(285 days)
Not Found
No
The device description and performance studies focus on standard biochemical reactions and electrical current measurements, with no mention of AI or ML algorithms for data processing or interpretation. The LDL calculation is a simple numerical formula.
No
The device measures total cholesterol, high density lipoprotein cholesterol, triglycerides, and glucose, which are diagnostic indicators, but it does not directly treat or provide therapy to patients.
Yes.
The device measures total cholesterol, HDL-C, triglyceride, and glucose, and these measurements are explicitly stated to be "used in the diagnosis and treatment of disorders" involving excess cholesterol, lipoprotein metabolism disorders, lipid disorders, atherosclerosis, and various liver and renal diseases, as well as an "aid in the management of diabetes." This statement directly indicates its use for diagnostic purposes.
No
The device description explicitly states that the LipidPro™ system consists of a meter and test strips, which are hardware components. It also describes the physical principles of measurement (reflection density and electrical current), indicating a reliance on physical interactions with the blood sample and the hardware.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Explicit Statement: The "Intended Use / Indications for Use" section explicitly states: "LipidPro™ system is intended for in home (self-testing) or health care professionals and for testing outside the body (in vitro diagnostic use only)."
- Testing Outside the Body: The description clearly indicates that the system measures substances (total cholesterol, HDL-C, triglyceride, and glucose) in capillary whole blood, which is a sample taken from the body and tested outside the body. This is a key characteristic of IVDs.
- Diagnostic Purpose: The intended use describes how the measurements are used in the "diagnosis and treatment" of various disorders related to cholesterol, lipids, and glucose metabolism. While the glucose measurement is not for initial diagnosis or screening, the lipid measurements are explicitly stated for diagnostic use.
- Control Solutions: The description of the control solutions also states they are for "testing outside the body (in vitro diagnostic use only)."
Therefore, based on the provided text, the LipidPro™ system is definitively an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
LipidPro™ system is intended for in home (self-testing) or health care professionals and for testing outside the body (in vitro diagnostic use only). LipidPro™ system which consists of meter and test strips, measures total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG) and glucose in capillary whole blood.
Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis and various liver and renal diseases. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. Calculated LDL-cholesterol values are reported only when triglycerides are 350 mg/dL, calculated LDL-cholesterol are not reported.
Glucose measurement is for the quantitative measurement of the concentration of glucose in capillary whole blood that can be taken from the fingertip, ventral palm, dorsal hand, upper arm, forearm, calf and/or thigh by diabetic patients or health care professionals as an aid in the management of diabetes. Glucose measurement is not to be used for the diagnosis of or screening for diabetes or for neonatal use. Alternate site testing should be done during steady- state times when glucose is not changing rapidly.
LipidPro™ Total Cholesterol control solution, HDL-cholesterol control solution, Triglyceride control solution, and Glucose control solution are used to test the precision of the LipidPro™ system and to detect systematic analytic deviations that may arise from reagent or analytical variation and they are for testing outside the body (in vitro diagnostic use only.) LipidPro™ control solutions are intended for in home and health care professionals use.
Product codes (comma separated list FDA assigned to the subject device)
CGA, NBW, JGY, CHH, LBR, JJX
Device Description
The LipidPro " system is a device which combines measuring systems for total cholesterol, high density lipoprotein cholesterol, low density lipoprotein cholesterol and triglyceride, and a blood glucose monitoring system into one convenient device. The glucose monitoring system function of the device utilizes Infopia's Evolution™ monitoring system cleared under K072369.
The LipidPro 10 system consists of a meter, five types of test strips, and four types of control solutions. The five types of the test strips are for total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG),a lipid profile (which combines TC, HDL-C, TG tests) and glucose test respectively. The four types of the control solutions are for total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG) and glucose test respectively.
LDL value is calculated from TC, HDL-C, and TG values using the following numerical formula. When TG is over 350 mg/dL, LDL is not calculated.
LDL=TC-HDL-(TG/5)
LipidPro " glucose test system is identical to the EVOLUTION™ glucose test system cleared under K072369. LipidPro™ glucose test strip is the same as the EVOLUTION™ glucose test strip except the cover printing design. Their size and components are the same. LipidPro™ glucose test meter is identical to the EVOLUTION " glucose test meter except the external design and having a lipid profile (TC/TG/HDL-C) test part on the other side of the meter. This does not affect to the safety and effectiveness for glucose test and data.
The test range of LipidPro TM system is 100-400mg/dL for total cholesterol, 25-80 mg/dL for HDLcholesterol. 70-600 mg/dL for Triglyceride and 20-600mg/dL for glucose. LDL cholesterol should be estimated from measurement of total cholesterol, HDL cholesterol and triglycerides.
LipidPro™ system has two kinds of test principle according to test items: one for glucose and the other for a lipid profile (TC, HDL-C, and TG).
Lipid profile test results are based on reading reflection density. When the blood is applied, the color changes in test area through an enzyme reaction. The meter records this change in color and converts the measurement signal to the displayed result using the data previously entered via the code. The deeper the color is, the higher the lipid level is.
Glucose in the blood sample will react to the electrodes in the glucose test strip, generating an electrical current that will stimulate a chemical reaction is measured by LipidPro™ meter and displayed as your blood glucose result.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
fingertip, ventral palm, dorsal hand, upper arm, forearm, calf and/or thigh
Indicated Patient Age Range
Not Found
Intended User / Care Setting
in home (self-testing) or health care professionals
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Clinical: The clinical performance evaluation using the LipidPro™ system was conducted for the purpose of validating the TC, HDL-C, LDL-C, TG, and Glucose measuring accuracy. Test results showed substantial equivalence.
Non-clinical: Verification, validation and testing activities were conducted to establish the performance, functionality and reliability characteristics of the LipidPro™ system with respect to the predicate device. Pass or fail criteria were based on the specifications cleared for the predicate device and results showed substantial equivalence.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
0
ﻬﺮ ﺑ
Infopia Co., Ltd.
510(K) Summary (K090405)
DEC - 1 2009
This summary of 510(k) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
1. Application date: 02/17/09
2. Submitter Information
Infopia Co.,Ltd.
#1603, Dongil Technotown A Bldg., 889-3, Kwanyang2-Dong, Dongan - Gu Anyang, Kyunggi 431-716, Korea Phone: +82-31-423-6170 Fax: +82-31-423-6171
ﺪ
3. U.S Agent/Contact Person
13340 E Firestone Blvd. Suite J, Santa Fe Springs, CA 90670 Priscilla Juhee Chung Fax: 562-404-2757 Email: agent.fda@gmail.com Phone: 562-404-8466
- Trade name: LipidPro™ system, LipidPro™ Total cholesterol (TC) control solution, LipidPro™ High density lipoprotein cholesterol (HDL-C) control solution, LipidPro™ Triglyceride (TG) control solution and LipidPro™ Glucose control solution
5. Classification name:
Total cholesterol test system (21 CFR Part 862.1175, CHH)
Lipoprotein test system (21 CFR Part 862.1475, LBR)
Triglyceride test system (21 CFR Part 862.1705, JGY)
Glucose test system (21 CFR Part 862.1345, NBW, CGA)
Quality control material (assayed and unassayed) (21 CFR Part 862.1660, JJX)
6. Predicate
Predicate for the Lipid profile test system ・
PTS PANELS Lipid Panel Test Strips (K023558) / Polymer Technology Systems, Inc.
- Predicate for the glucose test system -
Evolution™ Blood Glucose Test System (K072369) / HMD Biomedical LLC.
1
Predicate for the control solution
PTS Panels Multi-Chemistry Controls (K022401) / Polymer Technology Systems, Inc.
7. Device Description
The LipidPro " system is a device which combines measuring systems for total cholesterol, high density lipoprotein cholesterol, low density lipoprotein cholesterol and triglyceride, and a blood glucose monitoring system into one convenient device. The glucose monitoring system function of the device utilizes Infopia's Evolution™ monitoring system cleared under K072369.
The LipidPro 10 system consists of a meter, five types of test strips, and four types of control solutions. The five types of the test strips are for total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG),a lipid profile (which combines TC, HDL-C, TG tests) and glucose test respectively. The four types of the control solutions are for total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG) and glucose test respectively.
LDL value is calculated from TC, HDL-C, and TG values using the following numerical formula. When TG is over 350 mg/dL, LDL is not calculated.
LDL=TC-HDL-(TG/5)
LipidPro " glucose test system is identical to the EVOLUTION™ glucose test system cleared under K072369. LipidPro™ glucose test strip is the same as the EVOLUTION™ glucose test strip except the cover printing design. Their size and components are the same. LipidPro™ glucose test meter is identical to the EVOLUTION " glucose test meter except the external design and having a lipid profile (TC/TG/HDL-C) test part on the other side of the meter. This does not affect to the safety and effectiveness for glucose test and data.
The test range of LipidPro TM system is 100-400mg/dL for total cholesterol, 25-80 mg/dL for HDLcholesterol. 70-600 mg/dL for Triglyceride and 20-600mg/dL for glucose. LDL cholesterol should be estimated from measurement of total cholesterol, HDL cholesterol and triglycerides.
LipidPro™ system has two kinds of test principle according to test items: one for glucose and the other for a lipid profile (TC, HDL-C, and TG).
Lipid profile test results are based on reading reflection density. When the blood is applied, the color changes in test area through an enzyme reaction. The meter records this change in color and converts the measurement signal to the displayed result using the data previously entered via the code. The deeper the color is, the higher the lipid level is.
Glucose in the blood sample will react to the electrodes in the glucose test strip, generating an electrical current that will stimulate a chemical reaction is measured by LipidPro™ meter and displayed as your blood glucose result.
2
8. Intended Use
LipidPro™ system is intended for in home (self-testing) or health care professionals and for testing outside the body (in vitro diagnostic use only). LipidPro™ system which consists of meter and test strips, measures total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG) and glucose in capillary whole blood.
Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis and various liver and renal diseases. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. Calculated LDL-cholesterol values are reported only when triglycerides are 350 mg/dL, calculated LDL-cholesterol are not reported.
Glucose measurement is for the quantitative measurement of the concentration of glucose in capillary whole blood that can be taken from the fingertip, ventral palm, dorsal hand, upper arm, forearm, calf and/or thigh by diabetic patients or health care professionals as an aid in the management of diabetes. Glucose measurement is not to be used for the diagnosis of or screening for diabetes or for neonatal use. Alternate site testing should be done during steady- state times when glucose is not changing rapidly.
LipidPro™ Total Cholesterol control solution, HDL-cholesterol control solution, Triglyceride control solution, and Glucose control solution are used to test the precision of the LipidPro™ system and to detect systematic analytic deviations that may arise from reagent or analytical variation and they are for testing outside the body (in vitro diagnostic use only.) LipidPro™ control solutions are intended for in home and health care professionals use.
3
9. Comparison of Technological Characteristics with Predicate
The LipidPro™ system has the same technological characteristics as the current legally marketed predicate device, PTS PANELS Lipid Panel Test Strips (K023558), Evolution™ Blood Glucose Test System (K072369) and PTS Panels Multi-Chemistry Controls (K022401).
Similarities and Differences of the Lipid profile measuring system | ||
---|---|---|
Item | LipidPro™ lipid measuring system | PTS PANELS Lipid Panel Test Strips (K023558) |
Detection method | Spectrophotometry | Spectrophotometry |
Enzyme | TC: Cholesterol esterase, | |
Cholesterol oxidase, Peroxidase | ||
HDL-C: Cholesterol esterase, | ||
Cholesterol oxidase, Peroxidase | ||
TG: Lipase lipoprotein, | ||
Glycerol kinase, Glycerol-3- | ||
Phosphate Oxidase, Peroxidase | TC: Cholesterol esterase, | |
Cholesterol oxidase, Peroxidase | ||
HDL-C: Cholesterol esterase, | ||
Cholesterol oxidase, Peroxidase | ||
TG: Lipase lipoprotein, Glycerol | ||
kinase, Glycerol-3-Phosphate | ||
Oxidase, Peroxidase | ||
Color Fixatives | TC, HDL-C, TG: | |
4-aminoantipyrine | ||
Substituted aniline derivatives | TC, HDL-C: | |
4-aminoantipyrine | ||
Substituted aniline derivatives | ||
TG: | ||
4-aminoantipyrine | ||
N,N-disubstituted aniline | ||
Coding | Auto-coding by the RFID tag | MEMO Chip |
Sample type | Capillary whole blood | Capillary whole blood |
Venous whole blood | ||
Humidity range | 10~90% | 20 ~ 80% |
Power supply | 2 AAA 1.5 Volt alkaline | 2 AAA 1.5 Volt alkaline |
Battery lifetime | 1000tests | 300tests |
Test range | TC : 100-400 mg/dL | |
HDL-C: 25-80 mg/dL | ||
TG: 70-600 mg/dL | TC : 100-400 mg/dL | |
HDL-C: 25-85 mg/dL | ||
TG: 50-500 mg/dL | ||
Hematocrit range | TC, HDL-C, TG: | |
30 ~55% | TC: 30~50% | |
HDL-C: 30~45% | ||
TG: 15~55% | ||
Test time | 1-2 minutes | 1-2 minutes |
Sample volume | 5 μl | 15 μl |
Temperature range | 18-30°C (64~86°F) | 18-35°C (64.4-95°F) |
Similarities and Differences of the Lipid profile measuring system | ||
Item | LipidPro™ lipid measuring system | PTS PANELS Lipid Panel Test |
Strips (K023558) | ||
Size L x W x H (cm) | 10.9 X 6.04 X 2.29 | 13.97 X 7.62 X 2.54 |
Weight | 77.5 g | 121.9 g |
4
・
:
·
Similarities and Differences of the Blood Glucose System | ||
---|---|---|
Item | LipidPro™ glucose measuring | |
system | Evolution glucose system | |
(K072369) | ||
Detection method | Amperometry | Amperometry |
Enzyme | Glucose Oxidase | Glucose Oxidase |
Mediator | Hexaammineruthenium(III)Chloride | Hexaammineruthenium(III)Chloride |
Electrode | Carbon electrode | Carbon electrode |
Coding | Auto-coding by the color bar | Auto-coding by the color bar |
Sample type | Capillary whole blood | Capillary whole blood |
Humidity range | 10 ~ 90% | 20 ~ 80% |
Power supply | 2 AAA 1.5 Volt alkaline | 3V Li battery (CR2032×2) |
Battery lifetime | 1000 tests | 1 years |
Test range | 20 ~ 600 mg/dL | 20~600 mg/dL |
Hematocrit range | 30 ~55% | 20~60% |
Test time | 3 seconds | 3 seconds |
Sample volume | 0.3 uL | 0.3 uL |
Temperature range | 18-30°C (64~86°F) | 10 |
Memory capability | 200 tests | 7, 14, 21-day average and 365 tests |
in the memory | ||
Size L x W x H (cm) | 10.9 X 6.04 X 2.29 | 7.6 X 5.6 X 1.8 |
Weight | 77.5 g | 45g |
5
.
Similarities and Differences of the control solution | ||
---|---|---|
Item | LipidPro™ Control Solution | PTS panels multi-chemistry Control |
Solution (K022401) | ||
Indication for use | LipidPro™ Total Cholesterol | |
control solution, HDL- | ||
cholesterol control solution, | ||
Triglyceride control solution, | ||
and Glucose control solution are | ||
used to test the precision of the | ||
LipidPro™ system and to detect | ||
systematic analytic deviations | ||
that may arise from reagent or | ||
analytical variation and they are | ||
for testing outside the body (in | ||
vitro diagnostic use only.) | The PTS PANELS Multi-Chemistry | |
Controls are intended for use on the | ||
BioScanner2000 and CardioChek | ||
brand instruments to estimate | ||
precision and to detect systematic | ||
analytical deviations that may arise | ||
from reagent or analytical | ||
instrument variation and are | ||
intended for use by healthcare | ||
professionals in both physicians' | ||
offices and in acute and | ||
convalescent care facility bedside | ||
testing as well as consumers at | ||
home. | ||
Analyte | Glucose | |
TC: Total Cholesterol | ||
HDL-C: HDL-Cholesterol | ||
TG: Triglyceride | Cholesterol, Triglyceride, Glucose, | |
and Ketones | ||
Number of levels | TC: 2 Levels | |
(Normal, Abnormal) | ||
HDL-C: 2 Levels | ||
(Normal, Abnormal) | ||
TG: 2 Levels | ||
(Normal, Abnormal) | 2 Levels (Level 1, Level 2) | |
Container | Bottle | Bottle |
Color | Clear | Clear |
Temperature range | 8 | 20 |
Fill volume | 1.0mL | 1.5mL |
Matrix | TC: | |
Cholesterol, insert ingredients | ||
HDL-C: | ||
Cholesterol, insert ingredients | ||
TG: | ||
Triglyceride, insert ingredients | Glucose, Glycerol, Cholesterol, | |
DL-β-Hydroxybutyric Acid, | ||
MIT(methyl isothazoldone) | ||
and insert ingredients | ||
Target range | The expected range is printed on the Control Solution Range Card | The expected range is printed on the |
Quality Control Solution Range | ||
Card |
10. Performance Data
:
6
Clinical: The clinical performance evaluation using the LipidPro™ system was conducted for the purpose of validating the TC, HDL-C, LDL-C, TG, and Glucose measuring accuracy. Test results showed substantial equivalence.
Non-clinical: Verification, validation and testing activities were conducted to establish the performance, functionality and reliability characteristics of the LipidPro™ system with respect to the predicate device. Pass or fail criteria were based on the specifications cleared for the predicate device and results showed substantial equivalence.
11. Conclusion
The conclusion drawn from the clinical and the non clinical tests is that the LipidPro™ system is as safe, as effective, and performs as well as the legally marketed predicate device, the PTS PANELS™
7
7
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Infopia Co., Ltd. c/o Ms. Priscilla Chung Krodent DBA TMS 13340 E. Firestone Blvd., Suite J Santa Fe Springs, CA 90670
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
DEC - 1 2009
Re: Re: K090405
Trade Name: LipidPro™ system (ILM-0001A), LipidPro™ Total cholestero1 (TC) control solution, LipidPro™ High density lipoprotein cholesterol (HDL-C) control solution, LipidPro™ Triglyceride (TG) control solution and LipidPro™ glucose control solution
Regulation Number: 21 CFR §862.1345
Regulation Name: Glucose Test System.
Regulatory Class: Class II
Product Codes: CGA, NBW, JGY, CHH, LBR, JJX
Dated: November 18, 2009
Received: November 19, 2009
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
8
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Signature
Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
9
Indication for Use
510(k) Number (if known): K090405
Device Name: LipidPro™ system (ILM-0001A), LipidPro™ Total cholesterol (TC) control solution, LipidPro™ High density lipoprotein cholesterol (HDL-C) control solution, LipidPro™ Triglyceride (TG) control solution and LipidPro™ Glucose control solution
Indication For Use:
LipidPro™ system is intended for in home (self-testing) or health care professionals and for testing outside the body (in vitro diagnostic use only). LipidPro™ system which consists of meter and test strips, measures total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG) and glucose in capillary whole blood.
Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis and various liver and renal diseases. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. Calculated LDLcholesterol values are reported only when triglycerides are 350 mg/dL, calculated LDL-cholesterol are not reported.
Glucose measurement is for the quantitative measurement of the concentration of glucose in capillary whole blood that can be taken from the fingertip, ventral palm, dorsal hand, upper arm, forearm, calf and/or thigh by diabetic patients or health care professionals as an aid in the management of diabetes. Glucose measurement is not to be used for the diagnosis of or screening for diabetes or for neonatal use. Alternate site testing should be done during steady- state times when glucose is not changing rapidly.
Prescription Use (21 CFR Part 801 Subpart D) And/Or
· Over the Counter Use _ X . (21: CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carol Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K090405
10
Indication for Use
510(k) Number (if known): K090405
Device Name: LipidPro™ system (ILM-0001A), LipidPro™ Total cholesterol (TC) control solution, LipidPro " High density lipoprotein cholesterol (HDL-C) control solution, LipidPro™ Triglyceride (TG) control solution and LipidPro™ Glucose control solution
Indication For Use:
.
LipidPro™ Total Cholesterol control solution, HDL-cholesterol control solution, Triglyceride control solution, and Glucose control solution are used to test the precision of the LipidPro™ system and to detect systematic analytic deviations that may arise from reagent or analytical variation and they are for testing outside the body (in vitro diagnostic use only.) LipidPro™ control solutions are intended for in home and health care professionals use.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use X . (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carol C. Benaan
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K090405