K Number
K014203
Manufacturer
Date Cleared
2002-02-28

(69 days)

Product Code
Regulation Number
864.5200
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ABX MICROS 60 Hematology Analyzer is a fully automated (microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood specimens or blood cell concentrates. It operates in complete blood count (CBC) mode.

Device Description

The ABX MICROS 60 Hematology Analyzer is a bench top clinical laboratory instrument which analyzes in vitro samples of whole blood to provide complete blood count data using principles of cytochemistry, focused flow impedance, and light transmission technology. The system is equipped with automatic calibration; Automated smart cards based for: calibration, QC with 93 multiple range files and memory with 60 erasable results.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the ABX MICROS 60 Hematology Analyzer:

Due to the nature of the provided document (a 510(k) summary for a legacy device from 2001), some of the requested information (especially regarding modern AI/ML study practices like multi-reader multi-case studies, detailed ground truth establishment for training, or specific acceptance criteria tables with numerical thresholds) is not explicitly present in the document. The document focuses on demonstrating substantial equivalence to a predicate device and broadly describing performance, rather than detailing a rigorous, pre-defined acceptance criteria study in the way a modern AI/ML device would.

However, I will extract and present all available information and highlight what is missing based on your request.


Acceptance Criteria and Device Performance Study for ABX MICROS 60 Hematology Analyzer

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria in a table format. Instead, it describes performance characteristics and concludes that they are "acceptable" or "highly correlated." The predicate device comparison implies that the ABX MICROS 60 performance should be at least comparable to the BAKER SYSTEM 9110* PLUS.

Performance MetricAcceptance Criteria (Implied/General)Reported Device Performance
PrecisionAcceptable within-run, between-run, and between-day precision- Within-run, Between-run, and Between-day precision: Acceptable. Highest CV (Coefficient of Variation) was 3.0% for high-level control tested 10 times successively on a single day.
LinearityAcceptable linearity across the measuring range- Linearity: Quite linear from 250 to 5000 x 10^3/μL (for platelets, inferred).
Correlation with Predicate DeviceHighly correlated with the predicate device- Correlation: Results highly correlated with those from the Baker System 9110+ PLUS, although values tend to be somewhat lower.
Carry-overNo significant carry-over- Carry-over: No carry-over when samples with low platelet counts are assayed after those with substantially higher counts.
Overall ConclusionProvides reliable data for quantitating platelets in specified range (250 - 5000 10^3/µL)- Reliability: Provides reliable data when used to quantitate the number of platelets in preparations having relatively high concentrations of this analyte (250 - 5000 10^3/µL range).

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the test set. It mentions testing a "high level control 10 times in succession on a single day" for precision, but this is not the full test set.

The data provenance is also not explicitly stated. It can be inferred that the testing was conducted prospectively as part of the device development and submission process. The country of origin of the data is not mentioned, but given the company's US address (Irvine, CA), it is likely to be from the USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. For a hematology analyzer measuring objective parameters like cell counts, the "ground truth" would typically be established by established reference methods, manual microscopy by trained laboratory personnel, or internal laboratory standards and controls, rather than a panel of "experts" in the context of image interpretation.

4. Adjudication Method for the Test Set

This information is not applicable/provided. Adjudication methods (like 2+1, 3+1) are typically used for subjective assessments (e.g., image interpretation by multiple readers) to establish a consensus ground truth. For an automated hematology analyzer measuring numerical parameters, the "ground truth" would likely come from an objective reference method or a reference instrument, not through human adjudication of differing interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study (comparing human readers with and without AI assistance) is relevant for AI/ML devices where the algorithm interacts with human interpretation. The ABX MICROS 60 is a standalone automated analyzer, not an AI-assisted diagnostic tool.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the study described is a standalone performance study. The ABX MICROS 60 Hematology Analyzer is an automated device designed to measure parameters directly. The performance data presented (precision, linearity, correlation, carry-over) reflects the direct output of the instrument without human intervention in the primary measurement process itself.

7. The Type of Ground Truth Used

The ground truth for this device would have been established through a combination of:

  • Reference Methods: Comparison to established standard laboratory methods for cell counting. While not explicitly stated, this is standard practice.
  • Predicate Device Comparison: The study explicitly mentions correlation with the BAKER SYSTEM 9110+ PLUS, which serves as a type of "ground truth" or reference for establishing substantial equivalence.
  • Calibrators and Controls: Mention of "high level control" indicates the use of known reference materials.

Pathology reports or outcomes data are generally not the direct ground truth for individual numerical cell counts from a hematology analyzer, although clinical outcomes can validate the utility of the parameters measured. Expert consensus is also less direct for objective numeric measurements, compared to subjective interpretations.

8. The Sample Size for the Training Set

The document does not mention a "training set". This term is primarily associated with machine learning or artificial intelligence devices that are trained on data. The ABX MICROS 60, being a 2001 device, is a classical, rule-based or impedance-based instrument, not an AI/ML device that undergoes a training phase in the modern sense. Its development would have involved extensive calibration and internal validation rather than training on a separate dataset.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a "training set" in the context of an AI/ML device, this question is not applicable to the information provided. The "ground truth" for the development and calibration of such a device would have been established through traditional laboratory methods, reference instruments, and certified control materials during its engineering and validation phases.

{0}------------------------------------------------

KD14203

Premarket Notification [510(k)] Summary SECTION B

  • ABX Diagnostics, Inc. Company: 34 Bunsen Drive Irvine, CA 92618 Phone: (949) 453-0500 FAX: (949) 453-0600 Date Prepared: December 7, 2001
    Contact Person: Tom M. Phillips

Date of Preparation: December 19, 2001

Trade/Proprietary Name:ABX MICROS 60 Hematology Analyzer
Common or Usual Name:Automated cell counter andAutomated differential cell counter
Classification Name:Automated cell counter (864.5200) andAutomated differential cell counter (864.5220)
Product Code:GKZ (Hematology Analyzer)
Substantial Equivalence:The ABX MICROS 60 is substantially equivalent to theBAKER SYSTEM 9110⁺ PLUS Hematology Analyzer
Description:The ABX MICROS 60 Hematology Analyzer is abench top clinical laboratory instrument which analyzesin vitro samples of whole blood to provide completeblood count data using principles of cytochemistry,focused flow impedance, and light transmissiontechnology. The system is equipped with automaticcalibration; Automated smart cards based for:calibration, QC with 93 multiple range files andmemory with 60 erasable results.
Indication for Use:The ABX MICROS 60 Hematology Analyzer is a fullyautomated (microprocessor controlled) hematologyanalyzer used for the in vitro diagnostic testing of wholeblood specimens or blood cell concentrates. It operates incomplete blood count (CBC) mode.

Comparison to the Predicate Device:

The ABX MICROS 60 Hematology Analyzer is equivalent to the BAKER SYSTEM 9110* PLUS Hematology Analyzer; both instruments are fully automated hematology analyzers using whole blood to measure WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, and MPV. The ABX MICROS 60 also measures differential

{1}------------------------------------------------

counts whereas the BAKER SYSTEM 9110* PLUS does not.

Discussion of Performance Data:

Data provided in the 510(k) application demonstrate that within-run, between-run, and between-day precision for the ABX MICROS 60 is acceptable, with the highest CV (3.0%) being obtained for the high level control tested 10 times in succession on a single day, (B) the assay is quite linear from 250 to 5000 x 103/μL, (C) results from the ABX MICROS 60 assay are highly correlated with those derived from the Baker System 9110+ PLUS, although the values tend to be somewhat lower, and (D) there is no carry-over when samples having low platelet counts are assayed after those with substantially higher counts.

The ABX MICROS 60 Hematology Analyzer provides Conclusion: reliable data when used to quantitate the number of platelets in preparations having relatively high concentrations of this analyte (250 - 5000 103/µL range).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - IRA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - IRA" arranged around the perimeter of the circle. Inside the circle is a stylized image of what appears to be an eagle or other bird with three lines representing the body and wings.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

FEB 2 8 2002

Mr. Tom M. Phillips Manager of Regulatory Affairs ABX Diagnostics, Inc. 34 Bunsen Drive Irvine, CA 92618

Re: K014203

Trade/Device Name: ABX Micros 60 Hematology Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Dated: December 19, 2001 Received: December 21, 2001

Dear Mr. Phillips:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 more prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE STATEMENT

K014203 510(k) Number (if known):

Device Name: ABX MICROS 60 Hematology Analyzer

Indications For Use:

The ABX MICROS 60 Hematology Analyzer is a fully automated (microprocessor The ADA MICROD ov analyzer used for the in vitro diagnostic testing of whole Controlled) nomatorogy and your accurates. It operates in complete blood count (CBC) mode.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801. 109)

OR

Over-The-Counter Use

Josephine Bautista

510(k) Number K014203

§ 864.5200 Automated cell counter.

(a)
Identification. An automated cell counter is a fully-automated or semi-automated device used to count red blood cells, white blood cells, or blood platelets using a sample of the patient's peripheral blood (blood circulating in one of the body's extremities, such as the arm). These devices may also measure hemoglobin or hematocrit and may also calculate or measure one or more of the red cell indices (the erythrocyte mean corpuscular volume, the mean corpuscular hemoglobin, or the mean corpuscular hemoglobin concentration). These devices may use either an electronic particle counting method or an optical counting method.(b)
Classification. Class II (performance standards).