K Number
DEN200031

Validate with FDA (Live)

Date Cleared
2021-03-17

(302 days)

Product Code
Regulation Number
866.3981
Type
Direct
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioFire Respiratory Panel 2.1 (RP2.1) is a PCR-based multiplexed nucleic acid test intended for use with the BioFire FilmArray 2.0 or BioFire FilmArray Torch systems for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections, including COVID-19.

The following organism types and subtypes are identified using the BioFire RP2.1:

  • Adenovirus, ●
  • Coronavirus 229E. ●
  • Coronavirus HKU1, ●
  • Coronavirus NL63, ●
  • Coronavirus OC43.
  • Severe Acute Respiratory Syndrome Coronavirus (SARS-CoV-2), ●
  • Human Metapneumovirus.
  • Human Rhinovirus/Enterovirus,
  • Influenza A, including subtypes H1, H1-2009, and H3, ●
  • Influenza B,
  • Parainfluenza Virus 1,
  • Parainfluenza Virus 2,
  • Parainfluenza Virus 3, ●
  • Parainfluenza Virus 4. ●
  • Respiratory Syncytial Virus, ●
  • Bordetella parapertussis (IS1001),
  • Bordetella pertussis (ptxP), ●
  • Chlamydia pneumoniae, and ●
  • Mycoplasma pneumoniae ●

Nucleic acids from the respiratory viral and bacterial organisms identified by this test are generally detectable in NPS specimens during the acute phase of infection. The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of respiratory infection is indicative of the presence of the identified microorganism and aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by an NPS specimen. Positive results do not rule out coinfection with other organisms. The agent(s) detected by the BioFire RP2.1 may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

Device Description

The BioFire Respiratory Panel 2.1 is designed to simultaneously identify 22 different potential pathogens of the respiratory tract infection, including the novel coronavirus Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2), from a single NPS specimen in transport medium or saline. BioFire RP2.1 is compatible with BioFire's PCR-based in vitro diagnostic BioFire FilmArray 2.0 and BioFire FilmArray Torch systems for infectious disease testing. A specific software module (i.e., BioFire RP2.1 Pouch Module Software) is used to perform BioFire RP2.1 testing on these systems.

The RP2.1 reagent kit contains all the materials required to complete tests and includes the RP2.1 pouch, hydration solution, sample buffer, and sample handling components such as transfer pipettes. The RP2.1 pouches are used to test patient samples and is a closed-system disposable that stores all the necessary reagents for sample preparation reverse transcription. polymerase chain reaction (PCR), and detection in order to isolate, amplify, and detect nucleic acid from multiple pathogens within a single NPS specimen. The rigid plastic component ("fitment") of the pouch contains reagents in freeze-dried form. The flexible plastic portion of the pouch is divided into discrete segments ("blisters") where the required chemical processes are carried out. After sample collection, the user injections hydration solution and sample combined with BioFire Sample Buffer into the pouch, places the pouch into a FilmArray instrument, and starts the run. All other operations are automated.

The FilmArray instruments (FilmArray 2.0 and FilmArray Torch systems) interact with the pouch mechanically, thermally, and optically to drive the multi-step chemical process required for purification and detection of specific nucleic acid targets from the patient sample. FilmArray instruments follow a protocol defined in the BioFire RP2.1 Pouch Module Software that is downloaded from the host computer prior to runtime. The instrument protocol defines the specific sequence of the testing process, including the times and temperatures, as the instrument performs bead-based extraction/isolation/purification of nucleic acids, performs reverse transcription and a 2-stage nested PCR reaction, executes DNA melt and fluorescent signal detection, and monitors system performance in real time, and communicates results and errors to the user via software. The primary difference between the FilmArray 2.0 and FilmArray Torch systems is the external configuration of multiple modules in a system. Up to eight FilmArray 2.0 modules can be connected to one computer and pouch loading station, while up to 12 FilmArray Torch modules can be connected to one system base in a vertical stack to a computer and pouch loading station. In addition, the pouches are front-loaded via an automated mechanism for the Torch system whereas the pouches are manually inserted, removed, and there is pouch and lid sensing in the FilmArray 2.0.

Once a test run is completed, the software automatically interprets the results and displays a test report. The report can be printed and/or saved as a file. The test report is a single page containing three sections: Run Summary, Result Summary, and Run Details. An additional section, Change Summary, is present in specific situations. The overall layout of the report was previously described in the BioFire RP2 510(k) [K170604] and remains unchanged for the BioFire RP2.1.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study proving the BioFire Respiratory Panel 2.1 (RP2.1) meets those criteria, based on the provided text.

Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by the successful performance demonstrated in the various analytical and clinical studies. For qualitative tests like this, the key performance metrics are sensitivity (Positive Percent Agreement - PPA) and specificity (Negative Percent Agreement - NPA). The reproducibility study also defines implicit acceptance for consistent detection.

Table of Acceptance Criteria and Reported Device Performance (Summary for SARS-CoV-2 - a key new analyte)

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (SARS-CoV-2)
Prospective Clinical Study
PPA (Sensitivity)High agreement (e.g., >95%)98.4% (61/62) with 95% CI: 91.4-99.7%
NPA (Specificity)High agreement (e.g., >95%)98.9% (457/462) with 95% CI: 97.5-99.5%
Retrospective Clinical Study
PPA (Sensitivity)High agreement (e.g., >95%)98.0% (48/49) with 95% CI: 89.3-99.6%
NPA (Specificity)High agreement (e.g., >95%)100% (49/49) with 95% CI: 92.7-100%
Contrived Clinical Specimen Study
PPA (Sensitivity)High agreement (e.g., 100%)100% (50/50) with 95% CI: 92.9-100%
NPA (Specificity)High agreement (e.g., 100%)100% (10/10) with 95% CI: 72.2-100%
Analytical LoD (SARS-CoV-2 Inactivated)≥95% detection at 1x LoD, <95% at 0.1x LoD100% (20/20) at 1x LoD, 25% (5/20) at 0.1x LoD (5.0E+02 copies/mL LoD)
Analytical LoD (SARS-CoV-2 Infectious)≥95% detection at 1x LoD, <95% at 0.1x LoD100% (20/20) at 1x LoD, 55% (11/20) at 0.1x LoD (1.6E+02 copies/mL LoD)
Reproducibility (SARS-CoV-2)Consistent detection (e.g., 90-100%)Moderate Positive (3x LoD): 100% (60/60) Low Positive (1x LoD): 96.7% (58/60)
Cross-Reactivity (SARS-CoV-2 assays)No unexpected cross-reactivityNone with common respiratory flora/pathogens. Predicted cross-reactivity with closely related bat/pangolin coronaviruses not isolated from humans.

Study Details

2. Sample Size and Data Provenance

  • Test Set Sample Sizes:

    • Prospective Clinical Study: 524 valid NPS specimens.
    • Retrospective Clinical Study: 50 SARS-CoV-2 positive archived clinical specimens + 50 NPS specimens collected before Dec 2019 (expected negative for SARS-CoV-2) = 100 total specimens (98 evaluable due to instrument errors).
    • Contrived Clinical Specimen Study: 50 contrived clinical specimens (spiked with inactivated SARS-CoV-2) + 10 non-spiked specimens = 60 total specimens.
    • Clinical Comparison Study (RP2 vs RP2.1): 210 archived clinical NPS specimens.
    • Analytical Reproducibility: For each analyte tested, 120 valid runs per sample (total 360 valid runs for the entire study, across multiple analytes and concentrations).
    • Analytical LoD (SARS-CoV-2): 20 replicates each for 1x LoD and 0.1x LoD for inactivated virus; 20 replicates each for 1x LoD and 0.1x LoD for infectious virus.
  • Data Provenance:

    • Prospective Clinical Study: Prospectively collected NPS specimens from three collection sites in the U.S. (fresh/recent collection).
    • Retrospective Clinical Study: Retrospective, archived clinical specimens from March and April 2020 (U.S.) and NPS specimens collected before December 2019 (U.S. from three geographically distinct laboratories).
    • Contrived Clinical Specimen Study: Unique NPS specimens collected before December 2019 (expected negative for SARS-CoV-2).
    • Clinical Comparison Study: Archived nasopharyngeal swab (NPS) specimens collected prior to December 2019, obtained from BioFire specimen repository and external sources.
    • Analytical Studies (Reproducibility, LoD, Reactivity, Specificity): Primarily used contrived samples with known concentrations of isolates/strains. Provenance of source materials (isolates) is provided (e.g., ATCC, Zeptometrix, BEI, WHO International Standard).

3. Number of Experts and Qualifications for Ground Truth

  • The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the clinical test sets.
  • Ground Truth Determination for Clinical Studies:
    • Prospective Clinical Study (SARS-CoV-2): Ground truth was established using a composite comparator method of three U.S. FDA EUA tests. "Concordance for two out of three of the EUA tests were considered the final result for the comparator." This implies the "experts" were the results of these FDA EUA assays.
    • Retrospective Clinical Study (SARS-CoV-2): Ground truth (known positive/negative status) was based on previous characterization using the respective Emergency Use Authorization (EUA) assay employed at the collection site. This is also a comparator method.
    • Contrived Clinical Specimen Study (SARS-CoV-2): Ground truth was based on the known presence and concentration of the spiked inactivated SARS-CoV-2 isolate and the known SARS-CoV-2 negative status of the base clinical specimens. This is a known sample composition ground truth.
    • Clinical Comparison Study: Ground truth for analytes was based on results from the BioFire RP2 panel (for positive agreement) and the expected negative status for SARS-CoV-2 (for specificity).

4. Adjudication Method for the Test Set

  • For the Prospective Clinical Study, the "composite comparator" method described an adjudication-like process:
    • Rule 1: Pos/Pos/Any -> Positive (if 2 of 3 EUA tests are positive, regardless of the third)
    • Rule 2: Neg/Neg/Any -> Negative (if 2 of 3 EUA tests are negative, regardless of the third)
    • Rule 3: Pos/Neg/Inv -> Specimen excluded (no clear consensus)
    • Rule 4: Inv/Inv/Any -> Specimen excluded (no clear consensus)
  • No explicit human expert adjudication method is described. The adjudication relies on the consensus of the comparator FDA EUA tests.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done or reported.
  • This device is an automated in vitro diagnostic (IVD) test, not an AI-assisted diagnostic tool for human readers. Therefore, there's no "human readers improve with AI vs without AI assistance" effect size to report. The comparison studies are on the device's performance against comparator devices or known ground truth.

6. Standalone (Algorithm Only) Performance

  • Yes, standalone performance was done. The entire evaluation described in the "Performance Characteristics" section (analytical performance, comparison studies, and clinical studies) represents the standalone performance of the BioFire RP2.1 device, as it is an automated system that provides a qualitative result without human interpretation of raw signals. The software within the FilmArray instruments interprets the melt curve data and applies interpretation rules to determine the final test results.

7. Type of Ground Truth Used

  • For Clinical Studies (SARS-CoV-2):
    • Prospective Clinical Study: Comparator Gold Standard (composite of three U.S. FDA EUA tests).
    • Retrospective Clinical Study: Comparator Gold Standard (previous EUA assay results at collection sites).
    • Contrived Clinical Specimen Study: Known Sample Composition / Definitive Ground Truth (spiked inactivated virus at known concentrations, or known negative specimens).
  • For Analytical Studies (e.g., LoD, Reactivity, Specificity): Known Sample Composition / Definitive Ground Truth (contrived samples with specific isolates/strains at known concentrations).

8. Sample Size for the Training Set

  • The document does not explicitly state the sample size for the training set.
  • As an in vitro diagnostic (IVD) device, the development process typically involves internal analytical characterization and optimization (which can be considered analogous to "training" for a traditional assay, albeit not a machine learning model in the modern sense), followed by formal verification and validation studies. The analytical performance data (LoD, reactivity, specificity, etc.) are gathered on well-characterized samples to confirm the assay's performance.

9. How the Ground Truth for the Training Set Was Established

  • The concept of a "training set" with ground truth in the context of this traditional IVD device (not an AI/ML model) refers to the development and optimization of the assays themselves.
  • The ground truth for these developmental/training phases would be established through a combination of:
    • Defined concentrations of specific viral/bacterial isolates or nucleic acids: This is seen in the LoD and analytical reactivity studies where concentrations are precisely measured (e.g., copies/mL, TCID50/mL, CFU/mL) using established quantification methods (e.g., ddPCR, qPCR, plate counting).
    • Known sample compositions: Creating samples with specific organisms present or absent at controlled levels.
    • Previous BioFire RP2 data: The document notes that the RP2.1 builds upon the previous RP2 panel, meaning much of the "training" (i.e., assay design, optimization of reaction conditions, primer selection for common pathogens) for the shared analytes was likely carried over and validated.
    • In silico analysis: Used for SARS-CoV-2 primer design and reactivity prediction against sequence databases (NCBI, GISAID) as a form of "training" or initial validation for robustness against known variants. This informs assay design.

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR BIOFIRE RESPIRATORY PANEL 2.1 DECISION SUMMARY

A. De Novo Number:

DEN200031

B. Purpose for Submission:

De Novo request for evaluation of automatic class III designation for the BioFire Respiratory Panel 2.1 (RP2.1).

C. Measurands:

The assay detects and identifies nucleic acids of the following respiratory pathogens: Severe Acute Respiratory Syndrome Coronavirus (SARS-CoV-2), Adenovirus, Coronavirus 229E, Coronavirus HKU1, Coronavirus NL63, Coronavirus OC43, Human Metapneumovirus, Human Rhinovirus/Enterovirus, Influenza A, including subtypes H1, H1-2009, and H3, Influenza B, Parainfluenza Virus 1, Parainfluenza Virus 2, Parainfluenza Virus 3, Parainfluenza Virus 4, Respiratory Syncytial Virus, Bordetella parapertussis (IS1001), Bordetella pertussis (ptxP), Chlamydia pneumoniae, and Mycoplasma pneumoniae.

D. Type of Test:

A multiplexed nucleic acid test intended for use with the BioFire FilmArray Torch systems for the simultaneous qualitative detection of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections, including COVID-19.

E. Applicant:

BioFire Diagnostics, LLC

F. Proprietary and Established Names:

BioFire Respiratory Panel 2.1 (RP2.1)

G. Regulatory Information:

    1. Regulation section:
      21 CFR 866.3981
    1. Classification:
      Class II (special controls)

{1}------------------------------------------------

3. Product code(s):

QOF

    1. Panel:
      Microbiology (83)

H. Indications for Use:

    1. Indication(s) for use:
      The BioFire Respiratory Panel 2.1 (RP2.1) is a PCR-based multiplexed nucleic acid test intended for use with the BioFire FilmArray 2.0 or BioFire FilmArray Torch systems for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections, including COVID-19.

The following organism types and subtypes are identified using the BioFire RP2.1:

  • Adenovirus, ●
  • Coronavirus 229E. ●
  • Coronavirus HKU1, ●
  • Coronavirus NL63, ●
  • Coronavirus OC43.
  • Severe Acute Respiratory Syndrome Coronavirus (SARS-CoV-2), ●
  • Human Metapneumovirus.
  • Human Rhinovirus/Enterovirus,
  • Influenza A, including subtypes H1, H1-2009, and H3, ●
  • Influenza B,
  • Parainfluenza Virus 1,
  • Parainfluenza Virus 2,
  • Parainfluenza Virus 3, ●
  • Parainfluenza Virus 4. ●
  • Respiratory Syncytial Virus, ●
  • Bordetella parapertussis (IS1001),
  • Bordetella pertussis (ptxP), ●
  • Chlamydia pneumoniae, and ●
  • Mycoplasma pneumoniae ●

Nucleic acids from the respiratory viral and bacterial organisms identified by this test are generally detectable in NPS specimens during the acute phase of infection. The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of respiratory infection is indicative of the presence of the identified microorganism and aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test

{2}------------------------------------------------

should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by an NPS specimen. Positive results do not rule out coinfection with other organisms. The agent(s) detected by the BioFire RP2.1 may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

2. Special conditions for use statement(s):

For prescription use only.

For in vitro diagnostic use only.

    1. Special instrument requirements:
      FilmArray Respiratory Panel 2.1 (RP2.1) is performed on the FilmArray 2.0 or the FilmArray Torch systems.

I. Device Description:

The BioFire Respiratory Panel 2.1 is designed to simultaneously identify 22 different potential pathogens of the respiratory tract infection, including the novel coronavirus Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2), from a single NPS specimen in transport medium or saline. BioFire RP2.1 is compatible with BioFire's PCR-based in vitro diagnostic BioFire FilmArray 2.0 and BioFire FilmArray Torch systems for infectious disease testing. A specific software module (i.e., BioFire RP2.1 Pouch Module Software) is used to perform BioFire RP2.1 testing on these systems.

The RP2.1 reagent kit contains all the materials required to complete tests and includes the RP2.1 pouch, hydration solution, sample buffer, and sample handling components such as transfer pipettes. The RP2.1 pouches are used to test patient samples and is a closed-system disposable that stores all the necessary reagents for sample preparation reverse transcription. polymerase chain reaction (PCR), and detection in order to isolate, amplify, and detect nucleic acid from multiple pathogens within a single NPS specimen. The rigid plastic component ("fitment") of the pouch contains reagents in freeze-dried form. The flexible plastic portion of the pouch is divided into discrete segments ("blisters") where the required chemical processes are carried out. After sample collection, the user injections hydration solution and sample combined with BioFire Sample Buffer into the pouch, places the pouch into a FilmArray instrument, and starts the run. All other operations are automated.

The FilmArray instruments (FilmArray 2.0 and FilmArray Torch systems) interact with the pouch mechanically, thermally, and optically to drive the multi-step chemical process

{3}------------------------------------------------

required for purification and detection of specific nucleic acid targets from the patient sample. FilmArray instruments follow a protocol defined in the BioFire RP2.1 Pouch Module Software that is downloaded from the host computer prior to runtime. The instrument protocol defines the specific sequence of the testing process, including the times and temperatures, as the instrument performs bead-based extraction/isolation/purification of nucleic acids, performs reverse transcription and a 2-stage nested PCR reaction, executes DNA melt and fluorescent signal detection, and monitors system performance in real time, and communicates results and errors to the user via software. The primary difference between the FilmArray 2.0 and FilmArray Torch systems is the external configuration of multiple modules in a system. Up to eight FilmArray 2.0 modules can be connected to one computer and pouch loading station, while up to 12 FilmArray Torch modules can be connected to one system base in a vertical stack to a computer and pouch loading station. In addition, the pouches are front-loaded via an automated mechanism for the Torch system whereas the pouches are manually inserted, removed, and there is pouch and lid sensing in the FilmArray 2.0.

Once a test run is completed, the software automatically interprets the results and displays a test report. The report can be printed and/or saved as a file. The test report is a single page containing three sections: Run Summary, Result Summary, and Run Details. An additional section, Change Summary, is present in specific situations. The overall layout of the report was previously described in the BioFire RP2 510(k) [K170604] and remains unchanged for the BioFire RP2.1-

BioFire® Respiratory Panel 2.1
www.BioFireDx.com
Run Summary
Sample ID:RP2.1exampleRun Date:04 April 20205:21 PM
Detected:Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)Controls:Passed
Equivocal:➔Influenza A
Result Summary
Viruses
Not DetectedAdenovirus
Not DetectedCoronavirus 229E
Not DetectedCoronavirus HKU1
Not DetectedCoronavirus NL63
Not DetectedCoronavirus OC43
✓ DetectedSevere Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)
Not DetectedHuman Metapneumovirus
Not DetectedHuman Rhinovirus/Enterovirus
➔ EquivocalInfluenza A
Not DetectedInfluenza B
Not DetectedParainfluenza Virus 1
Not DetectedParainfluenza Virus 2
Not DetectedParainfluenza Virus 3
Not DetectedParainfluenza Virus 4
Not DetectedRespiratory Syncytial Virus
Bacteria
Not DetectedBordetella parapertussis (IS1001)
Not DetectedBordetella pertussis (ptxP)
Not DetectedChlamydia pneumoniae
Not DetectedMycoplasma pneumoniae
Run Details
Pouch:RP2.1 v1.0Protocol:NPS2 v3.2
Run Status:CompletedOperator:JDoe
Serial No.:01234567Instrument:TM8CCF3

Test results for the organisms included in the BioFire RP2.1 are provided in two locations on the report. The Result Summary section provides a complete list of the test results. Possible results include "Detected," "Not Detected," "Equivocal," and "Invalid." Positive (Detected) and Equivocal results are also displayed in the Run Summary section. The following table

{4}------------------------------------------------

provides an explanation for each interpretation and any follow-up necessary to obtain a final result.

ResultExplanationAction
DetectedaThe run was successfullycompleted ANDThe pouch controls were successful (Passed)ANDThe assay(s) for the organism werePOSITIVE(i.e., met the requirements for a positiveresult)Report results.
Not DetectedThe run was successfullycompleted ANDThe pouch controls were successful (Passed)ANDThe assay(s) for the organism were NEGATIVE(i.e., did not meet the requirements for a positiveresult)Report results.
EquivocalThe run was successfullycompleted ANDThe pouch controls were successful (Passed)ANDThe combination of positive and negative assay resultsfor Influenza A were inconclusiveRetest the original sample ONCE and reportthe result of the retest.
InvalidThe pouch controls were not successful(Failed)ORThe run was not successful (Run Status displayed as:Aborted, Incomplete, Instrument Error, or Software Error)See Interpretation of control fields on theBioFire RP2.1 test report for instruction.

Table 1. Explanation of Reported Results and Required Actions

a If four or more organisms are detected in a specimen, retesting is recommended to confirm the polymicrobial result.

For most organisms detected by the BioFire RP2.1. the organism is reported as Detected if a single corresponding assay is positive. For example, Human Metapneumovirus will have a test report result of "Human Metapneumovirus Detected" if at least two of the three replicates of the one Human Metapneumovirus assay (hMPV) have similar positive melt peaks with Tm values that are within the assay-specific Tm range.

In contrast, the test results for SARS-CoV-2, Adenovirus, and Influenza A depend on the interpretation of results from more than one assay. Interpretation results for all organisms detected by the BioFire RP2.1, except for SARS-CoV-2, are previously described in the BioFire RP2 510(k) submission [K170604] and remain unchanged for the BioFire RP2.1.

The BioFire RP2.1 pouch contains two different assays for the detection of the SARS-CoV-2 microorganism. The assays each target a spike protein (S) gene and membrane protein (M) gene respectively. The BioFire FilmArray software interprets each of these assays independently and the results are combined as a final test result for the virus. An assay is called positive if at least two of the three replicates within the pouch have similar positive melt peaks with Tm values that are within the assay-specific Tm range. If either one or both of the assays is called

{5}------------------------------------------------

positive, the test report will show Severe Acute Respiratory Coronavirus 2 (SARS-CoV-2) as Detected. If all assays are called negative. the test report will be Severe Acute Respiratory Coronavirus 2 (SARS-CoV-2) Not Detected.

J. Standard/Guidance Document Referenced (if applicable):

General

  • Guidance for Clinical Laboratories, Commercial Manufactures, and FDA Staff Policy for ● Coronavirus Disease-
  • 2019 Tests During the Public Health Emergency (2020) ●
  • . GHTF, Clinical Evidence for IVD Medical Devices - Clinical Performance Studies for In Vitro Diagnostic Medical
  • . Devices (November 2012)
  • . WMA Declaration of Helsinki, Ethical Principles for Medical Research Involving Human Subjects
  • . 2017/746 Regulation EU 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU
  • 2016/679 GDPR, Regulation EU 2016/679 of the European Parliament and of the Council of 27 April 2016 on the protection of natural persons with regard to the processing of personal data and on the free movement of such data, and repealing Directive 95/46/EC (General Data Protection Regulation)
  • . Guidance for Industry - Part 11. Electronic Records: Electronic Signatures - Scone and Application (August 2003)
  • Guidance for Industry Computerized Systems Used in Clinical Investigations (May 2007) .
  • . Guidance for Industry – Oversight of Clinical Investigations – A Risk-Based Approach to Monitoring (August 2013)
  • Guidance for Industry - Electronic Source Data in Clinical Investigations (September 2013)
  • . Guidance for IRBs, Clinical Investigators, and Sponsors - Informed Consent Information Sheet (July 2014)
  • FDA Draft Guidance Use of Electronic Records and Electronic Signatures in Clinical . Investigations Under 21 CFR Part 11 - Questions and Answers (June 2017)
  • . Guidance for Industry and FDA Staff – Acceptance of Clinical Data to Support Medical Device Applications and Submissions - Frequently Asked Questions (February 2018)
  • Guidance for Sponsors, Investigators, and IRBs – Impact of Certain Provisions of the Revised Common Rule on FDA-Regulated Clinical Investigations (October 2018)
  • ICH E6(R1) Guideline for Good Clinical Practice E6(R1) June 1996 ●
  • ICH E6(R2) Integrated Addendum to ICH E6(R1): Guideline for Good Clinical Practice E6(R2) ● – November 2016
  • . Guidance for Industry and Food and Drug Administration Staff – Highly Multiplexed Microbiological/Medical Countermeasure In Vitro Nucleic Acid Based Diagnostic Devices, (August 27, 2014)
  • . Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests, FDA Guidance Document (March 13, 2007)
  • User Protocol for Evaluation of Qualitative Test Performance, Clinical and Laboratory Standards ● Institute (CLSI) Approved Guideline - Second Edition, EP12-A2 (January 2008)
  • Molecular Diagnostic Methods for Infectious Diseases, Clinical and Laboratory Standards ● Institute (CLSI) Proposed Guideline, MM3-P2 (February 2006)

{6}------------------------------------------------

  • . Interference Testing in Clinical Chemistry, 3rd Edition, Clinical and Laboratory Standards Institute (CLSI) Approved Guideline, EP07 (April 2018).
  • . CLSI EP25-A, 'Evaluation of stability of in vitro diagnostic reagents; Approved Guidelines'.
  • Guidance for Sponsors. Institutional Review Boards. Clinical Investigators and FDA Staff . Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable, (April 2006)

Software

  • Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
  • . Off-The-Shelf Software Use in Medical Devices, Guidance for Industry and Food and Drug Administration Staff (September 27, 2019)
  • . General Principle of Software Validation; Final Guidance for Industry and FDA Staff (January 11, 2002)
  • . Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. Guidance for Industry and Food and Drug Administration Staff (October 2, 2014)

Labeling

  • Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use, FDA Guidance Document (November 30, 2004)
  • . Guidance for Industry and FDA on Alternative to Certain Prescription Device Labeling Requirements (January 1, 2000)

FDA-recognized Standards

  • ISO 14971:2007 'Medical devices – Application of risk management to medical devices'
  • EN 62366:2008/IEC 62366-1:2015, 'Medical device Application of usability engineering to . medical devices'
  • . ISO 62304:2006, 'Medical device software – Software life-cycle processes' – IEC 62304:2006, November 27, 2008
  • . ISO 15223-1:2012, 'Medical Devices – Symbols to be used with medical device labels, labeling and information to be supplied - Part 1: General requirements'

Non-recognized Standards

  • ISO 13485:2016/EN ISO 13485:2016, 'Medical devices Quality Management System ● Requirements for regulatory purposes'
  • ISO 20916:2019, 'In vitro diagnostic medical devices – Clinical performance studies using specimens from human subjects - Good study practice'
  • . EN 13612:2002. Performance evaluation of in vitro diagnostic medical devices (European Commission)
  • EN ISO 18113-1:2011, 'In vitro diagnostic medical devices - Information supplied by the manufacturer (labeling) - Part 1: Terms, definition and general requirements'
  • . EN ISO 18113-2:2011, 'In vitro diagnostic medical devices – Information supplied by the manufacturer (labeling) - Part 2: In vitro diagnostic reagents for professional use'
  • EN ISO 23640:2015, 'In vitro diagnostic medical devices – Evaluation of stability of in vitro diagnostic reagents'

{7}------------------------------------------------

K. Test Principle:

The BioFire RP2.1 test takes approximately 2 minutes of hands-on-time from the point of collection to the initiation of the automated test. Once the test is initiated, a test result is produced in approximately 45 minutes.

During a test. the FilmArray instrument. software, and pouch work together to generate assay results. The test works through automated sample processing and nested multiplex nucleic acid amplification (including reverse transcription as appropriate) followed by highresolution melt analysis to confirm the identity of the amplified product. The basic sequence of actions and their associated instrument functions are outlined in Figure 1-

Image /page/7/Figure/3 description: The image shows a diagram of a multi-step process. The process starts with "Sample Lysis", which involves bead beating, reagent addition, and liquid movement. The next step is "Nucleic Acid Isolation", which involves reagent addition, liquid handling, and magnetic beads. The process continues with "Reverse Transcription and 1st Stage PCR", which involves temperature control, reagent addition, and liquid handling, and ends with "2nd Stage PCR and Detection", which involves temperature control, LED control, and camera control.

Figure 1. Basic steps performed during BioFire RP2.1 testing

The pouch contains all the necessary PCR reagents and is where samples are automatically processed to generate test results. The instrument communicates with the host computer and the FilmArray software. The software provides instructions to the instrument to control the various test steps. The instrument drives the testing process by applying mechanical force on the pouch exterior to actuate liquid movement to various compartments and to seal or block off flow in particular channels. The instrument also thermally interacts with the pouch to perform the subsequent 2-stage nested PCR reactions.

Optical systems on-board the instrument that include a LED and digital camera allow illumination and recording of fluorescence generated in the second stage PCR. The fluorescence signal generated during DNA melting is automatically analyzed by the FilmArray software from replicate wells of each assay for the detection of amplicons with a specific Tm. The detection denotes the presence of specific bacterial or viral targets.

The BioFire RP2.1 pouch contains the same sample preparation and PCR reaction chemistry as the previously cleared BioFire FilmArray Respiratory Panel 2 (RP2) (K170604; cleared for use on both FilmArray 2.0 and FilmArray Torch Systems). The PCR1 primer multiplex is also the same, with the addition of SARS-CoV-2 primers. The PCR2 array is similar except with the additions and minor reconfiguration of wells to accommodate the two SARS-CoV-2 assays. In addition, the instrument protocol and the analysis parameters in the panel-specific pouch module are the same as for FilmArray RP2, with the additional analysis of the SARS-CoV-2 assays.

{8}------------------------------------------------

The BioFire RP2.1 procedure occurs in six steps below. This simple procedure minimizes specimen manipulation and reduces operator error.

  • . Step 1 - Place pouch into the FilmArray Pouch Loading Station.
  • Step 2 Hydrate pouch using a blue Hydration Injection Vial. ●
  • Step 3 Prepare sample in the red Sample Injection Vial: ●
    • o Dispense the Sample Buffer tube into the Sample Injection Vial.
    • With a transfer pipette, draw the NPS in transport media or saline sample to O the third line, then add it to the Sample Injection Vial.
    • o Mix by inversion.
  • Step 4 Load sample mix in pouch. .
  • Step 5 Insert pouch into the instrument. ●
  • Step 6 Enter sample information and start the run. The BioFire RP2.1 protocol will ● be automatically selected upon scanning the pouch barcode.

The FilmArray software uses the following steps to interpret the melt curve data generated from each FilmArray RP2.1 assay-

  • Analysis of Melt Curves ●
    • o The FilmArray RP2.1 Melt Detector first performs basic calculations on the melt data to determine if a PCR reaction occurred in each well. If the melt profile indicates that a PCR product is present, then the analysis software calculates one or two Tm values, depending on the number of melt curves present in the data, and the Tm values are compared against an expected melt range for the associate assay. If the software determines that the melt is positive and the melt curve falls inside the assay's specific melt range, thent he curve is called positive. If the software determines that the melt is negative or that it is not in the appropriate range, then the curve is called negative.
  • Analysis of Replicates ●
    • o The analysis software evaluates the replicates for each assay (target and control) to determine if the assay is positive or negative. For a positive, at least two of the three wells associated with an assay must have a positive melt curve and the Tm for the positive curves must be similar (i.e., within 1°C). Assays that do not meet these criteria are called negative.
  • Analysis of Controls ●
    • Results for control assays are compared to their expected values and are o reported as "Passed", "Failed" or "Invalid". Passed control result is for successful run completion AND both pouch controls were successful. Failed result is when the run was successfully completed BUT at least one of the pouch controls (RNA Process Control and/or PCR2 Control) failed. If the instrument detects an out-of-specification condition or a significant error, it will automatically abort the run. If this happens or if user aborts the run, the control result will display "Invalid" and all results in the Result Summary of the report will also be displayed as "Invalid." A Run Status indicating "Incomplete", "Aborted", "Software Error", or "Instrument Error" will be reported to the user and the operator is asked to consult with the manual for

{9}------------------------------------------------

specific instructions on resolving the error. The test should be repeated once error is corrected.

  • Interpretation of Assay Results ●
    • Once the results for the individual assays are determined, the software applies o interpretation rules to determine the final test results. For most organisms detected by the BioFire RP2.1, the organism is reported as Detected if a single corresponding assay is positive. The BioFire RP2.1 also includes test results for organisms (i.e., SARS-CoV-2, Adenovirus, and Influenza A) that depend on interpretation of results from more than one assay. See the Interpretation of Results section for more information on interpreting these test results.

L. Performance Characteristics (if/when applicable):

    1. Analytical performance:
    • a. Precision/Reproducibility:

A multi-variable study was performed to evaluate the reproducibility of BioFire RP2.1 analyte detection on FilmArray 2.0 and FilmArray Torch systems. This study was additive to the reproducibility evaluation performed for the BioFire RP2 device. with overlapping data for certain analytes to bridge results from the two panels and collect data for select analytes including the newly added SARS-CoV-2.

Contrived samples were used in this study to evaluate variability in between run, system, site, day, or lot. Three samples were prepared in a matrix of viral transport medium (Table 2) and data were collected representing a negative (no analyte) and those containing analytes at low positive (1x LoD) or moderate positive (3x LoD) concentrations. The positive samples included inactivated SARS-CoV-2, Coronavirus NL63, Influenza A H1-2009, and three analytes that had been previously evaluated for the BioFire RP2 reproducibility study (i.e., Adenovirus, Bordetella parapertussis (IS1001) and Respiratory Syncytial Virus).

Each sample was tested repeatedly in three (3) different testing sites over five days by different operators (at least two per site), on different systems (60 per system) and modules, using three different reagent kit lots. Twenty replicates per sample were tested at each site on both FilmArray systems for a total of 120 valid runs per sample and 360 valid runs in total for the entire study. Reproducibility of analyte detection was assessed as percent agreement with the expected Detected and Not Detected results for the positive and negative samples.

The performance of the FilmArray systems and BioFire RP2.1 Controls are summarized as follow. Valid results were obtained in 361 of the 363 runs that were initiated (361/363, 99.4%). There were 181 and 182 runs initiated on the FilmArray 2.0 and FilmArray Torch systems, respectively. There was one instrument error (FilmArray 2.0) and one aborted run (FilmArray Torch). This showed that performance of the controls was reproducible (no control failures) and valid results were obtained for all completed runs.

{10}------------------------------------------------

Reproducibility data for each BioFire RP2.1 analyte are summarized in Table 2. Results are organized by system type (i.e., FilmArray 2.0 or Torch), test site (Site A, B. C), and all sites/systems with the corresponding 95% confidence interval. The summary data are presented as a combination of results collected for reproducibility studies with the BioFire RP2.1 (gray highlight) and the previous RP2 devices.

Table 2. Reproducibility of Detection Results for BioFire RP2.1 Analytes

Highlighted data were collected with the BioFire RP2.1. Non-highted data was collected with the BioFire FilmArray RP2. The same number of replicates (120) were tested per sample on both panels, but testing was distributed differently between sites and systems.

Analyte(Isolate SourceID)ConcentrationTestedFilmArray 2.0FilmArray Torch
Site ASite BSite CSystemTotalSite ASite BSite CSystemTotal
Adenovirusa(NIBSC16/324)WHOInternationalStandardNegative(no analyte)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
ModeratePositive(3x LoD)$9.0E+03$IU/mL20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Low Positive(1x LoD)$3.0E+03$IU/mL20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Coronavirus229ENegative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
CoronavirusHKU1Negative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
CoronavirusNL63(BEI NR-470)Negative(no analyte)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
ModeratePositive(3x LoD)$7.5E-01$TCID50/mL(1.6E+02copies/mL)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Low Positive(1x LoD)$2.5E-01$TCID50/mL(5.4E+01copies/mL)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
CoronavirusOC43Negative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
ModeratePositive(3x LoD)$9.0E+01$TCID50/mL-29/30(96.7%)30/30(100%)59/60(98.3%)29/30(96.7%)-29/30(96.7%)58/60(96.7%)
Coronavirus OC43(ATCCVR-759)Low Positive-30/30(100%)27/30(90.0%)57/60(95.0%)30/30(100%)-30/30(100%)60/60(100%)

{11}------------------------------------------------

Agreement with Expected Result
Analyte(Isolate SourceConcentrationFilm Array 2.0FilmArray Torch
ID)TestedSite ASite BSite CSystemTotalSite ASite BSite CSystemTotal
(1× LoD)3.0E+01TCID50/mL
man MetapneumoviHumanMetapneumovirusNegative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
HumanMetapneumovirus(ZeptomeModeratePositive(3× LoD)3.0E+01TCID50/mL-30/30(100%)30/30(100%)60/60(100%)30/30(100%)-30/30(100%)60/60(100%)
trix0810161CF)Low Positive(1× LoD)1.0E+01TCID50/mL-28/30(93.3%)30/30(100%)58/60(96.7%)30/30(100%)-30/30(100%)60/60(100%)
nan Rhinovirus/EnteroviHumanRhinovirus/EnterovirusNegative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
HumanRhinovirusModeratePositive(3× LoD)3.0E-01TCID50/mL(1.1E+02copies/mL)-28/30(93.3%)30/30(100%)58/60(96.7%)30/30(100%)-30/30(100%)60/60(100%)
(Zeptometrix0810012CFN)Low Positive(1× LoD)1.0E-01TCID50/mL(3.8E+01copies/mL)-30/30(100%)30/30(100%)60/60(100%)30/30(100%)-30/30(100%)60/60(100%)
Influenza A H1Negative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
Negative(no analyte)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Influenza AH1-2009(Zeptometrix0810109CFN)ModeratePositive(3× LoD)1.5E+00TCID50/mL(9.9E+02copies/mL)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Low Positive(1× LoD)5.0E-01TCID50/mL(3.3E+02copies/mL)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
InflInfluenzaA H3Negative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
Analyte(Isolate Source ID)ConcentrationTestedFilmArray 2.0FilmArray Torch
Site ASite BSite CSystemTotalSite ASite BSite CSystemTotal
InfluenzaA H3(ATCCVR-810)ModeratePositive(3× LoD)$3.0E-01$TCID50/mL29/30(96.7%)30/30(100%)59/60(98.3%)30/30(100%)-30/30(100%)60/60(100%)
Low Positive(1× LoD)$1.0E-01$TCID50/mL30/30(100%)30/30(100%)60/60(100%)30/30(100%)-30/30(100%)60/60(100%)
Influenza BInfluenzaBNegative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
InfluenzaB(Zeptometrix0810037CF)ModeratePositive(3× LoD)$1.5E+01$TCID50/mL-30/30(100%)30/30(100%)60/60(100%)30/30(100%)-30/30(100%)60/60(100%)
Low Positive(1× LoD)$5.0E+00$TCID50/mL-30/30(100%)30/30(100%)60/60(100%)30/30(100%)-30/30(100%)60/60(100%)
ParainfluenzaVirus 1Negative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
Parainfluenza Virus 2ParainfluenzaVirus 2Negative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
ParainfluenzaVirus 2(Zeptometrix0810015CF)ModeratePositive(3× LoD)$1.5E+00$TCID50/mL-29/30(96.7%)30/30(100%)59/60(98.3%)30/30(100%)-29/30(96.7%)59/60(98.3%)
Low Positive(1× LoD)$5.0E-01$TCID50/mL-30/30(100%)27/30(90.0%)57/60(95.0%)30/30(100%)-29/30(96.7%)59/60(98.3%)
ParainfluenzaVirus 3Negative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
ParainfluenzaVirus 4Negative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
Parainfluenza Virus 4ParainfluenzaVirus 4(Zeptometrix0810060CF)ModeratePositive(3× LoD)$1.5E+02$TCID50/mL-30/30(100%)30/30(100%)60/60(100%)30/30(100%)-30/30(100%)60/60(100%)
Low Positive(1× LoD)$5.0E+01$TCID50/mL29/30(96.7%)30/30(100%)59/60(98.3%)30/30(100%)29/30(96.7%)59/60(98.3%)
Negative(no analyte)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Analyte(Isolate SourceID)ConcentrationTestedFilmArray 2.0Agreement with Expected ResultFilmArray Torch
Site ASite BSite CSystemTotalSite ASite BSite CSystemTotal
RespiratorySyncytialVirusb(Zeptometrix0810040ACF)ModeratePositive(3× LoD)6.0E-02TCID50/mL(2.7E+01copies/mL)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Low Positive(1× LoD)2.0E-02TCID50/mL(9.0E+00copies/mL)19/20(95%)20/20(100%)18/20(90%)57/60(95%)20/20(100%)20/20(100%)19/20(95%)59/60(98.3%)
Negative(no analyte)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Severe AcuteRespiratorySyndromeCoronavirus 2(SARS-CoV-2)(ATCC VR-1986HK)ModeratePositive(3× LoD)1.5E+03copies/mL20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Low Positive(1× LoD)5.0E+02copies/mL20/20(100%)19/20(95%)19/20(95%)58/60(96.7%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Negative(no analyte)20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)
Bordetellaparapertussisc(IS1001)(Zeptometrix0801461)ModeratePositive(3× LoD)1.8E+02IS1001copies/mL19/20(95%)20/20(100%)20/20(100%)59/60(98.3%)19/20(95%)19/20(95%)20/20(100%)58/60(96.7%)
Low Positive(1× LoD)6.0E+01IS1001copies/mL20/20(100%)20/20(100%)20/20(100%)60/60(100%)20/20(100%)19/20(95%)20/20(100%)59/60(98.3%)
Bordetella pertussis (ptxP)Bordetellapertussis(ptxP)Negative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
Bordetellapertussis(ptxP)(Zeptometrix0801459)ModeratePositive(3× LoD)3.0E+03CFU/mL30/30(100%)30/30(100%)60/60(100%)30/30(100%)30/30(100%)60/60(100%)
Low Positive(1× LoD)1.0E+03CFU/mL30/30(100%)30/30(100%)60/60(100%)28/30(93.3%)30/30(100%)58/60(96.7%)
Analyte(Isolate SourceID)ConcentrationTestedAgreement with Expected Result
Film Array 2.0FilmArray Torch
Site ASite BSite CSystemTotalSite ASite BSite CSystemTotal
ChlamydiapneumoniaeNegative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)
MycoplasmapneumoniaeNegative(no analyte)60/60(100%)60/60(100%)60/60(100%)180/180(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

a Reproducibility of Adenovirus detection with the BioFire RP2 was 98.3% and 99.2% for the low and moderate positive samples, respectively and 100% for the negative sample(s).

b Respiratory Syncytial Virus results in the BioFire RP2 reproducibility study agreed with the expected result in 98.3 - 100% of the positive sample replicates in 100% of the negative sample replicates.

· Bordetella parapertussis (IS 1001) was detected in 93.3% of the low positive sample replicates tested and in 99.2% of the moderate positive sample replicates tested in the BioFire RP2 reproducibility study. Agreement with the expected result was 100% for the negative sample(s).

For the three analytes that had been evaluated in both studies, the reproducibility of detection observed for the BioFire RP2.1 was overall similar to what was observed for BioFire RP2. Overall, there were ten Not Detected results when the analyte was known to be present in the test sample in the reproducibility evaluations for the BioFire RP2.1. The observed Not Detected frequency is consistent with the test levels (<5% Not Detected results when testing at or above LoD). No pattern in Not Detected results was observed in the study variables (site, system, day instrument/modules, operator, or reagent lot).

  • b. Linearity/assay reportable range:
    Not applicable

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Controls

Two process controls are included in each pouch:

RNA Process Control

The RNA Process Control assay targets an RNA transcript from the yeast Schizosaccharomyces pombe. The yeast is present in the pouch in a freeze-dried form and becomes rehydrated when sample is loaded. The control material is carried through all stages of the test process, including lysis, nucleic acid purification, reverse transcription, PCR1, dilution, PCR2, and DNA melting. A positive control result indicates that all steps carried out in the FilmArray RP2.1 pouch were successful.

PCR2 Control

The PCR2 Control assay detects a DNA target that is dried in the array along with the corresponding primers. A positive result indicates that PCR2 was successful.

{15}------------------------------------------------

Both control assays must be positive for the test run to pass. If the controls fail, the sample should be retested using a new pouch.

The FilmArray Software automatically fails the run if the melting temperature for either the RNA Process Control or the PCR2 Control is outside of an acceptable range.

The following is also described in the product package insert regarding to external controls:

External controls should be used in accordance with laboratory protocols and the appropriate accrediting organization requirements, as applicable. Transport media can be used as an external negative control. Previously characterized positive samples or negative samples spiked with well-characterized organisms can be used as external positive controls. Commercial external control materials may be available from other manufacturers; these should be used in accordance with the manufacturers' instructions and appropriate accrediting organization requirements, as applicable.

Calibrators

This device does not contain calibrators.

Specimen Stability

The BioFire RP2.1 test requires approximately 0.3mL of NPS specimen, collected according to standard technique and placed in transport media or saline. Samples in medium should be tested as soon as possible, but they may be stored at room temperature (approximately 23°C) for up to four hours, under refrigeration (approximately 4°C) for up to three days, or frozen (<-15 °C or <-70°C) for up to 30 davs.

Detailed documentation concerning NPS sample storage and transport was provided in the original FilmArray RP submissions (K103175, K110764, K120267) for NPS specimens stored in viral transport media. The BioFire RP2 and the BioFire RP2.1 utilize this same sample type and test principles and the additional organisms detected (B. parapertussis and SARS-CoV-2) are biologically similar to others detected by the FilmArray RP (i.e., a representative bacteria and virus). Therefore, the original FilmArray RP specimen stability study data are applicable to the BioFire RP2 and BioFire RP2.1 panels for samples stored in viral transport media. However, for establishing sample stability and storage conditions for NPS specimens in saline, an additional study was performed to validate claims.

For the study, natural NPS in saline matrix was prepared by eluting two NPS specimens collected from a single anonymous, asymptomatic volunteer in 6mL of 0.9% saline. Contrived organism mixes (Table 3) were prepared using analyte

{16}------------------------------------------------

negative individual donor natural NPS in saline matrices. For each contrived mix, a total of ten unique donor NPS in saline matrices were individually spiked with organisms corresponding to the RP2.1 panel to a final concentration based on the LoD (up to 5x). LoD of the analytes were determined in a separate limit of detection for saline samples study as described in the Limit of Detection. Immediately following sample preparation (TO), ten replicates (one from each donor) were tested to serve as a no storage control and to establish the expected Detected and Not Detected results.

SampleMixOrganismaStrain/Isolate/SerotypeConcentrationTestedbUnitsConcentrationRelative to LoDc
AdenovirusSpecies CSerotype 2WHO Int Std1.5E+04(1.5E+04)IU/mLd(copies/mL)e5x
Coronavirus NL63NL631.3E+00(2.7E+02)TCID50/mL(copies/mL)5x
Influenza AH1N1pdm09(H1-2009)A/SwineNY/3/20092.6E+00(1.7E+03)TCID50/mL(copies/mL)5x
M1BordetellaparapertussisA7472.1E+02(3.0E+02)CFU/mL(IS1001copies/mL)5x
Parainfluenza Virus3Type 38.1E+01(1.9E+02)TCID50/mL(copies/mL)5x
Severe AcuteRespiratorySyndromeCoronavirus 2(SARS-CoV-2)USA-WA1/2020(heat inactivated)2.5E+03hcopies/mL5x
Coronavirus 229E229E2.0E+00(3.3E+02)TCID50/mL(copies/mL)5x
HumanMetapneumovirusType 16,IA10-2003, A15.0E+01TCID50/mLi5x
Human Rhinovirus1A3.8E+00(1.9E+01)TCID50/mL(copies/mL)5x
M4Influenza A H3N2(H3)jHongKong/4801/141.3E+00(1.1E+00)TCID50/mL(copies/mL)5x
Parainfluenza Virus4Type 4a2.5E+02(8.0E+03)TCID50/mL(copies/mL)5x
MycoplasmapneumoniaeM1296.3E+00k(4.7E+02)kCCU/mLcopies/mL60xk
Middle EastRespiratorySyndromeCoronavirusEMC/2012(heat inactivated)6.7E+02copies/mL5x
Coronavirus OC43OC437.1E-02(2.8E+02)TCID50/mL(copies/mL)5x
EnterovirusD686.0E+02(5.2E+01)TCID50/mL(copies/mL)2xl
M5Influenza A H3N2(H3)jHongKong/4801/141.4E+00(1.1E+00)TCID50/mL(copies/mL)5x
Parainfluenza Virus2Type 22.5E+00(1.5E+02)TCID50/mL(copies/mL)5x
ChlamydiapneumoniaeAR-399.0E+01(6.7E+02)IFU/mL(copies/mL)5x
SampleMixOrganismaStrain/Isolate/SerotypeConcentrationTestedbUnitsConcentrationRelative to LoDc
M6Coronavirus HKU1Clinical NPSSpecimen 537271.0E+04hRNA copies/mL
Influenza A H1N1(H1)A/NewCaledonia/20/995.0E+03(7.0E+02)TCID50/mL(copies/mL)
Influenza BB/FL/04/062.5E+01(1.7E+02)TCID50/mL(copies/mL)
Parainfluenza Virus1Type 11.6E+00(5.0E+02g)TCID50/mL(copies/mL)
Bordetella pertussisA6392.0E+03CFU/mL2×1
RespiratorySyncytial VirusnType A2.1E-01(9.0E+01)TCID50/mL(copies/mL)10×

Table 3. Organism composition of each contrived sample mix for stability studies with saline samples

{17}------------------------------------------------

Following testing at T0. individual sample aliquots were prepared from each contrived donor mix and these were stored at ambient (25°C), refrigerated (8°C), or frozen ( <- 15°C) temperatures for the durations indicated in Table 4. Note, the T0 time point was collected on different days for different analytes but the frozen time point for all analytes was collected on the same day. This resulted in analytes in the different sample mixes to be tested for at least 30 days after the no storage time point.

At each time point, ten replicates (one from each donor) evenly distributed between the FilmArray 2.0 and FilmArray Torch instruments were tested. The reported results of Detected (D), Equivocal (E), and Not Detected (ND) were evaluated for each analyte across the storage conditions and compared to the results observed at TO. 0140 (b) (4)

Storage ConditionSample Size
TemperatureTime
N/ANo Storage Control (T0)10
Ambient(25°C)4 hours10
Refrigerated(8°C)2 days10
3 days10
Frozen(≤-15°C)≥30 days10
Total50

Table 4. Storage conditions and sample size to be tested at each condition.

A valid result (i.e., all internal pouch controls passing) was required for each pouch tested. Pouches with invalid results due to a control failure, instrument error, or software error were retested until a valid result was obtained. Only results from the valid pouches were considered in subsequent analyses. Samples with Influenza A/subtype Equivocal or Influenza A (no subtype detected) results were retested according to the intended result interpretation algorithm (see Table 1).

{18}------------------------------------------------

Any observed trending across conditions (i.e., time- and/or temperature-dependent shift in assay parameters) would be indicative of possible impact on sample stability.

The below table provides a summary of the saline sample stability study. As indicated in the summary table, some observations across different analytes and storage conditions did not meet the expected Detected results. Some of the missed detections did not correspond to the same donor samples and they were distributed such that it appears no trend was observed (e.g., an unexpected Not Detected result for T0 but Detected result for later time points and other storage conditions, etc.). However, for three analytes detected by the RP2.1 device (RSV, Parainfluenza Virus 2, and Mycoplasma Pneumoniae) an additional study was conducted to clarify any possible negative trends.

Table 5. Summary of analyte detection results observed for samples tested at TO (no storage control), ambient, refrigerated, and frozen conditions. Results are reported Detected results (D) in samples that contained the relevant analyte and as expected Not Detected (ND) results in samples that did not contain the relevant analyte.

OrganismSource IDConcentrationTestedaNo StorageControlT0Ambient(25°C)4 hoursRefrigerated(8°C)2 daysRefrigerated(8°C)3 daysFrozen(≤-15°C)≤30 daysb
DNDDNDDNDDNDDND
Adenovirus C(WHO IS)cNIBSC16/324$1.5E+04$ IU/mLd10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
Coronavirus 229EATCCVR-740$3.3E+02$copies/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
CoronavirusHKU1Clinical NPSSpecimen53727$1.0E+04$copies/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
Coronavirus NL63BEI NR-470$2.7E+02$copies/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
CoronavirusOC43Zeptometrix0810024CF$2.8E+02$copies/mL10/1030/3010/1030/309/1030/3010/1030/309/1030/30
HumanMetapneumovirusZeptometrix0810161CF$5.0E+01$TCID50/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
HumanEnterovirusATCCVR-1823$2.6E+02$copies/mLe10/1010/1010/1010/1010/10
HumanRhinovirusZeptometrix0810012CFN$1.9E+01$copies/mL10/1020/2010/1020/209/1020/209/1020/208/1020/20
Influenza A H1N1Zeptometrix0810036CF$7.0E+02$copies/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
Influenza A H1N1-2009Zeptometrix0810249CFf$1.7E+03$copies/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
Influenza AH3N2hZeptometrix0810526CF$1.1E+00$copies/mL10/1020/2010/1020/2010/1020/2010/1020/2010/1020/20
Influenza BZeptometrix0810255CF$1.7E+02$copies/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
ParainfluenzaVirus 1Zeptometrix0810014CF$5.0E+02$copies/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30

{19}------------------------------------------------

ConcentrationNo StorageControlT0Ambient(25°C)4 hoursRefrigerated(8°C)2 daysRefrigerated(8°C)3 daysFrozen(≤-15°C)<30 daysb
OrganismSource IDTestedaDNDDNDDNDDNDDND
ParainfluenzaVirus 2Zeptometrix0810015CF1.5E+02copies/mL7/1030/307/1030/309/1030/309/1030/305/1030/30
ParainfluenzaVirus 3Zeptometrix0810016CF1.9E+02copies/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
ParainfluenzaVirus 4Zeptometrix0810060CF8.0E+03copies/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
Severe AcuteRespiratorySyndromeCoronavirus 2ATCCVR-1986HK2.5E+03copies/mLe10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
BACTERIA
BordetellaparapertussisZeptometrix08014613.0E+02copies/mL10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
BordetellapertussisZeptometrix08014591.0E+04CFU/mLe10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30
ChlamydiapneumoniaeATCC535926.7E+02copies/mL9/1030/308/1030/309/1030/3010/1030/308/1030/30
MycoplasmapneumoniaeZeptometrix08015794.7E+03copies/mLe10/1030/3010/1030/3010/1030/3010/1030/3010/1030/30

" Unless otherwise noted, the concentration tested is 5× LoD was determined in the LoD study for saline sample type with the BioFre Respiratory Panel 2.1 (RP2.1).

"The storage claim is that samples can be frozen for up to 30 days. Although the T0 time point was collected on different analytes, the frozen ime point for all analytes was collected on this resulted in analytes being stored for variable durations and tested at different intervals after To. All analytes were tested at least 30 days containing Adenovins, Bordetella parapertussis, Coronavious NL63, Influenza A HINI-2009, Parainfluenza Virus 3, and Severe Acute Respiratory Syndrome Coronavirus 2 were tested 45 days after T0. Samples containing Chlamycia preumoniae, Coronavins OC43, Human Entervirus, Paranfluenza Virus 2 were tested 31 days after T0. Samples contaning Bordetella pertussis, Coronavins 229E, Coronavirus HKU1, Human Rhinoviros, Influenza A H1N1, Influenza A H1N1, Influenza B, Mycoplasma preumoniae, Parainfluenza Virus 1, Parainfinenza Virus 4, Middle East Respiratory Syndrome Coronavirus, and Respiratory Syncytal Virus were tested 30 days after TO

WHO IS = World Health Organization International Standard

4 TU = Intenational Unis. BioFire Diagnostics quantitive real-time PCR and denonstrated that 1.5E+04 IUinLis equivalent to 1.5E+04 copies/mL.

e Indicated concentration is 10× LoD.

4 Catalog or Source ID from Zeptometrix was previously 0810109CFN, as indicated on the stock was received.

6 Human Enterovins and Human Rhinovinus cannot be distingussed by the BloFire RP2.1 Panel. A detection of ether virus is reported as Hunan

Rhinovins Enterovins. Two organism mither virus, resulting in 20 expected Not Detected results at each time point.

4 Influenza A H3 was noorrectly spiked into one sample ma lower than expected organism concentration. It was then neluded in an additional mix at the appropriate concentration. Influenza A H3 results for the incorrectly formulated sample mix were excluded resulting in 10 expected Detected results and only 20 expected Not Detected results at each time point.

Additional saline specimen stability study (Mycoplasma pneumoniae, Parainfluenza Virus 2, and Respiratory Syncytial Virus (RSV))

In the original saline specimen stability study, three analytes were evaluated in concentrations that were not in accordance with the original study design and therefore vielded results that did not clarify possible stability trends. Therefore, an additional study was performed at the indicated 5x LoD as originally intended for those three analytes. A summary of the additional data collected for the three analytes are summarized in the below table (Table 6). Note that the conditions that were evaluated corresponded to sample storage claims, and included no storage control (TO), ambient (4 hours at 25℃), refrigerated (2-3 days at 8℃), and frozen (30 days at <- 15℃) conditions. Replicates at each test condition were evaluated and the performance was compared to the results obtained from the no storage control (TO) to indicate if there were any adverse sample stability observations in the saline media.

{20}------------------------------------------------

For this evaluation, unique natural NPS in saline specimen matrices were individually spiked with the three organisms to result in a final concentration of 5x LoD. The organism stock of Mycoplasma pneumoniae for this mix was the same used in previous evaluations for sample stability. However, due to insufficient volume remaining of the Parainfluenza Virus 2 and RSV stocks used in the initial testing. alternate stocks with the same source ID but different lot numbers were used in this mix. Replacement organism stock lots were used and quantified in copies/mL. Due to the slight variations in quantification methods used for the original and reestablished stock concentrations, different absolute values in copies/mL were observed compared to the original values. Therefore, the updated stock concentration was used to evaluate equivalency in panel performance with saline media for these analytes.

Table 6. Summary of analyte detection results observed for the additional sample mix at T0 (no storage control), ambient, refrigerated, and frozen conditions. Results are reported are number of detections vs. total replicates tested.

OrganismSource IDConcentrationTestedaNoStorageControlAmbient(25°C)Refrigerated(8°C)Frozen(≤-15°C)
T04 hours2 days3 days30 days
MycoplasmapneumoniaeZeptometrix0801579$2.3E+03$copies/mL9/99/99/99/99/9
ParainfluenzaVirus 2Zeptometrix0810015CF$5.0E+03$copies/mL9/99/99/99/99/9
RespiratorySyncytial VirusZeptometrix0810040ACF$4.5E+02$copies/mL9/99/99/99/99/9

ª All concentrations are 5× the limit of detection (LoD) with LoD updated from new stock evaluations for Parainfluenza Virus 2 and Respiratory Syncytial Virus (see Table 30).

For the Mycoplasma pneumoniae analyte, there were no observed trends that would indicate a sample stability issue with the saline specimens over the range of storage conditions. Compared to the Cp observed for the no storage control (mean Cp the change in Cp ranged from (b) to (b) with no notable trends. Therefore, (b) (4) it appeared that the storage conditions evaluated did not impact test performance for the analyte.

For the Parainfluenza Virus 2 analyte, compared to the Cp observed for the no storage control (mean Cp (b) (4) (4) the change in mean Cp ranged from (0) to (0) without any notable trends across the different storage conditions evaluated. All replicates were detected at the various test conditions.

For the RSV analyte, compared to the T0 baseline (mean Cp (b) (4) ) the change in mean Cp ranged from (b) to(b) . The changes across the different storage conditions evaluated do not show significant trends and all replicates were detected as expected. Further, based on the assumption that a 2-fold difference in concentration results in one cycle shift in Cp value, the variation is within the expected range for stochastic detection results.

{21}------------------------------------------------

Cumulatively, the reevaluations for these three analytes indicate no significant trends in sample stability in the saline media for the indicated sample storage conditions.

Fresh vs. Frozen Study

Detailed documentation concerning fresh vs. frozen NPS in VTM specimens was provided in the original FilmArray RP submissions (K103175, K110764, and K120267). The results of the analytical study conducted for the original FilmArray RP submissions demonstrated that preservation of samples (by freezing at <-70°C) does not affect the accuracy of test results compared to freshly collected or freshly prepared samples. Therefore, it is appropriate to utilize frozen archived prospective and retrospective clinical samples in the evaluation of BioFire RP2.1 to supplement the prospective (fresh) clinical study data, and to use frozen simulated samples in analytical studies for this submission.

Single-Spiked vs. Multi-Spiked Specimen Study

Detailed documentation concerning analytical performance with samples composed of multiple organisms in a specimen was provided in the original FilmArray RP submissions (K103175, K110764, and K120267). The results of the analytical study demonstrated that the LoDs from testing single-spiked and multi-spiked specimens were comparable. Therefore, it wasdeemed appropriate to utilize the multi-spiked approach in conducting the analytical studies and the clinical study testing contrived clinical specimens in support of this submission.

d. Detection limit:

The Limit of Detection (LoD) for all analytes on the BioFire RP2.1 were evaluated based on whether the analytes were new (i.e., SARS-CoV-2) or previously included in the BioFire RP2 panel, and in terms of the specimen media type (i.e., nasopharyngeal swab matrix in saline or viral transport media). LoD confirmation testing consisted of twenty replicates at the estimated LoD concentration (1×) and twenty replicates of a ten-fold dilution of the LoD sample (0.1×). Samples were contrived in either viral transport media (VTM) or in saline. In addition, saline specimens contained an artificial nasopharyngeal swab (aNPSs) matrix, and equivalency in detection between the previously established LoD of samples in VTM was evaluated for validation of NPS in saline as an appropriate sample type. When possible, testing was performed with the same stock/lot used to verify and confirm the LoD in VTM for the saline samples.

The LoD concentration was confirmed by detection in at least 95% of 20 replicates and detection in less than 95% of 20 replicates at a concentration below the LoD (0.1x) on FilmArray 2.0 and FilmArray Torch systems. The saline sample matrix was considered equivalent when the LoD was verified by detection in ≥95% (≥19/20) of replicates at the LoD (1×) concentration and detection in <95% (<18/20) of replicates below the LoD (0.1×) on FilmArray 2.0 and FilmArray Torch systems. When these criteria were not met, additional side-by-side testing was performed to compare

{22}------------------------------------------------

samples prepared in both saline/artificial NPS and VTM at the same concentrations in order to demonstrate that no significant difference existed between the sample matrices.

Limit of Detection for SARS-CoV-2

LoD estimate and confirmation testing for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) was performed with both inactivated and infectious cultures of the same virus isolate (USA-WA1/2020); see Table 8). Samples containing the inactivated virus (ATCC VR-1986HK, USA-WA1/2020) were prepared in Remel M4 viral transport medium as the sample matrix. Samples with cultured infectious virus (USA-WA1/2020) were prepared in a matrix of pooled clinical nasopharyngeal swab specimens (archived NPS specimens collected prior to the SARS-CoV-2 outbreak in Dec 2019). For aNPSs samples, heat inactivated virus was used to verify the LoD in contrived specimens.

The stock concentration of inactivated virus (VR-1986HK) in copies/mL and TCID50 mL was provided by ATCC. The concentration was assessed to be 1.60E+05 (i.e., 1.60 x 105)TCID50/mL and 1.16E+9 copies/mL, respectively by digital droplet PCR (ddPCR). The concentration of infectious virus (in genomic copies/mL and TCID50/mL) was provided by the contracted laboratory that cultured and tested the virus. For the infectious virus, the copies/mL concentration was determined by two different methods. Ouantitative real-time PCR (gPCR) was performed on intact virus (not extracted) using a WHO qualified assay with primers and probe targeting the E gene (WHO E. Charité. Germany: stock culture concentration of 2.4E+09 copies/mL) and qPCR on extracted genomic RNA was performed using the CDC 2019-Novel Coronavirus (2019-nCoV) Diagnostic Panel N1 assay primers and probe (CDC N1; stock culture concentration of 7.4E+08 copies/mL). Both copies/mL concentrations are indicated in the above table on isolate/culture information (Table 8), but the higher of the two stock concentrations (2.4E+09 copies/mL determined with the WHO/Charité E assay) was applied for determining the LoD concentration.

LoD estimate testing for inactivated and infectious SARS-CoV-2 consisted of multiple replicates of a ten-fold serial dilution. The SARS-CoV-2 Detected (D) or Not Detected (ND) results are shown for each replicate, (b) (4) (b) (4)

{23}------------------------------------------------

Table 7. LoD Estimate Test Results for Inactivated SARS-CoV-2 (ATCC VR-1986HK) The boxed data indicates the estimated LoD concentration.

Severe Acute Respiratory Syndrome Coronavirus (SARS-CoV-2) Assay and Detection Results b) (4)

a The concentration of inactivated virus (VR-1986HK) in copies/mL was provided by ATCC. The copies/mL value was determined by digital droplet PCR (ddPCR).

Based on the results of the LoD estimate testing, an estimated LoD concentration of 5.0E+02 copies/mL was selected for LoD confirmation testing.

LoD confirmation testing of contrived samples containing inactivated virus (prepared in Remel M4 viral transport medium) consisted of twenty replicates at the estimated LoD concentration (1×) and twenty replicates of a ten-fold dilution of the LoD sample (0.1×). The testing confirmed a LoD of 5.0E+02 copies/mL with detection of Severe Acute Respiratory Syndrome Coronavirus (SARS-COV-2) in 20/20 (100%) replicates at the 1× concentration and detection in <95% of the replicates (5/20, 25%) at the 0.1× concentration (Table 8).

Table 8. LoD BioFire RP2.1 LoD Confirmation Results for Inactivated SARS-CoV-2 (ATCC VR-1986HK)

{24}------------------------------------------------

The same method of LoD estimate and confirmation testing was carried out with contrived samples containing infectious virus in pooled clinical NPS (unknown transport media/medium). Six replicates were tested per dilution. Results from the LoD estimate titration are shown in Table 9 and an estimated LoD of 1.6E+02 copies/mL was selected for confirmation testing.

Severe Acute Respiratory Syndrome Coronavirus (SARS-CoV-2) Detection and Assay Results

Samples were prepared in pooled clinical NPS matrix.2 The boxed data indicates the estimated LoD concentration
------------------------------------------------------------------------------------------------------------------------

Image /page/24/Figure/3 description: The image shows the title "Infectious Virus (USA-WA1/2020)" at the top. Below the title, the text "(b) (4)" is visible. The background of the image is a solid gray color. The text is black and in a simple font.

ª Additional analyte detection (Influenza B, Parainfluenza Virus 3, and Human Rhinovirus/Enterovirus) from the sample matrix was observed in some replicates.

(b) (4)

{25}------------------------------------------------

Additional testing at the estimated LoD concentration and below confirmed a LoD of 1.6E+02 copies/mL for the infectious virus, with detection of Severe Acute Respiratory Syndrome Coronavirus (SARS-COV-2) in 20/20 (100%) replicates at the 1 × concentration and detection in <95% of the replicates (11/20, 55%) at the 0.1× concentration (Table 10).

Table 10. LoD Confirmation Results for Infectious SARS-CoV-2 (USA-WA1/2020; WRCEVA) Samples were prepared in pooled clinical NPS matrix. a

(b) (4)

ª Additional analyte detection (Parainfluenza Virus 4, and Human Rhinovirus/Enterovirus) from the sample matrix was observed in some replicates.

The confirmed LoD concentrations for inactivated and infectious virus (5.0E+02 copies/mL and 1.6E+02 copies/mL, respectively) in transport media specimens are within a ~3-fold difference and therefore are considered similar.

Initial testing of the heat-inactivated virus in saline sample evaluations generated only 90% detection (18/20) at the LoD concentration of 5.0E+02 copies/mL. A second round of testing was performed where replicates at 1x LoD concentration in samples prepared in NPS and VTM were compared. (b) (4)

b) (4)

{26}------------------------------------------------

le 11. SARS-CoV-2 LoD Verification Results - Round 2

Limit of Detection for other (non-SARS-CoV-2) BioFire RP2.1 Analytes (VTM specimens)

For all other analytes, the assays are shared with the previous version of the panel (i.e., BioFire RP2) and the new BioFire RP2.1. Therefore, LoD concentrations of VTM samples were evaluated and confirmed in the context of the BioFire RP2 LoD concentrations. To minimize uncertainty due to variability in isolate/culture quantification, LoD confirmation samples were prepared in transport medium at 1x and 0.1x LoD concentrations, using the same isolate culture stocks used in the BioFire RP2 LoD study, whenever possible. A clinical NPS specimen was used as the source material for the un-culturable Coronavirus HKU1 (viral RNA quantified in copies/mL by qPCR).

Substitution was required for another four of the original BioFire FilmArray RP2 study isolate cultures (Human Metapneumovirus, Human Rhinovirus, Bordetella

{27}------------------------------------------------

pertussis, and Mycoplasma pneumoniae) due to the original culture having been consumed in other testing. When a replacement culture was required, the analyte concentration tested on BioFire RP2.1 matched only one unit of measure for LoD concentration established with the BioFire FilmArray RP2 (1.e., either quantification by molecular methods in copies/mL or quantification by culture methods in TCID50 mL or CFU/mL) because the ratio of viable/infectivity units per mL to copies per mL varies from culture event to culture event. The units used to prepare the BioFire RP2.1 LoD confirmation samples for analytes other than SARS-CoV-2 are indicated in Table 13 below.

All samples (including those containing substituted isolate cultures) were first tested on BioFire FilmArray RP2 and then tested with the BioFire RP2.1. Data from both panels are shown in Table 12 on a combination of FilmArray 2.0 and FilmArray Torch systems. The analytes with additional information to note in the LoD evaluations between the FilmArray RP2 and BioFire RP2.1 panels are discussed below.

Table 12. Limit of Detection (LoD) for BioFire Respiratory Panel 2.1 (RP2.1) Analytes When a substitute culture was tested (grey shading), the LoD concentration is listed in only one unit.

AnalyteIsolateLoDConcentrationFilmArray RP2Detection ResultsBioFire RP2.1Detection Results
1×LoD0.1×LoD1×LoD0.1×LoD
Viruses
AdenovirusSpecies C Serotype 2WHO Int'l StandardNIBSC 16/324$3.0E+03$ IU/mLa( $3.0E+03$ copies/mL) b20/20100%12/2060.0%20/20100%14/2070.0%
Coronavirus229EATCC VR-740$4.0E-01$ TCID50/mL$6.5E+01$ copies/mL19/2095.0%12/2060.0%20/20100%10/2050.0%
CoronavirusHKU1Clinical NPS specimen(53727)$2.0E+03$ copies/mL20/20100%3/2015.0%19/20 c95.0%6/2030.0%
CoronavirusNL63BEI NR-470$2.5 E-01$ TCID50/mL$5.4E+01$ copies/mL20/20100%4/2020.0%20/20100%5/2025%
CoronavirusOC43ATCC VR-759$3.0E+01$ TCID50/mL$5.6E+02$ copies/mL20/20100%6/2030.0%20/20100%8/2040.0%
Severe AcuteRespiratory SyndromeCoronavirus 2(SARS-CoV-2)ATCC VR-1986HK(heat inactivated)$5.0E+02$ copies/mLd$6.9E-02$ TCID50/mLN/A20/20100%5/2025.0%
WRCEVA(infectious)$1.6E+02$ copies/mLf$1.1E-02$ TCID50/mL20/20100%11/2055.0%
HumanMetapneumovirus16, Type A1 IA10-2003Zeptometrix 0810161CF$1.0E+01$ TCID50/mL( $1.2E+03$ copies/mL)g19/20100%20/20100%20/20100%20/20100%
Human Rhinovirus/EnterovirusHuman RhinovirusType 1AZeptometrix 0810012CFN$3.8E+01$ copies/mL19/2095.0%10/2050.0%19/2095.0%9/2045.0%
Enterovirus D68ATCC VR-1823$3.0E+02$ TCID50/mL$2.6E+01$ copies/mL20/20100%11/2055.0%20/20100%14/2070.0%
Influenza AH1Influenza A H1N1A/New Caledonia/20/99Zeptometrix 0810036CF$1.0E+03$ TCID50/mL$1.4E+02$ copies/mL20/20100%12/20h60.0%20/20100%10/20h50.0%
Influenza AH1-2009Influenza A H1N1pdm09A/Swine/NY/03/2009Zeptometrix 0810249CF$5.0E-01$ TCID50/mL$3.3E+02$ copies/mL20/20100%5/20h25.0%20/20100%3/20h15.0%

{28}------------------------------------------------

AnalyteIsolateLoDConcentrationFilmArray RP2Detection ResultsBioFire RP2.1Detection Results
1×LoD0.1×LoD1×LoD0.1×LoD
Influenza AH3Influenza H3N2A/Port Chalmers/1/73ATCC VR-8101.0E-01 TCID50/mL2.1E+01 copies/mL18/20i90.0%0/20h0.0%20/20100%4/20h20.0%
Influenza BB/FL/04/06Zeptometrix 0810255CF5.0E+00 TCID50/mL3.4E+01 copies/mL20/20100%15/2075.0%20/20100%17/2085.0%
Parainfluenza Virus 1Type 1Zeptometrix 0810014CF5.0E+00 TCID50/mL1.0E+03 copies/mL20/20100%14/2070.0%20/20100%14/2070.0%
Parainfluenza Virus 2Type 2Zeptometrix 0810015CF5.0E-01 TCID50/mL3.0E+01 copies/mL20/20100%5/2025.0%19/2095.0%6/2030.0%
Parainfluenza Virus 3Type 3Zeptometrix 0810016CF2.5E+00 TCID50/mL3.8E+01 copies/mL19/2095.0%12/2060.0%20/20100%6/2030.0%
Parainfluenza Virus 4Type 4aZeptometrix 0810060CF5.0E+01 TCID50/mL1.6E+03 copies/mL20/20100%8/2040.0%19/2095.0%5/2025.0%
Respiratory SyncytialVirusType AZeptometrix 0810040ACF2.0E-02 TCID50/mL9.0E+00 copies/mL20/20100%6/2030.0%20/20100%3/2015.0%
Bacteria
Bordetella parapertussis(IS1001)A747Zeptometrix 08014616.0E+01 IS1001copies/mL4.1E+01 CFU/mL20/20100%6/2030%20/20100%3/2015.0%
Bordetella pertussis(ptxP)A639Zeptometrix 08014591.0E+03 CFU/mL20/20100%13/2065.0%19/2095.0%10/2050%
ChlamydiapneumoniaekTW183ATCC VR-22821.0E-01 TCID50/mL6.6E+01 copies/mL2.0E-01 TCID50/mL1.3E+02 copies/mL40/40100%9/4022.5%32/40j80.0%10/4025.0%
MycoplasmapneumoniaeM129Zeptometrix 08015794.6E+02 copies/mL20/20100%8/2040.0%20/20100%7/2035.0%

a IU = International Units.

b BioFire Diagnostics quantified the WHO International Standard by quantitative real-time PCR to demonstrate that 3.0E+03 IU/mL=3.0E+03 copies/nL. Two other adenovirus serotypes (B7 and F41) were also tested for LoD verification and confirmation. Bach was detected in ≥95% of replicates on both panels at a 1×LoD concentration less than 3.0E+03 copies/mL (8.7E+02 and 1.1E+03 copies/mL, respectively).

6 Results are from a second sample containing Coronavirus HKU1 at the 1× LoD concentration. In the first 1× sample tested on the BioFire RP2.1, Coronavirus HKU1 was detected in 18/20 (90%) replicates (5/20, 25.0% for the 0.1× sample).

4 Concentration of viral RNA in the culture was detemined by digital droplet PCR, as indicated on the Certificate of Analysis from ATCC.

e World Reference Center for Emerging Viruses and Arbovinuses; contributed by the United States Control (CDC).

4 Concentration of viral RNA in the culture was determined by quantitative real-time PCR using E gene primers and probe as deccribed on the World Health Organization (WHO) website: https://www.who.int/docs/default-source/coronaviruse/protocol-v2-1 pdf.

E A copies inL concentration for the substitute culture of Human Metapneumovirus that was tested has not been determined. The corresponding BioFire FilmArry RP2 LoD concentration for Human Metapneumovins at 1.0E+01 TCD30inL is 1.2E+03 copies/nL (based on testing of a different culture of the same isolate).

4 Detection for Influenza A subtypes is calculated based only on the correct Influenza A (subtype) Detected result. An Equivocal or Influenza A (no subtype detected) result is tallied as a Not Detected result.

1 2/20 replicates were Influenza A H3 Equivocal.

I Two different samples containing C. pneumoniae at the FilmArray RP2 LoD concentration of 6.6E+01 copies/mL were prepared and both samples were tested with BioFire FilmAray RP2 and BioFire RP2.1. Detection at the 1× level was 20/20 with BioFire FilmArrayRP2 and 16/20 with BioFire RP2.1 for both samples.

4 The LoD concentration for Chlanydia on the BioFire RP2.1 is two-fold higher than the LoD concentration on the BioFre FilmArray RP2.

LoD testing for Adenovirus was performed with a World Health Organization (WHO) International Standard developed by NIBSC (National Institute for Biological

{29}------------------------------------------------

Standards and Control) that was released in late 2018 (NIBSC 16/324, 1st WHO International Standard for human adenovirus DNA for nucleic acid amplification techniques, version 1.0, dated 12/13/2018). This is a whole virus preparation of HAdV type 2 with a standardized concentration expressed in International Units (IU)}. All adenovirus cultures previously quantified by different species-specific qRT-PCR kits and assays were re-evaluated using a single quantitative assay (Adenovirus R-GENE) that was verified against the international standard.

LoD confirmation for the Adenovirus international standard was first performed in 2019 with the FilmArray RP2 to establish an Adenovirus LoD concentration of 3.0E+03 IU/mL (data provided in Table 13 and Table 14; equivalent to 3.0E+03 copies/mL when quantified with the Adenovirus R-GENE (bioMerieux) quantitative real-time PCR kit). The same samples used in establishing the Adenovirus LoD on the FilmArray RP2 with the international standard was tested with the BioFire RP2.1. Similar results were observed (Table 15), confirming identical LoD for Adenovirus (3.0E+03 IU/mL or 3.0E+03 copies/mL) on each panel.

Table 13. LoD Estimate Test Results for WHO Adenovirus International Standard (NIBSC 16/324) on BioFire FilmArray

The boxed data indicates the estimated LoD concentration
(b) (4)

Table 14. LoD Confirmation Results for WHO Adenovirus International Standard (NIBSC 16/324) on BioFire FilmArrav (b) (4)

1 Fryer JF, Hockley JG, Govind S, Morris CL and the Collaborative Study Group. Collaborative Study to Evaluate the Proposed 1st WHO International Standard for Human Adenovirus (HAdV) DNA for Nucleic Acid Amplification Techniques (NAT). WHO ECBS Report 2018; WHO/BS/2018.2346

{30}------------------------------------------------

(b) (4)

Table 15. Adenvirus LoD Verification and Confirmation Results (Species C Serviype 2; WHO International Standard, M
NIBSC 16/32)

(b) (4)

{31}------------------------------------------------

Additional Adenovirus LoD Data

The BioFire RP2.1 contains five assays for Adenovirus detection and the WHO International Standard (type C2) is amplified by the Adeno2 and Adeno6 assays. Therefore, for completeness two additional Adenovirus isolates (serotypes B7A and F41) were also tested at the RP2 1x LoD concentrations (8.7E+02 copies/mL and 1.1E+03 copies/mL. respectively by R-GENE quantification) and at 0.1x LoD. in order to collect data for each of the five adenovirus assays. In these Adenovirus isolates testing, twenty replicates of each sample were evaluated with both panels and demonstrated 100% detection of replicates (20/20) at the 1x LoD concentration for the FilmArray RP2 and BioFire RP2.1 devices. The inclusion of two additional isolates in the testing allowed for data to be collected from each assay (Adeno2 and Adeno7.1 for B7 and Adeno 3 and Adeno8 for F41). Both of the additional isolates were tested and detected as expected (Table 16) at a copies/mL concentration lower than that of the WHO International Standard. The LoD for the Adenovirus WHO International Standard is the Adenovirus LoD claim for the panel.

AnalyteIsolateLoDConcentration a,bFilmArray RP2Detection ResultsBioFire RP2.1Detection Results
1×LoD0.1×LoD1×LoD0.1×LoD
AdenovirusSpecies B Serotype 7AZeptometrix 0810021CF5.0E-02 TCID50/mL3.9E+01 copies/mLa[8.7E+02 copies/mL]b20/20100%12/2060.0%20/20100%14/2070.0%
Species F Serotype 41ATCC VR-9301.0E+00 TCID50/mL1.2E+02 copies/mLa[1.1E+03 copies/mL]b20/20100%14/2070.0%20/20100%13/2065.0%

Table 16. Additional Adenovirus LoD Testing Data on FilmArray RP2 and BioFire RP2.1

4 Concentration based on quantification with the Genesig Adenovirus B kit (PrimerDesign, Ltd.) as described in K170604 b Concentration based on quantification with Adenovirus R-GENE (BioMerieux, ref. 69-010B)

The sample containing Human Metapneumovirus required a substitute culture that is only quantified in TCID50/mL units (same isolate, different culture event/lot#). Attempts to accurately quantify the substitute culture in copies/mL (similar to the original stock used in the RP2 LoD study) were not successful and therefore the sample could only be prepared based on the RP2 TCID50/mL LoD (1.0E+01 TCID50/mL). Testing at this concentration resulted in detection of the analyte in 19/20 and 20/20 replicates on FilmArray RP2 and BioFire RP2.1, respectively. However, the analyte was also detected in all replicates (20/20) on both panels at the 0.1× LoD concentration.

(b) (4)
(b) (4)more robust detection at the 0.1xlevel (Table 19) suggests that the substitute culture has a slightly higher nucleic acidconcentration compared to the original culture at the same TCID50/mL. Based on the

{32}------------------------------------------------

data provided, it appears that the BioFire RP2.1 hMPV amplification and detection are equivalent to FilmArray RP2 and the existing claimed LoD concentration (1.0E+01 TCID50mL and 1.2E+03 copies/mL) appears to be applicable to FilmArray RP2 and BioFire RP2.1.

Table 17. Comparison of Results at 1.0E+01 TCIDss/mL for Different Cultures of Human Metapneumovirus (Zeptometrix 0810161CF) (b) (4)

Table 18. Human Metapneumovirus LoD Verification and Confirmation Results (Zeptometrix 0810161CF) (BioFire RP2 and BioFire RP2 1)

{33}------------------------------------------------

(b) (4)

For Influenza A detection, the BioFire RP2.1 contains five assays, and including three assays used to differentiate the Influenza A hemagglutinin type. Testing for Influenza A H3 with BioFire FilmArray RP2 at the BioFire FilmArray RP2 LoD concentration generated 18/20 Influenza H3 Detected results, with Influenza A H3 Equivocal results for the other two replicates. Retesting is typically recommended for an Equivocal result, however, the retesting was not performed for these two sample replicates (Table 19). The BioFire RP2.1 data (20/20 at the 1× concentration and 4/20 at the 0.1× concentration) confirm that the concentration tested is an appropriate LoD for Influenza A H3 detection by BioFire RP2.1. The 2.1E+01 copies/mL confirmed LoD concentration is the same for each panel.

Table 19. Influenza A H3 LoD Verification and Confirmation Results (A/Port Chalmers/1/73, ATCC VR-810)

NSD = Influenza A (no subtype detected), InfA E = Influenza A Equivocal, H3 E = Influenza A H3 Equivocal (b) (4)

{34}------------------------------------------------

Image /page/34/Figure/0 description: The image shows the label '(b) (4)' in the upper left corner. The rest of the image is a gray rectangle. The OCR text indicates that the image is related to Chlamydia pneumoniae.

Chlannydia pneumoniae is the only analyte tested where results indicate a potential difference in LoD between the panels. The first sample prepared was detected in 20/20 replicates (100%) with the BioFire FilmArray RP2 test and in only 16/20 replicates in the BioFire RP2.1 testing. (b) (4)

, a second sample was prepared at the same concentration and tested on both panels again, with the same detection results (20/20 BioFire FilmArray RP2 and 16/20 BioFire RP2.1).

Table 20. Chlamydia pneumoniae LoD Testing Data on BioFire FilmArray RP2 and BioFire RP2.1 Pouches for Two Samples Prepared at the Same Concentration (1x LoD: 6.6E+01 copies/mL, 1.0E-01 TCID50/mL)

{35}------------------------------------------------

Testing was then performed with the BioFire RP2.1 pouches with a sample containing C. pneumoniae at a 2-fold higher concentration (1.3E+02 copies/mL) and the analyte was detected in all replicates (20/20, 100%) at the 1× concentration and in <95% of the replicates tested at the 0.1× concentration (9/20, 45% at 1.3E+01 copies/mL; Table 21). The data confirm a revised LoD for C. pneumoniae of 1.3E+02 copies/mL. Analytical reactivity (inclusivity) testing was performed for C. pneumoniae based on the revised LoD concentration.

Table 21. Chlamydia pneumoniae LoD Testing Data on BioFire RP2.1 at a Revised LoD Concentration (1.3E+02 copies/mL) (b) (4)

Limit of Detection for other (non-SARS-CoV-2) BioFire RP2.1 Analytes (Saline specimens)

{36}------------------------------------------------

When possible, testing was completed with the same stock/lot used to verify and confirm the LoD of the BioFire RP2.1 in VTM (results for 18/25 analytes of the RP2.1 panel are shown in Table 22). Among these, five analytes did not initially meet the acceptance criteria. These are indicated in the table by the addition of results from comparing to samples prepared in VTM. Additional evaluations were performed for these analytes including Adenovirus C, Coronavirus HKU1, and Respiratory Syncytial Virus. The SARS-CoV-2 detection rate in aNPSs is detailed above.

AnalyteIsolateLoDConcentrationDetection Results
SalineaVTMb
1× LoD0.1×LoD1× LoD0.1×LoD
Viruses
AdenovirusSpecies C Serotype 2WHO Int'l StandardNIBSC 16/324$3.0E+03 IU/mLc$$(3.0E+03 copies/mL)e$20/20d100%20/20d100%20/20100%18/2090%
Species F Serotype 41ATCC VR-9301.0E+00 TCID50/mL1.1E+03 copies/mL20/20100%7/2035%
Coronavirus229EATCC VR-7404.0E-01 TCID50/mL6.5E+01 copies/mL20/20100%8/2040%
CoronavirusHKU1Clinical NPS specimen(53727)2.0E+03 copies/mL18/2090%6/2030%17/2085%5/2025%
CoronavirusNL63BEI NR-4702.5 E-01 TCID50/mL5.4E+01 copies/mL20/20100%10/2050%10%
Severe AcuteRespiratory SyndromeCoronavirus 2(SARS-CoV-2)USA-WA1/2020ATCC VR-1986HK(heat inactivated)$5.0E+02 copies/mLf$2.1E-01 TCID50/mL17/20g85%10/20g50%20/20100%5/2025%
Human Metapneumovirus16, Type A1 IA10-2003Zeptometrix 0810161CF$1.0E+01 TCID50/mLh$20/20100%20/20100%
Human Rhinovirus/EnterovirusEnterovirus D68ATCC VR-18233.0E+02 TCID50/mL2.6E+01 copies/mL19/2095%6/2030%
Influenza A H1Influenza A H1N1A/New Caledonia/20/99Zeptometrix 0810036CF1.0E+03 TCID50/mL1.4E+02 copies/mL20/20100%14/2070%
Influenza A H1-2009iInfluenza AH1N1pdm09A/Swine/NY/03/2009Zeptometrix 0810249CF5.0E-01 TCID50/mL3.3E+02 copies/mL19/2095%0/200%
Influenza BB/FL/04/06Zeptometrix 0810255CF5.0E+00 TCID50/mL3.4E+01 copies/mL19/2095%10/2050%
Parainfluenza Virus 2Type 2Zeptometrix 0810015CF5.0E-01 TCID50/mL3.0E+01 copies/mL20/20100%6/2030%
Parainfluenza Virus 4Type 4aZeptometrix 0810060CF5.0E+01 TCID50/mL1.6E+03 copies/mL20/20100%4/2020%
Respiratory Syncytial VirusType AZeptometrix0810040ACF2.0E-02 TCID50/mL9.0E+00 copies/mL16/20i80%2/20J10%16/2080%7/2035%
Bacteria20/2011/20

Table 22. Limit of Detection (LoD) for BioFire Respiratory Panel 2.1 (RP2.1) Analytes in Saline.

{37}------------------------------------------------

Bordetella parapertussis(IS1001)A747Zeptometrix 08014616.0E+01 IS1001copies/mL4.1E+01 CFU/mL100%55%
Bordetella pertussis(ptxP)A639Zeptometrix 08014591.0E+03 CFU/mL20/20100%12/2060%
Mycoplasma pneumoniaeM129Zeptometrix 08015794.6E+02 copies/mL19/2095%9/2045%

a Testing was performed in aNPSs.

b Comparison to samples prepared in VTM are shown when initial testing results did not meet the acceptance criteria

C IU = International Units.

d Results shown are from a second sample containing Adenovirus C tested alongside sample prepared in VTM. In the first 0.1× sample

tested on the BioFire RP2.1/RP2.1plus. Adenovirus was Detected in 19/20 (95%) replicates (20/20.100% for the 1× sample). BioFire Diagnostics quantified the WHO International Standard by quantitative real-time PCR to demonstrate that 3.0E+03 IU/mL=3.0E+03 copies/mL.

4 Concentration of viral RNA in the culture was determined by digital droplet PCR, as indicated on the Certificate of Analysis from ATCC.

8 Results shown a second sample containing SARS-CoV-2 tested alongside sample prepared in VTM. In the first 1 × sample tested on the BioFire RP2.1. SARS-CoV-2 was Detected in 18/20 (90%) replicates (3/20, 15% for the 0.1× sample).

A copies/mL concentration for the culture of Human Metapneumovinus that was tested has not been determined.

1 Detection for Influenza A subtypes is calculated based only on the correct Influenza A (subtype) Detected result. An Equivocal or Influenza A (no subtype detected) result is tallied as a Not Detected result.

J Results shown are from a second sample containing Respiratory Syncytial Virus tested alongside sample prepared in VTM. In the first 1× sample tested on the BioFire RP2.1, Respiratory Syncytial Virus was Detected in 15/20 (75%) replicates (4/20, 20% for the 0.1× sample).

For seven analytes (Adenovirus, Coronavirus OC43, Human Rhinovirus/Enterovirus, Influenza A H3, Parainfluenza Virus 1, Parainfluenza Virus 3, and Chlamydia pneumoniae), it was necessary to perform testing using different organism stock cultures or lots than were previously evaluated on the RP2.1 device. For these analytes, LoD verification was completed by testing organism at a range of concentrations intended to bracket the LoD in samples prepared in both VTM and aNPSs. This procedure was used to preclude potential quantification differences in stocks used in the original study and subsequent saline media evaluations, that might confound a determination of the potential impact of media type on device performance. Detection results for these analytes indicated overall concordance at similar organism concentrations in both saline and VTM (Table 23).

Table 23. Limit of Detection (LoD) for BioFire Respiratory Panel 2.1 (RP2.1) Analytes that were Evaluated with New
Isolates in Saline and VTM.
1× LoD ConcentrationDetection Results
AnalyteIsolateSalineaVTM
1× LoD0.1× LoD1× LoD0.1× LoD
AdenovirusSpecies B Serotype 3Zeptometrix0810062CF1.2E+07 TCID50/mL3.9E+02 copies/mLb20/20100%9/2045%19/2095%12/2060%
Coronavirus OC43Zeptometrix0810024CF3.6E+05 TCID50/mL5.6E+01 copies/mLc20/20100%11/2055%19/2095%12/2060%
Human Rhinovirus/EnterovirusHuman RhinovirusType 1 AZeptometrix0810012CFN1.3E+06 TCID50/mL3.8E+01 copies/mLd20/20100%20/20100%20/20100%15/2075%
Influenza A H3Hong Kong/4801/14ZeptoMetrix0810526CF7.2E+05 TCID50/mL2.1E-01 copies/mLe19/2095%1/205%20/20100%3/2015%Parainfluenza Virus 1Type 1Zeptometrix4.2E+05 TCID50/mL1.0E+02 copies/mLc20/20100%3/2015%19/2095%8/2040%
Influenza A H3Hong Kong/4801/14ZeptoMetrix0810526CF7.2E+05 TCID50/mL2.1E-01 copies/mLe19/2095%1/205%20/20100%3/2015%
Parainfluenza Virus 1Type 1Zeptometrix4.2E+05 TCID50/mL1.0E+02 copies/mLc20/20100%3/2015%19/2095%8/2040%

{38}------------------------------------------------

AnalyteIsolate1× LoD ConcentrationDetection Results
SalineaVTM
1× LoD0.1× LoD1× LoD0.1× LoD
Parainfluenza Virus 30810014CFType 3Zeptometrix0810016CF3.4E+07 TCID50/mL3.8E+01 copies/mLd20/20100%9/2045%20/20100%12/2060%
ChlamydiapneumoniaeAR-39ATCC 535922.9E+07 IFU/mL1.3E+02 copies/mLd19/2095%7/2035%18/2090%0/200%

a Testing was performed in aNPSs.

b 1× LoD value is approximately 2-fold lower than the value established when testing original stock in VTM.

e 1× LoD value is 10-fold lower than value established when testing original stocks in VTM.

d 1× LoD value is equivalent to value established when testing original stocks in VTM.

e 1× LoD value is 100-fold lower than value established when testing an original stock in VTM.

e. Analytical Reactivity (Inclusivity):

The BioFire RP2.1 contains assays for the detection of SARS-CoV-2 and multiple other viral and bacterial respiratory pathogens. Assays are designed to detect sequences of clinically relevant strains, serotypes, and/or genotypes of species that cause respiratory illness.

The reactivity of each assay in the BioFire RP2.1 device has been previously evaluated by in silico analysis and laboratory testing in studies for the performance evaluation of the BioFire RP2, except for SARS-CoV-2. The reactivity assessment and limitations defined in the previous studies also apply to the same assays in the BioFire RP2.1 because the sample type, assay primers, pouch chemistry and reaction conditions are unchanged. In brief, similar to the previous BioFire RP2 analytical reactivity study, when testing the BioFire RP2.1, each isolate that was evaluated with RP2.1 was prepared as a contrived sample in transport medium at a concentration near (3x) LoD and then tested in triplicate. Reactivity was established when the isolate was detected at a near-LoD concentration in 3/3 or 4/5 replicates, and any isolate or sequence with an observed or predicted issue with detection (i.e., requiring more than 10-fold LoD concentration) is defined as an assay limitation.

The Detection Limit section describe the limits of detection for the analytes in the RP2.1 panel. The focus of this reactivity study for the RP2.1 device was on evaluating the reactivity of the SARS-CoV-2 assays and revising or expanding the reactivity assessment for a small number of analytes with updated limits of detection (i.e., Adenovirus and C. pneumoniae).

In Silico Reactivity Analysis of SARS-CoV-2 Sequences

Evaluation of inclusivity (analytical reactivity) for the BioFire RP2.1 SARS-CoV-2 assays (SARSCoV2-1 and SARSCoV2-2) was based on in silico analysis of sequences from the NCBI and GISAID databases as of February 21, 2021.

In total, 467,066 sequences from around the globe were aligned to the assay primers. The sequences evaluated included the following lineages and variants of

{39}------------------------------------------------

concern (VOC) or variants under investigation (VUI) that may have important epidemiological, immunological, or pathogenic properties from a public health perspective:

  • · A.23 lineage (Uganda)
    • o VUI-202102/01 (A.23.1 with E484K in Spike)
  • B.1.1, B1.1.7, B.1.258 lineages (United Kingdom; Δ69-70 and N501Y in Spike)
    • o VOC-202012/01 (B.1.1.7)
    • o VOC-202102/02 (B.1.1.7 with E484K in Spike)
  • B.1.1.28 lineage (Brazil) ●
    • o VOC-202101/02 P1 variant (Brazil/Japan)
    • o VUI-202101/01 P2 variant (Brazil)
  • B.1.1.318 (United Kingdom) ●
    • o VUI-202102/04
  • B.1.351 lineage (South Africa) ●
    • o VOC-202012/02 (501Y.V2 in Spike)
  • B.1.429 lineage (United States) ●
    • o CAL.20C variant
  • B.1.525 lineage (United Kingdom) ● o VUI-202102/03 or UK1188
    • B.1.526 (United States)

All lineages and variants of public health interest identified as of February 2021 are predicted to be detected. Approximately 1.2% of the sequences (5.405/467,066) have a mismatched base within the 3' half of a primer that may affect one assay, but will be detected by the second assay. Both SARS-CoV-2 assays of the BioFire RP2.1 device are predicted to be impaired for nine sequences (9/467,066) evaluated. In summary, this analysis determined that 99,998% (467,057/467,066) of the database sequences evaluated will be amplified by at least one of the SARS-CoV-2 assays in the BioFire RP2.1 device.

The analysis is summarized in the following table-

Predicted Assay ResultSARSCoV2-1# (%) sequences predicted to bedetected with no limitations(one or both assays positive)
+-
SARSCoV2-2+461,6524581b467,057/467,066
-8249c(99.998%)**

Table 24. In silico Prediction of SARS-CoV-2 Detection by the BioFire RP2.1 Assays®

a +/+ indicates detected by both assays with no impairment, +/- indicates detection by one assay with no impairment and potential for impaired detection by the other assay, -- indicates potential for impaired detection by both assays. b Includes sequences of lineage B.1.525 (VUI-202102/03), which has a mutation in the Spike gene that is predicted to impair detection by the SARS-CoV2-1 assay, but detection by the SARSCoV2-2 (Membrane gene) assay is predicted to be unaffected.

· Nine sequences have mismatches in the 3' half of primer(s) for both the SARSCoV2-1 and SARSCoV2-2 assays. The mismatches are predicted to impair detection for these sequences.

Empirical Evaluations for the Reactivity Analysis of SARS-CoV-2 Sequences

{40}------------------------------------------------

The in silico reactivity predictions were supplemented with empirical experiments to estimate the actual detection impact of the mismatches observed in two sequences (as indicated in the above In silico Prediction of SARS-CoV-2 Detection by the BioFire RP2.1 Assays table summary). Briefly, synthetic double-stranded gene fragments, (i.e., gBlocks) were synthesized for the regions of the Spike (S) gene amplified by the SARSCoV2-1 assay and the Membrane (M) gene amplified by the SARSCoV2-2 assay. One construct for each gene carried the consensus sequence with a perfect match to the assay primers (control) while the other was generated with the mismatches to primers, as indicated by the Oct 24, 2020 database alignment. Dilutions of the control and mismatch constructs were tested with the BioFire RP2.1 device.

Testing verified that the 3' terminal mismatch to one of the inner primers of the SARSCoV2-1 assay had a significant (10,000-fold) impact on amplification and detection relative to the control sequence. Further, a mismatch that is 7 bases in from the 3' end of a SARS-CoV2-2 inner primer had a much less impact (10-fold) on amplification and detection relative to the control sequence.

The analysis indicated that as of Oct 24, 2020, the BioFire RP2.1 device can detect the analyzed sequences though an impairment or limit of detection is predicted at low concentrations (≤10x LoD) for 2/130,788 sequences.

Reactivity for Other BioFire RP2.1 Analytes

The BioFire RP2.1 differs from the BioFire FilmArray RP2 by the addition of primers for the detection of SARS-CoV-2. The assay primers and reaction conditions for the viral and bacterial analytes shared between panels are unchanged, and testing has demonstrated that, with the exception of the updated Adenovirus evaluations based on the WHO International Standard and C. pneumoniae, the analytical LoD is unchanged for the assays shared between the two panels. The analytical inclusivity testing performed for the assessment of FilmArray RP2 assays (see K170604) were applied to the corresponding BioFire RP2.1 analytes with the following exceptions-

  • Testing of adenovirus isolates (same as in the RP2 study) but based on the . BioFire RP2.1 Adenovirus LoD newly established with the WHO International Standard (3.0E+03 IU/mL or copies/mL)
  • Testing of C. penumoniae isolates (same isolates as in RP2) but based on the ● LoD established for the BioFire RP2.1

Analytical reactivity testing for adenovirus with the BioFire RP2.1 included evaluating over 25 different isolates representing all but two for the known serotypes within the species associated with respiratory infection (i.e., B, C, and E) and representative serotypes for the other species. Adenovirus B serotype 55 and Adenovirus C serotype 57 were not tested but predicted to be detected efficiently with sequence analysis. At least one representative of each serotype within the Species A and F was tested, while only three different serotypes were tested as representative of species D supplemented with sequence analysis that predicted all other D serotypes

{41}------------------------------------------------

would be detected by the Adenovirus assays. An isolate of species G was not available for testing but in silico evaluation predicts that the Adenovirus assays react with species G, serotype 52. Overall, testing in triplicate demonstrated that all isolates were detected as expected on the BioFire RP2.1 at a concentration at or below 3x LoD (Table 26). Note that isolates tested in the Detection Limit study were not retested in this inclusivity testing but are listed for reference.

Table 25. Results for Adenovirus Inclusivity Testing on BioFire RP2.1 Isolates highlighted grey were tested in the BioFire RP2.1 LoD study and results are compiled from 20 replicates.

Isolates highlighted grey were tested in the BioFire RP2.1 LoD study and results are compiled from 20 replicates.
SpeciesSerotypeIsolate IDStrainxLoDTest Concentration(copies/mL)a#Detected/TotalResult(b) (4)
ASerotype 12ATCC VR-863Huie3x9.0E+03 copies/mL3/3D
Serotype 18ATCC VR-19Washington D.C./19543x9.0E+03 copies/mL3/3D
Serotype 31Zeptometrix 0810073CF-3x9.0E+03 copies/mL3/3D
Serotype 3Zeptometrix 0810062CF-3x9.0E+03 copies/mL3/3D
Serotype 7AZeptometrix 0810021CF-0.3x8.7E+02 copies/mL20/20D
Serotype 7d/d2UIRFIowa/20013x9.0E+03 copies/mL3/3D
Serotype 7hUIRFIowa/19993x9.0E+03 copies/mL3/3D
BSerotype 11ATCC VR-12-3x9.0E+03 copies/mL3/3
Serotype 14ATCC VR-15-3x9.0E+03 copies/mL3/3D
Serotype 16ATCC VR-17CH.79/Saudi Arabia/19553x9.0E+03 copies/mL3/3D
Serotype 21ATCC VR-1833-3x9.0E+03 copies/mL3/3D
Serotype 34ATCC VR-716Compton/19 723x9.0E+03 copies/mL3/3D
Serotype 35ATCC VR-718Holden3x9.0E+03 copies/mL3/3D

{42}------------------------------------------------

Serotype(b) (4)
50ATCC VR-1602Wan/Amsterdam/19883x9.0E+03 copies/mL3/3DDD
C1Zeptometrix0810050CF-3x9.0E+03 copies/mL3/3DDD
2ATCC VR-846Adenoid 63x9.0E+03 copies/mL3/3DDD
NIBSC16/324WHOInternationalStandard1x3.0E+03 copies/mL20/20DD
5Zeptometrix0810020CF-3x9.0E+03 copies/mL3/3DDD
6ATCC VR-6Tonsil99/Wash DC3x9.0E+03 copies/mL3/3DDD
D8Zeptometrix0810069CF-3x9.0E+03 copies/mL3/3DDD
20Zeptometrix0810115CF-3x9.0E+03 copies/mL3/3DDD
37Zeptometrix0810119CF-3x9.0E+03 copies/mL3/3DDD
E4aUIRF 4aS.Carolina/20043x9.0E+03 copies/mL3/3DDD
4Zeptometrix0810070CF-3x9.0E+03 copies/mL3/3DDD
F40Zeptometrix0810084CF-3x9.0E+03 copies/mL4/5DNDD
40NCPV0101141v-3x9.0E+03 copies/mL3/3DD
41ATCC VR-930Tak/73-3544/Netherlands/19730.4x1.1E+0320/20DD
Zeptometrix0810085CF-3x9.0E+03 copies/mL3/3DD

a All isolates were quantified by a quantitative real-time PCR assay (R-GENE, BioMerieux).

{43}------------------------------------------------

Analytical reactivity testing of C. pneumoniae isolates was performed according to the testing procedure followed for the FilmArray RP2 inclusivity study. Each isolate was prepared in VTM matrix at 3× the 1.3E+02 copies/mL BioFire RP2.1 LoD for this analyte and each was detected in the requisite 3/3 or 4/5 replicates tested (Table 27). The LoD isolate data (in grey) are shown for reference.

Source/Isolate IDStrain/Location/Year IsolatedConcentrationTestedxLoD# Detected/TotalResult(b) (4)
ATCC aVR-2282TW-1831.3E+02copies/mL1x20/20D
ATCCVR-1310CWL-0293.9E+02copies/mL3x4/5ND
D
D
D
ATCCVR-1360CM-1/Georgia3.9E+02copies/mL3x3/3D
D
D
ATCC53592AR-39/Seattle/19833.9E+02copies/mL3x3/3D
D
D

Table 26. BioFire RP2.1 Inclusivity Testing Results for C. pneumoniae Based on the Revised LoD

Summary of BioFire RP2.1 Analytical Reactivity

Cumulatively, through the in silico and empirical reactivity assessments, including those from shared assays in the BioFire RP2, 179 different viral and bacterial isolates were detected within 10x LoD.

It is notable that the Influenza A assays will react variably with non-human influenza A viruses and some rarely encountered human influenza A viruses that are not H1, H1-2009, or H3; generally producing Influenza A Equivocal or Influenza A (no subtype detected) results. Also, not evaluated in this study but in previous BioFire RP2 evaluations, the BioFire RP2.1 Bordetella pertussis. Influenza A (subtype). Influenza B and Human Rhinovirus/Enterovirus assays are predicted to react with the viral nucleic acids in B. pertussis, influenza, and poliovirus vaccines respectively. Thus, results should be interpreted with caution for specimens obtained from patients with such vaccinations.

The BioFire RP2.1 analytical reactivity test data is summarized as follows-

RP2.1 Analyte/Result# of Isolates DetectedxLoD DetectedDescription of Isolates Tested and DetectedLimitations
Viruses
AdenovirusSpecies A33xA12, A18, A31Noneb
Species B113xB3, B7, B11, B14, B16, B21, B34, B35, B50
Species C53xC1, C2, C5, C6
Species D33xD8, D20, D37
RP2.1 Analyte/Result# of Isolates DetectedxLoD DetectedDescription of Isolates Tested and DetectedLimitations
Species E23xE4 and E4a
Species F43xF40 and F41
Coronavirus 229E23xGroup 1 Coronavirus CoV-229E
Coronavirus HKU153xGroup 2 Coronavirus HKU1Clinical specimens from 2010 and 2015None
Coronavirus NL6323xGroup 1 Coronavirus CoV-NL63
Coronavirus OC4323xGroup 2 Coronavirus CoV-OC43
Severe Acute Respiratory SyndromeCoronavirus 2 (SARS-CoV-2)1c1xUSA-WA1/2020
Human Metapneumovirus113xGenotypes: A1, A2, B1, B2Subtypes: 3, 4, 5, 8, 9, 13, 16, 18, 20,22, 27None
Human Rhinovirus/EnterovirusHumanRhinovirus143xSpecies A:A1, A2, A7, A16, A34, A57, A77, A85Species B:B3, B14, B17, B27, B42, B83
Enterovirus113xSpecies: A, B, C, DCoxsackievirus A9, A10, A21, A24Coxsackievirus B3, B4Echovirus 6, 9, 11Enterovirus 71, D68 (2014)Noned
Influenza A H113Up to10xeHuman and swine Influenza A H1N1isolates from the 1930s to 2007 andH1N2 recombinant virusNone for Humanisolates.Variable detectionof swine and avianisolates, dependingon strain andconcentration
Influenza A H1-20099Up to10xfInfluenza A H1N1pdm09 isolates from2009 and 2012None
Influenza A H3103xInfluenza A H3N2 isolates from the1960s to 2012, attenuatedvaccine/recombinant H3N2 viruses, andswine variant H3N2 (H3N2v)None
Influenza A(no subtype detected)orInfluenza A (Equivocal)103xgHuman H2N2 recombinantAvian H2N3, H5N1, H5N2, H5N3,H5N8, H7N7, H7N9, H10N7 (includingisolates from human infection)Variable resultsreported,depending onstrain andconcentration
Influenza B133xInfluenza B isolates from the 1940s to2012 and attenuatedvaccine/recombinant virusesNone
Parainfluenza Virus 143xMultiple strains of PIV1
Parainfluenza Virus 223xMultiple strains of PIV2None
Parainfluenza Virus 333xMultiple strains of PIV3
Parainfluenza Virus 443xSubtypes: A and B
Respiratory Syncytial Virus7Up to10xhSubtypes: A and BNone
Bacteria
Bordetella parapertussis (IS1001)63xStrains of B. parapertussis and B. bronchiseptica containing IS1001iNonej
Bordetella pertussis (ptxP)93xStrains of B. pertussisNone
Chlamydia pneumoniae43xStrains of C. pneumoniaeNone
Mycoplasma pneumoniae93xStrains of M. pneumoniaeNone

Table 27. Summary of BioFire RP2.1 Analytical Reactivity Testing and in silico Analysis®

{44}------------------------------------------------

ª Includes limitations observed in tosting by in silico analyses. Data for Alenovins, Sever Acute Respiratory
Syndrome Coronavirus 2 (SARS-CoV-2), and C. presincesting with

{45}------------------------------------------------

  • b In silico analysis of available sequences predicts that the BioFire RP2.1 will react with all currently characterized serotypes of Adenovirus, including B55, C57, several species D serotypes and G52, which were not tested.
  • 6 The reactivity assessment for SARS-CoV-2 also included fifty clinical specimens from the 2020 pandemic collected in three geographic regions of the United States.
  • 4 In silico analysis of available sequences predicts that the HRV/EV assay with all currently characterized species and serotypes of Human Rhinovirus and Enterovirus, including Polioviruses.
  • 6 Influenza A H1N1 isolates ATCC VR-95 (H1N1 PR/8/1934) and swine variant Hsw1N1, ATCC VR-897 (HswN1 ANew Jersey/8/76) were detected as Influenza A H1 at 10×LoD. All other H1N1 isolates were detected as Influenza A H1 at 1× or 3× LoD.
  • f Influenza A H1N1pdm09 isolate BEI NR-44345 (Hong Kong/H090-761-V1(0)/2009) was detected as Influenza A H1-2009 at 10× LoD. All other H1N1pdm09 isolates tested were detected as Influenza A H1-2009 at 1× or 3× LoD.
  • & All human and avian isolates (or genomic RNA) tested were reported as either Influenza A (no subtyped detected) at 3× LoD. Three strains reported as Influenza A Equivocal at the 3× LoD concentration were reported as Influenza A (no subtype detected) when tested at higher concentrations of 10× LoD (H2N2 strains, BEI NR-2775) or 100× LoD (Avian H2N3 Mallard/Alberta/9/2003). Avian H10N7 (Chicken/Germany/N/49; BEI NR-2765) was only reported as Influenza A Equivocal, even at a concentration equal to 100× LoD.
  • b Respiratory Syncytial Virus, subtype B, ATCC VR-1580 (Washington DC/1962) was detected at 10× LoD. All other RSV isolates tested (subtypes A and B) were detected at 1× or 3× LoD.
  • i Reactivity with IS 001 sequences in B. bronchiseptica represents the intended reactivity of the analyte will be inaccurately reported as B. parapertussis. The assay is not expected to react with IS 1001-like sequences in B. holmesii.

Analytical Specificity/Cross-reactivity: f.

The potential for cross-reactivity between all BioFire RP2.1 assays (including SARS-CoV-2) and various on-panel or off-panel organisms that may be present in clinical respiratory specimens was evaluated with a combination of empirical testing and in silico analysis. The isolates tested represent all organisms evaluated on the previousgeneration respiratory panel (i.e., BioFire FilmArray Respiratory Panel 2 (RP2)) as well as additional isolates selected to assess the specificity of the novel SARS-CoV-2 assays.

Over 25 on-panel organisms (Table 29) were tested at high concentrations to assess the potential for intra-panel cross-reactivity. For off-panel testing, organisms were selected that represent normal respiratory flora and pathogens that may be present in the respiratory tract, as well as near-neighbors or species genetically related to the organisms detected by the panel. The off-panel isolates included more than 65 bacterial, viral, and fungal species, including common causes of pharyngitis and upper or lower respiratory infection as well as other SARS or SARS-like coronaviruses (Table 30). On-panel and off-panel isolates were tested at the highest concentration possible (generally >1.0E+07 units/mL for bacteria and fungi and ≥1.0E+05 units/mL for viruses). These concentrations are equal to or greater than those tested on the FilmArray RP2. Each isolate was tested in triplicate (once on each of the three different pouch/reagent lots) according to standard testing procedure.

Most bacterial and fungal isolates were grown in-house using traditional culture methods and quantified in CFU/mL (plate counting or turbidity standards) or cells/mL (optical density (OD600) reading). Some difficult-to-culture organisms, such as obligate intracellular bacteria, were cultured by an outside source laboratory and documented, with a Certificate of Analysis, in standard quantification units (CFU/mL or cells/mL), infectivity units (TCID50 or IFU), or other measures such as color changing units (CCU) or cell counts (i.e. nuclei/mL).

{46}------------------------------------------------

Viruses were cultured by outside source laboratories and provided as culture fluids quantified in TCID50/mL or other relevant units (provided on a Certificate of Analysis from the culture collection). Five coronaviruses (SARS-CoV, SARS-CoV-2, MERS-CoV, and two bat SARS-like CoVs) were cultured and tested in a contracted biosafety level 3 laboratory. Two viruses (bocavirus and CoV HKU1) that are not amenable to in vitro culture were acquired for testing as clinical specimens and quantified in RNA copies/mL.

Organisms that could not be acquired as intact organism cultures were tested as purified preparations of genomic DNA (gDNA) and reported in units of genomic equivalents per mL (GE/ mL) or were evaluated via directed in silico analysis of publicly available whole genome sequences.

Any samples generating unexpected results were retested to determine if the results were reproducible and/or investigated to rule in or rule out contamination, isolate misidentification, or cross-reactivity. Confirmed cross-reactivities or limitations identified in this and previous studies are included in the descriptions for the BioFire RP2.1.

Over 90 organisms were tested with the BioFire RP2.1, and only those few crossreactivities previously described for the BioFire RP2 assays were identified in the testing (e.g. Bordetella vertussis ptxP assay cross-reactivity with the ptxP pseudogene in other Bordetella species; see Table 32). In silico evaluation of the SARS-CoV-2 assays did indicate a risk of cross-reactivity with closely related SARS-like coronaviruses of bat and pangolin origin, although these viruses are not predicted to be present in human clinical specimens.

All known or predicted risks of cross-reactivity for the BioFire RP2.1 are summarized in Table 31. No additional cross-reactivities were identified in this study for previously existing panel assays or for the novel SARS-CoV-2 assays.

OrganismIsolate IDConcentration TestedCross-Reactivity Detected
Bacteria
Bordetella parapertussisZeptometrix 08014626.43E+09 CFU/mLBordetella pertussis (ptxp)a
Bordetella pertussisATCC 97975.50E+09 CFU/mLHumanRhinovirus/Enterovirusb
Chlamydia pneumoniaeATCC 535921.93E+07 IFU/mLNone
Mycoplasma pneumoniaeZeptometrix 08015792.65E+07 CCU/mLNone
Viruses
Adenovirus7A (species B)Zeptometrix 0810021CF1.02E+07 TCID50/mLNone
1 (species C)Zeptometrix 0810050CF2.26E+07 TCID50/mLNone
4 (species E)ATCC VR-15721.58E+06 TCID50/mLNone

Table 28. On-Panel Organisms Tested for Evaluation of BioFire RP2.1 Analytical Specificity

{47}------------------------------------------------

OrganismIsolate IDConcentrationTestedCross-ReactivityDetected
Coronavirus 229EZeptometrix 0810229CF1.13E+05 TCID50/mLNone
Coronavirus HKU1Clinical specimen8.94E+06 RNAcopies/mLNone
Coronavirus NL63Zeptometrix 0810228CF2.34E+05 TCID50/mLNone
Coronavirus OC43Zeptometrix 0810024CF6.37E+06 TCID50/mLNone
Severe Acute Respiratory SyndromeCoronavirus 2 (SARS-CoV-2)USA-WA1/20202.4E+09 copies/mLNone
Human MetapneumovirusZeptometrix 0810159CF1.05E+06 TCID50/mLNone
Human Rhinovirus(Type 1A)Zeptometrix 0810012CFN8.40E+05 TCID50/mLNone
Enterovirus (D68)ATCC VR-18231.58E+07 TCID50/mLNone
Influenza A H1N1(A1/FM/1/47)ATCC VR-971.58E+08 CEID50/mLNone
Influenza A Hsw N1(A/NewJersey/8/76)ATCC VR-8978.89E+06 CEID50/mLInfluenza A H1-2009 c
Influenza A (H1N1) pdm09(Michigan/45/15)Zeptometrix 0810538CF9.40E+04 TCID50/mLNone
Influenza A H3N2(A/Alice)ATCC VR-7763.33E+08 CEID50/mLNone
Influenza B(Massachusetts/2/12)Zeptometrix 0810239CF9.55E+05 TCID50/mLNone
Parainfluenza Virus 1Zeptometrix 0810014CF6.80E+07 TCID50/mLNone d
Parainfluenza Virus 2Zeptometrix 0810357CF4.57E+06 TCID50/mLNone d
Parainfluenza Virus 3ATCC VR-936.80E+07 TCID50/mLNone
Parainfluenza Virus 4ATCC VR-13774.17E+04 TCID50/mLNone
Respiratory Syncytial VirusZeptometrix 0810040ACF7.00E+05 TCID50/mLNone

ª Bordetella pertussis (ptxP) assay may amplify pertussis toxin pseudogene sequences from some strains of B. parapertussis at high concentration (>1.2E+09 CFU/mL).

b Human Rhinovirus Enterovinus assay may amplify non-target sequences from Bordetella species (B. parapertussis, and B. bronchiseptica) at a concentration >4.5E+07 CFU/mL.

6 The H1 hemagglutinin (HA) gene of Influenza A H1N1 strains of swine origin (prior to 2009) will be amplified by the H1 assay (Influenza A H1 Detected). However, some strains/sequences of swine origin may also be amplified by the H1-2009 assay (Influenza A H1-2009 Detected) at higher concentrations. Testing of this strain at 8.89E+06 CEIDsoinL generated an Influenza A H1 Detected result in 1/3 replicates and an Influenza A H1-2009 Detected in 2/3 replicates.

d Parainfluenza Virus 3 (PIV3) was detected in 2/5 replicates of the Parainfluenza Virus 1 (PIV1) isolate tested and in 2/4 replicates of the Parainfluenza 2 isolate tested. Sequencing of the amplicons generated match PIV3 sequences, indicating contamination of both the PIV1 and PIV2 isolate stocks with PIV3 nucleic acid.

Table 29. Off-Panel Organisms Tested for Evaluation of BioFire RP2.1 Analytical Specificity
-----------------------------------------------------------------------------------------------
OrganismIsolate IDConcentration TestedCross-ReactivityDetected/Predicted
Bacteria
Acinetobacter calcoaceticusATCC 230555.15E+09 CFU/mLNone
Arcanobacterium haemolyticumATCC 93455.70E+09 CFU/mLNone
Bacillus anthracisEvaluated in silicoNone
Bordetella aviumATCC 350861.88E+09 cells/mLNone
Bordetella bronchisepticaATCC 105802.09E+09 cells/mLBordetella pertussis (ptxp) a
OrganismIsolate IDConcentration TestedCross-ReactivityDetected/Predicted
Bordetella hinziiATCC 517834.30E+06 CFU/mLNone
Bordetella holmesiiATCC 7000523.15E+07 CFU/mLNone
Burkholderia cepaciaATCC 177625.04E+09 CFU/mLNone
Chlamydia trachomatisZeptometrix 08017751.67E+08 IFU/mLNone
Chlaymdia psittaciEvaluated in silicoEvaluated in silicoNone
Corynebacterium diphtheriaeZeptometrix 08018827.47E+08 CFU/mLNone
Corynebacterium striatumATCC BAA-12935.20E+09 CFU/mLNone
Coxiella burnetiiEvaluated in silicoEvaluated in silicoNone
Escherichia coliCDC AR Bank #05385.53E+09 CFU/mLNone
Fusobacterium necrophorumATCC 278521.33E+08 cells/mLNone
Haemophilus influenzaeATCC 333915.85E+09 CFU/mLNone
Klebsiella (Enterobacter) aerogenesCDC AR Bank #00746.83E+09 CFU/mLNone
Klebsiella oxytocaJMI 78185.60E+09 CFU/mLNone
Klebsiella pneumoniaeNCTC 134651.75E+08 CFU/mLNone
Lactobacillus acidophilusZeptometrix 08015401.60E+08 CFU/mLNone
Lactobacillus plantarumZeptometrix 08015071.20E+09 CFU/mLNone
Legionella (Fluoribacter)bozemanaeATCC 332173.24E+09 cells/mLNone
Legionella (Fluoribacter) dumoffiiATCC 332792.65E+09 cells/mLNone
Legionella feeleiiATCC 358491.49E+09 cells/mLNone
Legionella longbeachaeZeptometrix 08015771.93E+08 CFU/mLNone
Legionella (Tatlockia) micdadeiZeptometrix 08015761.80E+09 CFU/mLNone
Legionella pneumophilaZeptometrix 08015301.75E+09 CFU/mLNone b
Leptospira interrogansATCC BAA-1198D-5(genomic DNA)7.89E+08 GE/mLNone
Moraxella catarrhalisATCC 81765.73E+09 CFU/mLNone
Mycobacterium tuberculosisZeptometrix 0801660(avirulent strain)9.07E+06 CFU/mLNone
Mycoplasma genitaliumATCC 33530D(genomic DNA)8.40E+07 GE/mLNone
Mycoplasma hominisZeptometrix 08040112.11E+09 CCU/mLNone
Mycoplasma oraleATCC 195241.00E+07 CCU/mLNone
Neisseria elongataZeptometrix 08015101.99E+08 CFU/mLNone c
Neisseria gonorrhoeaeATCC 194242.31E+09 CFU/mLNone
Neisseria meningitidisATCC 130901.99E+09 CFU/mLNone
Proteus mirabilisATCC 124535.60E+09 CFU/mLNone
Pseudomonas aeruginosaATCC 278534.33E+09 CFU/mLNone
Serratia marcescensJMI 6974.75E+09 CFU/mLNone
Staphylococcus aureus (MRSA)ATCC 108321.88E+08 CFU/mLNone
Staphylococcus epidermidisATCC 298874.95E+09 CFU/mLNone
Stenotrophomonas maltophiliaATCC 7004754.93E+09 CFU/mLNone
Streptococcus agalactiaeATCC 138135.45E+09 CFU/mLNone
Streptococcus dysgalactiaeATCC 430785.70E+09 CFU/mLNone
OrganismIsolate IDConcentration TestedCross-ReactivityDetected/Predicted
Streptococcus pneumoniaeATCC BAA-3415.20E+09 CFU/mLNone
Streptococcus pyogenesATCC 196155.46E+07 CFU/mLNone
Streptococcus salivariusATCC 134194.92E+09 CFU/mLNone
Ureaplasma urealyticumATCC 276181.00E+08 CCU/mLNone
Viruses (SARS-CoV-2 Related Coronaviruses)
Bat SARS-like CoronavirusBEI NR-44009(Recombinant withSARS Urbani RBD)3.15E+06 TCID50/mLNone
Bat SARS-like Coronavirus HKU5BEI NR-48814(Recombinant withSARS Urbani SE)1.95E+06 TCID50/mLNone
Severe Acute Respiratory SyndromeCoronavirus (SARS)BEI NR-18925Urbani strain5.3E+09 copies/mLNone
Viruses
BocavirusClinical specimen1.40E+08 copies/mLNone
Cytomegalovirus (CMV)Zeptometrix 0810003CF7.67E+06 TCID50/mLNone
Epstein-Barr Virus (EBV)Zeptometrix 0810008CF3.65E+07 copies/mLNone
Herpes Simplex Virus 1 (HSV1)ATCC VR-17783.30E+08 copies/mLNone
Herpes Simplex Virus 2 (HSV2)Zeptometrix 0810217CF1.30E+07 TCID50/mLNone
Human Herpes Virus 6 (HHV6)Zeptometrix 0810072CF4.11E+08 copies/mLNone
Human Parechovirus (HPeV)Zeptometrix 0810147CF2.26E+07 TCID50/mLNone
Influenza CEvaluated in silicoNone
Measles VirusZeptometrix 0810025CF1.63E+05 TCID50/mLNone
MumpsZeptometrix 0810079CF4.83E+05 units/mLNone
Fungi
Aspergillus flavusZeptometrix 08015981.15E+08 CFU/mLNone
Aspergillus fumigatusZeptometrix 08017165.47E+07 CFU/mLNone
Blastomyces dermatitidisATCC 26199D-2(genomic DNA)7.05E+07 GE/mLNone
Candida albicansATCC 102311.19E+06 CFU/mLNone
Cryptococcus neoformansATCC MYA-45646.00E+07 CFU/mLNone
Histoplasma capsulatumEvaluated in silicoNone
Pneumocystis jirovecii (carinii)ATCC PRA-1596.67E+07 nuclei/mLNone

{48}------------------------------------------------

{49}------------------------------------------------

B Bordetella pertussis (ptxP) assay may amplify pertussis toxin pseudogene sequences from some strains of B. bronchiseptica at high concentration (≥1.2E+09 CFU/mL).

b Bordetella parapertussis (IS1001) was detected in 3/3 replicates. Amplification with an alternate B. parapertussis/151001 PCR assay confirmed the presence of IS1001 nucleic acid in the L. pneumophila stock (contamination).

& Mycoplasma pneumoniae was detected in 3/3 replication with an alternate M. pneumoniae PCR assay (gyrB gene) confirmed the presence M. pneumoniae nucleic acids in the N. elongata stock (contamination).

{50}------------------------------------------------

Table 30. Predicted and Observed Cross-Reactivity of the BioFire RP2.1
Cross-reactive Organism(s)/Sequence(s)BioFire RP2.1 ResultDescription
Bat coronavirus RaTG132(accession# MN996532)Pangolin coronavirus3(accession# MT08407)Bat SARS-like coronavirus(accession# MG772933 and MG772934)Severe Acute Respiratory SyndromeCoronavirus 2(SARS-CoV-2)The SARS-CoV-2 assays can amplify a small selectionof sequences from closely related Sarbecovirusesisolated from bats and pangolin. The SARSCOV2-2assay is predicted to cross-react with all foursequences, while the SARSCOV2-1 assay will likelyonly cross-react with the bat coronavirus RaTG13.
Non-pertussis Bordetella species(e.g. Bordetella parapertussis, Bordetellabronchisepticaa)Bordetella pertussis (ptxP)bThe Bordetella pertussis (ptxP) assay can amplifypertussis toxin pseudogene sequences in B.bronchiseptica and B. parapertussis, primarily whenpresent at high concentrations (≥1.2E+09 CFU/mL).
Bordetella bronchisepticaa(with IS1001 sequences)Bordetella parapertussis (IS1001)Some strains of B. bronchiseptica carry IS1001insertion sequences identical to those carried by B.parapertussis. These sequences will be efficientlyamplified by the IS1001 assay and reported by BioFireRP2.1 as Bordetella parapertussis (IS1001).
Bordetella pertussisBordetella parapertussisBordetella bronchisepticaHuman Rhinovirus/EnterovirusdThe Human Rhinovirus/Enterovirus assay may amplifyoff-target sequences found in strains of B. pertussis, B.bronchiseptica, and B. parapertussis when present athigh concentration. Cross-reactivity with B. pertussiswas observed at a concentration of 4.5E+07 CFU/mLor higher.
Influenza A H1N1(swine origin)Influenza A H1-2009eThe Influenza A H1-2009 assay may react with H1hemagglutinin gene sequences from viruses of swineorigin.BioFire RP2.1 will report either Influenza A H1 orInfluenza A H1-2009, depending on the strain andconcentration in the sample.

a B. bronchiseptica infection is rare in humans and more common in domesticated animals ('kennel cough').

Cross-reactivity between the Bordetella pertussis (ptxP) assay and B. parapertussis will be reported as a co-detection (Bordetella parapertussis (IS 1001) Detected and Bordetella pertussis (ptxP) Detected); while cross-reactivity with most strains of B. bronchiseptica (that do not carry IS 1001) will be reported only as Bordetella pertussis (pt.rP) Detected.

& Cross-reactivity with B. parapertussis and B. bronchiseptica is predicted based on in silico analysis but was not observed when tested at a concentration of 1.2E+09 CFU/mL.

  • 4 Cross-reactivity between the Human Rhinovirus assays and B. pertussis or B. parapertussis will be reported as a co-detection (Bordetella pertussis (pts.P) Detected and Human Rhinovirus Detected or Bordetella parabertussis (IS1001) Detected and Human Rhinovirus Detected); while cross-reactivity with most strains of B. bronchiseptica (that do not carry IS 1001) will be reported (falsely) only as Human Rhinovirus Detected.
  • 6 Swine origin Hsw1N1 (ANew Jersey/8/1976 ; ATCC VR-897) was detected as either Influenza A H1-2009 at a concentration of 8.9E+06 CEID50/mL.

Select organisms that could not be acquired for empirical testing were evaluated via in silico analysis to assess the potential for cross-reactivity between the novel SARS-CoV-2 assay primers and the organism whole genome sequences. The isolates evaluated include medically relevant select agents (Bacillus anthracis and

2 Peng Zhou et al., "A Pneunonia Outbreak Associated with a New Coronavirus of Probable Bat Origin," Nature 579, no. 7798 (March 2020): 270-73, https://doi.org/10.1038/s41586-020-2012-7.

Rachele Caglian et al., "Computational Inference of Selection Underlying the Novel Coronavinus, SARS-CoV-2," Journal of Virology, April 1, 2020, JVI.00411-20, jvi;JVI.00411-20v1, https://doi.org/10.1128/JVI.00411-20.

{51}------------------------------------------------

Coxiella burnetii), a BSL3 fungal respiratory pathogen (Histoplasma capsulatum), a rare zoonotic fungal pathogen (Chlamydia psittaci), and a relatively underdiagnosed cause of influenza infection (Influenza C virus).

Analysis involved retrieval of at least three representative whole genome sequences for each organism from the publicly available GenBank database and subsequent BLAST analysis to identify regions with greater than 80% homology to the SARS-CoV-2 assay primers. Homology was evaluated for both outer and inner assay primers, although only reactivity with inner primers is essential to generate a crossreactive result using the nested FilmArray system. When regions with >80% homology to at least one inner assay primer were identified, the sequences were further assessed to determine the potential for bidirectional amplification and generation of a detectable amplicon sequence (<2000 bp).

Although some regions with >80% homology to individual outer or inner primer sequences were identified (see Table 32), no primer binding sites were located in an orientation to allow for bi-directional amplification of complementary strands nor within sufficient proximity to generate an appropriately sized, detectable amplicon. As a result, no risk of cross-reactivity was identified for the organisms evaluated.

Table 31. In silico Evaluation of Potential Cross-Reactivity Between SARS-CoV-2 Assay Primers and Select Off-Panel
Organisms
Percent homology under SARS-CoV-2 assay outer forward (IF), inner reverse (IR), and outer forward (OF)
primers. Results for outer reaction primers are shaded grey. Sequences with >80% homology to an inner are shaded
green. Sequences with no homology to assay primers are denoted NH (no homology).
OrganismGenBank AccessionNo.SARSCoV2-1 AssayPrimersSARSCoV2-2 Assay PrimersPredictedCross-reactivity
OFIFIROROFIFIROR
BacillusanthracisNC_005945.1NHNHNHNH85%83%NH81%None
CP012728.1NHNHNHNH85%83%NH81%
NZ_KN050648.1NHNHNHNH85%83%NH81%
ChlamydiapsittaciNC_020248.1NHNHNHNH85%NHNH81%None
NC_015470.1NHNHNHNH85%NHNH81%
NZ_KE355746.1NHNHNHNH85%NHNH81%
CoxiellaburnetiiNC_00297181%NHNHNH80%NHNHNHNone
CP00073381%NHNHNH80%NHNHNH
CP00755581%NHNHNH80%NHNHNH
CP01815081%NHNHNH80%NHNHNH
HG82599081%NHNHNH80%NHNHNH
NC_004704NHNHNHNH80%NHNHNH
CP000735NHNHNHNH80%NHNHNH
CP000914NHNHNHNH80%NHNHNH
HistoplasmacapsulatumABBT0000000081%NHNHNH80%83.3%81.8%81%None
AAJI0000000081%NHNHNH80%83.3%81.8%81%

{52}------------------------------------------------

OrganismGenBank AccessionNo.SARSCoV2-1 Assay PrimersSARSCoV2-2 Assay PrimersPredictedCross-reactivity
OFIFIROROFIFIROR
ABBS00000000081%NHNHNH80%83.3%81.8%81%
ABRJ000000000081%NHNHNH80%83.3%81.8%81%
ABRK00000000081%NHNHNH80%83.3%81.8%81%
Influenza CVirus1419 SequencesEvaluatedaNHNHNHNHNHNHNHNHNone

a GenBank accession numbers evaluated for Influenza C cross-reactivity analysis were provided separately.

g. Assay cut-off:

Not applicable

  • h. Interfering Substances:
    Testing for possible interference from select substances that may be present in nasopharyngeal swabs (NPS) was performed with the BioFire RP2.1. The majority of interference data was collected in the interference study performed with the predecessor panel, BioFire FilmArray Respiratory Panel 2 (RP2). Both the BioFire RP2.1 and BioFire RP2 test the same sample type using the same pouch chemistry and cycling conditions. The only difference between the panels is the addition of the SARS-CoV-2 assays; therefore, the previous BioFire RP2 interference study results remain relevant and are applied to the BioFire RP2.1 as appropriate.

Substances evaluated for interference generally were categorized as-endogenous substances (i.e., biological substances naturally in NPS samples), competing microorganisms (i.e., pathogens or natural flora tested at high concentrations to evaluate polymicrobial NPS specimens), exogenous substances (i.e., non-native substances in NPS samples), or technique-specific substances (i.e., substances introduced during sample processing, collection, or testing).

The endogenous and technique-specific substances, as well as competing microorganisms, evaluated on the BioFire RP2.1 in this study were selected to assess the risk of possible interference with the SARS-CoV-2 assays as well as the other panel assays. Substances were also selected to reproduce a subset of the data collected in the BioFire RP2 study in order to verify that the data collected with one panel could be applied to the other.

Each substance was added to contrived samples containing representative organisms at concentrations near (2-3x) LoD (Table 33). The concentration of substance added to the samples was equal to or greater than the highest level expected to be in NPS specimens. The organisms in the sample included the new analyte (SARS-COV-2), two analytes included in the RP2 interference study (Adenovirus and Bordetella parapertussis), and three additional analytes not previously evaluated in the RP2 interference study (Coronavirus NL63, Influenza A H1N1pdm09, and Respiratory

{53}------------------------------------------------

Syncytial Virus). The test sample was composed to represent the types of analytes detected by the panel, including bacteria and enveloped or non-enveloped viruses with DNA and RNA genomes.

OrganismDescriptionSource3x LoD(Sample)Concentration
Severe Acute RespiratorySyndrome Coronavirus 2 (SARS-CoV-2)New analyte for RP2.1RNA genome (+ strand)enveloped virusATCC VR-1986HK(heat inactivated)1.5E+03copies/mL
Adenovirus (Type 2)DNA genome (double-stranded)non-enveloped virusWHO InternationalStandardNIBSC 16/3249.0E+03 IU/mL
Coronavirus NL63RNA genome (+ strand)enveloped virusBEI NR-4707.5E-01TCID50/mL
Influenza A H1N1pdm09 (H1-2009)RNA genome (- strand)enveloped virusZeptometrix0810109CFN1.5E+00TCID50/mL
Respiratory Syncytial VirusRNA genome (+ strand)enveloped virusZeptometrix 810040ACF6.0E-02TCID50/mL
Bordetella parapertussisGram-negative bacteriumDNA genomeZeptometrix 08014611.8E+02IS1001copies/mL

Table 32. Contrived Sample Composition for BioFire RP2.1 Interference Testing

Testing near LoD was to identify the effects of even minor interference on analyte detection. A control sample with no substance (positive control) was tested on each day of evaluation to demonstrate the expected detection without any potential interference. Each potential interferent was also added to a negative sample (substance only negative control) and tested in tandem with the corresponding positive sample to serve as a control for the substance alone (e.g. detection of an onpanel competing microorganism). Each type of sample (positive control, negative control, and sample with interfering substance) was tested in triplicate, with one replicate on each of three different pouch lots. The pouch control and analyte results from the spiked sample with potential interferents and negative control samples were compared to the results from the positive control sample replicates to evaluate whether there was interference in detection.

In addition, in silico analysis of the SARS-CoV-2 assays during assay development identified some closely related coronaviruses with >80% homology under one or more assay primers. Of those, three (Bat CoV KY770858, Bat CoV GU190215 and SARS-like coronavirus KR559017) show the indicated homology to only the outer forward primer and are, therefore, not expected to react with the SARS-CoV-2 assays nor present a risk of interference. However, four sequences from the proposed precursors to SARS-CoV-2 (Bat CoV RTG13 MN996532, Pangolin CoV MT084071, and Bat SARS-like coronavirus MG772933 and MG772934) are predicted to cross-react with one or more of the BioFire RP2.1 SARS-CoV-2 assays. These viruses could not be acquired for interference testing. However, none of these

{54}------------------------------------------------

viruses have been isolated from human infections. Therefore, the risk of interference with the presence of these coronaviruses in polymicrobial infections with SARS-CoV-2 appears to be negligible.

The combination of BioFire RP2.1 and BioFire RP2 interference testing evaluated a total of forty-three substances (Table 34). The notable results from testing are summarized as follows-a total of 41 pouch runs were performed on both FilmArray Torch and FilmArray 2.0 systems and all completed with valid results (i.e., no errors). Overall, testing with potentially interfering substances at high, "worst-case scenario" concentrations (Table 35) demonstrated no interference with pouch controls or detection of panel analytes. The only exception was with bleach, resulting in missed detection of various panel analytes likely due to damaged nucleic acids (this was also observed in the previous BioFire RP2 panel). A general warning to avoid contact between samples and bleach is noted.

Endogenous substance interference testing with the BioFire RP2.1 confirmed that presence of blood and human genomic DNA in samples had no effect on detection of low-level SARS-CoV-2 or other analytes, consistent with results obtained with these substances in the BioFire RP2 evaluation. In addition, no interference with SARS-CoV-2 detection was observed when testing a clinically relevant concentration of human peripheral blood mononuclear cells (PBMCs).

Competing microorganism testing with the BioFire RP2.1 showed that there was no effect with a high concentration of another human beta-coronavirus that can cause respiratory illness (Coronavirus OC43), nor a high concentration of a bacterium representing normal flora of the respiratory tract (Streptococcus salivarius). During evaluation of the potential Streptococcus salivarius interferent, one spiked substance replicate vielded a no detection with the Respiratory Syncvtial Virus analyte. However, the lack of detection was not reproducible in the two repeated replicates on the same pouch lot.

For technique-specific substance testing, no interference was observed when samples were prepared and tested in PrimeStore Molecular Transport Medium (MTM).

Table 33. Results from the Evaluation of Potentially Interfering Substance Effects on Analyte Detection - FilmArray RP2 on BioFire RP2.1

Substances tested with the BioFire RP2.1 are in bold fort. Results for substances tested only with the BioFire FilmArray RP2 previously are applied to BioFire RP2.1

Substance TestedConcentration TestedResult
Endogenous Substances
Human Whole Blood10% v/vNo Interference
Human Mucus (Sputum)1 swab/mL sampleNo Interference
Human Genomic DNA20 ng/µLNo Interference
Human Peripheral Blood Mononuclear Cells (PBMCs)1.0E+03 cells/µLNo Interference
Competitive Microorganisms
Coronavirus 229E1.7E+04 TCID50/mLNo Interference
Substance TestedConcentration TestedResult
Coronavirus OC43 (betacoronavirus)9.6E+05 TCID50/mLNo Interference
Adenovirus A128.9E+05 TCID50/mLNo Interference
Parainfluenza Virus 36.6E+05 TCID50/mLNo Interference
Bordetella pertussis5.8E+08 CFU/mLNo Interference
Enterovirus D681.6E+07 TCID50/mLNo Interference
Echovirus 61.0E+07 TCID50/mLNo Interference
Respiratory Syncytial Virus4.2E+04 TCID50/mLNo Interference
Staphylococcus aureus2.5E+07 CFU/mLNo Interference
Streptococcus pneumoniae1.7E+07 CFU/mLNo Interference
Streptococcus salivarius2.5E+09 CFU/mLNo Interference
Haemophilus influenzae6.2E+07 CFU/mLNo Interference
Candida albicans1.0E+06 CFU/mLNo Interference
Herpes Simplex Virus 11.6E+06 TCID50/mLNo Interference
Cytomegalovirus1.2E+06 TCID50/mLNo Interference
Exogenous Substances ª
Tobramycin (systemic antibiotic)0.6 mg/mLNo Interference
Mupirocin(active ingredient in anti-bacterial ointment)2% w/vNo Interference
Saline Nasal Spray with Preservatives(0.65% NaCl, Phenylcarbinol, Benzalkonium chloride)1% v/vNo Interference
Nasal Decongestant Spray(Oxymetazoline HCl 0.05%, Benzalkonium chloride, phosphate)1% v/vNo Interference
Analgesic ointment (Vicks®VapoRub®)1% w/vNo Interference
Petroleum Jelly (Vaseline®)1% w/vNo Interference
Snuff (Tobacco)1% w/vNo Interference
Disinfecting/Cleaning Substances
Bleach1% and 2% v/v[up to 1024 ppm chlorine]Interference b
Disinfecting wipes (ammonium chloride)½ in²No Interference
Ethanol7% v/vNo Interference
DNAZap (Ambion AM9891G & AM9892G)1% v/vNo Interference
RNaseZap (Ambion AM9782)1% v/vNo Interference
Specimen Collection Materials
Rayon Swabs (Copan 168C)N/ANo Interference
Nylon Flocked Swabs (Copan 553C)N/ANo Interference
Polyester Swabs (Copan 175KS01)N/ANo Interference
Calcium Alginate Swabs (Puritan 25-801 A 50)N/ANo Interference
M4 Transport Medium (Remel)100%No Interference
M4-RT Transport Medium (Remel)100%No Interference
M5 Transport Medium (Remel)100%No Interference
M6 Transport Medium (Remel)100%No Interference
Universal Viral Transport vial (BD)100%No Interference
PrimeStore Molecular Transport Medium70% v/vNo Interference
Sigma-Virocult Viral Collection and Transport System100%No Interference
Substance TestedConcentration TestedResult
(Swab and Transport Medium)
Copan ESwab Sample Collection and Delivery System(Swab and Liquid Amies Medium)100%No Interference

{55}------------------------------------------------

{56}------------------------------------------------

4 Nasal vaccines (e.g. FluMist) were not evaluated, but are predicted to be reactive with the BioFire RP2.1 Influenza A (subtype) and Influenza B assays.

b Not Detected results were reported for several analytes after incubation of the sample with 2% bleach for 10 minutes or ovenight. It was concluded that interference resulted prinarily from damage to the organisms hucleic acids in the sample, rather than inhibition or interference with pouch function(s).

i. Carry-Over Contamination:

A formal carry-over study in support of this regulatory submission for the BioFire RP2.1 was not performed since carry-over studies with high positive samples followed by negative samples have been performed for other FDA-cleared FilmArray Panels that are similar to the RP2.1 (i.e., BioFire FilmArray RP, BCID, and GI) for both the FilmArray 2.0 and the FilmArray Torch systems, and no significant carryover has been observed.

2. Comparison studies:

a. FDA SARS-CoV-2 Reference Panel Testing:

Quality assessments such as evaluating the BioFire RP2.1 device with the FDA SARS-CoV-2 reference panel, provide additional information on the relative sensitivity and specificity of the included SARS-CoV-2 assays. An evaluation of SARS-CoV-2 sensitivity and MERS-CoV cross-reactivity was performed using reference material, blinded samples, and a standard protocol provided by FDA. The study included a range finding study and a confirmatory study for LoD. Blinded sample testing was used to establish specificity and to confirm the LoD.

The results are summarized in the following table-

Table 34. Summary of LoD Confirmation Result using the FDA SARS-CoV-2 Reference Panel
Reference MaterialsProvided by FDASpecimen TypeLoD ConcentrationCross-Reactivity
SARS-CoV-2NPS intransport medium6.0E+03 NDU/mLaN/A
MERS-CoVNPS intransport mediumN/ANot Detected

4 NDUmL = RNA NAAT detectable units/mL

  • b. Matrix comparison:
    Not applicable

{57}------------------------------------------------

3. Clinical studies:

Prospective Clinical Study

A clinical evaluation of the BioFire RP2.1 panel was performed with prospectively collected NPS specimens. Specimens that were residual NPS in transport media left over from standard of care testing for SARS-CoV-2, and those that were either held at room temperature for <4 hours or 4℃ for <3 days before enrollment were accepted for this study. Specimens that were other than NPS in transport media or specimens that could not be tested within the defined storage parameters were excluded. Three collection sites were used in this prospective clinical study.

The BioFire RP2.1 was evaluated by comparing the test results for SARS-CoV-2 with a composite comparator of three U.S. FDA EUA tests. All results were interpreted according to the test's IFU.

Concordance for two out of three of the EUA tests were considered the final result for the comparator, and the interpretation rules for the reference method were outlined as follows-

Rule #EUA ResultsComposite Result
1Pos/Pos/AnyPositive
2Neg/Neg/AnyNegative
3Pos/Neg/Invspecimen excluded
4Inv/Inv/Anyspecimen excluded
Table 35. BioFire RP2.1 Prospective Clinical Evaluation
Composite Comparator Interpretations Rules3

a 'Any' may be positive, negative, or invalid. 'Inv' (invalid) results include any non-definitive result such as equivocal. indeterminate, unresolved, or inconclusive.

A total of 534 NPS specimens were acquired for the clinical study. In terms of the inclusion criteria for the study, a total of 527 specimens initially were included in the analysis. Of these 311 (59.0%) were run on BioFire 2.0 systems and 216 (41.0%) were run on BioFire Torch systems. Two tests did not complete the initial run resulting in a total instrument success rate of 99.6% (525/527). One specimen was able to be rerun. Ten (10) NPS specimens were excluded for reasons that included-not meeting inclusion criteria after enrollment (insufficient volume, N=1; stored at incorrect temperature, N=6), a run failure with insufficient volume for retesting (N=1), and inability to determine composite comparator interpretation for a specimen due to invalid comparator results (N=2).

{58}------------------------------------------------

The final data was comprised of 524 valid specimens with the following demographic information-

Overall
SexMale270 (52%)
Female251 (48%)
Unknown3 (<1%)
Age0-18 years55 (10%)
19-40 years170 (32%)
41-60 years146 (28%)
61+ years153 (29%)
Total524

Table 36. Demographic Data of Prospectively Collected Specimens

Note that two ECMs were provided to study sites for daily testing (one each as positive and negative controls). Instrument operators were required to complete a valid ECM run on each day of specimen testing. All ECM sample runs were completed and yielded the expected results.

A summary of the BioFire RP2.1 SARS-CoV-2 prospective clinical study performance is provided in Table 38. Positive Percent Agreement (PPA) was calculated as 100% × (TP / (TP + FN)). True positive (TP) indicates that both the BioFire RP2.1 and the comparator method had a positive result for the specific analyte, and false negative (FN) indicates that the BioFire RP2.1 was negative while the comparator result was positive. Negative Percent Agreement (NPA) was calculated as 100% × (TN + FP)). True negative (TN) indicates that both the BioFire RP2.1 and the comparator method had negative results, and false positive (FP) indicates that the BioFire RP2.1 was positive while the comparator result was negative.

Table 37. BioFire RP2.1 Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Overall Prospective Study Performance

AnalytePositive Percent AgreementNegative Percent Agreement
TP/(TP + FN)%95%CITN/(TN + FP)%95%CI
Severe Acute RespiratorySyndrome Coronavirus 2 (SARS-CoV-2)61/62a98.491.4-99.7%457/462b98.997.5-99.5%

a SARS-CoV-2 was detected in the single FN specimen with all three composite comparator methods.

b SARS-CoV-2 was detected in 4/5 FP specimens with only one of the three composite comparator methods. SARS-CoV-2 was detected in the remaining FP specimen (1/5) using an additional independent molecular method.

Overall, a single (1) False Negative and five (5) False Positive discrepant results were observed between the RP2.1 and the reference comparator methods in the prospective clinical study. Regarding the observed False Negative result, further investigation

{59}------------------------------------------------

indicated that all comparator assays had positive results, but the respective Ct and Cp values suggested that the analyte concentration in the sample may have been near the LoD according to the manufacturers' IFUs. Due to insufficient sample volume, this specimen could not be further investigated.

SCNDiscrepancySubjectInformationComparator EUA PCR Tests
Age(years)SexComparator 1Comparator 2Comparator 3
Composite EUA ResultResultT1(Nsp2) CtT2(N) CtResult(b) (4)ResultN1 CtN2 Ct
01-0301FN19-40MPositivePos31.9831.25DPos34.136.0
Table 38. False Negative SARS-CoV-2 RP2.1 Comparator Prospective Study Result
-------------------------------------------------------------------------------------

Investigation of the five False Positive specimens indicated that analyte concentrations may have been near LoD based on observed Cp values for all five RP2.1 runs. The SOC assay was negative for all five specimens. Four specimens had a Detected or Equivocal result on the COVID-19 reference assay with a varying number of positive assays for that device. For the fifth specimen. the study site initiated additional SOC testing based on their internal review of the amplification data. That specimen was further tested with another FDA authorized test, with a subsequent positive result. It is also noted that the observed mean Ct values for both reference tests suggested an analyte concentration near LoD, according to the manufacturer IFUs. Due to insufficient sample volume, these specimens could not be further investigated.

SubjectInformationBioFire RP2.1Comparator EUA PCR TestsAdditionalInvestigation
DiscrepanAge(years)SexMedian CpComparator 2FDA authorizedtest
SCNSARSCoV2-SARSCoV2-2CompositeEUA ResultComparatorResult(b) (4)(b) (4)Comparator3ResultT1(N2)CtT2(E)Ct
01-0042FP41-60M(b)NegativeNegENegNTa
01-0107FP0-18F(b)NegativeNegDNegNTa
01-0113FP19-40F(b)(b)NegativeNegDNegNTa
01-0143FP19-40E(b)NegativeNegENegNTa
01-0236FP61+F(b)-NegativeNegNDNegPos38.236.0

Table 39. False Positive SARS-CoV-2 RP2.1 Comparator Prospective Study Results

a NT = Not Tested

Retrospective Clinical Study

A clinical evaluation of the BioFire RP2.1 panel was performed using 50 natural retrospective leftover (archived) clinical specimens. All SARS-CoV-2 positive specimens were collected during March and April of 2020. Known demographic information for these collected samples is summarized in (Table 41). These specimens had been

{60}------------------------------------------------

previously characterized as positive for SARS-CoV-2 using the respective Emergency Use Authorization (EUA) assay employed at the collection site (i.e., the comparator method listed in the archived specimen testing). Specimens were obtained from three geographically distinct laboratories in the United States. Positive specimens were randomized and tested alongside 50 NPS specimens that were collected before December 2019 (i.e., expected to be negative for SARS-CoV-2). Positive Percent Agreement (PPA) was determined by comparing the observed test result to the expected test result based on previous laboratory testing, and Negative Percent Agreement (NPA) was determined by comparing the observed test result for SARS-CoV-2 negative specimens to the expected result of Not Detected. In addition, the LoD of RP2.1 SARS-CoV-2 assays were comparable to the reported LoDs of the EUA comparator tests (according to the manufacturer IFU) used during this retrospective clinical study.

Table 40. Demographic Summary for Positive SARS-CoV-2 Archived Specimens

Specimen Demographics (N=50)
Male (%)15 (30%)
SexFemale (%)20 (40%)
Unknown15 (30%)
0-18 years1 (2%)
19-40 years13 (26%)
Age Range41-60 years13 (26%)
61+ years8 (16%)
Unknown15 (30%)

In the course of testing, two specimens (one positive and one negative) were excluded due to instrument errors. Results from the remaining 98 evaluable specimens are shown in Table 46 below. The PPA was 98% (48/49) and NPA was 100% for the SARS-CoV-2 assay. One false negative (FN) result was observed (the specimen was positive upon retest). Further, 10.4% (5/48) of the 48 specimens with SARS-CoV-2 Detected results had additional analytes identified by BioFire RP2.1 in this retrospective study (Table 43).

Agreement with known analyte composition
Comparator MethodPPA: TP/(TP+FN)%95% CINPA: TN/(TN+FP)%95% CI
EUA 114/15a93.3[70.2-98.8%]N/AN/AN/A
EUA 215/15100[79.6-100%]N/AN/AN/A
EUA 319/19100[83.3-100%]N/AN/AN/A
Negative SpecimensN/AN/AN/A49/49100[92.7 - 100%]
Overall Agreement48/49a98.0[89.3 -99.6%]49/49100[92.7 - 100%]

Table 41. BioFire RP2.1 SARS-CoV-2 Archived Specimen Performance Data Summary

{61}------------------------------------------------

Table 42. Additional Analytes identified by BioFire RP2.1 in 48 specimens with SARS-CoV-2
Detected Results
Additional AnalytesNumber Observed (%)
Adenovirus1 (2.1%)
HRV/EV4 (8.3%)

Contrived Clinical Specimen Study

The BioFire RP2.1 clinical specimen study included testing of 50 contrived clinical specimens spiked with inactivated SARS-CoV-2 isolate USA-WA1/2020 at various levels of LoD (25 at 2× LoD, 15 at 3× LoD, and 10 at 5× LoD) and randomized with ten non-spiked specimens. Each specimen was a unique NPS specimen which had been collected before December 2019, and was therefore expected to be negative for SARS-CoV-2. PPA was determined by comparing the observed test results for samples contrived in unique clinical specimens to the expected Detected result. PPA and NPA are shown in Table 48. For SARS-CoV-2 contrived testing, both the PPA and NPA were 100%.

Table 43. Contrived SARS-CoV-2 Testing

Agreement with known analyte composition
PPA: TP/(TP+FN)0/0NPA:TN/(TN+FP)0/0
Overall Agreement50/50100%10/10100%
95% CI[92.9-100%][72.2-100%]

(b) (4)

{62}------------------------------------------------

Clinical Comparison Study

A clinical comparison study was performed to demonstrate equivalency between the existing BioFire RP2 and the new BioFire RP2.1 panel that is a modification of the RP2 panel with the addition of assays for the detection of SARS-CoV-2. The clinical comparison study used 210 clinical archived nasopharyngeal swab (NPS) specimens eluted in transport media. Note, the firm also evaluated 10 MERS-CoV contrived specimens as part of their internal characterizations for the comparison study. However, the BioFire RP2.1 does not report results for MERS-CoV. The clinical NPS specimens were previously obtained during prospective evaluations of other BioFire respiratory panels or acquired from external sources for the BioFire specimen repository. Specimens were chosen solely based on the analyte content, and analyte levels (if known) were not used as part of the specimen selection process. Specimens for use in this study were collected prior to December 2019, and were therefore presumed to be negative for SARS-CoV-2.

Specimens were stored frozen and maintained at <- 70°C until testing and were split into two aliquots for parallel testing with each panel. An attempt was made to test at least ten archived clinical specimens for every analyte shared between RP2 and RP2.1 panels. The archived and contrived specimens were evaluated at BioFire laboratories using FilmArray 2.0 and Torch systems. Prior to testing, specimens were randomized and the analyte contents remained blinded to the personnel performing the test procedures. Specimen

{63}------------------------------------------------

aliquots were subject to the same number of freeze/thaw cycles to prevent bias in testing.

The comparison performance is reported as Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) of the BioFire RP2 to BioFire RP2.1 panels. The results from the comparison evaluation is summarized in the below table and it was observed that there was overall a 97.7% positive percent agreement (PPA) and 99.8% negative percent agreement (NPA) (Table 45). The majority of discrepant results were due to unexpected positives by either panel (i.e., detection of analytes not previously identified by the source lab in addition to the expected analyte). These discrepant results are summarized below along with the corresponding expected analyte detection and Cp values (Table 46).

Positive AgreementNegative Agreement
AnalyteRP2 (+)RP2 (+)RP2.1 (-)PPA95% CIRP2 (-)RP2 (-)RP2.1 (+)NPA95% CI
RP2.1 (+)RP2.1 (-)
Viruses
Adenovirus14l14/15(93.3%)70.2-98.8%2032203/205(99%)96.5-99.7%
Coronavirus 229E10110/11(90.9%)62.3-98.4%2090209/209(100%)98.2-100%
Coronavirus HKU110010/10(100%)72.2-100%2082208/210(99%)96.6-99.7%
Coronavirus NL6310010/10(100%)72.2-100%2100210/210(100%)98.2-100%
Coronavirus OC4310010/10(100%)72.2-100%2100210/210(100%)98.2-100%
HumanMetapneumovirus12012/12(100%)75.8-100%2080208/208(100%)98.2-100%
HumanRhinovirus/Enterovirus19319/22(86.4%)66.7-95.3%1953195/198(98.5%)95.6-99.5%
Influenza A30030/30(100%)88.6-100%1900190/190(100%)98.0-100%
Influenza A H150ર્ડાર(100%)56.6-100%2150215/215(100%)98.2-100%
Influenza A H1-200912012/12(100%)75.8-100%2080208/208(100%)98.2-100%
Influenza A H313013/13(100%)77.2-100%2070207/207(100%)98.2-100%
Influenza B10010/10(100%)72.2-100%2100210/210(100%)98.2-100%

Table 44. Performance Comparison of the Modified BioFire RP2.1 to the Original BioFire FilmArray RP2 using Archived and Contrived Specimens

{64}------------------------------------------------

Positive AgreementNegative Agreement
AnalyteRP2 (+)RP2.1 (+)RP2 (+)RP2.1 (-)PPA95% CIRP2 (-)RP2.1 (-)RP2 (-)RP2.1 (+)NPA95% CI
Parainfluenza Virus 1909/9(100%)70.1-100%2110211/211(100%)98.2-100%
Parainfluenza Virus 211011/11(100%)74.1-100%2090209/209(100%)98.2-100%
Parainfluenza Virus 310110/11(90.9%)62.3-98.4%2081208/209(99.5%)97.3-99.9%
Parainfluenza Virus 411011/11(100%)74.1-100%2090209/209(100%)98.2-100%
Respiratory SyncytialVirus10010/10(100%)72.2-100%2100210/210(100%)98.2-100%
Middle EastRespiratory SyndromeCoronavirus (MERS-CoV )a10010/10(100%)72.2-100%2100210/210(100%)98.2-100%
Bacteria
Bordetella parapertussis(IS1001)10010/10(100%)72.2-100%2100210/210(100%)98.2-100%
Bordetella pertussis(ptxP)10010/10(100%)72.2-100%2100210/210(100%)98.2-100%
Chlamydia pneumoniae10010/10(100%)72.2-100%2100210/210(100%)98.2-100%
Mycoplasmapneumoniae10010/10(100%)72.2-100%2100210/210(100%)98.2-100%
Overall
Overall2566256/262(97.7%)95.1-98.9%457084570/4578(99.8%)99.7-99.9%

ª MERS-CoV is not reported by the RP2.1 device. The analyte was evaluated in contrived specimens with inactivated virus.

Table 45. Summary of Discrepant Results in Comparison Testing between BioFire RP2.1 and RP2 devices. Discrepant analytes are shown in bold text and highlighted specimens were positive for the expected analyte.

All Reported Results
SampleNumberExpected Analyte Based onSource Lab Result(b) (4)
Expected Analyte Detected by Both Methods
053953-026Coronavirus OC43(b) (4)
053953-033Parainfluenza Virus 1
053953-078Mycoplasma pneumoniae
053953-083Human Metapneumovirus

{65}------------------------------------------------

All Reported Results
SampleNumberExpected Analyte Based onSource Lab Result(b) (4)
053953-085Human Metapneumovirus(b) (4)
053953-099Parainfluenza Virus 4
053953-123RSV
053953-134Parainfluenza Virus 1
053953-178Bordetella parapertussis
053953-199Negative
053953-204Bordetella parapertussis
053953-219Coronavirus 229E
Expected Analyte Not Detected by At Least One Method
053953-002Parainfluenza Virus 3(b) (4)
053953-112Coronavirus 229E

This study also served to demonstrate SARS-CoV-2 assay specificity with testing of 220 NPS specimens collected before December 2019 (i.e., presumptively negative for SARS-CoV-2). The NPA for SARS-CoV-2 was 100% (220/220) as observed for the BioFire RP2.1 panel (Table 47).

Table 46. Overall BioFire RP2.1 NPA (Specificity) for SARS-CoV-2 in Clinical Comparison Study
-----------------------------------------------------------------------------------------------------------
NPA: TN/(TN+FP)%95% CI
220/220100%98.3 - 100%

4. Clinical cut-off:

Not applicable

    1. Expected values:
      Prevalence of RP2.1 results in the prospective study stratified by study site are presented in the following table. Only three analytes were observed-Adenovirus, SARS-CoV-2, and Human Rhinovirus/Enterovirus.

{66}------------------------------------------------

Overall (N=524)Site 1 (N=309)Site 2 (N=110)Site 3 (N=105)
BioFire RP2.1 Result#EV#i#EV#EV
Viruses
Adenovirus30.6%31.0%00%00%
Coronavirus 229E00%00%00%100%
Coronavirus HKU100%00%00%00%
Coronavirus NL6300%00%00%00%
Coronavirus OC4300%00%00%00%
Middle East Respiratory Syndrome Coronavirus (MERS-CoV)00%00%00%00%
Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)6612.6%4614.9%1210.9%87.6%
Human Metapneumovius00%00%00%00%
Human Rhinovirus/Enterovirus336.3%12$.9%1110.0%109.5%
Influenza A00%00%00%00%
Influenza A H100%00%00%100%
Influenza A H1-200900%00%00%100%
Influenza A H300%00%00%00%
Influenza B00%00%00%00%
Parainfluenza Virus 100%00%00%00%
Parainfluenza Virus 200%00%00%00%
Parainfluenza Virus 300%00%00%00%
Parainfluenza Virus 400%00%00%00%
Respiratory Syncytial Virus00%00%00%00%
Bacteria
Bordetella parapertussis (IS1001)00%00%00%00%
Bordetella pertussis (ptxP)00%00%00%00%
Chlamydia pneumoniae00%00%00%00%
Mycoplasma pneumoniae00%00%00%00%

Table 47. Prevalence of Detected Analytes Stratified by Site; #= Number; EV = Expected Value

{67}------------------------------------------------

The prevalence of RP2.1 detected analytes stratified by age are as follows-

Biofire RP21 ResultOverall (N=524)0-18 years (N=55)19-40 years (N=170) 41-60 years (N=146)61+ years (N=153)
EV#EVEV#ev*EV
Viruses
Adenovirus30.6%11.8%21.2%00%00%
Coronavirus 229E00%00%00%00%00%
Coronavirus HKU100%00%00%00%00%
Coronavirus NL6300%00%00%00%00%
Coronavirus OC4300%00%00%00%00%
Middle East Respiratory Syndrome Coronavirus (MERS-Col/)00%00%00%00%00%
Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)6612.6%59.1%2414.1%2215.1%159.8%
Human Metapneumovirus00%00%00%00%00%
Human Rhinovirus/Enterovirus336.3%1934.5%52.9%74.8%21.3%
Influonza A00%00%00%00%00%
Influenza A H100%00%00%00%00%
Influenza A H1-200900%00%00%00%00%
Influenza A H300%00%00%00%00%
Influenza B00%00%00%00%00%
Parainfluenza Virus 100%00%00%00%00%
Parainfluenza Virus 200%00%00%00%00%
Parainfluenza Virus 300%00%00%00%00%
Parainfluenza Virus 400%00%00%00%00%
Respiratory Syncytial Virus00%00%00%00%00%
Bacteria
Bordetella parapertussis (IS 1001)00%00%009600%00%
Bordetella pertussis (ptxP)00%00%00%00%00%
Chlamydia pneumoniae00%00%00%00%00%
Mycoplasma pneumoniae00%00%00%00%00%

Table 48. Prevalence of Detected Analytes Stratified by Age; # = Number; EV = Expected

During the prospective study, the RP2.1 reported a single specimen with multiple organism detections (1/524 total specimens). The distinct co-detection consisted of the SARS-CoV-2 analyte (True Positive result by comparator testing) and Adenovirus.

M. Instrument Name:

FilmArray 2.0 or FilmArray Torch System

N. System Descriptions:

    1. Modes of Operation:
      Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?

Yes Car No or No or No and X a ma

Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?

Yes or No or No X

{68}------------------------------------------------

    1. Software:
      FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes _____________ or No ______________________________________________________________________________________________________________________________________________________

    1. Specimen Identification:
      Specimen identification can be entered manually or via barcode
    1. Specimen Sampling and Handling:
      The BioFire RP2.1 is intended for use with nasopharyngeal swab (NPS) collected in transport media or saline. The operator places a Hydration Vial and a Sample Injection Vial into the BioFire Pouch Loading Station. The operator first hydrates the test pouch with the Hydration Injection Vial and then adds Sample Buffer into the Sample Injection Vial using the provided Sample Buffer ampoule. Using a transfer pipette provided in the kit, the operator adds ~300ul of specimen into the Sample Injection Vial, closes the Sample Injection Vial, removes the Sample Injection Vial containing the sample mixture from the Loading station, inverts the vial at least three times to mix, and then inserts it into the Loading Station port where the proper amount of specimen is pulled into the BioFire RP2.1 pouch by vacuum. The BioFire RP2.1 pouch is then placed in the FilmArray 2.0 instrument or the available module of a FilmArray Torch system for testing.
    1. Calibration:
      Not applicable.
    1. Quality Control:
      See Quality Control Section above ["Traceability, Expected Values (controls, calibrators, or methods)"]

O. Other Supportive Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Not Applicable

P. Proposed Labeling:

The labeling supports the decision to grant the De Novo request for this device. It consists of the instructions for use, a quick reference guide and Fact Sheets for Healthcare Providers and Patients.

{69}------------------------------------------------

Identified Risks to HealthMitigation Measures
Risk of an inaccurate testresult (false positive or falsenegative result) leading toimproper patientmanagementCertain labeling information, including limitations, warnings,device descriptions, explanation of procedures, and performanceinformation identified in special controls (1), (3), (5), and (6).Use of certain specimen collection devices identified in specialcontrol (2).Certain design verification and validation, documentation of
certain analytical studies and clinical studies, risk analysisstrategies, and device descriptions identified in special control(4).Testing of characterized viral samples and labeling informationidentified in special control (7).
Misinterpretation of testresults leading tomisdiagnosis and associatedrisk of false test resultsCertain labeling information, including limitations, warnings,device descriptions, explanation of procedures, resultsinterpretation information, and performance informationidentified in special controls (1), (3), and (5).Certain design verification and validation, documentation ofcertain analytical studies and clinical studies, risk analysisstrategies, and device descriptions identified in special control(4).
Failure to correctly operatethe device leading toinaccurate test resultsCertain labeling information, including limitations, warnings,device descriptions, explanation of procedures, and performanceinformation identified in special controls (1), (3), (5), and (6).Use of certain specimen collection devices identified in specialcontrol (2).Certain design verification and validation, documentation ofcertain analytical studies and clinical studies, risk analysisstrategies, and device descriptions identified in special control(4).

Q. Identified Risks to Health and Mitigation Measures

R. Benefit/Risk Analysis:

Summary of the Assessment of Benefit

The benefit of the assay is the ability to use a well-validated device for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections, including COVID-19. The device can detect and concurrently distinguish common and the novel respiratory pathogen, SARS-CoV-2, in human clinical specimens to aid in the differential diagnosis of the respiratory infections in conjunction with other clinical, epidemiological, and laboratory data. An assay to detect SARS-CoV-2 and distinguish it from multiple common viral and bacterial respiratory pathogens in

{70}------------------------------------------------

nasopharyngeal swabs that is properly validated and consistently provides accurate test results is a critical diagnostic element in the evaluation of patients suspected of having SARS-CoV-2. The analyte added to the existing FilmArray RP2 device and tested by the FilmArray RP2.1, SARS-CoV-2 PCR, is a pathogen known to cause novel infection that can lead to outbreaks, where rapid detection would have compelling public health and safety considerations. Aid in the diagnosis of this pathogen could help guide outbreak investigations. The device may also benefit laboratory personnel by helping to rapidly identify samples that require more stringent protective measures for staff to help avoid accidental exposure to contagious pathogens. The purpose of a correct diagnostic result is to guide the appropriate patient management and therapy when used in conjunction with other clinical and laboratory information. The benefit of a correct SARS-CoV-2 result and of results for multiple common viral and bacterial respiratory pathogens have the potential to be long lasting for those tested, particularly if the illness was life threatening. Positive SARS-CoV-2 results are indicative of the presence of SARS-CoV-2 RNA; clinical correlation with patient history and other diagnostic information is necessary to determine patient infection status and guide patient management. The test results may improve infection control measures and may aid in tracking and reducing transmission of infection. There is currently no acute SARS-CoV-2 diagnostic device that has undergone full FDA premarket review (i.e., not EUA authorized) to most definitively determine clinical truth for the method comparison study, however this uncertainty is acceptable, particularly because the sponsor used a composite comparator method comprised of multiple EUA-authorized devices, which represents the most reasonable alternative to establish clinical truth in the clinical study.

Summary of the Assessment of Risk

The risks associated with the device, when used as intended, are those related to the risk of false test results, failure to correctly interpret the test results, and failure to correctly operate the device. In general, the risks associated with inaccurate results for SARS-CoV-2 and the common viral and bacterial respiratory pathogens detected by the FilmArray RP2.1 in nasopharyngeal swabs are error in diagnosis and error or delay in treatment, and delay in diagnosing the patient's true disease. False positive SARS-CoV-2 results may lead to include improper patient management, including treatment for SARSCoV-2 with antiviral medication, monoclonal antibody treatment, or convalescent plasma. These treatments have risks including toxicity and more rarely allergic reactions. False positive SARS-CoV-2 results may also lead to unnecessary isolation or quarantine and additional health monitoring, mis-allocation of resources used for surveillance and prevention, and delayed diagnosis and treatment of other infections or health conditions. False negative SARS-CoV-2 results may lead to missing and not appropriately treating or monitoring a patient who has SARS-CoV-2 infection. Missing and not treating or appropriately monitoring a patient with SARS-CoV-2 infection whose clinical picture warrants treatment could result in the known sequelae of SARS-CoV-2 infection and may result in high morbidity and mortality in these patients. False negative SARS-CoV-2 results may also lead to unnecessary additional diagnostic evaluation or treatment and delay in correct diagnosis or further spread of disease, with the potential for novel cases of infection and concomitant increase in patient morbidity and mortality.

{71}------------------------------------------------

Summary of the Assessment of Benefit-Risk

The clinical benefits outweigh the risks for the proposed assay, considering the mitigations of the risks provided in the De Novo application, product labeling, risk analysis and documentation, special controls, as well as general controls. The required De Novo application helps to ensure that errors will be uncommon and will facilitate accurate assay implementation and interpretation of results. Further, a post-market monitoring plan was provided by the sponsor for continuous monitoring of the quality, safety, and performance during the entire lifecycle of the device. The plan includes assessment of publicly available information such as sequence analysis and other postmarket activities including further benchtop evaluations where appropriate. The plan summary appeared acceptable and addressed special controls for the device. In addition, the device's performance observed in the clinical study suggests that errors will be uncommon and that the assay will provide substantial benefits to patients as an aid in the diagnosis of SARS-CoV-2 or other respiratory infections when used in conjunction with other laboratory results and clinical information.

S. Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

T. Conclusion

The De Novo request is granted and the device is classified under the following and subject to the special controls identified in the letter granting the De Novo request:

Product Code(s): QOF

Device Type: Device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test

Class: II (special controls)

Regulation: 21 CFR 866.3981

§ 866.3981 Device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test.

(a)
Identification. A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
e.g., BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
e.g., when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
e.g., regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
e.g., saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
i.e., H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.