K Number
K250822

Validate with FDA (Live)

Date Cleared
2025-07-03

(107 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of X-ray transmission data.

The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment, and radiation therapy planning as well as for diagnostic and therapeutic interventions.

This CT system can be used for low dose lung cancer screening in high risk populations*.

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Device Description

Siemens intends to market a new software version, SOMARIS/10 syngo CT VB20 for the following SOMATOM Computed Tomography (CT) Scanner Systems:

a) Single Source CT Scanner systems (SOMATOM go. Platform):

  • SOMATOM go.Now
  • SOMATOM go.Up
  • SOMATOM go.All
  • SOMATOM go.Top
  • SOMATOM go.Sim
  • SOMATOM go.Open Pro

In this submission, the above listed CT scanner systems are jointly referred to as subject devices by "SOMATOM go. Platform" CT scanner systems.

b) Dual Source CT Scanner system:

  • SOMATOM Pro.Pulse

The above listed subject devices with SOMARIS/10 syngo CT VB20 are Computed Tomography X-ray Systems which feature one (Single Source) or two (Dual Source) continuously rotating tube-detector system and function according to the fan beam principle. The SOMATOM go. Platform and the SOMATOM Pro.Pulse with software SOMARIS/10 syngo CT VB20 produce CT images in DICOM format, which can be used by trained staff for software applications, e.g. post-processing applications, commercially distributed by Siemens Healthcare and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.

AI/ML Overview

The provided FDA 510(k) Clearance Letter for the SOMATOM CT Systems focuses heavily on establishing substantial equivalence to predicate devices through comparisons of technological characteristics, hardware, and software. It generally asserts that the device has met performance criteria through verification and validation testing, but it does not provide a detailed "Acceptance Criteria Table" with specific quantitative metrics and reported device performance. Similarly, it describes the types of studies performed (e.g., bench testing, retrospective blinded rater study), but it lacks the specific details requested regarding sample sizes, data provenance, expert qualifications, and effect sizes that would typically be found in a detailed study report.

Therefore, I will extract and synthesize the information that is available in the document and explicitly state where the requested information is not provided.


Understanding the Device and its Changes

The devices under review are Siemens SOMATOM CT Systems (SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM go.Open Pro, and SOMATOM Pro.Pulse) with a new software version, SOMARIS/10 syngo CT VB20. This new software version builds upon the previous VB10 version cleared in K233650 and K232206.

The submission focuses on modifications and new features introduced with VB20, including:

  • Eco Power Mode: New feature for reduced energy consumption during idle times (not supported on go.Now and go.Up).
  • Oncology Exchange: New feature for transferring prescription information from ARIA Oncology Information System.
  • myExam Contrast: New feature for exchanging contrast injection parameters.
  • FAST 3D Camera/FAST Integrated Workflow: Modifications including retrained algorithms, collision indication, and Centerline/Grid Overlay.
  • FAST Planning: Extended to detect additional body regions.
  • myExam Companion (myExam Compass/myExam Cockpit): Clinical decision trees now available for child protocols.
  • HD FoV 5.0: New extended field of view reconstruction algorithm (for go.Sim and go.Open Pro only).
  • CT guided intervention – myAblation Guide interface: New interface.
  • Flex 4D Spiral: Modifications regarding dynamic tube current modulation.
  • ZeeFree RT: New stack artifact reduced reconstruction for respiratory-related examinations (for go.Open Pro only).
  • DirectDensity: Modified to include stopping-power ratio (Kernel St).
  • DirectLaser: Patient Marking workflow improvement.
  • Respiratory Motion management - Open Online Interface: New interface for respiratory gating.
  • DirectSetup Notes: Enabled for certain SOMATOM go. Platform systems.

The core argument for clearance is substantial equivalence to predicate devices. This means that, despite modifications, the device is as safe and effective as a legally marketed device (the predicates).


1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not contain a specific table of quantitative acceptance criteria with corresponding reported device performance values. Instead, it describes general acceptance criteria related to verification and validation tests and then provides qualitative statements about the test results demonstrating comparability or improvement over predicate devices.

Here's a summary of the described performance evaluations:

Feature/MetricAcceptance Criteria (Qualitative)Reported Device Performance (Qualitative)
OverallMeet acceptance criteria for all software specifications. Enable safe and effective integration. Perform as intended in specified use conditions."All software specifications have met the acceptance criteria." "Verification and validation support the claims of substantial equivalence." "Perform(s) as intended in the specified use conditions." "As safe, as effective, and perform as well as or better than the predicate devices."
FAST 3D Camera Accuracy (Isocentering, Range, Direction)Comparable or better accuracy to predicate device for adults; extend support to adolescents."Overall, the subject devices with syngo CT VB20 delivers comparable or improved accuracy to the predicate devices with syngo CT VB10 predicate device for adults and extends the support to adolescents."
FAST Planning CorrectnessHigh fraction (percentage) of ranges calculated correctly and without needing change. Meets interactive requirements (fast calculation time)."For more than 90% of the ranges no editing action was necessary to cover standard ranges." "For more than 95%, the speed of the algorithm was sufficient."
HD FoV 5.0 Performance (vs. HD FoV 4.0)As safe and effective as HD FoV 4.0."Results obtained with the new HD FoV 5.0 algorithm are compared with its predecessor, the HD FoV 4.0 algorithm, based on physical and anthropomorphic phantoms...This comparison is conducted to demonstrate that the HD FoV 5.0 algorithm is as safe and effective as the HD FoV 4.0 algorithm." (No quantitative metrics provided in this document excerpt regarding this comparison's outcome).
Flex 4D Spiral Functionality & Image QualityProper function and acceptable image quality."The performed bench test report describes the technical background of Flex 4D Spiral and its functionalities with SOMATOM CT scanners, demonstrate the proper function of those, and assess the image quality of Flex 4D Spiral." (No quantitative metrics provided)
ZeeFree RT Reconstruction PerformanceNo relevant errors in CT values and noise in homogeneous phantoms. No relevant errors in CT values in tissue-equivalent phantoms. No relevant geometrical distortions in static phantoms. No relevant deteriorations of position/shape in dynamic phantoms. No relevant new artifacts. Maintain performance with iMAR. Independent of detector width."introduces no relevant errors in terms of CT values and noise levels measured in a homogeneous water phantom" "introduces no relevant errors in terms of CT values measured in a phantom with tissue-equivalent inserts, even in the presence of metals and in combination with the iMAR algorithm" "introduces no relevant geometrical distortions in a static torso phantom" "introduces no relevant deteriorations of the position or shape of a dynamic thorax phantom" "does not introduce relevant new artefacts" "can be successfully applied in combination with metal artifact correction (iMAR)" "is independent from the physical detector width"
DirectDensity Performance (iBHC variants)Reduced dependence on tube voltage and filtration for non-water-like tissues. Image values aligned with material properties."reduced dependence on tube voltage and filtration compared to the corresponding quantitative kernel (Qr) with iBHC Bone for non-water-like tissues, such as adipose and bone." "generate image value closely aligned with the respective material properties." "has been validated."

2. Sample Sizes Used for the Test Set and Data Provenance

The document provides very limited, qualitative information:

  • FAST 3D Camera: Optimized using "additional data from adults and adolescence patients." No specific number of patients or images mentioned.
  • FAST Planning: Evaluated on "patient data." No specific number of patients or images mentioned.
  • HD FoV 5.0: Evaluated with "physical and anthropomorphic phantoms."
  • Flex 4D Spiral: No specific sample size or data type mentioned for performance assessment.
  • ZeeFree RT: Evaluated with "homogeneous water phantom," "phantom with tissue-equivalent inserts," "static torso phantom," and "dynamic thorax phantom." Also, "retrospective blinded rater studies of respiratory 4D CT examinations performed at two institutions." No specific number of phantoms, images per phantom, or patient cases mentioned.
  • DirectDensity: Evaluated on "SOMATOM CT scanner models." No specific sample size or data type mentioned.

Data Provenance:

  • Country of Origin: Not specified for the patient data used for algorithm optimization/validation.
  • Retrospective or Prospective:
    • FAST 3D Camera: Implied retrospective as it uses "additional data."
    • FAST Planning: Implied retrospective as it uses "patient data."
    • HD FoV 5.0: Retrospective for the blinded rater study.
    • ZeeFree RT: Retrospective for the blinded rater study of clinical cases. The phantom tests are by nature not retrospective/prospective.

3. Number of Experts and Qualifications for Ground Truth

  • HD FoV 5.0: "board-approved radio-oncologists and medical physicists." The number of experts is not specified.
  • ZeeFree RT: "board-approved radio-oncologists and medical physicists." The number of experts is not specified.

For other tests, ground truth appears to be established by phantom measurements or internal engineering verification, rather than human expert reads validating clinical ground truth.


4. Adjudication Method for the Test Set

The document mentions "retrospective blinded rater study" for HD FoV 5.0 and ZeeFree RT. However, it does not specify the adjudication method used (e.g., 2+1, 3+1, none) for these studies. It only states they were "blinded."


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned for HD FoV 5.0 and ZeeFree RT. Both were "retrospective blinded rater studies."
  • Effect Size: The document does not report specific effect sizes (e.g., how much human readers improve with AI vs. without AI assistance). It only states that the purpose of the comparison was to "demonstrate that the HD FoV 5.0 algorithm is as safe and effective as the HD FoV 4.0 algorithm" and for ZeeFree RT that it "enables the optional reconstruction of stack artefact corrected images, which reduce the strength of misalignment artefacts." This implies an assessment of non-inferiority or improvement in image quality, but specific quantitative results for reader performance are not provided in this excerpt.

6. Standalone (Algorithm Only) Performance

The document describes tests for several algorithms (FAST 3D Camera, FAST Planning, HD FoV 5.0, Flex 4D Spiral, ZeeFree RT, DirectDensity) using phantoms and "patient data." These evaluations seem to be focused on the algorithm's performance in generating images or calculations, independent of human interpretation in some cases (e.g., accuracy of FAST 3D Camera, correctness percentage of FAST Planning).

However, it does not explicitly use the term "standalone performance" to differentiate these from human-in-the-loop assessments. The mention of "retrospective blinded rater studies" for HD FoV 5.0 and ZeeFree RT indicates a human-in-the-loop component for that specific evaluation, but the phantom testing mentioned alongside them would be considered standalone.


7. Type of Ground Truth Used

  • Phantom Data: For HD FoV 5.0, Flex 4D Spiral, ZeeFree RT, and DirectDensity, physical and/or anthropomorphic phantoms were used, implying the ground truth is precisely known physical characteristics or pre-defined phantom configurations.
  • Expert Consensus/Reads: For HD FoV 5.0 and ZeeFree RT, board-approved radio-oncologists and medical physicists performed retrospective blinded rater studies, implying their interpretations/ratings served as a form of ground truth or evaluation metric. It's not explicitly stated if this was against a clinical gold standard (e.g., pathology) or if it was a comparative assessment of image quality and clinical utility.
  • Internal Verification: For FAST 3D Camera, FAST Planning, accuracy was assessed, likely against internal system metrics or pre-defined ideal outcomes.

8. Sample Size for the Training Set

The document does not provide any specific information about the sample size used for training the algorithms (e.g., for FAST 3D Camera, FAST Planning, HD FoV 5.0, ZeeFree RT). It only states that FAST 3D Camera was "optimized using additional data" and FAST Planning's algorithm had "product development, validation, and verification on patient data."


9. How the Ground Truth for the Training Set Was Established

The document does not provide any specific information on how the ground truth for the training set was established. It only mentions the data types used for validation/verification (phantoms, patient data from two institutions, expert raters).

FDA 510(k) Clearance Letter - SOMATOM CT Systems

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

July 3, 2025

Siemens Medical Solutions USA, Inc.
℅ Kenny M Bello
Regulatory Affairs Professional
810 Innovation Drive
KNOXVILLE, TN 37932

Re: K250822
Trade/Device Name: SOMATOM go.Now; SOMATOM go.Up; SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM Pro.Pulse
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: JAK
Dated: May 16, 2025
Received: June 16, 2025

Dear Kenny M Bello:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250822 - Kenny M Bello Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250822 - Kenny M Bello Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250822

Device Name: SOMATOM go.Now; SOMATOM go.Up; SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM Pro.Pulse

Indications for Use (Describe)

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of X-ray transmission data.

The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment, and radiation therapy planning as well as for diagnostic and therapeutic interventions.

This CT system can be used for low dose lung cancer screening in high risk populations*.

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1

Page 5

510(k) Summary

SOMATOM go. Platform and SOMATOM Pro.Pulse
with software version SOMARIS/10 syngo CT VB20

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

I. Contact Details

Submitter
Siemens Medical Solutions USA, Inc.
810 Innovation Drive
Knoxville, TN 37932
Establishment Registration Number: 1034973

Importer/Distributor
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355
Establishment Registration Number: 2240869

Location of Manufacturing Site (1)
Siemens Healthineers AG
Siemensstr. 1 -OR- Rittigfeld 1
D-91301 Forchheim, Germany
Establishment Registration Number: 3004977335

Location of Manufacturing Site (2)
SIEMENS SHANGHAI, MEDICAL EQUIPMENT LTD
278 Zhou Zhu Rd
Shanghai, CHINA, 201318
Establishment Registration Number: 3003202425

Note: Descriptions in this submission use the short company name Siemens. It covers both manufacturing locations and names as listed above. Brand name on all products is Siemens Healthineers.

K250822

Page 6

Submitter Contact Person:
Kenny M Bello
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
Molecular Imaging
810 Innovation Drive
Knoxville, TN 37932
Phone: (205) 856-6099
monsuru.bello@siemens-healthineers.com

II. Device Name and Classification

Table 1: Subject Device Names

Product NameTrade Name
SOMATOM go.NowSOMATOM go.Now
SOMATOM go.UpSOMATOM go.Up
SOMATOM go.AllSOMATOM go.All
SOMATOM go.TopSOMATOM go.Top
SOMATOM go.SimSOMATOM go.Sim
SOMATOM go.Open ProSOMATOM go.Open Pro
SOMATOM Pro.PulseSOMATOM Pro.Pulse

Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK

III. Predicate Device

Primary predicate device:
Trade Name: SOMATOM go. Platform CT Scanner Systems with SOMARIS/10 syngo CT VB10 software
510(k) Number: K233650
Clearance Date: March 26, 2024
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
Recall Information: All predicate device recalls have been considered in the subject device design.

K250822

Page 7

Secondary predicate device:
Trade Name: SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 software
510(k) Number: K232206
Clearance Date: December 6, 2024
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
Recall Information: All predicate device recalls have been considered in the subject device design.

Note:
K233650 was a bundle submission with various Siemens CT Scanner Systems, including the CT scanner systems SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM go.Open Pro, SOMATOM X.cite, SOMATOM X.ceed.

IV. Device Description

Siemens intends to market a new software version, SOMARIS/10 syngo CT VB20 for the following SOMATOM Computed Tomography (CT) Scanner Systems:

a) Single Source CT Scanner systems (SOMATOM go. Platform):

  • SOMATOM go.Now
  • SOMATOM go.Up
  • SOMATOM go.All
  • SOMATOM go.Top
  • SOMATOM go.Sim
  • SOMATOM go.Open Pro

In this submission, the above listed CT scanner systems are jointly referred to as subject devices by "SOMATOM go. Platform" CT scanner systems.

b) Dual Source CT Scanner system:

  • SOMATOM Pro.Pulse

The above listed subject devices with SOMARIS/10 syngo CT VB20 are Computed Tomography X-ray Systems which feature one (Single Source) or two (Dual Source) continuously rotating tube-detector system and function according to the fan beam principle. The SOMATOM go. Platform and the SOMATOM Pro.Pulse with software SOMARIS/10 syngo CT VB20 produce CT images in DICOM format, which can be used by trained staff for software applications, e.g. post-processing applications, commercially distributed by Siemens Healthcare and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.

K250822

Page 8

Only trained and qualified users, certified in accordance with country-specific regulations, are authorized to operate the system. For example, physicians, radiologists, or technologists. The user must have the necessary U.S. qualifications in order to diagnose or treat the patient with the use of the images delivered by the system.

The software platform provides plugin software interfaces that allow for the use of specific commercially available post processing software algorithms in an unmodified form from the cleared stand-alone post processing version.

New software version syngo CT VB20 (SOMARIS/10 syngo CT VB20) is a modified software version based on syngo CT VB10 (SOMARIS/10 syngo CT VB10) which was cleared for the predicate devices in K233650 (SOMATOM go. Platform) and in K232206 (SOMATOM Pro.Pulse).

Software version SOMARIS/10 syngo CT VB20 will be offered ex-factory and as an optional upgrade for the applicable existing SOMATOM go. Platform and SOMATOM Pro.Pulse CT Scanner Systems.

The bundle approach is feasible for this submission since the subject devices have similar technological characteristics, software operating platform, and supported software characteristics. The supporting data are similar, primarily one review division/group will be involved, and the indications for use is the same between the devices. All subject devices will support previously cleared software and hardware features in addition to the applicable modifications as described within this submission. The intended use remains unchanged compared to the predicate devices.

V. Indications for Use

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment, and radiation therapy planning as well as for diagnostic and therapeutic interventions.

This CT system can be used for low dose lung cancer screening in high risk populations*.

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

VI. Indications for Use Comparison

Subject Device Indications for Use:

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.

This CT system can be used for low dose lung cancer screening in high risk populations*.

K250822

Page 9

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Primary Predicate Device Indications for Use:

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.

This CT system can be used for low dose lung cancer screening in high risk populations*.

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Secondary Predicate Device Indications for Use:

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.

This CT system can be used for low dose lung cancer screening in high risk populations*.

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Comparison:

The subject device Indications for Use is the exact same as the primary (K233650) and the secondary (K232206) predicate's Indications for Use. As such, the Indications for Use remains unchanged from the predicate devices and is substantially equivalent to the predicate devices.

VII. Comparison of Technological Characteristics with the Predicate Device

Supported by the subject device, SOMARIS/10 syngo CT VB20 software version is a further development of the SOMARIS/10 syngo CT VB10 software version which is cleared in K233650 and K232206.

The SOMATOM go. Platform and SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20 software version provides the same technological characteristics in terms of materials, energy source, and control mechanisms when compared to the predicate devices. The software features of the SOMATOM go. Platform and SOMATOM Pro.Pulse have been modified or improved in comparison to the predicate devices to support enhanced device functionality compared to the predicate devices.

K250822

Page 10

The new syngo CT VB20 software reuses all unmodified software features of the legacy software syngo CT VB10 cleared in K233650 and K232206. Additionally, no features present in the predicate device are descoped.

Software version SOMARIS/10 syngo CT VB20 is designed to reuse hardware independent extended functionalities and GO technologies provided by Siemens cleared software applications.

The intended use and fundamental scientific technology for the SOMATOM go. Platform and SOMATOM Pro.Pulse remain unchanged from the predicate devices.

At a high level, the subject and predicate devices are based on the same subset of technological elements:

  • Scanner Principle – Whole body X-Ray Computed Tomography Scanner
  • System Acquisition – Continuously rotating tube detector system
  • Iterative Reconstruction – Support of various iterative reconstruction principles
  • Workplaces – Support of workplaces that include reconstruction and image evaluation software
  • Patient table
  • Patient table foot switch for movement
  • Stellar Detector technology
  • Tin filtration technology
  • Athlon/Vectron X-ray Tube
  • Power Generator
  • Scan&GO
  • Mobile workflow (Tablet)
  • Optional injector arm
  • Optional support of CT guided intervention workflow (myNeedle Companion – myNeedle Guide)
  • Optional support of FAST 3D Camera for patient positioning workflow
  • Scanner display and control functionality
  • Remote Scan Control
  • Interfaces for Respiratory Gating Devices
  • Standard CARE technologies
  • Standard FAST technologies
  • Standard GO technologies
  • Precision matrix (large image matrices)
  • HD FoV
  • DirectDensity
  • Stack artefact reduced reconstruction

The terms in the following Table 2 are used to indicate various technological characteristics in comparison to the primary and secondary predicate devices information:

Table 2: Overview of term definition.

TermDefinition
newThe feature is newly supported for Siemens CT Scanners and newly introduced within the subject device submission

K250822

Page 11

TermDefinition
modifiedThis feature is modified from the predicate devices
enabledThis feature is currently supported by other cleared Siemens CT systems or cleared Siemens stand-alone software applications. This feature will be supported for the subject device with software version SOMARIS/10 syngo CT VB20 and is substantially equivalent compared to the cleared version of the predicate devices.
unmodifiedThe feature remains unchanged from the predicate device.
n.a.The feature is not supported for the subject/predicate device

The subject devices SOMATOM go. Platform and SOMATOM Pro.Pulse CT scanner systems with SOMARIS/10 syngo CT VB20 will support hardware and software modifications/further developments in comparison to the predicate devices as listed in Table 3. The column "HW/ SW change" in the table contains reference to a hardware (HW) or/and software (SW) change. The columns under "Predicate devices" of the Table 3 indicate the clearance of the technological properties in the corresponding predicate device.

K250822

Page 12

Table 3: Summary of modifications (hardware and software) of the subject devices SOMATOM go. Platform and SOMATOM Pro.Pulse with software version SOMARIS/10 syngo CT VB20 compared to the predicate devices.

Technological propertyHW/SW changeSubject devicesPredicate devices
Single SourceDual SourceSingle SourceDual Source
SOMATOM go. Platform with syngo CT VB20SOMATOM Pro.Pulse with syngo CT VB20SOMATOM go. Platform with syngo CT VB10 (K233650)SOMATOM Pro.Pulse with syngo CT VB10 (K232206)
1. CARE MoodlightHW/SWenabled¹ (same as in K233650)unmodifiedclearedcleared
2. Interfaces for respiratory gatingHWmodified² (same as in K233650, with exception of Open Online Interface which are introduced with syngo CT VB20)n.a.clearedn.a.
3. Eco Power ModeHW/SWnew³newn.a.n.a.

¹ In software version syngo CT VB20, the functionality CARE Moodlight is enabled for the subject devices SOMATOM go.Sim and SOMATOM go.Open Pro. For the subject devices SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, and SOMATOM go.Top, the functionality CARE Moodlight was cleared in K233650 and it remains unchanged in software version syngo CT VB20.

² SOMATOM go.Now does not support respiratory gating interfaces.

³ SOMATOM go.Now and SOMATOM go.Up do not support Eco Power Mode.

K250822

Page 13

Technological propertyHW/SW changeSubject devicesPredicate devices
Single SourceDual SourceSingle SourceDual Source
SOMATOM go. Platform with syngo CT VB20SOMATOM Pro.Pulse with syngo CT VB20SOMATOM go. Platform with syngo CT VB10 (K233650)SOMATOM Pro.Pulse with syngo CT VB10 (K232206)
4. FAST 3D Camera/ FAST Integrated WorkflowSWmodified (same as in K233650, with exception of retrained FAST 3D Camera algorithms, collision indication, and Centerline/Grid Overlay which are introduced with syngo CT VB20)modified (same as in K232206, with exception of retrained FAST 3D Camera algorithms, collision indication, and Centerline/Grid Overlay which are introduced with syngo CT VB20)clearedcleared
5. Oncology ExchangeSWnewnewn.a.n.a.
6. DirectSetup NotesSWenabled⁴ (same as in K233650)enabled⁴ (same as in K233650)clearedn.a.
7. FAST PlanningSWmodified (same as in K233650, with exception of the detection of additional body regions which are introduced with syngo CT VB20)modified (same as in K232206, with exception of the detection of additional body regions which are introduced with syngo CT VB20)clearedcleared

⁴ In software version syngo CT VB20, the functionality DirectSetup Notes is enabled for the subject devices SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, and SOMATOM Pro.Pulse. For the subject devices SOMATOM go.Sim and SOMATOM go.Open Pro, the functionality DirectSetup Notes was cleared in K233650 and it remains unchanged in software version syngo CT VB20.

K250822

Page 14

Technological propertyHW/SW changeSubject devicesPredicate devices
Single SourceDual SourceSingle SourceDual Source
SOMATOM go. Platform with syngo CT VB20SOMATOM Pro.Pulse with syngo CT VB20SOMATOM go. Platform with syngo CT VB10 (K233650)SOMATOM Pro.Pulse with syngo CT VB10 (K232206)
8. myExam Companion – myExam Compass/myExam CockpitSWmodified (same as in K233650, with exception of the clinical decision trees now also available for child protocols)modified (same as in K233650, with exception of the clinical decision trees now also available for child protocols)clearedcleared
9. HD FoV 5.0SWnew⁵n.a.n.a.n.a.
10. CT guided intervention – myAblation Guide interfaceSWnewnewn.a.n.a.
11. Flex 4D SpiralSWmodified⁶ (same as in K233650, with exception of the modifications regarding dynamic tube current modulation which is introduced with syngo CT VB20)modified (same as in K233650, with exception of the modifications regarding dynamic tube current modulation which is introduced with syngo CT VB20)cleared⁶cleared

⁵ In software version syngo CT VB20, the functionality HD FoV 5.0 is introduced only for the subject devices SOMATOM go.Sim and SOMATOM go.Open Pro.

⁶ The CT scanner CT systems SOMATOM go.Now and SOMATOM go.Up do not support Flex 4D Spiral.

K250822

Page 15

Technological propertyHW/SW changeSubject devicesPredicate devices
Single SourceDual SourceSingle SourceDual Source
SOMATOM go. Platform with syngo CT VB20SOMATOM Pro.Pulse with syngo CT VB20SOMATOM go. Platform with syngo CT VB10 (K233650)SOMATOM Pro.Pulse with syngo CT VB10 (K232206)
12. ZeeFree RTSWnew⁷n.a.n.a.n.a.
13. DirectDensitySWmodified (same as in K233650, with exception of the calculation of stopping-power ratio (Kernel St) which is introduced with syngo CT VB20)modified (same as in K233650, with exception of the calculation of stopping-power ratio (Kernel St) which is introduced with syngo CT VB20)clearedcleared
14. DirectLaserHW/SWmodified⁸ (same as in K233650, with exception of the Patient Marking workflow improvement introduced with syngo CT VB20)n.a.cleared⁸n.a.

⁷ In software version syngo CT VB20, the functionality ZeeFree RT is introduced only for the subject device SOMATOM go.Open Pro.

⁸ DirectLaser is available only for SOMATOM go.Sim and SOMATOM go.Open Pro.

K250822

Page 16

Technological propertyHW/SW changeSubject devicesPredicate devices
Single SourceDual SourceSingle SourceDual Source
SOMATOM go. Platform with syngo CT VB20SOMATOM Pro.Pulse with syngo CT VB20SOMATOM go. Platform with syngo CT VB10 (K233650)SOMATOM Pro.Pulse with syngo CT VB10 (K232206)
15. myExam ContrastSWnewnewn.a.n.a.

K250822

Page 17

A tabular summary of the comparable hardware and software properties between the subject devices SOMATOM go. Platform and SOMATOM Pro.Pulse with software version syngo CT VB20 and the predicate devices are listed in Table 4 through Table 8 below (modifications are in gray shaded sections).

Table 4 provides comparison of the hardware properties between the single source subject devices SOMATOM go.Now and SOMATOM go.Up with software version syngo CT VB20 and the predicate devices SOMATOM go.Now and SOMATOM go.Up with software version syngo CT VB10 (K233650).

Table 5 provides comparison of the hardware properties between the single source subject devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB20 and the predicate devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB10 (K233650).

Table 6 provides comparison of the hardware properties between the single source subject devices SOMATOM go.Sim and SOMATOM go.Open Pro with software version syngo CT VB20 and the predicate devices SOMATOM go.Sim and SOMATOM go.Open.Pro with software version syngo CT VB10 (K233650).

Table 7 provides comparison of the hardware properties between the dual source subject device SOMATOM Pro.Pulse with software version syngo CT VB20 and the predicate devices SOMATOM Pro.Pulse (K233650) and SOMATOM go.Top with software version syngo CT VB10 (233650).

Table 8 compares the software technological characteristics of the subject devices and the predicate devices.

Table 4: Comparison of technical hardware characteristics between the subject devices SOMATOM go.Now and SOMATOM go.Up with software version syngo CT VB20 and the primary predicate devices SOMATOM go.Now and SOMATOM go.Up with software version syngo CT VB10 (K233650).

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.NowSOMATOM go.UpSOMATOM go.NowSOMATOM go.Up
scannerWhole body Computed Tomography (CT) Scanner SystemWhole body Computed Tomography (CT) Scanner SystemSame as the primary predicate devices
System ConfigurationSingle SourceSingle SourceSame as the primary predicate devices

K250822

Page 18

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.NowSOMATOM go.UpSOMATOM go.NowSOMATOM go.Up
Environment of UseProfessional Healthcare FacilityProfessional Healthcare FacilitySame as the primary predicate devices
Generator Max. power (kW)3232Same as the primary predicate devices
Detector technologyStellar based on UFC (Ultra Fast Ceramic)Stellar based on UFC (Ultra Fast Ceramic)Same as the primary predicate devices
Detector volume coverage (mm)11.222.411.222.4Same as the primary predicate devices
Detector physical rows16321632Same as the primary predicate devices
Detector Slice width (mm)0.70.7Same as the primary predicate devices
Detector DAS channel no.768768Same as the primary predicate devices
Detector Image slices32643264Same as the primary predicate devices
Tube technologyChrononChrononSame as the primary predicate devices

K250822

Page 19

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.NowSOMATOM go.UpSOMATOM go.NowSOMATOM go.Up
Tube kV steps80, 110, 13080, 110, 130Same as the primary predicate devices
Tube Max. current (mA)400400Same as the primary predicate devices
Tube focus• 0.8 x 0.4 / 8° • 0.8 x 0.7 / 8°• 0.8 x 0.4 / 8° • 0.8 x 0.7 / 8°Same as the primary predicate devices
Tube Heat storage capacity (MHU)3.53.5Same as the primary predicate devices
Tube Cooling rate (kHU/ min)915915Same as the primary predicate devices
Gantry Bore size7070Same as the primary predicate devices
Gantry Scan FoV (cm)5050Same as the primary predicate devices
Gantry Extended FoV (cm)7070Same as the primary predicate devices
Gantry Rot. time (sec)0.5, 0.8, 1.0, 1.50.5, 0.8, 1.0, 1.5Same as the primary predicate devices

K250822

Page 20

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.NowSOMATOM go.UpSOMATOM go.NowSOMATOM go.Up
Gantry Tilt (degree)n.a.+/- 30° -30° to +28° with optional 307 kg patient tablen.a.+/- 30° -30° to +28° with optional 307 kg patient tableSame as the primary predicate devices
Patient table TypeVector (1250 mm), Vario 1 (1600 mm), Vario RT (1600 mm) with table extensionVario 1 (1600 mm), Vario 2 (2000 mm), Vario RT (1600 mm) with table extensionVector (1250 mm), Vario 1 (1600 mm), Vario RT (1600 mm) with table extensionVario 1 (1600 mm), Vario 2 (2000 mm), Vario RT (1600 mm) with table extensionSame as the primary predicate devices
Patient table Max. weight capacity (kg)Vector: 160 Vario 1: 227 Vario RT: 227Vario 1: 227 Vario 2: 307 Vario RT: 227Vector: 160 Vario 1: 227 Vario RT: 227Vario 1: 227 Vario 2: 307 Vario RT: 227Same as the primary predicate devices
Max. scan length topogram1680 mm with table extension1680 mm with table extension 2080 mm with table extension1680 mm with table extension1680 mm with table extension 2080 mm with table extensionSame as the primary predicate devices
Max. scan length Image acquisitionMax. 1600 mm with patient table extensionMax. 1600 mm with patient table extensionMax. 1600 mm with patient table extensionMax. 1600 mm with patient table extensionSame as the primary predicate devices

K250822

Page 21

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.NowSOMATOM go.UpSOMATOM go.NowSOMATOM go.Up
Max. 2000 mm with patient table extensionMax. 2000 mm with patient table extension
Spectral filtrationTin FilterTin FilterSame as the primary predicate devices
High Power 80yesyesyesyesSame as the primary predicate devices
FAST 3D Cameraoption for patient positioning with 3D Camera: • gantry mountedoption for patient positioning with 3D Camera: • ceiling mounted • gantry mountedoption for patient positioning with 3D Camera: • gantry mountedoption for patient positioning with 3D Camera: • ceiling mounted • gantry mountedSame as the primary predicate devices
x-ray foot switchOption to trigger hands-free scanningOption to trigger hands-free scanningSame as the primary predicate devices
Table foot switchOption for table patient movementOption for table patient movementSame as the primary predicate devices
i-joystickOption for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted.Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted.Same as the primary predicate devices

K250822

Page 22

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.NowSOMATOM go.UpSOMATOM go.NowSOMATOM go.Up
Tablet dockOption for mounting of the tablet on the patient table.Option for mounting of the tablet on the patient table.Same as the primary predicate devices
CARE Breathevisualization of breathing instructions to support three messages in respiratory motion management: • free breathing • preparation for breath hold • breath hold itself in combination with the visualization of a countdown visual patient instruction unit is placed on the gantry front and gantry back in the gantry funnelvisualization of breathing instructions to support three messages in respiratory motion management: • free breathing • preparation for breath hold • breath hold itself in combination with the visualization of a countdown visual patient instruction unit is placed on the gantry front and gantry back in the gantry funnelSame as the primary predicate devices
CARE 2D cameraone integrated into the gantry front and one mounted on the gantry back to monitor the patient during examinationone integrated into the gantry front and one mounted on the gantry back to monitor the patient during examinationSame as the primary predicate devices
CARE MoodlightLED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • ScanLED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Radiation onSame functionality as the primary predicate devices. In the subject devices with syngo CT VB20, the status "Radiation on" has been renamed to "Scan".

K250822

Page 23

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.NowSOMATOM go.UpSOMATOM go.NowSOMATOM go.Up
Respiratory Motion management - Interfaces for respiratory gatingn.a.acquisition of respiration-correlated scans supported Interface connectors for: • Anzai respiratory gating system Varian RGSC respiratory gating system • Open interface compatible respiratory gating systems • Open Online interfacen.a.acquisition of respiration-correlated scans supported Interface connectors for: • Anzai respiratory gating system Varian RGSC respiratory gating system • Open interface compatible respiratory gating systemsSame as the primary predicate device, the subject device SOMATOM go.Up with software version syngo CT VB20 support acquisition of respiration-correlated scans. For measurement of the breathing signal of the patient, interfaces for connecting of 3rd party respiratory gating devices are offered. Introduction of new Open Online Interface.

K250822

Page 24

Table 5: Comparison of technical hardware characteristics between the subject devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB20 and the primary predicate devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB10 (K233650).

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.AllSOMATOM go.TopSOMATOM go.AllSOMATOM go.Top
scannerWhole body Computed Tomography (CT) Scanner SystemWhole body Computed Tomography (CT) Scanner SystemSame as the primary predicate devices
System ConfigurationSingle SourceSingle SourceSame as the primary predicate devices
Environment of UseProfessional Healthcare FacilityProfessional Healthcare FacilitySame as the primary predicate devices
Generator Max. power (kW)7575Same as the primary predicate devices
Detector technologyStellar based on UFC (Ultra Fast Ceramic)Stellar based on UFC (Ultra Fast Ceramic)Same as the primary predicate devices
Detector volume coverage (mm)22.438.422.438.4Same as the primary predicate devices
Detector physical rows32643264Same as the primary predicate devices
Detector Slice width (mm)0.70.60.70.6Same as the primary predicate devices
Detector DAS channel no.768840768840Same as the primary predicate devices

K250822

Page 25

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.AllSOMATOM go.TopSOMATOM go.AllSOMATOM go.Top
Detector Image slices6412864128Same as the primary predicate devices
Tube technologyAthlonAthlonSame as the primary predicate devices
Tube kV steps70 to 140 (in 10 kV steps)70 to 140 (in 10 kV steps)Same as the primary predicate devices
Tube Max. current (mA)825825Same as the primary predicate devices
Tube focus• 1.1 x 1.2 / 8° • 0.8 x 0.8 / 8°• 1.1 x 1.2 / 8° • 0.8 x 0.8 / 8°Same as the primary predicate devices
Tube Heat storage capacity (MHU)77Same as the primary predicate devices
Tube Cooling rate (kHU/ min)Up to 1700Up to 1700Same as the primary predicate devices
Gantry Bore size7070Same as the primary predicate devices
Gantry Scan FoV (cm)5050Same as the primary predicate devices

K250822

Page 26

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.AllSOMATOM go.TopSOMATOM go.AllSOMATOM go.Top
Gantry Extended FoV (cm)7070Same as the primary predicate devices
Gantry Rot. time (sec)0.33, 0.5, 1.00.33, 0.5, 1.0Same as the primary predicate devices
Gantry Tilt (degree)+/- 30° -30° to +28° with optional 307 kg patient table+/- 30° -30° to +28° with optional 307 kg patient tableSame as the primary predicate devices
Patient table TypeVario 1 (1600 mm) Vario 2 (2000 mm) Vario RT (1600 mm) with table extensionVario 1 (1600 mm) Vario 2 (2000 mm) Vario RT (1600 mm) with table extensionSame as the primary predicate devices
Patient table Max. weight capacity (kg)Vario 1: 227 Vario 2: 307 Vario RT: 227Vario 1: 227 Vario 2: 307 Vario RT: 227Same as the primary predicate devices
Max. scan length topogram1680 mm with table extension 2080 mm with table extension1680 mm with table extension 2080 mm with table extensionSame as the primary predicate devices
Max. scan length Image acquisitionMax. 1600 mm with patient table extension Max. 2000 mm with patient table extensionMax. 1600 mm with patient table extension Max. 2000 mm with patient table extensionSame as the primary predicate devices

K250822

Page 27

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.AllSOMATOM go.TopSOMATOM go.AllSOMATOM go.Top
Spectral filtrationTin FilterCombined Tin Filter/Split filterTin FilterCombined Tin Filter/Split filterSame as the primary predicate devices
High Power 70/ High Power 80825 (@70 kV) 825 (@ 80 kV)825 (@70 kV) 825 (@ 80 kV)Same as the primary predicate devices
FAST 3D Cameraoption for patient positioning with FAST 3D Camera: • ceiling mounted • gantry mountedoption for patient positioning with FAST 3D Camera: • ceiling mounted • gantry mountedSame as the primary predicate devices
x-ray foot switchOption to trigger hands-free scanningOption to trigger hands-free scanningSame as the primary predicate devices
Table foot switchOption for table patient movementOption for table patient movementSame as the primary predicate devices
i-joystickOption for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted.Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted.Same as the primary predicate devices
Tablet dockOption for mounting of the tablet on the patient table.Option for mounting of the tablet on the patient table.Same as the primary predicate devices
CARE Breathevisualization of breathing instructions to support three messages in respiratory motion management:visualization of breathing instructions to support three messages in respiratory motion management:Same as the primary predicate devices

K250822

Page 28

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.AllSOMATOM go.TopSOMATOM go.AllSOMATOM go.Top
• free breathing • preparation for breath hold • breath hold itself in combination with the visualization of a countdown visual patient instruction unit is placed on the gantry front and gantry back in the gantry funnel• free breathing • preparation for breath hold • breath hold itself in combination with the visualization of a countdown visual patient instruction unit is placed on the gantry front and gantry back in the gantry funnel
CARE 2D cameraone integrated into the gantry front and one mounted on the gantry back to monitor the patient during examinationone integrated into the gantry front and one mounted on the gantry back to monitor the patient during examinationSame as the primary predicate devices
CARE MoodlightLED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • ScanLED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Radiation onSame functionality as the primary predicate devices In the subject devices with syngo CT VB20, the status "Radiation on" has been renamed to "Scan".
Respiratory Motion management - Interfaces for respiratory gatingacquisition of respiration-correlated scans supported Interface connectors for: • Anzai respiratory gating system • Varian RGSC respiratory gating systemacquisition of respiration-correlated scans supported Interface connectors for: • Anzai respiratory gating system • Varian RGSC respiratory gating systemSame as the primary predicate devices, the subject devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB20 support acquisition of respiration-correlated scans.

K250822

Page 29

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.AllSOMATOM go.TopSOMATOM go.AllSOMATOM go.Top
• Open interface compatible respiratory gating systems • Open Online interface• Open interface compatible respiratory gating systemsFor measurement of the breathing signal of the patient, interfaces for connecting of 3rd party respiratory gating devices are offered. Introduction of new Open Online Interface.

K250822

Page 30

Table 6: Comparison of technical hardware characteristics between the subject devices SOMATOM go.Sim and SOMATOM go.Open Pro with software version syngo CT VB20 and the primary predicate devices SOMATOM go.Sim and SOMATOM go.Open Pro with software version syngo CT VB10 (K233650).

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.SimSOMATOM go.Open ProSOMATOM go.SimSOMATOM go.Open Pro
scannerWhole body Computed Tomography (CT) Scanner SystemWhole body Computed Tomography (CT) Scanner SystemSame as the primary predicate devices
System ConfigurationSingle SourceSingle SourceSame as the primary predicate devices
Environment of UseProfessional Healthcare FacilityProfessional Healthcare FacilitySame as the primary predicate devices
Generator Max. power (kW)7575Same as the primary predicate devices
Detector technologyStellar based on UFC (Ultra Fast Ceramic)Stellar based on UFC (Ultra Fast Ceramic)Same as the primary predicate devices
Detector volume coverage (mm)19.238.419.238.4Same as the primary predicate devices
Detector physical rows32643264Same as the primary predicate devices
Detector Slice width (mm)0.60.60.60.6Same as the primary predicate devices

K250822

Page 31

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.SimSOMATOM go.Open ProSOMATOM go.SimSOMATOM go.Open Pro
Detector DAS channel no.920920920920Same as the primary predicate devices
Detector Image slices6412864128Same as the primary predicate devices
Tube technologyAthlonAthlonSame as the primary predicate devices
Tube kV steps70 to 140 (in 10 kV steps)70 to 140 (in 10 kV steps)Same as the primary predicate devices
Tube Max. current (mA)825825Same as the primary predicate devices
Tube focus• 1.1 x 1.2 / 7° • 0.8 x 0.8 / 7°• 1.1 x 1.2 / 7° • 0.8 x 0.8 / 7°Same as the primary predicate devices
Tube Heat storage capacity (MHU)77Same as the primary predicate devices
Tube Cooling rate (kHU/ min)Up to 1700Up to 1700Same as the primary predicate devices

K250822

Page 32

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.SimSOMATOM go.Open ProSOMATOM go.SimSOMATOM go.Open Pro
Gantry Bore size8585Same as the primary predicate devices
Gantry Scan FoV (cm)6060Same as the primary predicate devices
Gantry Extended FoV (cm)8585Same as the primary predicate devices
Gantry Rot. time (sec)0.35, 0.5, 1.00.35, 0.5, 1.0Same as the primary predicate devices
Gantry Tilt (degree)-30° to +25°-30° to +25°Same as the primary predicate devices
Patient table TypeVario 2 (2000 mm) Vario RT (1600 mm) with table extensionVario 2 (2000 mm) Vario RT (1600 mm) with table extensionSame as the primary predicate devices
Patient table Max. weight capacity (kg)Vario 2: 307 Vario RT: 227Vario 2: 307 Vario RT: 227Same as the primary predicate devices
Max. scan length topogram1680 mm with table extension 2080 mm with table extension1680 mm with table extension 2080 mm with table extensionSame as the primary predicate devices

K250822

Page 33

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.SimSOMATOM go.Open ProSOMATOM go.SimSOMATOM go.Open Pro
Max. scan length Image acquisitionMax. 1600 mm with patient table extension Max. 2000 mm with patient table extensionMax. 1600 mm with patient table extension Max. 2000 mm with patient table extensionSame as the primary predicate devices
Spectral filtrationTin FilterCombined Tin Filter/Split filterTin FilterCombined Tin Filter/Split filterSame as the primary predicate devices
High Power 70/ High Power 80825 (@70 kV) 825 (@ 80 kV)825 (@70 kV) 825 (@ 80 kV)Same as the primary predicate devices
FAST 3D Cameraoption for patient positioning with FAST 3D Camera: • ceiling mountedoption for patient positioning with FAST 3D Camera: • ceiling mountedSame as the primary predicate devices
CARE MoodlightLED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • ScanLED lights integrated on the gantry front cover indicating⁹: • System startup • Scan countdown • Radiation onCARE Moodlight is introduced for SOMATOM go.Sim and SOMATOM go.Open Pro In the subject devices with syngo CT VB20, the status "Radiation on" has been renamed to "Scan".

⁹ Added based on non-filing documentation (document-to-file) after the clearance of K233650. The modification has been considered as minor modification to the system and do not change the indications for use nor do significantly affect the safety and effectiveness of the device, hence non-filing documentation.

K250822

Page 34

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.SimSOMATOM go.Open ProSOMATOM go.SimSOMATOM go.Open Pro
x-ray foot switchOption to trigger hands-free scanningOption to trigger hands-free scanningSame as the primary predicate devices
Table foot switchOption for table patient movementOption for table patient movementSame as the primary predicate devices
i-joystickOption for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted.Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted.Same as the primary predicate devices
Tablet dockOption for mounting of the tablet on the patient table.Option for mounting of the tablet on the patient table.Same as the primary predicate devices
CARE Breathen.a.n.a.Same as the primary predicate devices
CARE 2D cameraPatient observation camera integrated on gantry frontPatient observation camera integrated on gantry frontSame as the primary predicate devices
Respiratory Motion management - Interfaces for respiratory gatingacquisition of respiration-correlated scans supported Interface connectors for:acquisition of respiration-correlated scans supported Interface connectors for:Same as the primary predicate devices, the subject devices SOMATOM go.Sim and SOMATOM go.Open Pro with software version syngo CT

K250822

Page 35

Hardware propertySubject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650)Assessment of the substantial equivalency (SE)
SOMATOM go.SimSOMATOM go.Open ProSOMATOM go.SimSOMATOM go.Open Pro
• Anzai respiratory gating system • Varian RGSC respiratory gating system • Open interface compatible respiratory gating systems • Open Online interface• Anzai respiratory gating system • Varian RGSC respiratory gating system • Open interface compatible respiratory gating systemsVB20 support acquisition of respiration-correlated scans. For measurement of the breathing signal of the patient, interfaces for connecting of 3rd party respiratory gating devices are offered. Introduction of new Open Online Interface.

K250822

Page 36

Table 7: Comparison of technical hardware characteristics between the subject device SOMATOM Pro.Pulse with software version syngo CT VB20 and the predicate devices SOMATOM Pro.Pulse (K232206) and SOMATOM go.Top with software version syngo CT VB10 (233650).

Hardware propertySubject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
scannerWhole body Computed Tomography (CT) Scanner SystemWhole body Computed Tomography (CT) Scanner SystemWhole body Computed Tomography (CT) Scanner SystemSame as the primary and the secondary predicate device
System ConfigurationDual SourceSingle SourceDual SourceSame as the secondary predicate device
Environment of UseProfessional Healthcare FacilityProfessional Healthcare FacilityProfessional Healthcare FacilitySame as the primary and the secondary predicate device
Generator Max. power (kW)150 (2 x 75)75150 (2 x 75)Same as the secondary predicate device
Detector technologyStellar detector based on Multislice UFC (Ultra Fast Ceramic)Stellar detector based on Multislice UFC (Ultra Fast Ceramic)Stellar detector based on Multislice UFC (Ultra Fast Ceramic)Same as the primary and the secondary predicate device
Detector volume coverage38.4 mm38.4 mm38.4 mmSame as the primary and secondary predicate device
Detector physical rows2 x 64642 x 64Same as the primary and the secondary predicate device

¹⁰ The primary predicate device of this eSTAR submission is the SOMATOM go. Platform with syngo CT VB10 (K233650). K233650 is a bundle 510(k) which the SOMATOM go.Top is part of.

K250822

Page 37

Hardware propertySubject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
Detector Slice width0.6 mm0.6 mm0.6 mmSame as the primary and secondary predicate device
Detector DAS channel no.840 (detector A) 600 (detector B)840840 (detector A) 600 (detector B)Same as the secondary predicate device
Detector Image slices2 x 128 (with Interleaved Volume Reconstruction, IVR) 2 x 384 (reconstructed slices)128 (with Interleaved Volume Reconstruction, IVR) 384 (reconstructed slices)2 x 128 (with Interleaved Volume Reconstruction, IVR) 2 x 384 (reconstructed slices)Same as the secondary predicate device
Tube technology2 x Athlon DS X-ray tubeAthlon X-ray tube2 x Athlon DS X-ray tubeSame as the secondary predicate device.
Tube collimatorTube collimator DS4-A, DS4-B • Equivalent to 0.5 mm Al in the isocenter • 1 mm Al with cardio wedgeTube collimator • Equivalent to 0.5 mm Al in the isocenter • 1 mm Al with cardio wedgeTube collimator DS4-A, DS4-B • Equivalent to 0.5 mm Al in the isocenter • 1 mm Al with cardio wedgeSame as the secondary predicate device.
Tube kV steps70–140 kV in 10 kV steps70–140 kV in 10 kV steps70–140 kV in 10 kV stepsSame as the primary and the secondary predicate device
Tube Max. currentSingle Source: 13-825 mA Dual Source: 26-1650 mAStandard range: 13 – 825 mASingle Source: 13-825 mA Dual Source: 26-1650 mASame as the secondary predicate device.

K250822

Page 38

Hardware propertySubject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
Tube focus• 0.8 × 0.8 / 7° • 1.0 × 1.2 / 7°• 0.8 × 0.8 / 7° • 1.0 × 1.2 / 7°• 0.8 × 0.8 / 7° • 1.0 × 1.2 / 7°Same as the primary and the secondary predicate device.
Tube Heat storage capacity2 x 7.0 MHU7.0 MHU2 x 7.0 MHUSame as the primary and the secondary predicate device.
Tube Cooling rate (MHU/ min)up to 1.7up to 1.7up to 1.7Same as the primary and the secondary predicate device.
Gantry Bore size707070Same as the primary and the secondary predicate device
Gantry Scan FoV (cm)505050Same as the primary and the secondary predicate device
Gantry Extended FoV (cm)707070Same as the primary and the secondary predicate device
Gantry Rot. time (sec)0.33 (optional), 0.5, 1.00.33 (optional), 0.5, 1.00.33 (optional), 0.5, 1.0Same as the primary and the secondary predicate device
Gantry Tilt (degree)n.a.± 30°n.a.Same as the secondary predicate device.
Maximum temporal resolution in ECG gated or triggered examination (ms)86 ms (mono-segment) 43 ms (bi-segment)165 ms (mono-segment) 82.5 ms (bi-segment)86 ms (mono-segment) 43 ms (bi-segment)Same as the secondary predicate device

K250822

Page 39

Hardware propertySubject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
Maximum scan speed at pitch (mm/s at pitch x)372 mm/s at pitch 3.2175 mm/s at pitch 1.5372 mm/s at pitch 3.2Same as the secondary predicate device
Patient table TypeVario 2 (2000 mm) Vario 2.D (2000 mm)Vario1 (1600 mm) Vario RT (1600 mm) Vario 2 (2000mm)Vario 2 (2000 mm) Vario 2.D (2000 mm)Same as the secondary predicate device.
Patient table Max. weight capacity (kg)307227 307307Same as the secondary predicate device.
Max. scan length topogram2080 mm with table extension1680 mm with table extension 2080 mm with table extension2080 mm with table extensionSame as the secondary predicate device.
Max. scan length Image acquisitionMax. 2000 mm with patient table extensionMax. 1600 mm with patient table extension Max. 2000 mm with patient table extensionMax. 2000 mm with patient table extensionSame as the secondary predicate device.

K250822

Page 40

Hardware propertySubject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
Spectral filtrationTin FilterCombined Tin Filter / Split FilterTin FilterSame as the secondary predicate device
FAST 3D Cameraoption for patient positioning with FAST 3D Camera: • Ceiling mounted • Gantry mountedoption for patient positioning with FAST 3D Camera: • ceiling mounted • gantry mountedoption for patient positioning with FAST 3D Camera: • Ceiling mounted • Gantry mountedSame as the primary and the secondary predicate device
High Power 70/ High Power 80up to 825 mA (@70 kV)/ up to 825 mA (@80 kV)up to 825 mA (@70 kV)/ up to 825 mA (@80 kV)up to 825 mA (@70 kV)/ up to 825 mA (@80 kV)Same as the primary and the secondary predicate device
x-ray foot switchOption to trigger hands-free scanningOption to trigger hands-free scanningOption to trigger hands-free scanningSame as primary and the secondary predicate device
Table foot switchOption for table patient movementOption for table patient movementOption for table patient movementSame as the primary and the secondary predicate device
i-joystickOption for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted.Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted.Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted.Same as the primary and the secondary predicate device
Tablet dockOption for mounting of the tablet on the patient table.Option for mounting of the tablet on the patient table.Option for mounting of the tablet on the patient table.Same as the primary and the secondary predicate device

K250822

Page 41

Hardware propertySubject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
CARE 2D cameraone integrated into the gantry front and one mounted on the gantry back to monitor the patient during examinationone integrated into the gantry front and one mounted on the gantry back to monitor the patient during examinationone integrated into the gantry front and one mounted on the gantry back to monitor the patient during examinationSame as the primary and the secondary predicate device
CARE MoodlightLED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • ScanLED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Radiation onLED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Radiation onSame as the primary and the secondary predicate device. In the subject device with syngo CT VB20, the status "Radiation on" has been renamed to "Scan".
Moodlight• Gantry ring Moodlight: ambient light during examination. • CARE Funnel Light: ambient light which illuminates the gantry funnel• Gantry ring Moodlight: ambient light during examination.• Gantry ring Moodlight: ambient light during examination. • CARE Funnel Light: ambient light which illuminates the gantry funnelSame as the secondary predicate device

K250822

Page 42

Table 8: Comparison of software characteristics between the subject devices with software version syngo CT VB20 and the predicate devices with software version syngo CT VB10 (K233650 and K232206).

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
Operating SystemWindows based SOMARIS/10 syngo CT VB20 Note: the short version syngo CT VB20 is also used as labeling informationWindows based SOMARIS/10 syngo CT VB20 Note: the short version syngo CT VB20 is also used as labeling informationWindows based SOMARIS/10 syngo CT VB10 Note: the short version syngo CT VB10 is also used as labeling informationWindows based SOMARIS/10 syngo CT VB10 Note: the short version syngo CT VB10 is also used as labeling informationSoftware version upgraded due to new functionalities.
Workplacesyngo Acquisition Workplace (ICS) named as "myExam Console" Image Reconstruction System (IRS)syngo Acquisition Workplace (ICS) named as "myExam Console" Image Reconstruction System (IRS)syngo Acquisition Workplace (ICS) named as "myExam Console" Image Reconstruction System (IRS)syngo Acquisition Workplace (ICS) named as "myExam Console" Image Reconstruction System (IRS)Same as the primary and secondary predicate devices

K250822

Page 43

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
2nd workplace option named as "myExam Satellite" with Remote Recon function2nd workplace option named as "myExam Satellite" with Remote Recon function2nd workplace option named as "myExam Satellite" with Remote Recon function2nd workplace option named as "myExam Satellite" with Remote Recon function
Standard system software• syngo Examination • syngo Viewing • syngo Filming • syngo Archiving & Network• syngo Examination • syngo Viewing • syngo Filming • syngo Archiving & Network• syngo Examination • syngo Viewing • syngo Filming • syngo Archiving & Network• syngo Examination • syngo Viewing • syngo Filming • syngo Archiving & NetworkSame as the primary and secondary predicate devices
Detector firmwareStellar detector firmware supportedStellar detector firmware supportedStellar detector firmware supportedStellar detector firmware supportedSame as the primary and secondary predicate devices

K250822

Page 44

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
TeamplaySupport of Teamplay protocolsSupport of Teamplay protocolsSupport of Teamplay protocolsSupport of Teamplay protocolsSame as the primary and secondary predicate devices
ProtocolsSupport of: • Protocols for Radiation Therapy Planning support patient marking • Protocols that allow scanning with support of an external respiratory gating systemSupport of: • Protocols for Radiation Therapy Planning support patient marking • Protocols supporting contrast bolus-triggered data acquisitionSupport of: • Protocols for Radiation Therapy Planning support patient marking • Protocols that allow scanning with support of an external respiratory gating systemSupport of: • Protocols for Radiation Therapy Planning support patient marking • Protocols supporting contrast bolus-triggered data acquisitionSOMATOM go. Platform: Same as the primary predicate devices SOMATOM Pro.Pulse:

K250822

Page 45

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
• Protocol supporting contrast bolus-triggered data acquisition • Contrast media protocols (including coronary CTA) • Pediatric Protocols • Flex Dose Profile • Turbo Flash Spiral • Dual Energy acquisition (TwinBeam DE and TwinSpiral DE) • Dynamic imaging (Flex 4D Spiral)• Contrast media protocols (including coronary CTA) • Pediatric Protocols • Flex Dose Profile • Dynamic imaging (Flex 4D Spiral) • interventional scan protocols (scan modes: i-sequence, i-spiral, i-Fluoro) • Protocols supporting Cardiac Scanning, Spectral imaging for child examination, Spectral imaging with high resolution• Protocol supporting contrast bolus-triggered data acquisition • Contrast media protocols (including coronary CTA) • Pediatric Protocols • Flex Dose Profile • Turbo Flash Spiral • Dual Energy acquisition (TwinBeam DE and TwinSpiral DE) • Dynamic imaging (Flex 4D Spiral)• Contrast media protocols (including coronary CTA) • Pediatric Protocols • Flex Dose Profile • Dynamic imaging (Flex 4D Spiral) • interventional scan protocols (scan modes: i-sequence, i-spiral, i-Fluoro) • Protocols supporting Cardiac Scanning, Spectral imaging for child examination, Spectral imaging with high resolutionSame as the secondary predicate device

K250822

Page 46

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
• Protocols supporting CT Intervention (scan modes: 2D i-sequence, 3D i-sequence, i-spiral, i-Fluoro) • Protocols supporting Cardiac Scanning • Protocols for DirectBreathhold • Turbo Flash Spiral • various i-spiral and i-sequence scan protocols (applying different scanning parameters such as different slice thicknesses, kV settings or reconstructions kernels to support different clinical scenarios) and using features like iMAR, CARE Dose4D and CARE kV• Protocols supporting CT Intervention (scan modes: i-sequence, i-spiral, i-Fluoro) • Protocols supporting Cardiac Scanning • Protocols for DirectBreathhold • Turbo Flash Spiral • various i-spiral and i-sequence scan protocols (applying different scanning parameters such as different slice thicknesses, kV settings or reconstructions kernels to support different clinical scenarios) and using features like iMAR, CARE Dose4D and CARE kV

K250822

Page 47

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
• TwinkV (Dual Source Dual Energy acquisition)• TwinkV (Dual Source Dual Energy acquisition)
Advanced ReconstructionRecon&GO: - Spectral Recon (Dual Energy Reconstruction including Virtual Unenhanced, Monoenergetic plus) - DE SPP (Spectral Post-Processing) - Inline Anatomical ranges (Parallel/Radial) incl. VirtualRecon&GO: - Spectral Recon (Dual Energy Reconstruction including None, low (L3D) and high (H3D) energy images) - DE SPP (Spectral Post-Processing) - Inline Anatomical ranges (Parallel/Radial) incl. VirtualRecon&GO: - Spectral Recon (Dual Energy Reconstruction including Virtual Unenhanced, Monoenergetic plus) - DE SPP (Spectral Post-Processing) - Inline Anatomical ranges (Parallel/Radial) incl. VirtualRecon&GO: - Spectral Recon (Dual Energy Reconstruction including None, low (L3D) and high (H3D) energy images) - DE SPP (Spectral Post-Processing) - Inline Anatomical ranges (Parallel/Radial) incl. VirtualSOMATOM go. Platform: Same as the primary predicate devices SOMATOM Pro.Pulse: Same as the secondary predicate device

K250822

Page 48

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
Unenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removalUnenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removalUnenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removalUnenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removal
Image viewingCT View&GO offers: - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI)CT View&GO offers: - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI)CT View&GO offers: - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI)CT View&GO offers: - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI)Same as the primary and the secondary predicate devices

K250822

Page 49

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
- Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average- Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average- Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average- Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average
Software interface• Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) cleared as medical devices in their own right.• Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) cleared as medical devices in their own right.• Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) cleared as medical devices in their own right.• Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) cleared as medical devices in their own right.In software version syngo CT VB20, the subject devices introduce plug-in interface for the stand alone software application myAblation Guide.

K250822

Page 50

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
• software interfaces for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. Note: The clearance of standalone Advanced Visualization Application software is mandatory precondition. These advanced visualization tools are designed to support the technician & physician in• software interfaces for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. Note: The clearance of standalone Advanced Visualization Application software is mandatory precondition. These advanced visualization tools are designed to support the technician & physician in• software interfaces for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. Note: The clearance of standalone Advanced Visualization Application software is mandatory precondition. These advanced visualization tools are designed to support the technician & physician in• software interfaces for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. Note: The clearance of standalone Advanced Visualization Application software is mandatory precondition. These advanced visualization tools are designed to support the technician & physician in

K250822

Page 51

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
the qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners. • Plug-in interface for myAblation Guidethe qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners. • Plug-in interface for myAblation Guidethe qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners.the qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners.
CybersecurityIT HardeningIT HardeningIT HardeningIT HardeningSame as the primary and the secondary predicate devices
HD FoVHD FoV 4.0HD FoV 4.0HD FoV 4.0HD FoV 4.0The subject devices

K250822

Page 52

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
HD FoV 5.0¹¹SOMATOM go. Sim and SOMATOM go.Open Pro with syngo CT VB20 support a new generation of extended field of view reconstruction algorithm: HD FoV 5.0.

¹¹ HD FoV 5.0: In software version syngo CT VB20, this version of HD FoV is available for SOMATOM go.Sim and SOMATOM go.Open Pro only.

K250822

Page 53

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
Standard technologies• FAST technologies • CARE technologies • GO technologies• FAST technologies • CARE technologies • GO technologies• FAST technologies • CARE technologies • GO technologies• FAST technologies • CARE technologies • GO technologiesSame as the primary and the secondary predicate devices
Iterative Reconstruction MethodsADMIRE iMAR SAFIREADMIRE iMARADMIRE iMAR SAFIREADMIRE iMARSOMATOM go. Platform: Same as the primary predicate devices SOMATOM Pro.Pulse:

K250822

Page 54

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
Same as the secondary
Matrix sizes256 x 256 pixels 512 x 512 pixels 768 x 768 pixels 1024 x 1024 pixels (Precision Matrix)256 x 256 pixels 512 x 512 pixels 768 x 768 pixels 1024 x 1024 pixels (Precision Matrix)256 x 256 pixels 512 x 512 pixels 768 x 768 pixels 1024 x 1024 pixels (Precision Matrix)256 x 256 pixels 512 x 512 pixels 768 x 768 pixels 1024 x 1024 pixels (Precision Matrix)Same as the primary and the secondary predicate device
DirectDensityincluding relative electron density, relative mass density, and stopping power ratioincluding relative electron density, relative mass density, and stopping power ratioincluding relative electron density, relative mass densityincluding relative electron density and relative mass densityCompared to the predicate devices, the subject devices with syngo CT VB20 support a

K250822

Page 55

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
modified DirectDensity functionality. Same as the secondary predicate
Matrix sizes256 x 256 pixels256 x 256 pixels256 x 256 pixels256 x 256 pixelsSame as the primary and the secondary predicate device
512 x 512 pixels512 x 512 pixels512 x 512 pixels512 x 512 pixels
768 x 768 pixels768 x 768 pixels768 x 768 pixels768 x 768 pixels
1024 x 1024 pixels (Precision Matrix)1024 x 1024 pixels (Precision Matrix)1024 x 1024 pixels (Precision Matrix)1024 x 1024 pixels (Precision Matrix)
DirectDensityincluding relative electron density, relative mass density, and stopping power ratioincluding relative electron density, relative mass density, and stopping power ratioincluding relative electron density, relative mass densityincluding relative electron density and relative mass densityCompared to the predicate devices, the subject devices with syngo CT VB20 support a

K250822

Page 56

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
Stack artifact reduced reconstructionProvides images with reduced stack artifacts after the reconstruction: ZeeFree: StackRecon types in cardiac-related examinations ZeeFree RT¹²: StackRecon types in respiratory-related examinationsProvides images with reduced stack artifacts after the reconstruction: ZeeFree: StackRecon types in cardiac-related examinationsProvides images with reduced stack artifacts after the reconstruction: ZeeFree: StackRecon types in cardiac-related examinationsProvides images with reduced stack artifacts after the reconstruction: ZeeFree: StackRecon types in cardiac-related examinationsZeeFree RT is a new functionality introduced with syngo CT VB20 for the subject device SOMATOM go.Open Pro.

¹² ZeeFree RT: In software version syngo CT VB20, ZeeFree RT is available for SOMATOM go.Open Pro only.

K250822

Page 57

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
Oncology exchangetransfer of prescription information from the ARIA Oncology Information System (OIS) to the CT scannertransfer of prescription information from the ARIA Oncology Information System (OIS) to the CT scannern.a.n.a.New functionality introduced with syngo CT VB20.
DirectSetup Notespatient and accessory setup documentation in the Scan&GO workflowpatient and accessory setup documentation in the Scan&GO workflown.a.n.a.The subject devices SOMATOM go. Platform CT scanner systems are enabled to support DirectSetup Notes.

K250822

Page 58

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
DirectSetup Notes was cleared in K233650 for SOMATOM go.Sim and SOMATOM go.Open Pro.
FAST Integrated Workflow (FAST 3D Camera)FAST integrated workflow: • FAST Range • FAST Isocentering • FAST Direction • Collision IndicationFAST integrated workflow: • FAST Range • FAST Isocentering • FAST Direction • Collision IndicationFAST integrated workflow: • FAST Range • FAST Isocentering • FAST DirectionFAST integrated workflow: • FAST Range • FAST Isocentering • FAST DirectionCompared to the predicate devices, the FAST 3D Camera of the subject devices supports in syngo CT VB20 a Collision indication and

K250822

Page 59

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
• Centerline/ Grid Overlay• Centerline/ Grid OverlayCenterline/Grid Overlay function. DirectSetup Notes was cleared in K233650 for SOMATOM go.Sim and SOMATOM go.Open Pro.
FAST Planningdetects globally used scan and recon ranges from the topogram image additional body regions: - Head/Neck (Vascular), Lateral/APPA - Vascular – Runoff Knee - Neck and chest - Neck and Lungdetects globally used scan and recon ranges from the topogram image additional body regions: - Head/Neck (Vascular), Lateral/APPA - Vascular – Runoff Knee - Neck and chest - Neck and Lungdetects globally used scan and recon ranges from the topogram imagedetects globally used scan and recon ranges from the topogram imageAs the primary and secondary devices, the subject devices with syngo VB20 support FAST Planning functionality with wide range of body regions. In syngo CT VB20, FAST Planning has been extended

K250822

Page 60

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
- Extremities – Both knees- Extremities – Both kneeswith additional body regions.
Eco Power Modereduce the energy consumption during idle times.¹³reduce the energy consumption during idle times.n.a.n.a.New feature introduced with syngo CT VB20 of the subject devices.
myExam Contrastexchange of contrast injection parameters, scan and patient parametersexchange of contrast injection parameters, scan and patient parametersn.a.n.a.New feature introduced with syngo CT VB20.

¹³ SOMATOM go.Now and SOMATOM go.Up do not support Eco Power Mode.

K250822

Page 61

Software propertySubject device with SOMARIS/10 syngo CT VB20Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650)Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206)Assessment of the substantial equivalency (SE)
SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.PulseSOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open ProSOMATOM Pro.Pulse
Breathhold techniqueRespiratory Motion Management support breath hold triggered spiral scans with manual breath hold triggered examinations. DirectBreathold: externally triggered workflow to start a spiral scan by receiving a trigger signal from an external gating device.n.a.Respiratory Motion Management support breath hold triggered spiral scans with manual breath hold triggered examinations. DirectBreathold: externally triggered workflow to start a spiral scan by receiving a trigger signal from an external gating device.n.a.SOMATOM go. Platform: Same as the primary predicate devices SOMATOM Pro.Pulse does not support DirectBreathhold

Any differences in technological characteristics do not raise different questions of safety and effectiveness. Testing and validation is completed. Test results show that the subject devices SOMATOM go. Platform and SOMATOM Pro.Pulse with syngo CT VB20 are comparable to the predicate devices in terms of technological characteristics and safety and effectiveness and therefore are substantially equivalent to the primary predicate devices.

K250822

Page 62

VIII. Performance Data

Non-Clinical Testing

Non-clinical testing, (integration and functional) including phantom tests were conducted for the SOMATOM go. Platform and SOMATOM Pro.Pulse CT scanner systems with syngo CT VB20 during product development. The modifications described in this Premarket Notification were supported with verification and validation testing.

The general purpose of each test is to verify and validate the functionality of the subject device modifications.

Testing will cover all related subsystems that contribute to the device modifications. Test levels are defined. For each test level several test activities are performed. The test specification and acceptance criteria are related to the corresponding requirements. Various test activities are performed to specific modifications on different test levels to ensure safe and effective integration in the system. Three test levels are defined:

System Validation test:

  • Acceptance test (workflow and user manual test)
  • Legal and Regulatory test

System Verification test:

  • System Integration Test (functional)
  • Functionality verification
  • Image Quality (IQ) Evaluation

Tests are conducted for all software components developed in product development and for the complete product itself. Several activities are considered for this process, including creation of test specifications that relate to software/hardware requirements including tests to address risk mitigations that are identified, documented, and traced by hazard keys.

Bench testing:

Additional evaluation tests are performed as bench tests to support the new device or device modification on Non-Clinical Performance Testing as listed in Table 9 below.

Table 9: Non-clinical performance testing (bench testing).

Feature/Non-clinical supportive testingBench Testing performed
FAST 3D Camera/ FAST Integrated WorkflowThe FAST 3D camera sub-features FAST Isocentering, FAST Range, and FAST Direction have been optimized using additional data from adults and adolescence patients. The bench test evaluates and compares the accuracy of the three sub-features in software version syngo CT VB20 to the accuracy of the predicate devices with syngo CT VB10. The objectives of the bench tests are to demonstrate that the FAST 3D camera achieves comparable or better results for both, adults and adolescents, as the predicate device for adults.

K250822

Page 63

Feature/Non-clinical supportive testingBench Testing performed
Overall, the subject devices with syngo CT VB20 delivers comparable or improved accuracy to the predicate devices with syngo CT VB10 predicate device for adults and extends the support to adolescents.
FAST PlanningThe purpose of the test is to provide a clear reporting on the applied algorithm, its product development, validation, and verification on patient data, which enable the claims. Objective of the test is to assess the fraction (percentage) of ranges calculated by the FAST Planning algorithm that are correct and can be applied without change. Additionally, calculation time was measured to check whether it meets interactive requirements. The test results show that the editing actions for the scanner technician can be reduced to a minimum and that the calculation time is fast enough for interactive speed during scanning. For more than 90% of the ranges no editing action was necessary to cover standard ranges. For more than 95%, the speed of the algorithm was sufficient.
HD FoV 5.0The bench test contains a detailed description and evaluation of the new HD FoV 5. 0 algorithm for extended field of view reconstruction. Results obtained with the new HD FoV 5. 0 algorithm are compared with its predecessor, the HD FoV 4. 0 algorithm, based on physical and anthropomorphic phantoms. In addition to the bench test, the performance of the algorithm was evaluated by board-approved radio-oncologists and medical physicists by means of retrospective blinded rater study. This comparison is conducted to demonstrate that the HD FoV 5. 0 algorithm is as safe and effective as the HD FoV 4. 0 algorithm.
Flex 4D SpiralThe performed bench test report describes the technical background of Flex 4D Spiral and its functionalities with SOMATOM CT scanners, demonstrate the proper function of those, and assess the image quality of Flex 4D Spiral.
ZeeFree RTThe bench test evaluates the performance of the ZeeFree RT reconstruction. The objectives of the tests are to demonstrate that compared to the Standard reconstruction, ZeeFree RT • introduces no relevant errors in terms of CT values and noise levels measured in a homogeneous water phantom, and • introduces no relevant errors in terms of CT values measured in a phantom with tissue-equivalent inserts, even in the presence of metals and in combination with the iMAR algorithm, and • introduces no relevant geometrical distortions in a static torso phantom, and

K250822

Page 64

Feature/Non-clinical supportive testingBench Testing performed
• introduces no relevant deteriorations of the position or shape of a dynamic thorax phantom when moving a spherical shape according to regular, irregular, and patient breathing motion. In addition to the bench tests, the performance of the algorithm was evaluated by board-approved radio-oncologists and medical physicists by means of retrospective blinded rater studies of respiratory 4D CT examinations performed at two institutions. The bench test results show that the ZeeFree RT reconstruction • can successfully be applied to 4D respiratory-gated sequence images (Direct i4D), and • enables the optional reconstruction of stack artefact corrected images, which reduce the strength of misalignment artefacts, if such stack alignment artefacts are identified in non-corrected standard images, and • does not introduce relevant new artefacts, which were previously not present in the non-corrected standard reconstruction, and • does not affect CT values and noise levels in a homogenous water phantom outside of stack-transition areas compared to the non-corrected standard reconstruction, and • can be successfully applied in combination with metal artifact correction (iMAR) and does not introduce new artifacts, which were previously not present in the non-corrected standard reconstruction, even in presence of metals, and • can be successfully applied to phantom data if derived from a suitable motion phantom demonstrating its correct technical function on the tested device, and • is independent from the physical detector width of the acquired data. By design, input images for the correction part of the algorithm are independent of kernel, slice thickness and increment since those parameters are fixed.
DirectDensityThe test results show for the iBHC variants Artificial120, eDDensity, mDDensity and StoppingPowerRatio a reduced dependence on tube voltage and filtration compared to the corresponding quantitative kernel (Qr) with iBHC Bone for non-water-like tissues, such as adipose and bone. Furthermore, the iBHC variants Artificial120, eDDensity, mDDensity and StoppingPowerRatio generate image value closely aligned with the respective material properties. In conclusion, the feature DirectDensity at any kV has been validated for the release Som/X VB20 on all supported SOMATOM CT scanner models

K250822

Page 65

A list of recognized consensus standards and general use consensus standards considered for the subject devices is provided as Table 10 and Table 11 below.

Table 10: Recognized Consensus Standards.

Date of EntryRecognition NumberStandard Developing OrganizationStandard Designation Number and DateTitle of Standard
12/18/202312-352NEMAPS 3.1 - 3.20 2023eDigital Imaging and Communications in Medicine (DICOM) Set
07/06/202012-325NEMAXR 25-2019Computed Tomography Dose Check
07/06/202012-330NEMAXR 28-2018Supplemental Requirements for User Information and System Function Related to Dose in CT
12/23/201912-328IEC61223-3-5 Edition 2.0 2019-09Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance tests and constancy tests - Imaging performance of computed tomography X-ray equipment [Including: Technical Corrigendum 1 (2006)]
03/14/201112-226IEC61223-2-6 Second Edition 2006-11Evaluation and routine testing in medical imaging departments - Part 2-6: Constancy tests - Imaging performance of computed tomography X-ray equipment
06/07/202112-336IEC60601-1-3 Edition 2.2 2021-01 CONSOLIDATED VERSIONMedical electrical equipment - Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment
06/27/201612-302IEC60601-2-44 Edition 3.2: 2016Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and

K250822

Page 66

Date of EntryRecognition NumberStandard Developing OrganizationStandard Designation Number and DateTitle of Standard
essential performance of x-ray equipment for computed tomography
12/23/20195-125ANSI AAMI ISO14971: 2019Medical devices - Applications of risk management to medical devices
ISO 14971 Third Edition 2019-12Medical devices - Application of risk management to medical devices
01/14/201913-79ANSI AAMI IEC62304:2006/A1:2016Medical device software - Software life cycle processes [Including Amendment 1 (2016)]
IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSIONMedical device software - Software life cycle processes
05/30/202219-46ANSI AAMIES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
12/21/202019-36ANSI AAMI IEC60601-1-2:2014 [Including AMD 1:2021]Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSIONMedical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

K250822

Page 67

Date of EntryRecognition NumberStandard Developing OrganizationStandard Designation Number and DateTitle of Standard
07/06/20205-129ANSI AAMI IEC62366-1:2015+AMD1:2020 (Consolidated Text)Medical devices Part 1: Application of usability engineering to medical devices, including Amendment 1
IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSIONMedical devices - Part 1: Application of usability engineering to medical devices
07/09/201412-273IEC60825-1 Edition 2.0 2007-03Safety of laser products - Part 1: Equipment classification, and requirements
12/21/20205-132IEC60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSIONMedical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
12/23/201912-309IEC60601-2-28 Edition 3.0 2017-06Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
12/20/202112-341IEC62563-1 Edition 1.2 2021-07 CONSOLIDATED VERSIONMedical electrical equipment - Medical image display systems - Part 1: Evaluation methods

Table 11: General Use Consensus Standards.

Standard Developing OrganizationStandard Designation Number and DateTitle of StandardHow was Standard Used
IEC60601-1:2005+A1:2012+A2:2020Medical electrical equipment - part 1: general requirements for basic safety and essential performanceANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 &

K250822

Page 68

Standard Developing OrganizationStandard Designation Number and DateTitle of StandardHow was Standard Used
A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]
IEC/ISO17050-1Conformity Assessment – Supplier's declaration of conformity – Part 1: General requirementsDeclaration of conformance to FDA recognized consensus standards.
IEC/ISO17050-2Conformity assessment – Supplier's declaration of conformity – Part 2: Supporting documentation.General consensus standards not currently recognized by FDA.

A list of applicable guidance documents considered for this submission is provided as Table 12 below.

Table 12: FDA Guidance Document and Effective Date.

FDA Guidance DocumentIssue date
User Fees and Refunds for Premarket Notification Submissions (510(k)s): Guidance for Industry and Food and Drug Administration Staff10/05/2022
Refuse to Accept Policy for 510(k)s: Guidance for Industry and Food and Drug Administration Staff04/21/2022
Electronic Submission Template for Medical Device 510(k) Submissions10/2/2023
Deciding When to Submit a 510(k) for a Change to an Existing Device10/25/2017
The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]07/28/2014
Content of Premarket Submissions for Software Contained in Medical Devices06/14/2023
Off-The-Shelf Software Use in Medical Devices09/27/2019
Applying Human Factors and Usability Engineering to Medical Devices02/03/2016
Pediatric Information for X-ray Imaging Device Premarket Notifications11/28/2017
Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions09/27/2023
Electromagnetic Compatibility (EMC) of Medical Devices06/06/2022
Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices09/06/2017

K250822

Page 69

FDA Guidance DocumentIssue date
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices09/14/2018

Verification and Validation

The performance data demonstrates continued conformance with special controls for medical devices containing software. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claims of substantial equivalence.

Cybersecurity

Siemens conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient.

Wireless Coexistence Testing

Additionally, Siemens conforms to the requirements for Radio Frequency Wireless Technology as defined in FDA guidance document "Radio Frequency Wireless Technology in Medical Devices, Guidance for Industry and Food and Drug Administration Staff" issued on August 14, 2013 by adhering to the EMC and risk based verification and validation requirements in design, testing, and labeling of the wireless remote control components of the subject devices.

The Radio Frequency Wireless Technology of the optional Remote Scan Control and supporting Control Device iPad for Scan&GO complies to 47 CFR part 15 subpart c – Intentional Radiators. All Radio device labels will show an FCC ID code to show compliance. Shielding requirement applicable to the SOMATOM go. Platform and SOMATOM Pro.Pulse and respective Scatter Radiation diagrams for typical room installations are provided in the User Documentation and Planning Guide of the intended Scanners in accordance with IEC60601-2-44.

Siemens has considered several measures to address wireless coexistence by design to ensure the safe operation of the wireless components in combination with the applicable system supported functionality. Wireless technology in the system setup to perform a task in a given shared environment where other systems have an ability to perform their tasks and may or may not be using the same set of rules has been considered. According to FDA guidance 'Radio Frequency Wireless Technology in Medical Devices" Siemens has addressed the safety, effectiveness, and high likelihood of coexistence with other devices of this technology in our product design by our Risk Management Process, Failure Mode and Effects Analysis (FMEA) Process, and Requirement Engineering Process. As part of the risk management process, hazardous situations associated with the Scan&GO and its connection to the host system via Wi-Fi were addressed as part of the Risk Management process.

Testing for co-existence considered for following scenarios:

  • Co-Channel Testing
  • Adjacent Channel Testing

K250822

Page 70

  • RF Interference Testing
  • Separation Distance/Location Testing

Scan&GO is designed to allow dynamic frequency selection and transmission power control by default in accordance with IEEE 802.11h. Adjacent channel testing is addressed by the fact that Scan&GO does not support shared medium access to Siemens Wi-Fi network. RF interference was tested by successfully ensuring that wireless communications were actively transmitting in situations where possible interference may exist. Recommended distance and router locations requirements are documented in the user documentation.

Summary

The features described in this premarket notification are supported with verification and validation testing, dosimetry and imaging performance, and analysis of phantom images to assess device and feature performance during product development. The risk analysis was completed, and risk control implemented to mitigate identified hazards. The test results show that all of the software specifications have met the acceptance criteria. Verification and validation testing of the device was found acceptable to support the claim of substantial equivalence.

General Safety and Effectiveness Concerns

The device labeling contains instructions for use as well as necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a system related risk analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing according to the risk management process. In order to minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.

IX. Conclusions

Verification and validation and phantom testing were performed. The non-clinical data supports the safety of the device and the hardware and software verification and validation demonstrates that the subject devices SOMATOM go. Platform and SOMATOM Pro.Pulse CT scanner systems with software version syngo CT VB20 perform as intended in the specified use conditions. The data included in this submission demonstrates that the SOMATOM go. Platform and SOMATOM Pro.Pulse CT scanner systems with software version syngo CT VB20 with the described modifications performs comparably to the predicate devices currently marketed for the same intended use. The conclusion drown from the non-clinical tests demonstrates that the subject devices are as safe, as effective, and perform as well as or better than the predicate devices.

K250822

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.