Search Results
Found 4 results
510(k) Data Aggregation
(176 days)
The Zimmer Biomet™ Select Ceranic Heads are indicated for the following conditions: (1) a severely painful and/or disabled hip joint as a result of osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia, (2) avascular necrosis of the femoral head, (3) acute traumatic fracture of the femoral head or neck, (4) failed previous hip surgery including joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement, (5) certain cases of ankylosis, (6) nonunions, correction of functional deformity, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
When used with constrained acetabular liners, the Zimmer Biomet™ Select Ceramic Heads are intended for primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered.
The Zimmer Biomet™ Select™ Ceramic Heads are modular ceramic heads with a Type 1, 12/14 or 6 Degree adapter sleeve. They are offered in a variety of head diameters and neck configurations. They are intended for mating with a variety of Titanium Alloy and Cobalt-Chromium Alloy, as well as Stainless Steel, femoral stems equipped with tapered necks.
This document is a 510(k) Premarket Notification from the FDA for the Zimmer Biomet™ Select™ Ceramic Heads. Since this is a submission for a medical device that does not involve artificial intelligence or machine learning, the concepts of acceptance criteria for algorithm performance, sample sizes for training/test sets, expert ground truth, adjudication methods, multi-reader multi-case studies, or standalone algorithm performance are not applicable.
Instead, the document focuses on demonstrating substantial equivalence to legally marketed predicate devices based on:
- Intended Use and Indications for Use: The subject device has the same intended use and similar indications for use as the predicate devices.
- Materials: The subject device uses similar ceramic materials that meet ISO 6472-2 standards, identical to a reference device (ICONACY™ I-Hip™ Ceramic Heads).
- Design Features: Both the subject and predicate devices mate with various femoral stems with tapered necks. The subject device offers a larger range of tapers and offsets.
- Sterilization: Identical to predicates.
The study that proves the device meets the acceptance criteria is a non-clinical performance study focusing on the mechanical and material characteristics of the implant. The "acceptance criteria" here are defined by the performance standards expected of such medical devices and comparison to predicate devices, rather than algorithm metrics.
Here's a breakdown of the requested information based on the provided document, acknowledging the non-AI context:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria Category | Specific Test/Characteristic | Reported Device Performance (Summary) |
---|---|---|
Intended Use | Identical to predicate | Same intended use as predicate devices. |
Indications for Use | Similar to predicate | Similar indications for use as predicate devices. |
Materials | Meets ISO 6472-2, Ceramic Composition | Made from alumina matrix composite. Meets ISO 6472-2. Identical to ICONACY™ I-Hip™ Ceramic Heads (a reference device). |
Design Features | Mating with femoral stems, range of tapers/offsets | Mates with various Titanium Alloy, Cobalt-Chromium Alloy, and Stainless Steel femoral stems with tapered necks. Offers a larger range of tapers and offsets compared to predicates. |
Sterilization | Method | Identical to predicates. |
Mechanical Performance | Static Compression | Non-clinical tests were conducted. |
Axial Fatigue and Post-Fatigue Compression | Non-clinical tests were conducted. | |
Axial Pull-Off | Non-clinical tests were conducted. | |
On-axis and Off-axis Impaction and Compression | Non-clinical tests were conducted. | |
Resistance to Wear and Head Retention | Considered in non-clinical tests. | |
Range of Motion | Non-clinical tests were conducted. | |
Safety and Effectiveness | No new questions raised compared to predicates | Performance data and analyses demonstrate that differences do not raise new questions of safety and effectiveness, and the device is at least as safe and effective as legally marketed predicate devices. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable in the context of this 510(k) for a physical medical device. There are no "test sets" of data in the AI/ML sense. The "testing" refers to mechanical and material characterization of the device components. The document does not specify the number of individual units tested for each mechanical test, nor does it specify a "country of origin for data" or whether the mechanical tests were "retrospective or prospective."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable. "Ground truth" in the AI/ML sense (e.g., expert-labeled images) is not relevant for this medical device submission. The "ground truth" for mechanical properties is established through standardized engineering tests and material specifications.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. Adjudication methods are used for resolving disagreements in expert labeling of data, which is not part of this device submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. MRMC studies are specific to evaluating the impact of AI on human performance (e.g., radiologists interpreting images), which is outside the scope of this mechanical implant device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. This device is a physical implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
As explained above, "ground truth" in the AI/ML context is not applicable. For this device, the "ground truth" is defined by:
- Standardized Material Specifications: e.g., ISO 6472-2 for ceramic materials.
- Engineering Mechanical Test Results: Bench testing of materials and device components for properties like compression, fatigue, pull-off, wear, and range of motion.
- Comparison to Predicate Devices: Demonstrating that the device's characteristics are similar or equivalent to those of previously cleared devices.
8. The sample size for the training set
This is not applicable. There is no "training set" in the AI/ML sense for this device.
9. How the ground truth for the training set was established
This is not applicable. There is no "training set" or corresponding "ground truth" establishment in the AI/ML sense for this device.
Ask a specific question about this device
(24 days)
- Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis.
- Correction of functional deformity.
- Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- Revision of previously failed total hip arthroplasty.
- Dislocation risks.
The ArComXL™ Active Articulation Head is a single-use implant, intended for uncemented applications.
The ArComXL™ Active Articulation belongs to the family of dual mobility acetabular implants: the presence of two articulating surfaces in the same joint device. The ArComXL™ Active Articulation Head fits over a femoral modular head, which articulates within the ArComXL™ Head. The resultant assembly then articulates within the acetabular metal shell. The ArComXL™ Head is designed to be used with several styles of acetabular shells that have been cleared in previous submissions: M²a Magnum™ (K042037), Magnum™ Tri-Spike (K062995), and M²a 38™ Flared Cups and Non-Flared Cups (K011110).
The ArComXL™ Active Articulation Heads are available in sizes 44-66mm (Note: Size 44-66mm references O.D. of mating shell; the actual head sizes are 38-60mm.) and are manufactured from highly cross-linked polyethylene, conforming to ASTM F648. ArComXL™ is not a new material; the material and manufacturing process were cleared in K042051, ArComXL™ Polyethylene Liners, as well as subsequent submissions. The ArComXL™ Active Articulation is designed for both primary and total revision surgeries, where all device components associated with the wear couple are removed and replaced. The system is intended for uncemented applications.
This document describes the Biomet ArComXL™ Active Articulation Head, a hip replacement component. This submission is for a material change from the predicate device.
Here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
Characteristic | Acceptance Criteria | Reported Device Performance |
---|---|---|
Push-In and Lever-Out | Equivalence to the Bipolar/Tri-Polar predicate (K991990). Meets or exceeds parameters. | Six samples demonstrated equivalence to the Bipolar/Tri-Polar predicate (K991990). Meets parameters. |
Wear Testing | Wear rates less than that of the ArCom™ 36mm liners (K032396). Meets or exceeds parameters. | Tested for 5 million cycles, demonstrated equivalence to K032396. Meets parameters. |
2. Sample Size and Data Provenance
- Test Set Sample Size:
- Push-In and Lever-Out: Six samples.
- Wear Testing: Not explicitly stated, but typically multiple samples are used in ISO 14242 testing. The document states "the ArComXL™ Active Articulation heads were tested".
- Data Provenance: Not explicitly stated, but assumed to be from laboratory testing conducted by Biomet Manufacturing Corp. (the submitter). These are non-clinical tests.
3. Number of Experts and Qualifications for Ground Truth of Test Set
Not applicable. The reported studies are non-clinical mechanical tests (Push-In and Lever-Out, Wear Testing), not clinical studies requiring expert ground truth for interpretation.
4. Adjudication Method for Test Set
Not applicable. This section relates to clinical studies and expert review, which were not performed for the determination of substantial equivalence for this device.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. No clinical studies were conducted, and therefore, no MRMC studies comparing human readers with and without AI assistance were performed. The device is a medical implant, not an AI-assisted diagnostic tool.
6. Standalone (Algorithm Only) Performance Study
Not applicable. This device is a physical implant, not an algorithm or software. The performance studies are non-clinical hardware tests.
7. Type of Ground Truth Used
- Push-In and Lever-Out: Comparison to the performance of a legally marketed predicate device (K991990), establishing functional equivalence.
- Wear Testing: Comparison to the wear rates of a legally marketed predicate device, ArCom™ 36mm liners (K032396), as per ISO 14242 standards.
8. Sample Size for Training Set
Not applicable. This device is a physical implant; there is no "training set" in the context of an algorithm or AI model development.
9. How Ground Truth for Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
(26 days)
Indications for use of the ConsensusTM Hip System or UNISYN™ Hip System
- A) Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
- B) Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
- C) Proximal femoral fractures.
- D) Avascular necrosis of the femoral head.
- E) Non-union of proximal femoral neck fractures.
- F) Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa vara, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities
Acetabular components are indicated for cemented and cementless use. Consensus femoral stems are indicated for cemented and cementless use. UniSyn femoral stems are indicated for cementless use only. HA coated implants are indicated for cementless use only.
The Consensus Hip 36mm CoCrMo (ASTM F799 or ASTM F1537) femoral head is offered with four different offsets, -5mm, neutral, +5mm and +10mm. Other than size, it incorporates no new design features. Female head taper is identical to the head taper design for the cleared devices.
The 36mm acetabular insert is offered in sizes 54 - 68 (2 mm increments), in neutral, 10 and 20 degree hooded configurations. Material for all 36mm insert sizes is crosslink UHMWPE (ASTM F648). Hooded configurations incorporate a titanium alloy x-ray marker (Ti 6Al-4V ELJ ASTM F136). Insert locking detail of the subject device is identical to the locking detail of the cleared devices.
The provided text is a 510(k) summary for a medical device: a 36mm CoCrMo Femoral Head and Acetabular Insert for the Consensus Hip System. This document focuses on demonstrating substantial equivalence to previously cleared devices, rather than presenting a study with acceptance criteria and a detailed performance evaluation of a new AI/software device.
Therefore, the information requested in your prompt (acceptance criteria, study details, sample sizes, ground truth, expert qualifications, adjudication, MRMC studies, standalone performance, training set details) is not applicable to this type of regulatory submission.
This 510(k) relies on the following for demonstrating safety and effectiveness:
- Comparison to Cleared Device: The primary argument is that the new 36mm femoral head and acetabular insert are substantially equivalent to existing, legally marketed devices. The only change is the addition of a new size (36mm).
- Intended Use, Material, Design Features, and Type of Interface: These aspects of the new components are stated to be identical or highly similar to predicate devices.
- Materials Data: The submission mentions "materials data" was provided to support safety and effectiveness, implying biocompatibility and mechanical properties testing relevant to the materials used (CoCrMo and UHMWPE) rather than clinical performance data.
In summary, this document does not contain the kind of study design and performance data typically associated with evaluating a novel AI/software medical device against specific acceptance criteria.
Ask a specific question about this device
(67 days)
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis.
- Correction of functional deformity.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- Revision of previously failed total hip arthroplasty.
Cemented and Uncemented Applications
Biomet Manufacturing Corp. is adding additional sizes and designs to the predicate ArComXL™ Acetabular Liners (K042051). The same manufacturing process used in the predicate results in a higher cross-linked polyethylene that Biomet will herein refer to as ArComXL™ . The femoral heads, sizes 38mm and 40mm, are a one-piece design with neck length variations ranging from -6mm to +9mm made from CoCrMo.
The provided document is a 510(k) Summary for a medical device (ArComXL™ Polyethylene Liners and 38/40mm Femoral Heads). It describes the device, its intended use, and claims substantial equivalence to predicate devices. Crucially, it states: "Clinical Testing: None provided as a basis for substantial equivalence."
Therefore, a study proving the device meets acceptance criteria as typically understood for AI/diagnostic devices (e.g., performance metrics, ground truth, expert adjudication, MRMC studies) was not conducted for this submission, as it is a medical device, not a diagnostic or AI product.
However, given the request's structure which assumes such a study, I can only extract information related to "Non-Clinical Testing" which serves a similar purpose of validating the device.
Here's a breakdown based on the provided text, primarily addressing the "Non-Clinical Testing" section since clinical trials were not performed.
Description of Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Non-Clinical) | Reported Device Performance (Non-Clinical) |
---|---|
Meet or exceed current standards or guidelines for ArComXL™ liners. | ArComXL™ liners met or exceeded current standards or guidelines. |
(Note: The document does not specify the exact "standards or guidelines" or quantitative performance metrics, only that they were met or exceeded.)
2. Sample size used for the test set and the data provenance
The document does not detail specific sample sizes for non-clinical testing. It simply states "Verification activities were performed on ArComXL™ liners." There is no mention of a "test set" in the context of clinical or diagnostic data, as this is a physical medical implant. Therefore, data provenance (country of origin, retrospective/prospective) is not applicable in the way it would be for a diagnostic study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This section is not applicable. Since no clinical or diagnostic study was performed, there was no "ground truth" to establish in the context of expert review of data/images. The "ground truth" for non-clinical testing of a physical implant would typically involve engineering specifications, material properties, and mechanical test results against established benchmarks or regulatory standards, which are not detailed here.
4. Adjudication method for the test set
This section is not applicable. No test set requiring adjudication by experts was utilized as no clinical or diagnostic study was conducted.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done
No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical Testing: None provided as a basis for substantial equivalence."
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This section is not applicable. The device is a physical medical implant (hip replacement components), not an algorithm or AI product.
7. The type of ground truth used
For the non-clinical testing, the "ground truth" would be related to engineering specifications, material science properties, and established regulatory/industry standards for medical implants (e.g., ASTM F-648 for UHMWPE). The document confirms that the device characteristics were "identical to the predicate (K042051)" and met "current standards or guidelines."
8. The sample size for the training set
This section is not applicable. There is no "training set" in the context of an AI/algorithm for this device.
9. How the ground truth for the training set was established
This section is not applicable. There is no "training set" or corresponding ground truth establishment in the context of an AI/algorithm for this device. The "ground truth" for the manufacturing materials and processes would be based on validated scientific and engineering principles and quality control measures.
Ask a specific question about this device
Page 1 of 1