K Number
K240252

Validate with FDA (Live)

Date Cleared
2024-07-05

(157 days)

Product Code
Regulation Number
864.5220
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The UniCel DxH 900/DxH 690T Coulter Cellular Analysis System is a quantitative, multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.

The DxH 900/DxH 690T analyzer identifies and enumerates the following parameters:

· Whole Blood (Venous or Capillary): WBC, RBC, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO%, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF
· Pre-Diluted Whole Blood (Venous or Capillary): WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV

· Body Fluids (cerebrospinal, serous or synovial): TNC and RBC

The UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.

Device Description

The UniCel DxH 900/DxH 690T System contains an automated hematology analyzer (DxH 900 or DxH 690T) designed for in vitro diagnostic use in screening patient populations by clinical laboratories. The system provides a Complete Blood Count (CBC), Leukocyte 5-Part Differential (Diff), Reticulocyte (Retic), Nucleated Red Blood Cell (NRBC) on whole blood, as well as, Total Nucleated Count (TNC), and Red Cell Count (RBC) on Body Fluids (cerebrospinal, serous and synovial).

The DxH Slidemaker Stainer II is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.

The DxH 900 System may consist of a workcell (multiple connected DxH 900 instruments with or without a DxH Slidemaker Stainer II), a stand-alone DxH 900, or a stand-alone DxH Slidemaker Stainer II. The DxH 690T System consists of a stand-alone DxH 690T instrument.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device submission (K240252) for the UniCel DxH 900/DxH 690T Coulter Cellular Analysis System and the UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria as would be the case for a novel device or a device requiring clinical efficacy trials.

Therefore, the acceptance criteria are largely implied by the performance of the predicate device and established CLSI (Clinical and Laboratory Standards Institute) guidelines for analytical performance. The "study" proving the device met acceptance criteria is a series of non-clinical performance verification tests designed to demonstrate that the new devices (DxH 900, DxH 690T, SMS II) perform "as well as or better than" the predicate devices (DxH 800, SMS) across various analytical parameters.

Here's an attempt to structure the information based on your request, understanding that the context is substantial equivalence testing, not a novel device demonstrating de novo clinical acceptance.


Device Under Evaluation for Substantial Equivalence:

  • UniCel DxH 900 Coulter Cellular Analysis System
  • UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System
  • UniCel DxH 690T Coulter Cellular Analysis System

Predicate Devices:

  • UniCel DxH 800 Coulter Cellular Analysis System (K193124)
  • UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System (K162414)

The "acceptance criteria" for the new devices are generally linked to demonstrating performance comparable to, or better than, the predicate devices, adhering to established analytical performance standards (e.g., CLSI guidelines). The "study" involves various analytical performance tests comparing the subject devices to the predicate.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by the predicate device's performance specifications and adherence to CLSI guidelines. The performance reported below demonstrates that the subject devices meet these implicit criteria (i.e., they perform comparably to the predicate). Due to the extensive list of parameters and a lack of explicit, single "acceptance limit" given for each, I will provide summary tables where possible, extrapolating from the comprehensive data provided in the text.

a. Repeatability (Within-run Imprecision)

  • Acceptance Criteria (Implied): Percent Coefficient of Variation (%CV) or Standard Deviation (SD) within specified limits, typically reflecting clinically acceptable imprecision and comparable to predicate device performance.
  • Reported Device Performance (DxH 900 - selected parameters; all passed):
ParameterUnitsLevelNTest Result MeanTest Result %CV or SDConclusion
WBCx10³ cells/µL5.000 - 10.000105.800.69% CVPass
RBCx10⁶ cells/µL4.500 - 5.500104.720.50% CVPass
Hgbg/dL14.00 - 16.001015.030.36% CVPass
Plateletx10³ cells/µL200.0 - 400.010256.801.35% CVPass
Neut %%50.0 - 60.01058.240.99 %CVPass
Retic %%0.000 - 1.500101.170.13 SDPass
BF-RBCcells/mm³10,000 - 15,0001012,6432.42 %CVPass
BF-TNCcells/mm³50-2,000105941.28 %CVPass

(Similar comprehensive data provided for DxH 690T, all passed.)

b. Reproducibility (Across-site/day/instrument Imprecision)

  • Acceptance Criteria (Implied): Reproducibility (Total CV%) to be within clinically acceptable limits, demonstrating consistent performance across different instruments, days, and runs. The test instruments met the reproducibility specifications for all parameters.
  • Reported Device Performance (DxH 900-3S workcell - examples for Level 1, all passed):
ParameterUnitN (total)Reproducibility CV%Conclusion
WBC10^3 cells/uL902.30Pass
RBC10^6 cells/uL900.66Pass
HGBg/dL900.52Pass
PLT10^3 cells/uL901.64Pass
BF TNCcells/mm^3907.28Pass
BF RBCcells/mm^3903.97Pass

c. Linearity

  • Acceptance Criteria: Deviation between measured and predicted values to be within specified acceptance limits for each parameter across the analytical measuring interval (AMI). All instances showed "Pass".
  • Reported Linearity Ranges (all passed on DxH 900-3S workcell):
ParameterUnitsLinearity Range Results
WBC10³ cells/µL0.064 - 408.5
RBC10⁶ cells/µL0.001 - 8.560
PLT10³ cells/µL3.2 - 3002
HGBg/dL0.04 - 26.070
BF-RBCcells/mm³1113.10 - 6,353,906
BF-TNCcells/mm³31.50 - 92,745

d. Carryover

  • Acceptance Criteria: Carryover to be below specified percentages/event counts (e.g., for WBC, RBC, Hgb, PLT limits <0.5% or <1.0%; for Diff, NRBC, Retic, event counts <200, <75, <600 respectively). All test instruments met these specifications.
  • Reported Device Performance (DxH 900 & DxH 690T - examples for WBC > 90):
CnDR ModeWBCRBCHgbPLTDiffNRBCRetic
DxH 690T0.11%0.03%0.26%0.07%22, 20, 3512, 5, 725, 20, 22
DxH 9000.09%0.05%0.23%0.17%11,15,377,1,310,6,3
Spec<0.5%<0.5%<1.0%<1.0%<200 Events<75 Events<600 Events

e. Performance Detection Capability Limits (LLoQ)

  • Acceptance Criteria: Limits of Quantitation (LoQ) to be ≤ specified goals. All results met the acceptance limits.
  • Reported LLoQ Results (all passed):
ParameterUnitsAcceptance limit (LoQ)LoQ resultConclusion
WBCx10³ cells/μL≤0.0500.019Pass
PLTx10³ cells/μL≤3.0000.757Pass
BF TNCcells/mm³≤20.00014.004Pass
BF RBCcells/mm³≤1000.000979.869Pass

f. Method Comparison (Accuracy vs. Predicate)

  • Acceptance Criteria (Implied): Statistical analysis (e.g., mean difference, bias, correlation coefficient) showing substantial equivalence to the predicate device within clinically acceptable limits for all parameters across the AMI. The data confirmed substantial equivalence.

  • Reported Device Performance (All Sites Combined - selected parameters, all passed comparisons to predicate):

    • Whole Blood (DxH 900 vs. DxH 800):
      • Difference (Mean): e.g., HCT: -0.297, HGB: -0.022, WBC: 0.055, PLT: -0.075, NE %: 0.014
      • Correlation (R): e.g., HCT: 0.99860, HGB: 0.99950, WBC: 0.99960, PLT: 0.99890, NE %: 0.99860
    • Body Fluid (BF-TNC, BF-RBC - DxH 900 vs. DxH 800):
      • Difference (Mean): BF RBC: 2601.802, BF TNC: 50.878
      • Correlation (R): BF RBC: 0.99991, BF TNC: 0.99989

g. Flagging Performance (Comparison to Predicate)

  • Acceptance Criteria (Implied): Negative Percent Agreement (NPA), Positive Percent Agreement (PPA), and Overall Percent Agreement (OPA) with the predicate device (DxH 800) should be high, demonstrating comparable flagging capabilities. All categories showed high agreement.
  • Reported Device Performance (DxH 900 vs. DxH 800):
Category of AbnormalitiesNPA (95% CI)PPA (95% CI)OPA (95% CI)Conclusion
Morphological Abnormalities for WB0.9555 (0.9341 to 0.9702)0.8853 (0.8362 to 0.9211)0.9347 (0.9145 to 0.9504)Pass
Distributional Study of WB0.8810 (0.8389 to 0.9131)0.9501 (0.9256 to 0.9668)0.9224 (0.9008 to 0.9397)Pass

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set (Samples/Analytes):

    • Repeatability: Varied. For whole blood, 10 aspirations per measurement. For some abnormal/low levels, contrived samples/control material used.
    • Reproducibility: 90 replicates per parameter (3 instruments x 5 days x 2 times/day x 3 shots/instrument).
    • Linearity: Minimum of seven (7) dilutions, tested in quadruplicate in random order on three (3) instruments for each measurand.
    • Carryover: Not specified for number of patient samples, but testing used three (3) high target value tubes and three (3) low target value tubes, repeated three (3) times.
    • Performance Detection Capability Limits (LoB, LLoD, LLoQ):
      • LoB: 120 whole blood cycles (DxH diluent as blank, 5 replicates x 20 total for whole blood, 120 single tube BF cycles for body fluid), using 2 lots of diluent and cell lyse reagents.
      • LLoD/LLoQ: 4 samples per parameter, 3 sets of 11 dilutions from stock, 1 set analyzed on each of 3 instruments. 5 replicates per dilution level.
    • Reference Ranges: Not explicitly stated but indicated as sufficient per CLSI EP28-A3c to verify reference intervals, usually requiring a sufficient number of healthy individuals (e.g., >120).
    • Method Comparison (Whole Blood): 735 whole blood specimens from 3 clinical sites (adult and pediatric samples).
    • Method Comparison (Body Fluid): 195 body fluid specimens (BF TNC), 130 body fluid specimens (BF RBC) from multiple sites.
    • Flagging Analysis: 735 whole blood samples (residual normal and abnormal) from three (3) clinical sites.
  • Data Provenance: Data collected from multiple clinical sites (indicated for method comparison and flagging analysis), and testing included analysis on workcell configurations (DxH 900-3S workcell) as well as stand-alone instruments. The data appears to be prospective as it involves active testing of the new devices. The countries of origin are not specified but typical of FDA submissions, implies US-based or international sites compliant with US regulations.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not mention experts and their qualifications for establishing ground truth. This is a common characteristic of analytical performance studies for IVD devices like hematology analyzers. The "ground truth" for quantitative measurements (e.g., WBC, RBC counts) is typically the measurement itself obtained from a reference method or predicate device, often with rigorous calibration and quality control. For qualitative aspects like flagging, the ground truth is established by comparing the flag output to the predicate device's flag output, assuming the predicate's performance is already validated. There is no indication of human "expert" review for individual case ground truth for these types of measurements.

4. Adjudication Method for the Test Set

Since ground truth is based on predefined analytical measurements against reference methods/predicate devices rather than human interpretation, an adjudication method like 2+1 or 3+1 (common in image-based AI studies) is not applicable and not mentioned in the document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. These types of studies are typically performed for AI-assisted diagnostic devices where the AI is intended to improve human reader performance (e.g., radiologists interpreting images). This submission is for an automated hematology analyzer, where the device performs the analysis directly without human interpretation in the loop in the same way. The evaluation is focused on the device's analytical performance (accuracy, precision, linearity, etc.) compared to its predicate device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the core of this submission is a standalone performance evaluation of the DxH 900/690T systems. The "algorithm" here refers to the instrument's internal processing of raw signals to derive reported parameters (e.g., cell counts, differentials). Its performance is assessed independently of human intervention during the measurement process, and its output is compared to a reference standard (predicate device) and expected analytical capabilities.

7. The Type of Ground Truth Used

The ground truth for the analytical performance studies (repeatability, linearity, method comparison, etc.) is based on:

  • Reference Method / Predicate Device Comparison: The performance of the UniCel DxH 900/DxH 690T is directly compared against the established performance of the legally marketed predicate UniCel DxH 800 and previously cleared UniCel DxH Slidemaker Stainer. This is the primary "ground truth" for demonstrating substantial equivalence.
  • CLSI Guidelines: Adherence to established CLSI (Clinical and Laboratory Standards Institute) protocols (e.g., EP09c, H26-A2, EP05-A3, EP06-A, EP17-A2, EP28-A3c) implicitly defines the "truth" or expected range of acceptable analytical performance for these types of in vitro diagnostic devices.
  • Control Materials and Calibrators: Certified reference materials and quality control products (e.g., COULTER 6C Cell Control, COULTER S-CAL Calibrator) are used to establish and verify instrument accuracy and precision, serving as a daily "ground truth" check.
  • Fresh Patient Samples: Used in method comparison and other studies to ensure real-world performance reflects clinical conditions.

There is no mention of "expert consensus," "pathology," or "outcomes data" being used as ground truth in the traditional sense for these analytical performance studies of a hematology analyzer.

8. The Sample Size for the Training Set

The document does not provide information on the training set size. Hematology analyzers, while highly sophisticated, are typically rule-based systems or calibrated analytical instruments, rather than machine learning/AI models that require explicit "training sets" in the modern sense (e.g., thousands of labeled images for deep learning). Their "training" involves instrument calibration using standardized calibrators. If there are adaptive algorithms or older machine learning components, their "training" data is internal to the manufacturer's development process and not typically disclosed in a 510(k) summary focused on analytical validation.

9. How the Ground Truth for the Training Set Was Established

As noted above, the concept of a "training set" and its "ground truth" in the context of a modern AI/ML device is not directly applicable to this traditional analytical instrument unless it has specific, undisclosed internal adaptive algorithms. The "ground truth" for the instrument's operational accuracy and precision is primarily established through its calibration process using certified calibrator materials and verified against quality control materials and comparisons to reference methods or predicate devices as part of its manufacturing and analytical validation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract image of a human figure, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" in a larger font size than the rest of the text.

July 5, 2024

Beckman Coulter, Inc Marie Steigerwalt Staff Regulatory Affairs Specialist 11800 S.W. 147th Avenue Miami, Florida 33196

Re: K240252

Trade/Device Name: UniCel DxH 900 Coulter Cellular Analysis System; UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System; UniCel DxH 690T Coulter Cellular Analysis System Regulation Number: 21 CFR 864.5220 Regulation Name: Automated Differential Cell Counter Regulatory Class: Class II Product Code: GKZ Dated: June 3, 2024 Received: June 4, 2024

Dear Marie Steigerwalt:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by YAN CAI -S
YAN CAI -S Date: 2024.07.05 13:46:27
-04'00'

for

Min Wu Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240252

Device Name UniCel DxH 900 Coulter Cellular Analysis System ;

UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System; UniCel DxH 690T Coulter Cellular Analysis System

Indications for Use (Describe)

The UniCel DxH 900/DxH 690T Coulter Cellular Analysis System is a quantitative, multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.

The DxH 900/DxH 690T analyzer identifies and enumerates the following parameters:

· Whole Blood (Venous or Capillary): WBC, RBC, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO%, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF · Pre-Diluted Whole Blood (Venous or Capillary): WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV

· Body Fluids (cerebrospinal, serous or synovial): TNC and RBC

The UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.#### This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K240252.

807.92 (a)(1): Contact Details

Name: Beckman Coulter, Inc Address: 11800 S.W. 147th Avenue Miami FL 33196 United States Phone: 305-380-3800 Email: msteigerwalt@beckman.com Contact: Marie Steigerwalt

807.92 (a)(2): Device Name

Trade Name:

UniCel DxH 900 Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System UniCel DxH 690T Coulter Cellular Analysis System Common Name: Automated differential cell counter Classification Name: Counter, Differential Cell Regulation Number: 21 CFR 864.5220 Product Code(s): GKZ

807.92 (a)(3): Identification of the Legally Marketed Predicate Devices

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K193124UniCel DxH 800 Coulter Cellular Analysis System with EarlySepsis Indicator ApplicationGKZ
K162414UniCel DxH Slidemaker Stainer Coulter Cellular Analysis SystemGKZ

807.92 (a)(4): Device Description Summary

The UniCel DxH 900/DxH 690T System contains an automated hematology analyzer (DxH 900 or DxH 690T) designed for in vitro diagnostic use in screening patient populations by clinical laboratories. The system provides a Complete Blood Count (CBC), Leukocyte 5-Part Differential (Diff), Reticulocyte (Retic), Nucleated Red Blood Cell (NRBC) on whole blood, as well as, Total Nucleated Count (TNC), and Red Cell Count (RBC) on Body Fluids (cerebrospinal, serous and synovial).

The DxH Slidemaker Stainer II is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.

The DxH 900 System may consist of a workcell (multiple connected DxH 900 instruments with or without a DxH Slidemaker Stainer II), a stand-alone DxH 900, or a stand-alone DxH Slidemaker Stainer II. The DxH 690T System consists of a stand-alone DxH 690T instrument. The available DxH 900 workcell systems are listed in the table below.

{4}------------------------------------------------

DxH 900, DxH 690T and DxH Slidemaker Stainer II Stand-alone Instruments, and DxH 900 and DxH Slidemaker Stainer II Workcell Configurations

ConfigurationTopologiesSystem ManagerReview Stations
DxH 900DxH 900Standard ComputerN/A
DxH 690TDxH 690TStandard ComputerN/A
DxHSlidemakerStainer IIDxH Slidemaker Stainer IIStandard ComputerN/A
DxH 900 SDxH Slidemaker Stainer IIDxH 900Standard ComputerOptional Stand-aloneReview Station withStandard Computer
DxH 900-2DxH 900DxH 900Power ComputerPower Computer
DxH 900-2 SDxH Slidemaker Stainer IIDxH 900DxH 900Power ComputerOne Review Stationrequired with aPower Computer.Optional ReviewStation withStandard Computer
DxH 900-3DxH 900DxH 900DxH 900Power ComputerTwo Review Stationsrequired, one withPower Computer andone with StandardComputer
DxH 900-3 SDxH Slidemaker Stainer IIDxH 900DxH 900DxH 900Power ComputerTwo Review Stationsrequired, one withPower Computer andone with StandardComputer. OptionalStand-aloneWorkstation withStandard Computer

{5}------------------------------------------------

Principle of Operation:

The Coulter Principle is used to count and size cells by detecting and measuring changes in electrical resistance when a particle (such as a cell), in a conductive liquid, passes through a small aperture as shown in Figure 1.

Figure 1: Coulter Principle

Image /page/5/Figure/3 description: The image shows a diagram of a blood cell counter. The diagram includes labels for different parts of the counter, including the aperture current (1), vacuum (2), internal electrode (3), blood cell suspension (4), detail of aperture (5), aperture tube (6), aperture (7), sample beaker (8), and external electrode (9). The diagram provides a visual representation of the components and their arrangement within the blood cell counter.

Each cell, suspended in a conductive liquid (diluent), acts as an insulator. As each cell goes through the aperture, it momentarily increases the resistance of the electrical path between two submerged electrodes on either side of the aperture. This causes a measurable electronic pulse. For counting, the vacuum used to pull the diluted suspension of cells through the aperture must be at a regulated volume. While the number of pulses indicates particle count, the size of the electrical pulse is proportional to the cell volume.

The lytic reagent used for the WBC prepares the blood so the system can count leukocytes and measure the amount of hemoglobin. The lytic reagent rapidly and simultaneously destroys the erythrocytes and converts a substantial proportion of the hemoglobin to a stable pigment while it leaves leukocyte nuclei intact. The absorbance of the pigment is directly proportional to the hemoglobin concentration of the sample. Hemoglobin is measured photometrically at 525 nm using the sample from the WBC analysis. Clean diluent is introduced into the cuvette during each operating cycle and is used as a blank in the calculation of the HGB.

The COULTER VCS technology is used to determine the white cell differential, nucleated red blood cell and reticulocyte parameters along with associated flags, messages, histograms and

{6}------------------------------------------------

data plots. As the particles pass through the sensing zone, a diode laser illuminates the particles. The flow cell measures volume, conductivity, multiple angles of light scatter, and axial light loss.

  • . For the WBC differential, sample preparation occurs at the Diff mix chamber where sample and reagents are added in the following order: Diff Lyse, blood, additional Diff Lyse followed by an air mix. Next, Diff preservative is added, followed by a second air mix, and an incubation period. The prepared sample is transferred to the module where cells are counted in an isometric sample stream. The algorithm analysis separates the WBC into five major populations.
  • . For NRBC, sample preparation occurs at the NRBC Diff mix chamber where sample and reagents are added in the following order: Diluent, blood, additional Diluent followed by an air mix. Next, DxH Cell Lyse is added, followed by a second air mix, and an incubation period. The prepared sample is transferred to the module where cells are counted in an isometric sample stream. The algorithm analysis separates NRBC from WBC.
  • Reticulocytes are immature, non-nucleated erythrocytes retaining a small network of . basophilic organelles, consisting of RNA and protoporphyrin. The enumeration of reticulocytes provides a simple, effective means to determine red cell production and regeneration. Reticulocyte immaturity is related to cell volume and light scatter. Since more immature reticulocytes are larger, contain more RNA and cause increased light scatter, the cell volume and light scatter will increase with immaturity of the cell.

The DxH Slidemaker Stainer II allows for adaptation of the smear appearance and stain methodology according to user preferences. Blood smears produced by the Slidemaker portion of the DxH Slidemaker Stainer II are moved to baskets for transfer to the Stainer portion by a robot. Microscopic examination of the stained blood smears can be used to help determine the hematologic status of a patient.

The aspiration probe aspirates the mixed sample, which is transported into the dispense probe. The probe moves to the drop placement position where blood is placed on the slide. A second slide is picked up and used to spread the drop on the first slide using the wedge technique. The prepared slide is transferred into the print shuttle where Information is thermally printed on the painted portion of the slide, and then into the basket elevator for drying. The dispense probe is cleaned in the dispense wash cup after drop placement on the slide.

For staining, the slide basket is dipped into each bath (1 to 5) for a staining time according to the active protocol. Each bath is configured to receive reagent from a predetermined supply source configured by the user. Slides are dried after staining.

{7}------------------------------------------------

807.92 (a)(5): Intended Use

The UniCel DxH 900/DxH 690T Coulter Cellular Analysis System is a quantitative, multiparameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.

The DxH 900/DxH 690T analyzer identifies and enumerates the following parameters:

· Whole Blood (Venous or Capillary); WBC. RBC. HGB. HCT. MCV. MCH. MCHC. RDW. RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF

· Pre-Diluted Whole Blood (Venous or Capillary): WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV

· Body Fluids (cerebrospinal, serous or synovial): TNC and RBC

The UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.

807.92 (a)(5): Indications for Use Comparison

The UniCel DxH 900 Coulter Cellular Analysis System. UniCel DxH 690T Coulter Cellular Analysis System, and UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System are substantially equivalent to the UniCel DxH 800 Coulter Cellular Analysis System and UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System cleared under the 510(k) numbers K193124 and K162414. There are no changes to the intended use or claims.

807.92 (a)(6): Technological Comparison

The UniCel DxH 900 Coulter Cellular Analysis System. UniCel DxH 690T Coulter Cellular Analysis System, and UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System have the same technological characteristics (principal of operation, energy source, reagents, calibrators, controls, and packaging) as the predicate devices.

CharacteristicPredicate DxH 800K193124Subject DxH 900/690T System
Intended UseAnalyzer:The UniCel DxH 800 Coulter Cellular Analysis Systemis a quantitative, multi-parameter, automatedhematology analyzer for in vitro diagnostic use inscreening patient populations found in clinicallaboratories.The DxH 800 hematology analyzer identifies andenumerates the following parameters:• Whole Blood (Venous or Capillary): WBC, RBC,HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%,MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#,RET%, RET#, MRV, IRF• Pre-Diluted Whole Blood (Venous or Capillary):WBC, RBC, HGB, HCT, MCV, MCH, MCHC,RDW, RDW-SD, PLT, MPVSame
CharacteristicPredicate DxH 800K193124Subject DxH 900/690T System
• Body Fluids (cerebrospinal, serous or synovial): TNCand RBC
Sampling MechanismSingle aspiration probe used for all sampling.Single tube presentation - open and closed vialsampling - specimen manually mixed.Automated presentation – closed vial sampling fromfive-position cassette accepting a variety of definedspecimen tubes. Cassette containing specimens mixedprior to starting sampling and between specimens.Maximum initial load capacity 20 cassettes - Systemwill continuously process racks as added.Same
Principles ofMeasurementWBC, RBC, MCV, PLT: Aperture impedance (Coulter®Principle)Hemoglobin: SpectrophotometricWBC Differential, Reticulocytes, NRBC:VCSn Technology using:• Aperture impedance (DC)• Conductivity (RF)• Laser Light Scatter (Multiple angles)• Laser Light AbsorbanceSame
ConsumablesAnalysis Reagents• COULTER DxH Diluent• COULTER DxH Diff Pack• COULTER DxH Cell Lyse• COULTER DxH Retic Pack• COULTER DxH CleanerQuality Control & Calibrators• COULTER 6C Cell Control• COULTER 6C Plus Cell Control• COULTER Latron CP-X Control• COULTER RETIC-X Cell Control• COULTER LIN-X Control• COULTER Body Fluids Control• COULTER S-CAL Calibrator kitSame, plus the option of the followingECO reagents:• DxH ECO Diluent (RTU)• DxH Concentrated ECODiluent (18x)• DxH ECO Cell Lyse
Pre-Analytic Features
SystemconfigurationPC based workstation running Microsoft Windows 7application specific software Handheld Barcode ScannerPrinterSame, but Windows 10
SamplingMechanismSingle tube presentation - open and closed vialsampling.Automated presentation - closed vial sampling from 5position cassette; Maximum initial load capacity 20racksSame
Mechanisms forprocessingMechanisms to achieve process of:Automated cassette transportation and specimen mixing(by rocking), sample aspiration, sample preparation,sample and reagent presentation to analytical modules,sample analysis, raw data collection, algorithmicprocessing and data reporting.Same
CharacteristicPredicate DxH 800K193124Subject DxH 900/690T System
allowing multi-directional moves and capability toreturn cassette to sampling position for repeat / reflextesting.
Sample IdentificationSample aspiration module (SAM) mounted barcodereader for automated barcode reading of cassetteand sample tube identifiers. Manual barcode scanning of sample tube identifier(handheld scanner) Manual keyboard entry of sample identifierSame
Sample Processing
Aspiration PathwaySingle sampling probe and common aspiration pathwayused for all sample presentation modes.Same
Sample AspirationVolumeAutomatic, cap-piercing: 165 µLSingle tube - open-vial and cap pierce:165 µLPre-dilute 165 µL - fixed ratio of 1 in 5 dilution of bloodwith diluentSame
ThroughputFor automatic mode:CBC at 100 specimens/hr CBC and Differential at 100 specimens/hr CBC and Differential with NRBC at 90specimens/hr Retic at 45 specimens/hrSame
Data ReportingWorkstation display graphics, hardcopy printing andtransmission to Laboratory Information System (LIS)Same
System Control and Software
System SoftwareSystem software (embedded and workstation) designedspecific to support all features of the DxH 800. Thesoftware system consists of a Data Manager component,a Universal System Manager component (includingalgorithms), the User Interface, all of which are residentin the system software in the workstation.In addition, an Embedded Application is resident in theanalyzer. The Embedded application uploads from theworkstation on system power-up. Extensive real timemonitoring and reporting of system status including:Component and module activities, System Voltages and Currents System Pressure and Vacuum System Temperatures Motor activity Mechanism Sensor status Reagent Pump Operation Raw data collection Single sampling probe and common aspiration pathwayused for all sample presentation modes.Same system software withenhancements to the instrument lookand feel, workflow, usability, qualitycontrol and reliability.
CharacteristicPredicate DxH SMSK162414Subject SMS II
Intended UseThe DxH Slidemaker Stainer is a fully automated slidepreparation and staining device that aspirates a whole-blood sample, smears a blood film on a cleanmicroscope slide, and delivers a variety of fixatives,stains, buffers, and rinse solutions to that blood smear.Same
CharacteristicPredicate DxH 800K193124Subject DxH 900/690T System
Specimen CollectionWhole venous blood in EDTASame
Blood FilmPreparationAutomatically prepared by DxH SMSSame
Blood FilmRequirementsSection 6.3.1 of CLSI H20-A2Same
ConsumablesStains and Buffers: Beckman Coulter TruColorReagents for use on the DxH Slidemaker Stainer• TruColor Wright Stain and TruColor Wright Buffer• TruColor Wright-Giemsa Stain and TruColor Wright-Giemsa BufferCoulter TruColor Giemsa and May Grunwald Stains arenot validated by Beckman Coulter for use throughcurrent default protocols on the system.Fixative: Methanol is used as a fixative for whole-bloodsmears in preparation for staining. Anhydrous methanol(chromatography grade, 99.8% or higher quality) isrecommended.Distilled Water: Distilled water is used to rinse thestained smears before drying. CLSI Type CLRW wateris recommended.Same

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

807.92 (b): Non-Clinical and/or Clinical Test Summary

  • a) Specimen Sampling Positions Comparability was tested on the DxH 800 (predicate) per EP09c, CLSI H26-A2, and CLSI H56 to establish equivalency for all sampling modes (Single tube open vial, Single tube closed vial, and Cassette presentations). Due to the similarities in the instrument hardware, data presented in K120771 continues to support the use of all specimen sampling positions as appropriate for the intended use of the device.
Test PanelReported ParametersSpecimen TypeSampling Method
Complete BloodCount (CBC)• WBC, RBC, HGB, HCT, MCV,MCH, MCHC, RDW, RDW-SD,PLT, MPVwhole blood• Automated cassette closed vial• Manual single tube closed vial• Manual single tube open vial
CBC andDifferential incNRBC(CD)• WBC, RBC, HGB, HCT, MCV,MCH, MCHC, RDW, RDW-SD,PLT, MPV• NE%, NE#, LY%, ly#, MO%, MO#,EO%, EO#, BA%, BA#, NRBC%,and NRBC#whole blood• Automated cassette closed vial• Manual single tube closed vial• Manual single tube open vial
CBC, Differentialinc NRBC andRetic(CDR)• WBC, RBC, HGB, HCT, MCV,MCH, MCHC, RDW, RDW-SD,PLT, MPV• NE%, NE#, LY%, ly#, MO%, MO#, MO#,EO%, EO#, BA%, BA#, NRBC%,and NRBC#• RET%, RET#, MRVM IRFwhole blood• Automated cassette closed vial• Manual single tube closed vial• Manual single tube open vial
CBC and Retic(CR)• WBC, RBC, HGB, HCT, MCV,MCH, MCHC, RDW, RDW-SD,PLT, MPV• RET%, RET#, MRVM IRFwhole blood• Automated cassette closed vial• Manual single tube closed vial• Manual single tube open vial
Retic(R)• RET%, RET#, MRVM IRFwhole blood• Automated cassette closed vial• Manual single tube closed vial• Manual single tube open vial

{11}------------------------------------------------

Test PanelReported ParametersSpecimen TypeSampling Method
Pre-dilute(PreDilx5)• WBC, RBC, HGB, HCT, MCV,MCH, MCHC, RDW, RDW-SD,PLT, MPV1 in 5 pre-dilutedwhole blood• Manual single tube open vial
Body Fluid(BF)• TNC, RBCbody fluid(cerebrospinal, serousand synovial)• Manual single tube open vial
  • b) Whole blood and body fluid repeatability precision was tested on the DxH 900-3S workcell (instruments BC03127, BC03129, and BC03130) using the instrument's repeatability mode using cassette presentation for whole blood and single-tube presentation sampling for body fluids. Testing was performed according to CLSI H26-A2 and CLSI EP05-A3 to demonstrate the precision of the DxH 900 using whole blood and body fluid specimens. The whole blood and body fluid precision data provided supports the precision claims in the instructions for use. Connection of instruments into a workcell configuration does not significantly impact the precision.
    Within run Repeatability of the parameters was performed using ten (10) aspirations of normal whole blood samples collected in K2EDTA. Specimens were selected for testing based on parameter value within specific ranges per parameter (measurand) to cover across the analytical measuring interval (AMI). Within run Imprecision (repeatability) was tested by performing replicate analysis of the specimens selected using the Repeatability function. Targeted specimens were used for more than one measurand. The coverage of the other measurand's measuring ranges was dependent on the spread of the samples collected to cover the measuring interval of the measurands. If normal and clinical specimens in the intended use populations were not found for estimation of precision profiles, contrived samples and control material were used.

Abnormal patient samples with low NRBC (1.00 - 2.00, >2.00 - 15.00), high NRBC (>15.00), and high Reticulocytes (>4.000 - 15.0) were not available during the test period. Instead, control material was used to assess Repeatability for these parameter values. Also, low levels for WBC 0.500 - 2.000 x103 cells/uL, Platelet 10.0 - 15.0 x103 cells/μL, Body fluid RBC 10,000 - 15,000 cells/mm3 and TNC 50-2,000 cells/mm3 were performed using altered whole blood samples due to no donor availability during the testing period as well.

ParameterUnitsLevelNTest Result MeanTest Result %CV orSDAcceptance
WBCx103 cells/µL5.000 - 10.000105.800.69% CVPass
WBCx103 cells/µL0.500 to 2.000100.592.02% CVPass
RBCx106 cells/µL4.500 - 5.500104.720.50% CVPass
Hgbg/dL14.00 - 16.001015.030.36% CVPass
MCVfL80.00 - 90.001085.390.22% CVPass
RDW %%12.00 - 14.001013.711.31% CVPass
RDW-SDfL33.00 - 48.001047.031.82% CVPass
Plateletx103 cells/µL200.0 - 400.010256.801.35% CVPass
Plateletx103 cells/µL10.0 to 15.01012.902.60% CVPass
MPVfL8.0 - 10.0108.131.15% CVPass
Neut %%50.0 - 60.01058.240.99 %CVPass
Lymph %%25.0- 35.01026.342.44% CVPass
Mono %%5.0 - 10.0106.227.66 % CVPass

Repeatability Results- Whole Blood CBC, DIFF, Retic

{12}------------------------------------------------

ParameterUnitsLevelNTest Result MeanTest Result %CV or SDAcceptance
Eos %%2.0 - 5.0102.215.06% CVPass
Baso %%0.5 - 1.5101.220.13 SDPass
NRBC%1.00 - 2.0010*1.460.20 SDPass
NRBC%>2.00 - 15.0010*9.214.69% CVPass
NRBC%> 15.0010*19.125.12% CVPass
Retic %%0.000 - 1.500101.170.13 SDPass
Retic %%> 1.500 - 4.000101.520.09 SDPass
Retic %%> 4.000 - 15.00010*10.071.44% CVPass
IRFN/A≥ 0.20100.477.20%Pass
MRVfL100.0 - 120.010118.401.19% CVPass

*Control Material was used

Repeatability Results for Body Fluids

ParameterUnitsLevelNTest ResultMeanTest Result %CVor SDAcceptance
BF-RBCcells/mm³10,000 - 15,0001012,6432.42Pass
BF-TNCcells/mm³50-2,000105941.28Pass
  • c) Reproducibility testing was performed on the DxH 900-3S workcell using a single lot of all three (3) levels of 6C Plus, Retic-X, and Body Fluid cell controls. The table below shows the control material used for each material.
Control Material UsedParameter
COULTER 6C PLUS Cell Control(3 levels)WBC
RBC
HGB
MCV
RDW
RDW-SD
PLT
MPV
NE
LY
MO
EO
BA
NRBC
COULTER RETIC-X Cell Control(3 levels)RET
COULTER Body Fluids Control(3 levels)BF TNC
BF RBC

The controls were analyzed on one (1) DxH900-3S workcell, five (5) days, two (2) different times on the same day and three (3) shots per instrument, two (2) hours apart, per CLSI EP05-A3. The test instruments met the reproducibility specifications for all parameters.

{13}------------------------------------------------

NMeanRepeatabilityBetween RunsBetween InstrumentBetween DaysReproducibility
LevelParameterUnitSDCV%SDCV%SDCV%SDCV%SDCV%
WBC10^3 cells/uL903.420.071.900.000.000.041.260.010.270.082.30
RBC10^6 cells/uL901.740.010.590.000.110.000.260.000.000.010.66
HGBg/dL904.620.020.430.000.000.010.260.010.130.020.52
MCVFl9079.480.140.180.190.240.230.290.190.240.380.48
RDW%9017.090.150.890.140.820.080.440.000.030.221.28
RDW-SD9049.500.491.000.190.380.100.210.100.200.551.11
PLT10^3 cells/uL9072.391.111.530.000.000.280.390.330.451.191.64
MPVfL909.130.060.660.000.000.030.300.030.340.070.80
LevelNE%9043.130.691.610.000.000.000.000.000.000.691.61
lLY%9046.240.721.560.000.000.120.250.000.000.731.58
MO%906.670.40ર તેર0.000.000.152.290.000.000.436.38
EO%903.940.348.580.143.670.000.000.000.000.379.33
BA%900.020.0274.920.000.000.000.000.0126.290.0279.39
NRBC%900.290.1034.200.000.000.0515.620.000.000.1137.60
RET%900.870.1011.180.022.710.1517.410.000.000.1820.86
BF TNCcells/mm^390120.746.515.393.633.013.412.833.182.648.797.28
BF RBCcells/mm^39012204.83398.753.270.000.00276.212.260.000.00485.073.97
WBC10^3 cells/uL9020.520.170.830.110.530.241.160.060.280.32ાં રેડ
RBC10^6 cells/uL903.960.030.740.020.450.020.490.010.220.041.02
HGBg/dL9011.800.050.450.020.180.070.610.010.110.090.79
MCVFl9087.380.190.220.050.050.250.290.150.170.350.40
RDW%9016.240.161.010.060.370.010.090.080.510.191.19
RDW-SDfL9052.730.671.270.140.260.080.140.170.320.711.35
PLT10^3 cells/uL90424.587.981.880.000.000.000.000.000.007.981.88
LevelMPVfL909.440.060.580.000.000.050.540.020.180.080.82
2NE%9065.120.681.050.000.000.000.000.120.190.691.07
LY%9015.710.503.210.070.460.130.800.090.600.533.39
MO%9014.160.523.650.261.840.000.000.000.000.584.09
EO%904.960.418.330.000.000.000.000.000.000.418.33
BA%900.060.0357.530.000.000.000.000.000.000.0357.53
NRBC%909.090.404.360.070.780.000.000.080.900.414.53
RET0%902.480.124.730.000.000.166.500.041.780.208.23
BF TNCcells/mm^3901526.9220.381.334.950.3211.010.7212.200.8026.641.74
BF RBCcells/mm^3901681333.928753.350.524306.830.2616235.200.97963.340.0618965.221.13
WBC10^3 cells/uL908.580.101.200.000.000.040.470.030.400.121.35
RBC10^6 cells/uL905.280.030.620.020.330.040.770.010.240.061.07
HGBg/dL9015.610.090.590.000.000.140.890.040.290.171.10
MCVF19087.590.220.260.000.000.240.280.140.160.360.41
RDW%9015.470.171.090.000.000.000.000.000.000.171.09
RDW-SDfL9050.860.601.190.000.000.000.000.000.000.601.19
PLT10^3 cells/uL90222.273.161.420.000.000.000.000.000.003.161.42
LevelMPVfL909.610.060.670.020.260.040.460.000.000.080.85
3NE%%9057.050.811.420.000.000.000.000.000.000.811.42
LY%909027.730.592.130.000.000.000.000.000.000.000.000.000.592.13
MO%8.360.404.780.000.000.000.000.404.78
EO%90906.820.470.026.8654.510.000.010.0013.070.000.000.009.390.000.000.000.470.026.86
BA%900.042.660.002.520.090.450.7256.83
NRBCRET%9019.4510.500.520.181.720.000.000.000.490.110.070.690.233.692.14
BF TNCcells/mm^39065778.99514.830.78265.680.40763.001.091.160.000.00958.021.46
BF RBCcells/mm^390 5639354.09 50090.170.890.000.0032100.780.571922.280.0359524.621.06

Analysis for Instrument Combined

{14}------------------------------------------------

ParameterUnitNMeanSDCV%
BF RBCcells/mm^311737248511.1
cells/mm^3107470611661.5
cells/mm^39126244073.2
BF-TNCcells/mm^31030993.1
cells/mm^3111270211.7
cells/mm^312696203.0
cells/mm^3103070652.1
cells/mm^310248792070.83
cells/mm^391281262.0
cells/mm^310214104.9

Additional Body Fluid Precision Runs

  • d) Linearity: Whole Blood and Body Fluid Linearity was tested on the DxH 900-3S workcell per CLSI EP06-A, 2nd Ed to assess system linearity of whole blood count parameters (WBC, RBC, PLT and HGB) and Body Fluids parameters (BF-RBC and BF-TNC) across the analytical measuring interval (AMI). Fresh whole blood was obtained and concentrated to achieve a high starting value near the upper limit of the AMI for each measurand. If values near the upper limit were not achieved using whole blood, surrogate (analog) cells were used instead. Dilutions were prepared of each measurand to cover the AMI and tested in quadruplicate in random order. Each dilution series contained a minimum of seven (7) dilutions and testing was performed on three (3) instruments. For count measurands, given the wide measuring interval, additional dilutions at the low end of the range were required.
DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.0120.0490.0700.081-0.01092-13.440.100Pass
0.0240.0980.1280.131-0.00217-1.660.100Pass
0.0490.2000.2370.2330.003891.670.100Pass
0.0980.4000.4600.4340.026136.020.100Pass
0.1950.7970.8640.8320.031363.770.100Pass
0.3911.5971.7381.6370.100976.170.174Pass
0.7813.1913.3713.2370.133374.120.200Pass
1.5636.3856.4666.4470.019650.300.200Pass
3.12512.76713.19112.8570.333752.600.396Pass
6.2525.53426.33825.6820.655502.550.790Pass
12.551.06851.62651.3330.292680.571.549Pass
25102.135103.036102.6340.401700.395.152Pass
50204.271204.357205.236-0.87860-0.4310.218Pass
100408.541408.541410.440-1.89885-0.4620.427Pass

Deviation from Linearity for WBC 10^3/μL on Instrument BC03127

Deviation from Linearity for WBC 10^3/µL on Instrument BC03129

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.0120.0510.0650.075-0.00995-13.330.100Pass
0.0240.1010.1250.126-0.00057-0.450.100Pass
0.0490.2070.2400.2320.008283.570.100Pass
0.0980.4140.4520.4400.012552.850.100Pass
0.1950.8240.8820.8520.030683.600.100Pass

{15}------------------------------------------------

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.3911.6531.7911.6840.107446.380.179Pass
0.7813.3023.4543.3400.114123.420.200Pass
1.5636.6076.7586.6600.097981.470.203Pass
3.12513.21113.53113.2930.238211.790.406Pass
6.2526.42227.18926.5620.627412.360.816Pass
12.552.84353.30853.1000.208140.391.599Pass
25105.686105.651106.177-0.52538-0.495.283Pass
50211.372212.634212.3300.304560.1410.632Pass
100422.744*422.744424.636-1.89156-0.4521.137Pass

Deviation from Linearity for WBC 10^3/μL on Instrument BC03130

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.0120.0490.0710.083-0.01241-14.940.100Pass
0.0240.0980.1290.132-0.00306-2.320.100Pass
0.0490.2000.2330.233-0.00007-0.030.100Pass
0.0980.4010.4650.4320.033307.710.100Pass
0.1950.7970.8850.8250.059747.240.100Pass
0.3911.5981.7311.6190.111536.890.173Pass
0.7813.1923.3833.2000.182215.690.200Pass
1.5636.3886.4876.3710.116121.820.200Pass
3.12512.77213.01012.7030.307072.420.390Pass
6.2525.54425.70525.3710.334561.320.771Pass
12.551.08850.63250.707-0.07478-0.151.519Pass
25102.176101.071101.380-0.30880-0.305.054Pass
50204.351198.604202.725-4.12083-2.039.930Pass
100408.702408.702405.4163.286430.8120.435Pass

Deviation from Linearity for RBC 10^6/µL on Instrument BC03127

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
00.0000.0130.0130.000000.000.050Pass
100.8580.8560.867-0.01116-1.290.050Pass
201.7161.7421.7210.021001.220.050Pass
302.5742.5672.574-0.00716-0.280.051Pass
403.4323.4333.4280.004340.130.069Pass
રે04.2904.3444.2820.062171.450.087Pass
605.1485.1985.1360.062341.210.104Pass
706.0065.9185.990-0.07216-1.200.118Pass
806.8656.8726.8440.028010.410.137Pass
907.7237.5837.697-0.11449-1.490.152Pass
1008.5818.5818.5510.029350.340.172Pass

{16}------------------------------------------------

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
00.0000.0010.001-0.00000-0.000.050Pass
100.8560.8570.8440.013361.580.050Pass
201.7121.6561.686-0.02994-1.780.050Pass
302.5682.5192.529-0.00992-0.390.050Pass
403.4243.4163.3720.044451.320.068Pass
504.2804.2014.214-0.01319-0.310.084Pass
605.1365.0625.0570.005170.100.101Pass
705.9925.8955.899-0.00480-0.080.118Pass
806.8486.6706.742-0.07244-1.070.133Pass
907.7047.5047.585-0.08075-1.060.150Pass
1008.5608.5608.4270.132951.580.171Pass

Deviation from Linearity for RBC 10^6/µL on Instrument BC03129

Deviation from Linearity for RBC 10^6/µL on Instrument BC03130

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
00.0000.0010.001-0.00000-0.000.050Pass
100.8590.8580.8430.015051.780.050Pass
201.7191.6431.686-0.04258-2.530.050Pass
302.5782.5102.528-0.01820-0.720.050Pass
403.4373.4193.3700.048521.440.068Pass
504.2974.2054.212-0.00777-0.180.084Pass
605.1565.0705.0550.014940.300.101Pass
706.0155.9505.8970.052990.900.119Pass
806.8756.6326.739-0.10730-1.590.133Pass
907.7347.5037.582-0.07892-1.040.150Pass
1008.5938.5938.4240.169462.010.172Pass

Deviation from Linearity for HGB g/dL on Instrument BC03127

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
00.0000.1200.285-0.16515-57.920.200Pass
102.6072.8232.860-0.03661-1.280.200Pass
205.2145.2875.435-0.14806-2.720.200Pass
307.8217.9138.010-0.09618-1.200.237Pass
4010.42810.80310.5840.219032.070.324Pass
5013.03513.48313.1590.324242.460.405Pass
6015.64215.98015.7340.246121.560.479Pass
7018.24918.56018.3090.251331.370.557Pass
8020.85620.68020.883-0.20345-0.970.620Pass
9023.46323.03023.458-0.42824-1.830.691Pass
10026.07026.07026.0330.036970.140.782Pass

{17}------------------------------------------------

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
00.0000.0570.122-0.06576-53.710.200Pass
102.6492.6972.749-0.05218-1.900.200Pass
205.2995.3605.375-0.01527-0.280.200Pass
307.9488.0478.0020.044970.560.241Pass
4010.59710.73010.6280.101880.960.322Pass
5013.24713.37713.2550.122120.920.401Pass
6015.89615.93715.8810.055700.350.478Pass
7018.54518.45318.507-0.05406-0.290.554Pass
8021.19521.03021.134-0.10382-0.490.631Pass
9023.84423.62023.760-0.14024-0.590.709Pass
10026.49326.49326.3870.106670.400.795Pass

Deviation from Linearity for HGB g/dL on Instrument BC03129

Deviation from Linearity for HGB g/dL on Instrument BC03130

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
00.0000.1400.242-0.10242-42.250.200Pass
102.6662.8972.8720.024610.860.200Pass
205.3315.4075.502-0.09503-1.730.200Pass
307.9978.0908.131-0.04133-0.510.243Pass
4010.66311.01310.7610.252362.350.330Pass
5013.32813.43713.3910.046060.340.403Pass
6015.99416.06016.0200.039760.250.482Pass
7018.66018.83318.6500.183450.980.565Pass
8021.32521.10321.280-0.17618-0.830.633Pass
9023.99123.66023.909-0.24915-1.040.710Pass
10026.65726.65726.5390.117880.440.800Pass

Deviation from Linearity for PLT 10^3/µL on Instrument BC03127

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.0982.9913.7004.633-0.93316-20.145.000Pass
0.1955.9516.7677.635-0.86840-11.375.000Pass
0.39111.93314.10013.7010.399222.915.000Pass
0.78123.83626.63325.7700.863033.355.000Pass
1.56347.70352.73349.9712.762095.535.000Pass
3.12595.375102.53398.3114.222094.295.127Pass
6.25190.750202.833195.0227.811164.0110.142Pass
12.5381.500402.867388.44414.422623.7120.143Pass
25763.000795.467775.28820.178872.6039.773Pass
501526.0001559.7001548.97510.724720.6977.985Pass
1003052.0003052.0003096.350-44.35026-1.43152.600Pass

{18}------------------------------------------------

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.0982.9423.4334.366-0.93287-21.375.000Pass
0.1955.8556.5337.332-0.79861-10.895.000Pass
0.39111.73913.16713.325-0.15790-1.195.000Pass
0.78123.44826.56725.2491.318005.225.000Pass
1.56346.92652.20049.1583.041986.195.000Pass
3.12593.823100.43396.9163.517763.635.022Pass
6.25187.646201.733192.4619.272074.8210.087Pass
12.5375.292403.633383.55320.080695.2420.182Pass
25750.583779.800765.73514.064611.8438.990Pass
501501.1671555.2671530.10125.165781.6477.763Pass
1003002.3333002.3333058.832-56.49855-1.85150.117Pass

Deviation from Linearity for PLT 10^3/µL on Instrument BC03129

Deviation from Linearity for PLT 10^3/μL on Instrument BC03130

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.0983.0293.3674.120-0.75307-18.285.000Pass
0.1956.0266.8007.145-0.34457-4.825.000Pass
0.39112.08412.90013.257-0.35660-2.695.000Pass
0.78124.13726.10025.4180.681702.685.000Pass
1.56348.30452.36749.8042.562595.155.000Pass
3.12596.577103.33398.5134.820094.895.167Pass
6.25193.154202.933195.9636.970563.5610.147Pass
12.5386.308404.667390.86213.804833.5320.233Pass
25772.617784.367780.6603.706710.4739.218Pass
501545.2331569.0331560.2568.777140.5678.452Pass
1003090.4673090.4673119.449-28.98200-0.93154.523Pass

Deviation from Linearity for TNC cells/mm^3 on Instrument BC03127

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.0218.55135.00030.8714.1294913.385.000Pass
0.0437.10245.33349.445-4.11131-8.315.000Pass
0.0655.65263.66768.019-4.35212-6.406.367Pass
0.0874.20389.66786.5933.073743.558.967Pass
0.192.75497.000105.167-8.16707-7.779.700Pass
109275.4009295.0009299.367-4.36665-0.05929.500Pass
2018550.80019726.00018586.4371139.563076.131972.600Pass
3027826.20027860.66727873.507-12.84054-0.052786.067Pass
4037101.60036818.33337160.577-342.24416-0.923681.833Pass
5046377.00046537.66746447.64890.018900.194653.767Pass
6055652.40055888.33355734.718153.615280.285588.833Pass
7064927.80064934.00065021.788-87.78833-0.146493.400Pass
8074203.20074464.66774308.859155.808060.217446.467Pass
9083478.60082905.00083595.929-690.92889-0.838290.500Pass
10092754.00092754.00092882.999-128.99917-0.149275.400Pass

{19}------------------------------------------------

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.0218.97227.33329.292-1.95844-6.695.000Pass
0.0437.94447.66748.503-0.83605-1.725.000Pass
0.0656.91667.00067.714-0.71366-1.056.700Pass
0.0875.88790.33386.9253.408733.929.033Pass
0.194.859102.000106.136-4.13555-3.9010.200Pass
109485.9339593.0009615.552-22.55231-0.23959.300Pass
2018971.86720269.33319221.0241048.309555.452026.933Pass
3028457.80028782.33328826.495-44.16193-0.152878.233Pass
4037943.73338453.66738431.96721.699930.063845.367Pass
5047429.66747965.66748037.438-71.77155-0.154796.567Pass
6056915.60057668.66757642.91025.756980.045766.867Pass
7066401.53366629.33367248.381-619.04783-0.926662.933Pass
8075887.46777711.33376853.853857.480691.127771.133Pass
9085373.40086751.33386459.324292.009210.348675.133Pass
10094859.33394859.33396064.796-1205.4623-1.259485.933Pass

Deviation from Linearity for TNC cells/mm^3 on Instrument BC03129

Deviation from Linearity for TNC cells/mm^3 on Instrument BC03130

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit ($\pm$)Conclusion
0.0218.88531.33331.494-0.16058-0.515.000Pass
0.0437.76950.00050.452-0.45154-0.895.000Pass
0.0656.65468.00069.409-1.40917-2.036.800Pass
0.0875.53990.66788.3672.299882.609.067Pass
0.194.42498.667107.324-8.65775-8.079.867Pass
109442.3679505.6679491.34914.318010.15950.567Pass
2018884.73320211.00018970.1611240.838986.542021.100Pass
3028327.10028414.66728448.973-34.30672-0.122841.467Pass
4037769.46737804.66737927.786-123.11908-0.323780.467Pass
5047211.83346924.00047406.598-482.59811-1.024692.400Pass
6056654.20057072.33356885.410186.922860.335707.233Pass
7066096.56765844.66766364.223-519.55617-0.786584.467Pass
8075538.93375947.33375843.035104.298130.147594.733Pass
9084981.30085579.33385321.848257.485770.308557.933Pass
10094423.66794423.66794800.660-376.99326-0.409442.367Pass

Deviation from Linearity for BF RBC on Instrument BC03127

DilutionExpectedValueMean ofMeasured ValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.015954.3281205.3331144.66960.663875.30500.000Pass
0.021272.4381705.3331463.115242.2180916.55500.000Pass
0.042544.8752654.3332736.898-82.56504-3.02500.000Pass
0.063817.3134231.3334010.682220.651835.50500.000Pass
0.085089.7515632.6675284.465348.202046.59500.000Pass
0.16362.1886485.3336558.248-72.91442-1.11500.000Pass
201272437.671277341.331273972.463368.872810.2663867.067Pass

{20}------------------------------------------------

DilutionExpectedValueMean ofMeasured ValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
301908656.501891584.671910864.02-19279.358-1.0194579.233Pass
402544875.332490719.332547755.59-57036.256-2.24124535.967Pass
503181094.173199415.673184647.1514768.51350.46159970.783Pass
603817313.003813229.673821538.72-8309.0507-0.22190661.483Pass
704453531.834495862.004458430.2837431.71840.84224793.100Pass
805089750.675112301.675095321.8516979.82090.33255615.083Pass
905725969.505736287.675732213.414074.256700.07286814.383Pass
1006362188.336362188.336369104.97-6916.6408-0.11318109.417Pass

Deviation from Linearity for BF RBC on Instrument BC03129

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.015957.1691145.0001122.94722.053251.96500.000Pass
0.021276.2251530.3331442.93087.403816.06500.000Pass
0.042552.4502803.3332722.86180.472732.96500.000Pass
0.063828.6753959.6674002.792-43.12502-1.08500.000Pass
0.085104.9005676.6675282.723393.943907.46500.000Pass
0.16381.1256627.0006562.65464.346160.98500.000Pass
201276225.001274035.671280094.08-6058.4121-0.4763701.783Pass
301914337.501892034.001920059.62-28025.619-1.4694601.700Pass
402552450.002528795.332560025.16-31229.826-1.22126439.767Pass
ર૦3190562.503237730.673199990.7037739.96751.18161886.533Pass
603828675.003843700.673839956.243744.427390.10192185.033Pass
704466787.504521090.674479921.7841168.88730.92226054.533Pass
805104900.005113788.005119887.32-6099.3196-0.12255689.400Pass
905743012.505759083.675759852.86-769.19302-0.01287954.183Pass
1006381125.006381125.006399818.40-18693.400-0.29319056.250Pass

Deviation from Linearity for BF RBC on Instrument BC03130

DilutionExpectedValueMean ofMeasuredValuePredictedValue fromlinearDeviation% DeviationAcceptanceLimit (±)Conclusion
0.015953.0861047.0001113.100-66.09964-5.94500.000Pass
0.021270.7811712.6671433.869278.7978919.44500.000Pass
0.042541.5632663.3332716.945-53.61197-1.97500.000Pass
0.063812.3444286.3334000.022286.311517.16500.000Pass
0.085083.1255300.0005283.09816.901650.32500.000Pass
0.16353.9066897.6676566.175331.491795.05500.000Pass
201270781.271266443.671283227.32-16783.653-1.3163322.183Pass
301906171.901899907.671924765.58-24857.916-1.2994995.383Pass
402541562.532527694.002566303.85-38609.847-1.50126384.700Pass
503176953.173219698.673207842.1111856.55650.37160984.933Pass
603812343.803871521.003849380.3722140.62630.58193576.050Pass
704447734.434590433.004490918.6499514.36272.22229521.650Pass
805083125.075142751.335132456.9010294.43240.20257137.567Pass
905718515.705764944.335773995.16-9050.8312-0.16288247.217Pass
1006353906.336353906.336415533.43-61627.095-0.96317695.317Pass

{21}------------------------------------------------

The data presented in this submission supports the linear range for whole blood and body fluids as described in the Instructions for Use. The data supports that connection of instruments into a workcell configuration does not significantly impact the linearity.

ParameterUnitsLinearity Range Results
WBC103 cells/µL0.064 - 408.5
RBC106 cells/µL0.001 - 8.560
PLT103 cells/µL3.2 - 3002
HGBg/dL0.04 - 26.070
BF-RBCcells/mm31113.10 - 6,353,906
BF-TNCcells/mm331.50 - 92,745
  • e) Carryover: Whole Blood and Body Fluid High to Low Carryover was conducted on the DxH 900-3S workcell per CLSI H26-A2. Three (3) test instruments were used for each of the WBC, RBC, Hgb, and PLT parameters. One (1) replicate from each of the three (3) High Target Value tubes followed by one (1) replicate from each of the three (3) Low Target Value tubes were analyzed. Each of the Low Target Value tubes was only aspirated once. This step was repeated three (3) times for each parameter. The WBC, RBC, Hgb, and PLT parameters for Carryover were calculated using the formula % carryover = [(LTV 1-LTV 3)(HTV 3-LTV 3)] X100, where HTV=high target value, LTV=low target value. The data presented in the submission supports there is no detectable carryover for whole blood or body fluids as described in the Instructions for Use. The data supports that connection of instruments into a workcell configuration does not impact carryover.
  • f) Performance Detection Capability Limits: Limit of Blank (LoB), Lower Limit of Detection (LLoD) and Lower Limit of Quantitation (LLoQ) was conducted on the DxH 900-3S workcell per CLSI EP17-A2 to evaluate the LoB, LLoQ for WBC, PLT, BF-RBC, and BF-TNC.

LoB was performed using DxH diluent as blank in the CBC, CDR, CD and CR test panel in cassette presentation, five (5) replicates of each panel were analyzed for a total of twenty (20) replicates in the AM and PM for three (3) days for a total of 120 whole blood cassette presentation cycles. Twenty (20) replicates of DxH diluent (blank) were analyzed in the single tube presentation as BF cycle in the AM and PM for three (3) days for a total of 120 single tube BF cycles. In both scenarios, two (2) lots of DxH diluent and DxH Cell Lyse reagents were used.

To test for LLoD and LLoQ, a dilution of a fresh whole blood sample was prepared to obtain a stock solution with a value in one of the following target ranges (WBC 60-70 x 103/μL, PLT 7-9 x 103/μL, TNC 45-50 cells/mm3, and BF-RBC 2450- 2470 cells/mm3). Four (4) samples per parameter were obtained for a total of sixteen (16) samples per instrument. Three (3) sets of eleven (11) dilutions of the stock solution were prepared in 10% increments from 0% to 100%. One (1) set of dilutions was analyzed on each of the test instruments. Five (5) replicates of each dilution level were analyzed on each test instrument in the CBC test panel for WBC and PLT and five (5) replicates of each were analyzed on each instrument in the BFC test panel for TNC and BF-RBC.

Bias was established by analyzing five (5) normal whole blood samples each day in duplicate on the test instruments and DxH 800 comparator using the CBC test panel.

{22}------------------------------------------------

The limit of quantitation data provided supports the intended use of the device and the lower limit of the analytical measuring interval per the instructions for use. The data supports that connection of instruments into a workcell configuration does not significantly impact the analytical measuring range. For WBC, PLT, BF-TNC, and BF-RBC, LLoQ are aligned with the background limit and the lower limit of the measuring range. These are parameters where there is a clinical interest (medical decision level) on very low or near zero values.

LLoQ Results

ParameterUnitsAcceptance limit(LoQ)LoQ result
WBCx103 cells/μL≤0.0500.019
PLTx103 cells/μL≤3.0000.757
BF TNCcells/mm3≤20.00014.004
BF RBCcells/mm3≤1000.000979.869
  • g) Reference Ranges: Studies to assess the Adult Reference Interval (Whole Blood) were conducted on the DxH 900 per CLSI EP28-A3c to verify the reference intervals for male and female adults on the DxH 900/DxH 690T Analyzer are the same as the DxH 800 Analyzer (predicate). The data verifies the DxH 800 references ranges presented in the IFU are transferable to the DxH 900/DxH 690T. The data supports that connected workcells have the same reference ranges as a stand-alone instrument.

Reference Ranges for Adults Combined Male and Female

ParameterUnitMean95% Confidence Lower Limit95% Confidence Upper Limit
WBCx 103 cells/µL6.33.611.2
RBCx 106 cells/µL4.523.735.50
Hemoglobing/dL13.411.415.9
HCT%39.033.345.7
MCVfL86.473.795.5
MCHpg29.624.333.2
MCHCg/dL34.232.535.8
RDW%13.812.317.0
RDW-SDfL41.437.147.8
Plateletx 103 cells/µL257159386
MPVfL9.27.511.2
Neutrophil%58.543.376.6
Lymphocyte%29.616.043.5
Monocyte%8.34.512.5
Eosinophil%2.80.67.9
Basophil%0.70.21.4
Neutrophilx 103 cells/µL3.71.87.8
Lymphocytex 103 cells/µL1.81.03.0
Monocytex 103 cells/µL0.50.31.0
Eosinophilx 103 cells/µL0.20.00.5
Basophilx 103 cells/µL0.00.00.1
Reticulocyte%1.100.502.17
Reticulocytex 106 cells/µL0.04980.02210.0963
MRVfL108.897.4120.2
IRFNA0.400.290.53
NRBC%0.10.00.4
NRBCx 103 cells/µL0.010.000.02

{23}------------------------------------------------

Reference Ranges for Adult Females

ParameterUnitMean95% Confidence Lower95% Confidence Upper
WBCx 103 cells/uL6.73.811.8
RBCx 106 cells/uL4.263.634.92
Hemoglobing/dL12.610.914.3
HCT%36.931.241.9
MCVfL86.875.595.3
MCHpg29.624.732.8
MCHCg/dL34.132.335.6
RDW%14.012.317.7
RDW-SDfL42.037.650.3
Plateletx 103 cells/uL278179408
MPVfL9.27.910.8
Neutrophil%59.742.776.8
Lymphocyte%29.416.045.9
Monocyte%7.64.310.9
Eosinophil%2.40.57.0
Basophil%0.70.21.3
Neutrophilx 103 cells/uL4.11.98.2
Lymphocytex 103 cells/uL1.91.13.1
Monocytex 103 cells/uL0.50.20.9
Eosinophilx 103 cells/uL0.20.00.5
Basophilx 103 cells/uL0.00.00.1
Reticulocyte%1.110.512.17
Reticulocytex 106 cells/uL0.040.02300.0935
MRVfL108.96.4118.0
IRFNA0.400.260.52
NRBC%0.10.00.3
NRBCx 103 cells/uL0.010.000.02

Reference Ranges for Adult Males

ParameterUnitMean95%Confidence Lower95%Confidence Upper
WBCx 103 cells/µL5.93.610.2
RBCx 106 cells/µL4.814.065.63
Hemoglobing/dL14.212.516.3
HCT%41.336.747.1
MCVfL86.173.096.2
MCHpg29.623.833.4
MCHCg/dL34.432.536.3
RDW%13.612.116.2
RDW-SDfL40.836.545.9
Plateletx 103 cells/µL234152348
MPVfL9.27.411.4
Neutrophil%57.343.573.5
Lymphocyte%29.815.243.3
Monocyte%9.05.513.7
Eosinophil%3.20.88.1
Basophil%0.70.21.5
Neutrophilx 103 cells/µL3.41.77.6
Lymphocytex 103 cells/µL1.71.03.2
Monocytex 103 cells/µL0.50.31.1
Eosinophilx 103 cells/µL0.20.00.5
Basophilx 103 cells/µL0.00.00.1
Reticulocyte%1.090.422.23
Reticulocytex 106 cells/µL0.05230.01880.1086
MRVfL109.597.5122.7

{24}------------------------------------------------

IRFNA0.410.300.54
NRBC%0.20.00.6
NRBCx 103 cells/µL0.010.000.02
  • h) Whole Blood Method Comparison was tested according to EP09c & CLSI H26-A2 to support substantial equivalency by comparison of the CBC, NRBC, DIFF and RETIC values obtained on the DxH 800 Analyzer (predicate) to the values obtained on the DxH 900 Analyzer . This included adult and pediatric samples. The data provided for whole blood method comparison is comprehensive, covering the analytical measurement interval (AMI). Data was collected from multiple sites on both stand-alone and connected workcell instrument configurations using both legacy and ECO reagents. Data were analyzed pooled and by site to support the substantial equivalence of the DxH 900 to the DxH 800 for CBC, NRBC, DIFF and RETIC parameters. This data supports the intended use of the DxH 900 in stand-alone and connected workcell configurations and supports the use of ECO reagents.
ParameterUnitNReferenceTestDifference95% Confidence Limits
MeanMinMaxMeanMinMaxLowerUpper
HCT%73432.658.4661.9432.358.4060.67-0.297-0.333-0.260
MCHpg73329.6213.7273.6329.7214.1473.760.0990.0670.131
MCHCg/dL73333.2325.3157.5033.4726.0957.500.2360.1990.273
NE10^3 cells/µL6736.690.0192.846.770.0194.690.0820.0300.134
LY10^3 cells/µL6734.330.04257.994.320.05257.91-0.020-0.0490.010
MO10^3 cells/µL6700.980.0056.470.990.0058.590.004-0.0100.017
EO10^3 cells/µL6470.190.004.710.180.001.92-0.005-0.0170.006
BA10^3 cells/µL6660.060.000.850.060.001.210.003-0.0010.007
RETc10^6 cells/µL7260.070.000.820.080.000.890.0030.0020.003
NRBCc10^3 cells/µL6160.110.0030.230.110.0034.920.008-0.0080.023
RBC10^6 cells/µL7343.730.948.203.710.938.06-0.020-0.023-0.016
HGBg/dL73410.832.9619.7610.812.9719.64-0.022-0.029-0.016
MCVfL73488.9754.20128.0588.6354.19128.29-0.342-0.386-0.297
RDW%73316.5011.9739.7816.4912.0839.15-0.004-0.0350.027
RDWSDfL73351.3034.1399.7551.0133.6998.44-0.288-0.377-0.199
PLT10^3 cells/µL687263.103.571477.44263.033.861449.60-0.075-0.7430.592
MPVfL6878.686.1412.838.756.2213.300.0760.0580.094
NE%68163.590.3197.6563.611.0297.620.014-0.0680.096
LY%68124.060.9798.4424.080.6298.090.014-0.0670.095
MO%6789.660.3497.449.650.3297.62-0.014-0.0750.048
EO%6552.180.0118.632.160.0118.26-0.017-0.0380.004
BA%6740.630.015.580.640.036.340.007-0.0170.031
RETp%7272.180.0222.152.260.0423.790.0750.0530.096
IRFN/A7230.440.020.860.450.030.830.002-0.0020.005
MRVfL727116.6075.50168.01115.4983.20169.45-1.115-1.373-0.858
WBC10^3 cells/µL72311.840.08273.5411.900.08282.900.0550.0110.098
NRBC%6190.620.01115.540.610.01130.63-0.005-0.0560.047

General Statistics for All Sites Combined

{25}------------------------------------------------

95% Confidence Limits95% Confidence Limits95% Confidence Limits
ParameterUnitSlopeLowerUpperInterceptLowerUpperCorrelationLowerUpper
HCT%0.9870.9830.9920.117-0.0100.2430.998600.998380.99879
MCHpg1.0190.9991.039-0.464-1.0510.1240.994200.993300.99498
MCHCg/dL1.0240.9821.066-0.571-1.9700.8280.952100.944830.95843
NE10^3 cells/µL1.0101.0041.0170.000-0.0000.0000.994900.994070.99562
LY10^3 cells/µL1.0030.9951.011-0.003-0.0090.0030.999800.999770.99983
MO10^3 cells/µL0.9940.9851.0030.0030.0010.0040.998500.998250.99871
EO10^3 cells/µL1.0080.9861.0290.00002-0.000330.000380.846600.823240.86710
BA10^3 cells/µL1.0100.9401.0800.0002-0.00140.00190.683700.641020.72216
RETc10^6 cells/µL1.0281.0161.0400.00040.00020.00060.987200.985210.98893
NRBCc10^3 cells/µL0.8340.7460.9220.00002-0.000400.000440.999300.999180.99940
RBC10^6 cells/µL0.9940.9900.9970.005-0.0060.0160.999200.999080.99931
HGBg/dL0.9950.9920.9970.0350.0090.0600.999500.999420.99957
MCVfL0.9980.9901.006-0.163-0.8720.5460.997800.997460.99810
RDW%0.9990.9671.0320.004-0.5010.5100.992600.991450.99360
RDWSDfL0.9800.9660.9930.7610.1271.3940.994000.993070.99481
PLT10^3 cells/µL1.0000.9961.0040.025-0.4570.5070.998900.998720.99905
MPVfL1.0120.9911.033-0.025-0.2000.1510.977700.974140.98077
NE%1.0010.9971.005-0.050-0.3210.2210.998600.998370.99880
LY%1.0010.9961.005-0.002-0.1170.1120.998400.998140.99862
MO%1.0050.9921.018-0.058-0.1740.0570.995800.995120.99639
EO%0.9850.9691.0000.016-0.0130.0460.993100.991960.99408
BA%0.8880.7091.0660.078-0.0220.1770.817900.791250.84145
RETp%1.0140.9841.0450.044-0.0110.0980.992100.990870.99317
IRFN/A0.9850.9551.0150.008-0.0060.0230.933500.923440.94228
MRVfL1.0010.9801.021-1.205-3.5711.1610.971600.967220.97540
WBC10^3 cells/µL1.0061.0031.0090.003-0.0090.0150.999600.999540.99965
NRBC%1.1150.8761.354-0.075-0.1740.0230.998300.998010.99855

Regression Statistics and Correlation for All Sites Combined

i) Body Fluid Method Comparison was tested according to EP09c, CLSI H26-A2 and CLSI H56 to establish equivalency of the Body Fluid obtained on the DxH 800 Analyzer (Predicate) to the values obtained on the DxH 900 Analyzer (subject). The data for body fluid method comparison is comprehensive data covering the AMI, from multiple sites including stand-alone and connected workcell instrument configurations using both legacy and ECO reagents. Data were analyzed pooled and by site to support the substantial equivalence of the DxH 900 to the DxH 800 for BF-TNC and BF-RBC parameters. This data supports the intended use of the DxH 900 in stand-alone and connected workcell configuration and supports the use of ECO reagents.

{26}------------------------------------------------

General Statistics of BF for All Sites Combined

ParameterUnitNReferenceTestDifference95% Confidence Limit
MeanMinMaxMeanMinMaxLowerUpper
BF RBCcells/mm^31301155683.061017.446001500.001158284.861085.996034566.002601.802-1857.2237060.827
BF TNCcells/mm^31952837.3120.6956226.962888.1825.0156066.8650.87829.92171.834

Regression Statistics and Correlation of BF for All Sites Combined

ParameterUnitSlope95% Confidence LimitsIntercept95% Confidence LimitsCorrelation95% Confidence Limits
LowerUpperLowerUpperLowerUpper
BF RBCcells/mm^30.9980.9911.005-18.721-95.80858.3670.999910.999880.99994
BF TNCcells/mm^31.0141.0001.0271.101-1.6263.8270.999890.999850.99992

{27}------------------------------------------------

  • Flagging analysis was conducted to evaluate the flagging capabilities of the DxH 900 using i) the flagging results obtained from the samples used in the method comparison study, from three (3) clinical sites, based on CLSI H20-A2 guidelines. A total of 735 residual normal (no flags, marked as negative) and abnormal (contained flags, marked as positive) whole blood samples were tested and compared to the predicate.
    The results were classified according to the following categories:

  • TN (True negative): DxH 900 result and predicate result are both negative (normal sample).

  • TP (True positive): DxH 900 result and predicate result are both positive (abnormal ● sample).

  • . FN (False Negative): DxH 900 gives a negative result (normal sample) whereas predicate gives a positive result (abnormal sample).

  • FP (False Positive): DxH 900 gives a positive result (abnormal sample) whereas predicate . gives a negative result (normal sample).

The NPA, PPA and OPA were calculated for each pathology and abnormality, using the following calculations:

Negative Percent Agreement (NPA): TN / (TN+FP) Positive Percent Agreement (PPA): TP / (TP+FN) Overall Percent Agreement (OPA): (TP+TN) / (TP+FN+FP+TN)

A summary of the results is presented in the table below. All of the distributional cases described in CLSI H20 are important medical decision levels and showed good clinical negative and positive percent agreement between the DxH 900 and DxH 800. The morphological agreement was based on DxH 800 (predicate) flagging for immature granulocytes, variant lymphocytes and left shift. Morphological results showed good clinical negative and positive percent agreement between the DxH 900 and DxH 800.

Category ofAbnormalitiesNTPFPFNTNNPA(95% ConfidenceLimits)PPA(95% ConfidenceLimits)OPA(95% ConfidenceLimits)
MorphologicalAbnormalitiesfor WB7351932325494.9555(0.9341 to 0.9702)0.8853(0.8362 to 0.9211)0.9347(0.9145 to 0.9504)
DistributionalStudy of WB7354193522259.8810(0.8389 to 0.9131)0.9501(0.9256 to 0.9668)0.9224(0.9008 to 0.9397)

Flagging Study NPA, PPA, OPA

  • k) DxH 690T The only differences between the UniCel DxH 900 Coulter Cellular Analysis System and UniCel DxH 690T Coulter Cellular Analysis System are the designed inability to connect as part of a workcell and the lack of the base cabinet with the quad diluent feature (ability to load 4 diluent boxes). All other hardware and software are identical on the DxH 900 and DxH 690T instruments. In order to show that analytical and clinical study data collected on a DxH 900 would also be representative of the DxH 690T, performance verification testing was completed. Testing was performed on one (1) DxH 690T along with one (1) DxH 900 and for accuracy, one (1) DxH 800. Test cases included Daily Checks, Daily Shutdown, Calibration and Initial Control Recover, Daily Control Recovery, Repeatability, Linearity using LIN-X Linearity Control, Comparability- whole blood

{28}------------------------------------------------

accuracy, and Carryover. All test cases met the predetermined acceptance criteria. Results of each test case are summarized here:

  • Daily Checks The Daily Checks cvcle was performed on days when testing was O performed before analysis of control material or blood specimens. Daily Checks results were within the instrument limits and background limits shown each day before the analysis of control material or blood specimens.
ParameterLimit
WBC$\leq$ 0.05 x $10^3$ /uL
RBC$\leq$ 0.005 x $10^6$ /uL
HGB$\leq$ 0.1 g/dL
PLT$\leq$ 3 x $10^3$ /uL
NRBC Region$\leq$ 10 Events
NRBC Total$\leq$ 60 Events
DIFF$\leq$ 100 Events
RET$\leq$ 600 Events

Background Limits

  • Daily Shutdown A shutdown cycle was performed each day prior to Daily O Checks according to the IFU. The shutdown cvcle was performed successfully each day following completion of the testing or in the morning prior to the Daily Checks cycle. The Shutdown cycle put cleaner into the system and removed cleaner and left the instrument ready for the Daily Checks cycle.
  • Calibration and Initial Control Recovery Initial calibration was performed on the o test instruments using the COULTER S-Cal Calibrator. Calibration was verified by analyzing each level of COULTER 6C Plus Cell Control. Calibration was initially performed using S-Cal lot 113161240, expiration 5/4/19. The HGB and MCV were adjusted on Instrument SN BC09305, and the RBC parameter was adjusted on Instrument SN BA51011. For verification, all three levels of 6C Plus Cell Control recovered within the expected ranges set for each level on all test instruments.
  • Daily Control Recovery A minimum of one replicate of COULTER Latron CPo X was analyzed each day after the Daily Checks cycle. A minimum of one replicate of each level of COULTER 6C Plus Cell Control, and COULTER Retic-X Cell Control was analyzed each day during the evaluation period. The controls were analyzed each day that testing was performed on an instrument. In the event of a flag, partial aspiration or non-numeric result on a control run, the control was repeated. In each case, a control result was recovered within the assigned ranges and without flags. For all test instruments, results for all controls were within the expected assay ranges prior to data collection each day data collection was performed.
  • Repeatability Within Run Repeatability for the CBC, Differential, NRBC and O Reticulocyte parameters was performed using ten aspirations (n=10) in the automatic CDR test panel. The customer "Repeatability" application in the software was used to capture the results. Repeatability testing was performed using a combination of normal whole blood samples, abnormal samples, manipulated samples, and COULTER 6C Plus Cell Control Level 2 and Level 3

{29}------------------------------------------------

to obtain the ranges for each parameter required. Testing was performed across three (3) days. The test instruments met the repeatability specifications for all parameters. The COULTER 6C Plus Cell Control Level 2 was used to obtain NRBC >2-5 and COULTER 6C Plus Cell Control Level 3 was used to obtain NRBC >15. Results are listed in the tables below.

ParameterLevelSpecificationNTest ResultMeanTest Result%CV orSDAcceptance
WBC5.00 - 10.00≤ 3.0% CV105.0061.12%Pass
RBC4.50 to 5.50≤ 1.5% CV104.670.51%Pass
Hgb14.00 - 16.00≤ 1.5% CV1014.980.37%Pass
MCV80.00 - 90.00≤ 1.0% CV1087.280.41%Pass
RDW %12.00 - 14.00≤ 2.5% CV1013.470.98%Pass
RDW-SD33.00 - 48.00≤ 2.5% CV1046.551.01%Pass
Platelet200.00 - 400.00≤ 3.5% CV10252.51.37%Pass
MPV8.00 - 10.00≤ 2.5% CV108.661.15%Pass
Neut %50.00 - 60.00≤ 3.5% CV1056.241.17%Pass
Lymph %25.00 - 35.00≤ 5.0% CV1025.482.04%Pass
Mono %5.00 - 10.00≤ 10.0% CV107.615.71%Pass
Eos %2.00 - 5.00≤ 13.5% CVor ≤ 0.5 SD102.586.64 %Pass
Baso %0.50 - 1.50≤ 0.5 SD100.950.23 SDPass
NRBC2.00 - 15.00≤ 20% CV109.174.89%Pass
NRBC>15.00≤ 15% CV1019.491.49%Pass
Retic %0.00 - 1.50≤ 0.25 SD101.1870.09 SDPass
IRF≥ 0.2≤ 20% CV100.427.15%Pass
MRV100.00 - 120.00≤ 5.0% CV10111.930.94%Pass

Repeatability DxH 900 Automatic CDR Mode

Repeatability DxH 690T Automatic CDR Mode

ParameterLevelSpecificationNTest ResultMeanTest Result%CV orSDAcceptance
WBC5.00 - 10.00≤ 3.0% CV107.831.12%Pass
RBC4.50 - 5.50≤ 1.5% CV104.850.64%Pass
Hgb14.00 - 16.00≤ 1.5% CV1015.300.41%Pass
MCV80.00 - 90.00≤ 1.0% CV1086.180.38%Pass
RDW %12.00 - 14.00≤ 2.5% CV1013.680.87Pass
RDW-SD33.00 - 48.00≤ 2.5% CV1046.771.68%Pass
Platelet200.00 - 400.00≤ 3.5% CV10230.81.17%Pass

{30}------------------------------------------------

ParameterLevelSpecificationNTest ResultMeanTest Result%CV orSDAcceptance
MPV8.00 - 10.00≤ 2.5% CV108.690.83%Pass
Neut %50.00 - 60.00≤ 3.5% CV1054.701.46%Pass
Lymph %25.00 - 35.00≤ 5.0% CV1027.392.61%Pass
Mono %5.00 - 10.00≤ 10.0% CV105.855.10%Pass
Eos %2.00 - 5.00≤ 13.5% CVor ≤ 0.5 SD102.138.14%Pass
Baso %0.50 - 1.50≤ 0.5 SD100.830.24 SDPass
NRBC2.00 - 15.00≤ 20% CV108.934.38%Pass
NRBC> 15.00≤ 15% CV1018.863.02%Pass
Retic %0.00 - 1.50≤ 0.25 SD101.220.13 SDPass
IRF≥ 0.20≤ 20% CV100.357.29%Pass
MRV100.0-120.0≤ 5.0% CV10109.771.20%Pass

Linearity using COULTER LIN-X Linearity Controls - Linearity was performed O in the CBC test panel for WBC, RBC, Hgb and PLT using LIN-X Linearity Control. Linearity with the COULTER LIN-X Linearity Control was performed on the DxH 900 and DxH 690T test instruments. LIN-X Linearity Control passed for all data points on both test instruments.

DxH 900 WBC Linearity

Image /page/30/Figure/3 description: The image shows a linearity plot of the test Wbc (BA51011,4/16/2019,Barracuda). The plot has assay values on the x-axis and values ranging from 81.8 to 413 on the y-axis. There are 6 data points on the plot, labeled as Level 1 through Level 6. The data points appear to follow a linear trend.

DxH 900 Hgb Linearity

Image /page/30/Figure/5 description: The image is a linearity plot of the test Hgb (BA51011,4/16/2019,Barracuda). The x-axis is labeled as 'Assay Values' and ranges from -1 to 27. The y-axis is labeled as 'Mean of aspirations per level' and ranges from -1 to 27. There is a linear trendline plotted on the graph.

DxH 900 RBC Linearity

Image /page/30/Figure/7 description: The image is a linearity plot of the test Rbc (BA51011,4/16/2019, Barracuda). The x-axis is labeled as "Assay Values", and the y-axis is labeled as "Mean of aspirations per level". The plot shows a linear relationship between the assay values and the mean of aspirations per level. There are 6 data points on the plot.

DxH 900 PLT Linearity

Image /page/30/Figure/9 description: The image is a linearity plot of the test pit (BA51011, 4/16/2019, Barracuda). The x-axis represents the assay values, ranging from -3 to 5074. The y-axis represents the mean of aspirations per level, also ranging from -3 to 5074. There are data points plotted on the graph, labeled as Level 2, Level 4, Level 6, Level 8, and Level 10.

{31}------------------------------------------------

DxH 690T WBC Linearity

Image /page/31/Figure/1 description: The image is a linearity plot of the test Wbc (BC09306,4/16/2019, Barracuda). The plot shows the mean of aspirations per level on the y-axis and assay values on the x-axis. There are several data points plotted on the graph, and a linear trend line is drawn through the data points. The plot also includes a confidence interval around the trend line.

Image /page/31/Figure/2 description: The image shows the text "DxH 690T Hgb Linearity". The text is in bold, black font. The text appears to be the title of a document or presentation.

DxH 690T RBC Linearity

Image /page/31/Figure/4 description: The image is a linearity plot of the test Rbc (BC09305,4/16/2019, Barracuda). The plot shows the relationship between the mean of aspirations per level and assay values. There are 5 data points on the plot, and they appear to be linearly related. The data points are labeled as Level 1, Level 2, Level 3, Level 4, and Level 5.

DxH 690T PLT Linearity

Image /page/31/Figure/6 description: The image contains two plots titled "Linearity Plot of the Test Hgb (BC09305,4/16/2019 Barracuda)" and "Linearity Plot of the Test Pit (BC09305,4/16/2019.Barracuda)". Both plots show a linear relationship between Assay Values and Mean of aspirations per level. The plots include a line of best fit and confidence intervals.

  • Accuracy A total of 382 whole blood specimens (60 normal, 322 abnormal) O were collected. Accuracy testing was performed in the automatic mode using the CDR test panel by analyzing each blood sample once through the DxH 900 and DxH 690T instruments and the DxH 800 (predicate). COULTER S-Cal Calibrator was analyzed on the test and comparator instruments each day of accuracy testing if zero calibration was required. Results for CBC, Differential, NRBC, RET, IRF, and MRV parameters were compared using the specifications of the DxH 800. Weighted Deming analysis was performed for WBC, RBC, and PLT, and Deming analysis was performed for the WBC Differential, RETIC, IRF, and MRV parameters, and the remaining CBC parameters. The average bias and upper and lower limits of bias for the complete data set, 25th, 50th, 75th percentiles, and medical decision points were compared to the DxH 800 accuracy specifications for each parameter. Although the COULTER S-Cal Calibrator was analyzed every day of testing, no adjustments were required. Regression analysis was performed for NRBC to obtain the coefficient of variation to compare to the DxH 800 specification. Parameters results with "R" flags, had non-numeric, or partial aspiration were excluded from the analysis.
    The CBC, WBC Differential, MRV, RET, IRF, and NRBC results for accuracy met the DxH 800 accuracy specifications for the DxH 900 and DxH 690T instruments. Mean difference and average lower and upper bias statistics for the CBC, WBC differential, MRV and IRF parameter comparisons follow. The acceptance limit was calculated from the DxH 800 accuracy specification. Results for the 25th, 50th and 75th percentiles and medical decision levels are not shown, but all recovered within the accuracy specification.

{32}------------------------------------------------

ParameterNClinical RangeMeanComp.MeanTestMeanDiffLowerBiasUpperBiasAcc.*Limit
WBC3760.06 - 148.839.889.82-0.07-0.10-0.04± 0.20
RBC3761.68 - 8.493.843.840.010.000.01± 0.07
Hgb3744.95 -24.8411.1211.10-0.02-0.03-0.01± 0.3
MCV37466.15 -115.0688.0587.66-0.39-0.43-0.34± 1.7
RDW37411.93 - 27.6016.5216.46-0.06-0.09-0.03± 0.80
RDW-SD37435.88 - 92.3150.9451.080.140.020.25± 5.00
PLT3387.80 -788.70230.05224.93-5.12-5.97-4.27± 16.10
MPV3386.58-12.218.738.900.170.150.19± 0.60
Neutrophil3460.66-96.264.2663.98-0.28-0.39-0.18± 6.40
Lymphocyte3461.03-9823.0123.420.410.310.51± 2.30
Monocyte3460.00-92.739.819.67-0.15-0.23-0.07± 1.00
Eosinophil3460.00 - 20.302.312.300.00-0.030.02± 0.50
Basophil3460.00 - 3.610.610.630.02-0.010.06± 0.50
RET3740.03-15.201.811.820.01-0.010.03± 0.50
MRV37476.42-170.24117.49117.28-0.22-0.610.18± 15.00
IRF3740.00-0.770.450.440.00-0.010.00± 0.20

Mean Difference/Bias Statistics DxH 900 vs DxH 800 (predicate)

  • = Acceptance limit calculated from Deming Regression analysis

Mean Difference/Bias Statistics DxH 690T vs DxH 800 (predicate)

ParameterNClinical RangeMeanComp.MeanTestMeanDiffLowerBiasUpperBiasAcc.*Limit
WBC3680.06 - 148.839.869.82-0.04-0.080.01$\pm$ 0.20
RBC3691.68 - 8.493.853.84-0.01-0.02-0.01$\pm$ 0.07
Hgb3674.95 -24.8411.1611.13-0.03-0.05-0.02$\pm$ 0.30
MCV36766.15 -115.0687.9987.71-0.28-0.32-0.23$\pm$ 1.70
RDW36711.93 - 27.6016.5116.49-0.02-0.040.01$\pm$ 0.80
RDW-SD36735.88 - 92.3150.8851.150.260.150.37$\pm$ 5.00
PLT3347.80 -788.70230.45231.481.030.251.81$\pm$ 16.10
MPV3356.58-12.218.738.68-0.06-0.09-0.03$\pm$ 0.60
Neutrophil3400.66-96.264.6964.42-0.27-0.38-0.16$\pm$ 6.40
Lymphocyte3401.03-9822.5622.940.380.280.47$\pm$ 2.20
Monocyte3400.00-92.739.799.68-0.11-0.20-0.02$\pm$ 1.00
Eosinophil3400.00 - 20.302.342.34-0.01-0.040.02$\pm$ 0.50
Basophil3400.00 - 3.610.610.630.02-0.030.07$\pm$ 0.50
RET3710.03-15.201.781.900.120.090.15$\pm$ 0.50
MRV37176.42-170.24117.27117.09-0.17-0.590.25$\pm$ 15.0
IRF3710.00-0.770.440.44-0.01-0.010.00$\pm$ 0.20
  • = Acceptance limit calculated from Deming Regression analysis

{33}------------------------------------------------

ParameterUnitSlope95% Confidence LimitsIntercept95% ConfidenceLimitsCorrelation95% ConfidenceLimits
LowerUpperLowerUpperLowerUpper
WBCx 103 cells/µL0.9940.9910.996-0.002-0.0040.0000.999690.999620.99974
RBCx 106 cells/µL1.0131.0101.017-0.042-0.054-0.0300.999550.999450.99963
Hemoglobing/dL0.9890.9860.9930.1010.0610.1410.999570.999470.99965
MCVfL0.9950.9881.0030.020-0.6230.6630.997980.997520.99835
RDW%0.9990.9871.012-0.042-0.2350.1520.99540.994370.99625
RDW-SDfL1.0050.9881.022-0.120-0.9470.7070.993950.992590.99506
Plateletx 103 cells/µL0.9790.9740.983-0.456-0.9940.0820.998960.998720.99916
MPVfL1.0190.9951.0430.009-0.1900.2090.982820.978750.98611
Neutrophil%1.0010.9951.007-0.347-0.7530.0590.998780.998490.99901
Lymphocyte%1.0070.9981.0150.2580.0850.4310.998550.99820.99882
Monocyte%0.9890.9651.012-0.036-0.2380.1660.997030.996330.9976
Eosinophil%0.9970.9821.0120.003-0.0280.0340.996170.995270.9969
Basophil%0.9640.6041.3230.045-0.1480.2380.61410.543880.67578
Reticulocyte%0.9870.9491.0260.033-0.0270.0930.99160.989710.99314
MRVfL0.9890.9601.0191.051-2.5084.6100.968920.962040.97457
IRF%0.9600.9111.0080.016-0.0090.0400.927730.912130.94064

Regression Statistics and Correlation for DxH 900 (Test) vs DxH 800 (Predicate)

Regression Statistics and Correlation for DxH 690T (Test) vs DxH 800 (Predicate)

Parameter95% Confidence Limits95% Confidence LimitsCorrelation95% Confidence Limits
UnitSlopeLowerUpperInterceptLowerUpperLowerUpper
WBCx 103 cells/µL0.9940.9920.9970.000-0.0020.0020.999450.999320.99955
RBCx 106 cells/µL1.0081.0041.012-0.039-0.053-0.0250.999550.999440.99963
Hemoglobing/dL0.9850.9820.9890.1300.0920.1680.999550.999450.99964
MCVfL1.0010.9941.008-0.342-0.9500.2670.997840.997350.99824
RDW%1.0030.9911.016-0.072-0.2690.1240.995930.995000.99668
RDW-SDfL1.0171.0011.034-0.627-1.4110.1580.994600.993380.99560
Plateletx 103 cells/µL1.0121.0011.024-0.854-2.2360.5270.999010.998780.99920
MPVfL0.9950.9701.020-0.015-0.2270.1960.968690.961320.97468
Neutrophil%1.0020.9961.009-0.428-0.832-0.0230.998620.998290.99889
Lymphocyte%1.0121.0041.0200.117-0.0650.2980.998540.998190.99882
Monocyte%0.9720.9590.9860.1580.0250.2900.996470.995640.99715
Eosinophil%0.9920.9721.0120.011-0.0290.0500.995260.994130.99617
Basophil%1.4810.3832.579-0.275-0.275-0.2750.473000.386060.55161
Reticulocyte%1.0310.9871.0740.062-0.0040.1290.987880.985150.99011
MRVfL0.9960.9611.0300.349-3.7924.4910.965120.957380.97148
IRF%1.0180.9711.065-0.014-0.0380.0090.924360.908010.93791

{34}------------------------------------------------

  • Carryover Within mode carryover was performed in the automatic CDR mode using o the customer "Carryover" application in the software. Normal whole blood, clinical samples, and blood altered by concentration, or the addition of analogs were used to achieve the following levels for testing:
    • WBC > 90.0

    • WBC > 300

    • ≫ RBC > 6.20
    • ≫ Hgb > 22.0
    • PLT > 900

Carryover was performed on the DxH 690T and DxH 900. Carryover results for both test instruments recovered within the DxH 800 carryover limits.

Carryover WBC > 90

CnDRModeWBCRBCHgbPLTDiffNRBCRetic
DxH 690T0.11%0.03%0.26%0.07%22, 20, 3512, 5, 725, 20, 22
DxH 9000.09%0.05%0.23%0.17%11,15,377,1,310,6,3
Spec<0.5%<0.5%<1.0%<1.0%<200 Events<75 Events<600 Events

Carryover WBC > 300

CnDRModeWBCRBCHgbPLTDiffNRBCRetic
DxH 690T0.18%0.04%0.18%0.02%42, 36, 1624, 12,1027, 20, 25
DxH 9000.13%0.05%0.13%0.06%31, 29, 2815, 4, 119, 6, 7
Spec≤0.5%≤0.5%≤1.0%≤1.0%≤200 Events≤100 Events≤600 Events

Carryover RBC > 6.20

CnDRModeWBCRBCHgbPLTDiffNRBCRetic
DxH 690T0.25%0.03%0.30%0.12%25, 15, 2412, 16, 923, 15, 16
DxH 9000.22%0.04%0.16%-0.03%30, 14, 3911, 2, 127, 8, 11
Spec≤0.5%≤0.5%≤1.0%≤1.0%≤200 Events≤75 Events≤600 Events

Carryover Hgb > 22.0

CnDRModeWBCRBCHgbPLTDiffNRBCRetic
DxH 690T0.20%0.03%0.17%0.03%31, 16, 159, 3, 644, 9, 4
DxH 9000.19%0.03%0.17%-0.23%14, 37, 281, 5, 337, 8, 3
Spec≤0.5%≤0.5%≤1.0%≤1.0%≤200 Events≤75 Events≤600 Events

{35}------------------------------------------------

Carryover Platelet > 900

CnDRModeWBCRBCHgbPLTDiffNRBCRetic
DxH 690T0.15%0.03%0.13%0.10%22, 15, 1411, 5, 812, 15, 11
DxH 9000.13%0.05%0.22%0.09%31, 26, 232, 2, 316, 13, 13
Spec≤0.5%≤0.5%≤1.0%≤1.0%≤200 Events≤75 Events≤600 Events

807.92 (b)(3): Conclusions from Analytical and Performance Data

All data provided for the nonclinical and clinical testing demonstrates that the UniCel DxH 900/DxH 690T Coulter Cellular Analysis System (DxH 900, DxH 690T) and the UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System (SMS II) are as safe, as effective, and perform as well as or better than the UniCel DxH 800 Coulter Cellular Analysis System (DxH 800) and the UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System (SMS).

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”