K Number
K992875

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-11-09

(75 days)

Product Code
Regulation Number
864.5220
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sysmex™ XE-2100 is a multi-parameter hematology analyzer intended to classify the following formed elements in anti-coagulated blood: WBC, Neut%/#, Lymph%/#, Mono%/#, Eo%/#, Baso%/#, NRBC%/#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, RET%/#, IRF, HFR*, MFR*, LFR*, PLT, MPV, PDW*, P-LCR*, PCT* (*Not Reportable in USA).

Device Description

The XE-2100 is an automated hematology analyzer which consists of four principal units: (1) Main Unit which aspirates, dilutes, mixes, and analyzes whole blood samples; (2) Sampler Unit which supplies samples to the Main Unit automatically; (3) IPU (Information Processing Unit) which processes data from the Main Unit and provides the operator interface with the system; (4) Pneumatic Unit which supplies pressure and vacuum from the Main Unit.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the Sysmex™ Automated Hematology Analyzer XE-2100, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary does not explicitly state numerical acceptance criteria for the performance of the Sysmex™ XE-2100. Instead, it relies on demonstrating substantial equivalence to predicate devices (Sysmex SE/RAM-1 and SF-3000) and established methods (manual differential, flow cytometry).

The reported device performance is described as "performance to manufacturer specifications" for carryover, precision, linearity, and sample stability. For other parameters, performance is considered "similar to the predicate devices."

Feature/ParameterAcceptance Criteria (Implied)Reported Device Performance
Overall DeviceSubstantial equivalence to predicate devices (SE/RAM-1, SF-3000), manual differential, and flow cytometry.Performance claims are similar to predicate devices. Supports substantial equivalence.
CarryoverManufacturer specificationsPerformance met manufacturer specifications
PrecisionManufacturer specificationsPerformance met manufacturer specifications
LinearityManufacturer specificationsPerformance met manufacturer specifications
Sample StabilityManufacturer specificationsPerformance met manufacturer specifications
Correlation Studies (All parameters listed for XE-2100)Results expected to be similar to predicate devices for relevant parameters.Specimens from healthy individuals and pathological conditions showed similar results to predicate devices.
WBC DifferentialCorrelation with manual differentials (NCCLS H20A) and flow cytometry.Correlated to results from manual differentials (NCCLS H20A) and flow cytometry.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document does not explicitly state the specific number of samples used in the correlation studies. It mentions that "specimens were evaluated from apparently healthy individuals and from patients with different pathological conditions." This implies a varied and likely representative sample, but the exact count is not given.
  • Data Provenance: The document does not state the country of origin of the data. The studies appear to be prospective as they were conducted to evaluate the performance of the XE-2100 against predicate devices and established methods.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: The document does not specify the number of experts used.
  • Qualifications of Experts: The document states that the WBC differential of the XE-2100 was correlated to results from "manual differentials performed according to NCCLS H20A." This implies that the manual differentials, which serve as a form of ground truth, were performed by trained laboratory professionals, adhereing to a recognized standard (NCCLS H20A). No specific experience level (e.g., radiologist with 10 years experience) is mentioned for these individuals.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (like 2+1 or 3+1) for establishing ground truth. The comparison for WBC differential was made against "manual differentials performed according to NCCLS H20A and to flow cytometry." This suggests a direct comparison to established methods rather than a consensus-based adjudication by multiple independent experts of the device's output.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No multi-reader multi-case (MRMC) comparative effectiveness study is mentioned in the provided text. The evaluation focuses on the device's performance against predicate devices and established laboratory methods, not on how the device assists human readers.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, the study described is a standalone performance evaluation. The Sysmex XE-2100 is an "Automated Hematology Analyzer," and the studies (carryover, precision, linearity, sample stability, and correlation studies) evaluate the device's output directly against reference methods or predicate devices, without human-in-the-loop interaction as part of the primary performance assessment.

7. Type of Ground Truth Used

The ground truth used in the studies includes:

  • Established Laboratory Methods:
    • Manual differentials performed according to NCCLS H20A for WBC differential.
    • Flow cytometry for WBC differential.
  • Predicate Devices: The Sysmex SE/RAM-1 and SF-3000 served as a comparative reference for many parameters. While not strictly "ground truth," their established performance was used as a benchmark for substantial equivalence.

8. Sample Size for the Training Set

The document does not provide any information about a training set or its sample size. This is typical for devices of this era, especially those demonstrating equivalence to predicate devices, where extensive AI/machine learning training sets as understood today were not a primary requirement for 510(k) submission. The device's operation is based on established analytical principles (e.g., flow cytometry, DC detection).

9. How the Ground Truth for the Training Set Was Established

As no training set is mentioned, there is no information on how its ground truth would have been established.

{0}------------------------------------------------

K992875

NOV - 9 1999

.

510(k) Summary of Safety and Effectiveness Information SYSMEX™ Automated Hematology Analyzer XE-2100 August 25, 1999

Sysmex Corporation of America Gilmer Road, 6699 RFD Long Grove, IL 60047-9596

Contact Person: Nina Gamperling at 847-726-3675 or by facsimile at 847-726-3559

Trade or Proprietary Name: Sysmex™ XE-2100

Automated Hematology Analyzer Common or Usual Name:

Automated Differential Cell Counter 21 CFR 864.5220 Classification Name:

Intended Use: The Sysmex™ XE-2100 is a multi-parameter hematology analyzer intended to classify the following formed elements in anti-coagulated blood:

WBCWhite Blood Cells countRBCRed Blood Cell or erythrocyte count
Neut%Neutrophil percentHGBHemoglobin concentration
Neut#Neutrophil absolute countHCTHematocrit
Lymph%Lymphocyte percentMCVMean Corpuscular Volume
Lymph#Lymphocyte absolute countMCHMean Cell Hemoglobin
Mono%Monocyte percentMCHCMean Cell Hemoglobin Concentration
Mono#Monocyte absolute countRDW-CVRed Cell Distribution Width-Coefficient of Variation
Eo%Eosinophil percentRDW-SDRed Cell Distribution Width-Standard Deviation
Eo#Eosinophil absolute countRET%Reticulocyte percent
Baso%Basophil percentRET#Reticulocyte absolute count
Baso#Basophil absolute countIRFImmature Reticulocyte Fraction
NRBC%Nucleated red blood cellpercentHFR*High Fluorescence Ratio
NRBC#Nucleated red blood cellabsolute countMFR*Medium Fluorescence Ratio
PLTPlatelet countLFR*Low Fluorescence Ratio
MPVMean Platelet Volume
PDW*Platelet Distribution Width
P-LCR*Platelet-Large Cell Ratio
PCT*Plateletcrit

*Not Reportable in USA

Device Description: The XE-2100 is an automated hematology analyzer which consists of four principal units: (1) Main Unit which aspirates, dilutes, mixes, and analyzes whole blood samples; (2) Sampler Unit which supplies samples to the Main Unit automatically; (3) IPU (Information Processing Unit) which processes data from the Main Unit and provides the operator interface with the system; (4) Pneumatic Unit which supplies pressure and vacuum from the Main Unit. Additional information on the XE-2100 is presented in Table 1.

510(k) Premarket Notification Sysmex™ Automated Hematology Analzyer XE-2100

{1}------------------------------------------------

Similarities and Differences to Predicate Devices: The Sysmex™ XE-2100 is substantially equivalent in intended use and technological characteristics to previous Sysmex™ instrumentation, the manual differential, and flow cytometry. A summary of the comparative features is presented in Table 1.

Supportive Data: Carryover, precision, linearity, and sample stability data show performance to manufacturer specifications. As demonstrated by correlation studies, the performance claims of the proposed device are similar to the predicate devices, the SE/RAM-1 and SF-3000. During these studies, specimens were evaluated from apparently healthy individuals and from patients with different pathological conditions which are expected to affect the results for particular parameters. In addition, the WBC differential of the XE-2100 was correlated to results from manual differentials performed according to NCCLS H20A and to flow cytometry. The data from these studies supports our claim that the XE-2100 is substantially equivalent to predicate methods.

Conclusion: The Sysmex™ XE-2100 is substantially equivalent to the predicate methods, the SE/RAM-1 and SF-3000.

{2}------------------------------------------------

Features(Submission #)XE-2100SE/RAM-1(K964375)SF-3000(K950508)
FDA Clearance---13-Mar-973-Nov-95
Intended UseAutomated blood cellanalyzerAutomated blood cellcounterAutomated blood cellanalyzer
Sample TypeWhole bloodWhole bloodWhole blood
Sample Volume200µL- Cap piercer130µL -Manual40µL-Capillary dilution250µL- Cap piercer100µL -Manual40µL-Capillary dilution270µL- Cap piercer170µL -Manual40µL-Capillary dilution
PerformanceSameProven performance inFDA submissionProven performance inFDA submission
ParametersWBC, Neut%/#,Lymph%/ #, Mono%/#,Eos%/#, Baso%/#,NRBC%/#, RBC, HGB,HCT, MCV, MCH,MCHC, RDW-CV,RDW-SD, RET%/#,IRF, HFR*, MFR*,LFR*, PLT, MPV,PDW*, P-LCR*, PCT**Not reportable in USAWBC, Neut%/#,Lymph%/ #, Mono%/#,Eos%/#, Baso%/#,RBC, HGB, HCT, MCV,MCH, MCHC, RDW-CV, RDW-SD, RET%/#,IRF, PLT, MPV.WBC, Neut%/#,Lymph%/#, Mono%/#,Eos%/#, Baso%/#,RBC, HGB, HCT, MCV,MCH, MCHC, RDW-CV, RDW-SD, PLT,PDW, MPV, P-LCR.
ReagentsCellpack, CellsheathStromatolyser-FB,Stromatolyser-4DL,Stromatolyser-4DS,Stromatolyser, NR,Stromatolyser-IM,Sulfolyser,Ret-Search IICellpack, Cellsheath,Cellpack-3D,StromatolyserWL(II),Stromatolyser-3D,StromatolyserEO(II),Stromalyser-BA,Stromatolyser-IM,Sulfolyser, Ret-Search,Ret-SheathCellpack, Sulfolyzer,Stromatolyser FD(I),Stromatolyser FD(II),Stromatolyser-FB
PrinciplesRBC, PLT: Sheath-flowDC detection method,WBC: Flow Cytometrymethod usingsemiconductor laserdetection methodRBC, PLT: Sheath-flowDC detection method,WBC: DC DetectionMethod, RF/DCDetection MethodRBC, PLT: DCdetection method,WBC: Flow usingsemiconductor laser
Dimensions(HxWxD) (mm)711x706x912720x636x820600x580x450
Weight (kg)938060
QC SystemL-J: 10 Files with 300points per fileL-J: 27 Files with 180points per fileL-J: 12 Files with 180points per file
Bar CodeYesYesYes
No. of Test / HrApprox. 112-150Approximately 120Approximately 80

Table 1: Comparative Features to SE/RAM-1 and SF-3000

:

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

NOV - 9 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Nina Gamperling Supervisor, Clinical and Regualtory Affairs Sysmex Corporation of America Gilmer Road, 6699 RFD Long Grove, Illinois 60047-9596

Re: K992875

Trade Name: Sysmex™ Automated Hematology Analyzer XE-2100 Regulatory Class: II Product Code: GKZ Dated: August 25, 1999 Received: August 26, 1999

Dear Ms. Gamperling:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): _ K 99 3825

Device Name: Sysmex ™ Automated Hematology Analyzer XE-2100

Indications For Use:

The Sysmex™ XE-2100 is a multi-parameter hematology analyzer intended to classify the following formed elements in EDTA anti-coagulated blood.

WBCWhite Blood Cells or leukocyte count
Neut%Neutrophil percent
Neut#Neutrophil absolute count
Lymph%Lymphocyte percent
Lymph#Lymphocyte absolute count
Mono%Monocyte percent
Mono#Monocyte absolute count
Eo%Eosinophil percent
Eo#Eosinophil absolute count
Baso%Basophil percent
Baso#Basophil absolute count
NRBC%Nucleated red blood cell percent
NRBC#Nucleated red blood cell absolute count
RBCRed Blood Cell or erythrocyte count
HGBHemoglobin concentration
HCTHematocrit
MCVMean Corpuscular Volume
MCHMean Cell Hemoglobin
MCHCMean Cell Hemoglobin Concentration
RDW-CVRed Cell Distribution Width-Coefficient of Variation
RDW-SDRed Cell Distribution Width-Standard Deviation
RET%Reticulocyte percent
RET#Reticulocyte absolute count
IRFImmature Reticulocyte Fraction
HFR*High Fluorescent Ratio
MFR*Medium Fluorescent Ratio
LFR*Low Fluorescent Ratio
PLTPlatelet count
MPVMean Platelet Volume
PDW*Platelet Distribution Width
P-LCR*Platelet- Large Cell Ratio
PCT*Plateletcrit
*Not reportable in the USA

Peter G. Maden

Laboratory Dev

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrenet of CHRD, Office of Device Evaluation (ODE) ﮯ Presription Use __ Over-The-Counter Use OR (Optional Format 1-2-96) (Per 21 CFR 801.109)

510(k) Premarket Notification Sysmex™ Automated Hematology Analzyer XE-2100

0000009

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”