K Number
K233650

Validate with FDA (Live)

Date Cleared
2024-03-26

(133 days)

Product Code
Regulation Number
892.1750
Age Range
12 - 65
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.

This CT system can be used for low dose lung cancer screening in high risk populations*. *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Device Description

Siemens intends to market a new software version, SOMARIS/10 syngo CT VB10 for the following SOMATOM Computed Tomography (CT) Scanner Systems:

SOMATOM go. Platform CT scanner systems:

  • . SOMATOM go.Up
  • . SOMATOM go.Now
  • SOMATOM go.All
  • SOMATOM go.Top
  • . SOMATOM go.Sim
  • . SOMATOM go.Open Pro

SOMATOM X. Platform CT scanner systems:

  • . SOMATOM X.cite
  • . SOMATOM X.ceed

In this submission, the above listed CT scanner systems are jointly referred to as subject devices by "SOMATOM go. Platform" and "SOMATOM X. Platform" CT scanner systems.

The subject devices SOMATOM go. Platform and SOMATOM X. Platform with SOMARIS/10 syngo CT VB10 are Computed Tomography X-ray Systems which feature one continuously rotating tubedetector system and function according to the fan beam principle (single source). The SOMATOM go. Platform and SOMATOM X. Platform with software SOMARIS/10 syngo CT VB10 produces CT images in DICOM format, which can be used by trained staff for post-processing applications commercially distributed by Siemens Healthcare and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.

Only trained and qualified users, certified in accordance with country-specific regulations, are authorized to operate the system. For example, physicians, radiologists, or technologists. The user must have the necessary U.S. qualifications in order to diagnose or treat the patient with the use of the images delivered by the system.

The platform software for the SOMATOM go. Platform and SOMATOM X. Platform, SOMARIS/10 synqo CT VB10, is a command-based program used for patient management, data management, Xray scan control, image reconstruction, and image archive/evaluation.

The software platform provides plugin software interfaces that allow for the use of specific commercially available post processing software algorithms in an unmodified form from the cleared stand-alone post processing version.

New software version syngo CT VB10 (SOMARIS/10 syngo CT VB10) is a modified software version based on syngo CT VA40 (SOMARIS/10 syngo CT VA40) which was cleared for the predicate device in K211373.

Software version SOMARIS/10 syngo CT VB10 will be offered ex-factory and as an optional upgrade for the applicable existing SOMATOM go. Platform and SOMATOM X. Platform CT Scanner Systems.

The bundle approach is feasible for this submission since the subject devices have similar technological characteristics, software operating platform, and supported software characteristics. The supporting data are similar, primarily one review division/group will be involved, and the indications for use is the same between the devices. All subject devices will support previously cleared software and hardware features in addition to the applicable modifications as described within this submission. The intended use remains unchanged compared to the predicate devices.

AI/ML Overview

The provided text is a 510(k) summary for a Computed Tomography (CT) system. It focuses on demonstrating substantial equivalence to previously cleared predicate devices, primarily through non-clinical testing and comparison of technological characteristics. The document does not contain information about comparative effectiveness studies, multi-reader multi-case (MRMC) studies, or detailed clinical study results with ground truth establishment as one might find for a novel AI/ML-driven diagnostic device.

Therefore, many of the requested items (e.g., sample size for the test set, number of experts, adjudication methods, MRMC study effect size, training set details) are not explicitly mentioned in this type of submission. The focus here is on the CT system itself and its software updates, not on a new AI algorithm for detection or diagnosis where such detailed performance metrics against ground truth would be paramount.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria & Reported Device Performance

The document describes "bench testing" as non-clinical supportive testing for specific features. The acceptance criteria are generally qualitative (e.g., "comparable accuracy," "reduce the number of alignment artefacts," "successfully detect needle-tips") rather than specific numerical thresholds.

Feature / Non-clinical Supportive TestingAcceptance Criteria (Implicit from Objectives)Reported Device Performance (Test Results)
FAST 3D Camera / FAST Integrated WorkflowAccuracy of FAST Isocentering, FAST Range, and FAST Direction comparable to predicate device (syngo CT VA40) with old camera hardware and ceiling mount.FAST Isocentering: Comparable accuracy to predicate, regardless of camera mounting. FAST Range: Detection accuracy of body region boundaries comparable. (Note: Legs can be occluded by torso in gantry mounting, not severe limitation as leg exams are usually feet-first). FAST Direction: Pose detection results comparable accuracy. Overall: SOMARIS/10 syngo CT VB10 delivers comparable accuracy to predicate for new FAST 3D Camera hardware, also in new gantry position.
Multi-Purpose TableSufficient freedom of movement for a mobile C-arm X-ray system for clinical routine without significant limitations for myNeedle Lasers or FAST 3D Camera when installed with enhanced distance (674 mm) to CT gantry and offering iCT mode functionality.Technical feasibility and possible limitations evaluated. Concluded that the CT scanner with a Multi-Purpose (Vitus) Patient Table, enhanced distance (674 mm) and iCT mode, provides sufficient freedom of movement for a mobile C-arm X-ray system to be used for clinical routine without any significant limitations.
Direct BreathholdA spiral scan can be automatically triggered from an external respiratory gating device, with the actual scan remaining unchanged and the object correctly depicted.Test results showed a spiral scan can be automatically triggered, actual scan remains unchanged, and object is correctly depicted.
ZeeFreeReduce number of artifacts attributed to stack misalignment; no new artifacts introduced; equivalent image quality in quantitative standard physics phantom-based measurements (noise, homogeneity, high-contrast resolution, slice thickness, CNR); equivalent image quality in quantitative and qualitative phantom-based measurements for metal objects; algorithm successfully applied to phantom data demonstrating correct technical function; algorithm independent from physical detector width.If misalignment artifacts identified, "Cardiac Stack Artefact Correction" (ZeeFree) enables optional stack artifact corrected images which reduce number of alignment artifacts. Does not introduce new artifacts. Realizes equivalent image quality in quantitative standard physics phantom-based measurements (ACR, Gammex phantom) in terms of noise, homogeneity, high-contrast resolution, slice thickness and CNR. Realizes equivalent image quality in quantitative and qualitative phantom-based measurements with respect to metal objects. Successfully applied to phantom data from a motion phantom. Independent from physical detector width.
myNeedle Guide (with myNeedle Detection)Clinical usability of the needle detection algorithm, accuracy of automatic needle detection, reduction of necessary user interactions for navigating to a needle-oriented view.Algorithm consistently detected needle-tips in 90.76% of cases over a wide variety of scans. Auto needle detection functionality reduces the number of interaction steps needed to generate a needle-aligned view. With successful AI-based needle tip detection, no user interaction is needed to achieve needle-aligned view during needle progression (manual adjustment always possible).

2. Sample Size for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated for any of the individual feature tests. The tests refer to "phantom tests" and "analysis of phantom images". For "myNeedle Guide," it mentions "a wide variety of scans," but no specific number.
  • Data Provenance: The document does not specify the country of origin for the test data (phantoms) or if any retrospective/prospective human data was used. Given the nature of these tests (bench testing on phantoms), human patient data is generally not the primary focus for these types of technical evaluations.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • This information is not provided as the testing primarily involves technical and phantom-based evaluations, not clinical reader studies requiring expert ground truth.

4. Adjudication Method for the Test Set

  • This information is not applicable/provided as detailed clinical studies with reader adjudication are not described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

  • No, an MRMC comparative effectiveness study is not mentioned in this 510(k) summary. The submission focuses on demonstrating substantial equivalence through technical testing and feature comparison, not on quantifying improvement in human reader performance with or without AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • For the "myNeedle Guide" feature, the 90.76% detection rate of needle tips might be considered a form of standalone performance for that specific algorithmic component, though it's still being evaluated in the context of aiding a human user. For other features, the tests are primarily system-level or component-level functional checks.

7. The Type of Ground Truth Used

  • For the non-clinical tests described, the "ground truth" would be established through phantom specifications and controlled experimental setups with known parameters (e.g., precise needle location in a phantom, known artifact presence/absence in a reconstructed image). This is typical for engineering verification and validation testing for CT systems.
  • For the "myNeedle Guide," the "ground truth" for needle tip detection would likely be based on the known, true location of the needle tip within the phantom or experimental setup.

8. The Sample Size for the Training Set

  • This information is not provided. The document describes software updates and system features, not the development of a new AI model from a training set. If the "myNeedle Guide" used machine learning, its training set details are not described here.

9. How the Ground Truth for the Training Set was Established

  • This information is not provided as no training sets are explicitly described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

March 26, 2024

Siemens Medical Solutions USA, Inc. % Clayton Ginn Regulatory Affairs Professional 810 Innovation Drive KNOXVILLE, TN 37932

Re: K233650

Trade/Device Name: SOMATOM go.Up: SOMATOM go.Now: SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM X.cite; SOMATOM X.ceed Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: February 23, 2024 Received: February 23, 2024

Dear Clayton Ginn:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K233650
Device Name
SOMATOM go.Up;
SOMATOM go.Now;
SOMATOM go.All;
SOMATOM go.Top;
SOMATOM go.Sim;
SOMATOM go.Open Pro;
SOMATOM X.cite;
SOMATOM X.ceed
Indications for Use (Describe)

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.

This CT system can be used for low dose lung cancer screening in high risk populations*. *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots in a pattern.

510(k) Summary for K233650

SOMATOM go. Platform and SOMATOM X. Platform CT Scanner Systems

with software version SOMARIS/10 syngo CT VB10

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

I. Contact Details

Submitter Siemens Medical Solutions USA, Inc. 810 Innovation Drive Knoxville, TN 37932 Establishment Registration Number: 1034973

Importer/Distributor Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number: 2240869

Location of Manufacturing Site (1) Siemens Healthcare GmbH Siemensstr. 1 -OR- Rittigfeld 1 D-91301 Forchheim, Germany Establishment Registration Number: 3004977335

Location of Manufacturing Site (2) SIEMENS SHANGHAI, MEDICAL EQUIPMENT LTD 278 Zhou Zhu Rd Shanghai, CHINA, 201318 Establishment Registration Number: 3003202425

Note: Descriptions in this submission use the short company name Siemens. It covers both manufacturing locations and names as listed above. Brand name on all products is Siemens Healthineers.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.

Submitter Contact Person:

Clayton Ginn Regulatory Affairs Professional Siemens Medical Solutions USA, Inc. Molecular Imaging 810 Innovation Drive Knoxville, TN 37932 Phone: (865) 898-2692 clayton.ginn@siemens-healthineers.com Backup Contact: Kenny Bello Regulatory Affairs Professional Siemens Medical Solutions USA, Inc. Molecular Imaging 810 Innovation Drive Knoxville, TN 37932 Phone: (205) 856-6099 monsuru.bello@siemens-healthineers.com

II. Device Name and Classification

Table 1: Subject Device Names

Product NameTrade Name
SOMATOM go.UpSOMATOM go.Up
SOMATOM go.NowSOMATOM go.Now
SOMATOM go.AllSOMATOM go.All
SOMATOM go.TopSOMATOM go.Top
SOMATOM go.SimSOMATOM go.Sim
SOMATOM go.Open ProSOMATOM go.Open Pro
SOMATOM X.citeSOMATOM X.cite
SOMATOM X.ceedSOMATOM X.ceed
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
Regulation Number:21 CFR §892.1750
Device Class:Class II
Product Code:JAK

III. Predicate Device

Primary predicate device:
Trade Name:SOMATOM CT Scanner Systems with SOMARIS/10 syngo CT VA40software
510(k) Number:K211373
Clearance Date:August 27, 2021
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
Regulation Number:21 CFR §892.1750
Device Class:Class II
Product Code:JAK
Recall Information:All predicate device recalls have been considered in the subject devicedesign.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots arranged in a pattern.

Reference device:
Trade Name:SOMATOM CT Scanner Systems with SOMARIS/7 syngo CT VB30software
510(k) Number:K230421
Clearance Date:June 16, 2023
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
Regulation Number:21 CFR §892.1750
Device Class:Class II
Product Code:JAK
Recall Information:All predicate device recalls have been considered in the subject devicedesign.

Note:

K211373 was a bundle submission with various Siemens CT Scanner Systems, including the CT scanner systems SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM go.Open Pro, SOMATOM X.cite, SOMATOM X.ceed.

K230421 was a bundle submission with various Siemens CT Scanner Systems, including the dual source CT scanner systems SOMATOM Drive, SOMATOM Force, SOMATOM Definition Flash and the single source CT scanner CT systems SOMATOM Edge Plus, SOMATOM Confidence, SOMATOM Definition Edge, SOMATOM Definition AS/AS+, SOMATOM Definition AS Open.

IV. Device Description

Siemens intends to market a new software version, SOMARIS/10 syngo CT VB10 for the following SOMATOM Computed Tomography (CT) Scanner Systems:

SOMATOM go. Platform CT scanner systems:

  • . SOMATOM go.Up
  • . SOMATOM go.Now
  • SOMATOM go.All
  • SOMATOM go.Top
  • . SOMATOM go.Sim
  • . SOMATOM go.Open Pro

SOMATOM X. Platform CT scanner systems:

  • . SOMATOM X.cite
  • . SOMATOM X.ceed

In this submission, the above listed CT scanner systems are jointly referred to as subject devices by "SOMATOM go. Platform" and "SOMATOM X. Platform" CT scanner systems.

The subject devices SOMATOM go. Platform and SOMATOM X. Platform with SOMARIS/10 syngo CT VB10 are Computed Tomography X-ray Systems which feature one continuously rotating tubedetector system and function according to the fan beam principle (single source). The SOMATOM go. Platform and SOMATOM X. Platform with software SOMARIS/10 syngo CT VB10 produces CT

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a cluster of orange dots of varying sizes.

images in DICOM format, which can be used by trained staff for post-processing applications commercially distributed by Siemens Healthcare and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.

Only trained and qualified users, certified in accordance with country-specific regulations, are authorized to operate the system. For example, physicians, radiologists, or technologists. The user must have the necessary U.S. qualifications in order to diagnose or treat the patient with the use of the images delivered by the system.

The platform software for the SOMATOM go. Platform and SOMATOM X. Platform, SOMARIS/10 synqo CT VB10, is a command-based program used for patient management, data management, Xray scan control, image reconstruction, and image archive/evaluation.

The software platform provides plugin software interfaces that allow for the use of specific commercially available post processing software algorithms in an unmodified form from the cleared stand-alone post processing version.

New software version syngo CT VB10 (SOMARIS/10 syngo CT VB10) is a modified software version based on syngo CT VA40 (SOMARIS/10 syngo CT VA40) which was cleared for the predicate device in K211373.

Software version SOMARIS/10 syngo CT VB10 will be offered ex-factory and as an optional upgrade for the applicable existing SOMATOM go. Platform and SOMATOM X. Platform CT Scanner Systems.

The bundle approach is feasible for this submission since the subject devices have similar technological characteristics, software operating platform, and supported software characteristics. The supporting data are similar, primarily one review division/group will be involved, and the indications for use is the same between the devices. All subject devices will support previously cleared software and hardware features in addition to the applicable modifications as described within this submission. The intended use remains unchanged compared to the predicate devices.

V. Indications for Use

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.

This CT system can be used for low dose lung cancer screening in high risk populations*.

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are several orange dots arranged in a circular pattern.

VI. Indications for Use Comparison

Subject Device Indications for Use:

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.

This CT system can be used for low dose lung cancer screening in high risk populations*.

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Primary Predicate Device Indications for Use:

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained physician as an aid in diagnosis.

The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning.

This CT system can be used for low dose lung cancer screening in high risk populations.

High risk populations are as defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Comparison:

The subject device Indications for Use is the exact same as the reference predicate's Indications for Use (K230421), however it is slightly different than the primary predicate's Indications for Use.

    1. The sentence "The images delivered by the system can be used by a trained physician as an aid in diagnosis" was removed, since the subsequent sentence restates the same, replacing 'trained physician' with 'trained staff'. It can be assumed a trained physician is part of the trained staff.
      To consider that in the US, the standard of care is for physicians to perform diagnosis/treatment, Siemens added the following statement in the Device Description above and an equivalent warning exists in the labeling:

"Only trained and qualified users, certified in accordance with country-specific regulations, are authorized to operate the system. For example, physicians, radiologists. The user must have the necessary U.S. qualifications in order to diagnose or treat the patient with the use of the images delivered by the system."

    1. The word 'preparation' has been removed from the phrase "treatment preparation and radiation therapy planning" since treatment planning better describes how the device is used.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.

    1. The phrase "as well as for diagnostic and therapeutic interventions" was added to the second sentence. This was done to express use of the device in standard practice. This device can and is commonly used as an aid in invasive procedures.
      None of the intended use includes computed tomography as the principal means of guidance in invasive procedures (involving the introduction of a device, such as a needle or a catheter into the body of the patient).

The subject devices SOMATOM go. Platform and SOMATOM X. Platform are not the principal means of guidance, because the CT Systems do not guide the invasive procedures, the needle orientation and the needle advance and handling is always done under the physicians control.

VII. Comparison of Technological Characteristics with the Predicate Device

Supported by the subject device, SOMARIS/10 syngo CT VB10 software version is a further development of the SOMARIS/10 syngo CT VA40 software version which is cleared in K211373.

The SOMATOM go. Platform and the SOMATOM X. Platform with SOMARIS/10 syngo CT VB10 software version provides the same technological characteristics in terms of materials, energy source, and control mechanisms when compared to the predicate devices. The software features of SOMATOM go. Platform and SOMATOM X. Platform have been modified or improved in comparison to the predicate devices to support enhanced device functionality compared to the predicate devices.

The new syngo CT VB10 software reuses all unmodified software features of the legacy software syngo CT VA40 cleared in K211373. Additionally, no features present in the predicate device are descoped.

Software version SOMARIS/10 syngo CT VB10 is designed to reuse hardware independent extended functionalities and GO technologies provided by Siemens cleared software applications.

The intended use and fundamental scientific technology for the SOMATOM go. Platform and SOMATOM X. Platform remain unchanged from the predicate devices.

At a high level, the subject and predicate devices are based on the same subset of technological elements:

  • Scanner Principle – Whole body X-Ray Computed Tomography Scanner
  • System Acquisition – Continuously rotating tube detector system
  • Iterative Reconstruction – Support of various iterative reconstruction principles
  • Workplaces Support of workplaces that include reconstruction and image evaluation ● software
  • Patient table
  • Patient table foot switch for movement
  • Stellar Detector technology
  • . Tin filtration technology
  • . Vectron/Athlon/Chronon X-ray Tube depending on the CT system
  • Power Generator
  • Scan&GO
  • Mobile workflow (Tablet)
  • Optional injector arm

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a cluster of orange dots that are arranged in a circular pattern.

  • Optional support of CT guided intervention workflow (myNeedle Guide) ●
  • . Optional support of FAST 3D Camera operation for fast patient positioning workflow
  • . Scanner display and control functionality
  • Remote Scan Control
  • . Respiratory Gating Devices
  • . Standard CARE technologies
  • Standard FAST technologies
  • Standard GO technologies
  • . Precision matrix (large image matrices)
  • HD FoV
  • DirectDensity™ ●

Compared to the primary predicate device referenced in this submission, the subject devices SOMATOM go. Platform and SOMATOM X. Platform CT Scanner Systems with SOMARIS/10 syngo CT VB10 support the modifications provided in the Table 3 under subsection 1) Modified Hardware and Table 4 under subsection 2) Modified Software.

The configuration table and comparison table use the following terms in Table 2 to describe various technological characteristics in comparison to the primary and secondary predicate devices information:

TermDefinition
N/AThe feature is not supported for the subject device
NewThe feature is newly supported for Siemens CT Scanners and newly introducedwithin the subject device submission
UnmodifiedThe feature remains unchanged from that within the predicate device. Therelevant modification was already present within the predicate device.
ModifiedThis feature is modified from the predicate / reference devices
EnabledThis feature is currently supported by other cleared Siemens CT systems orcleared Siemens stand-alone software applications. This feature will besupported for the subject device with software version SOMARIS/10 syngo CTVB10 and is substantially equivalent compared to the cleared version of theprimary/reference predicate device.

Table 2: Overview of term definition.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots.

1) Modified Hardware

Table 3: Overview of hardware modifications of SOMATOM go. Platform and SOMATOM X. Platform CT scanner systems with SOMARIS/10 syngo CT VB10.

#HardwarepropertySubject deviceswith SOMARIS/10 syngo CT VB10
SOMATOM go. PlatformSOMATOM X. Platform
SOMATOM go.NowSOMATOM go.UpSOMATOM go.AllSOMATOM go.TopSOMATOM go.SimSOMATOM go.Open ProSOMATOM X.citeSOMATOM X.ceed
1.FAST 3D Cameraenabledenabledenabled
2.CARE BreatheenabledN/Aunmodified
3.CARE 2D Cameraenabledunmodifiedunmodified
4.Moodlightenabledunmodifiedunmodified
5.CARE MoodlightenabledN/AN/A
6.Multi-PurposeTable (Vitus)N/AN/Amodified

2) Modified Software (syngo CT VB10)

Table 4: Overview of software modifications of SOMATOM go. Platform and SOMATOM X. Platform CT Scanner Systems with SOMARIS/10 syngo CT VB10.

#Software propertySubject deviceswith SOMARIS/10 syngo CT VB10
SOMATOM go. PlatformSOMATOMX. Platform
SOMATOMgo.NowSOMATOMgo.UpSOMATOMgo.AllSOMATOMgo.TopSOMATOMgo.SimSOMATOMgo.Open ProSOMATOMX.citeSOMATOMX.ceed
1.DirectBreathholdN/Aenabledenabledenabledenabledenabled
2.ZeeFreeN/AN/Aenabledenabledenabledenabled
3.myNeedle Guideenabledenabledenabledenabledenabledenabled
4.FAST IntegratedWorkflowenabledenabledenabledenabledenabledenabled
5.Direct i4DN/AN/AN/AN/Amodifiedmodified
6.myExamCompanionenabledmodifiedmodifiedmodifiedmodifiedmodified
7.myExam Satelite(Remote Recon)enabledenabledenabledenabledenabledenabled
8.Recon&GOmodifiedmodifiedmodifiedmodifiedmodifiedmodified

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots that are arranged in a circular pattern.

#Software propertySubject deviceswith SOMARIS/10 syngo CT VB10
SOMATOM go. PlatformSOMATOMX. Platform
SOMATOMgo.NowSOMATOMgo.UpSOMATOMgo.AllSOMATOMgo.TopSOMATOMgo.SimSOMATOMgo.Open ProSOMATOMX.citeSOMATOMX.ceed
9.CT View&GOenabledenabledenabledenabledenabledenabled
10.Scan&GOenabledenabledenabledenabledenabledenabled
11.Automatic Patientinstructionenabledenabledenabledenabledenabledenabled
12.DirectSetup NotesN/AN/AN/AnewnewN/A
13.ADMIREN/AN/Aunmodifiedenabledenabledunmodified

A tabular summary of the comparable hardware and software properties between the subject devices SOMATOM go. Platform and SOMATOM X. Platform with software version syngo CT VB10 and primary/reference predicate devices are listed in Table 9 below (modifications are in gray shaded sections).

Table 5: Comparison of technical hardware characteristics between subject devices SOMATOM go.Now and SOMATOM go.Up with software version syngo CT VB10 and the primary predicate devices with software version syngo CT VA40.

Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.NowSOMATOMgo.UpSOMATOMgo.NowSOMATOMgo.Up
scannerWhole body Computed Tomography(CT) Scanner SystemWhole body Computed Tomography(CT) Scanner SystemWhole body Computed Tomography(CT) Scanner SystemWhole body Computed Tomography(CT) Scanner System
SystemConfigurationSingle SourceSingle Source
Environment of UseProfessional Healthcare FacilityProfessional Healthcare Facility
GeneratorMax. power (kW)3232
DetectortechnologyStellarbased on UFC (Ultra Fast Ceramic)Stellarbased on UFC (Ultra Fast Ceramic)
Detectorvolume coverage(mm)11.222.411.222.4
Detector16321632
Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.NowSOMATOMgo.UpSOMATOMgo.NowSOMATOMgo.Up
physical rows
DetectorSlice width (mm)0.70.7
DetectorDAS channel no.768768
DetectorImage slices32643264
TubetechnologyChrononChronon
TubekV steps80, 110, 13080, 110, 130
TubeMax. current (mA)400400
Tubefocus• 0.8 x 0.4 / 8°• 0.8 x 0.7 / 8°• 0.8 x 0.4 / 8°• 0.8 x 0.7 / 8°
TubeHeat storagecapacity (MHU)3.53.5
TubeCooling rate (kHU/min)915915
GantryBore size7070
GantryScan FoV (cm)5050
GantryExtended FoV (cm)7070
GantryRot. time (sec)0.5, 0.8, 1.0, 1.50.8, 1.0, 1.5
GantryTilt (degree)N/A+/- 30°-30° to +28° withoptional 307 kgN/A+/- 30°-30° to +28° withoptional 307 kg
Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.NowSOMATOMgo.UpSOMATOMgo.NowSOMATOMgo.Up
patient tablepatient table
Patient tableTypeVector (1250mm),Vario 1 (1600mm),Vario RT (1600mm)with tableextensionVario 1 (1600mm),Vario 2 (2000mm),Vario RT (1600mm)with tableextensionVector (1250mm),Vario 1 (1600mm),Vario RT (1600mm)with tableextensionVario 1 (1600mm),Vario 2 (2000mm),Vario RT (1600mm)with tableextension
Patient tableMax. weightcapacity (kg)Vector: 160Vario 1: 227Vario RT: 227Vario 1: 227Vario 2: 307Vario RT: 227Vector: 160Vario 1: 227Vario RT: 227Vario 1: 227Vario 2: 307Vario RT: 227
Max. scan lengthtopogram1680 mm withtable extension1680 mm withtable extension2080 mm withtable extension1680 mm withtable extension1680 mm withtable extension2080 mm withtable extension
Max. scan lengthImage acquisitionMax. 1600 mmwith patienttable extensionMax. 1600 mmwith patienttable extensionMax. 2000 mmwith patienttable extensionMax. 1600 mmwith patienttable extensionMax. 1600 mmwith patienttable extensionMax. 2000 mmwith patienttable extension
Spectral filtrationTin FilterTin Filter
High Power 80yesyesyesyes
FAST 3D Cameraoption forpatientpositioning with3D Camera:• gantrymounted,modifieddesignoption forpatientpositioning with3D Camera:• ceilingmounted,modifieddesign• gantrymounted,modifieddesignn.a.option forpatientpositioning with3D Camera:• ceilingmounted
Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.NowSOMATOMgo.UpSOMATOMgo.NowSOMATOMgo.Up
x-ray foot switchOption to trigger hands-freescanningOption to trigger hands-freescanningOption to trigger hands-freescanningOption to trigger hands-freescanning
Table foot switchOption for table patient movementOption for table patient movementOption for table patient movementOption for table patient movement
i-joystickOption for patient tablemovements, including an electricalconnection for the tablet dockwhich allows charging the tabletwhen mounted.Option for patient tablemovements, including an electricalconnection for the tablet dockwhich allows charging the tabletwhen mounted.Option for patient tablemovements, including an electricalconnection for the tablet dockwhich allows charging the tabletwhen mounted.Option for patient tablemovements, including an electricalconnection for the tablet dockwhich allows charging the tabletwhen mounted.
Tablet dockOption for mounting of the tableton the patient table.Option for mounting of the tableton the patient table.Option for mounting of the tableton the patient table.Option for mounting of the tableton the patient table.
CARE Breathevisualization of breathinginstructions to support threemessages in respiratory motionmanagement:• free breathing• preparation for breath hold• breath hold itself incombination with thevisualization of a countdownvisual patient instruction unit isplaced on the gantry front andgantry back in the gantry funnelvisual countdown indicator toprovide visual instruction inrespiratory motion management:• free breathing• preparation for hold breath• breath holdvisual countdown indicator is placedon the halo
CARE 2D cameraone integrated into the gantry frontand one mounted on the gantryback to monitor the patient duringexamination2D camera* integrated on the haloplate*in syngo CT VA40 known as patientobservation camera
MoodlightIllumination Moodlight to enhancepatient comfort supportedintegrated as a ring with LED lightinto the gantry frontIllumination Moodlight to enhancepatient comfort supportedIntegrated on the Halo
CARE MoodlightLED lights integrated on the gantryfront cover indicating:• System startupn.a.
Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.NowSOMATOMgo.UpSOMATOMgo.NowSOMATOMgo.Up
• Scan countdown• Radiation on

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a series of orange dots that are arranged in a curved line.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots that are arranged in a pattern.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a series of orange dots arranged in a circular pattern.

Table 6: Comparison of technical hardware characteristics between subject devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB10 and the primary predicate devices with software version syngo CT VA40.

Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10SOMATOMgo.TopPrimary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)SOMATOMgo. Top
SOMATOMgo.AllSOMATOMgo.TopSOMATOMgo.AllSOMATOMgo.Top
scannerWhole body Computed Tomography(CT) Scanner SystemWhole body Computed Tomography(CT) Scanner SystemWhole body Computed Tomography(CT) Scanner SystemWhole body Computed Tomography(CT) Scanner System
SystemConfigurationSingle SourceSingle SourceSingle SourceSingle Source
Environment of UseProfessional Healthcare FacilityProfessional Healthcare FacilityProfessional Healthcare FacilityProfessional Healthcare Facility
GeneratorMax. power (kW)75757575
DetectortechnologyStellarbased on UFC (Ultra Fast Ceramic)Stellarbased on UFC (Ultra Fast Ceramic)Stellarbased on UFC (Ultra Fast Ceramic)Stellarbased on UFC (Ultra Fast Ceramic)
Detectorvolume coverage(mm)22.438.422.438.4
Detectorphysical rows32643264
DetectorSlice width (mm)0.70.60.70.6
DetectorDAS channel no.768840768840
DetectorImage slices6412864128
TubeAthlonAthlonAthlonAthlon
Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.AllSOMATOMgo.TopSOMATOMgo.AllSOMATOMgo.Top
technology
TubekV steps70 to 140 (in 10 kV steps)70 to 140 (in 10 kV steps)
TubeMax. current (mA)825825
Tubefocus• 1.1 x 1.2 / 8°• 0.8 x 0.8 / 8°• 1.1 x 1.2 / 8°• 0.8 x 0.8 / 8°
TubeHeat storagecapacity (MHU)77
TubeCooling rate (kHU/min)Up to 1700Up to 1700
GantryBore size7070
GantryScan FoV (cm)5050
GantryExtended FoV (cm)7070
GantryRot. time (sec)0.33, 0.5, 1.00.33, 0.5, 1.0
GantryTilt (degree)+/- 30°-30° to +28° with optional 307 kgpatient table+/- 30°-30° to +28° with optional 307 kgpatient table
Patient tableTypeVario 1 (1600 mm)Vario 2 (2000 mm)Vario RT (1600 mm)with table extensionVario 1 (1600 mm)Vario 2 (2000 mm)Vario RT (1600 mm)with table extension
Patient tableMax. weightcapacity (kg)Vario 1: 227Vario 2: 307Vario RT: 227Vario 1: 227Vario 2: 307Vario RT: 227
Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.AllSOMATOMgo.TopSOMATOMgo.AllSOMATOMgo.Top
Max. scan lengthtopogram1680 mm with table extension2080 mm with table extension1680 mm with table extension2080 mm with table extension
Max. scan lengthImage acquisitionMax. 1600 mm with patient tableextensionMax. 2000 mm with patient tableextensionMax. 1600 mm with patient tableextensionMax. 2000 mm with patient tableextension
Spectral filtrationTin FilterCombined TinFilter/Split filterTin FilterCombined TinFilter/Split filter
High Power 70/High Power 80825 (@70 kV)825 (@ 80 kV)825 (@70 kV)825 (@ 80 kV)
FAST 3D Cameraoption for patient positioning withFAST 3D Camera:• ceiling mounted, modified design• gantry mounted, modified designoption for patient positioning withFAST 3D Camera:• ceiling mounted
x-ray foot switchOption to trigger hands-freescanningOption to trigger hands-freescanning
Table foot switchOption for table patient movementOption for table patient movement
i-joystickOption for patient tablemovements, including an electricalconnection for the tablet dockwhich allows charging the tabletwhen mounted.Option for patient tablemovements, including an electricalconnection for the tablet dockwhich allows charging the tabletwhen mounted.
Tablet dockOption for mounting of the tableton the patient table.Option for mounting of the tableton the patient table.
CARE Breathevisualization of breathinginstructions to support threemessages in respiratory motionmanagement:• free breathing• preparation for breath hold• breath hold itself incombination with thevisualization of a countdownvisual countdown indicator toprovide visual instruction inrespiratory motion management:• free breathing• preparation for hold breath• breath holdvisual countdown indicator is placedon the halo
Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.AllSOMATOMgo.TopSOMATOMgo.AllSOMATOMgo.Top
visual patient instruction unit isplaced on the gantry front andgantry back in the gantry funnel
CARE 2D cameraone integrated into the gantry frontand one mounted on the gantryback to monitor the patient duringexamination2D camera* integrated on the haloplate*in syngo CT VA40 known as Patientobservation camera
MoodlightIllumination Moodlight to enhancepatient comfort supportedintegrated as a ring with LED lightinto the gantry frontIllumination Moodlight to enhancepatient comfort supportedIntegrated on the Halo
CARE MoodlightLED lights integrated on the gantryfront cover indicating:• System startup• Scan countdown• Radiation onn.a.
Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.SimSOMATOMgo.Open ProSOMATOMgo.SimSOMATOMgo.Open Pro
scannerWhole body Computed Tomography(CT) Scanner SystemWhole body Computed Tomography(CT) Scanner System
SystemConfigurationSingle SourceSingle Source
Environment of UseProfessional Healthcare FacilityProfessional Healthcare Facility
GeneratorMax. power (kW)7575
DetectorStellarStellar
technologybased on UFC (Ultra Fast Ceramic)based on UFC (Ultra Fast Ceramic)
Detectorvolume coverage(mm)19.238.419.238.4
Detectorphysical rows32643264
DetectorSlice width (mm)0.60.60.60.6
DetectorDAS channel no.920920920920
DetectorImage slices6412864128
TubetechnologyAthlonAthlon
TubekV steps70 to 140 (in 10 kV steps)70 to 140 (in 10 kV steps)
TubeMax. current (mA)825825
Tube• 1.1 x 1.2 / 7°• 0.8 x 0.8 / 7°• 1.1 x 1.2 / 7°• 0.8 x 0.8 / 7°
focus• 0.8 x 0.8 / 7°• 0.8 x 0.8 / 7°
Tube77
Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.SimSOMATOMgo.Open ProSOMATOMgo.SimSOMATOMgo.Open Pro
Heat storagecapacity (MHU)
TubeCooling rate (kHU/min)Up to 1700Up to 1700
GantryBore size8585
GantryScan FoV (cm)6060
GantryExtended FoV (cm)8585
GantryRot. time (sec)0.35, 0.5, 1.00.35, 0.5, 1.0
GantryTilt (degree)+/- 30-30° to +25° with optional 307 kgpatient table+/- 30-30° to +25° with optional 307 kgpatient table
Patient tableTypeVario 2 (2000 mm)Vario RT (1600 mm)with table extensionVario 2 (2000 mm)Vario RT (1600 mm)with table extension
Patient tableMax. weightcapacity (kg)Vario 2: 307Vario RT: 227Vario 2: 307Vario RT: 227
Max. scan lengthtopogram1680 mm with table extension2080 mm with table extension1680 mm with table extension2080 mm with table extension
Max. scan lengthImage acquisitionMax. 1600 mm with patient tableextensionMax. 2000 mm with patient tableextensionMax. 1600 mm with patient tableextensionMax. 2000 mm with patient tableextension
Spectral filtrationTin FilterCombined TinFilter/Split filterTin FilterCombined TinFilter/Split filter
High Power 70/High Power 80825 (@70 kV)825 (@ 80 kV)Combined TinFilter/Split filter825 (@70 kV)825 (@ 80 kV)Combined TinFilter/Split filter
Hardware propertySubject deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM go. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMgo.SimSOMATOMgo.Open ProSOMATOMgo.SimSOMATOMgo.Open Pro
FAST 3D Cameraoption for patient positioning withFAST 3D Camera:• ceiling mounted, modified designoption for patient positioning withFAST 3D Camera:• ceiling mounted
x-ray foot switchOption to trigger hands-freescanningOption to trigger hands-freescanning
Table foot switchOption for table patient movementOption for table patient movement
i-joystickOption for patient tablemovements, including an electricalconnection for the tablet dockwhich allows charging the tabletwhen mounted.Option for patient tablemovements, including an electricalconnection for the tablet dockwhich allows charging the tabletwhen mounted.
Tablet dockOption for mounting of the tableton the patient table.Option for mounting of the tableton the patient table.
CARE Breathen.a.n.a.
CARE 2D cameraPatient observation camera*integrated on gantry front*in syngo CT VB10 the patientobservation camera is renamed asCARE 2D cameraPatient observation cameraintegrated on gantry front

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern.

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" are several orange dots of varying sizes.

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots.

Table 7: Comparison of technical hardware characteristics between subject devices SOMATOM go.Sim and SOMATOM go.Open Pro with software version syngo CT VB10 and the primary predicate devices with software version syngo CT VA40.

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a series of orange dots arranged in a circular pattern.

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern.

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.

Table 8: Comparison of technical hardware characteristics between subject devices SOMATOM X.ceed with software version syngo CT VB10 and the primary predicate devices with software version syngo CT VA40 (K211373).

HardwarepropertySubject deviceSOMATOM X. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM X. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMX.citeSOMATOMX.ceed
scannerWhole body Computed Tomography(CT) Scanner SystemWhole body Computed Tomography(CT) Scanner System
SystemConfigurationSingle SourceSingle Source
Environment ofUseProfessional Healthcare FacilityProfessional Healthcare Facility
GeneratorMax. power (kW)90 kW or 105 kW105 kW or 120kW90 kW or 105 kW105 kW or 120kW
DetectortechnologyStellarInfinityStellarInfinity
Detectorvolume coverage38.4 mm38.4 mm
Detectorphysical rows64rows64
DetectorSlice width0.6 mm0.6 mm
DetectorDAS channel no.840920840920
DetectorImage slices128128
TubetechnologyVectronVectron
TubekV steps70–140 kV in 10 kV steps70–140 kV in 10 kV steps
TubeMax. current1100 mA (for 90kW)1200 mA (for 105kW)1200 mA (for 90kW)1300 mA (for 105kW)1100 mA (for 90kW)1200 mA (for 105kW)1200 mA (for 90kW)1300 mA (for 105kW)
Tubefocus• 0.6 x 0.7 / 8°• 0.8 x 1.1 / 8°• 0.4 x 0.5 / 8°• 0.6 x 0.7 / 8°• 0.6 x 0.7 / 8°• 0.8 x 1.1 / 8°• 0.4 x 0.5 / 8°• 0.6 x 0.7 / 8°
HardwarepropertySubject deviceSOMATOM X. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM X. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMX.citeSOMATOMX.ceedSOMATOMX.citeSOMATOMX.ceed
• 0.8 x 1.1 / 8°• 0.8 x 1.1 / 8°
TubeHeat storagecapacityhigher than 30 MHUhigher than 30 MHU
TubeCooling rate(MHU/ min)2.7 MHU / min2.7 MHU / min2.7 MHU / min2.7 MHU / min
GantryBore size8282
GantryScan FoV (cm)5050
GantryExtended FoV(cm)81.581.5
GantryRot. time (sec)0.3, 0.5, 1.00,25, 0.3, 0.5, 1.00.3, 0.5, 1.00,25, 0.3, 0.5, 1.0
GantryTilt (degree)+/- 25+/-30 (only with Vitus)+/- 25+/-30 (only with Vitus)
Patient tableTypeVario RT (1600mm)Vitus (2000 mm)Vario 2 (2000mm)Vario RT (1600mm)Vitus (2000 mm)Vario 2.D (2000mm)Vario RT (1600mm)Vitus (2000 mm)Vario 2 (2000mm)Vario RT (1600mm)Vitus (2000 mm)Vario 2.D (2000mm)
Patient tableInstallationoptionRegularinstallation(Vario RT, Vario 2and Vitus):474 mmInstallationoption withextendeddistance (Vitus):674 mm (474 mm)Regularinstallation(Vario RT, Vario 2and Vitus):474 mmInstallationoption withextendeddistance (Vitus):674 mm (474 mm)Regularinstallation(Vario RT, Vario 2and Vitus):474 mmRegularinstallation(Vario RT, Vario 2and Vitus):474 mm
HardwarepropertySubject deviceSOMATOM X. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM X. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMX.citeSOMATOMX.ceedSOMATOMX.citeSOMATOMX.ceed
Patient tableMax. weightcapacity (kg)227 (Vario RT)340 (Vitus)307 (Vario 2)227 (Vario RT)340 (Vitus)307 (Vario 2.D)227 (Vario RT)307 (Vitus)307 (Vario 2)227 (Vario RT)307 (Vitus)307 (Vario 2.D)
Max. scan lengthtopogram1680 mm (VarioRT)2080 mm withtable extension(Vario 2)2080 mm (Vitus)1680 mm (VarioRT)2080 mm withtable extension(Vario 2.D)2080 mm (Vitus)1680 mm (VarioRT)2080 mm withtable extension(Vario 2)2080 mm (Vitus)1680 mm (VarioRT)2080 mm withtable extension(Vario 2.D)2080 mm (Vitus)
Max. scan lengthImage acquisition1600 mm (VarioRT)2000 mm (Vitus)2000 mm (Vario2)with tableextension1600 mm (VarioRT)2000 mm (Vitus)2000 mm (Vario2.D)with tableextension1600 mm (VarioRT)2000 mm (Vitus)2000 mm (Vario2)with tableextension1600 mm (VarioRT)2000 mm (Vitus)2000 mm (Vario2.D)with tableextension
Spectral filtrationCombined TinFilter / SplitFilter, and extraTin FilterCombined TinFilter / SplitFilter, and extraTin FilterCombined TinFilter / SplitFilter, and extraTin FilterCombined TinFilter / SplitFilter, and extraTin Filter
FAST 3D Cameraoption for patient positioning withFAST 3D Camera:ceiling mounted, new designoption for patient positioning withFAST 3D Camera:ceiling mounted
High Power 70/High Power 80Generator 90kW:up to 1100 mA(@70 kV)up to 1100 mA(@80 kV)Generator 105kW:up to 1200 mA(@70 kV)Generator 105kW:up to 1200 mA(@70 kV)up to 1200 mA(@80 kV)Generator 120kW:up to 1300 mA(@70 kV)Generator 90kW:up to 1100 mA(@70 kV)up to 1100 mA(@80 kV)Generator 105kW:up to 1200 mA(@70 kV)Generator 105kW:up to 1200 mA(@70 kV)up to 1200 mA(@80 kV)Generator 120kW:up to 1300 mA(@70 kV)
HardwarepropertySubject deviceSOMATOM X. Platformwith SOMARIS/10 syngo CT VB10Primary predicate deviceSOMATOM X. Platformwith SOMARIS/10 syngo CT VA40(K211373)
SOMATOMX.citeSOMATOMX.ceedSOMATOMX.citeSOMATOMX.ceed
up to 1200 mA(@80 kV)up to 1300 mA(@80 kV)up to 1200 mA(@80 kV)up to 1300 mA(@80 kV)
x-ray foot switchOption to trigger hands-free scanningOption to trigger hands-free scanning
Table foot switchOption for table patient movementOption for table patient movement
i-joystickOption for patient table movements,including an electrical connection forthe tablet dock which allows chargingthe tablet when mounted.Option for patient table movements,including an electrical connection forthe tablet dock which allows chargingthe tablet when mounted.
Tablet dockOption for mounting of the tablet onthe patient table.Option for mounting of the tablet onthe patient table.
CARE 2D cameraPatient observation camera*integrated on gantry front*in syngo CT VB10 the patientobservation camera is renamed asCARE 2D cameraPatient observation cameraintegrated on gantry front

{23}------------------------------------------------

Image /page/23/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that are arranged in a circular pattern.

{24}------------------------------------------------

Image /page/24/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern.

{25}------------------------------------------------

Image /page/25/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern.

{26}------------------------------------------------

Image /page/26/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a graphic of orange dots arranged in a pattern.

Table 9: Comparison of software characteristics between the subject with software version syngo CT VB10 and predicate devices with software version syngo CT VA40 of the SOMATOM go. Platform.

Software propertySubject devicePrimary predicate device
SOMATOM go.NowSOMATOM go.Now
SOMATOM go.UpSOMATOM go.Up
SOMATOM go.AllSOMATOM go.All
SOMATOM go.TopSOMATOM go.Top
SOMATOM go.SimSOMATOM go.Sim
SOMATOM go.Open ProSOMATOM go.Open Pro
SOMATOM X.citeSOMATOM X.cite
SOMATOM X.ceedSOMATOM X.ceed
with SOMARIS/10 syngo CT VB10with SOMARIS/10 syngo CT VA40(K211373)
Operating SystemWindows basedSOMARIS/10 syngo CT VB10Note: the short version syngo CT VB10 is also used as labeling informationWindows basedSOMARIS/10 syngo CT VA40Note: the short version syngo CT VA40 is also used as labeling information
Workplacesyngo Acquisition Workplace (ICS) named as "myExam Console"Image Reconstruction System (IRS)2nd workplace option named as "myExam Satellite" with Remote Recon functionsyngo Acquisition Workplace (ICS) named as "myExam Console"Image Reconstruction System (IRS)2nd workplace option named as "myExam Satellite"
Standard systemsoftware• syngo Examination• syngo Viewing• syngo Filming• syngo Archiving & Network• syngo Examination• syngo Viewing• syngo Filming• syngo Archiving & Network
Detector firmwareStellar detector firmware supportedStellar detector firmware supported
TeamplaySupport of Teamplay protocolsSupport of Teamplay protocols
ProtocolsSupport of:• Protocols for Radiation Therapy Planning support patient marking• Protocols that allow scanning with support of an external respiratory gating system (ANZAI, Varian RGSC)Support of:• Protocols for Radiation Therapy Planning support patient marking• Protocols that allow scanning with support of an external respiratory gating system (ANZAI, Varian RGSC)
Software propertySubject devicePrimary predicate device
SOMATOM go.NowSOMATOM go.Now
SOMATOM go.UpSOMATOM go.Up
SOMATOM go.AllSOMATOM go.All
SOMATOM go.TopSOMATOM go.Top
SOMATOM go.SimSOMATOM go.Sim
SOMATOM go.Open ProSOMATOM go.Open Pro
SOMATOM X.citeSOMATOM X.cite
SOMATOM X.ceedSOMATOM X.ceed
with SOMARIS/10 syngo CT VB10with SOMARIS/10 syngo CT VA40(K211373)
Protocol supporting contrast bolus-triggered data acquisition Contrast media protocols (including coronary CTA) Pediatric Protocols Flex Dose Profile Turbo Flash Spiral Dual Energy acquisition (TwinBeam DE and TwinSpiral DE) Dynamic imaging (Flex 4D Spiral) Protocols supporting CT Intervention (scan modes: i-sequence, i-spiral, i-Fluoro) Protocols supporting Cardiac Scanning Protocols for DirectBreathholdProtocol supporting contrast bolus-triggered data acquisition Contrast media protocols (including coronary CTA) Pediatric Protocols Flex Dose Profile Turbo Flash Spiral Dual Energy acquisition (TwinBeam DE and TwinSpiral DE) Dynamic imaging (Flex 4D Spiral) Protocols supporting CT Intervention (scan modes: i-sequence, i-spiral, i-Fluoro) Protocols supporting Cardiac Scanning
Advanced ReconstructionRecon&GO:- Spectral Recon (Dual EnergyReconstruction from photon-counting data) / including VirtualUnenhanced, Monoenergetic plus- Inline Results DE SPP (SpectralPost-Processing with photon-counting image data)- Inline Anatomical ranges(Parallel/Radial) incl. VirtualUnenhanced, Monoenergetic plus- Inline Spine and Rib Ranges- Inline table and bone removalRecon&GO:- Spectral Recon (Dual EnergyReconstruction from photon-counting data) / including VirtualUnenhanced, Monoenergetic plus- Inline Results DE SPP (SpectralPost-Processing with photon-counting image data)
Image viewingCT View&GO offers:CT View&GO offers:
Software propertySubject devicePrimary predicate device
SOMATOM go.NowSOMATOM go.Now
SOMATOM go.UpSOMATOM go.Up
SOMATOM go.AllSOMATOM go.All
SOMATOM go.TopSOMATOM go.Top
SOMATOM go.SimSOMATOM go.Sim
SOMATOM go.Open ProSOMATOM go.Open Pro
SOMATOM X.citeSOMATOM X.cite
SOMATOM X.ceedSOMATOM X.ceed
with SOMARIS/10 syngo CT VB10with SOMARIS/10 syngo CT VA40(K211373)
- basic post-processing viewer (CTView&GO)- basic post-processing viewer (CTView&GO)
- 2D and 3D (MPR, VRT, MIP andminIP)- 2D and 3D (MPR, VRT, MIP andminIP)
- Evaluation tools, Filming, Printing- Evaluation tools, Filming, Printing
- Interactive Spectral Imaging (ISI)
- Basic visualization tools: EndoView
- Basic manipulation tools: DE ROI,ROI HU, Average
Post-Processinginterface• Recon&GO Inline ResultsSoftware interface to post-processing algorithms which areunmodified when loaded onto theCT scanners and 510(k) cleared asmedical devices in their own right.• Recon&GO Inline ResultsSoftware interface to post-processing algorithms which areunmodified when loaded onto theCT scanners and 510(k) cleared asmedical devices in their own right.
• software interfaces for post-processing functionalities toprovide advanced visualizationtools to prepare and processmedical images for diagnosticpurpose.• software interfaces for post-processing functionalities toprovide advanced visualizationtools to prepare and processmedical images for diagnosticpurpose.
Note: The clearance of standaloneAdvanced Visualization Applicationsoftware is mandatory precondition.Note: The clearance of standaloneAdvanced Visualization Applicationsoftware is mandatory precondition.
These advanced visualization toolsare designed to support thetechnician & physician in thequalitative and quantitativemeasurement & analysis of clinicaldata acquired and reconstructed byThese advanced visualization toolsare designed to support thetechnician & physician in thequalitative and quantitativemeasurement & analysis of clinicaldata acquired and reconstructed by
Software propertySubject devicePrimary predicate device
SOMATOM go.NowSOMATOM go.Now
SOMATOM go.UpSOMATOM go.Up
SOMATOM go.AllSOMATOM go.All
SOMATOM go.TopSOMATOM go.Top
SOMATOM go.SimSOMATOM go.Sim
SOMATOM go.Open ProSOMATOM go.Open Pro
SOMATOM X.citeSOMATOM X.cite
SOMATOM X.ceedSOMATOM X.ceed
with SOMARIS/10 syngo CT VB10with SOMARIS/10 syngo CT VA40(K211373)
Computed Tomography scanners.Additional information regardingthe points of interface and inputsfor this feature is provided inSection 16.Computed Tomography scanners.Additional information regardingthe points of interface and inputsfor this feature is provided inSection 16.
CybersecurityIT HardeningIT Hardening
HD FoVHD FoV 4.0HD FoV 4.0
Standardtechnologies• FAST technologies• CARE technologies• GO technologies• FAST technologies• CARE technologies• GO technologies
IterativeReconstructionMethodsADMIRE*SAFIREiMARADMIRE*SAFIREiMAR
*supported in SOMATOM go.All,SOMATOM go.Top, SOMATOMgo.Sim, SOMATOM go.Open Pro,SOMATOM X.cite and SOMATOMX.ceed*supported in SOMATOMgo.All,SOMATOM go.Top,SOMATOM X.cite and SOMATOMX.ceed
Matrix sizes256 x 256 pixels512 x 512 pixels768 x 768 pixels1024 x 1024 pixels (PrecisionMatrix)256 x 256 pixels512 x 512 pixels768 x 768 pixels1024 x 1024 pixels (PrecisionMatrix)
DirectDensity™DirectDensity™DirectDensity™
Software propertySubject devicePrimary predicate device
SOMATOM go.NowSOMATOM go.Now
SOMATOM go.UpSOMATOM go.Up
SOMATOM go.AllSOMATOM go.All
SOMATOM go.TopSOMATOM go.Top
SOMATOM go.SimSOMATOM go.Sim
SOMATOM go.Open ProSOMATOM go.Open Pro
SOMATOM X.citeSOMATOM X.cite
SOMATOM X.ceedSOMATOM X.ceed
with SOMARIS/10 syngo CT VB10with SOMARIS/10 syngo CT VA40(K211373)
(including relative electron densityand relative mass density)(including relative electron densityand relative mass density)
Breath-holdtechniqueRespiratory Motion Managementsupport breath hold triggered spiralscans with manual breath holdtriggered examinations.Respiratory Motion Managementsupport breath hold triggered spiralscans with manual breath holdtriggered examinations.
Direct Breathold: externallytriggered workflow to start a spiralscan by receiving a trigger signalfrom an external gating device.
*Note: Direct Breathold is notavailable for SOMATOM go.Now

{27}------------------------------------------------

Image /page/27/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a series of orange dots that are arranged in a curved line.

{28}------------------------------------------------

Image /page/28/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots that are arranged in a pattern.

{29}------------------------------------------------

Image /page/29/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots.

{30}------------------------------------------------

Image /page/30/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.

Any differences in technological characteristics do not raise different questions of safety and effectiveness. Testing and validation is completed. Test results show that the subject devices, the SOMATOM go. Platform and SOMATOM X. Platform with syngo CT VB10, are comparable to the predicate devices in terms of technological characteristics and safety and effectiveness and therefore are substantially equivalent to the predicate devices.

VIII. Performance Data

Non-Clinical Testing

Non-clinical testing, (integration and functional) including phantom tests were conducted for the SOMATOM go. Platform and SOMATOM X. Platform during product development. The modifications described in this Premarket Notification were supported with verification and validation testing.

The general purpose of each test is to verify and validate the functionality of the subject device modifications.

Testing will cover all related subsystems that contribute to the device modifications. Test levels are defined. For each test level several test activities are performed. The test specification and acceptance criteria are related to the corresponding requirements. Various test activities are

{31}------------------------------------------------

Image /page/31/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a cluster of orange dots of varying sizes.

performed to specific modifications on different test levels to ensure safe and effective integration in the system. Three test levels are defined:

System Validation test:

  • Acceptance test (workflow and user manual test) .
  • . Legal and Regulatory test

System Verification test:

  • System Integration Test (functional) ●
  • Functionality verification
  • . Image Quality (IQ) Evaluation

Tests are conducted for all software components developed in product development and for the complete product itself. Several activities are considered for this process, including creation of test specifications that relate to software/hardware requirements including tests to address risk mitigations that are identified, documented, and traced by hazard keys.

Additional evaluation tests are performed as bench tests to support the new device or device modification on Non-Clinical Performance Testing as listed in Table 10 below.

Table 10: Non-clinical performance testing (bench testing).
---------------------------------------------------------------------
Feature/Non-clinicalsupportive testingBench Testing performed
FAST 3D Camera /FAST IntegratedWorkflowThe bench test evaluates and compares the accuracy of the three sub-features FAST Isocentering, FAST Range, and FAST Direction to theaccuracy of the predicate device with syngo CT VA40 using the oldcamera hardware and the then only available ceiling mount.
The objectives of the bench test are to demonstrate that the FAST 3DCamera feature of the subject device with SOMARIS/10 syngo CT VB10,where the algorithms have been optimized for a new camera hardwarein two mounting positions, achieves comparable results as the predicatedevice with syngo CT VA40.
The FAST Isocentering accuracy of the subject device with syngo CT VB10is comparable to the predicate device with syngo CT VA40, regardless ofthe camera mounting position.
For the FAST Range feature, the detection accuracy of all body regionboundaries is comparable between the subject device with syngo CTVB10 and predicate device with syngo CT VA40. In the gantry mountingposition, the legs are often occluded by the torso when the patient islying head-first on the table. This is not a severe limitation, as legexaminations are usually performed feet-first.
The FAST Direction pose detection results are of comparable accuracy forsubject and predicate device, regardless of the camera mountingposition.
Overall, the SOMARIS/10 syngo CT VB10 delivers comparable accuracy tothe SOMARIS/10 syngo CT VA40 predicate for the new FAST 3D Camerahardware, also in the new gantry position.
Feature/Non-clinicalsupportive testingBench Testing performed
Multi-Purpose TableA CT system with extended distance between CT gantry and patient tablebase plus a mobile C-arm system were combined to evaluate thetechnical feasibility and possible limitations of this combination.The range of possible movement for the mobile C-arm in differentpositions between CT gantry and patient table was tested anddocumented by measurement of angles.Based on the test results it can be concluded that a CT scanner, equippedwith a Multi-Purpose (Vitus) Patient Table, which is installed withenhanced distance (674 mm) to the CT gantry and offers the iCT modefunctionality, provides sufficient freedom of movement for a mobile C-arm X-ray system to be used for clinical routine without any significantlimitations for the myNeedle Lasers or FAST 3D Camera.
Direct BreathholdThe test results showed that using the Direct Breathhold functionality, aspiral scan can automatically be triggered from an external respiratorygating device. The actual scan remains unchanged, and the object iscorrectly depicted in the resulting image.
ZeeFreeThe bench tests evaluate the performance of ZeeFree reconstruction.The objectives of the tests are to demonstrate:that the number of artefacts which can be attributed to a stackmisalignment (e.g. discontinuities in vessel structures,anatomical steps at air-soft-tissue interfaces, doubling of vesselor other anatomy) and which are often caused by incompletepatient breath-hold can be reduced in a "Cardiac Stack ArtefactCorrection" (SAC) reconstruction compared to the standardreconstruction with otherwise matching reconstructionparameters. that no artefacts are introduced by a SAC reconstruction. The test results show: If misalignment artefacts are identified in non-correctedstandard ECG-gated reconstructed sequence or spiral images,the feature "Cardiac Stack Artefact Correction" (SAC, marketingname: ZeeFree) enables optional stack artefact correctedimages, which reduce the number of alignment artefacts. The SAC reconstruction does not introduce new artefacts, whichwere previously not present in the non-corrected standardreconstruction. The SAC reconstruction does realize equivalent image quality inquantitative standard physics phantom-based measurements(ACR, Gammex phantom) in terms of noise, homogeneity, high-contrast resolution, slice thickness and CNR compared to a non-corrected standard reconstruction
Feature/Non-clinicalsupportive testingBench Testing performed
The SAC reconstruction does realize equivalent image quality in quantitative and qualitative phantom-based measurements with respect to metal objects compared to a non-corrected standard reconstruction. The SAC algorithm can be successfully applied to phantom data if derived from a suitable motion phantom demonstrating its correct technical function on the tested device. The SAC algorithm is independent from the physical detector width of the acquired data
myNeedle Guide(with myNeedleDetection)Tests were performed to ensure clinical usability of the myNeedle Guide needle detection algorithm. Two individual tests were performed. The accuracy of the automatic needle detection algorithm was tested. The reduction of necessary user interactions for navigating to a needle-oriented view with and without the support of the automatic needle detection algorithm was analyzed.It has been shown that the algorithm was able to consistently detect needle-tips over a wide variety of scans in 90.76% of cases.Further, the results of this bench test clearly show that the auto needle detection functionality reduces the number of interactions steps needed to generate a needle-aligned view in the CT Intervention SW.With successful Al-based needle tip detection, no user interaction is needed to achieve needle-aligned view during needle progression. Additional manual adjustment (fine tuning) of the needle-aligned view is always possible.

{32}------------------------------------------------

Image /page/32/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are several orange dots of varying sizes.

{33}------------------------------------------------

Image /page/33/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" are several orange dots arranged in a pattern.

A list of recognized and general consensus standards considered for the subject devices is provided as Table 11 and Table 12 below.

Table 11: Recognized Consensus Standards.

Date ofEntryRecognitionNumberStandardDevelopingOrganizationStandardDesignation Numberand DateTitle of Standard
12/19/202212-349NEMAPS 3.1 - 3.20 2022dDigital Imaging andCommunications in Medicine(DICOM) Set
07/06/202012-325NEMAXR 25-2019Computed Tomography DoseCheck
07/06/202012-330NEMAXR 28-2018Supplemental Requirementsfor User Information andSystem Function Related toDose in CT
Date ofEntryRecognitionNumberStandardDevelopingOrganizationStandardDesignation Numberand DateTitle of Standard
12/23/201912-328IEC61223-3-5 Edition2.0 2019-09Evaluation and routine testingin medical imagingdepartments - Part 3-5:Acceptance tests andconstancy tests - Imagingperformance of computedtomography X-ray equipment[Including: TechnicalCorrigendum 1 (2006)]
03/14/201112-226IEC61223-2-6 SecondEdition 2006-11Evaluation and routine testingin medical imagingdepartments - Part 2-6:Constancy tests - Imagingperformance of computedtomography X-ray equipment
01/14/201412-269IEC60601-1-3 Edition2.1 2013-04Medical electrical equipment- Part 1-3: Generalrequirements for basic safetyand essential performance -Collateral Standard: Radiationprotection in diagnostic X-rayequipment
06/27/201612-302IEC60601-2-44 Edition3.2: 2016Medical electrical equipment- Part 2-44: Particularrequirements for the basicsafety and essentialperformance of x-rayequipment for computedtomography
12/23/20195-125ANSI AAMIISO14971: 2019Medical devices - Applicationsof risk management tomedical devices
ISO14971 Third Edition2019-12Medical devices - Applicationof risk management tomedical devices
01/14/201913-79ANSI AAMIIEC62304:2006/A1:2016Medical device software -Software life cycle processes[Including Amendment 1(2016)]
IEC62304 Edition 1.12015-06Medical device software -Software life cycle processes
Date ofEntryRecognitionNumberStandardDevelopingOrganizationStandardDesignation Numberand DateTitle of Standard
CONSOLIDATEDVERSION
07/09/201419-46ANSI AAMIES60601-1:2005/(R)2012 &A1:2012,C1:2009/(R)2012 &A2:2010/(R)2012(Cons. Text) [Incl.AMD2:2021]Medical electrical equipment- Part 1: Generalrequirements for basic safetyand essential performance(IEC 60601-1:2005, MOD)[Including Amendment 2(2021)]
09/17/201819-36ANSI AAMIIEC60601-1-2:2014[Including AMD1:2021]Medical electrical equipment- Part 1-2: Generalrequirements for basic safetyand essential performance -Collateral Standard:Electromagnetic disturbances- Requirements and tests
IEC60601-1-2 Edition4.1 2020-09CONSOLIDATEDVERSIONMedical electrical equipment- Part 1-2: Generalrequirements for basic safetyand essential performance -Collateral Standard:Electromagnetic disturbances- Requirements and tests
12/23/20165-129ANSI AAMIIEC62366-1:2015+AMD1:2020(Consolidated Text)Medical devices Part 1:Application of usabilityengineering to medicaldevices, includingAmendment 1
IEC62366-1 Edition 1.12020-06CONSOLIDATEDVERSIONMedical devices - Part 1:Application of usabilityengineering to medicaldevices
07/09/201412-273IEC60825-1 Edition 2.02007-03Safety of laser products - Part1: Equipment classification,and requirements
12/21/20205-132IEC60601-1-6 Edition3.2 2020-07CONSOLIDATEDVERSIONMedical electrical equipment- Part 1-6: Generalrequirements for basic safetyand essential performance -Collateral standard: Usability
12/23/201912-309IEC60601-2-28 Edition2.0 2017-06Medical electrical equipment- Part 2-28: Particular
Date ofEntryRecognitionNumberStandardDevelopingOrganizationStandardDesignation Numberand DateTitle of Standard
requirements for the basicsafety and essentialperformance of X-ray tubeassemblies for medicaldiagnosis
12/20/202112-341IEC62563-1 Edition 1.22021-07CONSOLIDATEDVERSIONMedical electrical equipment- Medical image displaysystems - Part 1: Evaluationmethods

{34}------------------------------------------------

Image /page/34/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a pattern of orange dots.

{35}------------------------------------------------

Image /page/35/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are several orange dots of varying sizes.

{36}------------------------------------------------

Image /page/36/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.

Table 12: General Use Consensus Standards.

StandardDevelopingOrganizationStandard DesignationNumber and DateTitle of StandardHow was Standard Used
IEC60601-1:2005+A1:2012+A2:2020Medical electricalequipment - part 1:general requirementsfor basic safety andessential performanceANSI AAMI ES60601-1:2005/(R)2012 &A1:2012,C1:2009/(R)2012 &A2:2010/(R)2012 (Cons.Text) [Incl. AMD2:2021]
IEC/ISO17050-1Conformity Assessment– Supplier's declarationof conformity – Part 1:General requirementsDeclaration ofconformance to FDArecognized consensusstandards.
IEC/ISO17050-2Conformity assessment– Supplier's declarationof conformity – Part 2:Supportingdocumentation.General consensusstandards not currentlyrecognized by FDA.

{37}------------------------------------------------

Image /page/37/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that form an arrow shape.

A list of applicable guidance documents considered for this submission is provided as Table 13 below.

Table 13: FDA Guidance Document and Effective Date
FDA Guidance DocumentIssue date
User Fees and Refunds for Premarket Notification Submissions (510(k)s)10/05/2022
Refuse to Accept Policy for 510(k)s04/21/2022
Electronic Submission Template for Medical Device 510(k) Submissions10/2/2023
Deciding When to Submit a 510(k) for a Change to an Existing Device10/25/2017
Deciding When to Submit a 510(k) for a Software Change to an ExistingDevice10/25/2017
The 510(k) Program: Evaluating Substantial Equivalence in PremarketNotifications [510(k)]07/28/2014
Content of Premarket Submissions for Software Contained in MedicalDevices06/14/2023
Off-The-Shelf Software Use in Medical Devices08/11/2023
Applying Human Factors and Usability Engineering to Medical Devices02/03/2016
Pediatric Information for X-ray Imaging Device Premarket Notifications11/28/2017
Cybersecurity in Medical Devices: Quality System Considerations andContent of Premarket Submissions09/27/2023
Electromagnetic Compatibility (EMC) of Medical Devices06/06/2022
Design Considerations and Pre-market Submission Recommendations forInteroperable Medical Devices09/06/2017
Appropriate Use of Voluntary Consensus Standards in PremarketSubmissions for Medical Devices09/14/2018

Verification and Validation

Software documentation at an Enhanced Documentation level per FDA's Guidance Document "Content of Premarket Submissions for Software Contained in Medical Devices" issued on June 14, 2023 is included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claims of substantial equivalence.

{38}------------------------------------------------

Image /page/38/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are a series of orange dots arranged in a circular pattern.

Cybersecurity

Siemens conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with guidance document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" issued on September 27, 2023 is included within this submission.

Wireless Coexistence Testing

Additionally, Siemens conforms to the requirements for Radio Frequency Wireless Technology as defined in FDA guidance document "Radio Frequency Wireless Technology in Medical Devices, Guidance for Industry and Food and Drug Administration Staff" issued on August 14, 2013 by adhering to the EMC and risk based verification and validation requirements in design, testing, and labeling of the wireless remote control components of the subject devices.

The Radio Frequency Wireless Technology of the optional Remote Scan Control and supporting Control Device iPad for Scan&GO complies to 47 CFR part 15 subpart c – Intentional Radiators. All Radio device labels will show an FCC ID code to show compliance. Shielding requirement applicable to the SOMATOM go. Platform and SOMATOM X. Platform and respective Scatter Radiation diagrams for typical room installations are provided in the User Documentation and Planning Guide of the intended Scanners in accordance with IEC60601-2-44.

Siemens has considered several measures to address wireless coexistence by design to ensure the safe operation of the wireless components in combination with the applicable system supported functionality. Wireless technology in the system setup to perform a task in a given shared environment where other systems have an ability to perform their tasks and may or may not be using the same set of rules has been considered. According to FDA guidance 'Radio Frequency Wireless Technology in Medical Devices" Siemens has addressed the safety, effectiveness, and high likelihood of coexistence with other devices of this technology in our product design by our Risk Management Process, Failure Mode and Effects Analysis (FMEA) Process, and Requirement Engineering Process. As part of the risk management process, hazardous situations associated with the Scan&GO and its connection to the host system via Wi-Fi were addressed as part of the Risk Management process.

Testing for co-existence considered for following scenarios:

  • Co-Channel Testing
  • . Adjacent Channel Testing
  • . RF Interference Testing
  • . Separation Distance/Location Testing

Scan&GO is designed to allow dynamic frequency selection and transmission power control by default in accordance with IEEE 802.11h. Adjacent channel testing is addressed by the fact that Scan&GO does not support shared medium access to Siemens Wi-Fi network. RF interference was tested by successfully ensuring that wireless communications were actively transmitting in situations where possible interference may exist. Recommended distance and router locations requirements are documented in the user documentation.

{39}------------------------------------------------

Image /page/39/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" are several orange dots of varying sizes arranged in a diagonal line.

Summary

The features described in this premarket notification are supported with verification and validation testing, dosimetry and imaging performance, and analysis of phantom images to assess device and feature performance during product development. The risk analysis was completed, and risk control implemented to mitigate identified hazards. The test results show that all of the software specifications have met the acceptance criteria. Verification testing of the device was found acceptable to support the claim of substantial equivalence.

General Safety and Effectiveness Concerns

The device labeling contains instructions for use as well as necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a system related risk analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing according to the risk management process. In order to minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.

IX. Conclusions

The predicate/secondary devices were cleared based on the results of non-clinical and clinical testing including verification and validation, phantom tests, validation on clinical images, and supportive literature. The non-clinical and clinical data supports the safety of the device and the hardware and software verification and validation demonstrates that the subject devices SOMATOM go. Platform and SOMATOM X. Platform perform as intended in the specified use conditions. The data included in this submission demonstrates that the SOMATOM go. Platform and SOMATOM X. Platform with described modifications perform comparably to the predicate devices currently marketed for the same intended use. Siemens believes that the data generated from the SOMATOM go. Platform and SOMATOM X. Platform testing supports a finding of substantial equivalence.

N/A