K Number
K101976
Date Cleared
2010-09-14

(63 days)

Product Code
Regulation Number
888.3358
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Apex Hip System is intended for primary or revision total hip replacement. The femoral hip stems and acetabular cup are intended for uncemented fixation and single use implantation. The Apex Acetabular Cup liners, standard and ApeX-LNK Poly, are intended for use with the Apex Modular Acetabular Cup, in combination with the Apex Modular, Apex K2, Apex K2 mid length or Apex K1 Hip in total hip replacement procedures. The acetabular cup inserts are intended to articulate with a metal (cobalt chromium) or ceramic (alumina) femoral head. These prostheses may be used for the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis Rheumatoid arthritis Correction of functional deformity Congenital dislocation Revision procedures where other treatments or devices have failed Femoral neck and trochanteric fractures of the proximal femur

Device Description

Non hooded (0 deg neutral) and Hooded (10 deg) +4mm offsets for use with 28, 32, 36, 40mm Apex Modular heads.

AI/ML Overview

The provided text describes a 510(k) summary for the Interface Acetabular System, Acetabular Insert. The documentation primarily focuses on the device's design, materials, indications for use, and a comparison to predicate devices, along with non-clinical testing.

Here's an analysis of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
ROM analysis per ISO-21535-2007All samples tested met the acceptance criteria.
Extreme artificial aging per ASTM F2003-02All samples tested met the acceptance criteria.
Push-out & Lever-out TestAll samples tested met the acceptance criteria.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: The document does not specify the exact number of samples used for each non-clinical test (ROM analysis, artificial aging, push-out & lever-out). It only states "All samples tested."
  • Data Provenance: Not applicable in the context of this device and testing. The tests are in vitro mechanical/material tests, not data derived from human subjects or clinical settings based on a specific country of origin or retrospective/prospective nature.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not applicable as the document explicitly states, "No clinical studies were performed." The "ground truth" for the non-clinical tests would be the performance standards defined in the ASTM and ISO standards, which are engineering specifications, not expert medical consensus.

4. Adjudication method for the test set:

  • This information is not applicable as there were no clinical studies involving human readers or subjective interpretations of data that would require an adjudication method. The testing involved objective physical measurements against established engineering standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. This is a medical device (hip insert), not an AI diagnostic or image analysis tool. The document explicitly states, "No clinical studies were performed."

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This information is not applicable. This device is a mechanical implant, not an algorithm.

7. The type of ground truth used:

  • For the non-clinical tests, the "ground truth" was established by engineering performance standards specified by ASTM F2003-02 (extreme artificial aging) and ISO-21535-2007 (ROM analysis), along with industry-standard "Push-out & Lever-out Test" criteria.

8. The sample size for the training set:

  • This information is not applicable as no clinical studies or AI/machine learning models were involved in the evaluation of this device.

9. How the ground truth for the training set was established:

  • This information is not applicable as no clinical studies or AI/machine learning models were involved.

{0}------------------------------------------------

K101976 (pg 1/2)

.

ੋਂ ਨੂੰ

Appendix C

SEP 1 ,4 2010

510(k) Summary

Interface Acetabular System, Acetabular Insert17 Aug, 2010
SubmitterOMNI life science, Inc.50 O'Connell WayE. Taunton MA 02718ContactRadhika PondicherryRegulatory Affairs774-226-1852(508) 822-6030 (fax)
Preparation Date17 Aug 2010Device NameInterface Acetabular System, Acetabular Insert
Trade NameInterface Acetabular System, Acetabular Insert
Common/Classification NameHip joint metal/ceramic/polymer semi-constrained cemented or nonporousuncemented prosthesis
Regulatory ClassClass II per 21 CFR §888.3358
Product CodeLPH
Legally MarketedPredicateDevice(s)K031110 Apex Modular Acetabular Cup, May 22, 2003 K062489 Apex HCLA Acetabular Cup Liners August 15, 2007 K073150 ApeX-LNK Poly™ Acetabular Cup Liners K100555 ApeX-LNK Poly™ Acetabular Cup Liners, March 29, 2010
DeviceDescriptionNon hooded (0 deg neutral) and Hooded (10 deg) +4mm offsets for use with 28,32, 36, 40mm Apex Modular heads.
Indications forUseThe Apex Hip System is intended for primary or revision total hip replacement.The femoral hip stems and acetabular cups are intended for uncementedfixation and single use implantation. The Apex Acetabular Cup liners, standardand ApeX-LNK Poly, are intended for use with the Apex Modular Acetabular Cup,in combination with the Apex Modular, Apex K2, Apex K2 mid length or Apex K1Hip in total hip replacement procedures. The acetabular cup inserts are intendedto articulate with a metal (cobalt chromium) or ceramic (alumina) femoral head.These prostheses may be used for the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis Rheumatoid arthritis Correction of functional deformity Congenital dislocation Revision procedures where other treatments or devices have failed Femoral neck and trochanteric fractures of the proximal femur

.


·

{1}------------------------------------------------

Predicate Device Comparison

AcetabularInsert(subject device)Apex Hip SystemStandardUHMWPE(K031110)Apex Hip SystemCrosslinkedUHMWPE(K062489,K073150, K100555)
DESIGN
Insert engagement19° taper and PElocking ring19° taper and PElocking ring19° taper and PElocking ring
Inserts- Hooded/NeutralNeutral10° hoodedNeutral15° hoodedNeutral10° hooded
Insert offset+4mmNoneNone
Insert Thickness(minimum)Worst case sizehas a minimumthickness of8.6mm.Minimumthickness of 6mmat the loadbearing region.Minimum thicknessof 6mm at the loadbearing region.
Head Diameters28, 32, 36 and40mm28 and 32, mm28, 32, 36 and40mm
MATERIALS
Cross-linked UHMWPEYesNoYes
Standard UHMWPYesYesNo
StandardsASTM F648ASTM F648ASTM F648

Non-Clinical Test Summary

The following tests were conducted:

  • ROM analysis per ISO-21535-2007 .
  • . Extreme artificial aging per ASTM F2003-02
  • . Push-out & Lever-out Test

All samples tested met the acceptance criteria.

Clinical Test Summary Conclusions

No clinical studies were performed.

The Interface Acetabular System, Acetabular Insert is substantially equivalent to the predicate devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, in the center. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Omni Life Science, Inc. % Radhika Pondicherry 50 O'Connell Way East Taunton, MA 02718

SEP 1 4 2010

Re: K101976

Trade/Device Name: Interface Acetabular System, Acetabular Insert Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH Dated: August 17, 2010 Received: August 19, 2010

Dear Mr. Pondicherry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Radhika Pondicherry

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

For Peter Hamer, D.D.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K101976 (pg 1/1)

Indications for Use

510(k) Number:

Device Name: Interface Acetabular System, Acetabular Insert, +4 offset

The Apex Hip System is intended for primary or revision total hip replacement. The femoral hip stems and acetabular cup are intended for uncemented fixation and single use implantation. The Apex Acetabular Cup liners, standard and ApeX-LNK Poly, are intended for use with the Apex Modular Acetabular Cup, in combination with the Apex Modular, Apex K2, Apex K2 mid length or Apex K1 Hip in total hip replacement procedures. The acetabular cup inserts are intended to articulate with a neal (cobalt chromium) or ceramic (alumina) femoral head. These prostheses may be used for the following conditions, as appropriate:

  • Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; .
  • . Rheumatoid arthritis;
  • Correction of functional deformity;
  • . Congenital dislocation;
  • Revision procedures where other treatments or devices have failed; .
  • Femoral neck and trochanteric fractures of the proximal femur.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Santur
dir mxm

(Devision Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K,01976

§ 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.

(a)
Identification. A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.(b)
Classification. Class II.