Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K171156
    Manufacturer
    Date Cleared
    2017-08-17

    (119 days)

    Product Code
    Regulation Number
    888.3353
    Why did this record match?
    Reference Devices :

    Modular Acetabular Cup, K031110, Apex Modular Acetabular Cup, K112779, Apex HCLA Acetabular Cup Liners, K062489

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OMNI Anseris Hip Stem is intended for use as the femoral component of a primary or revision total hip replacement when used with the Apex Interface™ Acetabular System. The Apex Interface™ Acetabular System articulates with the Apex Modular Femoral Head (Cobalt Chromium or Ceramic). The femoral hip stem is intended for uncemented fixation and single use implantation. These prostheses may be used for hip arthroplasty to treat the following conditions, as appropriate:

    • Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
    • Rheumatoid arthritis;
    • Correction of functional deformity;
    • Congenital dislocation;
    • Revision procedures where other treatments or devices have failed;
    • Femoral neck and trochanteric fractures of the proximal femur.
      The OMNI Anseris Hip Stem is also intended for use in hemiarthroplasty when used with the Apex Bipolar Head. For further details, please refer to the Apex Bipolar Head Instructions for Use.
    Device Description

    The OMNI Anseris Stem is a "fit and fill" femoral stem with a tapered conical design and is intended for use as the femoral component of a primary or revision total hip replacement when used with the Apex Interface™ Acetabular System.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "OMNI Anseris Hip Stem." It details the manufacturer's claim of substantial equivalence to existing predicate devices, rather than proving the device meets acceptance criteria through a standalone study of its performance in a clinical setting.

    Therefore, the information typically found in a study proving acceptance criteria for an AI/algorithm-based device (like sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, or ground truth establishment) is not present in this document because it describes a physical implantable device, not a software algorithm.

    However, I can extract the information related to the device's non-clinical testing which serves as the "study" proving it meets specific engineering/physical acceptance criteria.

    Here's the information parsed from the provided text, focusing on the nearest equivalent to your request:

    Acceptance Criteria and Device Performance for OMNI Anseris Hip Stem (Non-Clinical Testing)

    This document describes the non-clinical testing performed to demonstrate substantial equivalence of the OMNI Anseris Hip Stem to predicate devices. The "acceptance criteria" here refer to meeting specific engineering and performance standards for a hip implant, rather than clinical performance metrics of an AI algorithm.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Test Standard / RequirementReported Device Performance (as stated in the document)
    Fatigue StrengthISO 7206-6 (Fatigue testing of hip joint prostheses – Part 6: Endurance performance of stemmed femoral components with application of torsion)Met the requirements for fatigue strength per ISO 7206-6
    Fatigue StrengthISO 7206-4 (Fatigue testing of hip joint prostheses – Part 4: Endurance performance of stemmed femoral components with application of a force)Met the requirements for fatigue strength per ISO 7206-4
    Range-of-Motion (ROM)ISO 21535 (Non-active surgical implants – Joint replacement implants – Specific requirements for hip-joint replacements)Met the range-of-motion requirement per ISO 21535
    BiocompatibilityEthylene Oxide (EO) SterilizationDevices undergo the same validated sterilization process, using EO, to a sterility assurance level (SAL) of 10-6. All ethylene oxide residuals are monitored and well below standard limits.
    BiocompatibilityEndotoxin LevelsPlan to test endotoxins on all OMNIlife science device groups through Limulus amebocyte lysate (LAL) testing. Products segregated into 8 product groups (based on manufacturing process/material type) tested quarterly on a yearly rotation. Product not released if 20 EU/device limit is exceeded.
    Design/MechanicalConical Stem FEA per ASTM F2996-13Testing performed.
    Design/MechanicalSolidworks Simulation VerificationVerification performed.
    Design/MechanicalConical Stem Bipolar Head Impingement RiskTesting performed.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated in terms of individual units tested, but the document refers to "the results of testing of the proposed Anseris Hip Stems." For mechanical testing, this typically refers to a specific number of samples required by the standard (e.g., n=6 or more for fatigue testing). The document does not specify the exact number of physical stems tested.
    • Data Provenance: The testing was carried out by the manufacturer, OMNIlife Science. The data is internal to the company's testing processes as part of their 510(k) submission. There is no mention of country of origin of data in a clinical sense, nor is it retrospective or prospective as it's mechanical testing of a physical device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is a physical device being cleared through substantial equivalence to existing predicate devices based on non-clinical (mechanical and material) testing. There is no "ground truth" derived from expert consensus like in an AI/imaging study. The "ground truth" here is the pass/fail criteria of established international and industry standards (ISO, ASTM).

    4. Adjudication method for the test set

    • Not Applicable. As there are no human readers or interpretations of data akin to an AI study, there's no adjudication method. The tests are designed to provide objective, quantifiable results against predefined engineering specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a physical hip implant, not an AI software or diagnostic imaging tool. Therefore, MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device, not an algorithm. The "standalone performance" is the performance of the hip stem in a mechanical testing environment.

    7. The type of ground truth used

    • For the non-clinical testing, the "ground truth" refers to compliance with established international and industry standards for hip implants, such as ISO 7206-6, ISO 7206-4, ISO 21535, and ASTM F2996-13, as well as biocompatibility standards (e.g., SAL of 10-6 for sterility, endotoxin limits).

    8. The sample size for the training set

    • Not Applicable. There is no "training set" as this is not an algorithm undergoing machine learning.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set, there is no ground truth for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K101976
    Date Cleared
    2010-09-14

    (63 days)

    Product Code
    Regulation Number
    888.3358
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K031110, K062489, K073150, K100555

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apex Hip System is intended for primary or revision total hip replacement. The femoral hip stems and acetabular cup are intended for uncemented fixation and single use implantation. The Apex Acetabular Cup liners, standard and ApeX-LNK Poly, are intended for use with the Apex Modular Acetabular Cup, in combination with the Apex Modular, Apex K2, Apex K2 mid length or Apex K1 Hip in total hip replacement procedures. The acetabular cup inserts are intended to articulate with a metal (cobalt chromium) or ceramic (alumina) femoral head. These prostheses may be used for the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis Rheumatoid arthritis Correction of functional deformity Congenital dislocation Revision procedures where other treatments or devices have failed Femoral neck and trochanteric fractures of the proximal femur

    Device Description

    Non hooded (0 deg neutral) and Hooded (10 deg) +4mm offsets for use with 28, 32, 36, 40mm Apex Modular heads.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Interface Acetabular System, Acetabular Insert. The documentation primarily focuses on the device's design, materials, indications for use, and a comparison to predicate devices, along with non-clinical testing.

    Here's an analysis of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    ROM analysis per ISO-21535-2007All samples tested met the acceptance criteria.
    Extreme artificial aging per ASTM F2003-02All samples tested met the acceptance criteria.
    Push-out & Lever-out TestAll samples tested met the acceptance criteria.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: The document does not specify the exact number of samples used for each non-clinical test (ROM analysis, artificial aging, push-out & lever-out). It only states "All samples tested."
    • Data Provenance: Not applicable in the context of this device and testing. The tests are in vitro mechanical/material tests, not data derived from human subjects or clinical settings based on a specific country of origin or retrospective/prospective nature.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable as the document explicitly states, "No clinical studies were performed." The "ground truth" for the non-clinical tests would be the performance standards defined in the ASTM and ISO standards, which are engineering specifications, not expert medical consensus.

    4. Adjudication method for the test set:

    • This information is not applicable as there were no clinical studies involving human readers or subjective interpretations of data that would require an adjudication method. The testing involved objective physical measurements against established engineering standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This is a medical device (hip insert), not an AI diagnostic or image analysis tool. The document explicitly states, "No clinical studies were performed."

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This information is not applicable. This device is a mechanical implant, not an algorithm.

    7. The type of ground truth used:

    • For the non-clinical tests, the "ground truth" was established by engineering performance standards specified by ASTM F2003-02 (extreme artificial aging) and ISO-21535-2007 (ROM analysis), along with industry-standard "Push-out & Lever-out Test" criteria.

    8. The sample size for the training set:

    • This information is not applicable as no clinical studies or AI/machine learning models were involved in the evaluation of this device.

    9. How the ground truth for the training set was established:

    • This information is not applicable as no clinical studies or AI/machine learning models were involved.
    Ask a Question

    Ask a specific question about this device

    K Number
    K090845
    Date Cleared
    2010-04-07

    (376 days)

    Product Code
    Regulation Number
    888.3353
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apex ARC Hip Stem is intended for use as the femoral component of a primary or revision total hip replacement. This femoral hip stem is intended for uncemented fixation and single use implantation. This prosthesis may be used for the following conditions, as appropriate:

    • Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
    • Rheumatoid arthritis;
    • Correction of functional deformity;
    • Congenital dislocation:
    • Revision procedures where other treatments or devices have failed;
    • Femoral neck and trochanteric fractures of the proximal femur.
    Device Description

    The Apex ARC Hip Stem consists of a curved, rectangular tapered stem, and modular necks that connect to the tapered hole in the stem. The femoral stems are manufactured from titanium alloy and the modular necks are manufactured from cobalt chromium alloy. Three neck sizes are offered, with a neutral, 8 degree, and 12 degree varus-valgus angle, respectively. The necks are compatible with the modular heads that are part of the Apex Modular and Apex K2 hip systems (K000788, K012918, and K073150) and may be used with head diameters and offsets up to a maximum offset of +7 mm. These configurations allow the user to choose a combination of stem, neck, and head components to appropriately fit the anatomy of the patient. The Apex ARC Hip Stem may be used in conjunction with the Apex Modular™ Acetabular Cup (K031110, K062489, and K073150) for total hip arthroplasty.

    AI/ML Overview

    The provided document describes the 510(k) Summary for the Apex ARC™ Hip Stem (K090845). This is a medical device submission to the FDA, and as such, it focuses on demonstrating "substantial equivalence" to predicate devices rather than proving performance against specific acceptance criteria through a traditional clinical study with defined endpoints as would be expected for a novel AI/software device.

    Therefore, the requested information elements (acceptance criteria, study details, sample sizes, expert ground truth, MRMC, standalone performance) are not directly applicable in the context of this device's submission type and the information provided. Traditional clinical trials with acceptance criteria and statistical analysis are typically not required for Class II devices demonstrating substantial equivalence through non-clinical testing.

    However, I can extract the relevant information regarding the non-clinical tests conducted which serve a similar purpose to demonstrating the device's functional integrity and safety, and compare them to the general characteristics of the predicate devices.

    Here's a breakdown based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) submission for a mechanical implant, "acceptance criteria" are not explicitly defined in terms the performance metrics commonly associated with AI/software (e.g., sensitivity, specificity, AUC). Instead, the device's "performance" is demonstrated through non-clinical mechanical and material testing based on recognized international and ASTM standards. The acceptance criterion for these tests is generally successful completion of the test without failure, meeting the requirements of the specified standard, and demonstrating equivalence to predicate devices.

    Acceptance Criterion (Implicit: Compliance with Standard and Predicate)Reported Device Performance
    Mechanical Strength & Durability:
    Distal fatigue strength per ISO 7206-4:1989, ISO 7206-8:1992, and ASTM 2068-09.Tests conducted and results deemed successful.
    Fretting potential per ISO 17853:2003.Tests conducted and results deemed successful.
    Disassembly strength after fatigue testing per ASTM F2009-00.Tests conducted and results deemed successful.
    Proximal fatigue strength per ISO 7206-6:1992 and ASTM 2068-09.Tests conducted and results deemed successful.
    Torsional strength of the modular neck.Tests conducted and results deemed successful.
    Burst test, fatigue test, post-fatigue burst test, pull-off, and rotational stability of the worst case modular neck-ceramic head combination per FDA Guidance for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems, and ISO 7206-10:2003.Tests conducted and results deemed successful.
    Functional Compatibility:
    Range of motion per ISO 21535:2007.Tests conducted and results deemed successful.
    Biocompatibility/Sterilization:
    ETO Residuals per ANSI/AAMI/ISO 10993-7.Tests conducted and results deemed successful.
    Hydroxyapatite coating clearance.Previously cleared in K043123.
    Design and Material Equivalence to Predicates:
    Demonstrated through detailed comparison table with Apex K2™ Hip System and Aesculap Metha® Hip System.Features and materials largely align; differences (e.g., modular neck material and junction) addressed through specific additional testing where needed (e.g., torsional strength, burst/fatigue tests for neck combination).

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified in terms of number of physical devices tested for each non-clinical test. For mechanical tests, representative samples are typically used to meet the requirements of the standards. For material properties, it refers to the composition. The document doesn't detail the number of physical units tested.
    • Data Provenance: The tests are non-clinical, meaning they were performed in a laboratory setting, not on human subjects. No country of origin is specified for the testing, but the submitter (OMNI life science, Inc.) is based in E. Taunton, MA, USA. This is a prospective assessment for market clearance, meaning the tests were conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this 510(k) summary. For mechanical devices, "ground truth" typically relates to objective physical measurements or material properties as defined by engineering standards, not expert consensus as in diagnostic imaging. The "experts" involved would be the engineers and technicians conducting and validating the tests, adhering to the specified ISO and ASTM standards. Their qualifications are inherent in the rigorous nature of these testing facilities.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for expert consensus on clinical findings, not for objective mechanical test results. The results of the non-clinical tests are binary (pass/fail compliance with the standard).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a total hip replacement device, not an AI/software diagnostic tool, and therefore no MRMC study was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's evaluation is primarily based on engineering standards (ISO, ASTM) for mechanical performance and material properties. This includes objective measurements of fatigue strength, fretting potential, disassembly strength, torsional strength, range of motion, and material composition. Biocompatibility (ETO residuals) is also assessed against established standards.

    8. The sample size for the training set

    Not applicable. This is a physical device, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.


    In summary: The Apex ARC™ Hip Stem, a Class II medical device, demonstrates its safety and effectiveness for market clearance by showing substantial equivalence to existing predicate devices (Apex K2™ Hip System and Aesculap Metha® Hip System) through a detailed comparison of design, materials, and intended use, and by successfully completing a series of non-clinical mechanical and material tests according to recognized international standards (ISO, ASTM) and FDA guidance documents. No clinical studies or AI-related evaluations were performed for this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092443
    Date Cleared
    2009-10-23

    (74 days)

    Product Code
    Regulation Number
    888.3358
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K031110, K062489, K073150, K934765, K953490, K972774

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Interface™ Acetabular Cup Liners are intended for use with the Interface™ Acetabular Cup, in combination with the Apex Modular™, Apex K2™, or Apex K1™ Hip Stem in total hip replacement procedures. The acetabular cup liners are intended to articulate with a metal (cobalt chromium) or ceramic (alumina) femoral head. This prosthesis is intended for single use implantation, and may be used for the following conditions, as appropriate:

    • . Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
    • e Rheumatoid arthritis:
    • . Correction of functional deformity;
    • . Congenital dislocation:
    • . Revision procedures where other treatments or devices have failed;
    • Femoral neck and trochanteric fractures of the proximal femur. .
    Device Description

    The Interface™ Acetabular Cup Liners are manufactured of compression molded ultrahigh molecular weight polyethylene, sterilized using ethylene oxide. The articular geometry of the liners are compatible with existing Apex Modular femoral heads, manufactured from cobalt chrome or alumina ceramic, 28 mm, 32 mm or 36 mm diameter. The subject device adds an option for a 20° elevated rim that was not previously offered.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Interface™ Acetabular Cup Liners." This submission focuses on establishing substantial equivalence to previously marketed predicate devices, rather than presenting a study of the device's clinical performance against specific acceptance criteria in the manner of a clinical trial for an AI/ML medical device.

    Therefore, the specific information requested about acceptance criteria, device performance, sample sizes, expert involvement, ground truth, and training sets (relevant to AI/ML device studies) is not applicable to this 510(k) submission for a physical orthopedic implant.

    Here's why and what information is available:

    • Nature of the Device: The Interface™ Acetabular Cup Liners are physical medical devices (orthopedic implants) used in total hip replacement procedures. They are not an AI/ML algorithm or software that performs diagnostic or prognostic functions.
    • Regulatory Pathway (510(k)): A 510(k) submission demonstrates "substantial equivalence" to a legally marketed predicate device. This typically involves comparing design, materials, manufacturing processes, intended use, and performance characteristics (often mechanical testing for implants) to established, safe, and effective devices. It does not generally require new clinical trials or performance studies against predefined clinical acceptance criteria in the same way an AI/ML device would.

    What is provided in the document and how it relates to "acceptance criteria" for a physical device:

    The "acceptance criteria" in this context are primarily the demonstration of substantial equivalence to predicate devices. The study proving this involves detailed comparisons rather than a clinical performance study.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The "acceptance criteria" here are implicitly meeting the design, material, and functional equivalence to the predicates. The "reported device performance" is the comparison showing these equivalences.

    Acceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance (Comparison from document)
    Intended Use: Similar to predicate device(s).Similar: "Modular liner in metal shell, primary and revision THA" (Yes, cementless for both subject and predicates)
    Design: Similar locking mechanism and articular geometry.Liner engagement: Subject device has "19° taper and PE locking ring," which is identical to Interface™ (K031110) and ApeX-LNK Poly™ (K062489 and K073150).
    Liner options: Subject device offers "20° elevated rim," which is equivalent to the 20° elevated rim option in the Zimmer Trilogy® Acetabular System. Predicates offered Neutral, 10°, 15°, and 20° options.
    Head diameters: Subject device accommodates "28, 32 and 36 mm" heads, consistent with ApeX-LNK Poly™ and Zimmer Trilogy®. Interface™ (K031110) accommodated 28 and 32 mm.
    Materials: Similar to predicate device(s).Cross-linked UHMWPE: Subject device uses No (i.e., standard UHMWPE), similar to Interface™ and Zimmer Trilogy®. ApeX-LNK Poly™ uses Yes (cross-linked). This is a difference, but the submission claims it's not a barrier to equivalence.
    Sterilization: Similar to predicate device(s).Ethylene oxide: Subject device uses Ethylene oxide, identical to Interface™ and ApeX-LNK Poly™. Zimmer Trilogy® uses Gas plasma. This is a difference, but the submission claims it's not a barrier to equivalence, likely due to established sterilization methods.

    2. Sample size used for the test set and the data provenance: Not applicable. This is not a clinical study involving a test set of patient data. The "test" is the comparison of device characteristics to predicates.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in the clinical sense, is not established for this type of submission. The "ground truth" is the established safety and effectiveness of the predicate devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is an orthopedic implant, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable in the context of AI/ML. For this 510(k) submission, the "ground truth" (or basis of comparison) is the known safety and effectiveness profiles of the predicate devices (Interface™ (K031110), ApeX-LNK Poly™ (K062489 and K073150), and Zimmer Trilogy® Acetabular System (K934765, K953490, and K972774) UHMWPE cup liners) as determined by their prior FDA clearances and long-standing use in the market.

    8. The sample size for the training set: Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable.

    In summary: The provided document is a 510(k) premarket notification that demonstrates substantial equivalence for a physical medical device (acetabular cup liners). The "study" proving it meets "acceptance criteria" involves a detailed comparison of its design, materials, and intended use against already cleared predicate devices. The concepts of test sets, training sets, expert ground truth, and AI-assisted performance are not relevant to this type of regulatory submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1