AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stat Profile Prime CCS Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of pH, PCO2, PO2, Hct, Na+. K-, CL-, iCa, Glu (Glucose), and Lac (Lactate) in heparinized whole blood.

PC02,P02,pH: Whole blood measurement of certain gases in whole blood,or pH of whole blood,is used in the diagnosis and treatment of life-threatening acid-base disturbances.

Hct: Whole blood measurements of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states,such as anemia and erythrocytosis (an increase in the number of red cells).

Na+: Sodium measurement is used in the diagnosis and treatment of aldosteronism,diabetes insipidus, adrenal hypertension, Addison's disease,dehydration,or diseases involving electrolyte imbalance.

K+: Potassium Measurement is used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high potassium levels.

Cl-: Chloride measurement is used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

iCa: Calcium measurements are used in the diagnosis and treatment of parathyroid disease,a variety of bone diseases,chronic renal disease and tetany (intermittent muscular contractions or spasms).

Glu: Glucose measurement is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus,neonatal hypoglycemia,and idiopathic hypoglycemia,and of pancreatic islet cell carcinoma.

Lac: Lactate (lactic acid) measurement is used to evaluate the acid-base status of patients lactic acidosis.

The Stat Profile Prime Auto QC Cartridge CCS is a quality control intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of the Stat Profile Prime CCS Analyzer.

The Stat Profile Prime Ampuled Control ABG/CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Stat Profile Prime CCS Analyzer.

The Stat Profile Prime Calibrator Cartridge CCS is intended for the calibration of pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, Glucose and Lactate using the Stat Profile Prime CCS Analyzer.

Linearity Standard Set A is intended for in vitro diagnostic use with Stat Profile Prime CCS Analyzers to verify calibration, analytical linearity, estimate test imprecision, and detect systematic analytical deviations that may arise from calibrator cartridge or analytical instrument variation.

Device Description

The Stat Profile Prime CCS Analyzer is a small, low cost blood gas, metabolite and electrolyte analyzer for laboratory use. The sensors and flow path have been integrated into one replaceable microsensor card, which is replaced periodically according to usage. The product, consumables, installation instructions and packaging are designed for easy customer installation.

Whole blood specimens are aspirated into the analyzer's microsensor card from syringes, tubes, or capillary blood collection devices using a peristaltic pump and a sampling probe. The disposable microsensor card contains the analytical flow path and all of the measurement sensors (pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, Glu (Glucose) and Lac (Lactate). Once the analysis measurement is complete, the whole blood specimen is automatically flushed out of the microsensor card flow path and into a self-contained waste collection bag contained within the disposable calibrator cartridge.

The Stat Profile Prime CCS Analyzer will have an enhanced test menu and multiple quality control options. Both traditional Internal and External liquid QC shall be offered, as well as an onboard Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly.

As with the predicate, the Stat Profile Prime CCS Analyzer is microprocessor-based and incorporates:

  • traditional sensor technology to measure blood pO2 ●
  • ion selective electrode technology to measure pH, pCO2, blood sodium, potassium, chloride, and ionized calcium
  • enzyme/Amperometric technology for glucose measurements

Liquid quality control materials are available as internal auto-cartridge quality control packs and as external ampules. The sampling, calibration and quality control functions are fully automated.

Internal Calibration standards with dissolved gases are provided in sealed pouches eliminating the need for users to calibrate the blood gas electrodes using external compressed gas cylinders. The Calibration Cartridges contain aqueous solutions within individual flexible bags housed in a cardboard box and a flexible waste bag. Each bag includes a fitment with septa that is pierced during the insertion of the cartridge into the analyzer. The Calibration Cartridge agueous solutions allow for 2 point calibration of each parameter as follows:

  • Calibrator A pH, PCO2, Na, K, Cl, iCa, Glu and Lactate (Volume > 500 mL) ●
  • Calibrator B pH, PO2, Na, K, Cl, iCa, Glu and Lactate (Volume > 250 mL) .
  • Calibrator F PCO2, PO2 (Volume > 720 mL) ●
  • Reference Solution KCI (Volume > 300 mL) .

The external glass ampule controls contain a buffered bicarbonate solution with a known pH and known levels of Na, K, Cl, iCa, Glucose (Glu) and Lactate. The solutions are equilibrated with known levels of 02, CO2, and N2. Each ampule contains 1.7 ml volume.

The internal auto QC cartridge consists of 3 flexible bags within a cardboard carton. Each bag contains an aqueous quality control material for monitoring the measurement of pH, PCO2, PO2, hematocrit (Hct) Na, K, Cl, iCa, Glucose (Glu) and Lactate. The aqueous quality control materials are composed of a buffered bicarbonate solution, each with a known pH and known level of Na, K, Cl. iCa. Glucose (Glu) and Lactate. Solutions are equilibrated with known levels of O2, CO2, and N2. Each bag contains a minimum volume of 100 mL. The aqueous quality control materials are formulated at three levels:

  • . Control 1: Acidosis, with High Electrolyte, Low Normal Glu, and Lactate
  • Control 2: Normal pH, Low-Normal Hct, Normal Electrolyte, High Glu, and Lactate ●
  • . Control 3: Alkalosis, High Hct, Low Electrolyte, High Abnormal Glu, and Lactate

Linearity Standard Set A consists of ampuled buffered solutions containing Ca++, Glu, Lactate, K+, and Cl-. Each ampule contains 1.8 ml volume.

The Stat Profile Analyzer accepts Lithium heparin whole blood sample from syringes, open tubes, small cups, and capillary tubes. The minimum sample size for both syringe and capillary samples analysis is 100 µL.

AI/ML Overview

Here's a summary of the acceptance criteria and the study details for the Stat Profile® Prime CCS Analyzer System, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state formal "acceptance criteria" with numerical thresholds for each parameter (e.g., a specific bias range for method comparison). Instead, it presents the results of equivalency studies against a predicate device and within the device itself (capillary vs. syringe, POC vs. trained healthcare professional). The implication is that the performance shown in these tables met the internal acceptance criteria for substantial equivalence to the predicate and for safe and effective use.

I will formulate a table focusing on the Method Comparison and Point-of-Care (POC) vs. Trained Healthcare Professional (THP) studies, as these directly compare the device's performance to established methods or user groups. The "Acceptance Criteria" column will represent the implicit expectation of strong correlation (high 'r' value, slope near 1, intercept near 0) for substantial equivalence and reliable performance.

Parameter (Analyte)Study TypeAcceptance Criteria (Implicit)Reported Device Performance (Slope, Intercept, r)
pHMethod Comparison (vs. Predicate)Slope ≈ 1, Intercept ≈ 0, r ≈ 1 (strong correlation)0.9976, 0.0099, 0.9985
Capillary vs. SyringeSlope ≈ 1, Intercept ≈ 0, r ≈ 11.0094, -0.0721, 0.9988
POC vs. THP (Syringe)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.983, 0.116, 0.997
POC vs. THP (Capillary)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.962, 0.275, 0.997
PCO2Method Comparison (vs. Predicate)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.9854, 0.9344, 0.9977
Capillary vs. SyringeSlope ≈ 1, Intercept ≈ 0, r ≈ 11.0026, -0.4347, 0.9989
POC vs. THP (Syringe)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.007, 0.750, 0.998
POC vs. THP (Capillary)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.989, 0.899, 0.998
PO2Method Comparison (vs. Predicate)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.9897, 1.4508, 0.9988
Capillary vs. SyringeSlope ≈ 1, Intercept ≈ 0, r ≈ 10.9942, 2.1791, 0.9996
POC vs. THP (Syringe)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.005, -0.094, 0.999
POC vs. THP (Capillary)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.979, 3.141, 0.999
HctMethod Comparison (vs. Predicate)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.0445, -1.9271, 0.9889
Capillary vs. SyringeSlope ≈ 1, Intercept ≈ 0, r ≈ 11.0013, 0.0485, 0.9963
POC vs. THP (Syringe)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.997, 0.395, 0.985
POC vs. THP (Capillary)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.978, 0.399, 0.984
Na+Method Comparison (vs. Predicate)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.0189, -2.2841, 0.9955
Capillary vs. SyringeSlope ≈ 1, Intercept ≈ 0, r ≈ 10.9995, -0.1711, 0.9978
POC vs. THP (Syringe)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.020, -2.540, 0.998
POC vs. THP (Capillary)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.010, -1.258, 0.997
K+Method Comparison (vs. Predicate)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.0163, -0.0371, 0.9996
Capillary vs. SyringeSlope ≈ 1, Intercept ≈ 0, r ≈ 10.9966, 0.0934, 0.9996
POC vs. THP (Syringe)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.974, 0.110, 0.999
POC vs. THP (Capillary)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.006, -0.025, 0.998
iCaMethod Comparison (vs. Predicate)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.9880, 0.0457, 0.9974
Capillary vs. SyringeSlope ≈ 1, Intercept ≈ 0, r ≈ 11.0228, -0.0603, 0.9855
POC vs. THP (Syringe)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.001, 0.004, 0.999
POC vs. THP (Capillary)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.977, 0.029, 0.996
Cl-Method Comparison (vs. Predicate)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.0003, 1.0158, 0.9955
Capillary vs. SyringeSlope ≈ 1, Intercept ≈ 0, r ≈ 10.9897, 0.1776, 0.9997
POC vs. THP (Syringe)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.000, -0.020, 0.999
POC vs. THP (Capillary)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.007, -0.710, 0.997
GluMethod Comparison (vs. Predicate)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.0007, -2.6844, 0.9892
Capillary vs. SyringeSlope ≈ 1, Intercept ≈ 0, r ≈ 10.9855, -0.4734, 0.9998
POC vs. THP (Syringe)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.989, 1.517, 0.998
POC vs. THP (Capillary)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.004, 0.036, 0.999
LacMethod Comparison (vs. Predicate)Slope ≈ 1, Intercept ≈ 0, r ≈ 10.9841, -0.0937, 0.9974
Capillary vs. SyringeSlope ≈ 1, Intercept ≈ 0, r ≈ 11.0034, -0.0120, 0.9994
POC vs. THP (Syringe)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.018, -0.093, 0.998
POC vs. THP (Capillary)Slope ≈ 1, Intercept ≈ 0, r ≈ 11.019, -0.127, 0.998
Lactate Limit of Detection (LoD)Total Error ≤ 0.3 (mmol/L)0.13 (mmol/L)
Lactate Linearity (r value for individual analyzer)r ≈ 10.9992, 0.9994, 0.9993 (for 3 analyzers)
GeneralInterference Testing (Lactate)Bias < ±10% for non-interfering substancesVarious substances tested with no significant interference. Glycolic acid (0.25 mmol/L) showed 11.7% bias, Hydroxyurea (0.2 mg/dL) showed 20.1% bias.
Total Imprecision (e.g., pH Level 1)Low SD and %CVpH: 0.007 SD, 0.10% CV
Total Imprecision (e.g., Lac Level 1)Low SD and %CVLac: 0.02 SD, 2.0% CV

2. Sample Sizes Used for the Test Set and Data Provenance

  • Method Comparison (vs. Predicate):
    • Sample Size: A minimum of 150 whole blood specimens for each parameter (ranging from 167 for pO2 to 186 for Cl).
    • Data Provenance: Not explicitly stated, but the samples were whole blood specimens. Given the context of seeking FDA clearance in the US, it's highly probable the data was collected within the United States. It used a predicate device (Stat Profile pHOx Ultra) for comparison, implying a retrospective comparison against results obtained from the predicate or concurrently with the predicate. The statement "Some samples were altered in order to achieve the hard-to-find sample range" suggests a controlled, laboratory-based study, likely prospective for the new device.
  • Capillary vs. Syringe Equivalency:
    • Sample Size: Approximately 100 whole blood samples for each parameter (ranging from 98 for iCa to 100 for others).
    • Data Provenance: Not explicitly stated, but implies a controlled, likely prospective study comparing two sampling modes on the same device.
  • Point-of-Care (POC) vs. Trained Healthcare Professional (THP):
    • Sample Size:
      • Syringe Mode: Ranged from 222 (Hct) to 234 (pH, PO2, iCa, Cl) specimens.
      • Capillary Mode: Ranged from 157 (Hct) to 173 (pH, PO2, iCa, Cl, Glu, Lac) specimens.
    • Data Provenance: The study was conducted in 3 POC sites (cardiovascular intensive care unit (CVICU), a medical intensive care unit (MICU), and a trauma/neuro intensive care unit). All testing used "discarded blood gas specimens." This implies a prospective study using real-world patient samples collected at these clinical sites.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Method Comparison: The predicate device, Stat Profile pHOx Ultra, was used as the comparative method. The "ground truth" for the new device's performance was established by comparing its results to the average of two results from the pHOx Ultra comparative method. This means the "experts" were the established and validated measurement capabilities of the predicate device, not human experts in interpreting results.
  • Capillary vs. Syringe: The "ground truth" for comparison was the syringe sample results from the same new device.
  • Point-of-Care (POC) vs. Trained Healthcare Professional (THP): The "ground truth" for comparison was the results obtained by "Trained Healthcare Professionals" (THP) also using the new device. The qualifications for THP are simply stated as "trained Healthcare Professionals." While the POC personnel included 43 respiratory therapy and 10 Nursing POC personnel from 3 POC settings, the specific qualifications of the "Trained Healthcare Professionals" are not detailed.

4. Adjudication Method for the Test Set

  • Method Comparison: The ground truth involved the average of 2 results from the predicate device (pHOx Ultra). If these two results were the reference method, it implies an internal comparison against a standard, not an adjudication process among human readers.
  • Capillary vs. Syringe & POC vs. THP: These studies involved direct comparison between two measurements from different methods/users. There's no mention of a separate adjudication process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This device is an automated in vitro diagnostic analyzer, not an AI-assisted diagnostic tool that interprets images or provides a diagnosis. Therefore, the concept of "human readers improving with AI vs. without AI assistance" does not apply here.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, this is essentially a standalone performance study for an automated in vitro diagnostic device. The device itself (its sensors and integrated system) performs the measurements. The comparison studies (Method Comparison, Capillary vs. Syringe) evaluate the analytical performance of the algorithm/device itself.
  • The POC study implicitly evaluates the human-in-the-loop aspect for correct operation of the device by different user groups, but the measurement itself is performed by the standalone instrument.

7. The Type of Ground Truth Used

  • Established Reference Method/Predicate Device: For the method comparison, the "ground truth" was established by the predicate device (Stat Profile pHOx Ultra), which is itself a legally marketed and presumably validated medical device. For Lactate linearity, the pHOx Ultra analyzers were used to establish the "target value" of each blood level.
  • Internal Device Consistency: For the capillary vs. syringe study, the syringe measurement mode on the new device served as the reference for the capillary mode.
  • Trained Professional Reference: For the Point-of-Care study, the measurements taken by "Trained Healthcare Professionals" using the new device served as the reference for measurements taken by POC personnel.

8. The Sample size for the training set

The document describes performance testing for substantial equivalence, not the development or training of an algorithm for the device. Therefore, a "training set" in the context of machine learning or AI is not applicable to this submission. The device relies on traditional sensor technology and measurement algorithms, which are engineered and validated rather than "trained" with data sets in the AI sense.

9. How the ground truth for the training set was established

As there is no explicit "training set" mentioned or implied for an AI algorithm, this question is not applicable to the information provided. The "ground truth" for the device's development and validation would have been established through a combination of engineering specifications, chemical/electrical standards, and established reference methods/materials for each analyte.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 29, 2015

NOVA BIOMEDICAL CORPORATION PAUL MACDONALD CHIEF QUALITY AND REGULATORY AFFAIRS OFFICER 200 PROSPECT ST WALTHAM MA 02454

Re: K141907 Trade/Device Name: Stat Profile® Prime CCS Analyzer System, Stat Profile Prime Auto QC Cartridge CCS, Stat Profile Prime Ampuled Control ABG/CCS, Stat Profile Prime Calibrator Cartridge CCS/CCS Comp and Nova Linearity Standard Set A Regulation Number: 21 CFR 862.1120 Regulation Name: Blood gases (PCO2, PO2) and blood pH test system Regulatory Class: II Product Code: CHL, JGS, CEM, JFP, CGZ, CGA, GKF, KHP, JIX, JJS Dated: April 24, 2015 Received: April 27, 2015

Dear Mr. Paul Macdonald:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{1}------------------------------------------------

as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Katherine Serrano -S

For :

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141907

Device Name

Stat Profile® Prime CCS Analyzer System, Stat Profile Prime Auto QC Cartridge CCS, Stat Profile Prime Ampuled Control ABG/ CCS,Stat Profile Prime Calibrator Cartridge CCS/CCS Comp, Nova Linearity Standard Set A

Indications for Use (Describe)

The Stat Profile Prime CCS Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of pH, PCO2, PO2, Hct, Na+. K-, CL-, iCa, Glu (Glucose), and Lac (Lactate) in heparinized whole blood.

PC02,P02,pH: Whole blood measurement of certain gases in whole blood,or pH of whole blood,is used in the diagnosis and treatment of life-threatening acid-base disturbances.

Hct: Whole blood measurements of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states,such as anemia and erythrocytosis (an increase in the number of red cells).

Na+: Sodium measurement is used in the diagnosis and treatment of aldosteronism,diabetes insipidus, adrenal hypertension, Addison's disease,dehydration,or diseases involving electrolyte imbalance.

K+: Potassium Measurement is used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high potassium levels.

Cl-: Chloride measurement is used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

iCa: Calcium measurements are used in the diagnosis and treatment of parathyroid disease,a variety of bone diseases,chronic renal disease and tetany (intermittent muscular contractions or spasms).

Glu: Glucose measurement is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus,neonatal hypoglycemia,and idiopathic hypoglycemia,and of pancreatic islet cell carcinoma.

Lac: Lactate (lactic acid) measurement is used to evaluate the acid-base status of patients lactic acidosis.

The Stat Profile Prime Auto QC Cartridge CCS is a quality control intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of the Stat Profile Prime CCS Analyzer.

The Stat Profile Prime Ampuled Control ABG/CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Stat Profile Prime CCS Analyzer.

The Stat Profile Prime Calibrator Cartridge CCS is intended for the calibration of pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, Glucose and Lactate using the Stat Profile Prime CCS Analyzer.

Linearity Standard Set A is intended for in vitro diagnostic use with Stat Profile Prime CCS Analyzers to verify calibration, analytical linearity, estimate test imprecision, and detect systematic analytical deviations that may arise from calibrator cartridge or analytical instrument variation.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
X Prescription Use (Part 21 CER 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

510(k) Number:K141907
510(k) Owner:Nova Biomedical Corporation
Registration Number:1219029
Address:200 Prospect St.
Waltham, MA 02454
Phone:781-894-0800
Fax Number:784-891-4806
Contact Person:Paul W. MacDonald
Date Prepared:22 May, 2015

Proprietary Name: Stat Profile® Prime CCS Analyzer System, Stat Profile Prime Auto QC cartridges CCS, Ampuled Control ABG/CCS, Calibrator Cartridge CCS/CCS Comp, and the Nova Linearity Standard Set A

Common or Usual Name: Blood Gas/Electrolyte/Metabolite/CO-Oximetry Analyzer

Classification Name: Multiple

Classification Names:Class No.Reg. No.Class
Blood Gases and Blood pH system75CHL862.1120II
Sodium Test System75JGS862.1665II
Potassium Test System75CEM862.1600II
Calcium Test System75JFP862.1145II
Chloride Test System75CGZ862.1170II
Glucose Test System75CGA862.1345II
Instrument, Hematocrit, Automated81GKF864.5600II
Lactic Acid Test System75KHP862.1450I
Calibrators75JIX862.1150II
Quality Control Materials75JJS862.1660I

Product Codes: CHL, JGS, CEM, JFP, CGZ, CGA, GKF, JIX, JJS, KHP

Predicate Device: K131703 - Nova Stat Profile Prime CCS Analyzer System (including controls, calibrators and linearity standards)

Device Description:

The Stat Profile Prime CCS Analyzer is a small, low cost blood gas, metabolite and electrolyte analyzer for laboratory use. The sensors and flow path have been integrated into one replaceable microsensor card, which is replaced periodically according to usage. The product, consumables, installation instructions and packaging are designed for easy customer installation.

Whole blood specimens are aspirated into the analyzer's microsensor card from syringes, tubes, or capillary blood collection devices using a peristaltic pump and a sampling probe. The disposable microsensor card contains the analytical flow path and all of the measurement sensors (pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, Glu (Glucose) and Lac (Lactate). Once the analysis measurement is complete, the whole blood specimen is automatically flushed out of the microsensor card flow path and into a self-contained waste collection bag contained within the disposable calibrator cartridge.

{5}------------------------------------------------

The Stat Profile Prime CCS Analyzer will have an enhanced test menu and multiple quality control options. Both traditional Internal and External liquid QC shall be offered, as well as an onboard Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly.

As with the predicate, the Stat Profile Prime CCS Analyzer is microprocessor-based and incorporates:

  • traditional sensor technology to measure blood pO2 ●
  • ion selective electrode technology to measure pH, pCO2, blood sodium, potassium, chloride, and ionized calcium
  • enzyme/Amperometric technology for glucose measurements

Liquid quality control materials are available as internal auto-cartridge quality control packs and as external ampules. The sampling, calibration and quality control functions are fully automated.

Internal Calibration standards with dissolved gases are provided in sealed pouches eliminating the need for users to calibrate the blood gas electrodes using external compressed gas cylinders. The Calibration Cartridges contain aqueous solutions within individual flexible bags housed in a cardboard box and a flexible waste bag. Each bag includes a fitment with septa that is pierced during the insertion of the cartridge into the analyzer. The Calibration Cartridge agueous solutions allow for 2 point calibration of each parameter as follows:

  • Calibrator A pH, PCO2, Na, K, Cl, iCa, Glu and Lactate (Volume > 500 mL) ●
  • Calibrator B pH, PO2, Na, K, Cl, iCa, Glu and Lactate (Volume > 250 mL) .
  • Calibrator F PCO2, PO2 (Volume > 720 mL) ●
  • Reference Solution KCI (Volume > 300 mL) .

The external glass ampule controls contain a buffered bicarbonate solution with a known pH and known levels of Na, K, Cl, iCa, Glucose (Glu) and Lactate. The solutions are equilibrated with known levels of 02, CO2, and N2. Each ampule contains 1.7 ml volume.

The internal auto QC cartridge consists of 3 flexible bags within a cardboard carton. Each bag contains an aqueous quality control material for monitoring the measurement of pH, PCO2, PO2, hematocrit (Hct) Na, K, Cl, iCa, Glucose (Glu) and Lactate. The aqueous quality control materials are composed of a buffered bicarbonate solution, each with a known pH and known level of Na, K, Cl. iCa. Glucose (Glu) and Lactate. Solutions are equilibrated with known levels of O2, CO2, and N2. Each bag contains a minimum volume of 100 mL. The aqueous quality control materials are formulated at three levels:

  • . Control 1: Acidosis, with High Electrolyte, Low Normal Glu, and Lactate
  • Control 2: Normal pH, Low-Normal Hct, Normal Electrolyte, High Glu, and Lactate ●
  • . Control 3: Alkalosis, High Hct, Low Electrolyte, High Abnormal Glu, and Lactate

Linearity Standard Set A consists of ampuled buffered solutions containing Ca++, Glu, Lactate, K+, and Cl-. Each ampule contains 1.8 ml volume.

The Stat Profile Analyzer accepts Lithium heparin whole blood sample from syringes, open tubes, small cups, and capillary tubes. The minimum sample size for both syringe and capillary samples analysis is 100 µL.

Measured Parameters:

The Stat Profile Prime CCS Analyzer measures pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, Glu and Lactate (Note: Glucose and Lactate are optional).

{6}------------------------------------------------

Calculated Parameters:

  • pH, PCO2, PO2 (corrected to patient temperature) ●
  • Bicarbonate level (HCO3-) ●
  • . Total Carbon Dioxide (TCO2)
  • Base Excess of the blood (BE-b)
  • Base Excess of extracellular fluid (BE-ecf) .
  • Standard Bicarbonate Concentration (SBC) ●
  • Oxygen Content (O2Ct) ●
  • . Oxygen Capacity (O₂Cap)
  • Alveolar Oxygen (A) .
  • Arterial Alveolar Oxygen Tension Gradient (AaDO2) ●
  • Arterial Alveolar Oxygen Tension Ratio (a/A) .
  • Respiratory Index (RI)
  • . P50
  • PO2/FIO2 ratio .
  • Oxygen Saturation (SO2%) .
  • Hemoqlobin
  • . Anion Gap
  • Normalized Calcium, nCa ●

Intended Use:

The Stat Profile Prime CCS Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and for point-of-care usage for the quantitative determination of pH, PCO2, PO2, Hct, Na*, K*, Clr, iCa, Glu (Glucose), and Lac (Lactate) in heparinized whole blood.

PCO2,PO2,pHWhole blood measurement of certain gases in whole blood, or pH of whole blood, isused in the diagnosis and treatment of life-threatening acid-base disturbances.
HctWhole blood measurements of the packed red cell volume of a blood sample are usedto distinguish normal from abnormal states, such as anemia and erythrocytosis (anincrease in the number of red cells).
Na+Sodium measurement is used in the diagnosis and treatment of aldosteronism, diabetesinsipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involvingelectrolyte imbalance.
K+Potassium Measurement is used to monitor electrolyte balance in the diagnosis andtreatment of disease conditions characterized by low or high potassium levels.
Cl-Chloride measurement is used in the diagnosis and treatment of electrolyte andmetabolic disorders such as cystic fibrosis and diabetic acidosis.
iCaCalcium measurements are used in the diagnosis and treatment of parathyroid disease,a variety of bone diseases, chronic renal disease and tetany (intermittent muscularcontractions or spasms).
GluGlucose measurement is used in the diagnosis and treatment of carbohydratemetabolism disturbances including diabetes mellitus, neonatal hypoglycemia, andidiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
LacLactate (lactic acid) measurement is used to evaluate the acid-base status of patientssuspected of having lactic acidosis.

Stat Profile Prime Auto QC Cartridge CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of the Stat Profile Prime CCS Analyzer.

{7}------------------------------------------------

Stat Profile Prime Ampuled Control ABG/CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Stat Profile Prime CCS Analyzer.

Stat Profile Prime Calibrator Cartridge CCS is intended for the calibration of pH, PCO2, PO2, Hct, Na*, K*, Cl*, iCa, Glucose, and Lactate, using the Stat Profile Prime CCS Analyzer.

Linearity Standard Set A is intended for in vitro diagnostic use with Stat Profile Prime CCS Analyzers to verify calibration, analytical linearity, estimate test imprecision, and detect systematic analytical deviations that may arise from calibrator cartridge or analytical instrument variation.

Summary of the Technological Characteristics:

The Stat Profile Prime CCS Analyzer is substantially equivalent to the previously cleared for market Stat Profile Prime CCS Analyzer System in intended use. It uses the same sensor technology and measurement algorithms, and the formulations of the internal and external controls and the calibration cartridge are the same for the tested parameters. The linearity standards for use with the Stat Profile Prime CCS Analyzer are identical to those cleared for use with the predicate Stat Profile Prime CCS Analyzer System.

Summary of Performance Testing:

Testing was completed to demonstrate that the Stat Profile Prime CCS Analyzer is substantially equivalent in performance, safety and efficacy to the Stat Profile pHOx Ultra Analyzer System.

In this submission, only Lactate data for detection limit. Iinearity, and interference/specificity was reviewed and cleared. Performance data for all other analytes was presented and reviewed in the predicate device submission (K131703 - Stat Profile Prime CCS Analyzer).

Method Comparison

Method comparison studies were performed. A minimum of 150 whole blood specimens were analyzed for each parameter in svringe collection devices. Some samples were altered in order to achieve the hard-to-find sample range. The samples were analyzed on each of the Stat Profile Prime CCS analyzers and on each of the pHOx Ultra analyzers. The Stat Profile Prime CCS results for each analyzer were compared to the average of the 2 results from the pHOx Ultra comparative method.

Table 1: Syringe method comparison study results vs. the predicate device (Stat Profile pHOx Ultra)

TestParameterTotal #samplesSample range testedSlopeInterceptr
pH1726.523 - 7.862 pH units0.99760.00990.9985
pCO21793.4 - 200.0 mmHg0.98540.93440.9977
pO216729.5 - 593.2 mmHg0.98971.45080.9988
Hct17412 - 70%1.0445-1.92710.9889
Na18085.5 - 195.7 mmol/L1.0189-2.28410.9955
K1791.11 - 19.75 mmol/L1.0163-0.03710.9996
iCa1810.25 - 2.48 mmol/L0.98800.04570.9974
Cl18652.8 - 189.3 mmol/L1.00031.01580.9955
Glu18539 - 474 mg/dL1.0007-2.68440.9892
Lac1820.4 - 17.80.9841-0.09370.9974

{8}------------------------------------------------

Whole blood patient samples were evaluated to demonstrate that syringe sample mode and capillary sample mode are equivalent. The syringe sample was analyzed directly using the syringe mode. After the measurement, samples were transferred from the syringe into the capillary tube and then analyzed using the capillary mode. Approximately 100 whole blood samples were analyzed for each parameter. The capillary test result was compared to the syringe test result and linear regression analysis was performed. Results are summarized in the table below.

Parametertotal #samplessample rangetestedSlopeInterceptr
pHpH units1006.787 - 7.6831.0094-0.07210.9988
pCO2mmHg10017.7 - 111.01.0026-0.43470.9989
pO2mmHg10025.5 - 435.20.99422.17910.9996
Hct%10014 - 691.00130.04850.9963
Nammol/L10085.0 - 198.10.9995-0.17110.9978
Kmmol/L1002.70 - 19.370.99660.09340.9996
iCammol/L980.33 - 2.511.0228-0.06030.9855
Clmmol/L10055.8 - 197.10.98970.17760.9997
Glumg/dL10017 - 4880.9855-0.47340.9998
Lacmmol/L1001.1 - 18.11.0034-0.01200.9994

Table 2: Method Comparison Results - Capillary vs. Syringe

Specificity / Interference Testing Protocol:

Li-heparin whole blood samples were used in the interference study. For the initial testing, each interferent was spiked at 20 times the recommended concentration by CLSI EP-7 A2. Each sample containing interferent was evaluated against the same whole blood sample without the interferent. If interference was observed, a dose response study was performed at two analytes concentrations (low and high) to determine the highest substance concentration where no interference was observed. The sponsor's definition of significant interference is ±10% bias for Lactate. The following table represents substances that were tested without demonstrating a significant interference on test results:

{9}------------------------------------------------

Interfering SubstanceHighest Concentration Tested
Acetaminophen20.0 mg/dL
Acetoacetate2.0 mmol/L
Acetylsalicylic acid3.62 mmol/L
Ammonium Chloride107.0 µmol/L
Ascorbic Acid50 mg/dL
Benzylkonium Chloride10.0 mg/L
BetaHydroxybutyrate2.0 mmol/L
Bilirubin20.0 mg/dL
D-Galactose1.0 mmol/L
Dobutamine2.0 mg/dL
Dopamine Hydrochloride5.87 µmol/L
EDTA3.4 µmol/L
Ethanol86.8 mmol/L
Fluoride105 µmol/L
Glucose1,000 mg/dL
Glucosamine30.0 µmol/L
Hemoglobin2.0 g/L
Heparin100 IU/mL
Ibuprofen2.4 mmol/L
Intralipid10.0 mg/mL
Maltose13.0 mmol/L
Mannose1.0 mmol/L
Potassium Thiocyanate2,064 µmol/L
Pyruvate309 µmol/L
Salicylic Acid4.34 mmol/L
Sodium Bromide37.5 mmol/L
Sodium Citrate12.0 mmol/L
Sodium Oxalate500 mg/dL
Thiocyanate6.8 mmol/L
Urea40.0 mg/dL
Uric Acid1.4 mmol/L
Xylose25.0 mg/dL

Table 3: Substances Tested for Lactate Interference

{10}------------------------------------------------

The following table represents substances that were tested that demonstrated a significant interference on test results:

ParameterInterferingSubstanceConcentration ofinterfering substanceInterference
LactateGlycolic acid0.0 mmol/LNo interference observed
0.25 mmol/LBias of 11.7%
Hydroxyurea0.0 mg/dLNo interference observed
0.2 mg/dLBias of 20.1%

Table 4: Lactate Interfering Substances

Determination of Limit of Detection: Limit of Detection was determined by obtaining the standard deviation of sample measurements from repeated measurements of samples with a relevant low concentration. Three blood samples with low level lactate concentration were used to estimate the limit of detection.

Table 5: Lactate Limit of Detection

Limit of Detection
LoBLoDLoQAverageTotal ErrorAcceptanceCriteria forTotal ErrorClaimedMeasurementRange
Lac2.70 mg/dL0.30 mmol/L3.60 mg/dL0.40 mmol/L3.60 mg/dL0.40 mmol/L0.13≤ 0.33.6 - 178.6 mg/dL0.4 - 20.0 mmol/L

Lactate Linearity: Samples were prepared by tonometering, spiking or diluting whole blood to span the analytical measurement range for lactate. Each blood level sample was analyzed in triplicate on each of the three (3) test analyzers and on the pHOx Ultra analyzers. The pHOx Ultra analyzers were used to establish the target value of each blood level for lactate.

Table 6: Lactate Linearity Results

Individual Analyzer Performance Data
Parametertotal #levelsSpecimenrangeAnalyzerSlopeInterceptr
Lac110.2 - 24.9#10.99460.11780.9992
0.2 - 24.7#20.99200.14310.9994
0.2 - 24.8#30.99110.21170.9993

{11}------------------------------------------------

Point-of-Care

A Point-of-Care study was conducted to demonstrate that the system was safe and effective for use in the POC setting. The testing compared results obtained by trained Healthcare Professionals to results obtained by POC personnel on the same specimens using the same analyzer. The Stat Profile Prime CCS Analyzer was evaluated by point-of-care (POC) personnel in 3 POC sites including a cardiovascular intensive care unit (CVICU), a medical intensive care unit (MICU) and a trauma/neuro intensive care unit. A total of 43 respiratory therapy and 10 Nursing POC personnel participated from the 3 POC settings over the study. The personnel represent trained, qualified staff found in typical POC sites where blood gas analyzers are utilized. All testing was performed using discarded blood gas specimens.

Combined data from all 3 POC settings is summarized in Tables 7 and 8.

ParameterTotal #specimensWhole BloodRangeSlopeInterceptr
рн2346.874 - 7.6650.9830.1160.997
PCO2 mmHg2304.1 - 195.51.0070.7500.998
PO2 mmHg23415.2 - 714.51.005-0.0940.999
Hct %22212 - 700.9970.3950.985
Na mmol/L22983.2 - 192.31.020-2.5400.998
K mmol/L2311.10 - 18.800.9740.1100.999
iCa mmol/L2340.26 - 2.551.0010.0040.999
Cl mmol/L23453.0 - 188.71.000-0.0200.999
Glu mg/dL23317 - 4780.9891.5170.998
Lac mmol/L2330.6 - 19.51.018-0.0930.998

Table 7: Prime CCS: POC v Trained Healthcare Professional (THP) - Svringe Mode

Table 8: Prime CCS: POC v Trained Healthcare Professional (THP) - Capillary Mode

ParameterTotal #specimensWhole BloodRangeSlopeInterceptr
pH1736.881 - 7.7800.9620.2750.997
PCO2 mmHg1703.2 - 181.40.9890.8990.998
PO2 mmHg17322.8 - 597.30.9793.1410.999
Hct %15713 - 680.9780.3990.984
Na mmol/L16983.2 - 197.01.010-1.2580.997
K mmol/L1681.15 - 19.471.006-0.0250.998
iCa mmol/L1730.32 - 2.450.9770.0290.996
Cl mmol/L17355.9 - 188.11.007-0.7100.997
Glu mg/dL17315 - 4841.0040.0360.999
Lac mmol/L1730.6 - 18.41.019-0.1270.998

Total Imprecision Performance

The total imprecision data included in the following table was obtained from different POC site personnel running 3 levels of Stat Profile Prime External Quality Control material (Levels 1-3) in duplicate each day for a total of 20 runs on 3 Stat Profile Prime CCS analyzers. The protocol was based upon methods described in CLSI "Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second edition," CLSI EP5-A2. The test data is representative of the expected total imprecision between analyzer performance obtainable by POC personnel using the Stat Profile Prime CCS analyzer using external quality control materials.

{12}------------------------------------------------

Combined Total Imprecision Data - Level 1
ParameterPooledMeanNWithin Run SD (Sr)Within Run %CVTotal Imprecision SD(St)Total Imprecision %CV
pH7.1441200.0040.060.0070.10
PCO264.61200.81.31.52.3
PO256.21201.83.22.64.7
Hct331200.51.60.61.9
Na163.61200.60.41.10.7
K5.691200.050.90.081.4
Cl128.21200.30.21.31.0
iCa1.561200.0070.50.0231.5
Glu81.01201.11.41.41.8
Lac1.01200.022.00.022.0
Combined Total Imprecision Data - Level 2
pH7.3681200.0020.030.0050.07
PCO242.51200.41.00.82.0
PO298.01200.91.02.22.3
Hct511200.50.90.61.2
Na139.71200.50.31.00.7
K3.731200.020.50.051.3
Cl102.11200.20.20.40.4
iCa1.001200.0050.50.0090.9
Glu203.01202.21.14.92.4
Lac2.81200.051.80.062.1
Combined Total Imprecision Data - Level 3
pH7.5621200.0080.110.0090.12
PCO225.11200.72.71.35.2
PO2140.81201.20.93.22.3
Hct651200.50.80.71.0
Na117.61200.30.20.50.4
K1.871200.010.50.021.1
Cl87.01200.40.50.91.1
iCa0.551200.0040.70.0071.3
Glu318.01204.21.39.02.8
Lac7.11200.10.70.121.7
Combined Total Imprecision Data - Low Hematocrit
Hct191200.42.20.52.6

Table 9: Prime CCS: Total Imprecision Results – Combined – External Controls

{13}------------------------------------------------

Table 10: Comparison of Predicate and Proposed devices
------------------------------------------------------------------------------
CharacteristicPredicate: K131703 Stat Profile Prime CCS AnalyzerProposed: Stat Profile Prime CCS Analyzer
Indication For UseThe Stat Profile Prime CCS Analyzer System is intended for in vitrodiagnostic use by health care professionals in clinical laboratorysettings for the quantitative determination of pH, PCO2, PO2, Hct,Na+, K+, Cl-, iCa, and Glu (Glucose), in heparinized whole blood.The Stat Profile Prime CCS Analyzer System isintended for in vitro diagnostic use by health careprofessionals in clinical laboratory settings and forpoint-of-care usage for the quantitative determination ofpH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, Glu (Glucose),and Lac (Lactate) in heparinized whole blood.
PCO2,PO2,pHWhole blood measurement of blood gases is used in thediagnosis and treatment of life-threatening acid-basedisturbances in critically ill patients with numerous metabolicand pulmonary diseases.Same
HctWhole blood measurements of hematocrit are used toestimate that red blood cells are present in sufficientquantity to carry oxygen and carbon dioxide.Same
Na+Sodium measurement is used in the diagnosis andtreatment of aldosteronism, diabetes insipidus, adrenalhypertension, Addison's disease, dehydration, or diseasesinvolving electrolyte imbalance.Same
K+Potassium Measurement is used to monitor electrolytebalance in the diagnosis and treatment of diseaseconditions characterized by low or high potassium levels.Same
Cl-Chloride measurement is used in the diagnosis andtreatment of electrolyte and metabolic disorders such ascystic fibrosis and diabetic acidosis.Same
iCaIonized Calcium measurement is used in the diagnosis andtreatment of hypertension, renal disease, and vitamin Drelated disorders. Also useful in the diagnosis and treatmentof patients with increased total protein and/or albuminlevels, as in dehydration.Same
CharacteristicPredicate: K131703 Stat Profile Prime CCS AnalyzerProposed: Stat Profile Prime CCS Analyzer
GluGlucose measurement is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.Same
LacLactate (lactic acid) measurement is used to evaluate the acid-base status of patients suspected of having lactic acidosis.Same
Acceptable SamplesLithium heparinized whole blood from syringes, open tubes, small cups, and capillary tubes.Same
Sample Volumes100µL (syringe and capillary)Same
Measurement Range
pH6.500-8.000Same
PCO23.0 -200 mmHgSame
PO25-765 mmHgSame
Hct12%-70%Same
Na+80-200 mmol/LSame
K+1.0-20.0 mmol/LSame
Cl-50-200 mmol/LSame
iCa (Ca++)0.20-2.70 mmol/LSame
Glu15-500 mg/dLSame
Lac0.3-20.0 mmol/LSame
Principles of Measurement
pHHydrogen ion-selective sensorSame
CharacteristicPredicate: K131703 Stat Profile Prime CCS AnalyzerProposed: Stat Profile Prime CCS Analyzer
PCO2Severinghaus-type sensorSame
PO2Polarographic Clark-type sensorSame
HctImpedance sensorSame
Na+Sodium ion-selective sensorSame
K+Potassium ion-selective sensorSame
Cl-Chloride ion-selective sensorSame
iCa (Ca++)Calcium ion-selective sensorSame
GluGlucose Oxidase Enzymatic sensorSame
LacLactate Oxidase Enzymatic sensorSame
Touch Screen5.7" VGA full color display with LED backlight and integrated touchpanelSame
MenuFully configurable test menu based on above sensorsSame
Bar Code ScannerInternal Integrated 1D/2DSame
Printer2" Roll, Thermal TransferSame
PumpPeristaltic Pump w/ Pressure Plate, TPE Tubing (Pharmed BPT)Same
Analog BoardPrecision low level analog front end w/ amperometric andpotentiometric amplifiers, air detector circuitry and temperaturecontrol circuitrySame

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

CharacteristicPredicate: K131703 Stat Profile Prime Calibrator Cartridge CCSProposed: Stat Profile Prime CalibratorCartridge CCS
Indication For UseThe Stat Profile Prime Calibrator Cartridge CCS is intended for thecalibration of pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, and Glucose using theStat Profile Prime CCS Analyzer.The Stat Profile Prime Calibrator Cartridge CCSis intended for the calibration of pH, PCO2, PO2,Hct, Na+, K+, Cl-, iCa, Glucose (Glu), and Lactate(Lac) using the Stat Profile Prime CCS Analyzer.
SettingsClinical LaboratoriesClinical Laboratories and/or Point Of Care
Configuration2 level calibration standards per analyte, and reference solutionSame
PackagingLiquid in Mylar bags inside cardboard container. Includes a wastecollection bag. Self-contained, disposable packaging.Same
CharacteristicPredicate: K131703 Stat Profile Prime Auto QC Cartridge CCSProposed: Stat Profile Prime Auto QC Cartridge CCS
Indication For UseThe Stat Profile Prime Auto QC Cartridge CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of the Stat Profile Prime CCS Analyzer.Same
SettingsClinical LaboratoriesClinical Laboratories and/or Point Of Care
Configuration3 level aqueous electrolyte, metabolite and gas solutions.Same
PackagingCartridge: Solution in Mylar bags inside cardboard container. Includes a waste collection bag. Self-contained, disposable packaging.Same
CharacteristicPredicate: K131703 Stat Profile Prime Ampuled Control ABG/CCSProposed: Stat Profile Prime AmpuledControl ABG/CCS
Indication For UseThe Stat Profile Prime Ampuled Control ABG/CCS is a quality controlmaterial intended for in vitro diagnostic use by healthcare professionals formonitoring the performance of Stat Profile Prime CCS Analyzer.Same
SettingsClinical LaboratoriesClinical Laboratories and/or Point Of Care
Configuration3 level aqueous electrolyte, metabolite and gas solutions.Same
PackagingAmpules: Each glass ampule contains 1.7 ml volume.Same

{17}------------------------------------------------

CharacteristicPredicate: K131703 Linearity Standard Set AProposed: Linearity Standard Set A
Indication For UseThe Linearity Standard Set A is intended for in vitro diagnostic use withStat Profile Prime CCS Analyzers to verify calibration, analytical linearity,estimate test imprecision, and detect systematic analytical deviations thatmay arise from calibrator cartridge or analytical instrument variation.Same
SettingsClinical LaboratoriesClinical Laboratories and/or Point Of Care
Configuration4 level aqueous solutions in glass ampules. Contain electrolyte, metaboliteand gas solutions.Same
PackagingAmpules: Each glass ampule contains 1.8 ml volumeSame

Conclusion:

The results of software validation and perfication testing confirmed that the Stat Profile Prime CCS Analyzer is safe and effective for its intended purpose and that the Stat Profile Prime CCS Analyzer System is substantially equivalent to the Profile Prine CCS Analyzer System (K131703).

§ 862.1120 Blood gases (P

CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.