K Number
K160429
Device Name
BC-5390 Auto Hematology Analyzer
Date Cleared
2016-09-01

(198 days)

Product Code
Regulation Number
864.5220
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology Analyzer for in vitro diagnostic use in clinical laboratories. The BC-5390 Auto Hematology Analyzer provides complete blood count (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV) and leukocyte 5-Part differential (Neut, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%) for whole blood specimens collected in a salt of EDTA [dipotassium (K2) or tripotassium (K3)) obtained from venous or capillary blood collection. The purpose of the BC-5390 Auto Hematology Analyzer is to identify the normal human patient, with normal system-generated parameters, from patients whose results require additional studies.
Device Description
The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter for In Vitro Diagnostic Use in clinical laboratories. It is only to be used by trained laboratory professionals to identify the normal patient, with all normal system-generated parameters, and to flag or identify patient results that require additional studies. The BC-5390 Auto Hematology Analyzer system consists of: - Instrument: Sample Processing Unit (SPU) and Data Managing Unit (DMU) - . Reagents M-53D DILUENT M-5LEO(I) LYSE M-5LEO(II) LYSE M-53LH LYSE PROBE CLEANSER - Controls BC-5D Hematology Control (High, Normal, Low, Pending) Note: Controls for BC-5390 Auto Hematology Analyzer will be submitted in parallel with this 510(k) by R&D Systems as separate 510(k). - . Calibrator SC-CAL PLUS Hematology Calibrator (cleared as K955925) The Analyzer provides analysis results of 21 parameters, 3 histograms and 1 scattergram of human blood. It supports two test panels: CBC and CBC+DIFF.
More Information

Not Found

No
The summary describes a standard automated hematology analyzer using established methods for blood cell counting and differentiation. There is no mention of AI, ML, or related concepts in the device description, intended use, or performance studies. The flagging ability is based on system-generated parameters, not explicitly described as AI/ML-driven.

No
The device is an in vitro diagnostic device used for analyzing blood components and is not intended for treating or preventing disease.

Yes

Explanation: The "Intended Use / Indications for Use" section explicitly states that the device is "for in vitro diagnostic use in clinical laboratories" and its purpose is to "identify the normal human patient, with normal system-generated parameters, from patients whose results require additional studies." This clearly indicates its role in diagnosing or aiding in the diagnosis of medical conditions by analyzing blood samples.

No

The device description explicitly states the system consists of an "Instrument: Sample Processing Unit (SPU) and Data Managing Unit (DMU)", reagents, controls, and a calibrator, indicating it includes significant hardware components beyond just software.

Yes, this device is an IVD (In Vitro Diagnostic).

The document explicitly states in the "Intended Use / Indications for Use" section: "The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology Analyzer for in vitro diagnostic use in clinical laboratories."

It is designed to analyze biological specimens (whole blood) outside of the body to provide information for diagnostic purposes.

N/A

Intended Use / Indications for Use

The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology analyzer for in vitro diagnostic use in clinical laboratories. The BC-5390 Auto Hematology Analyzer provides complete blood count (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV) and leukocyte 5-Part differential (Neut, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%) for whole blood specimens collected in a salt of EDTA [dipotassium (K2) or tripotassium (K3)) obtained from venous or capillary blood collection. The purpose of the BC-5390 Auto Hematology Analyzer is to identify the normal human patient, with normal system-generated parameters, from patients whose results require additional studies.

Product codes (comma separated list FDA assigned to the subject device)

GKZ

Device Description

The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter for In Vitro Diagnostic Use in clinical laboratories. It is only to be used by trained laboratory professionals to identify the normal patient, with all normal system-generated parameters, and to flag or identify patient results that require additional studies.

The BC-5390 Auto Hematology Analyzer system consists of:

  • Instrument: Sample Processing Unit (SPU) and Data Managing Unit (DMU)
  • . Reagents M-53D DILUENT M-5LEO(I) LYSE M-5LEO(II) LYSE M-53LH LYSE PROBE CLEANSER
  • Controls
    BC-5D Hematology Control (High, Normal, Low, Pending) Note: Controls for BC-5390 Auto Hematology Analyzer will be submitted in parallel with this 510(k) by R&D Systems as separate 510(k).
  • . Calibrator
    SC-CAL PLUS Hematology Calibrator (cleared as K955925)

The Analyzer provides analysis results of 21 parameters, 3 histograms and 1 scattergram of human blood. It supports two test panels: CBC and CBC+DIFF.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

The purpose of the BC-5390 Auto Hematology Analyzer is to identify the normal human patient, with normal system-generated parameters, from patients whose results require additional studies.
These analyzers may be used in adult and pediatric population.

Intended User / Care Setting

In vitro diagnostic use in clinical laboratories.
trained laboratory professionals

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Method Comparison: A total of 1531 whole blood samples (collected in K2EDTA) were analyzed at three actual user sites. Five instruments and multiple operators were used for the study. The subjects of the study with age range from 1 day to 100 years old, and with 652 females and 874 males (5 subjects gender was not recorded). Patient's samples covered the normal and most abnormal conditions for all parameters. For each whole blood sample, three manual wedge smears were prepared and stained with Wright-Giemsa stain. A 400 cell WBC differential was performed on two smears per CLSI H20-A2. All samples were testing on whole blood mode in parallel with BC-5390 analyzer and the predicate device.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Method Comparison:

  • Study Type: Method Comparison, Whole Blood Accuracy and Regression, Flagging ability
  • Sample Size: 1531 whole blood samples
  • Key Results: The result demonstrated the BC-5390 analyzer met the pre-defined specification for the difference limits. Whole blood accuracy and regression vs predicate and the flagging ability of BC-5390 vs Manual differential comparison were performed. The following tables showed that the results are comparable.

Precision/Repeatability and Reproducibility:

  • Study Type: Repeatability and Reproducibility
  • Sample Size: Not explicitly stated for repeatability, but for reproducibility, a total of 240 replicates for each level control across three sites (80 replicates per site for each sample level).
  • Key Results: Repeatability study met predefined performance acceptance specifications. Reproducibility study results obtained were within specifications.

Linearity Range:

  • Study Type: Linearity Study
  • Sample Size: Not explicitly stated, but involved 7 subsequent dilutions for WBC, PLT, RBC/HGB.
  • Key Results: The results indicated that BC-5390 Auto Hematology Analyzer exhibits linearity across the claimed range.

Carryover:

  • Study Type: Carryover performance
  • Sample Size: Not explicitly stated, samples tested in triplicates.
  • Key Results: BC-5390 Analyzer demonstrated minimum carryover level within the defined specification for both Whole Blood and Predilute mode (≤ 1.0% for WBC, RBC, HGB, HCT and PLT).

Interference:

  • Study Type: Interference evaluation
  • Sample Size: Not explicitly stated.
  • Key Results: Results demonstrated no significant interference of bilirubin, WBC, and PLT for the results of BC-5390, but high elevated concentration of Triglyceride (TG) and hemoglobin exhibits minor impact to the HGB, MCH, MCHC parameters.

Other Compatibility Studies:

  • Comparison of Whole Blood Mode and Predilute Mode: 124 leftover whole blood samples. No difference when testing samples in either mode.
  • Comparison of CBC and CBC+DIFF Mode: 103 leftover whole blood samples for whole blood modes, 75 samples for predilute modes. No difference when testing samples in either mode.
  • Comparison of Capillary and Venous blood: 57 paired specimens. Performance characteristics were comparable.
  • Comparison of K2EDTA and K3EDTA Anticoagulants Samples: 70 paired fresh whole blood samples. Performance characteristics were comparable.

Sample Stability:

  • Study Type: Specimen stability study
  • Sample Size: Fresh samples covering normal and medical decision range.
  • Key Results: Study results met predefined acceptance criteria supporting the sample stability claims (36 hours at 2-8°C, 24 hours at 18-26°C for whole blood; 25 minutes at 18-26°C for predilute).

Reference Interval:

  • Study Type: Assessment of Adult Reference Ranges and Pediatric Literature Reference Interval Verification
  • Sample Size: 251 donors (121 adult male, 130 adult female) for adult reference range; 161 pediatric samples (45 neonate, 26 infant, 57 child, 33 adolescent) for pediatric verification.
  • Key Results: Adult reference ranges calculated. Pediatric intervals published in Mayo and IOWA literature were verified.

Determination of Limit of Blank (LoB), Limits of Detection (LoD) and Limit of Quantitation (LoQ):

  • Study Type: LoB, LoD, LoQ determination
  • Sample Size: Five blank samples, five low levels of samples (each tested 12 times on subject device, 5 times on predicate device).
  • Key Results: The maximum LoB, LoD, or LoQ value of the three (3) analyzers is taken as the reported value.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

WBC Morphology Flagging Ability (vs Manual):

  • True Positive: 283
  • True Negative: 937
  • False Positive: 260
  • False Negative: 34
  • Total: 1514
  • Sensitivity (TP %): 89.3%
  • Specificity (TN %): 78.3%
  • Efficiency: 80.6%

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

SYSMEX XE-2100 Automated Hematology Analyzer, K992875

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

0

Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health and Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 1, 2016

Mindray Bio-medical Electronics Co., LTD c/o Jinjie Hu, Ph.D. Biologics Consulting Group 400 N. Washington St. Suite 100 Alexandria, VA 22314

Re: K160429

Trade/Device Name: BC-5390 Auto Hematology Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated Differential Cell Counter Regulatory Class: Class II Product Code: GKZ Dated: August 30, 2016 Received: August 31, 2016

Dear Dr. Hu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

1

Page 2 - Jinjie Hu, Ph.D.

CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Kelly Oliner AS

For

Leonthena R. Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known)

K160429

Device Name

BC-5390 Auto Hematology Analyzer

Indications for Use (Describe)

The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology Analyzer for in vitro diagnostic use in clinical laboratories. The BC-5390 Auto Hematology Analyzer provides complete blood count (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV) and leukocyte 5-Part differential (Neut, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%) for whole blood specimens collected in a salt of EDTA [dipotassium (K2) or tripotassium (K3)) obtained from venous or capillary blood collection. The purpose of the BC-5390 Auto Hematology Analyzer is to identify the normal human patient, with normal system-generated parameters, from patients whose results require additional studies.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

3

510(k)PremarketNotification

510(K) SUMMARY

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the BC-5390 Auto Hematology Analyzer is provided below.

Device Common Name:Auto Hematology Analyzer
Device Proprietary Name:BC-5390 Auto Hematology Analyzer
Submitter:Mindray Bio-medical Electronics Co., LTD
Mindray Building, Keji 12th Road South, High-tech
Industrial Park,
Nanshan, Shenzhen 518057, P.R. China
Contact:Jinjie Hu Ph.D.
400 N. Washington Street
Suite 100
Alexandria, VA 22314
Tel: 301-814-4985
Email: jhu@bcg-usa.com
Classification Regulation:21 CFR 864.5220, Class II
Panel:Hematology
Product Code:GKZ

Predicate Device:

SYSMEX XE-2100 Automated Hematology Analyzer, K992875

Indication for Use:

The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology analyzer for in vitro diagnostic use in clinical laboratories. The BC-5390 Auto Hematology Analyzer provides complete blood count (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV) and leukocyte 5-Part differential ((Neu#, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%) for whole blood specimens collected in a salt of EDTA [dipotassium (K2) or tripotassium (K3)] obtained from venous or capillary blood collection. The purpose of the BC-5390 Auto Hematology Analyzer is to identify the normal human patient, with normal system-generated parameters, from patients whose results require additional studies.

Device Description:

4

The BC-5390 Auto Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter for In Vitro Diagnostic Use in clinical laboratories. It is only to be used by trained laboratory professionals to identify the normal patient, with all normal system-generated parameters, and to flag or identify patient results that require additional studies.

The BC-5390 Auto Hematology Analyzer system consists of:

  • Instrument: Sample Processing Unit (SPU) and Data Managing Unit (DMU)
  • . Reagents M-53D DILUENT M-5LEO(I) LYSE M-5LEO(II) LYSE M-53LH LYSE PROBE CLEANSER
  • Controls

BC-5D Hematology Control (High, Normal, Low, Pending) Note: Controls for BC-5390 Auto Hematology Analyzer will be submitted in parallel with this 510(k) by R&D Systems as separate 510(k).

  • . Calibrator
    SC-CAL PLUS Hematology Calibrator (cleared as K955925)

The Analyzer provides analysis results of 21 parameters, 3 histograms and 1 scattergram of human blood. It supports two test panels: CBC and CBC+DIFF. The abbreviations for all parameters are listed in Table 1, Table 2 and Table 3.

Analysis ParameterAbbreviationCBCCBC + DIFF
White Blood Cell CountWBC
Neutrophil CountNeu#/
Lymphocyte CountLym#/
Monocyte CountMon#/
Eosinophil CountEos#/
Basophil CountBas#/
Neutrophil PercentNeu%/
Lymphocyte PercentLym%/
Monocyte PercentMon%/
Eosinophil PercentEos%/
Basophil PercentBas%/
Red Blood Cell CountRBC
Hemoglobin ConcentrationHGB
HematocritHCT
Mean Corpuscular VolumeMCV
Table 1: Parameters of BC-5390 Auto Hematology Analyzer
-----------------------------------------------------------

5

510(k)PremarketNotification

| Mean Corpuscular

HemoglobinMCH
Mean Corpuscular
Hemoglobin ConcentrationMCHC
Red Blood Cell Distribution
Width (Coefficient of
Variation)RDW-CV
Red Blood Cell Distribution
Width (Standard Deviation)RDW-SD
Platelet CountPLT
Mean Platelet VolumeMPV

Table 2: Histograms of BC-5390 Auto Hematology Analyzer

NameAbbreviationCBCCBC + DIFF
White Blood Cell/ BasophilsWBC/BASO/
HistogramHistogram
White Blood Cell HistogramWBC Histogram/
Red Blood Cell HistogramRBC Histogram
Platelet HistogramPLT Histogram

Table 3: Scattergram of BC-5390 Auto Hematology Analyzer

NameAbbreviation.CBCCBC + DIFF
Differential ScattergramDiff Scattergram/

Comparison of Technological Characteristics:

The subject BC-5390 Auto Hematology Analyzer has similar technological characteristics, compared to the predicate device.

Both devices have similar intended uses as quantitative, automated hematology analyzers and leukocyte differential counters to be used in clinical laboratories for Invitro Diagnostic purpose.

Safety testing was conducted on the subject BC-5390 Auto Hematology Analyzer per IEC 61010 and passed the standard requirement.

Substantial equivalence studies performed and demonstrated that the BC-5390 Auto Hematology Analyzer is similar to the predicate Sysmex XE-2100 (K992875).

Principle of Operation:

The patient whole blood sample or the prediluted sample with diluent is aspirated and delivered to the measuring bath to be measurement methods used in this analyzer are:

6

  • The Electrical Impedance method for determining the WBC/BAS, RBC and PLT by 1. counting the electric pulses that occur in each cell size category. The BC-5390 Auto Hematology Analyzer determines the blood cell volume and identifies rare and pathological cells by creating and analyzing histograms of the various cell populations using their respective pulse data.
  • The colorimetric method is used for determining the Hemoglobin. 2.
    1. Flow Cytometry by laser is used to determine the WBC differential. When the blood cells suspended in the diluent pass through the flow cell, they are exposed to a laser beam. The intensity of scatter light reflects the blood cell size and intracellular density. A 2-dimensional distribution (scattergram) is drawn and analyzed to obtain the differential results.

Analysis Mode:

The BC-5390 provides the Closed Whole Blood Mode, Autoloader Whole Blood Mode and Closed Predilute Mode.

Test Panel:

The BC-5390 provides CBC and CBC+DIFF (CD) panels.

Specimen identification:

Specimen identification input is manual (by operator) or by barcode reader.

Specimen sampling and handling:

Samples are manually mixed and loaded into a sample compartment one at a time; or ten samples in a rack are automatically loaded, mixed and sampled for autoloader analysis. The BC-5390 processes anti-coagulated whole blood collected in a K2EDTA or K3EDTA on all modes.

Reagents, Calibrator and Controls

The details for reagents, calibrator and controls are presented below:

Reagents:

  • M-53D DILUENT ●
  • . M-5LEO(1) LYSE
  • M-5LEO(II) LYSE ●
  • M-53LH LYSE ●

7

  • . PROBE CLEANSER

Controls: BC-5D Hematology Controls (High, Normal, Low) (K160606)

Three levels of Controls with low, normal and high levels are provided. It is recommended to perform the quality control check using these controls at intervals established by the laboratory procedures and local or national regulations.

Quality control performance was evaluated using BC-5D Hematology Controls.

Calibrator: SC-CAL PLUS Hematology Calibrator (cleared as K955925)

Calibration and verification of Calibration are performed with the previously cleared Calibrator SC-CAL PLUS Hematology Calibrator. The calibration and quality control should be performed according to the instruction for the Calibrator and to laboratory procedures and local or national regulations.

Software

The BC-5390 software system carries out the instrument function by taking the input information, triggering the functional modules, controlling and managing fluidic sequencing and measuring various blood cell parameters and recording the signals measured. Additionally, the software system also manages the counting results and sample information. Through its QA/QC program, the software system also provides daily maintenance functions. The software is designed to be intuitive user interface and to guide user workflows including error handling procedures. The software development processes and Hazard Analysis were conducted per FDA guidance.

Four Off-the-shelf (OTS) software products are used in BC-5390 (Windows 8.1 Professional 64bit, SQLite Database,Linux real-time operating system,Libxml2 Open Source file). The safety and effectiveness of the OTS software are assessed and verified in accordance with FDA Guidance "Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices, September 9, 1999".Complete software documentation are submitted in accordance with Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", 2005.

8

Special Control and Guidances

Special Controls and Standard Guidance were used as references for the BC-5390 Auto Analyzer design, verification and validation studies. See Table 4 for the list of references.

| FDA
Guidance | Class II Special Controls Guidance Document: Premarket
Notifications for Automated Differential Cell Counters for
Immature or Abnormal Blood Cells; Final Guidance for Industry
and FDA | 2001 |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| H20-A2 | Reference Leukocyte (WBC) Differential Count (Proportional) and
Evaluation of Instrumental Methods; Approved Standard - Second
Edition | 2007 |
| H26-A2 | Validation, Verification and Quality Assurance of Automated
Hematology Analyzers; Approved Standard-Second Edition | 2010 |
| EP05-A2 | Evaluation of Precision Performance of Quantitative Measurement
Methods; Approved Guideline—Second Edition | 2004 |
| EP6-A | Evaluation of the Linearity of Quantitative Measurement
Procedure :A statistical Approach; Approved Guideline | 2003 |
| EP07-A2 | Interference Testing in Clinical Chemistry-Approved Guideline—
Second Edition | 2005 |
| EP09-A3 | Measurement Procedure Comparison and Bias Estimation Using
Patient Samples; Approved Guideline - Third Edition | 2013 |
| EP28-A3c | Defining, Establishing, and Verifying Reference Intervals in the
Clinical Laboratory; Approved Guideline—Third Edition | 2010 |
| ICSH
literature | ICSH guidelines for the evaluation of blood cell analysers
including those used for differential leucocyte and reticulocyte
counting. International Council for Standardization in Hematology.
Int. Jnl. Lab. Hem. 2014, 36, 613-627 | 2014 |
| EP25-A | Evaluation of Stability of In Vitro Diagnostic Reagents; Approved
Guideline. | 2009 |
| EP17-A | Protocols for Determination of Limits of Detection and Limits of
Quantitation. Approved Guideline | 2004 |

Table 4: Special Control and Guidance Document Referenced

Performance Characteristics:

9

Method Comparison:

A total of 1531 whole blood samples (collected in K2EDTA) were analyzed at three actual user sites. Five instruments and multiple operators were used for the study. The subjects of the study with age range from 1 day to 100 years old, and with 652 females and 874 males (5 subjects gender was not recorded). Patient's samples covered the normal and most abnormal conditions for all parameters. For each whole blood sample, three manual wedge smears were prepared and stained with Wright-Giemsa stain. A 400 cell WBC differential was performed on two smears per CLSI H20-A2. All samples were testing on whole blood mode in parallel with BC-5390 analyzer and the predicate device. An estimation of the bias was determined for each parameter (Exclude R). The result demonstrated the BC-5390 analyzer met the pre-defined specification for the difference limits. Whole blood accuracy and regression vs predicate and the flagging ability of BC-5390 vs Manual differential comparison were performed. The following tables showed that the results are comparable.

| Parameters | N | Result Range | r | slope
(95%CI) | intercept
(95%CI) |
|------------|------|--------------|-------|---------------------------|-----------------------------------|
| WBC | 1091 | 0.06-177.15 | 0.999 | 1.006
(0.9903-1.021) | -0.07511
(-0.2123-0.06207) |
| Neu# | 789 | 0.27-59.17 | 0.999 | 1.011
(0.9971-1.025) | -0.2916
(-0.09452-0.03620) |
| Lym# | 864 | 0.00-9.68 | 0.993 | 0.9750
(0.9636-0.9863) | -0.01264
(-0.03416-0.008883) |
| Mon# | 799 | 0.00-2.21 | 0.966 | 0.9303
(0.8912-0.9693) | 0.02702
(0.01080-0.04324) |
| Eos# | 932 | 0.00-3.39 | 0.995 | 0.9939
(0.9603-1.028) | 9.386E-04
(-0.004361-0.006238) |
| Bas# | 877 | 0.00-0.69 | 0.624 | 0.9156
(0.2266-1.605) | 0.006836
(-0.006382-0.02005) |
| Neu% | 792 | 19.6-99.6 | 0.995 | 0.9840
(0.9762-0.9919) | 1.487
(0.9603-2.013) |
| Lym% | 866 | 0.0-77.6 | 0.995 | 0.9897
(0.9814-0.9979) | -0.2723
(-0.4864--0.05818) |
| Mon% | 792 | 0.20-16.40 | 0.930 | 0.9652
(0.9300-1.000) | 0.2260
(0.03859-0.4134) |
| Eos% | 936 | 0.00-37.40 | 0.994 | 0.9877
(0.9689-1.006) | 0.02701
(-0.01416-0.06817) |
| Bas% | 881 | 0.00-3.20 | 0.415 | 0.4308
(0.2936-0.5679) | 0.2147
(0.1821-0.2472) |
| RBC | 1451 | 1.13-7.75 | 0.997 | 1.004
(0.9999-1.008) | -0.001771
(-0.01777-0.01423) |

Table 5: The Correlation and Estimated Bias of BC-5390 (vs XE-2100, Combined)

10

| Parameters | N | Result Range | r | slope
(95%CI) | intercept
(95%CI) |
|------------|------|--------------|-------|---------------------------|--------------------------------|
| HGB | 1486 | 3.30-23.1 | 0.998 | 1.010
(1.005-1.014) | -0.05500
(-0.1023-0.007685) |
| HCT | 1452 | 10.50-67.80 | 0.995 | 1.013
(1.006-1.019) | -0.5671
(-0.7991-0.3351) |
| MCV | 1452 | 53.70-127.10 | 0.993 | 1.036
(1.029-1.043) | -3.763
(-4.368--3.158) |
| MCH | 1409 | 13.90-41.20 | 0.977 | 1.020
(1.008-1.032) | -0.5451
(-0.8854-0.2048) |
| MCHC | 1408 | 23.80-37.70 | 0.825 | 1.184
(1.130-1.238) | -5.834
(-7.657-4.010) |
| RDW-CV | 1453 | 11.10-29.70 | 0.978 | 1.024
(1.004-1.044) | -0.1895
(-0.4807-0.1018) |
| RDW-SD | 1453 | 29.20-103.70 | 0.988 | 0.9878
(0.9760-0.9997) | -0.08935
(-0.6338-0.4551) |
| PLT | 1199 | 4.00-1879.00 | 0.996 | 1.019
(1.012-1.026) | -1.157
(-2.215-0.1002) |
| MPV | 1238 | 7.70-14.50 | 0.951 | 1.057
(1.037-1.078) | -0.7742
(-0.9964-0.5520) |

The WBC Morphology flagging rate for the BC-5390 was compared to the WBC manual differential for the same population of samples as shown in table below.

FlagsBC-5390
True Positive283
True Negative937
False Positive260
False Negative34
Total1514
Sensitivity (TP %)89.3%
Specificity (TN %)78.3%
Efficiency80.6%
Table 6: The WBC Morphology Flagging Ability of BC-5390 (vs Manual)
-----------------------------------------------------------------------------

Precision/Repeatability and Reproducibility

11

Repeatability study was conducted for each reported analyte parameter using whole blood samples collected in K2EDTA anticoagulant tubes either with normal reference intervals or around medical decision levels. Each of the samples was tested 10 times consecutively within the same day of sample collection. Three analyzers in one clinical site with at least two professional laboratory technicians were used to conduct the studies. The SD and CV of the different parameters of those samples were calculated. The results met the predefined performance acceptance specifications.

Reproducibility study was conducted to verify the potential performance variation derived from testing sites, operator, reagent lots. Multiple lots reagent (five lots of Diluent, two lots of LH Lyse and LEO(II) lyse, and three lots of LEO(I) lyse), three BC-5390 Analyzers and two operators at each sites were used to conduct the study. Three levels of samples were prepared using commercial control material BC-5D and tested in duplicate for each run, twice each day for 20 days. A total of 80 (2 duplicates, 2 runs and 20 days) replicates for each sample were tested at each clinical site and 240 replicates together for all three sites for each level control. Standard deviation and CV% were calculated for each measurand and the results obtained were within the specifications.

Linearity Range

WBC and PLT high-value analogs, from commercialized materials, were diluted to different values to cover the AMR of WBC and PLT respectively. RBC/HGB linearity was performed using dilutions prepared from fresh whole blood. The whole blood was concentrated to obtain specimens to test the high linearity limit. Proportional dilutions were prepared using the diluent for each parameter to create 7 subsequent dilutions. The expected values of the diluted samples were considered the true values. Each level of the samples was run in triplicate. The mean of three runs from each of the 7 dilutions across the linear range were used. Acceptable performance is indicated by the data fitting a linear regression line with a coefficient of determination (R2) of >0.95 and the parameters measured recovering within the bias limits for each parameters based on CLSI EP06-A. The results indicated that BC-5390 Auto Hematology Analyzer exhibits linearity across the claimed range.

Carrvover

Carryover performance was determined for parameters WBC, RBC, HGB, HCT and PLT. Testing was performed based on the different analytical cycle combinations of within mode and mode to mode. For each analytical cycle combination, whole blood and predilute sample with extremely elevated blood components (high sample) and with decreased blood components (low sample) were tested in triplicates in turn according to H26-A2.

For whole blood and predilute sampling, within mode to mode for sampling carryover was calculated and the results were within specifications (≤ 1.0%) for WBC, RBC, HGB, HCT and PLT. BC-5390 Analyzer demonstrated minimum carryover level within the defined specification for both Whole Blood and Predilute mode.

Interference

12

The potential interfering substances of BC-5390 were evaluated by measuring the whole blood samples and after adding different level of potential interfering substances. The paired difference testing was compared to the acceptance criteria. The impact of those substances were analyzed and the results demonstrated that there is no significant interference of bilirubin, WBC, and PLT for the results of BC-5390, but the high elevated concentration of Triglyceride (TG) and hemoglobin exhibits minor impact to the the HGB, MCH, MCHC parameters on the BC-5390. The information is disclosed in the BC-5390 Auto Hematology Analyzer Operator's Manual.

Other Compatibility Studies

Comparison of Whole Blood Mode and Predilute Mode

To demonstrate that BC-5390 analyzer performs equally on samples tested in different analysis modes (Whole blood and Predilute modes), 124 leftover whole blood samples representing the normal and medical conditions were collected in K2EDTA collection tube. All the samples were tested in duplicate on BC-5390 analyzer with Closed or Autoloader Whole blood CD mode and Closed Predilute CD mode. The results were analyzed according to CLSI EP09-A3. It showed that there was no difference when testing samples in either mode or when the sample contained parameters in normal or abnormal range due to medical conditions.

Comparison of CBC and CBC+DIFF Mode

To demonstrate that BC-5390 analyzer performs equally on samples tested in different analysis modes (CBC and CBC+DIFF modes), 103 leftover whole blood samples representing the normal and medical conditions were collected in K2EDTA collection tube for the comparison of Whole blood CBC and CD modes. Meanwhile, a total of 75 samples were collected in the same site for the comparison of Predilute CBC and CD modes. All the samples were tested in duplicate on BC-5390 analyzer with Closed or Autoloader Whole blood CD mode or Closed Predilute CD mode. The results were analyzed according to CLSI EP09-A3. It showed that there was no difference when testing samples in either mode or when the sample contained parameters in normal or abnormal range due to medical conditions.

Comparison of Capillary and Venous blood

To demonstrate the comparable performance between capillary and venous samples, 57 paired specimens were collected from donors by capillary method in K2EDTA microtainers and venipuncture method in K2EDTA collection tubes. Specimen levels were selected to cover the analytical measuring interval and medical decision levels for each parameter. Each sample was tested in the Closed whole blood CD mode at the U.S. site. The results were analyzed according to CLSI EP09-A3 and showed that the performance characteristics of the two specimen types were comparable.

13

Comparison of K2EDTA and K3EDTA Anticoagulants Samples

To evaluate whether samples collected in K2EDTA and K3EDTA have the same performance characteristics on BC-5390 analyzer, 70 paired fresh whole blood samples collected in K2EDTA and K3EDTA anticoagulant tubes were tested. The samples were selected with the targets to cover the analytical measuring interval and medical decision levels for each parameter. Each sample was tested in Autoloader/Closed whole blood CD mode on BC-5390 in the clinical laboratory in U.S. The results were analyzed according to CLSI EP09-A3 and showed that the performance characteristics of the specimens collected in the two anticoagulants are comparable.

Sample Stability

Specimen stability study was conducted to support the specimen stability claim as for whole blood is 36 hours at refrigerated temperature (2-8°C) and 24 hours when stored at 18-26°C, and for predilute sample the stability is 25 minutes when stored at 18-26°C before testing. Fresh samples covering normal and medical decision range were stored at at 2-8°C and then tested at defined intervals of immediate (0-1 hour), 4, 8, 12, 24, 26, 36 and 38 hours or stored at 18-26°C and then tested at the time points of immediate (0-1hour), 4, 8, 12, 14, 24 and 26 hours. Prediluted samples were also stored at 18-26°C and then tested at the time points of 5, 10, 15, 20, 25, 30, 35 minutes after sample preparation. The study results met the pre-defined acceptance criteria supporting the sample stability claims.

Reference Interval

A study was performed to assess the Adult Reference Ranges for BC-5390 analyzer using whole blood samples collected from 251 donors. In the study 121 adult male donors and 130 adult female donors between the ages of 19-70 were included. The non-parametric method and 95% confidence were used to calculate the lower and upper limits of the reference range according to EP28-A3c. The results are all shown in Table 7.

It is recommended that laboratories establish their own reference range based on the actual current patient population.

Partitions*MaleFemaleCombined
Parameters(Female vs
Male)NLow Limit
(90%CI)Upper Limit
(90%CI)NLow Limit
(90%CI)Upper Limit
(90%CI)NLow Limit
(90%CI)Upper Limit
(90%CI)
WBCYES1213.782
(3.560-4.000)12.049
(11.420-12.580)1304.374
(4.130-5.000)11.193
(10.380-11.850)2513.972
(3.780-4.330)11.726
(11.120-12.060)
Neu#NO1211.744
(1.460-2.050)9.728
(7.870-9.780)1302.229
(2.060-2.580)8.363
(7.230-8.690)2512.029
(1.740-2.190)8.578
(7.860-9.280)
Lym#NO1211.181
(1.150-1.400)4.268
(3.220-4.920)1301.288
(0.920-1.460)3.778
(3.410-4.110)2511.214
(1.150-1.400)3.858
(3.430-4.280)
Mon#YES1210.210
(0.150-0.250)0.869
(0.780-0.920)1300.211
(0.160-0.240)0.667
(0.630-0.710)2510.210
(0.160-0.240)0.787
(0.710-0.870)
Partitions*MaleFemaleCombined
Parameters(Female vsNLow LimitUpper LimitNLow LimitUpper LimitNLow LimitUpper Limit
Male)(90%CI)(90%CI)(90%CI)(90%CI)(90%CI)(90%CI)
Eos#NO1210.031
(0.000-0.050)0.599
(0.540-0.740)1300.023
(0.020-0.030)0.487
(0.420-1.090)2510.030
(0.020-0.040)0.557
(0.490-0.630)
Bas#YES1210.010
(0.000-0.010)0.100
(0.070-0.150)1300.010
(0.010-0.010)0.080
(0.060-0.120)2510.010
(0.010-0.010)0.080
(0.070-0.110)
Neu%NO12139.06
(37.80-42.30)78.14
(72.40-80.00)13041.94
(41.20-43.70)76.49
(74.60-83.60)25141.33
(39.00-42.40)76.90
(74.70-79.00)
Lym%NO12114.13
(11.80-18.30)50.47
(47.40-54.40)13016.14
(8.90-19.50)49.67
(48.20-52.30)25115.13
(12.00-18.30)49.84
(48.70-51.10)
Mon%YES1213.61
(2.60-4.00)10.90
(9.70-13.80)1303.63
(3.20-3.90)8.87
(7.60-9.30)2513.63
(3.40-3.90)9.77
(8.80-10.90)
Eos%NO1210.41
(0.00-0.60)8.82
(6.60-10.50)1300.40
(0.30-0.50)8.12
(6.00-13.90)2510.40
(0.30-0.50)8.10
(6.60-9.10)
Bas%YES1210.20
(0.10-0.20)1.20
(0.90-2.20)1300.20
(0.10-0.20)1.00
(0.90-1.20)2510.20
(0.10-0.20)1.10
(0.90-1.20)
RBCYES1214.231
(3.660-4.370)6.219
(6.000-6.380)1303.728
(3.540-4.010)5.454
(5.310-5.940)2513.907
(3.660-4.130)6.021
(5.920-6.220)
HGBYES12113.01
(11.10-13.20)18.28
(17.40-19.70)13010.18
(9.80-11.10)16.27
(15.90-16.50)25110.68
(10.10-11.50)17.47
(17.10-18.30)
HCTYES12138.50
(34.60-39.20)54.13
(51.40-56.70)13031.93
(31.70-34.30)47.30
(46.70-48.70)25133.81
(31.90-35.40)51.47
(50.60-54.20)
MCVNO12174.28
(68.20-79.70)100.68
(97.10-107.00)13074.33
(68.90-75.90)96.87
(94.00-98.80)25174.33
(71.10-76.20)97.64
(96.00-100.80)
MCHNO12124.96
(22.90-26.60)34.54
(32.80-36.50)13023.36
(22.40-25.10)32.40
(31.90-34.50)25123.90
(22.90-25.10)33.18
(32.40-34.60)
MCHCYES12132.01
(30.10-32.50)36.00
(35.80-36.20)13030.80
(29.40-31.40)35.87
(35.20-36.10)25131.02
(30.30-31.90)35.90
(35.70-36.10)
RDW-CVYES12112.01
(11.90-12.30)17.39
(15.80-18.50)13011.80
(11.50-12.10)18.63
(17.50-20.50)25112.00
(11.80-12.20)17.67
(17.20-18.90)
RDW-SDNO12136.74
(35.20-37.90)55.63
(51.30-58.30)13037.08
(34.50-39.10)56.00
(52.60-57.90)25137.09
(35.90-38.00)55.55
(52.60-57.70)
PLTYES121141.1
(120.0-154.0)456.6
(373.0-608.0)130186.4
(169.0-203.0)405.5
(385.0-481.0)251147.7
(141.0-171.0)435.4
(387.0-461.0)
MPVNO1219.30
(9.10-9.40)14.26
(13.00-15.90)1309.30
(9.20-9.40)13.59
(12.70-14.20)2519.30
(9.20-9.40)13.61
(12.90-14.30)

Table 7: The Reference Intervals of BC-5390

14

Note; * in this table denoted that if significantly difference exists between Male and Female according to CLSI EP28-A3c. If noted "Yes", it indicated that a separate reference interval should be considered for the Male and Female.

In addition, a literature reference interval verification study was performed in one clinical site for a total of 161 pediatric samples, including neonate (N=45), infant (N=26), child (N=57), and adolescent (N=33). These samples were collected from pediatric patients

15

ranging from neonates to 21 years of age. The results of this study verified the pediatric intervals published in Mayo and IOWA literature.

Please note, to substantiate a pediatric claim, we have included pediatric subjects in the method comparison and validated literature reference intervals according to FDA's recommendation in the pre-sub.

Determination of Limit of Blank (LoB), Limits of Detection (LoD) and Limit of Quantitation (LoQ)

Limit of Blank was determined using five blank samples (diluent) which just contain the dilute solution. To determine the Limits of Detection and Limit of Quantitation, five low levels of samples were created by adding whole blood samples to diluent to reach approximately 1-4 times of the LoB. Each blank sample or low level sample was tested 12 times at one clinical laboratory in the whole blood mode and predilute mode respectively on three BC-5390 analyzers. Each of the low level samples were tested 5 times on the Sysmex XE-2100 analyzer. The results were analyzed according to CLSI EP17-A. The maximum LoB, LoD, or LoQ value of the three (3) analyzers is taken as the reported value.

Substantial Equivalence

Table 8 and Table 9 summarize the similarities and differences between the BC-5390 Auto Hematology Analyzer and the predicate device XE-2100 Automated Hematology Analyzer.

Similarities
ItemSubject Device BC-5390Predicate XE-2100 (K992875)
Overview
layoutImage: BC-5390Image: XE-2100
Intended
UseThe BC-5390 Auto Hematology
Analyzer is a quantitative,
automated hematology analyzer
for in vitro diagnostic use in
clinical laboratories. The BC-
5390 Auto Hematology Analyzer
provides complete blood count
(WBC, RBC, HGB, HCT, MCV,The Sysmex XE-2100 is a multi-
parameter hematology analyzer
intended to classify the following
formed elements in anti-coagulated
blood.
WBC, Neut%/#, Lymph%/#,
Mono%/#, Eos%/#, Baso%/#,
NRBC%/# , RBC, HGB, HCT,
Similarities
MCH, MCHC, RDW-CV, RDW-
SD, PLT, MPV) and leukocyte 5-
Part differential (Neu#, Lym#,
Mon#, Eos#, Bas#, Neu%,
Lym%, Mon%, Eos%, Bas%) for
whole blood specimens, collected
in a salt of EDTA [dipotassium
(K2) or tripotassium (K3)]
obtained from venous or
capillary blood collection. The
purpose of the BC-5390 Auto
Hematology Analyzer is to
identify the normal human
patient, with normal system-
generated parameters, from
patients whose results require
additional studies.MCV, MCH, MCHC, RDW-CV,
RDW-SD, RET%/#, IRF, PLT, MPV
ParametersParameters(21):
WBC, Neu%, Lym%, Mon%,
Eos%, Bas%, Neu#, Lym#,
Mon#, Eos#, Bas#, RBC, HGB,
HCT, MCV, MCH, MCHC,
RDW-CV, RDW-SD, PLT, MPVSame
Test
Principle
(RBC/PLT)Impedance methodSame
Test
Principle
(HGB)Colorimetric methodSame
Sample
TypeWhole bloodSame
Sample
ProcessingUtilizes an automatic sampling,
diluting and mixing device for
sample processingSame
Similarities
Data
AnalysisRaw data analyzed by algorithms
to generate reported parametersSame
Data
ReportingDisplay, printing and
transmission of data to LIS/HIS.Same
Sample IDManual or Automatic barcode
scan of sample tube identifier or
manual keyboard entry of sample
identifierSame
Software
Risk LevelModerateSame
I/O
InterfacesOne LAN interfaceSame
SoftwareDMU (Data Management Unit)
Function: Managing data, such as
editing worklist, reviewing
results, generating reports,
communicating with LIS/HIS,
etc.IPU (Information Processing Unit)
Function: Same

Table 8: Device Comparison Table - Similarities

16

17

Table 9: Device Comparison Table - Differences

Differences
ItemDevice BC-5390Predicate XE-2100 (K992875)
Parameters
(Reticulocytes,
Nucleated
RBC)Not AvailableRET%/#, IRF, NRBC%/#
ReagentsM-53D DILUENT
M-5LEO(I) LYSE
M-5LEO(II) LYSE
M-53LH LYSE
PROBE CLEANSERCELLPACK, CELLSHEATH
STROMATOLYSER-4DS
STROMATOLYSER-4DL
STROMATOLYSER-FB, SULFOLYSER
CELLCLEAN
Differences
ItemDevice BC-5390Predicate XE-2100 (K992875)
Plus Reagents with IMI, RET, NRBC
(STROMATOLYSER-NR,
STROMATOLYSER-IM,
RET-SEARCH II)
Test PanelCBC
CBC+DIFFSame
Plus RET and NRBC
Test
Principle
(WBC)WBC: Flow Cytometry by
Laser and Electrical
Impedance MethodWBC: Flow Cytometry method using
semiconductor laser detection method
ThroughputClosed Whole Blood Mode:
Upper capacity of 51
samples/hour
Autoloader Whole Blood
Mode: Upper capacity of
60 samples/hour
Closed Predilute Mode:
Upper capacity of 53
samples/hourCBC, CBC+DIFF: Approx. 150
samples/hour
Mode with NRBC, RET: Approx. 113-150
samples/hour
Sample
VolumeAutoloader Whole Blood
Mode and Closed Whole
Blood Mode: 33μL
Closed Predilute Mode:
20μLClosed and Sampler Mode: Approx.
200μL
Manual Mode: Approx. 130μL
Capillary Mode: 40μL
Sample
assignment
mechanismUsing the sample
assignment Syringe to
assign the sample to the
baths in sequenceUsing the sample rotary valve (SRV) to assign
the sample to the baths simultaneously
Drive
Source UnitPumpPneumatic unit
Analysis
ModeClosed Whole Blood Mode
Autoloader Whole Blood
Mode
Closed Predilute ModeClosed Mode
Autoloader Mode
Predilute Mode and
Manual Mode
Sample
Anticoagul
antK2EDTA or K3EDTAK2EDTA or K3EDTA And EDTA-2Na
Differences
ItemDevice BC-5390Predicate XE-2100 (K992875)
ControlsBC-5D Hematology
Control - 3 Levels of High,
Normal and Lowe-CHECK – 3 Levels
CalibratorSC-CAL PLUS
Hematology CalibratorSCS-1000 Calibrator
Quality
Control
TechniquesL-J QC Program
X-B QC ProgramL-J QC Program
X-M QC Program
Data
StorageAnalysis data with
histograms and
scattergrams: 100,000
samples
Patient information:
100,000 persons
Order information: 4,000
samples
Quality control (QC) files: 60
filesAnalysis data with histograms and
scattergrams: 10,000 samples
Patient information: 5,000 persons
Order information: 1,000 samples
Quality control (QC) files: 20 files

18

19

Substantial Equivalence Conclusion

The BC-5390 Auto Hematology Analyzer has the same intended use as the predicate device, the XE-2100 as cleared in K992875. They are the quantitative, automated hematology analyzer and leukocyte differential counter for In Vitro Diagnostic Use in clinical laboratories. These analyzers may be used in adult and pediatric population. Both analyzers are only to be used by trained laboratory professionals to identify the normal patient, with all normal system-generated parameters, and to flag or identify patient results that require additional studies. The analyzer provides analysis results of 21 parameters of human blood, 3 histograms and 1 scattergram. The performance comparison study results demonstrated that the BC-5390 Auto Hematology Analyzer is substantially equivalent to the predicate device, the XE-2100 Automated Hematology Analyzer (K992875) and the differences do not raise new questions of the safety and effectiveness.