K Number
K241806

Validate with FDA (Live)

Date Cleared
2025-01-08

(201 days)

Product Code
Regulation Number
866.3981
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) in vitro diagnostic test for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza B virus, and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections.

Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infective of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.

The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organism(s) detected by the Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections.

The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is intended for use by qualified and trained clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.

Device Description

The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) test. The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, influenza B, respiratory syncytial virus (RSV) A/B and RNase P primer and probe sets are designed to detect viral RNA in nasopharyngeal (NP) and anterior nasal (AN) swab specimens from individuals exhibiting signs and symptoms of a respiratory tract infection.

Each TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel includes the following components:

  • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Assay-Multiplex assays that contain . primer and probe sets specific to the following targets:
    • Three SARS-CoV-2 targets (Orfla, Orf1b, and N genes) .
    • . Two influenza A virus targets (PB1 & M genes)
      • I Two influenza B virus targets (M & NS genes)
      • . Three RSV targets (NP, M, and L protein genes)
      • l RNase P (internal human sample collection control)
  • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Positive Control—Inactivated viral . control that contains SARS-CoV-2, influenza A, influenza B, and RSV.
  • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Negative Control—MS2 packaged RNA . control that contains targets specific to RNase P genomic regions targeted by the assay.
  • TaqPath™ 1-Step Select Master Mix (No ROX)—Ready-to-use PCR mix, including . reverse transcriptase, polymerase, dNTPs, salts, and buffer.
  • . Package Insert -- Provides the instructions and the link to download the instructions for use and other assets (including the ADF)
  • An Assay Definition File (ADF) applicable to the instrument used in the workflow . (available via download).

In addition to the SARS-CoV-2, influenza A, influenza B and RSV viral assay targets, the assay portion of the panel includes RNase P, which serves as an endogenous internal process control to monitor extraction and amplification of each clinical sample. The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel also contains external process positive and negative controls. The positive control (PC) component included is an inactivated viral control that contains SARS-CoV-2, influenza A, influenza B, and RSV viruses. The PC monitors extraction and real-time RT-PCR by demonstrating that each of the four viruses can be detected when present and that RNase P is not detected when absent. The negative control (NC) component included is an MS2 packaged RNA control that contains targets specific to RNase P genomic regions targeted by the assay. The NC also monitors extraction and real-time RT-PCR by demonstrating RNase P can be detected when present and that the four viruses are not detected when absent. The TaqPath™ 1-Step Select Master Mix (No ROX) included as a component of the kit is a ready-to-use PCR mix which contains a deoxyribonucleotide triphosphate mix (dNTPs), enzymes, and other components to permit reverse transcription and amplification of the assay targets. The TagPath™ 1-Step Select Master Mix (No ROX) also contains ribonuclease (RNase) inhibitors as well as deoxyuridine triphosphate (dUTP) and uracil N-glycosylase (known as UNG or UDG).

The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is provided in two overall kit configurations: either 200 reactions or 1.000 reactions.

AI/ML Overview

The provided text describes the analytical and clinical performance studies of the "Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel" (hereafter referred to as "the Device"). This device is an in vitro diagnostic test for the qualitative detection and differentiation of SARS-CoV-2, influenza A, influenza B, and respiratory syncytial virus (RSV).

Below is a breakdown of the acceptance criteria and study details based on the provided information, presented in the requested format.


Acceptance Criteria and Device Performance

The acceptance criteria for this diagnostic device are primarily demonstrated through various analytical performance evaluations (Limit of Detection, Inclusivity/Reactivity, Precision, Interfering Substances, Cross-Reactivity, Stability, Carryover) and clinical performance studies (PPA and NPA against a comparator device). The studies aim to show that the device performs as expected and is "substantially equivalent" to a legally marketed predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

Study Aspect / Performance MetricAcceptance Criteria (Implied/Directly Stated where available)Reported Device Performance (Summary)
Limit of Detection (LoD)≥ 95% positive detection at the determined LoD concentrations.All viruses confirmed at ≥ 95% positive detection at the determined LoD concentrations (e.g., SARS-CoV-2 at 107 GCE/mL; Flu A H1N1 at 340 GCE/mL). WHO Standard LoD for SARS-CoV-2 confirmed at 150.36 IU/mL.
Inclusivity (in silico)Less than 1% (SARS-CoV-2) and less than 5% (Flu A, Flu B, and RSV) of publicly available sequences have mutations that reduce melting temperature below annealing temperature for all gene targets.> 99.99% of SARS-CoV-2, 99.9% of Flu B, and 99.49% of RSV strains met the criteria. For Flu A, 95.27% of strains were considered reactive after wet testing.
Reactivity (wet-lab)100% detection at approximately 3x LoD.Most tested strains showed 100% detection at approximately 3x LoD. Strains not at 100% initially were further tested until 100% detection was achieved at higher concentrations.
Precision (Within-Laboratory)N/A (Statistical measures like %CV and %PPA are reported).Positivity Rate (Negative Sample): 0%.
Overall PPA (1x LoD, 3x LoD): 100.00% for all targets.
Observed Precision and Repeatability %CV (1x and 3x LoD): < 2%CV for all targets.
Site-to-Site ReproducibilityQualitative PPA performance: Generally high (e.g., above 90-95% at 1x LoD, 100% at 3x LoD). Quantitative %CV for Cq variability: Generally low (e.g., usually < 5%).NPA (Negative Sample): 99.63%.
Overall Site-to-Site Reproducibility PPA (1x LoD): 92.59% (SARS-CoV-2) to 98.89% (Flu A).
Overall Reproducibility PPA (3x LoD): 100.00% for all targets.
Observed Reproducibility, Within-lab precision, Repeatability %CV (1x & 3x LoD): < 5% CV for all targets.
Interfering SubstancesAll positive samples remain positive; negative samples remain negative. No false positives (FP) or false negatives (FN).All conditions produced expected results. No interference observed from 26 substances, 2 solvents, 5 VTMs, or extraction reagent carryover, except for expected positive results from FluMist® due to live influenza components.
Competitive Interference100% positivity for both low-concentration and high-concentration viruses in multi-target samples. No off-target false positives.All conditions satisfied acceptance criteria, showing 100% positivity for both viruses in pairs and no off-target false positives.
Cross-Reactivity & Microbial InterferenceNo false positive results from tested microbes in negative samples. No inhibition of PCR or interference with extraction/RNase P detection in positive samples with microbes.None of the 49 tested microorganisms produced false positive results or interfered with detection. One of 39 CoV-HKU1 clinical samples produced a positive SARS-CoV-2 result, the rest were negative.
Specimen Stability100% positivity for all targets at specified storage conditions and time points.100% positivity for NP and AN specimens at all time points tested (RT for 24h, refrigerated for 72h, ≤ -70°C for 30 days).
Eluate Stability100% positivity for contrived positive sample eluates at specified conditions.100% of contrived positive sample eluates produced expected results for up to 24 hours at 2-8°C, 5 days at ≤ -70°C, and after one freeze/thaw cycle.
Fresh vs Frozen EquivalencySupport the use of frozen samples in clinical studies and stated freeze/thaw cycles in IFU.All samples showed 100% concordance with expected results, supporting use of frozen samples and specific F/T cycles (up to 2 F/T for Copan/BD, 1 F/T for others).
In-Use and Hold Time StabilityAll components meet specified stability criteria for storage conditions and freeze/thaw cycles.Assay Mix Tube and Master Mix stable for 72h at 2-8°C and up to 4 F/T cycles. PC and NC stable for 24h at 2-8°C.
Carryover Cross-ContaminationCarryover/cross-contamination rate of 0%.0% carryover/cross-contamination rate reported among 492 replicates tested.
RNase P Internal Control Cutoff ConfirmationOverall PPA of 98.07% (NPA 100%) for NP swabs and 92.30% (NPA 100%) for AN swabs. Improved AN swab collection technique reduces invalid rate and shows 100% RNase P detection.NP PPA: 98.07%, NPA: 100%. AN PPA: 92.30%, NPA: 100%. Subsequent study with improved AN collection showed 100% RNase P detection and 100% validity for AN samples.
Matrix Equivalency100% concordance between NP and AN swab specimens.All samples produced 100% concordance with expected results.
Clinical Performance (Prospective Study)Demonstrate substantial equivalence based on PPA and NPA compared to a comparator device. Target typically high PPA/NPA (e.g., >85-90%).SARS-CoV-2: NP Overall PPA 96.6%, NPA 96.47%; AN Overall PPA 95.3%, NPA 96.46%.
Flu A: NP Overall PPA 94.9%, NPA 99.36%; AN Overall PPA 97.8%, NPA 99.06%.
Flu B: NP Overall PPA 98.0%, NPA 99.87%; AN Overall PPA 100.0%, NPA 99.87%.
RSV: NP Overall PPA 93.8%, NPA 99.75%; AN Overall PPA 100.0%, NPA 99.6%.
Clinical Performance (Enrichment Study)Maintain high PPA/NPA for Flu A, Flu B, RSV (where prospective numbers might be low).Flu A: NP Overall PPA 97.9%, NPA 100.0%; AN Overall PPA 100.0%, NPA 95.5%.
Flu B: NP Overall PPA 92.3%, NPA 98.2%; AN Overall PPA 92.3%, NPA 100.0%.
RSV: NP Overall PPA 100.0%, NPA 94.9%; AN Overall PPA 100.0%, NPA 96.6%.

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Performance Study (Prospective Cohort):

    • Subjects: 1,909 subjects enrolled (1,840 prospectively, 69 enrichment).
    • Test Set (Evaluable Specimens): 1,620 Nasopharyngeal (NP) swabs and 1,541 Anterior Nasal (AN) swabs were included in the final data analysis for the prospective cohort. (Minor adjustments for inconclusive comparator results in Flu A, specifically 1,539 AN and 1,616 NP for Flu A analysis).
    • Enrichment Study: 69 subjects, yielding 69 NP swabs and 68 AN swabs for analysis.
    • Data Provenance:
      • Country of Origin: Not explicitly stated as a single country, but "14 different collection sites across the U.S." indicate United States data.
      • Retrospective or Prospective: Primarily prospective data, with an "enrichment phase" also involving prospective collection from subjects with confirmed positive PCR results for Flu A, Flu B, and/or RSV.
      • Specimen Handling: Specimens were tested either fresh (Category I) or frozen (Category II).
  • Analytical Studies (Test Sets):

    • LoD Confirmation: At least 24 replicates per virus strain/specimen type.
    • Precision (within-lab): 96 replicates for each panel member.
    • Reproducibility (site-to-site): 9 replicates per panel member (3 sites x 3 reagent lots x 1 replicate/lot). Total ~270 replicates per analyte level across 3 sites.
    • Interfering Substances: 3 replicates per substance/condition.
    • Competitive Interference: 3 replicates per condition.
    • Cross-Reactivity & Microbial Interference: 3 replicates per organism/condition.
    • Specimen Stability: Multiple replicates per time point/storage condition.
    • Eluate Stability: Multiple replicates per time point/storage condition.
    • Fresh vs Frozen Equivalency: Multiple replicates per F/T cycle and concentration.
    • In-Use and Hold Time Stability: Multiple replicates per time point/F/T cycle.
    • Carryover Cross-Contamination: 492 replicates (negative wells).
    • RNase P Internal Control Cutoff Confirmation: Individually collected NP and AN swabs (specific number not given for the initial confirmation, but the follow-up AN study used 138 AN samples).
    • Matrix Equivalency: Multiple replicates per matrix and concentration.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • For the clinical performance study, the ground truth appears to be established by a comparator FDA-cleared molecular test.
  • The document does not specify the number or qualifications of human experts (e.g., radiologists, pathologists) involved in establishing the ground truth for any of the studies. This is typical for molecular diagnostic assays where the "ground truth" is often defined by the results of a highly sensitive and specific reference method (e.g., another molecular test or a composite reference method incorporating multiple tests or clinical findings).

4. Adjudication Method for the Test Set

  • For the clinical performance study, samples that produced a discordant call between the Device and the comparator test were further tested with "another FDA cleared molecular assay" (a resolver assay). This implies a reconciliation method where a third test acts as an adjudicator for discordant results between the device under evaluation and the primary comparator.
  • The specific rules for how the resolver assay's result determined the final ground truth (e.g., 2+1 majority rule, or if the resolver result was the final truth) are not explicitly detailed but are common in diagnostic test evaluation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done. The evaluated device is a molecular diagnostic test (RT-PCR kit), not an imaging AI algorithm that typically involves human readers. The performance is assessed based on the device's ability to detect viral nucleic acids against a comparator molecular test, not on improvements in human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, the primary evaluation of this device is a standalone performance study. The "TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel" is a laboratory-based RT-PCR diagnostic kit. Its performance, as detailed in the analytical and clinical studies, is measured by its output (detection/differentiation of viral targets) directly from patient samples, independent of human interpretation or "human-in-the-loop" assistance for the test result itself. The Diomni™ Software provides automated interpretation, making it an algorithm-only result at the interpretation stage.

7. The Type of Ground Truth Used

  • For analytical studies (e.g., LoD, Inclusivity, Precision, Interference, Cross-Reactivity, Stability), the ground truth was established by contrived samples with known concentrations of viral material or known presence/absence of interfering/cross-reactive substances. This is a common and appropriate method for analytical validation.
  • For the clinical performance study, the ground truth was established by an FDA-cleared molecular test (comparator device), supplemented by a resolver FDA-cleared molecular assay for discordant results. This constitutes a form of "composite reference standard" where the truth is determined by a highly reliable, already approved diagnostic method.

8. The Sample Size for the Training Set

  • The document describes the validation of a diagnostic kit, not an AI model requiring a separate "training set" in the machine learning sense. The information provided heavily details the test set performance (analytical and clinical validation).
  • While the "Diomni™ Software" performs data analysis and interpretation based on "assay-specific parameters from the Assay Definition File (ADF)," the document does not specify a distinct training set size typically associated with machine learning model development. The "training" of such a system would typically involve internal development data used to establish parameters like Cq cutoffs, but this is not detailed as a separate "training set" with specific sample numbers. The development study that determined the RNase P IC cutoff of Cq = 33.0 could be considered part of the "training" or parameter establishment phase, but no sample size is given for it.

9. How the Ground Truth for the Training Set Was Established

  • As noted above, a distinct "training set" for an AI/algorithm is not explicitly described in the context of this diagnostic kit's submission. The "ground truth" for establishing device parameters (like Cq cutoffs) would generally come from extensive analytical experiments and internal validation studies using characterized samples (e.g., known concentrations of viruses, negative controls).
  • For example, the preliminary LoD determined by probit analysis and the RNase P IC cutoff determined in a development study are examples of how internal parameters were established. These processes rely on highly controlled experiments with well-defined inputs (the "ground truth" being the known composition of the contrived samples).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 8, 2025

Life Technologies Corporation Stacey Moltchanoff Regulatory Affairs Manager 6055 Sunol Boulevard Pleasanton, California 94566

Re: K241806

Trade/Device Name: Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel Regulation Number: 21 CFR 866.3981 Regulation Name: Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The Sars-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test Regulatory Class: Class II Product Code: QOF Dated: June 21, 2024

Received: June 21, 2024

Dear Stacey Moltchanoff:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Anna M. Mielech -S

Anna Mielech, PhD. Deputy Branch Chief (Acting) Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241806

Device Name

Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel

Indications for Use (Describe)

The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) in vitro diagnostic test for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza B virus, and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections.

Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infective of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.

The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organism(s) detected by the Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections.

The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is intended for use by qualified and trained clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Submitter:Life Technologies Corporation (a legal entity of Thermo FisherScientific)
Address:6055 Sunol Blvd, Pleasanton, CA 94566, USA
Phone:(1) 760-603-7200
Fax:N/A
Contact Person:Stacey Moltchanoff
Contact email:stacey.moltchanoff@thermofisher.com
Contact Phone Number:1 (707) 494-1756
Date:June 21, 2024
Proprietary Name:Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B,RSV Select Panel
Measurand:The assay detects and identifies RNA from: SARS-CoV-2,influenza A virus, influenza B virus, and respiratory syncytialvirus (RSV)
Regulation:866.3981
Classification:Class II
Product Code:QOF
Panel:MI - Microbiology
Predicate Device:BioFire® Respiratory Panel 2.1 (DEN200031)

The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) test. The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, influenza B, respiratory syncytial virus (RSV) A/B and RNase P primer and probe sets are designed to detect viral RNA in nasopharyngeal (NP) and anterior nasal (AN) swab specimens from individuals exhibiting signs and symptoms of a respiratory tract infection.

Each TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel includes the following components:

  • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Assay-Multiplex assays that contain . primer and probe sets specific to the following targets:
    • Three SARS-CoV-2 targets (Orfla, Orf1b, and N genes) .
    • . Two influenza A virus targets (PB1 & M genes)
      • I Two influenza B virus targets (M & NS genes)
      • . Three RSV targets (NP, M, and L protein genes)
      • l RNase P (internal human sample collection control)

{5}------------------------------------------------

  • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Positive Control—Inactivated viral . control that contains SARS-CoV-2, influenza A, influenza B, and RSV.
  • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Negative Control—MS2 packaged RNA . control that contains targets specific to RNase P genomic regions targeted by the assay.
  • TaqPath™ 1-Step Select Master Mix (No ROX)—Ready-to-use PCR mix, including . reverse transcriptase, polymerase, dNTPs, salts, and buffer.
  • . Package Insert -- Provides the instructions and the link to download the instructions for use and other assets (including the ADF)
  • An Assay Definition File (ADF) applicable to the instrument used in the workflow . (available via download).

In addition to the SARS-CoV-2, influenza A, influenza B and RSV viral assay targets, the assay portion of the panel includes RNase P, which serves as an endogenous internal process control to monitor extraction and amplification of each clinical sample. The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel also contains external process positive and negative controls. The positive control (PC) component included is an inactivated viral control that contains SARS-CoV-2, influenza A, influenza B, and RSV viruses. The PC monitors extraction and real-time RT-PCR by demonstrating that each of the four viruses can be detected when present and that RNase P is not detected when absent. The negative control (NC) component included is an MS2 packaged RNA control that contains targets specific to RNase P genomic regions targeted by the assay. The NC also monitors extraction and real-time RT-PCR by demonstrating RNase P can be detected when present and that the four viruses are not detected when absent. The TaqPath™ 1-Step Select Master Mix (No ROX) included as a component of the kit is a ready-to-use PCR mix which contains a deoxyribonucleotide triphosphate mix (dNTPs), enzymes, and other components to permit reverse transcription and amplification of the assay targets. The TagPath™ 1-Step Select Master Mix (No ROX) also contains ribonuclease (RNase) inhibitors as well as deoxyuridine triphosphate (dUTP) and uracil N-glycosylase (known as UNG or UDG).

The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is provided in two overall kit configurations: either 200 reactions or 1.000 reactions. The materials provided within each TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel kit are shown in Table 1 (200 reactions, Cat. No. A56123) and Table 2 (1,000 reactions, Cat. No. A56124).

{6}------------------------------------------------

Table 1: TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel Components (200 reactions, Cat. No. A56123)

ComponentSKU (boxSKU)QuantityVolume pertube or bottle
TaqPath™ COVID 19, Flu A, Flu B, RSV SelectAssayA564402 tubes150 µL
TaqPath™ COVID 19, Flu A, Flu B, RSV SelectPositive Control9636126 tubes250 µL
TaqPath™ COVID 19, Flu A, Flu B, RSV SelectNegative Control9636096 tubes250 µL
TaqPath™ 1-Step Select Master Mix (No ROX)A565182 tubes1 mL

Table 2: TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel Components (1,000 reactions, Cat. No. A56124)

ComponentSKU (boxSKU)QuantityVolume pertube or bottle
TaqPath™ COVID 19, Flu A, Flu B, RSV SelectAssayA5644110 tubes150 µL
TaqPath™ COVID 19, Flu A, Flu B, RSV SelectPositive Control96361312 tubes250 µL
TaqPath™ COVID 19, Flu A, Flu B, RSV SelectNegative Control96360812 tubes250 µL
TaqPath™ 1-Step Select Master Mix (No ROX)A5651910 tubes1 mL

Materials required but not provided with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel:

  • Applied Biosystems™ QuantStudio™ 5 Dx Real-Time PCR Instrument ●
  • . MagMAX™ Dx Viral/Pathogen NA Isolation Kit
  • KingFisher™ Apex Dx Purification System .
  • . Additional reagents and materials listed in the instructions for use

A high-level overview of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel workflow is shown in Figure 1 below.

{7}------------------------------------------------

Image /page/7/Figure/0 description: The image shows a diagram of a process with four steps: Sample Collection, Sample Preparation, RT-PCR, and Data Analysis. In the Sample Collection step, RNA is extracted from a patient sample. In the Sample Preparation step, nucleic acids are isolated and purified from the specimens using the MagMAX Dx Viral/Pathogen NA Isolation Kit with the KingFisher Apex Dx Purification System. In the RT-PCR step, RT-PCR is run on the QuantStudio 5 Dx Real-Time PCR Instrument. In the Data Analysis step, results are analyzed and interpreted by the Diomni Software according to the settings in the Assay Definition File.

Figure 1: TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel Workflow

The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel workflow begins with nucleic acid extraction from nasopharyngeal and anterior nasal swab specimens that arrive at the testing site stored in viral transport media (VTM) or universal transport medium (UTM. Nucleic acids are isolated and purified from the specimens using the MagMAX™ Dx Viral/Pathogen NA Isolation Kit with the KingFisher™ Apex Dx Purification System.

The purified nucleic acid is reverse transcribed into cDNA and amplified using the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel and an Applied Biosystems™ QuantStudio™ 5 Dx Real Time PCR Instrument. The Applied Biosystems™ Diomni™ Software, which is used for results interpretation, is supplied with the PCR instrument.

Once the reaction plate is loaded onto the real-time PCR instrument, RNA is reverse transcribed into cDNA using the same gene-specific primers used for PCR. After heat inactivation of the reverse transcriptase enzyme, thermal cycling for PCR amplification commences.

During the anneal/extension phase of a PCR cycle, primers and probes specific to the target genes bind to corresponding DNA sequences in the indicated targets. The 5' nuclease activity of Taq polymerase degrades the probe, causing the reporter dye to separate from the quencher dye, generating a fluorescent signal. With each cycle, additional reporter dye molecules are cleaved

{8}------------------------------------------------

from their respective probes, increasing the overall fluorescence intensity. Fluorescence intensity is monitored at each PCR cycle by the real-time PCR instrument.

The Diomni™ Software implements the thermal cycling program, secondary analysis and clinical interpretation of the results using assay-specific parameters from the Assay Definition File (ADF) that is installed in the software.

Following PCR, results are analyzed and interpreted by the Diomni™ Software according to the settings in the ADF. A detailed explanation of the calling rules, plate validity and Cq cutoff values for assay targets is provided in the instructions for use. A minimum of one negative control and one positive control must be present for each run for Diomni™ Software to call the run valid. Furthermore, all control wells must pass for the real-time RT-PCR plate to be considered valid.

Control well interpretive rules are show in Table 3 below.

Negative Control (NC)Well CallStatus
C19Flu AFlu BRSV ABRNase P
NEGNEGNEGNEGPOSNegativePassed
All other scenariosInvalidFailed

Table 3: Control Well Interpretive Rules

Positive Control (PC)Well CallStatus
C19Flu AFlu BRSV ABRNase P
POSPOSPOSPOSNEGPositivePassed
All other scenariosInvalidFailed

Results interpretation for patient samples is described in Table 4 below. A Positive call for RNase P is only required for samples that are Negative for all four viral targets; samples with no amplification are called Invalid.

{9}------------------------------------------------

TargetCallResult
C19Flu AFlu BRSV ABRNase PCallResult
POSPOSPOSPOSPOS or NEGPositiveSARS-CoV-2, Flu A, Flu B, RSVAB Detected
POSPOSPOSNEGPOS or NEGPositiveSARS-CoV-2, Flu A, Flu B Detected
POSPOSNEGPOSPOS or NEGPositiveSARS-CoV-2, Flu A, RSVAB Detected
POSPOSNEGNEGPOS or NEGPositiveSARS-CoV-2, Flu A Detected
POSNEGPOSPOSPOS or NEGPositiveSARS-CoV-2, Flu B, RSVAB Detected
POSNEGPOSNEGPOS or NEGPositiveSARS-CoV-2, Flu B Detected
POSNEGNEGPOSPOS or NEGPositiveSARS-CoV-2, RSVAB Detected
POSNEGNEGNEGPOS or NEGPositiveSARS-CoV-2 Detected
NEGPOSPOSPOSPOS or NEGPositiveFlu A, Flu B, RSVAB Detected
NEGPOSPOSNEGPOS or NEGPositiveFlu A, Flu B Detected
NEGPOSNEGPOSPOS or NEGPositiveFlu A, RSVAB Detected
NEGPOSNEGNEGPOS or NEGPositiveFlu A Detected
NEGNEGPOSPOSPOS or NEGPositiveFlu B, RSVAB Detected
NEGNEGPOSNEGPOS or NEGPositiveFlu B Detected
NEGNEGNEGPOSPOS or NEGPositiveRSVAB Detected
NEGNEGNEGNEGPOSNegativeSARS-CoV-2, Flu A, Flu B, RSVAB NotDetected
NEGNEGNEGNEGNEGInvalidRETEST[1]

Table 4: Patient Samples Interpretive Rules

[1] Retesting must be performed by re-extracting the original sample and repeating the real-time PCR. If the repeat result remains invalid, collection of a new specimen should be considered.

Intended Use

The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) in vitro diagnostic test for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus, influenza B virus, and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2,

{10}------------------------------------------------

influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections.

Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.

The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms. The organism(s) detected by the Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections.

The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is intended for use by qualified and trained clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.

Predicate Device Comparison

The similarities and differences between TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel and the predicate device, BioFire® Respiratory Panel 2.1 (DEN20031) are described in Table 5 and Table 6 below.

{11}------------------------------------------------

Device & Predicate DeviceNew DeviceDEN200031
Device Trade NameTaqPath™ COVID-19, Flu A, FluB, RSV Select PanelBioFire® Respiratory Panel 2.1(RP2.1)
General DeviceCharacteristic Similarities
Organisms DetectedSARS-CoV-2, Influenza A,Influenza B, Respiratory SyncytialVirusSARS-CoV-2, Influenza A,Influenza B, Respiratory SyncytialVirus
AnalyteRNARNA
Technological Principlesreal-time reverse transcriptionPCRreverse transcription PCR
Specimen Typesnasopharyngeal swabsnasopharyngeal swabs
Test InterpretationAutomated test interpretationAutomated test interpretation
Transport MediaUniversal Transport Medium(UTM) / Viral Transport Medium(VTM)Universal Transport Medium(UTM) / Viral Transport Medium(VTM)
General DeviceCharacteristic DifferencesNew DeviceDEN200031
Device Trade NameTaqPath™ COVID-19, Flu A, Flu B, RSV SelectPanelBioFire® Respiratory Panel 2.1 (RP2.1)
Intended Use/ IndicationsFor UseThe Applied Biosystems™ TaqPath™ COVID-19,Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction(RT-PCR) in vitro diagnostic test for the qualitativedetection and differentiation of severe acuterespiratory syndrome coronavirus 2 (SARS-CoV-2),influenza A virus, influenza B virus, and respiratorysyncytial virus (RSV). Nucleic acids are isolated andpurified from nasopharyngeal (NP) swab and anteriornasal (AN) swab specimens obtained from individualsexhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections.Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus and aids in diagnosis ifThe BioFire® Respiratory Panel 2.1 (RP2.1) is aPCR-based multiplexed nucleic acid test intendedfor use with the BioFire FilmArray® 2.0 orBioFire FilmArray® Torch systems for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections, including COVID-19.The following organism types and subtypes are identified using the BioFire® RP2.1:Adenovirus, Coronavirus 229E, Coronavirus HKU1, Coronavirus NL63, Coronavirus OC43, Severe Acute Respiratory SyndromeCoronavirus (SARS-CoV-2), Human Metapneumovirus, Human Rhinovirus/Enterovirus, Influenza A, including subtypes H1, H1-2009,and H3, Influenza B,
General DeviceCharacteristic DifferencesNew DeviceDEN200031
used in conjunction with other clinical andepidemiological information, and laboratory findings.The results of this test should not be used as the solebasis for diagnosis, treatment, or other patientmanagement decisions. Positive results do not rule outcoinfection with other organisms. The organism(s)detected by the Applied Biosystems™ TaqPath™COVID-19, Flu A, Flu B, RSV Select Panel may notbe the definite cause of disease. Negative results donot preclude SARS-CoV-2, influenza A virus,influenza B virus, or RSV infections.The Applied Biosystems™ TaqPath™ COVID-19,Flu A, Flu B, RSV Select Panel is intended for use byqualified and trained clinical laboratory personnelspecifically instructed and trained in the techniques ofreal-time PCR and in vitro diagnostic procedures.Parainfluenza Virus 1, Parainfluenza Virus 2, Parainfluenza Virus 3, Parainfluenza Virus 4, Respiratory Syncytial Virus, Bordetella parapertussis (IS1001), Bordetella pertussis (ptxP), Chlamydia pneumoniae, and Mycoplasma pneumoniae Nucleic acids from the respiratory viral andbacterial organisms identified by this test aregenerally detectable in NPS specimens during theacute phase of infection. The detection andidentification of specific viral and bacterial nucleicacids from individuals exhibiting signs and/orsymptoms of respiratory infection is indicative ofthe presence of the identified microorganism andaids in the diagnosis of respiratory infection ifused in conjunction with other clinical andepidemiological information. The results of thistest should not be used as the sole basis fordiagnosis, treatment, or other patient managementdecisions.Negative results in the setting of a respiratoryillness may be due to infection with pathogens thatare not detected by this test, or lower respiratorytract infection that may not be detected by an NPS
General DeviceCharacteristic DifferencesNew DeviceDEN200031
specimen. Positive results do not rule outcoinfection with other organisms. The agent(s)detected by the BioFire® RP2.1 may not be thedefinite cause of disease. Additional laboratorytesting (e.g. bacterial and viral culture,immunofluorescence, and radiography) may benecessary when evaluating a patient with possiblerespiratory tract infection.
Specimen TypesAnterior Nasal Swabs and Nasopharyngeal swabsNasopharyngeal swabs only
InstrumentationSample Prep/Nucleic Acid Purification:• MagMAX™ Dx Viral/Pathogen NA Isolation Kit(magnetic bead-based sample prep chemistry)• KingFisher™ Apex Dx Purification System(sample prep instrumentation)Real-time PCR instrument:• QuantStudio 5 DxSample Prep/Nucleic Acid Purification:• Magnetic bead-based chemistry is containedwithin the BioFire® RP2.1 pouchInstruments:• FilmArray® 2.0• FilmArray® Torch
ControlsOne negative control and one positive control are runfor each plate (external process controls). The assaydetects endogenous RNase P present in the specimenas an internal positive/process control.Two internal controls are included in each reagentpouch for quality control of sample processing andboth PCR stages and melt analysis.

Table 5: Predicate Device Similarities

{12}------------------------------------------------

Table 6: Predicate Device Differences

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

Analytical/Clinical Performance

Note that the terms “Cq” and “Ct” are used interchangeably throughout the documentation and refer to equivalent values. Quantification cycle (Cq) is an algorithm-agnostic term across different methods for determining the fractional cycle used for PCR reaction quantification, whereas "Ct", or threshold cycle, refers specifically to the baseline-threshold algorithm term used by the software for determining this same value.

Analytical Performance Summaries

The analytical performance testing that was conducted to support substantial equivalence is summarized below.

Limit of Detection Determination

The preliminary LoD was determined by probit analysis at a hit rate of 0.95. LoD was determined for two (2) strains of influenza A (Flu A) H1N1, two (2) strains of Flu A H3N2, one (1) strain of influenza B (Flu B) of Victoria lineage, two (2) strains of Flu B of Yamagata lineage, one (1) strain of SARS-CoV-2, one (1) strain of RSV A and one (1) strain of RSV B. Each virus was spiked into pooled, negative nasopharyngeal (NP) swab specimens in VTM/UTM, extracted on the KingFisher™ Apex Purification System, then real-time RT-PCR was performed on the QuantStudio™ 5 Dx (QS5 Dx) PCR instrument across three (3) days, with two (2) reagent lots. LoDs for all viruses except the Flu B Victoria lineage strain were also determined in anterior nasal (AN) swab specimens in VTM/UTM. The NP and AN swab specimens in VTM/UTM were obtained as de-identified clinical remnants which were confirmed negative and pooled. In this study, the preliminary LoD was determined by probit analysis separately for each virus in each specimen type with each reagent lot, and the highest (i.e., worst-case) LoD value across reagent lots was selected as the preliminary LoD for that virus in that specimen type. In parallel, the Limit of Blank (LoB) was assigned as zero and confirmed as such in this LoD Determination Study.

Probit analysis was used to determine the preliminary LoD at a hit rate of 0.95 for each virus strain tested. The preliminary LoDs for each virus strain and specimen type across reagent lots is

{16}------------------------------------------------

provided in Table 7. The concentrations below are provided in both GCE/mL (as determined by in-house ddPCR) and in TCID50/mL (as reported by the virus supplier in a Certificate of Analysis).

VirusStrainLoD Concentration(GCE/mL)LoD Concentration(TCID50/mL)
NPANNPAN
SARS-CoV-2USA-WA1/202082820.1210.120
Flu A H1N1 strain #1Michigan/45/151663050.0220.041
Flu A H1N1 strain #2California/07/09881830.0080.017
Flu A H3N2 strain #1South Australia/55/142033350.0250.041
Flu A H3N2 strain #2Texas/50/121422320.0420.068
Flu B Victoria strainMalaysia/2506/20044287N/A0.092N/A
Flu B Yamagata strain#1Victoria/504/005475520.0470.047
Flu B Yamagata strain#2Florida/04/062322270.0240.024
RSV A2013 Isolate Culture Fluid9827530.0080.006
RSV BCH93(18)-186646140.0200.019

Table 7: Preliminary LoD as determined by probit analysis for each specimen type across reagent lots on the QS5Dx.

Limit of Detection Confirmation

The preliminary LoDs determined by probit analysis were confirmed in the LoD Confirmation Study. The previously determined preliminary LoDs were confirmed for the same two (2) strains of Flu A H1N1, two (2) strains of Flu A H3N2, one (1) strain of Flu B of Victoria lineage, two (2) strains of Flu B of Yamagata lineage, one (1) strain of SARS-CoV-2, one (1) strain of RSV A and one (1) strain of RSV B. Each virus was spiked into pooled, negative NP (all viruses) and AN (all except Flu B Victoria lineage) swab specimens in VTM/UTM and extracted on the KingFisher™ Apex Purification System, then real-time RT-PCR performed on the QS5 Dx instrument across three (3) days with one (1) reagent lot. The preliminary LoD was confirmed as the lowest concentration (i.e., most dilute) that results in at least 95% positive detection (≥95% positive) when tested with at least 24 replicates. Viral concentrations are reported in both GCE/mL and TCID50/mL.

{17}------------------------------------------------

The LoD was confirmed for each virus strain and specimen type as the lowest concentration (i.e., most dilute) that resulted in at least 95% positive detection (≥95% positive) and is presented in both GCE/mL and TCID50/mL. The overall LoD concentrations (i.e., worst case) across matrices (NP and AN) are provided in Table 8 below.

VirusStrainLoD Concentration(GCE/mL)LoD Concentration(TCID50/mL)
SARS-CoV-2USA-WA1/20201070.158
Flu A H1N1 strain #1Michigan/45/153400.045
Flu A H1N1 strain #2California/07/092820.026
Flu A H3N2 strain #1South Australia/55/142810.034
Flu A H3N2 strain #2Texas/50/122680.079
Flu B Victoria strainMalaysia/2506/200435000.075
Flu B Yamagata strain #1Victoria/504/0014050.120
Flu B Yamagata strain #2Florida/04/064640.048
RSV A2013 Isolate Culture Fluid16140.013
RSV BCH93(18)-1814780.046

Table 8: Overall LoD Values for the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel

Limit of Detection with WHO Standard

The study was performed using two-phase approach to determine the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel LoD using the WHO Standard on the QS5Dx PCR instrument. Contrived specimens were formulated with the First WHO International Standard for SARS-CoV-2 RNA spiked into pooled, negative anterior nasal (AN) swab specimens in VTM/UTM at various concentrations. Once formulated, contrived specimens were extracted and detected using the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel in Phase I (range finding). The results of Phase I were used to establish the preliminary LoD (which was the lowest concentration to generate 100% positivity when tested in triplicate). In Phase II, the preliminary LoD was confirmed as the lowest concentration that results in at least 95% positive detection (≥95% positive) when tested with 20 replicates. Phase II testing was performed in a blinded format.

{18}------------------------------------------------

The results of Phase I testing produced a preliminary LoD of $5.012 × 10^1$ IU/mL, which was the lowest (most dilute) concentration that yielded 100% positivity. Results are summarized in Table 9 below.

ConcentrationPositive / Total Replicates (% Positive)
5.012 × 105 IU/mL3 / 3 (100%)
5.012 × 104 IU/mL3 / 3 (100%)
5.012 × 103 IU/mL3 / 3 (100%)
5.012 × 102 IU/mL3 / 3 (100%)
5.012 × 101 IU/mL3 / 3 (100%)
5.012 × 100 IU/mL1 / 3 (33.3%)
5.012 × 10-1 IU/mL0 / 3 (0%)
5.012 × 10-2 IU/mL0 / 3 (0%)
5.012 × 10-3 IU/mL0 / 3 (0%)
Negative0 / 3 (0%)

Table 9: Summary of Phase I Results with WHO Standard

In Phase II, the LoD was confirmed at 150.36 IU/mL as the lowest concentration that resulted in ≥ 95% positive detection as summarized in Table 10 below.

18 / 20 (90%)

12 / 20 (60%)

0 / 3 (0%)

ConcentrationPositive / Total Replicates(% Positive)
451.08 IU/mL (9x Preliminary LoD)20 / 20 (100%)
150.36 IU/mL (3x Preliminary LoD)20 / 20 (100%)

Table 10: Summary of Phase II Results with WHO Standard

In silico Reactivity (Inclusivity)

50.12 IU/mL (Preliminary LoD)

16.71 IU/mL (0.33x Preliminary LoD)

Negative

In silico analysis was conducted for RSV A/B, Flu A, Flu B, COVID-19 assay primers and probes using BLAST against known strains/isolates in the GISAID and GenBank (NCBI) databases to determine reactivity (inclusivity) of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel. The study included a summary of mismatch and melting temperature analysis. The acceptance criteria for this study was that less than 1% (SARS-CoV-2) and less than 5% (Flu A, Flu B, and RSV) of publicly available sequences have mutations that map to all gene targets and also reduce the melting temperature below annealing temperature.

{19}------------------------------------------------

A strain must have 100% homology or a predicted melting temperature (Tm) above the PCR annealing temperature (Tg) to at least one of the individual assays targeting that organism to be considered a positive for inclusivity. Greater than 99.99% of SARS-CoV-2 had Tm > Ta for at least one of the three SARS-CoV-2 assays. Likewise, 99.9% of Flu B strains and 99.49% of RSV strains had at least one assay with Tm > Ta. Whereas 10.43% of the Flu A strains had mismatches that cause the predicted Tm to fall below the T a for both the MP and PB1 gene assays, the most frequent mismatch in the MP gene was wet tested and found to not impact detection. After accounting for the number of strains with the mismatch in the MP gene, 95.27% of Flu A strains can be considered reactive with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel.

Reactivity (Inclusivity)

Reactivity (inclusivity) was determined for eleven (11) strains of Flu A (H1N1), twelve (12) strains of Flu A (H3N2), five (5) strains of influenza B (two (2) of Victoria lineage and three (3) of Yamagata lineage), six (6) strains of RSV (three (3) each of RSV A and B), five (5) strains of SARS-CoV-2 and five (5) avian influenza strains (one (1) each of H5N1, H5N2, H7N2, H7N7 and H7N9). Each virus was introduced at near-LoD levels (approximately 3x LoD) into negative, pooled NP swab specimens in VTM/UTM, then extracted and detected with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel. Sample extraction was performed using the KingFisher™ Apex Purification System, and real-time RT-PCR was performed with the QS5Dx using one (1) lot of TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel reagents. Each sample was formulated with live virus, inactivated virus or genomic RNA depending on availability, and tested in triplicate (N=3). Virus strains that did not produce 100% detection at the initial test concentration were further tested at progressively higher concentrations until 100% detection was achieved.

Each virus was tested and observed to be reactive (unless otherwise noted) with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel at approximately 3x LoD. As multiple strains per species were used for LoD determination for Flu A, Flu B and RSV, there was not a single LoD value for these viruses. For Flu A viruses, in which the LoD was determined for two strains each for subtypes H1N1 and H3N2, the worst-case LoD for the same subtype was used as the LoD. For

{20}------------------------------------------------

avian flu, the worst case for Flu A across both subtypes was used as the LoD. For Flu B and RSV, the subtype- or lineage-specific LoD was used. Results are shown in Table 11 below.

StrainPos Replicates / Total (% Positivity)GCE/mLTCID50/mL
Influenza A H1N1 (Georgia/M5081/2012)3/3 (100%)1020 GCE/mL42.83 TCID50/mL
Influenza A H1N1 (New Caledonia/20/99)3/3 (100%)1020 GCE/mL0.44 TCID50/mL
Influenza A H1N1 (NY/03/09) [1],{2]3/3 (100%) [1]4080 GCE/mL [1]6.88 TCID50/mL [1]
Influenza A H1N1 (Singapore/63/04)3/3 (100%)1020 GCE/mL0.05 TCID50/mL
Influenza A H1N1 (Brisbane/02/2018)[2]3/3 (100%)1020 GCE/mL0.11 TCID50/mL
Influenza A H1N1 (Solomon Islands/3/2006)3/3 (100%)1020 GCE/mL1.67 TCID50/mL
Influenza A H1N1 (Puerto Rico/08/1934)3/3 (100%)1020 GCE/mL0.11 TCID50/mL
Influenza A H1N1 (Mexico/4108/09) [2]3/3 (100%)1020 GCE/mL7.75 TCID50/mL
Influenza A H1N1 (Taiwan/42/06)3/3 (100%)1020 GCE/mL0.99 TCID50/mL
Influenza A H1N1 (Brisbane/59/07)3/3 (100%)1020 GCE/mL0.07 TCID50/mL
Influenza A H1N1 (Victoria/2570/19)[2]3/3 (100%)1020 GCE/mL0.03 TCID50/mL
Influenza A H3N2 (Wisconsin/67/2005)3/3 (100%)1278 GCE/mL0.20 TCID50/mL
Influenza A H3N2 (Switzerland/9715293/13)3/3 (100%)1278 GCE/mL0.007 TCID50/mL
Influenza A H3N2 (Kansas/14/2017)3/3 (100%)1278 GCE/mL0.69 TCID50/mL
Influenza A H3N2 (Singapore/INFIMH-16-0019/16)3/3 (100%)1278 GCE/mL0.57 TCID50/mL
Influenza A H3N2 (Perth/16/09)3/3 (100%)1278 GCE/mL0.07 TCID50/mL
Influenza A H3N2 (Victoria/361/2011)3/3 (100%)1278 GCE/mL0.02 TCID50/mL
Influenza A H3N2 (Hong Kong/8/68)3/3 (100%)1278 GCE/mL0.12 TCID50/mL
Influenza A H3N2 (Brisbane/10/07)3/3 (100%)1278 GCE/mL0.20 TCID50/mL
Influenza A H3N2 (California/7/04)3/3 (100%)1278 GCE/mL0.72 TCID50/mL
Influenza A H3N2 (Hong Kong/4801/14)3/3 (100%)1278 GCE/mL2.32 TCID50/mL
Influenza A H3N2 (Hong Kong/2671/19)3/3 (100%)1280 GCE/mL1.07 TCID50/mL
Influenza A H3N2 (Macha/O1237/2021)3/3 (100%)1280 GCE/mL2.62 TCID50/mL
Influenza B Victoria (Colorado/06/2017)3/3 (100%)4215 GCE/mL [4]0.08 TCID50/mL
Influenza B Victoria (Victoria/2/87)3/3 (100%)4215 GCE/mL [4]1.74 TCID50/mL
Influenza B Yamagata (Massachusetts/02/2012)3/3 (100%)1392 GCE/mL0.02 TCID50/mL
Influenza B Yamagata (Phuket/3073/13)3/3 (100%)1392 GCE/mL0.004 TCID50/mL
Influenza B Yamagata (Wisconsin/01/2010)3/3 (100%)1392 GCE/mL0.06 TCID50/mL
RSV A (Long)3/3 (100%)4842 GCE/mL4.56 TCID50/mL
RSV A (2014 Isolate 342)3/3 (100%)4842 GCE/mL0.26 TCID50/mL
RSV A (2006 isolate)3/3 (100%)4842 GCE/mL0.05 TCID50/mL
RSV B (9320)[5]3/3 (100%)4434 GCE/mLUnknown [3]
StrainPos Replicates /Total (%Positivity)GCE/mLTCID50/mL
RSV B (3/2015 Isolate #2)3 / 3 (100%)4434 GCE/mL2.07 TCID50/mL
RSV B (12/2014 Isolate #1)3 / 3 (100%)4434 GCE/mL0.56 TCID50/mL
SARS-CoV-2 (Alpha Variant (B.1.1.7)England/204820464/2020)[5]3 / 3 (100%)321 GCE/mL0.26 TCID50/mL
SARS-CoV-2 (Beta (B.1.351) South Africa/KRISP-K005325/2020)[5]3 / 3 (100%)321 GCE/mL0.09 TCID50/mL
SARS-CoV-2 (Gamma (P1) Japan/TY7-503/2021)[5]3 / 3 (100%)321 GCE/mL0.03 TCID50/mL
SARS-CoV-2 (Delta (B.1.617.2) USA/PHC658/2021)[5]3 / 3 (100%)321 GCE/mL0.02 TCID50/mL
SARS-CoV-2 (Omicron (BA.2.3) USA/MD-HP24556/2022)[5]3 / 3 (100%)321 GCE/mL0.02 TCID50/mL
Avian influenza H5N1 (India/NIV/2006)[5]3 / 3 (100%)1278 GCE/mLUnknown[3]
Avian influenza H5N2 (pheasant/New Jersey/1355/1998)[5]3 / 3 (100%)1278 GCE/mLUnknown[3]
Avian influenza H7N2 (turkey/Virginia/4529/2002)[5]3 / 3 (100%)1278 GCE/mLUnknown[3]
Avian influenza H7N7 (mallard/Netherlands/12/2000)[5]3 / 3 (100%)1278 GCE/mLUnknown[3]
Avian influenza H7N9 (Anhui/1/2013)[5]3 / 3 (100%)1278 GCE/mLUnknown[3]

Table 11: Reactivity (Inclusivity) Study Results

{21}------------------------------------------------

[1] Reactive with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel at 12x LoD.

[2] H1N1pdm09 strain.

[3] TCID50/mL concentrations unknown (not reported by supplier).

[4] 1.2x LoD relative to LoD of 3500 GCE/mL determined for Flu B Victoria lineage strain Malaysia/2506/2004.

[5] Viral genomic RNA or inactivated virus

Within-Laboratory Precision - Testing Extracted RNA

The Within-Laboratory Precision Study was performed using a panel of extracted RNA from nine (9) specimens contrived by spiking quantified virus from one (1) representative viral strain each of SARS-CoV-2, Flu A, Flu B, and RSV into a negative pool of nasopharyngeal swab (NP) specimen in VTM/UTM at concentrations relative to the LoD confirmed for that strain. RNA from each sample was extracted, pooled, aliquoted, and frozen pending use in the study. Use of extracted RNA for this study provides a separate precision value for the real-time PCR portion of the workflow alone and independent of sample extraction. Repeatability and within-lab precision values that include the sample extraction portion of the workflow were calculated from the Siteto-site Reproducibility study. NP was used to evaluate within-lab precision as the representative and more challenging sample type.

The study was performed by two (2) operators each completing two (2) real-time PCR runs per day (4 runs total per day) at one (1) internal test site on 12 testing days spread over a ≥ 20-day

{22}------------------------------------------------

study duration. All testing was completed using one (1) real-time PCR reagent lot and on one (1) QuantStudio™ 5 Dx (QS5 Dx) real-time PCR instrument. Each panel member was tested with two (2) replicates in every run with at least one (1) Negative Control (NC) and one (1) Positive Control (PC) included in each reaction plate as run controls.

The positivity rate for the negative sample was 0% (100% NPA) as expected. The overall PPA for all targets at approximately 1x LoD and 3x LoD on the QS5Dx was 100.00%. The observed precision and repeatability %CV at 1x and 3x LoD were less than 2 %CV for all targets. The results of the Cq Variability Analysis are shown in Table 12.

Panel MemberNMean CqRandom EffectVariance95%Lower95%UpperSDCV (%)
Negative96/9625.16Operator0.00-0.010.010.050.18
Day0.01-0.010.020.080.31
Run0.030.010.050.170.68
Repeatability0.010.010.020.100.40
Within-LabPrecision0.050.030.080.220.87
SARS-CoV-21x LoD96/9635.52Operator0.00-0.030.040.060.18
Day0.04-0.020.110.210.59
Run-0.03-0.110.050.000.00
Repeatability0.270.190.420.521.47
Within-LabPrecision0.320.220.500.561.59
SARS-CoV-23x LoD96/9634.11Operator0.00-0.010.010.040.13
Day-0.03-0.06-0.010.000.00
Run0.03-0.020.090.180.54
Repeatability0.100.070.150.310.92
Within-LabPrecision0.140.100.210.371.08
Flu A1x LoD96/9633.05Operator0.000.000.000.000.00
Day0.01-0.010.030.100.30
Run0.00-0.020.030.070.20
Repeatability0.060.040.090.240.73
Within-LabPrecision0.070.050.100.270.82
Panel MemberNMean CqRandom EffectVariance95%Lower95%UpperSDCV (%)
Flu A3x LoD96/9632.48Operator0.00-0.010.010.040.13
Day0.00-0.010.010.050.17
Run0.00-0.020.010.000.00
Repeatability0.050.030.070.210.66
Within-LabPrecision0.050.030.080.220.69
Flu B1x LoD96/9634.03Operator0.00-0.010.010.040.10
Day-0.01-0.030.000.000.00
Run0.02-0.010.050.130.38
Repeatability0.060.040.100.250.73
Within-LabPrecision0.080.060.120.280.83
Flu B3x LoD96/9632.35Operator0.000.000.000.000.00
Day0.010.000.020.090.27
Run0.01-0.010.020.080.23
Repeatability0.020.020.040.150.47
Within-LabPrecision0.040.030.050.190.59
RSV1x LoD96/9634.28Operator0.00-0.010.000.000.00
Day-0.01-0.050.030.000.00
Run0.02-0.050.090.150.45
Repeatability0.170.120.270.421.22
Within-LabPrecision0.200.140.290.441.30
RSV3x LoD96/9632.30Operator-0.01-0.010.000.000.00
Day0.060.000.120.250.78
Run0.03-0.010.070.170.52
Repeatability0.080.050.120.280.87
Within-LabPrecision0.170.120.260.411.28

Table 12: Summary of Within-Laboratory Precision (Testing Extracted RNA) – Cq Variability Analysis Results

{23}------------------------------------------------

{24}------------------------------------------------

Site-to-Site Reproducibility

The Site-to-Site Reproducibility study was performed using a panel of nine (9) specimens contrived by spiking quantified virus into negative pool of NP swab specimens in VTM/UTM (NP specimen pool). One (1) representative viral strain each of SARS-CoV-2, Flu, Flu B, and RSV was selected to be spiked into negative NP specimen pool at concentrations relative to the LoD determined for each analyte strain.

Reproducibility of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel was assessed for the QuantStudio™ 5 Dx (QS5 Dx) real-time PCR instrument. Testing was performed at three (3) test sites across five (5) non-consecutive testing days with two (2) instruments per site. Two of the testing sites were external to Thermo Fisher Scientific and one site was internal. Each panel member was tested with each of the three (3) real-time PCR reagent lots in triplicate, for a total of nine (9) replicates per panel member in every run. At least one (1) Negative Control (NC) and one (1) Positive Control (PC) were included on each plate as run controls.

Two operators per site tested the samples following the full TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel workflow, with each site using a different lot of nucleic acid extraction kit (three (3) lots total) and a single KingFisher™ Apex Purification System at each site for performing the automated RNA extractions. Each operator performed one (1) run on the PCR instrument per day.

The NPA for the negative sample was 99.63%. The overall Site-to-Site reproducibility PPA for all targets at 1x LoD ranged from 92.59% to 98.89% and at 3x LoD, the reproducibility PPA was 100.00% for all targets (Table 13). The observed reproducibility, within-lab precision and repeatability % CV at approximately 1x and 3x LoD was less than 5% CV for all targets. Results are shown in Table 14 below.

{25}------------------------------------------------

AnalyteTest LevelAgreement/Expected (% Agreement)Site 1Agreement/Expected (% Agreement)Site 2Agreement/Expected (% Agreement)Site 3Overall Agreement(95% Score CI)
N/ANegative90/90(100%)90/90(100%)89/90(98.89%)[1]99.63%(97.93%, 99.93%)
SARS-CoV-2~1x LoD85/90(94.44%)82/90(91.11%)83/90(92.22%)92.59%(88.84%, 95.15%)
SARS-CoV-2~3x LoD90/90(100%)90/90(100%)90/90(100%)100%(98.60%, 100%)
Flu A~1x LoD88/90(97.78%)90/90(100%)89/90(98.89%)98.89%(96.78%, 99.62%)
Flu A~3x LoD90/90(100%)90/90(100%)90/90(100%)100%(98.60%, 100%)
Flu B~1x LoD86/90(95.56%)89/90(98.89%)81/90(90.00%)94.81%(91.49%, 96.89%)
Flu B~3x LoD90/90(100%)90/90(100%)90/90(100%)100%(98.60%, 100%)
RSV~1x LoD87/90(96.67%)87/90(96.67%)90/90(100%)97.78%(95.24%, 98.98%)
RSV~3x LoD90/90(100%)90/90(100%)90/90(100%)100%(98.60%, 100%)

Table 13: Reproducibility Study – Qualitative Results

[1] One replicate was called positive for Flu A.

{26}------------------------------------------------

Panel MemberNMean CqRandom EffectVariance95% Lower95% UpperSDCV (%)
NegativeRNase P269/27025.35Site0.08-0.080.240.281.11
Lot0.00-0.010.010.060.24
Day-0.11-0.17-0.060.000.00
Instrument/Operator0.250.150.360.501.99
Repeatability0.010.010.010.100.39
Within-lab Precision0.270.522.04
Reproducibility0.350.210.650.592.32
SARS-CoV-21x LoD250/27036.09Site0.08-0.090.260.290.79
Lot0.01-0.030.060.120.33
Day-0.05-0.130.030.000.00
Instrument/Operator0.160.020.290.391.09
Repeatability0.430.350.540.661.82
Within-lab Precision0.600.772.14
Reproducibility0.680.500.980.822.29
SARS-CoV-23x LoD270/27034.50Site0.07-0.090.230.260.76
Lot0.01-0.030.040.080.23
Day0.07-0.010.140.260.75
Instrument/Operator0.05-0.020.130.230.67
Repeatability0.370.310.460.611.77
Within-lab Precision0.500.712.05
Reproducibility0.570.420.800.752.18
Flu A1x LoD267/27034.75Site0.01-0.030.050.100.28
Lot-0.01-0.02-0.010.000.00
Day-0.06-0.180.060.000.00
Instrument/Operator0.260.060.460.511.46
Repeatability0.610.500.760.782.25
Panel MemberNMean CqRandom EffectVariance95% Lower95% UpperSDCV (%)
Within-lab Precision0.870.932.69
Reproducibility0.880.701.140.942.70
Site0.00-0.010.020.050.16
Lot0.00-0.010.000.000.00
Day-0.08-0.160.000.000.00
Flu A3x LoD270/27033.05Instrument/Operator0.250.100.400.501.51
Repeatability0.300.240.370.541.65
Within-lab Precision0.540.742.23
Reproducibility0.550.420.740.742.24
256/27035.01Site0.11-0.140.360.330.96
Lot-0.01-0.020.000.000.00
Day0.05-0.060.160.220.63
Flu Blx LoDInstrument/Operator0.04-0.110.180.200.56
Repeatability0.770.630.970.882.51
Within-lab Precision0.860.932.65
Reproducibility0.970.741.340.982.81
270/27033.50Site0.22-0.230.670.471.40
Lot-0.01-0.010.000.000.00
Day-0.06-0.170.040.000.00
Flu B3x LoDInstrument/Operator0.250.080.420.501.50
Repeatability0.480.400.600.702.08
Within-lab Precision0.740.862.56
Reproducibility0.960.611.710.982.92
264/27035.43Site0.09-0.140.320.300.86
Lot-0.02-0.030.000.000.00
RSVlx LoDDay-0.02-0.240.200.000.00
Instrument/Operator0.24-0.090.570.491.37
Repeatability1.541.261.921.243.50
Panel MemberNMean CqRandom EffectVariance95% Lower95% UpperSDCV (%)
Within-lab Precision1.771.333.76
Reproducibility1.861.492.401.373.85
RSV3x LoD270/27033.81Site0.00-0.020.030.040.11
Lot0.02-0.040.080.140.42
Day-0.26-0.45-0.070.000.00
Instrument/Operator0.430.060.790.651.93
Repeatability1.281.051.591.133.34
Within-lab Precision1.721.313.88
Reproducibility1.721.382.201.313.88

Table 14: Summary of Site-to-site Reproducibility - Ct Variability Analysis Results

{27}------------------------------------------------

{28}------------------------------------------------

{29}------------------------------------------------

Interfering Substances

Interference was assessed by testing contrived positive samples consisting of one strain each of SARS-CoV-2, Flu A, Flu B and RSV at 3x their respective limits of detection (LoDs) as well as negative samples in the presence or absence of the twenty-six (26) interfering substances, two (2) solvents, one (1) no interferent control, and five (5) viral transport media (VTM). Additionally, potential interference due to extraction reagent carryover was assessed with four (4) carryover conditions and one (1) no-carryover control. For all potential interferents other than VTM, contrived samples were prepared in pooled negative NP specimens. VTM testing used pooled negative AN swab specimens that had been collected in the VTM being tested. Each substance was tested in triplicate (N=3) for each positive and negative sample through the full workflow with one (1) lot of reagents on the QuantStudio™ 5 Dx (QS5 Dx) real-time PCR instrument.

Each of the conditions tested produced the expected results (i.e., each of the Positive samples was "Positive" for its relevant target, and each of the Negative samples was "Negative"). Detection of both Flu A and Flu B in all samples tested with FluMist® (0.05%, 0.01% and 0.005%) was expected since it is a live-attenuated virus vaccine against influenza A and influenza B viruses. Importantly, the presence of FluMist® did not result in false positive (FP) or false negative (FN) results for the other targets at any of the three concentrations tested. None of the other exogenous or endogenous substances, transport media or extraction kit reagents tested interfered with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel. Results showing positivity and mean Cq values by sample are shown in Table 15 below. Concentrations (Conc.) listed are per volume of sample before extraction for all conditions except Extraction Reagent Carryover. Extraction Reagent Carryover substances were added directly to the post-extraction eluate at the indicated concentrations per volume of eluate.

{30}------------------------------------------------

NEGFlu AFlu BSARS-CoV-2RSV
Substance DescriptionConc.% Positive(nPositive/Total)MeanCq% Positive(nPositive/Total)MeanCq% Positive(nPositive/Total)MeanCq% Positive(nPositive/Total)MeanCq% Positive(nPositive/Total)MeanCq
No interferent control(NP)N/A0% (0/9)33.20100% (6/6)33.32100% (9/9)34.20100% (6/6)33.42
Mucin5 mg/mL0% (0/3)33.94100% (3/3)33.23100% (3/3)35.20100% (3/3)33.23
Whole blood5% (v/v)0% (0/3)32.76100% (3/3)33.08100% (3/3)34.09100% (3/3)32.78
Jurkat cell gDNA7 µg/mL0% (0/3)33.65100% (3/3)33.09100% (3/3)34.02100% (3/3)33.15
Nasal decongestant -Phenylephrine (Equate™)10% (v/v)0% (0/3)32.23100% (3/3)33.30100% (3/3)34.72100% (3/3)33.60
Nasal decongestant -Oxymetazoline (Afrin®)10% (v/v)0% (0/3)30.50100% (3/3)31.20100% (3/3)34.42100% (3/3)32.83
Saline nasal spray(Equate™)10% (v/v)0% (0/3)32.69100% (3/3)33.04100% (3/3)34.04100% (3/3)33.29
Bronchodilator(Ventolin® Evohaler®)0.83mg/mL0% (0/3)33.67100% (3/3)33.35100% (3/3)34.26100% (3/3)33.51
Nasal corticosteroids -Dexamethasone1.5 mg/mL0% (0/3)33.60100% (3/3)32.87100% (3/3)34.44100% (3/3)33.48
Nasal corticosteroids -Flunisolide2 mg/mL0% (0/3)34.16100% (3/3)32.93100% (3/3)34.78100% (3/3)31.27
Nasal corticosteroids -Budesonide (Apotex®)1% (v/v)0% (0/3)33.71100% (3/3)33.00100% (3/3)34.10100% (3/3)32.73
Nasal corticosteroids -Fluticasone (Equate™)1% (v/v)0% (0/3)33.17100% (3/3)33.57100% (3/3)33.74100% (3/3)33.11
Nasal gel (NeilMed®Nasogel®)1% (w/v)0% (0/3)33.38100% (3/3)33.27100% (3/3)34.60100% (3/3)33.29
Homeopathic allergyrelief medicine (Zicam®)1% (w/v)0% (0/3)33.36100% (3/3)33.84100% (3/3)34.74100% (3/3)33.12
NEGFlu AFlu BSARS-CoV-2RSV
Substance DescriptionConc.% Positive(nPositive/Total)% Positive(nPositive/Total)MeanCq% Positive(nPositive/Total)MeanCq% Positive(nPositive/Total)MeanCq% Positive(nPositive/Total)MeanCq
Throat lozenges(Chloraseptic® Max)2.2 mg/mLmethanol;3.3 mg/mLbenzocaine0% (0/3)100% (3/3)33.71100% (3/3)33.46100% (3/3)34.33100% (3/3)33.25
Zinc lozenges (LifeExtension®)7.5 mg/mL0% (0/3)100% (3/3)33.69100% (3/3)33.17100% (3/3)34.06100% (3/3)33.20
Antiviral - Zanamivir5.5 mg/mL0% (0/6)100% (4/4)33.46100% (6/6)32.95100% (3/3)34.33100% (3/3)33.39
Antiviral - Oseltamivirphosphate33 µg/mL0% (0/3)100% (3/3)33.25100% (3/3)31.93100% (3/3)34.52100% (3/3)33.38
Antiviral - Remdesivir6.7 µg/mL0% (0/3)100% (3/3)33.45100% (3/3)33.48100% (3/3)33.88100% (3/3)34.03
Topical antibiotic(Pseudomonic Acid)3.3 mg/mL0% (0/3)100% (3/3)32.09100% (3/3)33.06100% (3/3)34.56100% (3/3)32.43
Systemic antibiotic(Tobramycin)4 µg/mL0% (0/3)100% (3/3)33.33100% (3/3)33.24100% (3/3)34.40100% (3/3)32.92
Flu vaccine: FluMist®0.05%(v/v)100% (3/3) [1]100% (3/3)21.59100% (3/3)20.85100% (3/3)33.19100% (3/3)32.82
0.01%(v/v)100% (3/3) [1]100% (3/3)23.89100% (3/3)23.36100% (3/3)32.85100% (3/3)32.99
0.005%(v/v)100% (3/3) [1]100% (3/3)24.86100% (3/3)24.24100% (3/3)33.37100% (3/3)32.88
Analgesic (Ibuprofen)220 µg/mL0% (0/3)100% (3/3)33.30100% (3/3)32.69100% (3/3)34.00100% (3/3)33.48
Tobacco (Camel Snus®)1% (w/v)0% (0/3)100% (3/3)32.93100% (3/3)32.99100% (3/3)33.85100% (3/3)32.82
Solvent DMSO20% (v/v)0% (0/6)100% (3/3)33.53100% (6/6)32.74100% (3/3)34.69100% (3/3)31.27
Solvent Ethanol20% (v/v)0% (0/3)100% (3/3)32.34100% (3/3)32.79100% (3/3)34.01100% (3/3)33.60
VTM Remel® M4RTNeat[2]0% (0/3)100% (3/3)33.35100% (3/3)33.37100% (3/3)34.73100% (3/3)32.10
NEGFlu AFlu BSARS-CoV-2RSV
Substance DescriptionConc.% Positive(nPositive/Total)% Positive(nPositive/Total)MeanCq% Positive(nPositive/Total)MeanCq% Positive(nPositive/Total)MeanCq% Positive(nPositive/Total)MeanCq
VTM Remel® M5Neat[2]0% (0/3)100% (3/3)33.00100% (3/3)33.19100% (3/3)34.76100% (3/3)33.06
VTM Remel® M6Neat[2]0% (0/3)100% (3/3)33.31100% (3/3)33.05100% (3/3)35.42100% (3/3)33.15
VTM Copan® UTM-RTNeat[2]0% (0/3)100% (3/3)32.88100% (3/3)31.92100% (3/3)34.57100% (3/3)32.76
VTM BD® UVTNeat[2]0% (0/3)100% (3/3)31.77100% (3/3)33.40100% (3/3)34.47100% (3/3)32.96
No Extraction ReagentCarryover ControlN/A0% (0/3)100% (3/3)33.43100% (3/3)32.97100% (3/3)33.94100% (3/3)33.26
Extraction ReagentCarryover - Sample Plate1% (v/v)[3]0% (0/3)100% (3/3)33.47100% (3/3)33.06100% (3/3)34.59100% (3/3)33.52
Extraction ReagentCarryover - Wash 11% (v/v) [3]0% (0/3)100% (3/3)33.67100% (3/3)33.44100% (3/3)34.25100% (3/3)33.33
Extraction Reagent6% (v/v) [3]0% (0/3)100% (3/3)33.90100% (3/3)33.13100% (3/3)34.02100% (3/3)33.54
Carryover - Wash 23% (v/v) [3]0% (0/3)100% (3/3)33.32100% (3/3)32.79100% (3/3)34.41100% (3/3)33.33

Table 15: Summary of Interfering Substances Testing Results for the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel

{31}------------------------------------------------

{32}------------------------------------------------

[1] As expected, NEG samples containing FluMist® were positive for Flu A and Flu B and negative for SARS-CoV-2 and RSV

[2] Neat = undiluted; AN samples originally collected in indicated VTM

[2] Neat = undiluted; AN samples originally collected in indicated VTM
[3] Interferent added to eluates (post-extraction), and concentrations are per volume of eluate.

{33}------------------------------------------------

Competitive Interference

Competitive interference testing was performed using contrived samples containing pairs of Flu A, Flu B, SARS-CoV-2 or RSV with one virus at a low concentration (3x LoD) and the other virus at a high concentration ($1.0 \n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n $TCID$50$/mL) spiked into pooled negative NP specimens. In parallel, viruses were tested at both levels individually to serve as controls and demonstrate within-panel specificity of each of the target channels. Sample extraction was performed using the KingFisher™ Apex Purification System and real-time RT-PCR was performed with the QS5Dx real-time PCR instrument with one (1) lot of TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel reagents. Note that different virus strains were used to prepare the highconcentration virus samples because the quantified virus stocks used for the 3x LoD level were either insufficiently concentrated to achieve the target of $105$ TCID$50$/mL or had limited stock.

The results of each competitive interference condition tested satisfied the acceptance criteria. Three positive test results among the three total replicates (N=3) (100% positivity) for each viral target for both low-concentration virus and the high-concentration virus were produced. There were no off-target false-positive results in any target channel observed for the within-panel specificity testing component of the study (i.e., single targets present at $105$ TCID$50$/mL). The negative, pooled NP specimens (alone, with no virus added) were tested in the workflow as a negative control condition and yielded negative test results, as expected. The results of the study are summarized in Table 16.

Table 16: Summary of Competitive Interference Testing Results for the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel

Viral Test ConditionCombinationLow Target (Group 1)High Target (Group 2)
Virus(Concentration)nPositive/Total (%Positive)Virus(Concentration)nPositive/Total (%Positive)
SARS-CoV-2 (low), Flu A(high)SARS-CoV-2 (3xLoD)3 / 3 (100%)Flu A (105TCID50/mL)3 / 3 (100%)
SARS-CoV-2 (low), Flu B(high)SARS-CoV-2 (3xLoD)3 / 3 (100%)Flu B (105TCID50/mL)3 / 3 (100%)
SARS-CoV-2 (low), RSV(high)SARS-CoV-2 (3xLoD)3 / 3 (100%)RSV (105TCID50/mL)3 / 3 (100%)
Low Target (Group 1)High Target (Group 2)
Viral Test ConditionCombinationVirus(Concentration)nPositive/Total (%Positive)Virus(Concentration)nPositive/Total (%Positive)
Flu A (low), SARS-CoV-2(high)Flu A (3x LoD)3 / 3 (100%)SARS-CoV-2 (105TCID50/mL)3 / 3 (100%)
Flu A (low), Flu B (high)Flu A (3x LoD)3 / 3 (100%)Flu B (105TCID50/mL)3 / 3 (100%)
Flu A (low), RSV (high)Flu A (3x LoD)3 / 3 (100%)RSV (105TCID50/mL)3 / 3 (100%)
Flu B (low), SARS-CoV-2(high)Flu B (3x LoD)3 / 3 (100%)SARS-CoV-2 (105TCID50/mL)3 / 3 (100%)
Flu B (low), Flu A (high)Flu B (3x LoD)3 / 3 (100%)Flu A (105TCID50/mL)3 / 3 (100%)
Flu B (low), RSV (high)Flu B (3x LoD)3 / 3 (100%)RSV (105TCID50/mL)3 / 3 (100%)
RSV (low), SARS-CoV-2(high)RSV (3x LoD)3 / 3 (100%)SARS-CoV-2 (105TCID50/mL)3 / 3 (100%)
RSV (low), Flu A (high)RSV (3x LoD)3 / 3 (100%)Flu A (105TCID50/mL)3 / 3 (100%)
RSV (low), Flu B (high)RSV (3x LoD)3 / 3 (100%)Flu B (105TCID50/mL)3 / 3 (100%)
SARS-CoV-2 (low)SARS-CoV-2 (3xLoD)3 / 3 (100%)N/AN/A
SARS-CoV-2 (high)N/AN/ASARS-CoV-2 (105TCID50/mL)3 / 3 (100%)
Flu A (low)Flu A (3x LoD)3 / 3 (100%)N/AN/A
Flu A (high)N/AN/AFlu A (105TCID50/mL)3 / 3 (100%)
Flu B (low)Flu B (3x LoD)3 / 3 (100%)N/AN/A
Flu B (high)N/AN/AFlu B (105TCID50/mL)3 / 3 (100%)
RSV (low)RSV (3x LoD)3 / 3 (100%)N/AN/A
RSV (high)N/AN/ARSV (105TCID50/mL)3 / 3 (100%)
Negative (NP alone)N/AN/AN/AN/A

{34}------------------------------------------------

{35}------------------------------------------------

Cross-Reactivity & Microbial Interference

Contrived samples were prepared by spiking one strain each of 49 microorganisms into negative pooled NP samples to determine whether false positive (FP) results would be produced for any of the targets of the test. In parallel, the same microorganisms were tested in the presence of one representative viral strain for each of SARS-CoV-2, Flu A, Flu B, and RSV at 3x LoD to determine whether false negative (FN) results would be produced for these targets. Potentially cross-reactive microbes were either tested individually or in groups (multi-spikes). Each sample was tested in triplicate (N=3) through the full workflow with one (1) lot of reagents on one (1) representative PCR instrument model. Microbes with stock concentrations too low to achieve the target concentration were tested at 20% (v/v) in negative pooled NP samples. In addition to contrived samples, results from samples tested during the Clinical Validation study that had been called positive for coronavirus HKU1 (CoV-HKU1) by the comparator test were assessed for evidence of cross-reactivity with SARS-CoV-2, Flu A, Flu B or RSV.

Cross-reactivity was evaluated by assessing the occurrence of false positive results in negative pooled NP samples spiked with the microbes tested; pooled human nasal wash was tested alone. None of the potentially cross-reactive organisms tested produced FP results. Moreover, none of the microbes tested inhibited PCR or interfered with extraction or RNase P detection as evidenced by the RNase P signal in each sample. Within-panel cross-reactivity was assessed as part of the Competitive Interference Study. When each of the viruses was present at 10° TCID50/mL, FPs were not observed for the other targets. All but one of the 39 CoV-HKU1-positive samples from the Clinical Performance study were called negative for all virus targets, with one sample producing a positive result of SARS-CoV-2 below the Cg cutoff for positivity. Results are shown in Table 17 below.

{36}------------------------------------------------

Table 17: Cross-Reactivity & Microbial Interference Study Results for the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel

NegativeSamplesContrived Positive Samples at 3x LoD
OrganismSpeciesTestNumberNumber Positive / Total (% Positive)
TypeConcentrationNegative /Total (%Negative)SARS-CoV-2Flu AFlu BRSV
BacteriaBordetellaparapertussis1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaBordetella pertussis1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaChlamydophilapneumoniae1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaCorynebacteriumdiphtheriae1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaEscherichia coli1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaFusobacteriumnecrophorum1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaHaemophilusinfluenzae type b1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaKlebsiella pneumoniae1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaLactobacillusacidophilus1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaLegionellapneumophila1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaMoraxella catarrhalis1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaMycobacteriumtuberculosis5.4 x 107copies/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaMycoplasmagenitalium4 x 105cells/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaMycoplasmapneumoniae1 x 106CCU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaNeisseria elongata1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaNeisseria meningitidisserogroup A1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaPseudomonasaeruginosa1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaStaphylococcus aureusMRSA1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaStaphylococcusepidermidis MRSE1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
NegativeSamplesContrived Positive Samples at 3x LoD
OrganismSpeciesTestNumberNumber Positive / Total (% Positive)
TypeConcentrationNegative /Total (%Negative)SARS-CoV-2Flu AFlu BRSV
BacteriaStreptococcuspneumoniae1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaStreptococcuspyogenes1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BacteriaStreptococcussalivarius1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
FungiAspergillus flavus1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
FungiCandida albicans1 x 106CFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
FungiPneumocystis carinii1 x 106cells/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusAdenovirus type 1,species C1 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusMastadenovirus Btype 7, Gomen1 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusCytomegalovirus(HHV-5)3.2 x 104TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusEnterovirus type 681 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusEpstein-Barr virus(HHV-4)1.57 x 107copies/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusHepatitis B virus4.2 x 107IU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusHepatitis C virus3.3 x 107IU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusHerpes simplex virus1 (HSV-1)1 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusHuman coronavirus229E2.82 x 104TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusHuman coronavirusHKU1[1]2 x 106copies/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusHuman coronavirusHKU139 positiveclinicalspecimens38/39(97.4%)[2]NottestedNottestedNottestedNot tested
VirusHuman coronavirusNL632.34 x 104TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusHuman coronavirusOC431 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
NegativeSamplesContrived Positive Samples at 3x LoD
OrganismTypeSpeciesTestNumberNumber Positive / Total (% Positive)
ConcentrationNegative /Total (%Negative)SARS-CoV-2Flu AFlu BRSV
VirusHumanimmunodeficiencyvirus type IIIB1 x 104 IU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusHumanmetapneumovirus1 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusMeasles virus1 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusMERS coronavirusstrain Florida/USA-2Saudi Arabia 2014Not provided[3]3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusMERS coronavirusstrain EMC/20121 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusMumps virus1 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusParainfluenza virustype 11 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusParainfluenza virustype 21 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusParainfluenza virustype 31 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusParainfluenza virustype 4A1 x 105TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusRhinovirus type 1A2.82 x 104TCID50/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusSARS coronavirusstrain 2003-00592Not provided[3]3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusSARS coronavirusstrain Urbani1 x 105PFU/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
VirusVaricella-zoster virus(HHV-3)1.94 x 108copies/mL3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)
BiofluidPooled human nasalwashNeat3/3 (100%)3/3(100%)3/3(100%)3/3(100%)3/3(100%)

{37}------------------------------------------------

{38}------------------------------------------------

[1] Subgenomic synthetic control (ATCC VR-3262SD)

[2] One sample produced a positive for SARS-CoV-2.

[3] Concentration not provided by vendor.

In silico Cross-Reactivity

In silico analysis was conducted for the COVID-19, Flu A, Flu B, RSV and RNase P assay primers and probes using BLAST against known strains/isolates from the GenBank database, as of March

{39}------------------------------------------------

29, 2023. Cross-reactivity was evaluated in silico by determining whether any of the microbes had at least 80% homology with primers or probes and produced an amplicon of less than 10,000 bases.

No isolate out of the 1,155 tested matched to 3 or more individual components of any TaqPath™ COVID-19, Flu A. Flu B. RSV Select Panel assay. An amplicon will not be detected when less than 3 individual assay components bind to an isolate and an amplicon will not be produced unless two primers bind; therefore, all sequences were predicted to be negative for cross-reactivity.

Specimen Stability

Specimen stability was assessed by examining the stability of contrived NP and AN swab specimens collected in VTM/UTM at the following storage conditions:

  • . Upper end of Room temperature (30°C) for up to 24 hours
  • Upper end of Refrigerated temperature (8°C) for up to 72 hours .
  • Low frozen temperature (≤ -70°C) for up to 30 days .

Contrived samples were created by pooling individual negative clinical NP and AN swab specimens collected in VTM/UTM respectively, then aliquoting the pooled matrix and spiking with one representative strain for each of SARS-CoV-2, Flu A, Flu B and RSV at 3x LoD. Upon contriving, samples were aliquoted and stored at the various storage conditions prior to testing. At each timepoint, samples were removed from their storage condition and moved to ≤ -70°C to prevent further aging. An isochronous design was employed where possible to reduce variability in testing. When ready to conduct testing, extraction was performed for all samples being tested together using the MagMAX™ Dx Viral/Pathogen Nucleic Acid Isolation Kit. Negative NP and AN samples were extracted in parallel. All samples were tested with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel on the Applied Biosystems™ QuantStudio™ 5 Dx. Specimen stability testing utilized one lot of extraction and PCR reagents for all storage conditions.

All targets for both NP and AN specimens produced 100% positivity at all the time points tested for the 8°C, 30°C and the ≤ -70°C temperatures and therefore, support the claimed specimen stability in the Instruction for Use of 24 hours at room temperature (15–25°C), 72 hours refrigerated (2–8°C) and 30 days at ≤-70°C.

{40}------------------------------------------------

Eluate Stability

The Eluate Stability Study evaluated the performance of refrigerated and frozen-and-thawed nucleic acid eluates with the TaqPath™ COVID-19, Flu A. Flu B. RSV Select Panel. NP and AN specimens contrived with SARS-CoV-2, Flu A, Flu B and RSV to 3x LoD were used with the MagMAX™ Dx Viral/Pathogen Nucleic Acid Isolation Kit to collect the eluates. Eluates were either tested fresh or stored at +8°C for 20 and 27 hours, then underwent two (2) freeze-and-thaw cycles (-70°C to +8°C) before testing with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel.

Eluate stability was assessed by the occurrence of FN results in contrived positive samples, each containing SARS-CoV-2, Flu A, Flu B or RSV at 3x LoD. 100% of contrived positive sample eluates produced the expected results (correct calls) at all three (3) time points, supporting the eluates storage claims for up to 24 hours at 2°C – 8°C, 5 days at ≤ -70°C and after one (1) freeze/thaw cycle, cumulatively.

Fresh vs Frozen Equivalency

The Fresh Versus Frozen Equivalency Study examined equivalency of fresh and frozen NP and AN swab specimens collected in VTM/UTM. The study served two purposes: first, to support the use of frozen samples during the Clinical Validation study and second, to evaluate freeze/thaw stability of specimens. Positived samples were created by pooling individual negative clinical NP and AN swab specimens collected in VTM/UTM respectively, then aliquoting the pooled matrix and spiking with one representative strain for each of SARS-CoV-2, Flu A, Flu B and RSV at 2x LoD and 5x LoD. Negative samples from the same NP and AN specimen pools were also tested. Upon contriving, samples were aliquoted for immediate testing for the "fresh" study arm, and the remaining sample aliquots were stored at ≤ -70°C to undergo the freeze/thaw cycles prior to testing. Samples were frozen at ≤ -70°C for at least 2 hours (maximum 24 hours) or until fully frozen. After thaw event (maximum 4 hours or until completely thawed), the aliquots were removed and stored for testing. Testing was conducted for samples that underwent two (2), three (3) and six (6) freeze/thaw (F/T) cycles.

{41}------------------------------------------------

Samples undergoing freeze/thaw were stored at ≤ -70°C upon completion of their respective F/T cycles and tested together to reduce variability in testing. Extraction was performed using the MagMAX™ Dx Viral/Pathogen Nucleic Acid Isolation Kit. Negative NP and AN samples were also extracted in parallel. All samples were tested with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel on the Applied Biosystems™ QuantStudio™ 5 Dx. Fresh vs frozen equivalency testing utilized one lot of extraction and real-time PCR reagents for all conditions.

The results of this study support the use of frozen specimens in the clinical performance study and the claim in the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel Instructions for Use that specimens collected in Copan® UTM Universal Transport Medium (UTM) or BD® Universal Viral Transport (UVT) Medium may undergo up to 2 freeze/thaw cycles. Do not exceed 1 freeze/thaw cycle for specimens collected in other transport media types.

In-Use and Hold Time Stability

The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel has a recommended storage temperature of -30°C to -10°C and the kit components require thawing prior to use. The in-use stability of the reagents during typical in-use conditions were evaluated while executing the workflow, and the performance of the kit components was compared to the untreated baseline (i.e., time point T0 or freeze/thaw cycle FT0).

Assay Mix Tube and Master Mix were tested using contrived positive samples in NP matrix containing SARS-CoV-2, Flu A, Flu B or RSV at 3x LoD as well as negative samples consisting of NP matrix alone. Positive Control (PC) and Negative Control (NC) were used unstressed as batch controls for Master Mix and Assay Mix Tube testing. PC and NC were also stressed to determine their own in-use stability parameters. Testing utilized the isochronous design where possible in which reagents were returned to frozen storage at various time points and tested together at the end. A minimum of three time points or freeze/thaw cycles were tested per condition, including T0/FT0, and all testing extended beyond the claimed stability period. Samples were prepared in advance and extracted with the MagMAX™ Dx Viral/Pathogen Nucleic Acid Isolation Kit on the KingFisher™ Apex Purification System, and testing was performed on the QuantStudio™ 5 Dx. Data were analyzed qualitatively to determine if positive samples remained

{42}------------------------------------------------

positive, negative samples remained valid negative and PC and NC gave the required results after in-use testing.

Data from these studies support the storage conditions and freeze thaw cycles specified the in Instructions for Use for the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel for the Assay Tube Mix, Master Mix and PC and NC.

The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel components met all the acceptance criteria for all the test points in all the tested conditions and summarized as follows:

  • The Assay Mix Tube and Master Mix demonstrated stability when held at 80 hours at 8°C, . and thus the Assay Mix Tube and Master Mix are deemed stable when each are stored at 2°C to 8°C for up to 72 hours.
  • The Positive Control and Negative Control demonstrated stability when held at 27 hours . at 8°C prior to extraction, and therefore the Positive Control and Negative Control are deemed stable when each are stored at 2℃ to 8℃ for up to 24 hours.
  • The Assay Tube Mix and Master Mix demonstrated stability to the fifth freeze/thaw cycle, . and are deemed stable up to 4 freeze/thaw cycles (inclusive) from -30°C to -10°C to 2°C to 8°C.

Carryover Cross-Contamination

In the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel workflow, PCR plate setup is performed manually and only the extraction step utilizes instrumentation that could result in carryover and/or cross-contamination. To this end, assessment of carryover and crosscontamination was performed beginning from sample extraction. A modified checkerboard layout was extracted in which 12 high-concentration samples containing inactivated SARS-CoV-2 viral material at 105 TCIDss/mL in pooled negative NP samples were interspersed with negative NP samples.

Six full plates were extracted and tested sequentially by two operators on two different KingFisher™ Apex Purification System instruments. After extraction, the eluates were tested

{43}------------------------------------------------

using the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel. SARS-CoV-2 was used as a model for all targets because (a) SARS-CoV-2 has higher clinically relevant concentrations and (b) the SARS-CoV-2 LoD is lower (more sensitive) that the other target viruses. Of the 492 replicates run, 0 of the negative sample wells produced a positive call for SARS-CoV-2, providing a carryover/cross-contamination rate of 0%.

RNase P Internal Control Cutoff Confirmation

In the RNase P Internal Control Cutoff Confirmation study, viral target-negative "used" (i.e., collected from individual donors and containing RNase P) nasopharyngeal (NP) swab specimens and "unused" (i.e., unopened and not containing RNase P) NP swabs were tested according to the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel Instructions for Use. Sample extraction was performed using the KingFisher™ Apex Purification System, and RT-PCR was performed on the QS5Dx with one (1) lot of TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel reagents. The RNase P IC cutoff was determined to be Cq = 33.0 in a development study and was confirmed in the execution of this study.

The IC cutoff was confirmed at Cq = 33.0 while testing individually collected NP swabs on the QS5Dx instrument. The results of the experiment demonstrated an overall positive percent agreement (PPA) of 98.07% (95% CI of 97.31-98.63%) and negative percent agreement (NPA) of 100% (95% CI of 97.49-100%) for RNase P detection in NP swabs. For AN swabs, PPA for RNase P detection was 92.30% (95% CI of 90.83-93.54%) and NPA was 100% (95% CI of 97.50-100%).

Due to the lower than expected PPA observed for AN specimens in the RNase P Cutoff confirmation study, an additional study was conducted utilizing an improved AN swab collection technique. All of the evaluated samples produced valid calls. Two (2) out of the one hundred thirtyeight (138) AN samples were positive for COVID-19, while the rest were negative for all viral targets. RNase P was detected in 100% of both negative and positive samples. The results are summarized in Table 18.

{44}------------------------------------------------

SampleCallRNase PCallNumberofSamplesPercentValidityRNase P (Cq)
MeanMedianSD
NegativePositive136100%27.126.72.1
COVID-19PositivePositive2100%26.926.91.9
AllPositive138100%27.126.72.1

Table 18: Invalid Rate in AN Specimen Swab Collection Study Results

This study has demonstrated that the improved AN swab collection technique reduces the invalid rate for AN swabs when tested with the TaqPath COVID-19, Flu A, Flu B, RSV Select Panel. The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel Instructions for Use include the improved AN swab collection instructions.

Matrix Equivalency

The matrix equivalency study evaluated the performance of NP and AN swab specimens (collected in VTM/UTM) with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel. Contrived samples prepared with one strain each of SARS-CoV-2, Flu A, Flu B and RSV at two concentrations (2x and 5x LoD) in NP and AN sample matrices were tested with one lot of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel on the QS5Dx real-time PCR instrument. The selected virus concentrations (2x and 5x relative to LoD) represent low positive and moderate positive specimens, respectively. Negative specimens were also tested. All samples produced 100% concordance with the expected results; thus, matrix equivalency between NP and AN swab specimens collected in VTM/UTM was demonstrated for the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel.

Clinical Cut-Off

The clinical cut-off study was conducted in two phases. The first phase was a preliminary benchmarking investigation into the performance of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel against each two comparator devices to assess performance near the analytical limits of the TaqPath™ COVID-19, Flu A. Flu B. RSV Select Panel. The second testing phase, the

{45}------------------------------------------------

Receiver Operating Characteristic (ROC) justification study, was designed to further evaluate the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel with clinical and contrived samples against each comparator device for the purpose of defining the clinical cut-off prior to design validation. The method used to determine the most appropriate clinical cutoff was the Receiver Operating Characteristic analysis according to CLSI EP24-A2.

The benchmarking and ROC justification studies established cutoffs for the TagPath™ COVID-19. Flu A. Flu B. RSV Select Panel by comparing to the predicate device.

Clinical Performance

The clinical performance testing that was conducted to support substantial equivalence is summarized below.

Clinical Performance (Prospective Study)

A clinical performance study was conducted using the comparator approach to evaluate the clinical performance of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel (investigational test) in nasopharyngeal (NP) swabs and anterior nasal (AN) swabs from individuals with signs and symptoms of respiratory tract infections including SARS-CoV-2, influenza A, influenza B, or RSV. The comparator device chosen in this study was an FDA cleared molecular test that detects and differentiates between SARS-CoV-2, influenza A virus, influenza B virus, and respiratory syncytial virus (RSV). NP and AN swab specimens were collected prospectively and during an enrichment phase from eligible subjects who provided consent or assent to participate in the study. Specimens collected in this study were tested by the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel and the comparator device.

The study population was composed of participants who were experiencing signs and symptoms of respiratory viral infection. A total of one thousand nine (1,909) subjects from 14 different collection sites were enrolled in this study, of which 1,840 subjects were prospectively enrolled from all comers who met the inclusion criteria and 69 subjects were enrolled during an enrichment phase of the study. One nasopharyngeal swab and one anterior nasal swab in Universal

{46}------------------------------------------------

Viral Transport Medium was collected from each participant. Sample collection was completed between February 20, 2023 and February 29, 2024.

After applicable exclusions, a total of 1,620 NP and 1,541 AN swabs were included in final data analysis for the prospective cohort. Two (2) of the 1,541 AN specimens and four (4) of the 1,620 NP specimens observed inconclusive comparator results for Flu A, leaving 1,539 evaluable AN specimens and 1,616 NP specimens for Flu A. Specimens were tested either fresh as Category I samples or frozen as Category II samples. The performance comparison of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel and the comparator test results are presented in Table 19. Performance characteristics of the study included Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) for each clinical target using NP and AN swab specimen types on the QuantStudio™ 5 Dx (QS5 Dx) instrument. PPA and NPA are defined as follows:

$$PPA (%) = 100 \times \frac{Number \ of \ Positives \ by \ Both \ Evaluation \ Method \ and \ Its \ Comparator}{Number \ of \ Positives \ by \ Comparator \ Method}$$

$$NPA (%) = 100 \times \frac{Number \ of \ Negatives \ by \ Both \ Evaluation \ Method \ and \ Its \ Comparator}{Number \ of \ Negatives \ by \ Comparator \ Method}$$

Two-sided, Wilson Score confidence intervals (CI) around each estimate were also calculated. Samples that produced a discordant call between the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel and the comparator test were further tested with another FDA cleared molecular assay. Results from discordant sample testing are presented as footnotes in each table.

Of the 1,840 subjects enrolled in the prospective study, 1,647 had their AN swab specimens evaluated with the candidate device. One hundred seventy nine (179) of these specimens were invalid by the candidate device during testing, for an initial invalid rate of 10.9% (179/1,647). Upon retesting, the invalid rate for AN specimens decreased to 6.4% (105/1,647). Of the 1,840 subjects enrolled in the prospective study, 1,647 had their NP swab specimens evaluated with the candidate device. Forty five (45) of these specimens were invalid by the candidate device during testing, for an initial invalid rate of 2.7% (45/1,647). Upon retesting, the invalid rate for NP specimens decreased to 1.6% (27/1.647).

{47}------------------------------------------------

TargetSwab TypeSpecimenCollectionNumber ofSpecimensTruePositiveFalseNegativeTrueNegativeFalsePositivePPA(%)95% Two-SidedConfidenceIntervalNPA(%)95% Two-SidedConfidenceInterval
SARS-CoV-2NasopharyngealSwabFresh124686311144396.6%90.6% - 98.9%96.28%95.0% - 97.2%
NasopharyngealSwabFrozen374562307996.6%88.3% - 99.1%97.15%94.7% - 98.5%
Overall16201425a142152b96.6%92.3% - 98.5%96.47%95.4% - 97.3%
SARS-CoV-2Anterior NasalSwabFresh119280310684196.4%89.9% - 98.8%96.3%95.0% - 97.3%
Anterior NasalSwabFrozen349413296993.2%81.8% - 97.7%97.05%94.5% - 98.4%
Overall15411216c136450d95.3%90.1% - 97.8%96.46%95.4% - 97.3%
Flu ANasopharyngealSwabFresh12424311191797.7%88.2% - 99.6%99.42%98.8% - 99.7%
NasopharyngealSwabFrozen374132356386.7%62.1% - 96.3%99.16%97.6% - 99.7%
Overall1616563e154710f94.9%86.1% - 98.3%99.36%98.8% - 99.7%
Flu AAnterior NasalSwabFresh1191390114012100.0%91.0% - 100.0%98.96%98.2% - 99.4%
Anterior NasalSwabFrozen34861339285.7%48.7% - 97.4%99.41%97.9% - 99.8%
Overall1539451g147914h97.8%88.7%-99.6%99.06%98.4% - 99.4%
Flu BNasopharyngealSwabFresh12463111213196.9%84.3% - 99.5%99.92%99.5% - 100.0%
NasopharyngealSwabFrozen3741803551100.0%82.4% - 100.0%99.72%98.4% - 100.0%
Overall1620491i15682j98.0%89.5%-99.7%99.87%99.5% - 100.0%
Flu BAnterior NasalSwabFresh119230011602100.0%88.7% - 100.0%99.83%99.4% - 100.0%
Anterior NasalSwabFrozen3491703320100.0%81.6% - 100.0%100.0%98.9% - 100.0%
Overall154147014922k100.0%92.4% - 100.0%99.87%99.5% - 100.0%
RSVNasopharyngealSwabFresh12462511216496.2%81.1%-99.3%99.67%99.2% - 99.9%
NasopharyngealSwabFrozen37451368083.3%43.7% - 97.0%100.0%99.0% - 100.0%
Overall1620302l15844m93.8%79.9%-98.3%99.75%99.4%-99.9%
RSVAnterior NasalSwabFresh119224011635100.0%86.2% - 100.0%99.57%99.0% - 99.8%
Anterior NasalSwabFrozen349503431100.0%56.6% - 100.0%99.71%98.4% - 100.0%
Overall154129015066n100.0%88.3% - 100.0%99.6%99.1%-99.8%

Table 19: Performance Estimates for the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel (Prospective Cohort)

a Discordant test results based on resolver testing: 3/5 SARS-CoV-2 Positive and 2/5 SARS-CoV-2 Negative

b Discordant test results based on resolver testing: 38/52 SARS-CoV-2 Positive and 14/52 SARS-CoV-2 Negative

{48}------------------------------------------------

& Discordant test results based on resolver testing: 5/6 SARS-CoV-2 Positive and 1/6 SARS-CoV-2 Negative

d Discordant test results based on resolver testing: 37/50 SARS-CoV-2 Positive and 13/50 SARS-CoV-2 Negative

  • e Discordant test results based on resolver testing: 1/3 Flu A Positive and 2/3 Flu A Negative

  • f Discordant test results based on resolver testing: 3/10 Flu A Positive and 7/10 Flu A Negative

  • 8 Discordant test results based on resolver testing: 1/1 Flu A Negative

  • i Discordant test results based on resolver testing: 1/1 Flu B Negative

  • J Discordant test results based on resolver testing: 1/2 Flu B Positive and 1/2 Flu B Negative

  • k Discordant test results based on resolver testing: 2/2 Flu B Positive

  • 1 Discordant test results based on resolver testing: 2/2 RSV Negative

  • m Discordant test results based on resolver testing: 4/4 RSV Negative

™ Discordant test results based on resolver testing: 1/6 RSV Positive and 5/6 RSV Negative

h Discordant test results based on resolver testing 4/14 Flu A Positive and 10/14 Flu A Negative.

{49}------------------------------------------------

Enrichment Study

After one thousand (1,000) subjects were enrolled in the prospective study and the desired minimum number of SARS-CoV-2 positive samples were collected, an enrichment phase of the study was initiated to supplement the prospective data for Flu A, Flu B and RSV. Three (3) of the fourteen (14) collection sites enrolled subjects for the enrichment phase. Subjects who were experiencing signs and symptoms of respiratory viral infection with a positive PCR test result for Flu A, Flu B and/or RSV within three (3) days prior to enrollment were included in the study.

A total of sixty-nine (69) subjects were enrolled during the enrichment phase of this study, one AN and one NP swab was collected from each of the 69 subjects. One AN swab was excluded from the performance analysis due to a collection site protocol deviation. Sixty-eight (68) AN and 69 NP swabs were included in the performance analysis for the enriched sample cohort. Sixty (60) specimens were tested fresh with the TaqPath COVID-19, Flu A, Flu B, RSV Select Panel and 9 NP swabs and 8 AN swabs specimens were tested after freezing. No invalid results were observed in the enrichment phase for an invalid rate of 0.0% for both AN and NP specimens. The performance comparison of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel and the comparator test results are presented in Table 20.

{50}------------------------------------------------

TargetSwab TypeSpecimenCollectionNumberofSpecimensTruePositiveFalseNegativeTrueNegativeFalsePositivePPA(%)95% Two-SidedConfidenceIntervalNPA(%)95% Two-SidedConfidenceInterval
Flu ANasopharyngealSwabFresh60451a14097.8%88.7% - 99.6%100.0%78.5% - 100.0%
NasopharyngealSwabFrozen92070100.0%34.2% - 100.0%100.0%64.6% - 100.0%
NasopharyngealSwabOverall6947121097.9%89.1% - 99.6%100.0%84.5% - 100.0%
Flu AAnterior NasalSwabFresh60450141100.0%92.1% - 100.0%93.3%70.2% - 98.8%
Anterior NasalSwabFrozen81070100.0%20.7% - 100.0%100.0%64.6% - 100.0%
Anterior NasalSwabOverall68460211b100.0%92.3% - 100.0%95.5%78.2% - 99.2%
Flu BNasopharyngealSwabFresh608150188.9%56.5% - 98.0%98.0%89.7% - 99.7%
NasopharyngealSwabFrozen94050100.0%51.0% - 100.0%100.0%56.6% - 100.0%
NasopharyngealSwabOverall69121c551d92.3%66.7% - 98.6%98.2%90.6% - 99.7%
Flu BAnterior NasalSwabFresh608151088.9%56.5% - 98.0%100.0%93.0% - 100.0%
Anterior NasalSwabFrozen84040100.0%51.0% - 100.0%100.0%51.0% - 100.0%
Anterior NasalSwabOverall68121c55092.3%66.7% - 98.6%100.0%93.5% - 100.0%
RSVNasopharyngealSwabFresh6070521100.0%64.6% - 100.0%98.1%90.1% - 99.7%
NasopharyngealSwabFrozen93042100.0%43.9% - 100.0%66.7%30.0% - 90.3%
NasopharyngealSwabOverall69100563f100.0%72.3% - 100.0%94.9%86.1% - 98.3%
RSVAnterior NasalSwabFresh6070521100.0%64.6% - 100.0%98.1%90.1% - 99.7%
Anterior NasalSwabFrozen83041100.0%43.9% - 100.0%80.0%37.6% - 96.4%
Anterior NasalSwabOverall68100562g100.0%72.3% - 100.0%96.6%88.3% - 99.1%

Table 20: Agreement of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel with the Comparator (Enriched Sample Cohort)

ª Discordant test results based on resolver testing: 1/1 Flu A Negative

b Discordant test results based on resolver testing 1/1 Flu A Negative

& Discordant test results based on resolver testing: 1/1 Flu B Positive

d Discordant test results based on resolver testing: 1/1 Flu B Negative

€ Discordant test results based on resolver testing: 1/1 Flu B Positive

f Discordant test results based on resolver testing: 1/3 RSV Positive and 2/3 RSV Negative

8 Discordant test results based on resolver testing: 1/2 RSV Positive and 1/2 RSV Negative

{51}------------------------------------------------

Reference Ranges/Expected Values

The prospective cohort of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel clinical performance study tested a total of one thousand six hundred twenty (1,620) nasopharyngeal swabs (NP) and one thousand five hundred forty one (1,541) anterior nasal swabs (AN) after exclusions. Samples were collected from 14 geographically diverse collection sites across the U.S. and tested at three qualified testing sites. The stratification of positivity rates (as determined by the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel) per collection site for the prospective samples is shown in Table 21.

{52}------------------------------------------------

CollectionSiteSARS-CoV-2Flu AFlu BRSV
NPANNPANNPANNPAN
Site 114.2% (23/162)13.5% (20/148)0.0% (0/162)0.0% (0/148)0.0% (0/162)0.0% (0/148)0.0% (0/162)0.0% (0/148)
Site 26.7% (9/135)9.5% (12/126)0.0% (0/135)0.0% (0/126)0.0% (0/135)0.0% (0/126)0.7% (1/135)0.8% (1/126)
Site 313.8% (49/356)10.9% (38/348)2.2% (8/356)1.7% (6/348)0.8% (3/356)0.9% (3/348)2.0% (7/356)2.0% (7/348)
Site 411.0% (9/82)18.2% (14/77)0.0% (0/82)0.0% (0/77)0.0% (0/82)0.0% (0/77)1.2% (1/82)1.3% (1/77)
Site 59.0% (17/188)3.9% (7/178)0.0% (0/188)0.0% (0/178)0.0% (0/188)0.0% (0/178)0.0% (0/188)0.0% (0/178)
Site 67.1% (13/182)6.4% (11/171)2.7% (5/182)2.3% (4/171)0.0% (0/182)0.0% (0/171)0.5% (1/182)0.6% (1/171)
Site 721.7% (10/46)15.6% (7/45)10.9% (5/46)6.7% (3/45)15.2% (7/46)15.6% (7/45)2.2% (1/46)2.2% (1/45)
Site88.3% (1/12)8.3% (1/12)0.0% (0/12)0.0% (0/12)0.0% (0/12)0.0% (0/12)0.0% (0/12)0.0% (0/12)
Site 90.0% (0/10)0.0% (0/7)0.0% (0/10)0.0% (0/7)0.0% (0/10)0.0% (0/7)0.0% (0/10)0.0% (0/7)
Site 1013.9% (28/202)14.7% (29/197)7.9% (16/202)8.1% (16/197)12.4% (25/202)12.7% (25/197)6.4% (13/202)5.6% (11/197)
Site 118.0% (6/75)6.6% (5/76)28.0% (21/75)30.3% (23/76)12.0% (9/75)11.8% (9/76)9.3% (7/75)11.8% (9/76)
Site 1218.8% (15/80)18.8% (15/80)1.3% (1/80)2.5% (2/80)1.3% (1/80)0.0% (0/80)1.3% (1/80)0.0% (0/80)
Site 1322.7% (5/22)16.7% (3/18)9.1% (2/22)11.1% (2/18)9.1% (2/22)11.1% (2/18)4.5% (1/22)5.6% (1/18)
Site 1413.2% (9/68)15.5% (9/58)14.7% (10/68)8.6% (5/58)5.9% (4/68)5.2% (3/58)1.5% (1/68)5.2% (3/58)
All12.0%(194/1620)11.1%(171/1541)4.2%(68/1620)4.0%(61/1541)3.1%(51/1620)3.2%(49/1541)2.1%(34/1620)2.3%(35/1541)

Table 21: TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel – Expected Values by Specimen Collection Site for NP and AN Swab Specimens

{53}------------------------------------------------

Conclusion

The results of the analytical and clinical performance studies summarized above demonstrated that the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is substantially equivalent to the predicate device.

§ 866.3981 Device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test.

(a)
Identification. A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
e.g., BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
e.g., when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
e.g., regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
e.g., saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
i.e., H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.