(198 days)
The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology Analyzer for in vitro diagnostic use in clinical laboratories. The BC-5390 Auto Hematology Analyzer provides complete blood count (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV) and leukocyte 5-Part differential (Neut, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%) for whole blood specimens collected in a salt of EDTA [dipotassium (K2) or tripotassium (K3)) obtained from venous or capillary blood collection. The purpose of the BC-5390 Auto Hematology Analyzer is to identify the normal human patient, with normal system-generated parameters, from patients whose results require additional studies.
The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter for In Vitro Diagnostic Use in clinical laboratories. It is only to be used by trained laboratory professionals to identify the normal patient, with all normal system-generated parameters, and to flag or identify patient results that require additional studies.
The BC-5390 Auto Hematology Analyzer system consists of:
- Instrument: Sample Processing Unit (SPU) and Data Managing Unit (DMU)
- . Reagents M-53D DILUENT M-5LEO(I) LYSE M-5LEO(II) LYSE M-53LH LYSE PROBE CLEANSER
- Controls
BC-5D Hematology Control (High, Normal, Low, Pending) Note: Controls for BC-5390 Auto Hematology Analyzer will be submitted in parallel with this 510(k) by R&D Systems as separate 510(k).
- . Calibrator
SC-CAL PLUS Hematology Calibrator (cleared as K955925)
The Analyzer provides analysis results of 21 parameters, 3 histograms and 1 scattergram of human blood. It supports two test panels: CBC and CBC+DIFF.
Here's an analysis of the acceptance criteria and study details for the Mindray BC-5390 Auto Hematology Analyzer, based on the provided text:
Key Takeaways on Device Performance:
The device's performance was primarily evaluated through a method comparison study against a predicate device (Sysmex XE-2100) and various studies adhering to CLSI (Clinical and Laboratory Standards Institute) guidelines. The results generally indicate good correlation and adherence to predefined specifications, supporting the device's substantial equivalence.
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly list "acceptance criteria" in a single, dedicated table with numerical targets, but instead describes performance studies and states that results "met the pre-defined specification" or "were within the specifications." The table below synthesizes the implicit acceptance criteria based on the reported study results, primarily from the method comparison against the predicate device (Sysmex XE-2100).
| Performance Metric Category | Implicit/Derived Acceptance Criteria (based on predicate comparison) | Reported Device Performance (from Method Comparison vs. Predicate) |
|---|---|---|
| Method Comparison vs. Predicate (Correlation & Bias) | r (correlation coefficient) generally close to 1 (e.g., >0.95 for most parameters, some lower may be acceptable depending on parameter and range) and slope close to 1 with intercept close to 0, ensuring bias is within pre-defined limits. | As presented in Table 5 (pages 9-10), most parameters show r > 0.99. Lower r values for Bas# (0.624) and Bas% (0.415), and Mon% (0.930), MCHC (0.825) were observed, but the overall conclusion stated the analyzer met pre-defined specification for difference limits. Specific slope and intercept confidence intervals are provided for each parameter. |
| WBC Morphology Flagging Ability (vs. Manual) | High sensitivity, specificity, and efficiency for identifying abnormal samples. (No explicit numeric targets given, but implied clinical utility) | Sensitivity (TP %): 89.3% Specificity (TN %): 78.3% Efficiency: 80.6% |
| Precision/Repeatability & Reproducibility | SD and CV of different parameters to be within predefined performance acceptance specifications. | Results met the predefined performance acceptance specifications. (No explicit numeric targets provided in this summary) |
| Linearity Range | Data fitting a linear regression line with a coefficient of determination (R²) of >0.95 and parameters recovering within bias limits. | Results indicated that BC-5390 Auto Hematology Analyzer exhibits linearity across the claimed range. |
| Carryover | Carryover level within defined specification (e.g., ≤ 1.0% for WBC, RBC, HGB, HCT, PLT). | Results were within specifications (≤ 1.0%) for WBC, RBC, HGB, HCT and PLT. Device demonstrated minimum carryover level. |
| Interference | No significant interference from tested substances to key parameters. | No significant interference from bilirubin, WBC, and PLT. High elevated concentrations of Triglyceride (TG) and hemoglobin exhibit minor impact to HGB, MCH, MCHC (disclosed in manual). |
| Sample Stability | Parameter results stable for claimed durations at specified temperatures. | Study results met the pre-defined acceptance criteria supporting claims of 36 hours at 2-8°C, 24 hours at 18-26°C for whole blood, and 25 minutes at 18-26°C for predilute samples. |
| Reference Interval Verification (Adult) | Established reference ranges to be supported by the study performed. | Non-parametric method used to calculate lower and upper limits. Results presented in Table 7 (pages 13-14). Recommended that laboratories establish their own reference range. |
| Reference Interval Verification (Pediatric) | Intervals to be verified against published literature. | Study verified the pediatric intervals published in Mayo and IOWA literature. |
| Limit of Blank (LoB), Limits of Detection (LoD), Limit of Quantitation (LoQ) | Values determined according to CLSI EP17-A. | Maximum LoB, LoD, or LoQ value of the three analyzers is reported. (Specific values not provided in this summary) |
2. Sample Size Used for the Test Set and Data Provenance
-
Test Set Sample Size:
- Method Comparison: 1531 whole blood samples (collected in K2EDTA).
- WBC Morphology Flagging: 1514 samples.
- Precision/Repeatability: Not explicitly stated, but "whole blood samples" for repeatability, and "three levels of samples" from commercial control material BC-5D (tested in duplicate, twice daily for 20 days across 3 sites) for reproducibility (total 240 replicates per level across all sites).
- Linearity: Dilutions prepared from commercial analogs (WBC, PLT) or fresh whole blood (RBC/HGB), with 7 subsequent dilutions, each run in triplicate.
- Carryover: High and low samples tested in triplicates.
- Interference: Whole blood samples with added interfering substances.
- Comparison of Whole Blood Mode and Predilute Mode: 124 leftover whole blood samples.
- Comparison of CBC and CBC+DIFF Mode: 103 leftover whole blood samples (Whole blood CBC vs CD) and 75 samples (Predilute CBC vs CD).
- Comparison of Capillary and Venous Blood: 57 paired specimens.
- Comparison of K2EDTA and K3EDTA Anticoagulants: 70 paired fresh whole blood samples.
- Sample Stability: Fresh samples covering normal and medical decision range.
- Adult Reference Interval: 251 donors (121 male, 130 female).
- Pediatric Reference Interval: 161 pediatric samples (45 neonate, 26 infant, 57 child, 33 adolescent).
- LoB, LoD, LoQ: Five blank samples and five low-level samples, each tested 12 times.
-
Data Provenance:
- Country of Origin: Not explicitly stated for all studies. The method comparison was done at "three actual user sites." The comparison of capillary and venous blood, and K2EDTA vs K3EDTA anticoagulants were done at "the U.S. site" or "clinical laboratory in U.S." This implies at least some data originated from the US, but a global origin is not ruled out for all studies.
- Retrospective or Prospective: Not explicitly stated. The description "whole blood samples... were analyzed," "patient's samples covered the normal and most abnormal conditions," "leftover whole blood samples," and collection of "fresh samples" suggests a mix, and for studies like method comparison and flagging, samples from a patient population would likely be collected prospectively or retrospectively from a clinical lab biobank. The reference interval studies explicitly describe collecting samples, indicating these were prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: For the WBC manual differential (used as ground truth for WBC morphology flagging rate), "three manual wedge smears were prepared and stained with Wright-Giemsa stain. A 400 cell WBC differential was performed on two smears per CLSI H20-A2." This implies at least two experts (or a single expert doing two reads, which is less likely for independent confirmation) were involved in the manual differential portion. The document doesn't specify if these were consensus reads or independent.
- Qualifications of Experts: Not explicitly stated for the manual differential readers. It mentions "trained laboratory professionals" in the device description, implying that these professionals would be performing the manual differentials. CLSI H20-A2 typically refers to the standard for manual differential counting, which implies trained medical technologists or clinical laboratory scientists.
4. Adjudication Method for the Test Set
- Method Comparison and Flagging: For the WBC manual differential, it states "A 400 cell WBC differential was performed on two smears per CLSI H20-A2." It does not explicitly mention an adjudication method (like 2+1 or 3+1). It's possible that if two readers performed the differential, a consensus or third reader was used for discordant results, but the document doesn't detail this. It simply states the manual differential was performed "per CLSI H20-A2," which provides guidelines for the procedure.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done.
- This study evaluates the improvement in human reader performance (e.g., accuracy, efficiency) with the assistance of an AI algorithm compared to performing the task without AI assistance. The document describes a standalone device, an automated hematology analyzer, which performs the cell counting and differential analysis itself. The comparison is between the device and a predicate device, and between the device and manual methods, not human readers with and without AI assistance.
6. Standalone (Algorithm Only) Performance
- Yes, a standalone study was done.
- The "Method Comparison" study (pages 9-10) directly compares the BC-5390 Auto Hematology Analyzer's quantitative results for 21 parameters against a legally marketed predicate device, the Sysmex XE-2100 Automated Hematology Analyzer. This is a standalone performance evaluation of the new device against an established one.
- The "WBC Morphology Flagging Ability" (page 10), which compares the BC-5390's flagging rate to manual WBC differential, is also a standalone performance assessment of the device's ability to identify abnormal samples.
- All the other performance characteristic studies (Precision, Linearity, Carryover, Interference, Sample Stability, LoD/LoQ/LoB) evaluate the BC-5390's performance in isolation or against defined standards, without human intervention in the primary measurement process once the sample is loaded.
7. Type of Ground Truth Used
- Predicate Device Measurements: Primarily used for most quantitative parameter comparisons (e.g., WBC, RBC, HGB, etc.), where the Sysmex XE-2100 (K992875) was the reference. This represents an established and FDA-cleared measurement method.
- Expert Manual Differential: Used for the WBC morphology flagging ability study (400-cell WBC differential performed on manual wedge smears per CLSI H20-A2). This is a form of expert consensus/manual reference.
- Dilution Series / Established Standards: For linearity, carryover, and LoD/LoQ/LoB studies, the ground truth was based on prepared dilution series and reference materials.
- Literature Reference Intervals: For pediatric reference interval verification, the ground truth was published pediatric intervals from "Mayo and IOWA literature."
8. Sample Size for the Training Set
- The document does not provide information on the sample size used for the training set. This is typical for a 510(k) summary, which focuses on device validation/verification testing rather than details of the algorithm's development (training).
9. How the Ground Truth for the Training Set Was Established
- As the document does not provide information on the training set, it also does not explain how the ground truth for any potential training set was established.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health and Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 1, 2016
Mindray Bio-medical Electronics Co., LTD c/o Jinjie Hu, Ph.D. Biologics Consulting Group 400 N. Washington St. Suite 100 Alexandria, VA 22314
Re: K160429
Trade/Device Name: BC-5390 Auto Hematology Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated Differential Cell Counter Regulatory Class: Class II Product Code: GKZ Dated: August 30, 2016 Received: August 31, 2016
Dear Dr. Hu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
Page 2 - Jinjie Hu, Ph.D.
CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Kelly Oliner AS
For
Leonthena R. Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
BC-5390 Auto Hematology Analyzer
Indications for Use (Describe)
The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology Analyzer for in vitro diagnostic use in clinical laboratories. The BC-5390 Auto Hematology Analyzer provides complete blood count (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV) and leukocyte 5-Part differential (Neut, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%) for whole blood specimens collected in a salt of EDTA [dipotassium (K2) or tripotassium (K3)) obtained from venous or capillary blood collection. The purpose of the BC-5390 Auto Hematology Analyzer is to identify the normal human patient, with normal system-generated parameters, from patients whose results require additional studies.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
510(k)PremarketNotification
510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the BC-5390 Auto Hematology Analyzer is provided below.
| Device Common Name: | Auto Hematology Analyzer |
|---|---|
| Device Proprietary Name: | BC-5390 Auto Hematology Analyzer |
| Submitter: | Mindray Bio-medical Electronics Co., LTDMindray Building, Keji 12th Road South, High-techIndustrial Park,Nanshan, Shenzhen 518057, P.R. China |
| Contact: | Jinjie Hu Ph.D.400 N. Washington StreetSuite 100Alexandria, VA 22314Tel: 301-814-4985Email: jhu@bcg-usa.com |
| Classification Regulation: | 21 CFR 864.5220, Class II |
| Panel: | Hematology |
| Product Code: | GKZ |
Predicate Device:
SYSMEX XE-2100 Automated Hematology Analyzer, K992875
Indication for Use:
The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology analyzer for in vitro diagnostic use in clinical laboratories. The BC-5390 Auto Hematology Analyzer provides complete blood count (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV) and leukocyte 5-Part differential ((Neu#, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%) for whole blood specimens collected in a salt of EDTA [dipotassium (K2) or tripotassium (K3)] obtained from venous or capillary blood collection. The purpose of the BC-5390 Auto Hematology Analyzer is to identify the normal human patient, with normal system-generated parameters, from patients whose results require additional studies.
Device Description:
{4}------------------------------------------------
The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter for In Vitro Diagnostic Use in clinical laboratories. It is only to be used by trained laboratory professionals to identify the normal patient, with all normal system-generated parameters, and to flag or identify patient results that require additional studies.
The BC-5390 Auto Hematology Analyzer system consists of:
- Instrument: Sample Processing Unit (SPU) and Data Managing Unit (DMU)
- . Reagents M-53D DILUENT M-5LEO(I) LYSE M-5LEO(II) LYSE M-53LH LYSE PROBE CLEANSER
- Controls
BC-5D Hematology Control (High, Normal, Low, Pending) Note: Controls for BC-5390 Auto Hematology Analyzer will be submitted in parallel with this 510(k) by R&D Systems as separate 510(k).
- . Calibrator
SC-CAL PLUS Hematology Calibrator (cleared as K955925)
The Analyzer provides analysis results of 21 parameters, 3 histograms and 1 scattergram of human blood. It supports two test panels: CBC and CBC+DIFF. The abbreviations for all parameters are listed in Table 1, Table 2 and Table 3.
| Analysis Parameter | Abbreviation | CBC | CBC + DIFF |
|---|---|---|---|
| White Blood Cell Count | WBC | √ | √ |
| Neutrophil Count | Neu# | / | √ |
| Lymphocyte Count | Lym# | / | √ |
| Monocyte Count | Mon# | / | √ |
| Eosinophil Count | Eos# | / | √ |
| Basophil Count | Bas# | / | √ |
| Neutrophil Percent | Neu% | / | √ |
| Lymphocyte Percent | Lym% | / | √ |
| Monocyte Percent | Mon% | / | √ |
| Eosinophil Percent | Eos% | / | √ |
| Basophil Percent | Bas% | / | √ |
| Red Blood Cell Count | RBC | √ | √ |
| Hemoglobin Concentration | HGB | √ | √ |
| Hematocrit | HCT | √ | √ |
| Mean Corpuscular Volume | MCV | √ | √ |
| Table 1: Parameters of BC-5390 Auto Hematology Analyzer | |
|---|---|
| --------------------------------------------------------- | -- |
{5}------------------------------------------------
510(k)PremarketNotification
| Mean CorpuscularHemoglobin | MCH | √ | √ |
|---|---|---|---|
| Mean CorpuscularHemoglobin Concentration | MCHC | √ | √ |
| Red Blood Cell DistributionWidth (Coefficient ofVariation) | RDW-CV | √ | √ |
| Red Blood Cell DistributionWidth (Standard Deviation) | RDW-SD | √ | √ |
| Platelet Count | PLT | √ | √ |
| Mean Platelet Volume | MPV | √ | √ |
Table 2: Histograms of BC-5390 Auto Hematology Analyzer
| Name | Abbreviation | CBC | CBC + DIFF |
|---|---|---|---|
| White Blood Cell/ Basophils | WBC/BASO | / | √ |
| Histogram | Histogram | ||
| White Blood Cell Histogram | WBC Histogram | √ | / |
| Red Blood Cell Histogram | RBC Histogram | √ | √ |
| Platelet Histogram | PLT Histogram | √ | √ |
Table 3: Scattergram of BC-5390 Auto Hematology Analyzer
| Name | Abbreviation. | CBC | CBC + DIFF |
|---|---|---|---|
| Differential Scattergram | Diff Scattergram | / | √ |
Comparison of Technological Characteristics:
The subject BC-5390 Auto Hematology Analyzer has similar technological characteristics, compared to the predicate device.
Both devices have similar intended uses as quantitative, automated hematology analyzers and leukocyte differential counters to be used in clinical laboratories for Invitro Diagnostic purpose.
Safety testing was conducted on the subject BC-5390 Auto Hematology Analyzer per IEC 61010 and passed the standard requirement.
Substantial equivalence studies performed and demonstrated that the BC-5390 Auto Hematology Analyzer is similar to the predicate Sysmex XE-2100 (K992875).
Principle of Operation:
The patient whole blood sample or the prediluted sample with diluent is aspirated and delivered to the measuring bath to be measurement methods used in this analyzer are:
{6}------------------------------------------------
- The Electrical Impedance method for determining the WBC/BAS, RBC and PLT by 1. counting the electric pulses that occur in each cell size category. The BC-5390 Auto Hematology Analyzer determines the blood cell volume and identifies rare and pathological cells by creating and analyzing histograms of the various cell populations using their respective pulse data.
- The colorimetric method is used for determining the Hemoglobin. 2.
-
- Flow Cytometry by laser is used to determine the WBC differential. When the blood cells suspended in the diluent pass through the flow cell, they are exposed to a laser beam. The intensity of scatter light reflects the blood cell size and intracellular density. A 2-dimensional distribution (scattergram) is drawn and analyzed to obtain the differential results.
Analysis Mode:
The BC-5390 provides the Closed Whole Blood Mode, Autoloader Whole Blood Mode and Closed Predilute Mode.
Test Panel:
The BC-5390 provides CBC and CBC+DIFF (CD) panels.
Specimen identification:
Specimen identification input is manual (by operator) or by barcode reader.
Specimen sampling and handling:
Samples are manually mixed and loaded into a sample compartment one at a time; or ten samples in a rack are automatically loaded, mixed and sampled for autoloader analysis. The BC-5390 processes anti-coagulated whole blood collected in a K2EDTA or K3EDTA on all modes.
Reagents, Calibrator and Controls
The details for reagents, calibrator and controls are presented below:
Reagents:
- M-53D DILUENT ●
- . M-5LEO(1) LYSE
- M-5LEO(II) LYSE ●
- M-53LH LYSE ●
{7}------------------------------------------------
- . PROBE CLEANSER
Controls: BC-5D Hematology Controls (High, Normal, Low) (K160606)
Three levels of Controls with low, normal and high levels are provided. It is recommended to perform the quality control check using these controls at intervals established by the laboratory procedures and local or national regulations.
Quality control performance was evaluated using BC-5D Hematology Controls.
Calibrator: SC-CAL PLUS Hematology Calibrator (cleared as K955925)
Calibration and verification of Calibration are performed with the previously cleared Calibrator SC-CAL PLUS Hematology Calibrator. The calibration and quality control should be performed according to the instruction for the Calibrator and to laboratory procedures and local or national regulations.
Software
The BC-5390 software system carries out the instrument function by taking the input information, triggering the functional modules, controlling and managing fluidic sequencing and measuring various blood cell parameters and recording the signals measured. Additionally, the software system also manages the counting results and sample information. Through its QA/QC program, the software system also provides daily maintenance functions. The software is designed to be intuitive user interface and to guide user workflows including error handling procedures. The software development processes and Hazard Analysis were conducted per FDA guidance.
Four Off-the-shelf (OTS) software products are used in BC-5390 (Windows 8.1 Professional 64bit, SQLite Database,Linux real-time operating system,Libxml2 Open Source file). The safety and effectiveness of the OTS software are assessed and verified in accordance with FDA Guidance "Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices, September 9, 1999".Complete software documentation are submitted in accordance with Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", 2005.
{8}------------------------------------------------
Special Control and Guidances
Special Controls and Standard Guidance were used as references for the BC-5390 Auto Analyzer design, verification and validation studies. See Table 4 for the list of references.
| FDAGuidance | Class II Special Controls Guidance Document: PremarketNotifications for Automated Differential Cell Counters forImmature or Abnormal Blood Cells; Final Guidance for Industryand FDA | 2001 |
|---|---|---|
| H20-A2 | Reference Leukocyte (WBC) Differential Count (Proportional) andEvaluation of Instrumental Methods; Approved Standard - SecondEdition | 2007 |
| H26-A2 | Validation, Verification and Quality Assurance of AutomatedHematology Analyzers; Approved Standard-Second Edition | 2010 |
| EP05-A2 | Evaluation of Precision Performance of Quantitative MeasurementMethods; Approved Guideline—Second Edition | 2004 |
| EP6-A | Evaluation of the Linearity of Quantitative MeasurementProcedure :A statistical Approach; Approved Guideline | 2003 |
| EP07-A2 | Interference Testing in Clinical Chemistry-Approved Guideline—Second Edition | 2005 |
| EP09-A3 | Measurement Procedure Comparison and Bias Estimation UsingPatient Samples; Approved Guideline - Third Edition | 2013 |
| EP28-A3c | Defining, Establishing, and Verifying Reference Intervals in theClinical Laboratory; Approved Guideline—Third Edition | 2010 |
| ICSHliterature | ICSH guidelines for the evaluation of blood cell analysersincluding those used for differential leucocyte and reticulocytecounting. International Council for Standardization in Hematology.Int. Jnl. Lab. Hem. 2014, 36, 613-627 | 2014 |
| EP25-A | Evaluation of Stability of In Vitro Diagnostic Reagents; ApprovedGuideline. | 2009 |
| EP17-A | Protocols for Determination of Limits of Detection and Limits ofQuantitation. Approved Guideline | 2004 |
Table 4: Special Control and Guidance Document Referenced
Performance Characteristics:
{9}------------------------------------------------
Method Comparison:
A total of 1531 whole blood samples (collected in K2EDTA) were analyzed at three actual user sites. Five instruments and multiple operators were used for the study. The subjects of the study with age range from 1 day to 100 years old, and with 652 females and 874 males (5 subjects gender was not recorded). Patient's samples covered the normal and most abnormal conditions for all parameters. For each whole blood sample, three manual wedge smears were prepared and stained with Wright-Giemsa stain. A 400 cell WBC differential was performed on two smears per CLSI H20-A2. All samples were testing on whole blood mode in parallel with BC-5390 analyzer and the predicate device. An estimation of the bias was determined for each parameter (Exclude R). The result demonstrated the BC-5390 analyzer met the pre-defined specification for the difference limits. Whole blood accuracy and regression vs predicate and the flagging ability of BC-5390 vs Manual differential comparison were performed. The following tables showed that the results are comparable.
| Parameters | N | Result Range | r | slope(95%CI) | intercept(95%CI) |
|---|---|---|---|---|---|
| WBC | 1091 | 0.06-177.15 | 0.999 | 1.006(0.9903-1.021) | -0.07511(-0.2123-0.06207) |
| Neu# | 789 | 0.27-59.17 | 0.999 | 1.011(0.9971-1.025) | -0.2916(-0.09452-0.03620) |
| Lym# | 864 | 0.00-9.68 | 0.993 | 0.9750(0.9636-0.9863) | -0.01264(-0.03416-0.008883) |
| Mon# | 799 | 0.00-2.21 | 0.966 | 0.9303(0.8912-0.9693) | 0.02702(0.01080-0.04324) |
| Eos# | 932 | 0.00-3.39 | 0.995 | 0.9939(0.9603-1.028) | 9.386E-04(-0.004361-0.006238) |
| Bas# | 877 | 0.00-0.69 | 0.624 | 0.9156(0.2266-1.605) | 0.006836(-0.006382-0.02005) |
| Neu% | 792 | 19.6-99.6 | 0.995 | 0.9840(0.9762-0.9919) | 1.487(0.9603-2.013) |
| Lym% | 866 | 0.0-77.6 | 0.995 | 0.9897(0.9814-0.9979) | -0.2723(-0.4864--0.05818) |
| Mon% | 792 | 0.20-16.40 | 0.930 | 0.9652(0.9300-1.000) | 0.2260(0.03859-0.4134) |
| Eos% | 936 | 0.00-37.40 | 0.994 | 0.9877(0.9689-1.006) | 0.02701(-0.01416-0.06817) |
| Bas% | 881 | 0.00-3.20 | 0.415 | 0.4308(0.2936-0.5679) | 0.2147(0.1821-0.2472) |
| RBC | 1451 | 1.13-7.75 | 0.997 | 1.004(0.9999-1.008) | -0.001771(-0.01777-0.01423) |
Table 5: The Correlation and Estimated Bias of BC-5390 (vs XE-2100, Combined)
{10}------------------------------------------------
| Parameters | N | Result Range | r | slope(95%CI) | intercept(95%CI) |
|---|---|---|---|---|---|
| HGB | 1486 | 3.30-23.1 | 0.998 | 1.010(1.005-1.014) | -0.05500(-0.1023-0.007685) |
| HCT | 1452 | 10.50-67.80 | 0.995 | 1.013(1.006-1.019) | -0.5671(-0.7991-0.3351) |
| MCV | 1452 | 53.70-127.10 | 0.993 | 1.036(1.029-1.043) | -3.763(-4.368--3.158) |
| MCH | 1409 | 13.90-41.20 | 0.977 | 1.020(1.008-1.032) | -0.5451(-0.8854-0.2048) |
| MCHC | 1408 | 23.80-37.70 | 0.825 | 1.184(1.130-1.238) | -5.834(-7.657-4.010) |
| RDW-CV | 1453 | 11.10-29.70 | 0.978 | 1.024(1.004-1.044) | -0.1895(-0.4807-0.1018) |
| RDW-SD | 1453 | 29.20-103.70 | 0.988 | 0.9878(0.9760-0.9997) | -0.08935(-0.6338-0.4551) |
| PLT | 1199 | 4.00-1879.00 | 0.996 | 1.019(1.012-1.026) | -1.157(-2.215-0.1002) |
| MPV | 1238 | 7.70-14.50 | 0.951 | 1.057(1.037-1.078) | -0.7742(-0.9964-0.5520) |
The WBC Morphology flagging rate for the BC-5390 was compared to the WBC manual differential for the same population of samples as shown in table below.
| Flags | BC-5390 |
|---|---|
| True Positive | 283 |
| True Negative | 937 |
| False Positive | 260 |
| False Negative | 34 |
| Total | 1514 |
| Sensitivity (TP %) | 89.3% |
| Specificity (TN %) | 78.3% |
| Efficiency | 80.6% |
| Table 6: The WBC Morphology Flagging Ability of BC-5390 (vs Manual) | ||||
|---|---|---|---|---|
| -- | --------------------------------------------------------------------- | -- | -- | -- |
Precision/Repeatability and Reproducibility
{11}------------------------------------------------
Repeatability study was conducted for each reported analyte parameter using whole blood samples collected in K2EDTA anticoagulant tubes either with normal reference intervals or around medical decision levels. Each of the samples was tested 10 times consecutively within the same day of sample collection. Three analyzers in one clinical site with at least two professional laboratory technicians were used to conduct the studies. The SD and CV of the different parameters of those samples were calculated. The results met the predefined performance acceptance specifications.
Reproducibility study was conducted to verify the potential performance variation derived from testing sites, operator, reagent lots. Multiple lots reagent (five lots of Diluent, two lots of LH Lyse and LEO(II) lyse, and three lots of LEO(I) lyse), three BC-5390 Analyzers and two operators at each sites were used to conduct the study. Three levels of samples were prepared using commercial control material BC-5D and tested in duplicate for each run, twice each day for 20 days. A total of 80 (2 duplicates, 2 runs and 20 days) replicates for each sample were tested at each clinical site and 240 replicates together for all three sites for each level control. Standard deviation and CV% were calculated for each measurand and the results obtained were within the specifications.
Linearity Range
WBC and PLT high-value analogs, from commercialized materials, were diluted to different values to cover the AMR of WBC and PLT respectively. RBC/HGB linearity was performed using dilutions prepared from fresh whole blood. The whole blood was concentrated to obtain specimens to test the high linearity limit. Proportional dilutions were prepared using the diluent for each parameter to create 7 subsequent dilutions. The expected values of the diluted samples were considered the true values. Each level of the samples was run in triplicate. The mean of three runs from each of the 7 dilutions across the linear range were used. Acceptable performance is indicated by the data fitting a linear regression line with a coefficient of determination (R2) of >0.95 and the parameters measured recovering within the bias limits for each parameters based on CLSI EP06-A. The results indicated that BC-5390 Auto Hematology Analyzer exhibits linearity across the claimed range.
Carrvover
Carryover performance was determined for parameters WBC, RBC, HGB, HCT and PLT. Testing was performed based on the different analytical cycle combinations of within mode and mode to mode. For each analytical cycle combination, whole blood and predilute sample with extremely elevated blood components (high sample) and with decreased blood components (low sample) were tested in triplicates in turn according to H26-A2.
For whole blood and predilute sampling, within mode to mode for sampling carryover was calculated and the results were within specifications (≤ 1.0%) for WBC, RBC, HGB, HCT and PLT. BC-5390 Analyzer demonstrated minimum carryover level within the defined specification for both Whole Blood and Predilute mode.
Interference
{12}------------------------------------------------
The potential interfering substances of BC-5390 were evaluated by measuring the whole blood samples and after adding different level of potential interfering substances. The paired difference testing was compared to the acceptance criteria. The impact of those substances were analyzed and the results demonstrated that there is no significant interference of bilirubin, WBC, and PLT for the results of BC-5390, but the high elevated concentration of Triglyceride (TG) and hemoglobin exhibits minor impact to the the HGB, MCH, MCHC parameters on the BC-5390. The information is disclosed in the BC-5390 Auto Hematology Analyzer Operator's Manual.
Other Compatibility Studies
Comparison of Whole Blood Mode and Predilute Mode
To demonstrate that BC-5390 analyzer performs equally on samples tested in different analysis modes (Whole blood and Predilute modes), 124 leftover whole blood samples representing the normal and medical conditions were collected in K2EDTA collection tube. All the samples were tested in duplicate on BC-5390 analyzer with Closed or Autoloader Whole blood CD mode and Closed Predilute CD mode. The results were analyzed according to CLSI EP09-A3. It showed that there was no difference when testing samples in either mode or when the sample contained parameters in normal or abnormal range due to medical conditions.
Comparison of CBC and CBC+DIFF Mode
To demonstrate that BC-5390 analyzer performs equally on samples tested in different analysis modes (CBC and CBC+DIFF modes), 103 leftover whole blood samples representing the normal and medical conditions were collected in K2EDTA collection tube for the comparison of Whole blood CBC and CD modes. Meanwhile, a total of 75 samples were collected in the same site for the comparison of Predilute CBC and CD modes. All the samples were tested in duplicate on BC-5390 analyzer with Closed or Autoloader Whole blood CD mode or Closed Predilute CD mode. The results were analyzed according to CLSI EP09-A3. It showed that there was no difference when testing samples in either mode or when the sample contained parameters in normal or abnormal range due to medical conditions.
Comparison of Capillary and Venous blood
To demonstrate the comparable performance between capillary and venous samples, 57 paired specimens were collected from donors by capillary method in K2EDTA microtainers and venipuncture method in K2EDTA collection tubes. Specimen levels were selected to cover the analytical measuring interval and medical decision levels for each parameter. Each sample was tested in the Closed whole blood CD mode at the U.S. site. The results were analyzed according to CLSI EP09-A3 and showed that the performance characteristics of the two specimen types were comparable.
{13}------------------------------------------------
Comparison of K2EDTA and K3EDTA Anticoagulants Samples
To evaluate whether samples collected in K2EDTA and K3EDTA have the same performance characteristics on BC-5390 analyzer, 70 paired fresh whole blood samples collected in K2EDTA and K3EDTA anticoagulant tubes were tested. The samples were selected with the targets to cover the analytical measuring interval and medical decision levels for each parameter. Each sample was tested in Autoloader/Closed whole blood CD mode on BC-5390 in the clinical laboratory in U.S. The results were analyzed according to CLSI EP09-A3 and showed that the performance characteristics of the specimens collected in the two anticoagulants are comparable.
Sample Stability
Specimen stability study was conducted to support the specimen stability claim as for whole blood is 36 hours at refrigerated temperature (2-8°C) and 24 hours when stored at 18-26°C, and for predilute sample the stability is 25 minutes when stored at 18-26°C before testing. Fresh samples covering normal and medical decision range were stored at at 2-8°C and then tested at defined intervals of immediate (0-1 hour), 4, 8, 12, 24, 26, 36 and 38 hours or stored at 18-26°C and then tested at the time points of immediate (0-1hour), 4, 8, 12, 14, 24 and 26 hours. Prediluted samples were also stored at 18-26°C and then tested at the time points of 5, 10, 15, 20, 25, 30, 35 minutes after sample preparation. The study results met the pre-defined acceptance criteria supporting the sample stability claims.
Reference Interval
A study was performed to assess the Adult Reference Ranges for BC-5390 analyzer using whole blood samples collected from 251 donors. In the study 121 adult male donors and 130 adult female donors between the ages of 19-70 were included. The non-parametric method and 95% confidence were used to calculate the lower and upper limits of the reference range according to EP28-A3c. The results are all shown in Table 7.
It is recommended that laboratories establish their own reference range based on the actual current patient population.
| Partitions* | Male | Female | Combined | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Parameters | (Female vsMale) | N | Low Limit(90%CI) | Upper Limit(90%CI) | N | Low Limit(90%CI) | Upper Limit(90%CI) | N | Low Limit(90%CI) | Upper Limit(90%CI) |
| WBC | YES | 121 | 3.782(3.560-4.000) | 12.049(11.420-12.580) | 130 | 4.374(4.130-5.000) | 11.193(10.380-11.850) | 251 | 3.972(3.780-4.330) | 11.726(11.120-12.060) |
| Neu# | NO | 121 | 1.744(1.460-2.050) | 9.728(7.870-9.780) | 130 | 2.229(2.060-2.580) | 8.363(7.230-8.690) | 251 | 2.029(1.740-2.190) | 8.578(7.860-9.280) |
| Lym# | NO | 121 | 1.181(1.150-1.400) | 4.268(3.220-4.920) | 130 | 1.288(0.920-1.460) | 3.778(3.410-4.110) | 251 | 1.214(1.150-1.400) | 3.858(3.430-4.280) |
| Mon# | YES | 121 | 0.210(0.150-0.250) | 0.869(0.780-0.920) | 130 | 0.211(0.160-0.240) | 0.667(0.630-0.710) | 251 | 0.210(0.160-0.240) | 0.787(0.710-0.870) |
| Partitions* | Male | Female | Combined | |||||||
| Parameters | (Female vs | N | Low Limit | Upper Limit | N | Low Limit | Upper Limit | N | Low Limit | Upper Limit |
| Male) | (90%CI) | (90%CI) | (90%CI) | (90%CI) | (90%CI) | (90%CI) | ||||
| Eos# | NO | 121 | 0.031(0.000-0.050) | 0.599(0.540-0.740) | 130 | 0.023(0.020-0.030) | 0.487(0.420-1.090) | 251 | 0.030(0.020-0.040) | 0.557(0.490-0.630) |
| Bas# | YES | 121 | 0.010(0.000-0.010) | 0.100(0.070-0.150) | 130 | 0.010(0.010-0.010) | 0.080(0.060-0.120) | 251 | 0.010(0.010-0.010) | 0.080(0.070-0.110) |
| Neu% | NO | 121 | 39.06(37.80-42.30) | 78.14(72.40-80.00) | 130 | 41.94(41.20-43.70) | 76.49(74.60-83.60) | 251 | 41.33(39.00-42.40) | 76.90(74.70-79.00) |
| Lym% | NO | 121 | 14.13(11.80-18.30) | 50.47(47.40-54.40) | 130 | 16.14(8.90-19.50) | 49.67(48.20-52.30) | 251 | 15.13(12.00-18.30) | 49.84(48.70-51.10) |
| Mon% | YES | 121 | 3.61(2.60-4.00) | 10.90(9.70-13.80) | 130 | 3.63(3.20-3.90) | 8.87(7.60-9.30) | 251 | 3.63(3.40-3.90) | 9.77(8.80-10.90) |
| Eos% | NO | 121 | 0.41(0.00-0.60) | 8.82(6.60-10.50) | 130 | 0.40(0.30-0.50) | 8.12(6.00-13.90) | 251 | 0.40(0.30-0.50) | 8.10(6.60-9.10) |
| Bas% | YES | 121 | 0.20(0.10-0.20) | 1.20(0.90-2.20) | 130 | 0.20(0.10-0.20) | 1.00(0.90-1.20) | 251 | 0.20(0.10-0.20) | 1.10(0.90-1.20) |
| RBC | YES | 121 | 4.231(3.660-4.370) | 6.219(6.000-6.380) | 130 | 3.728(3.540-4.010) | 5.454(5.310-5.940) | 251 | 3.907(3.660-4.130) | 6.021(5.920-6.220) |
| HGB | YES | 121 | 13.01(11.10-13.20) | 18.28(17.40-19.70) | 130 | 10.18(9.80-11.10) | 16.27(15.90-16.50) | 251 | 10.68(10.10-11.50) | 17.47(17.10-18.30) |
| HCT | YES | 121 | 38.50(34.60-39.20) | 54.13(51.40-56.70) | 130 | 31.93(31.70-34.30) | 47.30(46.70-48.70) | 251 | 33.81(31.90-35.40) | 51.47(50.60-54.20) |
| MCV | NO | 121 | 74.28(68.20-79.70) | 100.68(97.10-107.00) | 130 | 74.33(68.90-75.90) | 96.87(94.00-98.80) | 251 | 74.33(71.10-76.20) | 97.64(96.00-100.80) |
| MCH | NO | 121 | 24.96(22.90-26.60) | 34.54(32.80-36.50) | 130 | 23.36(22.40-25.10) | 32.40(31.90-34.50) | 251 | 23.90(22.90-25.10) | 33.18(32.40-34.60) |
| MCHC | YES | 121 | 32.01(30.10-32.50) | 36.00(35.80-36.20) | 130 | 30.80(29.40-31.40) | 35.87(35.20-36.10) | 251 | 31.02(30.30-31.90) | 35.90(35.70-36.10) |
| RDW-CV | YES | 121 | 12.01(11.90-12.30) | 17.39(15.80-18.50) | 130 | 11.80(11.50-12.10) | 18.63(17.50-20.50) | 251 | 12.00(11.80-12.20) | 17.67(17.20-18.90) |
| RDW-SD | NO | 121 | 36.74(35.20-37.90) | 55.63(51.30-58.30) | 130 | 37.08(34.50-39.10) | 56.00(52.60-57.90) | 251 | 37.09(35.90-38.00) | 55.55(52.60-57.70) |
| PLT | YES | 121 | 141.1(120.0-154.0) | 456.6(373.0-608.0) | 130 | 186.4(169.0-203.0) | 405.5(385.0-481.0) | 251 | 147.7(141.0-171.0) | 435.4(387.0-461.0) |
| MPV | NO | 121 | 9.30(9.10-9.40) | 14.26(13.00-15.90) | 130 | 9.30(9.20-9.40) | 13.59(12.70-14.20) | 251 | 9.30(9.20-9.40) | 13.61(12.90-14.30) |
Table 7: The Reference Intervals of BC-5390
{14}------------------------------------------------
Note; * in this table denoted that if significantly difference exists between Male and Female according to CLSI EP28-A3c. If noted "Yes", it indicated that a separate reference interval should be considered for the Male and Female.
In addition, a literature reference interval verification study was performed in one clinical site for a total of 161 pediatric samples, including neonate (N=45), infant (N=26), child (N=57), and adolescent (N=33). These samples were collected from pediatric patients
{15}------------------------------------------------
ranging from neonates to 21 years of age. The results of this study verified the pediatric intervals published in Mayo and IOWA literature.
Please note, to substantiate a pediatric claim, we have included pediatric subjects in the method comparison and validated literature reference intervals according to FDA's recommendation in the pre-sub.
Determination of Limit of Blank (LoB), Limits of Detection (LoD) and Limit of Quantitation (LoQ)
Limit of Blank was determined using five blank samples (diluent) which just contain the dilute solution. To determine the Limits of Detection and Limit of Quantitation, five low levels of samples were created by adding whole blood samples to diluent to reach approximately 1-4 times of the LoB. Each blank sample or low level sample was tested 12 times at one clinical laboratory in the whole blood mode and predilute mode respectively on three BC-5390 analyzers. Each of the low level samples were tested 5 times on the Sysmex XE-2100 analyzer. The results were analyzed according to CLSI EP17-A. The maximum LoB, LoD, or LoQ value of the three (3) analyzers is taken as the reported value.
Substantial Equivalence
Table 8 and Table 9 summarize the similarities and differences between the BC-5390 Auto Hematology Analyzer and the predicate device XE-2100 Automated Hematology Analyzer.
| Similarities | ||
|---|---|---|
| Item | Subject Device BC-5390 | Predicate XE-2100 (K992875) |
| Overviewlayout | Image: BC-5390 | Image: XE-2100 |
| IntendedUse | The BC-5390 Auto HematologyAnalyzer is a quantitative,automated hematology analyzerfor in vitro diagnostic use inclinical laboratories. The BC-5390 Auto Hematology Analyzerprovides complete blood count(WBC, RBC, HGB, HCT, MCV, | The Sysmex XE-2100 is a multi-parameter hematology analyzerintended to classify the followingformed elements in anti-coagulatedblood.WBC, Neut%/#, Lymph%/#,Mono%/#, Eos%/#, Baso%/#,NRBC%/# , RBC, HGB, HCT, |
| Similarities | ||
| MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV) and leukocyte 5-Part differential (Neu#, Lym#,Mon#, Eos#, Bas#, Neu%,Lym%, Mon%, Eos%, Bas%) forwhole blood specimens, collectedin a salt of EDTA [dipotassium(K2) or tripotassium (K3)]obtained from venous orcapillary blood collection. Thepurpose of the BC-5390 AutoHematology Analyzer is toidentify the normal humanpatient, with normal system-generated parameters, frompatients whose results requireadditional studies. | MCV, MCH, MCHC, RDW-CV,RDW-SD, RET%/#, IRF, PLT, MPV | |
| Parameters | Parameters(21):WBC, Neu%, Lym%, Mon%,Eos%, Bas%, Neu#, Lym#,Mon#, Eos#, Bas#, RBC, HGB,HCT, MCV, MCH, MCHC,RDW-CV, RDW-SD, PLT, MPV | Same |
| TestPrinciple(RBC/PLT) | Impedance method | Same |
| TestPrinciple(HGB) | Colorimetric method | Same |
| SampleType | Whole blood | Same |
| SampleProcessing | Utilizes an automatic sampling,diluting and mixing device forsample processing | Same |
| Similarities | ||
| DataAnalysis | Raw data analyzed by algorithmsto generate reported parameters | Same |
| DataReporting | Display, printing andtransmission of data to LIS/HIS. | Same |
| Sample ID | Manual or Automatic barcodescan of sample tube identifier ormanual keyboard entry of sampleidentifier | Same |
| SoftwareRisk Level | Moderate | Same |
| I/OInterfaces | One LAN interface | Same |
| Software | DMU (Data Management Unit)Function: Managing data, such asediting worklist, reviewingresults, generating reports,communicating with LIS/HIS,etc. | IPU (Information Processing Unit)Function: Same |
Table 8: Device Comparison Table - Similarities
{16}------------------------------------------------
{17}------------------------------------------------
Table 9: Device Comparison Table - Differences
| Differences | ||
|---|---|---|
| Item | Device BC-5390 | Predicate XE-2100 (K992875) |
| Parameters(Reticulocytes,NucleatedRBC) | Not Available | RET%/#, IRF, NRBC%/# |
| Reagents | M-53D DILUENTM-5LEO(I) LYSEM-5LEO(II) LYSEM-53LH LYSEPROBE CLEANSER | CELLPACK, CELLSHEATHSTROMATOLYSER-4DSSTROMATOLYSER-4DLSTROMATOLYSER-FB, SULFOLYSERCELLCLEAN |
| Differences | ||
| Item | Device BC-5390 | Predicate XE-2100 (K992875) |
| Plus Reagents with IMI, RET, NRBC(STROMATOLYSER-NR,STROMATOLYSER-IM,RET-SEARCH II) | ||
| Test Panel | CBCCBC+DIFF | SamePlus RET and NRBC |
| TestPrinciple(WBC) | WBC: Flow Cytometry byLaser and ElectricalImpedance Method | WBC: Flow Cytometry method usingsemiconductor laser detection method |
| Throughput | Closed Whole Blood Mode:Upper capacity of 51samples/hourAutoloader Whole BloodMode: Upper capacity of60 samples/hourClosed Predilute Mode:Upper capacity of 53samples/hour | CBC, CBC+DIFF: Approx. 150samples/hourMode with NRBC, RET: Approx. 113-150samples/hour |
| SampleVolume | Autoloader Whole BloodMode and Closed WholeBlood Mode: 33μLClosed Predilute Mode:20μL | Closed and Sampler Mode: Approx.200μLManual Mode: Approx. 130μLCapillary Mode: 40μL |
| Sampleassignmentmechanism | Using the sampleassignment Syringe toassign the sample to thebaths in sequence | Using the sample rotary valve (SRV) to assignthe sample to the baths simultaneously |
| DriveSource Unit | Pump | Pneumatic unit |
| AnalysisMode | Closed Whole Blood ModeAutoloader Whole BloodModeClosed Predilute Mode | Closed ModeAutoloader ModePredilute Mode andManual Mode |
| SampleAnticoagulant | K2EDTA or K3EDTA | K2EDTA or K3EDTA And EDTA-2Na |
| Differences | ||
| Item | Device BC-5390 | Predicate XE-2100 (K992875) |
| Controls | BC-5D HematologyControl - 3 Levels of High,Normal and Low | e-CHECK – 3 Levels |
| Calibrator | SC-CAL PLUSHematology Calibrator | SCS-1000 Calibrator |
| QualityControlTechniques | L-J QC ProgramX-B QC Program | L-J QC ProgramX-M QC Program |
| DataStorage | Analysis data withhistograms andscattergrams: 100,000samplesPatient information:100,000 personsOrder information: 4,000samplesQuality control (QC) files: 60files | Analysis data with histograms andscattergrams: 10,000 samplesPatient information: 5,000 personsOrder information: 1,000 samplesQuality control (QC) files: 20 files |
{18}------------------------------------------------
{19}------------------------------------------------
Substantial Equivalence Conclusion
The BC-5390 Auto Hematology Analyzer has the same intended use as the predicate device, the XE-2100 as cleared in K992875. They are the quantitative, automated hematology analyzer and leukocyte differential counter for In Vitro Diagnostic Use in clinical laboratories. These analyzers may be used in adult and pediatric population. Both analyzers are only to be used by trained laboratory professionals to identify the normal patient, with all normal system-generated parameters, and to flag or identify patient results that require additional studies. The analyzer provides analysis results of 21 parameters of human blood, 3 histograms and 1 scattergram. The performance comparison study results demonstrated that the BC-5390 Auto Hematology Analyzer is substantially equivalent to the predicate device, the XE-2100 Automated Hematology Analyzer (K992875) and the differences do not raise new questions of the safety and effectiveness.
§ 864.5220 Automated differential cell counter.
(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”