(205 days)
The Fixion® Hip System is intended for use as a hemi-hip or total hip replacement. The Fixion® Hip System is indicated as a unipolar hemi-hip replacement in cases of:
- Femoral head and/or neck fractures or non-unions.
- Aseptic necrosis of the femoral head and/or neck.
- Osteo-, rheumatoid-, and/or post-traumatic arthritis of the hip, with minimal acetabular involvement.
The Fixion® Hip System is indicated as a total hip replacement in cases of:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
The Fixion® Hip Stem is intended for cemented or non-cemented use.
The Fixion® Hip System is composed of an implant, inflation device (pump) and an instrumentation set.
- Implant comprises femoral stem and head ball, and for total hip replacement - acetabular cup and liner and cancellous bone screws, as detailed below:
- A stainless steel Stem, including stem body and neck, which is inserted into the femoral medullary canal and expanded following insertion.
- A Head Ball, assembled on the stem neck and articulates within the acetabulum (in hemi-hip replacement) or acetabular cup and liner (in total hip replacement). A unipolar stainless steel head (for hemi- hip replacement), and a cobalt chromium head (for total hip replacement) are available.
- A porous coated, cobalt chromium Acetabular Cup, inserted and fixated within the acetabulum.
- A UHMWPE Acetabular Cup Liner, which is seated inside the acetabular cup.
- Cobalt chromium Cancellous Bone Screws, intended to fixate the acetabular cup to the acetabulum, in cases additional stabilization of the cup is required.
- Instrumentation Set - a set of accessories to be used with the Fixion® Hip implants.
- Inflation Device (Pump) - a manual pump used to expand the Fixion® stem.
The provided text describes the 510(k) summary for the Disc-O-Tech Medical Technologies Ltd. Fixion® Hip System, which is a medical device application. Medical device regulatory submissions like 510(k) summaries do not typically include the detailed study design elements you are asking for, such as specific acceptance criteria and detailed performance reporting from a clinical study, sample sizes for test and training sets, ground truth establishment methods, or multi-reader multi-case (MRMC) study results with effect sizes.
Instead, 510(k) submissions demonstrate "substantial equivalence" to legally marketed predicate devices. This is typically achieved through:
- Comparison of technological characteristics: Showing that the new device has the same intended use, design features, materials, and operating principles as one or more predicate devices.
- Performance testing (benchtop/mechanical): Ensuring the device meets relevant industry consensus standards (e.g., ISO, ASTM) to demonstrate safety and effectiveness for its intended use, often through engineering tests rather than clinical trials with human subjects.
- Biocompatibility testing: As required.
The document states: "The Fixion® Hip System intended use, design, materials, technological characteristics and principles of operation are substantially equivalent to those of Disc-O-Tech's Fixion® Unipolar Modular Hemi-Hip System (K014072, K030972), Fixion® Interlocking PF Intramedullary Nailing System (K010988, K012967, K023437), United Orthopedic Corporation' U1 Hip System (K994078 Orthopaedics' Elite® Modular Hip System (K871867) and S-ROM® Total Hip System (K851422, K913231, K954935, K961939 and others), and the Link America Inc.' Link® MP® Hip System (K955296)."
It also lists several Performance Standards (ISO and ASTM) that were referred to. These standards define the mechanical and material properties that such an implant should meet. Meeting these standards comprises the "acceptance criteria" for the mechanical performance of the device.
Therefore, I cannot provide the specific details you requested because they are not typically part of a 510(k) summary for a hip implant system, which focuses on substantial equivalence rather than a clinical study demonstrating performance against a detailed set of acceptance criteria with human subjects or AI algorithms.
However, I can extract the closest analogous information from the document regarding acceptance criteria and performance:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Standards Referred) | Reported Device Performance |
|---|---|
| ISO 7206-1 (Draft, 1995): Implants for Surgery - Partial and Total Hip Joint Prostheses - Part 1: Classification and Description of Dimensions. | The device's dimensions and classification conform to this standard, implying its design and size ranges are within accepted specifications for hip prostheses. |
| ISO 7206-4 (Draft, 1999): Implants for Surgery Partial and Total Hip Joint Prostheses - Part 4: Determination of Endurance Properties of Stemmed Femoral Components. | The device's stemmed femoral components meet the endurance properties defined by this standard, indicating sufficient fatigue strength under cyclic loading conditions experienced in the body. |
| ISO 7206-8 (1995): Implants for Surgery Partial and Total Hip Joint Prostheses - Part 8: Endurance Performance of Stemmed Femoral Components with Application of Torsion. | The device's stemmed femoral components demonstrate adequate endurance performance under torsional loads, ensuring stability and integrity during hip movement. |
| Guidance Document for Femoral Stem Prostheses (Draft), ORDB/DGRD/CDRH/FDA, August 1, 1995. | The device's femoral stem components comply with the recommendations outlined in this FDA guidance, ensuring its design and performance address specific regulatory concerns for safety and effectiveness. |
| Guidance Document for Testing Non-Articulating, "Mechanically Locked", Modular Implant Components (Draft), ORDB/DGRD/CDRH/FDA, May 1, 1995. | The modular components of the device (e.g., head-neck junction, cup/liner assembly) meet the testing requirements for mechanical locking mechanisms, assuring their secure and stable assembly. |
| ASTM F 138-2000: Standard Specification for Stainless Steel Bar and Wire for Surgical Implants. | The stainless steel components (e.g., stem, unipolar head) are made from materials conforming to this standard, ensuring appropriate mechanical properties and biocompatibility for surgical implants. |
| ASTM F 565-2000: Standard Practice for Care and Handling of Implants and Instruments. | The labeling and instructions for the device align with this standard, providing proper guidance for its care and handling to maintain sterility and integrity prior to and during implantation. |
| ASTM F 1586-95: Wrought Nitrogen Strengthened 21 Chromium 10 Nickel - 3 Manganese - 2.5. Molybdenum Stainless Steel Bar for Surgical Implants. | Where applicable, stainless steel components are fabricated from materials meeting this specific standard, indicating use of high-strength, corrosion-resistant steel. |
| ISO 5832-1 (1997): Implants for Surgery Metallic Materials Part 1: Wrought High Nitrogen Stainless Steel. | Where applicable, metallic components are made from materials conforming to this standard. |
| ISO 5832-9 (1995): Implants for Surgery Metallic Materials Part 9: Wrought High Nitrogen Stainless Steel. | Where applicable, metallic components are made from materials conforming to this standard. |
| BS EN 12563: 1999: Non-active surgical implant Joint replacement implants - Specific requirements for hip joint replacement implants. | The device meets the specific requirements outlined in this European standard for hip joint replacement implants, covering various aspects of design, materials, and performance. |
Note: The "reported device performance" is inferred from the statement of "substantial equivalence" and the reference to these standards. Regulatory bodies accept that if a device design, materials, and mechanical properties meet these established standards, it is inherently performing acceptably for its intended use, similar to predicate devices that have also met these standards.
The following information cannot be extracted from the provided text for the reasons explained above, as this document is a 510(k) summary focused on substantial equivalence rather than a clinical trial report:
- Sample size used for the test set and the data provenance.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts.
- Adjudication method for the test set.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
- The sample size for the training set.
- How the ground truth for the training set was established.
{0}------------------------------------------------
510(K) Summary MAR 2 6 2004
Disc-O-Tech Medical Technologies Ltd. Fixion® Hip System
Company Name
Disc-O-Tech Medical Technologies Ltd.
3 Hasadnaot St.
Herzliya 46728, Israel
Submitter's Name and Contact Person
Hila Wachsler-Avrahami Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St. Herzliya 46728, Israel Tel: + 972-9-9511511 Fax: + 972-9-9548939
Date Prepared
August 2003
Trade/Proprietary Name
Fixion® Hip System
Classification
Class II
{1}------------------------------------------------
Predicate Devices
公
Fixion® Unipolar Modular Hemi-Hip System (Fixion® MH, K014072, K030972) by Disc-O-Tech Medical Technologies Ltd.
- U1 Hip System (K994078) by United Orthopedic Corporation. ﺮ
- Elite® Modular Hip System (K871867) by Depuy Orthopeadics. ア
- S-ROM® Total Hip System (K851422, K913231, K954935, K961939) by Depuy A Orthopeadics.
- Link® MP® Hip System (K955296) by Link America Inc. 公
- Fixion® Interlocking Proximal Femoral Intramedullary Nailing System 公 (Fixion® PF, K010988, K012967, K023437) by Disc-O-Tech Medical Technologies Ltd.
Performance Standards
The following standards were referred:
- ン ISO 7206-1 (Draft, 1995): Implants for Surgery - Partial and Total Hip Joint Prostheses - Part 1: Classification and Description of Dimensions.
- ISO 7206-4 (Draft, 1999): Implants for Surgery Partial and Total Hip Joint 公 Prostheses - Part 4: Determination of Endurance Properties of Stemmed Femoral Components.
- ISO 7206-8 (1995): Implants for Surgery Partial and Total Hip Joint A Prostheses - Part 8: Endurance Performance of Stemmed Femoral Components with Application of Torsion.
- ア Guidance Document for for Femoral Stem Prostheses (Draft), ORDB/DGRD/CDRH/FDA, August 1, 1995.
- 公 Guidance Document for Testing Non-Articulating, "Mechanically Locked", Modular Implant Components (Draft), ORDB/DGRD/CDRH/FDA, May 1, 1995.
- ASTM F 138-2000: Standard Specification for Stainless Steel Bar and Wire ﺮ for Surgical Implants.
3 11asadnao! St., Herzliya 46728, Israel; Tel: 972-9511511; Fax: 972-9-9548939; F-mail: hila@disc-o-tech.com Page 19 of 23
{2}------------------------------------------------
- ASTM F 565-2000: Standard Practice for Care and Handling of تز Implants and Instruments.
- ASTM F 1586-95: Wrought Nitrogen Strengthened 21 Chromium 10 A Nickel - 3 Manganese - 2.5. Molybdenum Stainless Steel Bar for Surgical Implants.
- ISO 5832-1 (1997): Implants for Surgery Metallic Materials Part 1: ﺮ Wrought High Nitrogen Stainless Steel.
- ISO 5832-9 (1995): Implants for Surgery Metallic Materials Part 9: V Wrought High Nitrogen Stainless Steel.
- BS EN 12563: 1999 Non-active surgical implant Joint replacement ア implants - Specific requirements for hip joint replacement implants.
Intended Use
The Fixion® Hip System is intended for use as a hemi-hip or total hip replacement. The Fixion® Hip System is indicated as a unipolar hemi-hip replacement in cases of:
- ア Femoral head and/or neck fractures or non-unions.
- Aseptic necrosis of the femoral head and/or neck. ﺮ
- Osteo-, rheumatoid-, and/or post-traumatic arthritis of the hip, with V minimal acetabular involvement.
The Fixion® Hip System is indicated as a total hip replacement in cases of:
- Painful, disabling joint disease of the hip resulting from: degenerative A arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup ﺮ arthroplasty or other procedure.
- A management problems where arthrodesis or alternative Clinical reconstructive techniques are less likely to achieve satisfactory results.
The Fixion® Hip Stem is intended for cemented or non-cemented use.
3 Fasadnaot St., Herzliya 46728, Jsrael, Tel: 972-9-9511511; Fax: 972-9-9548939; E-mail: hila@disc-o-tech.com Page 20 of 23
{3}------------------------------------------------
page 4 of 5
K032730
System Description
The Fixion® Hip System is composed of an implant, inflation device (pump) and an instrumentation set.
- Implant comprises femoral stem and head ball, and for total hip >> replacement - acetabular cup and liner and cancellous bone screws, as detailed below:
- A stainless steel Stem, including stem body and neck, which is inserted . into the femoral medullary canal and expanded following insertion.
- A Head Ball, assembled on the stem neck and articulates within the . acetabulum (in hemi-hip replacement) or acetabular cup and liner (in total hip replacement). A unipolar stainless steel head (for hemi- hip replacement), and a cobalt chromium head (for total hip replacement) are available.
- A porous coated, cobalt chromium Acetabular Cup, inserted and . fixated within the acetabulum.
- A UHMWPE Acetabular Cup Liner, which is seated inside the . acetabular cup.
- Cobalt chromium Cancellous Bone Screws, intended to fixate the . acetabular cup to the acetabulum, in cases additional stabilization of the cup is required.
- ﯩﺰ Instrumentation Set - a set of accessories to be used with the Fixion® Hip implants.
-
Inflation Device (Pump) - a manual pump used to expand the Fixion® stem.
Substantial Equivalence
Hip System intended use, design, materials, technological The Fixion® characteristics and principles of operation are substantially equivalent to those of Disc-O-Tech's Fixion® Unipolar Modular Hemi-Hip System (K014072, K030972), Fixion® Interlocking PF Intramedullary Nailing System (K010988, K012967,
3 Hasadnaot St., Herzliya 46728, Israel; Tel: 972-9511511; Fax: 972-9-9548939; E-mail: hila@disc-o-tech.com; www.disc-o-tech.com Page 21 of 23
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the text "page 5 of 5" and "K032730" written in black ink. The text "page 5 of 5" is underlined. The text appears to be handwritten.
K023437), United Orthopedic Corporation' U1 Hip System (K994078 Orthopaedics' Elite® Modular Hip System (K871867) and S-ROM® Total Hip System (K851422, K913231, K954935, K961939 and others), and the Link America Inc.' Link® MP® Hip System (K955296).
3 Hasadnavt St., Herzliya 46728, Israel; Tel: 972-9-9511511; Fax: 972-9-9548939; E-mail: hila@disc-o-tech.com; www.disco-tech.com Page 22 of 23
==============================================================================================================================================================================
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three stripes on its wings, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 6 2004
Ms. Hila Wachsler-Avrahami Regulatory Affairs Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot Street Herzliya 46728, Israel
Re: K032730 Trade/Device Name: Fixion® Hip System Regulation Numbers: 21 CFR 888.3350; 21 CFR 888.3358; 21 CFR 888.3360 Regulation Names: Hip joint metal/polymer semi-constrained cemented prosthesis; Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis; Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis. Regulatory Class: II
Product Codes: JDI, LPH, MBL, KWL, JDG, LWJ Dated: December 25, 2003 Received: December 29, 2003
Dear Ms. Wachsler-Avrahami:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Mark N Milliman
'elia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indication for Use Mark N. Milliner
Division Sign Off:
510(K) Number (if known): K032730
Device Name: Fixion® Hip System
Indication for Use:
Division of General, Restorative,
K032730
510(a) Number
The Fixion® Hip System is intended for use as a hemi-hip or total hip replacement.
The Fixion® Hip System is indicated as a unipolar hemi-hip replacement in cases of:
- A Femoral head and/or neck fractures or non-unions.
- A Aseptic necrosis of the femoral head and/or neck.
- A Osteo-, rheumatoid-, and/or post-traumatic arthritis of the hip, with minimal acetabular involvement.
The Fixion® Hip System is indicated as a total hip replacement in cases of:
- A Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup A arthroplasty or other procedure.
- A Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
The Fixion® Hip Stem is intended for cemented or non-cemented use.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
3 Hasadnaot St., Herzliya 46728, Israel; Tel: 972-9-9511511; Fax: 972-9-9548939; E-mail: hila@disc-o-tech.com Page 16 of 23
§ 888.3350 Hip joint metal/polymer semi-constrained cemented prosthesis.
(a)
Identification. A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.