Acid images, and Uric-Acid-No-Calcium
images Virtual Non-Contrast (VNC) images Effective Atomic Number Beam Harding Artifact Reduction |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Clinical
Testing: | Clinical images were collected and analyzed, to ensure that images
constructed by the IQon Spectral CT system meet user needs. This
evaluation demonstrated that spectral images were useful for the
visualization and analysis of anatomical and pathological
structures. |
| Conclusion: | Philips Medical Systems (Cleveland), Inc. believes that, based
on the information provided in this submission, the Philips IQon
Spectral CT is substantially equivalent to its predicate devices
without raising any new safety and/or effectiveness concerns. |