K Number
K970742

Validate with FDA (Live)

Date Cleared
1997-10-22

(236 days)

Product Code
Regulation Number
864.5220
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use to identify and enumerate percentages and absolute counts of T and B lymphocytes in bload.

Indications for Use

  • For use with any flow cytometer with specified detection ranges .
  • For use with erythrocyte lysed whole blood .
  • For use with or without an isotype control .
  • For in vitro diagnostic use .
  • To identify and enumerate percentages and absolute counts of CD3+ and CD19+ ● lymphocytes
  • To characterize and monitor some forms of immunodeficiency .
  • To characterize and monitor some forms of autoimmune diseases .
Device Description

The BDIS TriTEST CD3 fluorescein isothiocyanate (FITC)/CD19 phycoerythrin (PE)/CD45 peridinin chlorophyll protein (PeCP) reagent is a three-color, direct immunofluorescence reagent for identifying and enumerating percentages of T lymphocytes (CD3+) and B lymphocytes (CD19+) in eythrocyte-lysed whole blood (LWB). When used with TRUCOUNT Absolute Count Tubes, the product will yield absolute counts in cells/uL. The Becton Dickinson Tril EST / TRUCOUNT system for immunophenotyping consists of a flow cytometer (cither from BDIS or from another manufacturer), conjugated monoclonal reagent (TriTEST CD3 FITC/CD19 PE/CD45 PeCP) and TRUCOUNT Absolute Count Tubes.

The process to obtain lymphocyte subset percentages includes: 1) obtaining a whole blood sample, 2) cell-surface antigen staining with three-color monoclonal antibody reagents, 3) erythrocyte lysis, and 4) flow cytometric acquisition and analysis of list mode data. Analysis involves computing the ratio of reagent-positive events (CD3+ and CD19+) to the CD45 positive events, and expressing the ratio as a percentage.

To obtain absolute counts, the TriTEST reagent and whole blood are added directly to an Absolute Count Tube prior to lysis. The remaining process, until analysis, is identical to that for percentages. Analysis for absolute counts requires that an additional region, the bead region, be identified and the events in this region counted. The proportion of reagent positive events to be identified and are trated. The absolute count is P x (beads/pellet)/(volume of blood sample).

When monoclonal antibody reagents are added to human whole blood, the fluorochromelabeled antibodies bind specifically to antigens on the surface of leucocytes, thus identifying lymphocyte populations. The patient blood sample is added to the counting bead peller and is treated with fluorochrome-labeled antibodies and the erythrocytes are lysed with FACS® Lysing Solution. The flow cytometer is set up so that cell populations for most samples occupy approximately the same region of fluorescence space. The sample is then introduced into the flow cytometer and the stained cells and beads fluoresce when excited by a laser beam.

The three-color reagent permits identification of lymphocyte subsets using fluorescence gating instead of forward scatter gating. This three-color reagent allows direct gating on the CD45-positive population using a combination of fluorescence and side scatter parameters. By gating on the CD45-positive population, a maximum number of lymphocytes may be captured in the gate and non-lymphocyte contamination may be minimized.

AI/ML Overview

The Becton Dickinson TriTEST reagent CD3 FITC/CD19 PE/CD45 PerCP with TRUCOUNT Absolute Count Tubes received 510(k) clearance (K970742) in 1997. The device is intended for in vitro diagnostic use to identify and enumerate percentages and absolute counts of T and B lymphocytes in erythrocyte-lysed whole blood, and to characterize and monitor some forms of immunodeficiency and autoimmune diseases.

Here's a breakdown of the acceptance criteria and study information:

1. Acceptance Criteria and Reported Device Performance

CriteriaAcceptance Criteria (Implied)Reported Device Performance
AccuracyEquivalence to predicate device (Simultest/ADCC).Accuracy data demonstrated the TriTEST/TRUCOUNT product's equivalence to Simultest/ADCC.
Isotype ControlUsability with or without an isotype control.Data indicated that the reagent may be used with or without an isotype control.
Reference RangeSites must be able to determine their own reference range.Acknowledged that variables like age, sex, and geographical location influence reference ranges, requiring each site to determine its own.
StabilityAcceptable stability for percentages and absolute counts.For absolute counts, samples should be stained and analyzed within 6 hours of draw. (Specific stability for percentages is not explicitly detailed but implied as acceptable).
Reproducibility (Within-specimen)Acceptable within-sample reproducibility.BDIS Internal Study: 10 replicates from 1 high, 1 medium, and 1 low T and B lymphocyte samples were assessed. Clinical Site Study: 3 aliquots from each donor (n=92 for CD3+ and n=80 for CD19+) were assessed at 3 clinical sites. Results: Demonstrated acceptable within-sample reproducibility for both.
LinearityLinear response over a specified concentration range.Linear response determined using blood samples from 3 normal donors diluted to 5 concentrations, ranging from 16,700 to 200 lymphocytes/uL and from 31,000 to 2,500 WBC/uL. Results indicate a linear response over this range.
Cross-ReactivitySpecificity of clones should not be affected by conjugation and formulation.Cross-reactivity of clones is reported in the literature, and conjugation and product formulation have not changed their specificity.
Cross-Platform ReproducibilityUsability with flow cytometers not made by Becton Dickinson.Results indicated that TriTEST CD3 FITC/CD19 PE/CD45 PerCP with or without TRUCOUNT tubes may be used with flow cytometers not made by Becton Dickinson.

2. Sample Size Used for the Test Set and Data Provenance

  • Accuracy: Not explicitly stated for accuracy comparison, but implies a sufficient number of samples were tested to demonstrate equivalence.
  • Isotype Control: Sample size not explicitly stated.
  • Reproducibility (Within-specimen):
    • BDIS Internal Study: 3 samples (1 high, 1 medium, 1 low concentration of T and B lymphocytes), with 10 replicates each.
    • Clinical Site Study: 92 donors for CD3+ and 80 donors for CD19+, with 3 aliquots assessed from each donor.
  • Linearity: Blood samples from 3 normal donors.
  • Data Provenance: Studies were performed at Cleveland Clinic, Johns Hopkins Hospital, Institute for Medical Research at the University of North Carolina, and at Becton Dickinson Immunocytometry Systems laboratories in San Jose, California. This indicates a mix of external clinical sites and internal company labs, suggesting prospective data collection for these specific studies.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The provided text does not specify the number or qualifications of experts used to establish the ground truth for the test set. Given the nature of the device (flow cytometry for lymphocyte enumeration), the ground truth would likely be established through comparisons with established methods and/or expert review of flow cytometry plots.

4. Adjudication Method for the Test Set

The provided text does not specify any explicit adjudication method (e.g., 2+1, 3+1, none) for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC study was not done. The device's performance was compared to a predicate device (Simultest™ IMK-Lymphocyte Tube C plus ADCC and the hematology analyzer) and internal reproducibility studies, not through a study involving multiple human readers with and without AI assistance to measure improvement in human performance.

6. Standalone (Algorithm Only) Performance

This device is a reagent and tube system for use with a flow cytometer. Its performance is intrinsically linked to the overall system (reagent, sample preparation, flow cytometer, and analysis software). Therefore, a "standalone algorithm only" performance study in the sense of a pure AI algorithm is not applicable to this type of medical device. The performance data presented (accuracy, reproducibility, linearity, etc.) inherently reflect the "algorithm only" performance within the context of the complete TriTEST/TRUCOUNT system.

7. Type of Ground Truth Used

The primary ground truth used for the performance studies appears to be:

  • Comparison to Predicate Device: For accuracy, the device was compared against the established Simultest™ IMK-Lymphocyte Tube C plus ADCC. This implies that the predicate device's results serve as a form of "ground truth" for demonstrating equivalence.
  • Internal Consistency/Expected Biological Ranges: For reproducibility, stability, and linearity, the ground truth is based on the expectation of consistent results from the same sample or linear responses over dilution ranges, consistent with established biological principles and laboratory practices.

8. Sample Size for the Training Set

The document does not specify a separate "training set" sample size. For in vitro diagnostic reagents and systems like this, the development process (which might involve data analogous to a training set) is not typically detailed in terms of sample size in a 510(k) summary. The performance data presented are from validation studies, which are akin to testing.

9. How the Ground Truth for the Training Set Was Established

As no explicit "training set" is mentioned for an AI/algorithm-based device, information on how its ground truth was established is not applicable in this context. The product is a diagnostic reagent and counting tube system, not a machine learning algorithm that requires a labeled training dataset in the modern sense.

{0}------------------------------------------------

Attachment F

K9 70742
Oct. 22, 1997

Summary of Safety and Effectiveness

Submitter Information (21 CFR 807,92(a)(1))

Submitter:Becton Dickinson Immunocytometry Systems
2350 Qume Drive
San Jose, CA 95131-1807
  • Anna Longwell, Esq. Contact: Director, Regulatory Affairs - Corporate (408) 954-2254
    Summary date: February 26, 1997

Name of Device and Classification (21 CFR 807.92(a)(2))

  • Becton Dickinson TriTEST™ reagent CD3 FITC/CD19 PE/CD45 Name: PerCP; TRUCOUNT TM Absolute Count Tubes
    Class II Classification:

Predicate Device (21 CFR 807.92(a)(3))

The BDIS TriTEST™ CD3 FITC/CD19 PE/CD45 PerCP reagent with TRUCOUNT Absolute Count Tubes is substantially equivalent to Simultest™ IMK-Lymphocyte Tube C (cleved to market under 510(k) K913192) plus ADCC. TriTEST reagent CD3/CD19/CD45, when used to enumerate percentages of T and B lymphocytes was cleared to market under 510(k) K950342.

Description of the Device (21 CFR 807.92(2)(4))

The BDIS TriTEST CD3 fluorescein isothiocyanate (FITC)/CD19 phycoerythrin (PE)/CD45 peridinin chlorophyll protein (PeCP) reagent is a three-color, direct immunofluorescence reagent for identifying and enumerating percentages of T lymphocytes (CD3+) and B lymphocytes (CD19+) in eythrocyte-lysed whole blood (LWB). When used with TRUCOUNT Absolute Count Tubes, the product will yield absolute counts in cells/uL. The Becton Dickinson Tril EST / TRUCOUNT system for immunophenotyping consists of a flow cytometer (cither from BDIS or from another manufacturer), conjugated monoclonal reagent (TriTEST CD3 FITC/CD19 PE/CD45 PeCP) and TRUCOUNT Absolute Count Tubes.

The process to obtain lymphocyte subset percentages includes: 1) obtaining a whole blood sample, 2) cell-surface antigen staining with three-color monoclonal antibody reagents, 3) erythrocyte lysis, and 4) flow cytometric acquisition and analysis of list mode data. Analysis involves computing the ratio of reagent-positive events (CD3+ and CD19+) to the CD45 positive events, and expressing the ratio as a percentage.

To obtain absolute counts, the TriTEST reagent and whole blood are added directly to an Absolute Count Tube prior to lysis. The remaining process, until analysis, is identical to that for percentages. Analysis for absolute counts requires that an additional region, the bead region,

306

{1}------------------------------------------------

Summary of Safety and Effectiveness

be identified and the events in this region counted. The proportion of reagent positive events to be identified and are trated. The absolute count is P x (beads/pellet)/(volume of blood sample).

When monoclonal antibody reagents are added to human whole blood, the fluorochromelabeled antibodies bind specifically to antigens on the surface of leucocytes, thus identifying lymphocyte populations. The patient blood sample is added to the counting bead peller and is treated with fluorochrome-labeled antibodies and the erythrocytes are lysed with FACS® Lysing Solution. The flow cytometer is set up so that cell populations for most samples occupy approximately the same region of fluorescence space. The sample is then introduced into the flow cytometer and the stained cells and beads fluoresce when excited by a laser beam.

The three-color reagent permits identification of lymphocyte subsets using fluorescence gating instead of forward scatter gating. This three-color reagent allows direct gating on the CD45positive population using a combination of fluorescence and side scatter parameters. By gating on the CD45-positive population, a maximum number of lymphocytes may be captured in the gate and non-lymphocyte contamination may be minimized.

Intended Use (21 CFR 807.92(a)(5))

For in vitro diagnostic use to identify and enumerate percentages and absolute counts of T and B lymphocytes in bload.

Indications for Use

  • For use with any flow cytometer with specified detection ranges .
  • For use with erythrocyte lysed whole blood .
  • For use with or without an isotype control .
  • For in vitro diagnostic use .
  • To identify and enumerate percentages and absolute counts of CD3+ and CD19+ ● lymphocytes
  • To characterize and monitor some forms of immunodeficiency .
  • To characterize and monitor some forms of autoimmune diseases .

Clinical Utility

The enumeration of T (CD3+) and B (CD19+) lymphocytes has been found useful in monitoring some forms of immunodeficiency and autoimmune disease.

Comparison to Predicate Device (21 CFR 807.92(a)(6))

The three-color reagent is substantially equivalent to the predicate: Simultest™ IMK-Lymphocyte Tube C plus hematology analyzer in that they share the same intended uses. The Simultest portion of the predicate and the TriTEST reagent use essentially the same monoclonal antibody/flow cytometric methodology. The TriTEST and Simultest products differ in the steps used to determine analysis gates to identify the lymphocyte population. The TriTEST /TRUCOUNT system compared to the predicate, requires less sample handling, a simpler sample preparation scheme and fewer sample tubes to achieve the same end result, Results demonstrate that the products yield essentially equivalent performance characteristics.

307

{2}------------------------------------------------

Summary of Safety and Effectiveness

Performance Data (21 CFR 807.92(b)(2))

Performance of the product was established by testing at Cleveland Clinic, Johns Hopkins I enomance of the problem was events of North Carolina, and at Becton I rospital, Institute or ometry Systems laboratories in San Jose, California.

Several studies were performed:

  • · Accuracy was determined by comparison to Simultest/ADCC. Accuracy data demonstrated the Tril EST / TRUCOUNT product's equivalence to Simultest/ADCC.
  • · Use of an Isotype Control was studied. Data was analyzed for percent T lymphocytes and Ose of an 1990 per Souther, was control to set quadrant markers and then using only the stained sample to set quadrant markers. Data indicated that the reagent may be used with or without an isotype control.
  • · Reference range studies were performed. Many variables such as age, sex and geographical location may influence the reference range. Each site must determine its own reference range.
  • · A stability study was conducted to assess the time effect relating to age of blood (timefrom-draw) and the time effect relating to the age of the s-ain (time-from-sample preparation), as well as the combined effect of both. Stability was determined for both percentages and absolute counts. Results indicate that for absolute counts, samples should be stained and analyzed within 6 hours of draw.
  • · Within-specimen reproducibility was performed at BDIS; 10 replicates from 1 high, 1 medium, and 1 low (with respect to T and B lymphocytes) were assessed. Within-specimen reproducibility was also performed at 3 clinical sites; 3 aliquots from each donor (n=92 for CD3+ and n=80 for CD19+) were assessed. Results demonstrated acceptable within-sample reproducibility.
  • Lincarity was determined using blood samples from 3 normal donors diluted to 5 concentrations, ranging from 16,700 to 200 lymphocytes/uL and from 31,000 to 2,500 WBC/uL. Results indicate a linear response over this range.
  • · Cross reactivity of these clones is reported in the literature. Conjugation and product formulation have not changed their specificity.
  • Results from a goss platform reproducibility study indicated that TriTEST CD3 FITC/CD19 PE/CD45 PerCP with or without TRUCOUNT tubes may be used with flow cytometers not made by Becton Dickinson.

Performance Data - Conclusions (21 CFR 807.92(b)(3))

The results of the clinical studies demonstrate that the device is as safe and effective as the predicate device.

CD3/CD19/CD45 TRUCOUNT 510(k) Notification Attachment F: Summary of Safety and Effectiveness

308
/

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Anna Longwell, Esq. Director, Regulatory Affairs - Corporate Becton Dickinson Immunocytometry Systems 2350 Qume Drive San Jose, CA 95131-1807

OCT 2 2 1997

Re : K970742 Becton Dickinson TriTEST™ Reagent CD3 FITC/CD19 Trade Name: PE/CD45 PerCP; TruCOUNT™ Absolute Count Tubes Regulatory Class: II Product Code: GKZ Dated: August 29, 1997 Received: September 02, 1997

Dear Ms. Longwell:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. go determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) Additionally, for questions on the promotion and 594-4588. advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices ... Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

| of_ Page_

K910742 (if known):-

dons For Use:

6

.

For use with any flow cytometer with specified detection ranges
t

For use with erythrocyte lysed whole blood

For use with or without an isotype control

For in vitro diagnostic use

For in viris cases of absolute counts of CD3+ and absolute counts of CD3+ and CD19+ lymphocytes

SE of the program of monitor some forms of immunodeficiency

To characterize and monitor some forms of autoimmune diseases .

IF YOU STILL NEED MORE SPACE ON THIS LINE, CONTINUE ON ANOTHER PAGE IF NEEDED).

MEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Peter E. Marin

(Division Sign-Of
Division of Clinical Lang
510(k) Number

OR

Over-The-Counter Use_

Prescription Use Or 21 CFR 801.109)

(Optional Format 1-2-96).

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”