AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fixion Interlocking Proximal Femoral Nailing System ("Fixion PF") is intended for use in fixation of fractures in the femur. The Fixion PF is indicated for use in fractures in the femur shaft, proximal fractures, and combinations of these fractures. Proximal femoral fractures include stable and unstable pertrochanteric, intertrochanteric and subtrochanteric (with and without break-off of the minor trochanter), high subtrochanteric fractures and combinations of these fractures. The long Fixion PF may also be used in mid shaft fractures, 5cm below the surgical neck to 5cm proximal to the distal end of the medullar canal.

The Fixion PF is also indicated for use in osteotomy, nonunions and malunions, bone reconstruction following tumor resection, grafting and pathological fractures, revision procedures.

Device Description

The Fixion Intramedullary Proximal Femoral Nailing System consist of the following main components:

  1. The Nail Implant is an expandable non-slotted stainless steel cylindrical tube, with a cap protected, female threaded proximal end with holes for Femoral Neck Peg (Hip Peg) and Hip Pin. Other implantable components provided are the Femoral Neck Peg (Hip Peg) and the Hip Pin.

  2. The Instrument Set consists of a few accessories used during insertion and removal (if required) of the implant.

  3. The Inflation Device (Pump) is a single-use manual plastic pump that is filled with sterile inflation liquid and used to expand the Nail implant and Femoral Neck (Hip) Peg

Once the Nail and the Femoral Neck Peg are positioned within the medullary canal and femur neck respectively, rotation of the pump handle allows for peg and nail diameter increase to their intended diameter under x-ray and controlled pressure. The Hip Pin may be inserted as well.

AI/ML Overview

The provided text describes a 510(k) summary for a medical device (Fixion™ Interlocking Proximal Femoral Intramedullary Nailing System), which focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through clinical studies. Therefore, much of the requested information (e.g., acceptance criteria table, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types) is not available in the given document.

The document primarily outlines:

  • The device's description and intended use.
  • Its classification and predicate devices.
  • The performance standards it meets (primarily ASTM standards for material composition and mechanical testing).
  • The claim of substantial equivalence to a previously cleared device (K010988) based on similar intended use, operating principles, design, materials, processes, sterilization, and packaging.

Based on the information provided, here is what can be inferred or stated:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Material Composition Standards
ASTM F138 - Standard Specification for Stainless steel Bar and Wire for Surgical Implants (316L Stainless Steel)The Fixion PF Nail is manufactured from 316L Stainless Steel, which meets the requirements of ASTM F138.
Surgical grade stainless steel and silicone for accessoriesThe Fixion PF Nailing System accessories incorporate surgical grade stainless steel and silicone.
Handling and Design Standards
ASTM F565 Standard practice for Care and Handling of Orthopedic Implants and InstrumentsThe Fixion PF Nail is designed to meet the requirements of ASTM F565.
Mechanical Performance Standards
ASTM F1264-99 - Standard for Mechanical Performance Considerations for Intramedullary Fixation Devices (4 point bending)4 point bending mechanical testing was performed according to ASTM F1264-99. (Specific performance values not provided, but implies meeting the standard).
ASTM F384-99 Standard Specification for Metallic Angled Orthopedic Fracture Fixation Devices (Hip Peg testing)Hip Peg testing was performed according to ASTM F384-99. (Specific performance values not provided, but implies meeting the standard).
Substantial Equivalence to Predicate Device (K010988)
Same intended useThe modified Fixion PF Nail has the same intended use.
Same operating principlesBasically has the same operating principles.
Same designBasically incorporates the same design.
Same materials and processesIncorporates the same materials and processes.
Same sterilization and packagingIs sterilized and packed, basically, in the same manner.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document refers to mechanical testing against ASTM standards, but does not detail sample sizes for these tests, nor does it refer to human clinical test sets.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. The assessment relies on engineering and material standards, not expert clinical ground truth from a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided. Not applicable to the type of data presented (mechanical and material compliance).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A MRMC comparative effectiveness study was not conducted and is not applicable to this type of device (an orthopedic implant) which does not involve "human readers" or "AI assistance" in its direct function. The device is a surgical implant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

A standalone performance study of an algorithm was not conducted and is not applicable to this device. The device is a physical intramedullary nailing system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's acceptance is based on:

  • Compliance with ASTM material standards (e.g., F138 for stainless steel).
  • Compliance with ASTM mechanical performance standards for orthopedic devices (e.g., F1264-99 for intramedullary fixation, F384-99 for hip pegs).
  • Demonstration of substantial equivalence to existing legally marketed predicate devices, implying similar safety and effectiveness based on similar design, materials, and intended use.

8. The sample size for the training set

This information is not provided and is not applicable. There is no "training set" in the context of this device's mechanical and material compliance testing or substantial equivalence demonstration.

9. How the ground truth for the training set was established

This information is not provided and is not applicable as there is no training set mentioned or relevant for this type of device submission.

{0}------------------------------------------------

OCT - 4 2001

510(K) Summary

Disc-O-Tech Medical Technologies, Ltd. Fixion Interlocking Proximal Femoral Intramedullary Nailing System

Company Name

Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St., Herzelia, 46728, Israel

Submitter's Name and Contact Person

Elad Magal Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St., Herzelia, 46728, Israel Tel: +972 9 9511511 Fax: +972 9 9548939

Date Prepared

August 2001

Trade/Proprietary Name

Fixion™ Interlocking Proximal Femoral Intramedullary Nailing System (Fixion PF)

Classification Name

Intramedullary Fixation Rod 21 CFR § 888.3020 Class II

Predicate Devices

    1. Fixion Interlocking PF Intramedullary Nailing System (K010988) by Disc-O-Tech
    1. Fixion Intramedullary Nailing System (K990717, K003212, K003215, K010901) by Disc-O-Tech
    1. Fixion Interlocking Intramedullary Nailing System (K002783) by Disc-O-Tech.
    1. Proximal Femoral Nail (K970097) by Synthes.

Performance Standards

The following standards were used:

    1. The Fixion PF Nail is manufactured from 316L Stainless Steel, which meets the requirements of ASTM F138 - Standard Specification for Stainless steel Bar and Wire for Surgical Implants.
    1. The Fixion PF Nailing System accessories incorporate surgical grade stainless steel and silicone.
    1. The Fixion PF Nail is designed to meet the requirements of ASTM F565 Standard practice for Care and Handling of Orthopedic Implants and Instruments.
    1. The 4 point bending mechanical testing was performed according to ASTM

{1}------------------------------------------------

for Mechanical Performance Considerations for F1264-99 - - Standard Intramedullary Fixation Devices.

    1. The Hip Peg testing was performed according to ASTM F384-99 Standard Specification for Metallic Angled Orthopedic Fracture Fixation Devices.

Intended Use

The Fixion Interlocking Proximal Femoral Nailing System ("Fixion PF") is intended for use in fixation of fractures in the femur. The Fixion PF is indicated for use in fractures in the femur shaft, proximal fractures, and combinations of these fractures. Proximal femoral fractures include stable and unstable pertrochanteric, intertrochanteric and subtrochanteric (with and without break-off of the minor trochanter), high subtrochanteric fractures and combinations of these fractures. The long Fixion PF may also be used in mid shaft fractures, 5cm below the surgical neck to 5cm proximal to the distal end of the medullar canal.

The Fixion PF is also indicated for use in osteotomy, nonunions and malunions, bone reconstruction following tumor resection, grafting and pathological fractures, revision procedures.

System Description

The Fixion Intramedullary Proximal Femoral Nailing System consist of the following main components:

  1. The Nail Implant is an expandable non-slotted stainless steel cylindrical tube, with a cap protected, female threaded proximal end with holes for Femoral Neck Peg (Hip Peg) and Hip Pin. Other implantable components provided are the Femoral Neck Peg (Hip Peg) and the Hip Pin.

  2. The Instrument Set consists of a few accessories used during insertion and removal (if required) of the implant.

  3. The Inflation Device (Pump) is a single-use manual plastic pump that is filled with sterile inflation liquid and used to expand the Nail implant and Femoral Neck (Hip) Peg

Once the Nail and the Femoral Neck Peg are positioned within the medullary canal and femur neck respectively, rotation of the pump handle allows for peg and nail diameter increase to their intended diameter under x-ray and controlled pressure. The Hip Pin may be inserted as well.

Substantial Equivalence

The Fixion PF Nail is substantially equivalent to the Fixion PF Nail currently cleared for marketing under 510(k) K010988.

The modified Fixion PF Nail has the following similarities to that which previously received 510(k) concurrence:

  • Has the same intended use

  • Basically has the same operating principles

  • Basically incorporates the same design

{2}------------------------------------------------

  • 》 Incorporates the same materials and processes
  • Is sterilized and packed, basically, in the same manner.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three curved lines resembling a bird or a wing-like shape.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT - 4 2001

Mr. Elad Magal General Manager Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot Street Herzelia, 46728 Israel

Re: K012967

Trade/Device Name: Fixion™ Interlocking Proximal Femoral Intramedullary Nailing System Regulation Number: 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Code: HSB Dated: August 30, 2001 Received: September 4, 2001

Dear Mr. Magal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Elad Magal

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Sincerely, yours,

L. Mark N. Wilkerson

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication for Use

510(K) Number (if known): K012967

Device Name: Fixion™ Interlocking Proximal Femoral Intramedullary Nailing System

Indication for Use:

The Fixion Interlocking Proximal Femoral Nailing System ("Fixion PF") is intended for The I chon intervociang 1 rouman estur. The Fixion PF is indicated for use in fractures in the femur shaft, proximal femoral fractures, and combinations of these fractures. Proximal femoral fractures include stable and unstable pertrochanteric, intertrochanteric and subtrochanteric (with and without break-off of the minor trochanter), high subtrochanteric Subtrochanterie (Mini timse fractures. The long Fixion PF may also be used in mid shaft fractures, 5cm below the surgical neck to 5cm proximal to the distal end of the medullar canal.

The Fixion PF is also indicated for use in osteotomy, nonunions and malunions, bone The I baon 11 - b abound tumor resection, grafting and pathological fractures, revision procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(per 21 CFR 801.109)

OR

Over the Counter Use _________________________________________________________________________________________________________________________________________________________

Page 22 of 27

for Mark N. Miller

and Neurological Device

CONFIDENTIAL-

510(k) Number.

§ 888.3020 Intramedullary fixation rod.

(a)
Identification. An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.(b)
Classification. Class II.