(466 days)
The Athelas Home is indicated for the quantitative determination of white blood cells (WBC) and Neutrophil percentages (NEUT%) in capillary whole blood from fingerstick for patient self-testing with results viewable by healthcare professionals, and in capillary or K2EDTA venous whole blood for multiple-patient use in point-of-care settings. For self-testing, the Athelas Home is intended to be used by a single person and should not be shared. The Athelas Home is only to be used with Athelas Test Strips. The Athelas Home is intended for adult patients (aged 21 and older) at risk of neutropenia. For self-testing patients with psychiatric conditions, clinical judgment should be exercised when deciding the end-user and based on the instructions for use (IFU), the treating physician should determine which patients are competent to perform the test by themselves. Results obtained with the Athelas Home should not be the sole basis for patient diagnosis, treatment, or management of leukopenia and neutropenia, and all results should be evaluated by a healthcare provider. Prescription Use Only.
The Athelas Home is identical to the Athelas Home (K200828) and Athelas One (K181288) in its technology, core software, the analytes measured, all components, and usage process.
The Athelas Home is an automated diagnostic device intended to perform tests on whole blood samples collected in K2EDTA or capillary finger stick samples collected directly into the Athelas test strip. Athelas Home returns quantitative WBC and Neut% metrics from the blood sample. The Athelas Home system is for In Vitro Diagnostic use and for prescription use only.
The Athelas Home is only to be used with Athelas Test Strips. The Athelas Home is indicated for use by patients and caregivers within home settings and at Point-of-Care settings with results viewable by health care professionals. The Athelas Home is indicated for adult populations only (aged 21 and older).
The Athelas Home is not indicated to completely replace a patient's monitoring regimen on reference lab systems. Prescribers should continue monitoring patients using traditional lab-based means as deemed necessary and possible.
The Athelas Home is not indicated to directly diagnose or assist in disease management. Rather, the Athelas Home should be used to quantify WBC and Neut% only, with all decision making owned by the health care professional as per standard of care.
The Athelas Home uses image processing and microfluidics in order to measure WBC and NEUT% values from whole blood in point-of-care and home settings. A drop of blood is placed on the Athelas test strip either directly from the finger or via pipette from K2EDTA whole blood tube. The device uses single-use, disposable test strips, ensuring there is no direct patient contact or contamination with the device itself.
The Athelas Home test strip stains and creates a monolayer of the blood sample within the chamber. The strip is inserted into the test strip slot of the Athelas Home device where a built-in camera captures images of the sample. The images are automatically transmitted to the cloud server where they are processed and analyzed by an image processing algorithm. The quantitative results of WBC, NEUT% results are based on the concentrations and types of cells present in the sample.
Here's a breakdown of the acceptance criteria and study information for the Athelas Home device, based on the provided FDA 510(k) clearance letter:
1. Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance (WBC) | Reported Device Performance (Neut%) |
|---|---|---|
| WBC Precision: | ||
| - 7.5% CV above 2K/µL WBC | Within-run Precision/Reproducibility: 2.20 K/µL (5.87% CV), 3.75 K/µL (5.46% CV), 4.12 K/µL (5.09% CV), 5.11 K/µL (5.96% CV), 7.89 K/µL (4.82% CV), 10.01 K/µL (5.45% CV), 14.64 K/µL (4.69% CV), 17.52 K/µL (4.34% CV), 23.33 K/µL (5.81% CV). All met. | |
| - 0.25 K/µL SD below 2K/µL WBC | Not explicitly detailed for each sample below 2K/µL, but overall "All results met the predefined specifications (CV%) for precision." "Overall reproducibility levels were found to meet the acceptance criteria of 7.5% CV for WBC." (Implicitly including SD for lower range). Low-level QC (2.746 K/µL) showed a SD of 0.153 K/µL. | |
| WBC Bias/Error: | ||
| - ±7.5% error above 2K/µL WBC | Method Comparison: -2.31% Mean % Bias (1.1 - 23 K/µL range). Satisfied. | |
| - ±0.25 K/µL error below 2K/µL WBC | Not explicitly detailed for each sample below 2K/µL. | |
| Neutrophil % Precision: | ||
| - 5% SD OR 15% CV | Within-run Precision/Reproducibility: Not explicitly detailed for each sample, but overall "All results met the predefined specifications...for precision." (Implicitly including Neut%). | |
| Reproducibility (Overall Summary Table - Low): 2.680 SD (5.278% CV) Within-run. Total SD 3.443 (6.780% CV). Satisfied. | ||
| Reproducibility (Overall Summary Table - Medium): 3.015 SD (6.030% CV) Within-run. Total SD 3.344 (6.689% CV). Satisfied. | ||
| Reproducibility (Overall Summary Table - High): 2.967 SD (5.838% CV) Within-run. Total SD 3.316 (6.525% CV). Satisfied. | ||
| Neutrophil % Bias: | ||
| - ±10% bias or ±5% Neut% total error (whichever larger) | Method Comparison: 1.18% Mean % Bias (8 - 92.89% range). Satisfied. | |
| At-Home Precision (WBC) | At-Home Precision Repeatability: 5.4% CV (combined patient blood and QC material). Pass (Evaluation Criteria: 7.5% CV). | |
| At-Home Precision (Neut%) | At-Home Precision Repeatability: 1.2% SD (combined patient blood and QC material). Pass (Evaluation Criteria: 5% SD or 15% CV). | |
| Linearity (WBC) | R² = 0.997, Slope = 1.013, Intercept = 0.0449, CVr = 5.08% as per OLS regression. Demonstrated to be linear from lower limit to upper limit and within measured allowable max % diff for each interval. | |
| Flagging Accuracy | % Positive Agreement (Sensitivity): 90.91%. % Negative Agreement (Specificity): 96.71%. % Overall Agreement: 94.87%. Met the specification of >= 90%. | |
| Matrix Comparability (WBC - venous vs capillary) | WBC: Slope 1.026 (1.000, 1.055), Intercept -0.145 (-0.319, 0.018), Mean Bias 0.056 K/µL, Mean % Bias -0.588. Met all evaluation criteria. | |
| Matrix Comparability (Neut% - venous vs capillary) | Neut%: Slope 0.999 (0.938, 1.058), Intercept 0.457 (-2.502, 3.614), Mean Bias 0.162 Percentage Points, Mean % Bias -0.333. Met all evaluation criteria. | |
| At-Home Method Comparison (WBC - Athelas One vs Athelas Home) | WBC: Slope 1.0 (0.95, 1.03), Intercept 0 (-0.16, 0.31), Mean Bias 0, Mean % Bias 0.0. Met all evaluation criteria. | |
| At-Home Method Comparison (Neut% - Athelas One vs Athelas Home) | Neut%: Slope 1.0 (0.98, 1.09), Intercept -1.0 (-6.09, 0.0), Mean Bias -0.5, Mean % Bias 1.3. Met all evaluation criteria. | |
| At-Home Error Rate | Over 94% of tests run successfully with no errors. Errors that were found were resolved after running a single additional test-strip. | |
| At-Home Usability | Large majority of patients found the Athelas Home easy to use and had very little difficulty running the test end to end. (e.g., Lancet use: 32/32 Yes; Test-strip fill: 30/32 Yes; No errors (non-code): 31/32 Yes; App use: 32/32 Yes). | |
| Fill-Volume Testing | Robust to variable volumes of blood. Errors returned for 0µL. Accurate results within bias criteria for 1.4µL, 3.4µL, and 5.4µL fill volumes across various sample concentrations. | Robust to variable volumes of blood. Errors returned for 0µL. Accurate results within bias criteria for 1.4µL, 3.4µL, and 5.4µL fill volumes across various sample concentrations. |
| Device Tilt Testing | Not affected by tilted placement. No significant bias across tilt orientations and samples. All WBC criteria met. | Not affected by tilted placement. No significant bias across tilt orientations and samples. All Neut% criteria met. |
| Device Drop Testing | Robust to drop and vibration schedules. All 12 devices passed functional tests after testing, with no critical physical damage or functional defects, and WBC results remained within reference range. | Robust to drop and vibration schedules. All 12 devices passed functional tests after testing, with no critical physical damage or functional defects, and Neut% results remained within reference range. |
| Device/Test-Strip Temperature Testing | Both device and test-strip continued to meet performance specifications after exposure to extreme temperatures. All WBC bias criteria met. | Both device and test-strip continued to meet performance specifications after exposure to extreme temperatures. All Neut% bias criteria met. |
| Humidity Flex Testing (Test-Strips) | Test-strips continued to function within specifications after exposure to extreme humidity conditions. No evidence of condensation, fill-time within criteria (< 25 seconds), and no color change present for all 8 test strips across schedules. | |
| Vibration Testing | System successfully met evaluation criteria post vibration test. | |
| Cleaning and Disinfection | No change in performance or external materials after 10,000 cleaning and disinfection cycles using Caviwipe. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Bench Testing (Within-run Precision):
- Sample Size: 9 whole blood samples. 10 replicates per sample, per operator, per test strip lot. Total of 90 tests per sample (90 tests x 9 samples = 810 tests).
- Data Provenance: Not specified, but generally from a laboratory setting.
- Bench Testing (Between-run Reproducibility):
- Sample Size: Three levels of ATH-check quality control material (Low, Normal, High). 240 readings per level (2 runs x 20 days x 2 replicates x 3 sites). Total of 720 readings across all levels.
- Data Provenance: Not specified, but implied to be from laboratory/controlled settings across three sites.
- At-Home Precision Study (Patient In-home Whole Blood):
- Sample Size: 24 patients, 2 replicates per patient. Total of 48 individual tests.
- Data Provenance: From at-home users in home environments in the US (implied by "at-home settings in the US" in subsequent sections and the study name).
- At-Home Precision Study (Patient In-home QC Material):
- Sample Size: 10 patients, 3 replicates per patient. Total of 30 individual tests.
- Data Provenance: From at-home users in home environments in the US.
- Linearity:
- Sample Size: 10 samples run in 4 replicates on 4 devices. Total of 160 tests.
- Data Provenance: Samples obtained by pooling "one low WBC concentration fresh whole blood sample one high WBC concentration sample." Generally laboratory prepared samples.
- Interfering Substances:
- Sample Size: Not explicitly stated how many individual samples were spiked and tested for each substance, but a list of 16 substances were tested.
- Data Provenance: Prepared whole blood samples, some naturally occurring, some spiked.
- Reference Intervals:
- Sample Size: 120 healthy donors.
- Data Provenance: Not specified, implied to be based on healthy donors.
- Limit of Blank (LoB):
- Sample Size: 120 total repeated measurements of blank samples.
- Limit of Detection (LoD):
- Sample Size: 2 test strip lots (60 strips each), 5 low-level samples, 2 replicates per sample. Total of 20 low-level tests for LoD calculation (across each lot, 5 samples x 2 replicates).
- Limit of Quantification (LoQ):
- Sample Size: 2 test-strip lots, 1 instrument, 3 days, 3 replicates per sample, 4 independent low-level whole blood samples. Total of 36 low-level tests (4 samples x 3 replicates x 3 days?) + more for the two lots.
- Data Provenance: Distinct whole blood samples acquired from patients in K2EDTA tubes, presumably from a laboratory or clinical setting. Diluted to target concentrations.
- Specimen Stability (Whole Blood):
- Sample Size: 9 different venous blood samples (low, normal, high WBC levels).
- Data Provenance: Venous blood samples.
- Specimen Stability (Test Strip):
- Sample Size: Two test strip lots compared over 26 months, 3 replicates, multiple QC lots. Three replicates per two lots of strips across 3 levels of control fluid run every 60 days.
- Method Comparison:
- Sample Size: 312 patient samples.
- Data Provenance: Taken at 3 point-of-care sites in the US.
- Flagging Comparison:
- Sample Size: 312 samples (same as Method Comparison).
- Data Provenance: Same as Method Comparison (3 point-of-care sites in the US).
- Matrix Comparison (Venous vs. Capillary):
- Sample Size: 59 patients.
- Data Provenance: Capillary finger-prick blood samples and K2EDTA venous whole blood samples from the same patients.
- At-Home Method Comparison:
- Sample Size: 165 samples.
- Data Provenance: Taken in at-home settings in the US.
- At-Home Error Analysis:
- Sample Size: 81 tests (derived from percentages: 76 successful, 3+2+1+1=7 errors, 76/0.94 = 80.85; 7/0.06 = 116.6; approximation derived from given percentages. The actual number is not precisely 81 based on the sum, but implies around this amount.)
- Data Provenance: At-home testing.
- At-Home Usability Analysis:
- Sample Size: 32 users.
- Data Provenance: At-home users during at-home visits.
- Fill-Volume Testing:
- Sample Size: 3 samples tested across 4 fill volumes (0µL, 1.4µL, 3.4µL, 5.4µL). Number of replicates not explicitly stated, but common for such studies.
- Data Provenance: K2EDTA whole blood samples for predicate comparison.
- Device Tilt Testing:
- Sample Size: 3 samples tested across 4 tilt conditions. Number of replicates not explicitly stated.
- Data Provenance: K2EDTA whole blood samples for predicate comparison.
- Device Drop Testing:
- Sample Size: 12 devices.
- Data Provenance: Device testing.
- Device and Test-Strip Temperature Testing:
- Sample Size: Device testing: 3 devices, 9 replicates each, across 3 environmental conditions. Test-strip testing: 3 test-strip lots, 9 replicates each, across 3 environmental conditions.
- Data Provenance: Device and test strips exposed to controlled environmental conditions.
- Vibration Testing:
- Sample Size: Not specified, but refers to "The system".
- Data Provenance: Third party ISTA certified test lab Quanta Testing.
- Humidity Flex Testing:
- Sample Size: 8 test-strips, subject to specified schedules.
- Data Provenance: Test-strips exposed to controlled humidity chambers at Quanta Laboratories.
3. Number of Experts Used to Establish Ground Truth and Qualifications
The document does not explicitly state the number or qualifications of experts used to establish ground truth.
However, it implicitly relies on:
- Predicate devices: The Sysmex XE-5000 and Athelas One are frequently used as the "ground truth" or reference method for comparison. These are legally marketed, validated automated hematology analyzers.
- Clinical Laboratory Standards Institute (CLSI) recommendations: The studies (precision, linearity, LoQ, stability) frequently refer to CLSI guidelines (e.g., CLSI EP05-A3, EP06-A, EP17-A2, EP25-A) as the basis for their methodologies and acceptance criteria, implying adherence to established best practices in laboratory medicine.
- Medical decision levels: The testing was conducted around "medical decision levels," which are established by clinical experts.
- Morphological flagging: This implies comparison against expert (e.g., pathologist, trained medical technologist) review of blood smears, though not directly stated.
Without explicit mention, it's not possible to state "X number of radiologists with Y years of experience," as the device is for hematology, not radiology.
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method for the test set (e.g., 2+1, 3+1, none).
- For method comparison studies, the predicate device (Sysmex XE-5000 or Athelas One) serves as the reference, which performs automated analysis. The "ground truth" is effectively the results from this established, high-accuracy device.
- For flagging comparison, the Sysmex XE-5000's flagging is used as the reference, which would typically be based on its internal algorithms and potentially validated against expert manual differential counts in its own clearance studies.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. The studies focus on the device's analytical performance against predicate devices or its own established metrics, not on how human readers improve with or without AI assistance. The Athelas Home is an automated system providing quantitative results, not an AI-assisted diagnostic aid for human interpretation.
6. Standalone (Algorithm Only) Performance
Yes, the studies primarily describe standalone performance of the algorithm.
- The device captures images, transmits them to a cloud server, where they are "processed and analyzed by an image processing algorithm."
- The results for WBC and NEUT% are "based on the concentrations and types of cells present in the sample" as determined by this algorithm.
- All the performance metrics (precision, bias, linearity, LoD, LoQ, method comparison, flagging) describe the output of the device (i.e., the algorithm and hardware combined) compared to a reference method, without direct human-in-the-loop assistance for interpretation of individual results. Human healthcare professionals evaluate the overall results and provide diagnosis/treatment.
7. Type of Ground Truth Used
- Predicate device measurements: The primary ground truth for quantitative measurements (WBC, NEUT%) was the Sysmex XE-5000 analyzer (a widely accepted automated hematology analyzer) and the Athelas One (the previous generation predicate Athelas device).
- Quality Control (QC) material values: For precision and linearity studies, established values of QC materials (ATH-CHECK) were also used.
- Blank samples: For Limit of Blank.
- Diluted patient samples: For Limit of Quantification.
- Expert consensus/pathology (implied): For flagging comparison and verification of morphological flags, the reference Sysmex XE-5000's flagging implies an underlying ground truth (likely derived from expert manual differentials or pathology review during its own validation), which the Athelas One then compares to. This is not directly stated for Athelas Home validation, however.
- Pre-established reference intervals: For verifying reference intervals against Sysmex XE-5000.
- Outcomes data (No): Not mentioned as a ground truth.
8. Sample Size for the Training Set
The document does not specify the sample size for the training set. It focuses entirely on the validation/test set performance. It mentions that the "images are automatically transmitted to the cloud server where they are processed and analyzed by an image processing algorithm," implying a machine learning component, but provides no details on its development or training data.
9. How the Ground Truth for the Training Set Was Established
As the training set sample size is not specified, neither is the method for establishing its ground truth. However, given the nature of the device (automated cell counter), it would typically involve:
- Expert manual differential counts: Morphological classification and counting of various WBC types on blood smears by highly trained medical technologists or pathologists.
- Reference laboratory analyzer results: Pairing manual counts with results from established automated hematology analyzers for quantitative parameters.
- Annotated images: Images used for training would require meticulous annotation by experts to identify and classify different cell types.
FDA 510(k) Clearance Letter - Athelas Home
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.02
February 6, 2026
Athelas Inc.
Nicole Clausen
GM Athelas Home & Head of Pharmacy
1300 Terra Bella Ave, Suite 200
Mountain View, California 94043
Re: K243348
Trade/Device Name: Athelas Home
Regulation Number: 21 CFR 864.5220
Regulation Name: Automated differential cell counter
Regulatory Class: Class II
Product Code: GKZ
Dated: January 6, 2026
Received: January 7, 2026
Dear Nicole Clausen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K243348 - Nicole Clausen Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
Page 3
K243348 - Nicole Clausen Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Takeesha Taylor-Bell
Takeesha Taylor-Bell
Deputy Director
Division of Immunology and Hematology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K243348
Device Name: Athelas Home
Indications for Use (Describe)
The Athelas Home is indicated for the quantitative determination of white blood cells (WBC) and Neutrophil percentages (NEUT%) in capillary whole blood from fingerstick for patient self-testing with results viewable by healthcare professionals, and in capillary or K2EDTA venous whole blood for multiple-patient use in point-of-care settings. For self-testing, the Athelas Home is intended to be used by a single person and should not be shared. The Athelas Home is only to be used with Athelas Test Strips. The Athelas Home is intended for adult patients (aged 21 and older) at risk of neutropenia. For self-testing patients with psychiatric conditions, clinical judgment should be exercised when deciding the end-user and based on the instructions for use (IFU), the treating physician should determine which patients are competent to perform the test by themselves. Results obtained with the Athelas Home should not be the sole basis for patient diagnosis, treatment, or management of leukopenia and neutropenia, and all results should be evaluated by a healthcare provider. Prescription Use Only.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
Page 5
510(k) Summary
Date Prepared: January 29th, 2026
Submitter: Tanay Tandon (Athelas)
Company: Athelas Inc.
Address: 1300 Terra Bella Suite 200, Mountain View, CA 94043
Email Address: tanay@athelas.com
Proprietary Trade name: Athelas Home
Common name: Automated differential cell counter.
Classification name: Automated differential cell counter (21 CFR 864.5220, Product Code GKZ)
Type of Test or Tests Performed: WBC and NEUT% on Whole Blood (venous K2EDTA and capillary whole blood). Identical to predicate Athelas One K181288 and Athelas Home K200828 for home-use indication.
Indicated use: The Athelas Home is indicated for the quantitative determination of white blood cells (WBC) and Neutrophil percentages (NEUT%) in capillary whole blood from fingerstick for patient self-testing with results viewable by healthcare professionals, and in capillary or K2EDTA venous whole blood for multiple-patient use in point-of-care settings. For self-testing, the Athelas Home is intended to be used by a single person and should not be shared. The Athelas Home is only to be used with Athelas Test Strips. The Athelas Home is intended for adult patients (aged 21 and older) at risk of neutropenia. For self-testing patients with psychiatric conditions, clinical judgment should be exercised when deciding the end-user and based on the instructions for use (IFU), the treating physician should determine which patients are competent to perform the test by themselves. Results obtained with the Athelas Home should not be the sole basis for patient diagnosis, treatment, or management of leukopenia and neutropenia, and all results should be evaluated by a healthcare provider. Prescription Use Only.
Substantial Equivalence: Predicate: Athelas Home, Automated Haematology Analyzer (K200828); Athelas One, Automated Haematology Analyzer (K181288)
Page 6
Comparison with predicate:
Similarities
| Item | Device: Athelas Home | Predicate: Athelas Home | Predicate: Athelas One |
|---|---|---|---|
| Intended Use | The Athelas Home is indicated for the quantitative determination of white blood cells (WBC) and Neutrophil percentages (NEUT%) in capillary whole blood from fingerstick for patient self-testing with results viewable by healthcare professionals, and in capillary or K2EDTA venous whole blood for multiple-patient use in point-of-care settings. For self-testing, the Athelas Home is intended to be used by a single person and should not be shared. The Athelas Home is only to be used with Athelas Test Strips. The Athelas Home is intended for adult patients (aged 21 and older) at risk of neutropenia. For self-testing patients with psychiatric conditions, clinical judgment should be exercised when deciding the end-user and based on the instructions for use (IFU), the treating physician should determine which patients are competent to perform the test by themselves. Results obtained with the Athelas Home should not be the sole basis for patient diagnosis, treatment, or management of leukopenia and neutropenia, and all results should be evaluated by a healthcare provider. Prescription Use Only. | The Athelas Home is indicated for the quantitative determination of white blood cells (WBC) and Neutrophil percentages (NEUT%) in capillary whole blood. The Athelas Home system is for In Vitro Diagnostic use and for prescription use only. The Athelas Home is only to be used with Athelas Home Test Strips. The Athelas Home is indicated for use by patients and caregivers within home settings with results viewable by health care professionals. The Athelas Home is intended for patients at risk of neutropenia. For patients with psychiatric conditions, clinical judgment should be exercised when deciding the end-user. Based on the instructions for use (IFU) of Athelas Home, the treating physician should determine which patients are competent to perform the test by themselves. The Athelas Home is indicated for adult populations only (aged 21 and older). Results obtained with the Athelas Home should not be the sole basis for patient diagnosis, treatment or management of leukopenia and neutropenia. All results should be evaluated by a healthcare provider. The Athelas Home is intended to be used by a single person and should not be shared. Prescription Use Only. | Athelas One is indicated for use for quantitative determination of white blood cells (WBC) and Neutrophil percentages (NEUT%) in capillary or K2EDTA venous whole blood. The Athelas One system is for In Vitro Diagnostic use only. The Athelas One is only to be used with Athelas One Test Strips. The Athelas One is indicated for use in clinical laboratories and for point of care settings. The Athelas One is only indicated for use in adult populations (aged 21 and older). |
| Specimen Type | Capillary and Venous Whole Blood | Capillary Whole Blood | Capillary and Venous Whole Blood |
| Modes of Operation | Single mode of operation for both venous and capillary samples | Single mode of operation for capillary samples | Single mode of operation for both venous and capillary samples |
| Software/Hardware | Equivalent | Internet connected device for processing results on cloud server. | Internet connected device for processing results on cloud server. |
| Sample Volume | Equivalent | 3.5 uL | 3.5 uL |
Page 7
| Measurement Range | Equivalent | WBC: 1 - 25K/μL | WBC: 1 - 25K/μL |
| Calibration | Equivalent | Factory calibrated and automatic calibration at the beginning of each test. No manual calibration by the end user. | Factory calibrated and automatic calibration at the beginning of each test. No manual calibration by the end user. |
| Test Principle | Equivalent | A microfluidic test strip channel creates a stained monolayer of white blood cells. Multiple images are taken of the monolayer and the cells are counted and classified by computer vision based image analysis. | A microfluidic test strip channel creates a stained monolayer of white blood cells. Multiple images are taken of the monolayer and the cells are counted and classified by computer vision based image analysis. |
| Parameters | Equivalent | WBC, NEUT% | WBC, NEUT% |
| Reagents | Equivalent | Cresyl Violet, Methylene Blue Stain (pre-loaded/coated dry and contained in test strip) | Cresyl Violet, Methylene Blue Stain (pre-loaded/coated dry and contained in test strip) |
| Controls/ Calibrators | Auto-Check 3-Level Gridded Strip OR ATH-CHECK (3 level control) | Auto-Check 3-Level Gridded Strip OR ATH-CHECK (3 level control) | ATH-CHECK (3 level control) |
Differences
| Item Name | Device: Athelas Home | Predicate: Athelas Home | Predicate: Athelas One |
|---|---|---|---|
| Intended use sites | Home, Point-of-Care | Home | Point of Care, Clinical Laboratory |
| Result Viewing | Results are remotely and securely transmitted to a web-based application with role-based access controls. | Results are remotely and securely transmitted to the prescribing physician or healthcare provider. | Results are shown on the mobile/tablet application used for system operation. |
| Remote Lockout Capability | Yes | Yes | No |
| Target Population | Adult patients (21 and older) at risk of neutropenia. | Adult patients (21 and older) at risk of neutropenia. | Adults (21 and older). |
Description of Device:
The Athelas Home is identical to the Athelas Home (K200828) and Athelas One (K181288) in its technology, core software, the analytes measured, all components, and usage process.
The Athelas Home is an automated diagnostic device intended to perform tests on whole blood samples collected in K2EDTA or capillary finger stick samples collected directly into the Athelas test strip. Athelas Home returns quantitative WBC and Neut% metrics from the blood sample. The Athelas Home system is for In Vitro Diagnostic use and for prescription use only.
The Athelas Home is only to be used with Athelas Test Strips. The Athelas Home is indicated for use by patients and caregivers within home settings and at Point-of-Care settings with results viewable by health care professionals. The Athelas Home is indicated for adult populations only (aged 21 and older).
Page 8
The Athelas Home is not indicated to completely replace a patient's monitoring regimen on reference lab systems. Prescribers should continue monitoring patients using traditional lab-based means as deemed necessary and possible.
The Athelas Home is not indicated to directly diagnose or assist in disease management. Rather, the Athelas Home should be used to quantify WBC and Neut% only, with all decision making owned by the health care professional as per standard of care.
The Athelas Home uses image processing and microfluidics in order to measure WBC and NEUT% values from whole blood in point-of-care and home settings. A drop of blood is placed on the Athelas test strip either directly from the finger or via pipette from K2EDTA whole blood tube. The device uses single-use, disposable test strips, ensuring there is no direct patient contact or contamination with the device itself.
The Athelas Home test strip stains and creates a monolayer of the blood sample within the chamber. The strip is inserted into the test strip slot of the Athelas Home device where a built-in camera captures images of the sample. The images are automatically transmitted to the cloud server where they are processed and analyzed by an image processing algorithm. The quantitative results of WBC, NEUT% results are based on the concentrations and types of cells present in the sample.
For further technological details refer to Athelas One K181288 application and Athelas Home K200828 application.
The precision/bias goals of the device are provided below:
Target Evaluation Criteria:
- WBC Precision: 7.5% CV above 2K/µL WBC. 0.25 K/µL SD below 2K/µL WBC.
- WBC Bias/Error: ±7.5% error above 2K/µL WBC. ±0.25 K/µL error below 2K/µL WBC.
- Neutrophil % Precision: 5% SD OR 15% CV
- Neutrophil % Bias: ±10% bias or ±5% Neut% total error (whichever larger).
These targets were chosen based on CLSI clinical precision/bias recommendations for WBC and Neut%. Constant value targets were set to account for asymptotically increasing percent imprecision towards lower concentrations as per the poisson distribution.
Description of Intended Setting:
The Athelas Home is indicated for use by patients and caregivers within home settings and at Point-of-Care settings with results viewable by health care professionals. The device should be placed in a stationary and dry location. The Athelas Home labeling highlights the specific conditions (temperature, humidity) for storage and operation, which are identical to the FDA-cleared Athelas Home (K200828) and Athelas One (K181288).
Page 9
Summary of Technological Characteristics Compared to Predicate Devices:
The Athelas Home is identical to the Athelas Home (K200828) and Athelas One (K181288) in its technology, core software, the analytes measured, all components, and usage process.
Testing: Bench testing, Clinical testing, and Flex Testing from the Athelas One K181288 and Athelas Home K200828 were used to support the Athelas Home for PoC use because the devices are identical in its technology, core software, the analytes measured, all components, and usage process.
Bench
Within-run Precision/Reproducibility
Within-run Precision Repeatability
Precision studies were performed using K2EDTA whole blood samples around medical decision levels and the upper and lower limit of the analytical measuring range. The study was conducted with nine whole blood samples, three different operators and three different test strip lots. Care was taken to include both normal and abnormal samples.
Three operators and three test strip lots were used in conjunction. Per operator, per lot, ten replicates of each sample were tested and the results were recorded. 90 tests were run per sample, with 810 tests run in total. The mean, standard deviation (SD), and coefficient of variation (CV) were calculated for each sample. All results met the predefined specifications (CV%) for precision (targets shown on below table as well).
WBC Summarized
| Repeatability | Between-Lot | Between-Instrument | Between-Operator | Total | Target Evaluation | |
|---|---|---|---|---|---|---|
| Sample | Mean Value (K/µL) | N | SD (K/µL) | CV (%) | SD (K/µL) | CV (%) |
| 1 | 2.20 | 90 | 0.12 | 5.62 | 0.00 | 0.00 |
| 2 | 3.75 | 90 | 0.20 | 5.42 | 0.00 | 0.00 |
| 3 | 4.12 | 90 | 0.20 | 4.78 | 0.00 | 0.00 |
| 4 | 5.11 | 90 | 0.25 | 4.96 | 0.00 | 0.00 |
| 5 | 7.89 | 90 | 0.33 | 4.18 | 0.06 | 0.78 |
| 6 | 10.01 | 90 | 0.50 | 5.01 | 0.00 | 0.00 |
| 7 | 14.64 | 90 | 0.66 | 4.51 | 0.19 | 1.27 |
| 8 | 17.52 | 90 | 0.70 | 3.97 | 0.00 | 0.00 |
| 9 | 23.33 | 90 | 1.01 | 4.33 | 0.77 | 3.31 |
Page 10
Precision Reproducibility (between run)
Precision/Reproducibility studies were performed on the Athelas One Whole Blood Manual Mode using three levels of ATH-check quality control material (Low, Normal and High). The 20-day study used three sites, with three analyzers (one instrument per site), three lots of test strips, and one lot of quality control materials (all levels). We performed two runs per day and two replicates per run with two operators per site. A total of 80 readings were generated at each site for each level of control (2 runs x 20 days x 2 replicates = 80 readings). As per Agency recommendation results were analyzed using ANOVA analysis and taking into account the nested design of the study.
Low
Overall Summary Table
| Measurand | Mean Value | N | Within-Run | Between-Run | Between-Day | Between-Site | Total |
|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | |||
| WBC | 2.746 | 240 | 0.140 | 5.097 | 0.000 | 0.000 | 0.060 |
| NEUT % | 50.781 | 240 | 2.680 | 5.278 | 1.239 | 2.440 | 1.305 |
Medium
Overall Summary Table
| Measurand | Mean Value | N | Within-Run | Between-Run | Between-Day | Between-Site | Total |
|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | |||
| WBC | 7.546 | 240 | 0.339 | 4.492 | 0.112 | 1.479 | 0.080 |
| NEUT % | 49.990 | 240 | 3.015 | 6.030 | 0.932 | 1.865 | 0.833 |
High
Page 11
Overall Summary Table
| Measurand | Mean Value | N | Within-Run | Between-Run | Between-Day | Between-Site | Total |
|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | |||
| WBC | 15.246 | 240 | 0.656 | 4.300 | 0.000 | 0.000 | 0.460 |
| NEUT % | 50.823 | 240 | 2.967 | 5.838 | 0.000 | 0.000 | 0.947 |
A site by site analysis was also conducted as per Agency recommendation, also falling within evaluation criteria. Overall reproducibility levels were found to meet the acceptance criteria of 7.5% CV for WBC and 5% SD or 15% CV for Neutrophils across all three levels and all sites.
At-Home Precision Study
An additional precision repeatability (within-run) study was conducted to showcase the precision performance of the Athelas Home specifically by at-home users. The study design recommended by the Agency was for samples to be collected in duplicate by home users in home environments and analyzed for CV% to confirm that evaluation criteria for precision were still met.
Patients were instructed to conduct two tests on the Athelas Home following the labeling process. Patients were instructed to collect both samples within 5 minutes of each other following the Athelas Home labeling, filling two unique test strips. This was known as the in-home whole blood self-test group.
A separate group of patients were provided with Athelas Home Quality Control material and asked to use the dropper on the Quality Control Material vial to conduct 3 replicates for the assigned sample to the given patient. This was known as the in-home quality control material self-testing group.
The evaluation criteria is shown below:
WBC Targets (Evaluation Criteria)
| Range (WBC) | Target | Statistic |
|---|---|---|
| 0 - 2K | 0.25 K/µL | Standard Deviation |
| 2K + | 7.5% | CV % |
Page 12
CV% was calculated for the two groups (patient blood samples, and quality control material) combined as well as independently in order to evaluate against the Evaluation Criteria.
Per-recommendation of CLSI EP05-A3, the primary analyte (WBC) was used for precision evaluation.
The results indicate that the At-Home Precision Repeatability studies met their pre-set evaluation criteria for WBC CV% and NEUT% precision (7.5% CV for WBC and 5% SD for NEUT% or 15% CV for NEUT%).
Patient In-home Whole Blood Studies Self-Test
The in-home whole blood study consisted of 24 patients, and 2 replicates per sample.
| Patient Run Number | Replicate 1 WBC | Replicate 1 NEUT% | Replicate 2 WBC | Replicate 2 NEUT% | WBC Average | CV WBC | NEUT Average | NEUT SD | NEUT% CV |
|---|---|---|---|---|---|---|---|---|---|
| 1 | 5 | 57% | 5.1 | 56% | 5.1 | 1.4% | 57% | 0.71% | 1% |
| 2 | 6.1 | 63% | 6.5 | 66% | 6.3 | 4.5% | 65% | 2.12% | 3% |
| 3 | 7.1 | 59% | 6.1 | 65% | 6.6 | 10.7% | 62% | 4.24% | 7% |
| 4 | 5.5 | 57% | 6 | 57% | 5.8 | 6.1% | 57% | 0.00% | 0% |
| 5 | 1 | 32% | 1.1 | 36% | 1.1 | 6.7% | 34% | 2.83% | 8% |
| 6 | 7.2 | 75% | 7.4 | 71% | 7.3 | 1.9% | 73% | 2.83% | 4% |
| 7 | 1.4 | 31% | 1.5 | 31% | 1.5 | 4.9% | 31% | 0.00% | 0% |
| 8 | 2.6 | 79% | 2.7 | 80% | 2.7 | 2.7% | 80% | 0.71% | 1% |
| 9 | 5.6 | 66% | 5.2 | 63% | 5.4 | 5.2% | 65% | 2.12% | 3% |
| 10 | 5.7 | 63% | 6.2 | 66% | 6 | 5.9% | 65% | 2.12% | 3% |
| 11 | 5.4 | 62% | 5.7 | 59% | 5.6 | 3.8% | 61% | 2.12% | 4% |
| 12 | 6.6 | 61% | 6.1 | 66% | 6.4 | 5.6% | 64% | 3.54% | 6% |
| 13 | 4.7 | 58% | 5.1 | 54% | 4.9 | 5.8% | 56% | 2.83% | 5% |
| 14 | 6.2 | 45% | 5.5 | 44% | 5.9 | 8.5% | 45% | 0.71% | 2% |
| 15 | 4.2 | 27% | 4.3 | 30% | 4.3 | 1.7% | 29% | 2.12% | 7% |
| 16 | 6.1 | 33% | 5.7 | 28% | 5.9 | 4.8% | 31% | 3.54% | 12% |
| 17 | 7.4 | 37% | 7.2 | 37% | 7.3 | 1.9% | 37% | 0.00% | 0% |
| 18 | 3.3 | 31% | 3.2 | 34% | 3.3 | 2.2% | 33% | 2.12% | 7% |
| 19 | 5.1 | 63% | 6.2 | 55% | 5.7 | 13.8% | 59% | 5.66% | 10% |
| 20 | 7.3 | 55% | 7.4 | 59% | 7.4 | 1.0% | 57% | 2.83% | 5% |
| 21 | 5.2 | 45% | 5.8 | 48% | 5.5 | 7.7% | 47% | 2.12% | 5% |
| 22 | 7.1 | 52% | 7.6 | 51% | 7.4 | 4.8% | 52% | 0.71% | 1% |
| 23 | 12.5 | 72% | 13.9 | 68% | 13.2 | 7.5% | 70% | 2.83% | 4% |
Page 13
| 24 | 9.8 | 65% | 10.9 | 60% | 10.4 | 7.5% | 63% | 3.54% | 6% |
Patient in-home QC Material Repeatability Self-Test
The in-home QC material repeatability study consisted of 10 patients, and 3 replicates per sample.
| Replicate 1 WBC | Replicate 1 NEUT% | Replicate 2 WBC | Replicate 2 NEUT% | Replicate 3 WBC | Replicate 3 NEUT% | Average NEUT% | CV WBC | Average NEUT% | STD NEUT% |
|---|---|---|---|---|---|---|---|---|---|
| 2.4 | 8.00% | 2.6 | 11.00% | 2.7 | 9.00% | 2.6 | 6.0% | 9.33% | 1.53% |
| 2.9 | 11.00% | 2.9 | 12.00% | 3.1 | 8.00% | 3 | 3.9% | 10.33% | 2.08% |
| 7.8 | 8.00% | 7.2 | 6.00% | 7.9 | 8.00% | 7.6 | 5.0% | 7.33% | 1.15% |
| 8 | 8.00% | 8.3 | 8.00% | 8.2 | 7.00% | 8.2 | 1.9% | 7.67% | 0.58% |
| 20 | 8.00% | 23 | 9.00% | 22 | 9.00% | 21.7 | 7.1% | 8.67% | 0.58% |
| 23.8 | 9.00% | 23.8 | 6.00% | 24.1 | 9.00% | 23.9 | 0.7% | 8.00% | 1.73% |
| 2.7 | 12.00% | 2.8 | 11.00% | 2.8 | 13.00% | 2.8 | 2.1% | 12.00% | 1.00% |
| 2.9 | 13.00% | 2.9 | 14.00% | 3 | 12.00% | 2.9 | 2.0% | 13.00% | 1.00% |
| 2 | 15.00% | 3 | 18.00% | 3.1 | 17.00% | 2.7 | 22.5% | 16.67% | 1.53% |
| 7.2 | 7.00% | 7.6 | 9.00% | 7.6 | 8.00% | 7.5 | 3.1% | 8.00% | 1.00% |
Results with Evaluation Criteria (see Specifications)
| Value | Eval Criteria | Pass/Fail | |
|---|---|---|---|
| WBC CV% | 5.4% CV | 7.5% CV | Pass |
| NEUT% SD | 1.2% SD | 5% SD or 15% CV | Pass |
Linearity
10 samples across the reporting range were run in 4 replicates across 4 devices and one test strip lot. The samples were obtained by pooling together one low WBC concentration fresh whole blood sample one high WBC concentration sample in different volumes which is a recommended option in CLSI EP6-A.
The method has been demonstrated to be linear from lower limit to upper limit and within measured allowable max % diff for each interval. The primary analyte (WBC) was analyzed in linearity as per CLSI EP06-A guidance. The CLSI recommended OLS regression of the linearity data is provided below:
Page 14
| Parameter | N | R² | Slope | Intercept | CVr |
|---|---|---|---|---|---|
| WBC | 10 | 0.997 | 1.013 | 0.0449 | 5.08% |
Athelas One WBC vs True Concentration
[Chart showing linear relationship with equation y = 1.01x + 0.0449 R² = 0.997, Pearson's r = 0.996]
Interfering Substances
Interference studies were performed taking sample abnormalities, drugs, metabolites, sample additives and dietary substances into consideration. A list of substances were tested and found naturally occurring in patient samples, or were spiked in whole blood. This list is included below (with highest concentration or percentages tested). None of the substances were found to interfere with the measurements of the Athelas One device. This was evaluated through analysis of interferon impact on bias and precision performance. All interferon spiked samples continued to perform within the evaluation criteria.
- Triglyceride Rich Lipoproteins
- Hemolysate
- Protein
- Levodopa
- Methyldopa
- Metronidazole
- Acetylsalicylic Acid
- Phenylbutazone
- Rifampicin
Page 15
- Cyclosporine
- Acetaminophen
- Heparin
- Ibuprofin
- Bilirubin C
- Bilirubin F
- Reticulocytes
- Platelet Aggregates
- Nucleated Red Blood Cells
- Giant Platelets
- Myelodysplastic Syndrome
- Macrocytic Anemia
- Infectious Mononucleosis
- Bacterial Infections
- Chronic Myeloid Leukemia
- Patients on Chemotherapy
- Patients treated with GCSFs
Reference Intervals
Adult reference intervals for the Athelas One were verified across 120 healthy donors in comparison to pre-established reference intervals for the Sysmex XE-5000 utilizing whole blood samples.
Limit of Blank (LoB) – Whole Blood
The LoB was obtained from 120 total repeated measurements of blank samples analyzed in the sampling modes for whole blood. The mean, SD and LoB were calculated for the analyzer. The LoB for whole blood parameters WBC is zero (0) K/μL.
Limit of Detection (LoD) – Whole Blood
Limit of Detection was conducted using run as follows:
- 2 test strip lots (60 strips each)
- 2 instruments
- 3 days
- 5 low level samples
- 2 replicates per sample
Test Strip Lot 1 Summarized Data
| N | LoB (from LoB section) | cp | Pooled Standard Deviation (K/µL) | LoD (K/µL) |
|---|---|---|---|---|
| 60 | 0 | 1.653 | 0.033 | 0.055 |
Page 16
Test Strip Lot 2 Summarized Data
| N | LoB (from LoB section) | cp | Pooled Standard Deviation (K/µL) | LoD (K/µL) |
|---|---|---|---|---|
| 60 | 0 | 1.653 | 0.048 | 0.079 |
As shown in the Limit of Blank section, the LoB was found to be 0.
The LoD of Lot 1 was 0.055 K/µL while the LoD of Lot 2 was 0.079 K/µL. The maximum LoD of the two lots was established as the final LoD. Hence, the LoD was set to be 0.079 K/µL.
The LoD was found to be 0.079 K/µL which is well below the LoQ of 0.44 K WBC/µL.
Limit of Quantification (LoQ) – Whole Blood
The Limit of Quantification study was designed to estimate the lowest measurement concentration that can be measured on the Athelas One and meet the predefined total error goals. The study was conducted over two test-strip lots, 1 instrument, 3 days, 3 replicates per sample, and 4 independent low-level whole blood samples. The samples were distinct whole blood samples acquired from patients in K2EDTA tubes. The original concentrations were obtained from the predicate analyzer (Sysmex XE-5000). These samples were then diluted to reach an approximate concentration of 0.45 K WBC/µL. The original concentrations, the volume of diluent, and the final concentrations of the samples are shown in the table below.
| Sample | WBC Initial (K/µL) | Volume of Sample in Aliquot (µL) | Volume of Diluent in Aliquot (µL) | Final Aliquot Concentration (K/µL) | Final Aliquot Volume (µL) |
|---|---|---|---|---|---|
| 1 | 3.97 | 100 | 800 | 0.44 | 900 |
| 2 | 3.92 | 100 | 800 | 0.44 | 900 |
| 3 | 4.16 | 100 | 800 | 0.46 | 900 |
| 4 | 3.13 | 100 | 600 | 0.45 | 700 |
The Limit of Quantification of the Athelas One was determined to be 0.44 K WBC/µL as per CLSI EP17-A2. This is below the lower limit of the Athelas One's reporting range of 1.0K WBC/µL.
Page 17
Specimen Stability Studies
Whole blood stability - Room temperature vs Refrigerated temperature
Whole blood stability was evaluated on the Athelas One device by conducting a 48 hour stability study on nine different venous blood samples with low (0.5-3 K/µL, normal (4-10 K/µL), and high (>10 K/µL) WBC levels. The study was designed in accordance with recommendations from CLSI EP25-A.
Based on results evaluating deviation from pre-defined bias and precision criteria, the stability duration for all samples can be set to 24 hours with observed degradation at 48 hours for both the WBC and Neutrophil % parameters. Overall, all samples tested within 24 hours of collection time were found to meet the pre-set acceptance criteria of 7.5% CV for WBC precision, ± 7.5% Bias for WBC, 5% SD OR 15% CV for Neutrophils (updated based on Sysmex XW-100 Neut% CV criteria), and ± 10% Bias across the whole study.
Test Strip Stability over time
A 26 month test strip stability study was conducted to assess the shelf life stability of the Athelas test strips. The strips were tested every 60 days over 26 months to justify a 24 month expiration. The study was designed in accordance with recommendations from CLSI EP25-A.
In particular, the study was conducted over 26 months, across two test strip lots, 4 analyzers, three replicates, and multiple lots of quality control materials. Every 60 days, three replicates per two lots of strips across 3 levels of control fluid were run.
From the study it can be observed that none of the analyzed test strips manifested statistically significant degradation/drift over the 26 month study (p > 0.05) as recommended by the CLSI EP25-A regression slope analysis. The stability duration for all samples can be set to 24 months with no observed degradation at 26 months, the last time point of the study.
Clinical
Method Comparison
Method comparison studies were performed to assess the performance of the Athelas One analyzer when compared to the predicate Sysmex XE-5000 analyzer on a total of 312 patient samples taken at 3 point of care sites in the US. All samples were run in the Automated Sampling Modes in singlet on the XE-5000.
Samples covered clinical medical decision levels, and included normal and abnormal samples. The full reportable measuring ranges of the Athelas One analyzer were tested in method comparison. All evaluation criteria (slope, r, bias at medical decision levels, overall bias, bland altman analysis, intercept) were met, and even more specifically:
Page 18
manufacture-defined criteria for equivalence between two predicate Sysmex XE-5000 devices was also met by the Athelas One.
The overall results across three sites are shown in the table below.
Method Comparison Combined Site WBC and Neutrophil Results
| Slope | Intercept | |||||||
|---|---|---|---|---|---|---|---|---|
| Parameter | Interval | N | (r) | Estimate | LCI (2.5%) | UCI (97.5%) | Estimate | LCI (2.5%) |
| WBC | 1.1 - 23 (K/µL) | 312 | 0.99 | 0.978 | 0.958 | 1.000 | -0.042 | -0.159 |
| Neutrophil % | 8 - 92.89 (%) | 312 | 0.96 | 0.980 | 0.953 | 1.003 | 1.855 | -0.530 |
Flagging Comparison
A Flagging Comparison study was conducted to determine the ability for the Athelas One to successfully flag abnormal samples for manual review. The study design of a Sysmex XN series analyzer (FDA 510k K112605) was used, and a similar evaluation criteria of 90% flagging accuracy.
Morphological Flags
The results of the Athelas One morphological flagging compared to the XE-5000 flagging evaluation were divided into two categories: (1) No Flags, Negative Judgement (2) Patients with positive Morphology/Differential abnormalities - Flags present, Positive Judgement. The study was limited to WBC and DIFF related flags. The results obtained from the flagging comparison study met the specification of ≥ 90%.
| Sysmex XE - 5000 | |||
|---|---|---|---|
| Positive (Abnormal) | Negative (Normal) | ||
| Athelas One | Positive (Abnormal) | 90 | 7 |
| Negative (Normal) | 9 | 206 | |
| Total | 99 | 213 |
% Positive Agreement (Sensitivity) = 100 * 90/(90 + 9) = 90.91%
% Negative Agreement (Specificity) = 100 * 206/(206 + 7) = 96.71%
% Overall Agreement = 100 * (90 + 206) / (90 + 206 + 9 + 7) = 94.87%
Page 19
Evaluation criteria for morphological flagging accuracy was met, and a similar analysis was conducted for distributional flagging as well. The flagging study covered a variety of abnormal samples that are flagged by the predicate, with a breakdown as follows:
| Predicate Flag | Count |
|---|---|
| PLT Clumps? | 15 |
| Abn Lympho/L_Blasts? | 51 |
| Blasts? | 46 |
| Immature Gran? | 29 |
| Left Shift? | 14 |
Matrix Comparison
Experiments were conducted to determine the comparability of venous and capillary whole blood on the Athelas One device across 59 patients. These were in addition to the method comparison trials that used a combination of venous and capillary blood tested against venous samples run on the predicate device.
Capillary finger-prick blood samples were placed onto Athelas test strips from subjects. Venous whole blood samples were collected in K2EDTA vacutainer tubes for the same patients. The capillary whole blood and the venous whole blood samples were analyzed on Athelas One devices. It was concluded that there is no statistical difference between the systems when analyzing venous and capillary samples.
The regression parameters and 95% confidence interval of bias at the medical decision levels were calculated as per recommendation of CLSI EP5-A3 and H26-A2 and met all evaluation criteria. The information is summarized on the tables below:
| Parameter | Interval | N | (r) | Slope, 95% CI | Intercept | Mean Bias | Mean % Bias |
|---|---|---|---|---|---|---|---|
| WBC | 1.33 - 23.38 | 59 | 0.996 | 1.026 (1.000, 1.055) | -0.145 (-0.319, 0.018) | 0.056 (K/µL) | -0.588 |
| Neutrophil % | 5.8 - 74.6 | 59 | 0.969 | 0.999 (0.938, 1.058) | 0.457 (-2.502, 3.614) | 0.162 (Percentage Points) | -0.333 |
Page 20
WBC Passing-Bablok Regression.
[Chart showing WBC correlation with Pearson's r = 0.996]
Neutrophil % Passing-Bablok Regression.
[Chart showing Neutrophil % correlation with Pearson's r = 0.969]
Page 21
At-Home Method Comparison
Method comparison studies were performed to assess the performance of the Athelas Home analyzer in at-home settings when compared to the predicate Athelas One analyzer on a total of 165 samples taken in at-home settings in the US.
The full reportable measuring ranges of the Athelas Home analyzer were tested in method comparison. All evaluation criteria (slope, r, bias at medical decision levels, overall bias, bland altman analysis, intercept) were met, and even more specifically: manufacture-defined criteria for equivalence between two predicate Sysmex XE-5000 devices was also met by the Athelas Home in home settings and the Athelas One.
The overall results are shown in the table below.
Method Comparison WBC and Neutrophil Results
| Slope | Intercept | Mean Bias (K/µL) | Mean % Bias | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Parameter | Interval | N | (r) | Estimate | LCI (2.5%) | UCI (97.5%) | Estimate | LCI (2.5%) | UCI (97.5%) | ||
| WBC | 1.0 - 16.9 | 165 | 0.97 | 1.0 | 0.95 | 1.03 | 0 | -0.16 | 0.31 | 0 | 0.0 |
| Neutrophil % | 21-78 | 165 | 0.92 | 1.0 | 0.98 | 1.09 | -1.0 | -6.09 | 0.0 | -0.5 | 1.3 |
WBC Athelas One v. Athelas Home
[Chart showing correlation with Pearson's r = 0.968]
Page 22
Neut% Athelas One v Athelas Home
[Chart showing correlation with Pearson's r = 0.924]
At-Home Error Analysis
During At-Home testing, error analysis was conducted to verify the usability and ease of use of the Athelas Home when used by patients in the intended use settings. Over 94% of tests were run successfully with no errors. Errors that were found were resolved after running a single additional test-strip. The results are shown below:
| Error Type | Number of Occurrences | Percent of Tests |
|---|---|---|
| No Error | 76 | 94 |
| Test-Strip/Optical Error | 3 | 3.7 |
| Incorrect Insertion Direction | 2 | 2.5 |
| HH! (shows as success to the patient) | 1 | 1.2 |
| LL! (shows as success to the patient) | 1 | 1.2 |
At-Home Usability Analysis
Page 23
During the first At-Home visits, study investigators were asked to conduct usability surveys to ensure at-home users had no difficulties in setting up and understanding how to run the test. The following questions were asked per user:
- Was the user able to use the lancet correctly? Mark yes if the user was able to open the lancet, puncture the correct site on their finger, and dispose of the lancet correctly.
- Did the user fill up the test-strip fully? Mark yes if the user placed enough blood such that the blood reached the end of the channel on the test-strip.
- Did the user face any errors when running the test (not error codes)?
- Did the user correctly know how to run the test using the application?
- Was this test performed by the patient (yes) or caregiver (no)? ('Yes' if conducted by patient, 'No' if conducted by caregiver)
The results, shown below, indicate a large majority of patients found the Athelas Home easy to use and had very little difficulty running the test end to end.
| Survey Question | "Yes" Responses | "No" Responses |
|---|---|---|
| 1 | 32 | 0 |
| 2 | 30 | 2 |
| 3 | 31 | 1 |
| 4 | 32 | 0 |
| 5 | 10 | 22 |
Flex Testing
Fill-Volume Testing
A study was conducted to determine whether the Athelas Home correctly handles variable amounts of blood inserted into the test-strip. In particular, the study aimed to verify that should an at-home user place either too little or too much blood in the test-strip inlet, one of the following occurs:
- An error code is returned with the appropriate message.
- The results are accurate and meet the acceptance criteria.
The study found that the Athelas Home is robust to variable volumes of blood placed on the test-strip. The results for multiple fill volumes tested across different samples are shown below:
| Sample | Fill-Volume (µL) | Mean WBC (K/µL) | Mean Neut (%) | Predicate WBC (K/µL) | Predicate Neut (%) | WBC Bias (%) | Neut% Bias (percentage points) | WBC bias criteria met | Neut % bias criteria met |
Page 24
| 1 | 0 | error | error | 2.7 | 42 | N/A | N/A | Yes | Yes |
| 1 | 1.4 | 2.7 | 39 | 2.7 | 42 | 0.00% | -3 | Yes | Yes |
| 1 | 3.4 | 2.8 | 43 | 2.7 | 42 | 3.70% | 1 | Yes | Yes |
| 1 | 5.4 | 2.7 | 37.5 | 2.7 | 42 | 0.00% | -4.5 | Yes | Yes |
| 2 | 0 | error | error | 8.0 | 48 | N/A | N/A | Yes | Yes |
| 2 | 1.4 | 7.6 | 49.5 | 8.0 | 48 | -5.00% | 1.5 | Yes | Yes |
| 2 | 3.4 | 8.2 | 46 | 8.0 | 48 | 2.50% | -2 | Yes | Yes |
| 2 | 5.4 | 7.7 | 52.5 | 8.0 | 48 | -3.75% | 4.5 | Yes | Yes |
| 3 | 0 | error | error | 18.2 | 50 | N/A | N/A | Yes | Yes |
| 3 | 1.4 | 17.7 | 48 | 18.2 | 50 | -2.75% | -2 | Yes | Yes |
| 3 | 3.4 | 18.4 | 49.5 | 18.2 | 50 | 1.10% | -0.5 | Yes | Yes |
| 3 | 5.4 | 17.6 | 45.5 | 18.2 | 50 | -3.30% | -4.5 | Yes | Yes |
Device Tilt Testing
A study was conducted to determine whether the Athelas Home is robust against being placed on tilted surfaces in different orientations. In particular, the study aimed to verify that should an at-home user place place the device on some tilted surface and runs a test, one of the following occurs:
- An error code is returned with the appropriate message.
- OR The results are accurate and meet the acceptance criteria.
The study found that the Athelas Home is not affected by tilted placement of the device. The results indicating no significant bias in both parameters across tilt orientations and samples is shown below:
| Sample | Condition | Average WBC (K/mL) | Average Neut (%) | Predicate WBC (K/µL) | Predicate Neut (%) | WBC Bias (%) | Neut% Bias (percentage points) | WBC Criteria Met | Neut% Criteria Met |
|---|---|---|---|---|---|---|---|---|---|
| 1 | 1 | 3.0 | 40 | 2.9 | 42 | 1.72% | -2 | Yes | Yes |
| 1 | 2 | 3.1 | 42 | 2.9 | 42 | 6.90% | 0 | Yes | Yes |
| 1 | 3 | 2.9 | 41.5 | 2.9 | 42 | -1.72% | -0.5 | Yes | Yes |
| 1 | 4 | 3.1 | 40.5 | 2.9 | 42 | 6.90% | -1.5 | Yes | Yes |
| 2 | 1 | 6.8 | 46 | 7.2 | 47 | -5.56% | -1 | Yes | Yes |
| 2 | 2 | 7.2 | 49 | 7.2 | 47 | 0.21% | 2 | Yes | Yes |
Page 25
| 2 | 3 | 6.9 | 45 | 7.2 | 47 | -4.86% | -2 | Yes | Yes |
| 2 | 4 | 7.5 | 46 | 7.2 | 47 | 3.47% | -1 | Yes | Yes |
| 3 | 1 | 17.0 | 47 | 16.3 | 50 | 3.99% | -3 | Yes | Yes |
| 3 | 2 | 16.1 | 49 | 16.3 | 50 | -1.07% | -1 | Yes | Yes |
| 3 | 3 | 16.0 | 49.5 | 16.3 | 50 | -1.60% | -0.5 | Yes | Yes |
| 3 | 4 | 16.5 | 50.5 | 16.3 | 50 | 0.92% | 0.5 | Yes | Yes |
Device Drop Testing
A study was conducted to determine whether the Athelas Home continues to return accurate results after being dropped from multiple orientations and heights. In particular, the study aimed to verify that should an at-home user drop the device, the following occurs:
- An error code is returned with the appropriate message.
- The results are accurate and meet the acceptance criteria.
The study was conducted in accordance with ASTM 1469 - D03 across 12 devices.
The study found that the Athelas Home is robust to the given drop and vibration schedules. The table below shows that all devices continued to pass functional tests after being subjected to the full testing schedules:
| Device | Critical Physical Damage Present? | No functional defects? | Auto-Check WBC (K/µL) | Auto-Check Neut% | Reference WBC Range (K/µL) | Reference Neut% Range (%) | WBC in range? | Neut% In range? |
|---|---|---|---|---|---|---|---|---|
| 1 | No | Yes | 3.4 | 41 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
| 2 | No | Yes | 3.1 | 38 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
| 3 | No | Yes | 3.3 | 42 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
| 4 | No | Yes | 3.3 | 41 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
| 5 | No | Yes | 3.3 | 41 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
| 6 | No | Yes | 3.3 | 39 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
| 7 | No | Yes | 3.3 | 41 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
| 8 | No | Yes | 2.9 | 42 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
| 9 | No | Yes | 3.0 | 40 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
Page 26
| 10 | No | Yes | 3.1 | 45 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
| 11 | No | Yes | 3.0 | 42 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
| 12 | No | Yes | 3.1 | 38 | 2.9 - 3.4 | 35 - 45 | Yes | Yes |
Device and Test-Strip Temperature Testing
A study was conducted to verify the performance of the Athelas Home device and test-strips after extreme and varying temperature. The goal of the study was to confirm that the Athelas Home retains performance characteristics after prolonged exposure to such conditions. Flex temperatures may arise in home-use environments.
Both the Athelas Home device and Athelas Home test-strip were found to continue to meet performance specifications after exposure to extreme temperatures. The results are shown below.
Device Results
| Environment Condition | Device | N | Average WBC Bias | Average Neut% Bias | WBC Meets Criteria | Neut% Meets Criteria |
|---|---|---|---|---|---|---|
| 1 | 1 | 9 | 1.26% | 1.70 | Yes | Yes |
| 1 | 2 | 9 | -0.33% | 1.48 | Yes | Yes |
| 1 | 3 | 9 | 2.34% | 2.59 | Yes | Yes |
| 2 | 1 | 9 | 0.93% | -1.52 | Yes | Yes |
| 2 | 2 | 9 | 0.49% | -0.97 | Yes | Yes |
| 2 | 3 | 9 | 6.27% | 0.81 | Yes | Yes |
| 3 | 1 | 9 | 6.15% | 0.37 | Yes | Yes |
| 3 | 2 | 9 | 3.94% | -1.74 | Yes | Yes |
| 3 | 3 | 9 | 0.37% | -1.97 | Yes | Yes |
Test-Strip Results
| Environment Condition | Test-Strip Lot | N | Average WBC Bias | Average Neut% Bias | WBC Meets Criteria | Neut% Meets Criteria |
|---|---|---|---|---|---|---|
| 1 | 1 | 9 | 3.01% | 1.26 | Yes | Yes |
| 1 | 2 | 9 | 3.08% | 2.14 | Yes | Yes |
| 1 | 3 | 9 | -2.69% | 1.81 | Yes | Yes |
| 2 | 1 | 9 | 0.46% | 0.59 | Yes | Yes |
| 2 | 2 | 9 | -0.99% | 0.14 | Yes | Yes |
Page 27
| 2 | 3 | 9 | -3.78% | 0.70 | Yes | Yes |
| 3 | 1 | 9 | 6.01% | -2.52 | Yes | Yes |
| 3 | 2 | 9 | 5.58% | -1.08 | Yes | Yes |
| 3 | 3 | 9 | -0.97% | -0.08 | Yes | Yes |
Vibration Testing
Vibration testing conducted to ensure that the Athelas Home device can continue to function or return appropriate error codes after exposure to extensive vibration in handling by a patient or shipping.
Vibration testing was conducted by third party ISTA certified test lab Quanta Testing. The system successfully met evaluation criteria post vibration test.
Humidity Flex Testing
Objective
Humidity Flex Testing was part of the final list of flex tests decided by Athelas and the Agency as part of the Athelas Home Presubmission (Q190305 Attachment B). A study was conducted to verify the performance of the Athelas Home test-strips under extreme humidity conditions.
Methodology
The study aimed to showcase functional performance of the Athelas Home Test-Strips after exposure of test-strips to the following humidity conditions:
| Condition | Temperature (C) | Humidity (%) |
|---|---|---|
| X | 38º | 85 |
| Y | -29 | Uncontrolled |
Each of eight test-strips were subject to the following test schedules based on the above conditions:
| Test-Strip Number | Schedules |
|---|---|
| 1 | Condition X for 72 hrs |
| 2 | Condition X for 72 hrs |
| 3 | Condition X for 72 hrs |
Page 28
| 4 | Condition X for 72 hrs |
| 5 | Condition X for 72 hrs followed by Y for 72 hrs |
| 6 | Condition X for 72 hrs followed by Y for 72 hrs |
| 7 | Condition X for 72 hrs followed by Y for 72 hrs |
| 8 | Condition X for 72 hrs followed by Y for 72 hrs |
The conditions were implemented using humidity chambers at Quanta Laboratories, an ISTA certified environmental testing laboratory.
After exposure to the testing schedules, the following functional tests were conducted on the test-strips:
-
Change in Color
a. Verify qualitatively that no change in test-strip color is present.
i. This was conducted by comparing a reference test-strip that hadn't been placed into the testing conditions against the test-strip placed under the conditions. -
Fill-Time
a. Measurement of time for 3.4 µL of a single vial of Ath-Check control fluid to flow to the end of the channel.
b. A stopwatch was started when a single drop of the fluid was placed on the channel inlet and stopped when the fluid met the side edges of the T channel on the test-strip.
These tests were chosen due to risk analyses which found that humidity primarily affects the test-strips by increasing fill-time through decreasing capillary pressure in the channel, and changing the test-strip color due to presence of condensation affecting stain color.
As such, verification of flow-time and color change were found to be sufficient checks for no performance degradation of the test-strip.
Results
The below table showcases the results of humidity related environment testing on the eight test-strips:
Page 29
| Test-Strip Number | Schedules | Fill Time (seconds) | Fill Time Criteria (seconds) | Fill Time Within Specification? | No Color Change present? |
|---|---|---|---|---|---|
| 1 | X for 72 hrs | 7.32 | < 25 | Yes | Yes |
| 2 | X for 72 hrs | 7.72 | < 25 | Yes | Yes |
| 3 | X for 72 hrs | 7.70 | < 25 | Yes | Yes |
| 4 | X for 72 hrs | 7.58 | < 25 | Yes | Yes |
| 5 | X for 72 hrs followed by Y for 72 hrs | 8.16 | < 25 | Yes | Yes |
| 6 | X for 72 hrs followed by Y for 72 hrs | 7.66 | < 25 | Yes | Yes |
| 7 | X for 72 hrs followed by Y for 72 hrs | 7.21 | < 25 | Yes | Yes |
| 8 | X for 72 hrs followed by Y for 72 hrs | 7.08 | < 25 | Yes | Yes |
Discussion
Based on the results, it was concluded that the Athelas Home test-strips continued to function within specifications after exposure to extreme humidity conditions. In particular, no evidence was found of condensation in the test-strip which would be manifested in increased fill-time and changes in stain color.
All eight test-strips across all schedules were found to be within specifications for both fill time and color. This is due to specially designed packaging for the Athelas Home test-strips that ensure humidity resistance. The Athelas Home Test-Strip packaging includes three separate layers designed specifically to reduce humidity effects:
- PET Layer
- Aluminum Layer
- Desiccant Layer
As such, the relative humidity within the package is consistently less than 10% in extreme conditions and ensures the test-strip is protected from humidity related effects.
Page 30
Cleaning Study
In point-of-care settings, the device is intended for multiple-patient use. Robustness studies were performed by the sponsor demonstrating that there was no change in performance or in the external materials of the device after 10,000 cleaning and disinfection cycles using the chosen disinfectant, Caviwipe (EPA registration # 46781-8). The robustness studies were designed to simulate cleaning and disinfection over the 2-year multiple-patient use life of the device. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
Conclusion
Through a range of CLSI recommended studies in clinical and bench settings, the Athelas Home has been shown to be safe and effective for at-home use and POC settings.
§ 864.5220 Automated differential cell counter.
(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”