(118 days)
The SFI-Bar® is intended to be used with the implant manufacturer's implant (Table 1) to provide support for fixation of overdentures.
The SFI-Bar® provides the connection between compatible dental implant systems for the fixation of removable overdentures. The SFI-Bar@ consists of an implant adapter (abutinent) and a stress-free bar for the fixation of removable overdentures. The implant adapter is screwed into the dental implant.
Here's a breakdown of the acceptance criteria and study information for the SFI-Bar® based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Performance Characteristics) | Reported Device Performance |
|---|---|
| Torque tests | Met acceptance criteria |
| Application testing | Met acceptance criteria |
| Functional testing | Met acceptance criteria |
| Performance equivalence to predicate SFI-Bar® | Equivalent |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a separate "test set" and its sample size in the context of clinical data for the SFI-Bar®. The performance evaluation relies on non-clinical test data.
Data Provenance: The testing was conducted as part of a 510(k) submission, originating from Switzerland (Cendres + Métaux SA, Biel/Bienne, Switzerland). The data is retrospective in the sense that it's an evaluation against established performance metrics and comparison to predicate devices, rather than a prospective clinical trial.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided as the ground truth for this device's performance evaluation was based on engineering and performance testing rather than expert-derived clinical ground truth.
4. Adjudication Method for the Test Set
This information is not applicable as the evaluation did not involve human interpretation or adjudication processes like those in clinical image analysis studies. The adjudication for performance was based on meeting technical specifications from engineering tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices, which is not the nature of the SFI-Bar®.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This concept is not applicable to the SFI-Bar®. The device is a physical dental abutment, not an algorithm. The performance evaluation was of the physical device's mechanical properties and function.
7. The Type of Ground Truth Used
The ground truth for evaluating the SFI-Bar®'s performance was based on:
- Engineering specifications and standards: Adherence to established mechanical properties (e.g., torque, application, functional requirements).
- Performance characteristics of predicate devices: The SFI-Bar® was compared for substantial equivalence to legally marketed predicate devices, meaning their established safe and effective performance served as a benchmark.
8. The Sample Size for the Training Set
This information is not applicable. The SFI-Bar® is a physical medical device, not an AI/ML algorithm that requires a "training set."
9. How the Ground Truth for the Training Set was Established
This information is not applicable as there is no "training set" for this device.
{0}------------------------------------------------
KBIS26
Image /page/0/Picture/1 description: The image shows a logo with the text "CENDRES+ MÉTAUX". To the left of the text is a stylized symbol that appears to be a combination of the letters 'C' and 'M'. The text is in a bold, sans-serif font, with "CENDRES" on top and "MÉTAUX" below it. The overall design is simple and modern.
May 23, 2013
Image /page/0/Picture/4 description: The image shows a logo with the letters "CM" on the left and the words "CENDRES" and "MÉTAUX" on the right. The letters "CM" are in a bold, sans-serif font and are stacked on top of each other. The word "CENDRES" is in a smaller, sans-serif font and is stacked on top of the word "MÉTAUX". The logo is black and white.
Traditional 510(k) - SFI-Bar® 510(k) Summary
May 23, 2013
510(k) Summary
Applicant's Name and Address
Submitter:
Rue de Boujean 122 2501 Biel/Bienne, Switzerland Phone: +41 58 360 20 00 Fax: +41 58 360 20 10
Cendres + Métaux SA
Tanja Bongni Regulatory Affairs Manager
Date of Submission:
Name of the Device
Contact Person:
Trade Name: Common Name: Classification Name: Regulation Number:
Predicate Device(s):
SFI-Bar Br Abutment, Dental, Endosseous implants Endosseous Dental Implant Abutment 21 CFR 872.3630
Legally Marketed Device to which Equivalence is Claimed (Predicate Device)
May 23, 2013
K083876, K111390, K071638, K042429, K073268, K053152, K073282
Description of the Device
Device Description:
The SFI-Bar® provides the connection between compatible dental implant systems for the fixation of removable overdentures. The SFI-Bar@ consists of an implant adapter (abutinent) and a stress-free bar for the fixation of removable overdentures. The implant adapter is screwed into the dental implant.
The implant adapter (abutment) fit the Thommen SPI® Element Plattorn Ø 4.0 mm / the Neoss ProActive Implant Ø 3.5 / 4.0 / 4.5 / 5.0 / 5.5 mm and the Straumann cental implants / ITI Dental Implant System@ Standard Ø 4.1 and Ø 4.8 mm / Standard Plus Ø 4.1 mm and 0 4.8 mm / Tapered Effect 0 4.1 and 0 4.8 mm and Regular Neck (RN) Ø 4.8 mm and the BioHorizons Implant Systems Tapered Internal Implants / Internal Implants - bone level / Single-stage Implants - tissue level (Implant Platform: Ø 3.5 mm, Ø 4.5 mm, Ø 5.7 mm).
Page 1 of 6
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Cendres+Metaux. The logo consists of a stylized letter "C" on the left, followed by the text "CENDRES+" on the top line and "MÉTAUX" on the bottom line. The text is in a bold, sans-serif font.
May 23, 2013
Image /page/1/Picture/3 description: The image shows a logo with the text "CENDRES+ METAUX". To the left of the text is a stylized letter "M". The text is stacked, with "CENDRES+" on top and "METAUX" on the bottom. The logo is black and white.
Traditional 510(k) - SFI-Bar® 510(k) Summary
May 23, 2013
Intended Use of the Device:
The SFI-Bar® is intended to be used with the implant manufacturer's (Table 1) implant lo provide support for lixation of overdentures.
Table 1 Compatible Commercial Implant Manufacturers:
| Implant Manufacturer | Implant System | Implant Platform Diameter: |
|---|---|---|
| Institut Straumann | ITI Dental Implant System® | Standard 4.1 and 4.8 mm /Standard Plus 4.1 and 4.8 mm /Tapered Effect 4.1 and 4.8 mm /Regular Neck (RN) 4.8 mm |
| Thommen Medical | SPI® Element Platform | 4.0 mm |
| Neoss | Neoss ProActive Implant | 3.5 / 4.0 / 4.5 / 5.0 / 5.5 mm |
| BioHorizons | Tapered internal Implant - bonelevel | Ø 3.5 mm, Ø 4.5 mm, Ø 5.7 mm |
| Internal Implants -bone level | ||
| Single-stage Implants - tissuelevel |
Summary Technological Characteristics:
The proposed implant adapters are substantially equivalent to the currently marketed predicate devices. The intended use, basic design, fundamental operating principles and manufacturing procedures are the same as the predicate devices.
The material of the implant adapters conform to ASTM F 136, Whought Titanium-6Aluminium-4Vanadium ELI Alloy for Surgica! Implant applications (UNS R 56401). The parts for the SFI-Bar® System are rnanufactured from wires.
Comparison /Campatibility Substantially Equivatence:
The proposed implant adapters are substantially equivalent to the currently marketed predicate devices. The intended use, basic design. fundamental operating principles and manufacturing procedures are the same as the predicate device.
To ensure compatibility the following process was carried out: The implant adapters are developed and manufactured in close cooperation with the implant companies (see Table 1, column "Implant Company").
There are Quality Agreements between Cendres-Mistaux and the implant companies in place. Those agreements handle arnong other things the Design Control. Change Control, Complaint Handling and Post Market Surveillance,
Table 2 summarizes the substantial equivalence comparison to the predicate device:
Page 2 of 6
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with a stylized letter "M" on the left and the words "CENDRES+" and "MÉTAUX" stacked on the right. The letter "M" is bold and has a textured appearance. The text is in a simple, sans-serif font and is also in bold. The overall design is clean and professional.
May 23, 2013 100
Image /page/2/Picture/3 description: The image shows a logo with the letters 'CM' stacked on top of each other on the left side. To the right of the letters, the word 'CENDRES+' is written in all caps on the top line. Below that, the word 'MÉTAUX' is written in all caps on the second line. The logo appears to be for a company that deals with ashes and metals.
.
Traditional 510(k) – SFI-Bar® 510(k) Summary
.
. .
May 23, 2013
Table 2 Substantial Equivalence Comparison to Predicate Devices:
| SFI-Bar® | Predicate Devices | Comparison/Performance"Predicate Devices" | ||
|---|---|---|---|---|
| 510(k) No. | Candidate | K083876,K111390 | K071638,K042429,K073268,K053152,K073282 | |
| ProprietyName | SFI-Bar® | SFI-Bar® | "BioHorizonsTapered InternalImplant System","The ProdigySystem™ DentalImplants","BioHorizonsSingle-stageImplant","BioHorizonsSingle-stageImplant Systems","BioHorizonsInternal ImplantSystem" | |
| Model no orSystem | 05001022,05001021,05001020,05001019,05001023,05001027,05001026,05001025,05001024,05001028,05001032,05001031,05001030,05001029,05001033 | 05000576,05000577,05000578,05000579 | Tapered internalImplants - bonelevel,$Ø$ 3.8 (y)./ 4.6 (g)/ 5.8 (h).Internal Implants -bone.level,$Ø$ 3.5 (y) / 4.0 (g)/ 5.0 (b) / 6.0 (b).Single- stageImplants - tissuelevel,$Ø$ 3.5 (y) / 4.0(y&g) / 5.0 (g&b) /6.0 (b). | |
| CommonName: | Abutment forEndosseous Implant(Dental) | Abutment forEndosseous Implant(Dental) | Endosseous dentalimplant | |
| Manufacturer | Cendres + MétauxSA | Cendres + Métaux SA | BioHorizonsImplant Systems,Inc. | |
| ProductCodes | NHA | NHA | DZE | Both SFI-Bar®sare within thesame product |
Page 3 of 6
.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the text "CENDRES + MÉTAUX". To the left of the text is a stylized letter "M". The text is in a bold, sans-serif font. The logo appears to be for a company or organization related to ashes and metals.
May 23, 2013
.
Image /page/3/Picture/3 description: The image shows a logo with the letters "CM" on the left and the words "CENDRES+" and "MÉTAUX" on the right. The letters "CM" are in a bold, stylized font, and the words "CENDRES+" and "MÉTAUX" are in a smaller, sans-serif font. The logo is black and white. The plus sign is located to the right of the word "CENDRES".
Traditional 510(k) – SFI-Bar® 510(k) Summary
:
1000 - 1000
May 23, 2013
:
| SFI-Bar® | Predicate Devices | Comparison/Performance"Predicate Devices" | ||
|---|---|---|---|---|
| 510(k) No. | Candidate | K083876,K111390 | K071638,K042429.K073268.K053152.K073282 | |
| code. All predicatedevices are withina dentalapplication. | ||||
| RegulationIndicateduse | 872.3630Intended to be usedwith the implantmanufacturer'simplant (Table 1) toprovide support forfixation ofoverdentures. | 872.3630Intended to be usedwith the implantmanufacturer'simplant (Table 1:SPI® ElementPlatform 4.0 mmimplant) to providesupport for fixation ofoverdentures. | 872.3640Dental implant | Equivalent toSFI-Bar® |
| Manu-facturingProcess | Machined. Surfacetreated (anodized). | Machined. | Machined. Surfacetreated (anodized). | All SFI-Bar®Implant adapter aremachined. Thenew devices withinthis submission arealso partiallyanodized. Theanodizationprocess has noinfluence on thesafety &effectiveness of thedevice. Therefore,it can be statedthat themanufacturingprocess isequivalent to theSFI-Bar®(predicate devices).Equivalent toSFI-Bar®BioHorizonsimplants are alsosurface treated(same color |
Page 4 of 6
.
.
.
.
.
·
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the text "CENDRES+ METAUX". To the left of the text is a stylized letter "C". The text is in a bold, sans-serif font. The plus sign is smaller and slightly raised compared to the other letters.
.
·
.
: ·
Traditional 510(k) Submission USA: SFI-Bar® Format: Based on STED (GHTF/SG1/N011:2008) ·
:
May 23, 2013
·
.....
Image /page/4/Picture/4 description: The image shows a logo with the text "CENDRES+ METAUX". To the left of the text is a stylized symbol that appears to be a combination of the letters 'C' and 'M'. The text is stacked, with "CENDRES+" on top and "METAUX" below it. The logo is in black and white.
Traditional 510(k) - SFI-Bar® 510(k) Summary
: · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
·
May 23, 2013
.
| SFI-Bar® | Predicate Devices | Comparison/Performance"Predicate Devices" | ||
|---|---|---|---|---|
| 510(k) No. | Candidate | K083876,K1.11390 | K071638,K042429,K073268,K053152,K073282 | |
| OperatingPrinciple /BasicDesign | Impression taking:Optional,preassembled (plug-in connection). | Impression taking:Optional,preassembled (plug-inconnection). | Not relevant | ≒ Equivalent toSFI-Bar® |
| Abutment implantconnection: Screwfixation. | Abutment implantconnection: Screwfixation. | |||
| Connecting principleto overdenture:Retentive system. | Connecting principleto overdenture:Retentive system. | |||
| Cleaning proceduresfor patient: Commonprocedure for oralhygiene. | Cleaning proceduresfor patient: Commonprocedure for oralhygiene. | |||
| Patient handling:Common cleaningand insertion ofprosthesis. | Patient handling:Common cleaning andinsertion of denture. | |||
| Materialsincorporatedinto keyfunctionalelements | Titanium | Titanium | Not relevant | ≒ Equivalent toSFI-Bar® |
| Shelf Life | 95% after 10 years | 95% after 10 years | Not known byCendres + MétauxSA | ≒ Equivalent toSFI-Bar® |
| Packagingandsterilization | Dentalblister, non-sterile. | Dentalblister, non-sterile.Tubular bag, non-sterile | Not relevant | ≒ Equivalent toSFI-Bar® |
Pertormance Data:
.
·
Torque tests, application lesting and lunctional lesting have been conducted to evaluate the performance characteristics of the additional SFI-Rarib. The test methods used were the same as in the predicate device. Testing has shown that the SFI-BarQi is equivalent in performance characteristics to the predicate SFI-Bair®. The acceptance culeria were met.
. Page 5 of 6
・
:
CONFIDENTIAL
. . . .
、
·
:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains a logo with the text "CENDRES+ METAUX". To the left of the text is a stylized letter "C" that is bolded. The text is stacked, with "CENDRES+" on top and "METAUX" on the bottom. The text and the "C" are all in black.
1
.
·
May 23, 2013
Image /page/5/Picture/3 description: The image shows a logo with the text "CENDRES+" on the top line and "MÉTAUX" on the bottom line. To the left of the text is a stylized symbol that appears to be a combination of the letters 'C' and 'M'. The text and symbol are in black, and the background is white. The logo is simple and clean, with a focus on the text.
Traditional 510{k} - SFI-Bar® 510(k) Summary
May 23, 2013
Summary of Testing to Oernonstrate Safety and Effectiveness / Conclusion:
Non-clinical test data was used to support the substantially equivalence claim. Clinical testing was not necessary. Non-clinical testing consisted of analysis of platforms to identity worst-case test samples. Faligue lesting was not done as the basic cesign is the same than the predicate device. The evaluation was based on FDA guidance "Class II Soccial Controls Guidance Document: Root form Endosseous Dental Implants and Endosseous Dental Implant Abutments." Torque tests, application and functional tests have been carried out.
The summary of technological characteristics as well as the torque test, application and functional testing indicate that the cevice is safe and . effective for its intended use and performis as well or better than the predicate devices.
Page 6 of 6
{6}------------------------------------------------
Image /page/6/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
September 23, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Cendres & Metaux SA Ms. Tanja Bongni Regulatory Affairs Manager Rue De Boujean 122 2501 Biel/Bienne, Switzerland
Re: K131526
Trade/Device Name: SFI-Bar® Regulation Number: 21 CFR 872.3630 Regulation Name: Endosscous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: September 11, 2013 Received: September 13, 2013
Dear Ms. Bongni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.lda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicallDevices/Safets/Reportal/roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Richard C. Chapman 2013.09.23 13:19:52 -04'00'
for
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo with the text "CENDRES+ METAUX". To the left of the text is a stylized letter "C" that is connected to the letter "M". The text is stacked, with "CENDRES+" on top and "METAUX" on the bottom.
May 23, 2013
Indications for Use
KISIS26 5 10(k) Number (if known):
Device Name:
SFI-Bark
Indications for Use:
The SFI-Bans is intended to be used with the implant translations in the called in to provide support ler lixation of overdentures.
Only the imp ant systems listed in Table I may be used will SFI-Mar-y.
The compatible implant systems are specified in Table . below
| Implant Manufacturer | Implant System | Implant Platform Diameter |
|---|---|---|
| Institut Straumann | ITI Dental Implant System | Standard 4.1 and 4.8 mm /Standard Plus 4.1 and 4.8 mm /Tapered Effect 4.1 and 4.8 mm /Regular Neck (RN) 4.8 mm |
| Thommen Medica | SPI Element Platform | 4.0 mm |
| Neoss | Neoss ProActive Implant | 3.5 / 4.0 / 4.5 / 5.0 / 5.5 mm |
| BioHorizons | Tapered Internal Implant - bone level | ⌀ 3.5 mm, ⌀ 4.5 mm, ⌀ 5.7 mm |
| Internal Implants - bone level | ||
| Single-stage Implants - tissue level |
Table I Compatible Implani Systems
AND / OR Prescription Jsc × (Pari 21 CFR 801 Subpart D)
Over-The-Counter Jsc (2) CFR 801 Subpart C)
(PLEASE DO NOT WRITE RELOW THIS : INE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CORH. Office of Device Evaluation (OOE)
Andrew I. Steen -S 2013.09.23 08:00:13 -04'00'
:Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Fage : ci l
K131524 510(k) Number:
§ 872.3630 Endosseous dental implant abutment.
(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)