K Number
K232962

Validate with FDA (Live)

Date Cleared
2024-04-19

(211 days)

Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The iX series Patient Monitors including iX10, iX12, iX15 are intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults and pediatics (including neonates). The monitors are intended for use by trained healthcare professionals in hospital environments.

The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), and Anaesthesia gas (AG).

The arrhythmia detection and ST Segment analysis are intended for adult patients.

The NIBP monitoring supports iCUFS algorithm and iFAST algorithm is intended for adult, pediatric and neonatal patients. The iFAST algorithm is intended for adult and pediatic patients (≥3 years of age). Both measurement algorithms are also intended for use with pregnant women, including pre-eclamptic patients. NIBP MAP is not applicable to pregnant women.

The Spot Temp with T2A module can only measure temperature of adult and pediatric (> 1 year of age) patients. The monitors are not intended for MRI environments.

The cardiac output (C.O.) is only intended for adult patients.

Device Description

The iX series Patient Monitors including iX10, iX12, iX15 can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormalities so that doctors and nurses can respond to the patient's situation as appropriate.

AI/ML Overview

This document describes the premarket notification (510(k)) for the Edan Instruments, Inc. Patient Monitor (iX10, iX12, iX15) and its equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the new device were generally established by compliance with various international standards for medical electrical equipment and specific performance characteristics. The reported device performance indicates that the device meets these standards and its specified accuracy.

For the Quick Temp Module (T2A), the provided text details specific clinical study results. Other parameters were tested for compliance with relevant standards through non-clinical testing.

Parameter/CharacteristicAcceptance Criteria (Reference Standard/Specification)Reported Device Performance
General Safety and PerformanceANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]Complied with the standard
Electromagnetic Compatibility (EMC)IEC 60601-1-2:2014+A1:2020Complied with the standard
Alarm SystemsIEC 60601-1-8:2020Met relevant consensus standards
Electrocardiographs (ECG)IEC 60601-2-25:2011Met relevant consensus standards
ECG Monitoring EquipmentIEC 60601-2-27:2011Met relevant consensus standards
Automated Non-Invasive Sphygmomanometers (NIBP)IEC 80601-2-30:2018Met relevant consensus standards
Invasive Blood Pressure Monitoring Equipment (IBP)IEC 60601-2-34:2011Met relevant consensus standards
Multi-function Patient Monitoring EquipmentIEC 60601-2-49:2018Met relevant consensus standards
Respiratory Gas Monitors (CO2, AG)ISO 80601-2-55:2018Met relevant consensus standards
Clinical Thermometers (TEMP - General)ISO 80601-2-56:2017+A1:2018Met relevant consensus standards
Pulse Oximeter Equipment (SpO2)ISO 80601-2-61:2017Met relevant consensus standards
Wireless CoexistenceIEEE ANSI C63.27:2017Met relevant consensus standards
Cardiac Rhythm and ST Segment Measurement AlgorithmsANSI AAMI EC57:2012Met relevant consensus standards
Quick Temp Module (T2A) Predict Mode Clinical AccuracyISO 80601-2-56:2017/Amd.1:2018(E) - Temperature measurement requirementsMeets the requirements of the standard and acceptance criteria in clinical protocol

2. Sample Size for Test Set and Data Provenance

For the Quick Temp Module (T2A) Predict Mode clinical accuracy study:

  • Sample Size: 142 valid cases for sublingual and axillary temperature measurements.
  • Data Provenance: Clinical investigation (prospective study). No country of origin is explicitly stated, but the company is based in China.

For other parameters, specific sample sizes for non-clinical (bench) testing are not provided, but the document states "Edan has conducted functional and system level testing to validate the performance of the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards."

3. Number of Experts and their Qualifications for Ground Truth

The document does not specify the number or qualifications of experts used to establish ground truth for the clinical study of the Quick Temp Module (T2A). It only mentions that the study compared the new module against the "direct mode of F3000 Temp Module of M3A Vital signs monitor," implying the predicate device served as a reference for accuracy.

For other non-clinical tests, the ground truth is implicitly defined by the specifications and performance requirements outlined in the referenced international standards.

4. Adjudication Method for the Test Set

The document does not describe a specific adjudication method (e.g., 2+1, 3+1) for the clinical study or any other test sets.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No multi-reader multi-case (MRMC) comparative effectiveness study is mentioned in the provided text. The device is a patient monitor, and its performance evaluation typically focuses on the accuracy and reliability of its physiological parameter measurements, not on improving human reader performance with AI assistance.

6. Standalone (Algorithm Only) Performance Study

Yes, standalone performance was evaluated for the components of the device. The non-clinical data section describes extensive functional and system-level testing to validate the performance against relevant consensus standards. This includes specific tests for algorithms like cardiac rhythm and ST segment measurement. The clinical study for the Quick Temp Module (T2A) also assesses the standalone performance of that specific module.

7. Type of Ground Truth Used

  • For the Quick Temp Module (T2A) clinical study: The ground truth was established by comparison to a reference device's direct temperature measurement mode ("direct mode of F3000 Temp Module of M3A Vital signs monitor"). This is a form of reference standard comparison (using a predicate device as the reference).
  • For other parameters (e.g., ECG, NIBP, SpO2, CO2, AG): The ground truth for non-clinical testing is implicitly based on the specifications and performance requirements outlined in the referenced international consensus standards (e.g., IEC 60601 series, ISO 80601 series, AAMI standards).

8. Sample Size for the Training Set

The document does not provide information about a training set since this is a patient monitoring device and not a machine learning algorithm as typically understood in the context of large-scale image-based diagnostics. The "algorithms" mentioned (e.g., arrhythmia detection, NIBP algorithms) are likely engineered signal processing algorithms rather than deep learning models requiring large training datasets.

9. How Ground Truth for Training Set was Established

Not applicable, as no training set for a machine learning model is described in the provided text. The algorithms for the patient monitor's functions are developed and validated against established physiological principles and engineering standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 19, 2024

Edan Instruments, Inc Tracy Yue Regulatory Affairs Engineer 15 Jinhui Road, Jinsha Community, Kengzi Sub-district Pingshan District Shenzhen. Shenzhen 518122 China

Re: K232962

Trade/Device Name: Patient Monitor (iX10, iX12, iX15) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, DPS, DSI, MLD, DRT, DXN, DSK, FLL, DQA, CCK, CBS, CBR, CCL, CBQ, NHO, NHQ, NHP Dated: March 22, 2024 Received: March 22, 2024

Dear Tracy Yue:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer W. Shih -S

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices

{2}------------------------------------------------

Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232962

Device Name Patient Monitor, Model: iX10, iX12, iX15

Indications for Use (Describe)

The iX series Patient Monitors including iX10, iX12, iX15 are intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults and pediatics (including neonates). The monitors are intended for use by trained healthcare professionals in hospital environments.

The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), and Anaesthesia gas (AG).

The arrhythmia detection and ST Segment analysis are intended for adult patients.

The NIBP monitoring supports iCUFS algorithm and iFAST algorithm is intended for adult, pediatric and neonatal patients. The iFAST algorithm is intended for adult and pediatic patients (≥3 years of age). Both measurement algorithms are also intended for use with pregnant women, including pre-eclamptic patients. NIBP MAP is not applicable to pregnant women.

The Spot Temp with T2A module can only measure temperature of adult and pediatric (> 1 year of age) patients. The monitors are not intended for MRI environments.

The cardiac output (C.O.) is only intended for adult patients.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92

1. Submitter:

Applicant:Edan Instruments, Inc.#15 Jinhui Road, Jinsha Community,Kengzi Sub-District, Pingshan District,Shenzhen, 518122 P.R.China.Tel: +86(0755) 26858736Fax: +86(0755) 26882223
Contact person:Tracy Yue
Contact person:Tracy Yue
Preparing date:September 21, 2023

2. Device name and classification:

Trade name:Patient Monitor, Model: iX10, iX12, iX15
Common/Usual Name:Patient Monitor
Classification Name:21 CFR 870.1025Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Regulatory class:Class II
Primary product code:MHX-Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)
Subsequent product code:Regulation number/Device Product Code 21 CFR 868.2375 DPS
Regulation number/DeviceProduct Code
21 CFR 868.2375ElectrocardiographDPS
21 CFR 870.2340Detector and Alarm, ArrhythmiaDSI
21 CFR 870.1025Monitor, ST Segment with AlarmMLD
21 CFR 870.2300Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm)DRT
21 CFR 870.1130System, Measurement, Blood-Pressure, Non-InvasiveDXN
21 CFR 870.1110Computer, Blood-PressureDSK
21 CFR 880.2910Thermometer, Electronic, ClinicalFLL
21 CFR 870.2700OximeterDQA

{5}------------------------------------------------

21 CFR 868.2375Monitor, Breathing FrequencyBZQ
21 CFR 870.1400Analyzer, Gas, Carbon-Dioxide, Gaseous-PhaseCCK
21 CFR 868.1620Analyzer, Gas, Halothane, Gaseous-Phase (AnestheticConc.)CBS
21 CFR 868.1700Analyzer, Gas, Nitrous-Oxide, Gaseous Phase (AnestheticConc.)CBR
21 CFR 868.1720Analyzer, Gas, Oxygen, Gaseous-PhaseCCL
21 CFR 868.1500Analyzer, Gas, Enflurane, Gaseous-Phase (AnestheticConcentration)CBQ
21 CFR 868.1500Analyzer, Gas, Desflurane, Gaseous-Phase (AnestheticConcentration)NHO
21 CFR 868.1500Analyzer, Gas, Isoflurane, Gaseous-Phase (AnestheticConcentration)NHQ
21 CFR 868.1500Analyzer, Gas, Sevoflurane, Gaseous-Phase (AnestheticConcentration)NHP

3. Predicate Device(s):

    1. Edan Instruments, Inc., Patient Monitor Model iM50, iM60, iM70, iM80, K202336 (Primary)
    1. Shenzhen Mindray Bio-Medical Electronics Co., LTD., VS9 Vital Signs Monitor, K211475 (Reference)
    1. Edan Instruments, Inc., Vital signs monitor Model iM3, K180380 (Reference)
    1. Edan Instruments, Inc., Vital signs monitor Model iM3s, K202892 (Reference)
    1. Edan Instruments, Inc., Patient Monitor Model RespArray, K220308 (Reference)
    1. Shenzhen Mindray Bio-Medical Electronics Co., LTD., BeneVision N Series Patient Monitors, K182075 (Reference)

{6}------------------------------------------------

4. Device Description:The iX series Patient Monitors including iX10, iX12, iX15 can perform long-timecontinuous monitoring of multiple physiological parameters. Also, it is capable ofstoring, displaying, analyzing and controlling measurements, and it will indicatealarms in case of abnormalities so that doctors and nurses can respond to the patient'ssituation as appropriate.
5. Indication for UseThe iX series Patient Monitors including iX10, iX12, iX15 are intended to be used formonitoring, storing, recording, and reviewing of, and to generate alarms for, multiplephysiological parameters of adults and pediatrics (including neonates). The monitorsare intended for use by trained healthcare professionals in hospital environments.The monitored physiological parameters include: ECG, respiration (RESP), temperature(TEMP), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR),non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2),cardiac output (C.O.), and Anaesthesia gas (AG).The arrhythmia detection and ST Segment analysis are intended for adult patients.The NIBP monitoring supports iCUFS algorithm and iFAST algorithm. The iCUFSalgorithm is intended for adult, pediatric and neonatal patients. The iFAST algorithm isintended for adult and pediatric patients (≥3 years of age). Both measurement algorithmsare also intended for use with pregnant women, including pre-eclamptic patients. NIBPMAP is not applicable to pregnant women.The Spot Temp with T2A module can only measure temperature of adult and pediatric (>1year of age) patients.The monitors are not intended for MRI environments.The cardiac output (C.O.) is only intended for adult patients.

6. Predicate Device Comparison

The table below compares the indication for use and key technological feature of the subject devices to the predicate device (Patient Monitor Model iM50, iM60, iM70, iM80, K202336).

Note: "/" means no such functions in predicate device.

K#Subject Device(iX10, iX12, iX15)Predicate Device(iM50, iM60, iM70, iM80)
K232962K232336K202336
DescriptionThe iX series Patient Monitorsincluding iX10, iX12, iX15 areintended to be used for monitoring,storing, recording, and reviewing of,and to generate alarms for, multiplephysiological parameters of adultsand pediatrics (including neonates)The monitors are intended for useby trained healthcare professionals inhospital environments. The monitoredphysiological parametersThe monitors are intended to be used formonitoring, storing, recording, andreviewing of, and to generate alarms for,multiple physiological parameters of adults,pediatrics and neonates. The monitors areintended for use by trained healthcareprofessionals in hospital environments.
include: ECG, respiration (RESP), temperature(TEMP), functional oxygen saturation ofarterial hemoglobin (SpO2), pulse rate (PR),non-invasive blood pressure (NIBP), invasiveblood pressure (IBP), carbon dioxide (CO2),cardiac output (C.O.), and Anaesthesia gas(AG).The monitored physiological parametersinclude: ECG, respiration (RESP),temperature (TEMP), oxygen saturation ofarterial blood (SpO2), pulse rate (PR),non-invasive blood pressure (NIBP),invasive blood pressure (IBP), carbondioxide (CO2), cardiac output (C.O.), andAnaesthesia gas(AG).
The arrhythmia detection and ST Segmentanalysis are intended for adult patients.The arrhythmia detection and ST Segmentanalysis are intended for adult patients.
The NIBP monitoring supports iCUFSalgorithm and iFAST algorithm. The iCUFSalgorithm is intended for adult, pediatricand neonatal patients. The iFAST algorithmis intended for adult and pediatric patients(≥3 years of age). Both measurementalgorithms are also intended for use withpregnant women, including pre-eclampticpatients. NIBP MAP is not applicable topregnant women.The monitors are not intended for MRIenvironments.
The Spot Temp with T2A module can onlymeasure temperature of adult and pediatric(> 1 year of age) patients.
The monitors are not intended for MRIenvironments.
The cardiac output (C.O.) is only intendedfor adult patients.
ECG module
Lead Mode3 Electrodes; 5 Electrodes; 6 Electrodes; 10Electrodes3 Electrodes; 5 Electrodes; 6 Electrodes; 10Electrodes
Arrhythmia analysesASYSTOLE, VFIB/VTAC, COUPLET, VT >2, BIGEMINY, TRIGEMINY, VENT, R on T,PVC, TACHY, BRADY, MISSED BEATS,IRR, VBRADY, PNC, PNPASYSTOLE, VFIB/VTAC, COUPLET,VT > 2, BIGEMINY, TRIGEMINY, VENT,R on T, PVC, TACHY, BRADY, MISSEDBEATS, IRR, VBRADY, PNC, PNP
The arrhythmia detection algorithm is the same as predicate device K202336
ST value
Measurement Range-2.0 mV to +2.0 mV-2.0 mV to +2.0 mV
Pace
Pulse IndicatorAmplitude: ±2 mV to ±700 mVWidth: 0.1 ms to 2.0 msAscending time: 10 us to 100 usAmplitude: ±2 mV to ±700 mVWidth: 0.1 ms to 2.0 msAscending time: 10 us to 100 us
PVC
RangeADU: (0 to 300) PVCs/ minADU: (0 to 300) PVCs/ min
PED/NEO: (0 to 350) PVCs/ minPED/NEO: (0 to 350) PVCs/ min
HR
Measurement rangeADU: 15 bpm to 300 bpmADU: 15 bpm to 300 bpm
PED/NEO: 15 bpm to 350 bpmPED/NEO: 15 bpm to 350 bpm
QT Analysis
QT/QTc/ΔQTcmeasurementQT Range: 200 ms ~ 800 msQTc Range :200ms ~ 800 ms△ QTc Range: -600 ms ~ 600 msQT Range: 200 ms ~ 800 msQTc Range :200ms ~ 800 ms△ QTc Range: -600 ms ~ 600 ms
RESP module
Principle of OperationImpedance between RA-LL, RA-LAImpedance between RA-LL, RA-LA
Measurement Range0 rpm to 200 rpmAdult: 0 to 120 rpmPediatric/neonate: 0 rpm to 150rpm
NIBP module (EDAN)
TechniqueOscillometryOscillometry
Measurement ModeiCUFS, iFASTiCUFS
EDAN-NIBP(iCUFS)
Intended usersAdult, pediatric, neonatesAdult, pediatric, neonates
Measuring parameterSYS, DIA, MAP, PR
Measurement RangeAdult Pediatric NeonateSystolic 25-290 25-240 25-140Diastolic 10-250 10-200 10-115Mean 15-260 15-215 15-125Adult Pediatric NeonateSystolic 25-290 25-240 25-140Diastolic 10-250 10-200 10-115Mean 15-260 15-215 15-125
PR from NIBP
Measurement range40 bpm to 240 bpm40 to 240 bpm
EDAN-NIBP(iFAST)
Intended usersAdult, pediatric
Measuring parameterSYS, DIA, MAP
Measurement RangeAdult Mode:SYS: 25 mmHg to 290 mmHgDIA: 10 mmHg to 250 mmHgMAP: 15 mmHg to 260 mmHg
Pediatric Mode:
SYS: 25 mmHg to 240 mmHg

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

DIA: 10 mmHg to 200 mmHg
MAP: 15 mmHg to 215 mmHg
SpO2 module(EDAN)
Measurement RangeSpO20% to 100%SpO20% to 100%
Pulse Rate25 to 300 bpmPulse Rate25 to 300 bpm
Nellcor SpO2 module is the same as K202336
Temperature(TEMP) module
Number of channels22
Measurement Range0 °C to 50 °C(32 °F to 122 °F)0 °C to 50 °C(32 °F to 122 °F)
Spot Temp module
Measurement site: oral/axillary/rectal
Quick Temp (T2Amodule)Monitor mode: 25 °C ~45 °C
Predict mode: 35.5 °C ~42 °C
Accuracy (Monitor mode): 0.1 °C (25 °C ~45 °C)
The Quick Temp (T2A) module is similar to K180380, and the performance of the Predict mode of the module is
validated by the clinical trial.
Infrared Temp(TAT-5000S-RS232Thermometer)Measuring Range : 16 °C~ 43 °C (61 °F ~110 °F)
The Infrared Temp module is the same as K202892
The Quick Temp module and Infrared Temp module is applicable to iX10 and iX12
IBP module
ART, Ao, UAP, BAP, FAP, LV, P1-P4: (-50 to+400) mmHgArt: (0 to +300) mmHg
Measurement RangePA/PAWP: (-6 to +120) mmHgPA/PAWP: (-6 to +120) mmHg
CVP/RAP/LAP/ICP: (-10 to +40) mmHgCVP/RAP/LAP/ICP: (-10 to +40) mmHgP1/P2: (-50 to +300) mmHg
C.O. Module
TechniqueThermodilution TechniqueThermodilution Technique
C.O.: 0.1 to 20L/minC.O.: 0.1 to 20L/min
Measurement rangeTB: 23 °C to 43 °C(73.4 °F to 109.4 °F)TB: 23° C to 43 °C(73.4 °F to 109.4 °F)
TI: -1 °C to 27 °C(30.2 °F to 80.6 °F)TI: -1 °C to 27 °C(30.2 °F to 80.6 °F)
CO2 Module (EDAN G2)
Intended PatientAdult, pediatric, neonatalAdult, pediatric, neonatal
Measure ParametersEtCO2, FiCO2, AwRREtCO2, FiCO2, AwRR
Measuring RangeCO2:0 mmHg to 150 mmHg (0 % to 20%)CO2:0 mmHg to 150 mmHg (0 % to 20%)
AwRR: 0 rpm to 150 rpmAwRR: 2 rpm to 150 rpm
Respironics and Masimo CO2 module are the same as K202336
AG module (EDAN G7 module)
Measure ParametersCO2、N2O、O2、HAL、ISO、ENF、SEV、CO2、N2O、O2、HAL、ISO、ENF、SEV、

{10}------------------------------------------------

DES、AwRR、MACDES、AwRR、MAC
Measuring RangeCO2: 0%~15Vol %CO2: 0%~15Vol %
N2O: 0%~100 Vol %N2O: 0%~100 Vol %
HAL/ISO: 0%-8Vol %HAL/ISO: 0%-8Vol %
ENF: 0%-8 Vol %ENF: 0%-8 Vol %
SEV: 0%-10 Vol %SEV: 0%-10 Vol %
DES: 0%-20 Vol %DES: 0%-20 Vol %
O2: 0%~100 Vol %O2: 0%~100 Vol %
AwRR: 2rpm-100rpmAwRR: 2rpm-100rpm
Masimo AG module are the same as K202336
WI-FI
IEEE802.11a/b/g/n802.11a/b/g/n
Frequency Band2.4 GHz ISM band & 5 G ISM band2.4 GHz ISM band & 5 G ISM band
e-Link
e-LinkBluetooth Low Energy 4.0/
Power supply
AC power
Requirement100-240V, 50/60Hz100-240V, 50/60 Hz
Battery
RechargeableBatteryYesYes

As seen in the comparison tables, the subject and predicate device have similar design features and performance specifications. The technological differences between the subject and predicate device do not raise different questions of safety or effectiveness.

7. Performance Data:

Non-clinical data:

Electrical safety and electromagnetic compatibility (EMC)

iX Series Patient Monitors were assessed for conformity with the relevant requirements of the following standards and found to comply:

  • ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
  • IEC 60601-1-2:2014+A1:2020 Medical electrical equipment Part 1-2: General requirements for basic ● safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests.

{11}------------------------------------------------

Performance testing-Bench

Edan has conducted functional and system level testing to validate the performance of the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.

  • IEC 60601-1-8:2020 Medical electrical equipment - part 1-8: general requirements for basic safety and essential performance - collateral standard: general requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
  • IEC 60601-2-25:2011 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs
  • IEC 60601-2-27:2011 Medical electrical equipment--Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
  • IEC 80601-2-30:2018 Medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
  • IEC 60601-2-34:2011 Medical electrical equipment - part 2-34: particular requirements for the basic safety, including essential performance, of invasive blood pressure monitoring equipment
  • IEC 60601-2-49:2018 Medical electrical equipment –Part 2–49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
  • ISO 80601-2-55:2018 Medical electrical equipment - part 2-55: particular requirements for the basic safety and essential performance of respiratory gas monitors
  • ISO 80601-2-56:2017+A1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
  • ISO 80601-2-61:2017 Medical electrical equipment - part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment
  • IEEE ANSI C63.27:2017 American National Standard for Evaluation of Wireless Coexistence
  • ANSI AAMI EC57:2012 Testing and reporting performance results of cardiac rhythm and ST segment measurement algorithms

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and Food and Drug Administration Staff, "Content of Premarket Submissions for Device Software Functions".

Clinical data:

To meet the requirements for the validity and accuracy of the TEMP measurement of the Quick Temp Module (T2A), EDAN conducted clinical investigation according to the requirements of the standard ISO 80601-2-56

{12}------------------------------------------------

Second edition 2017-03 + AMD1:2018: Medical electrical equipment- Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. A clinical accuracy study evaluated 142 valid cases of sublingual and axillary temperature, which were performed on the following two age groups: >1 and <5 years old and ≥5 years old in accordance with ISO 80601-2-56:2017/Amd.1:2018(E) to compare the direct mode of F3000 Temp Module of M3A Vital signs monitor. The age of subjects is from >1 year old to 73 years old. The total number of hyperthermia subjects shall be no less than 30% and no more than 50% of all subjects in the selected age group. Statistical results show that, the temperature measurement function of the T2A module in Predict mode meets the requirements of standard ISO 80601-2-56:2017+AMD1:2018 for temperature measurement and meets the acceptance criteria in clinical protocol. During the entire clinical trial, all subjects are in good condition, and none of the subjects have discomfort symptoms or adverse reactions.

Summarv

Both the clinical and non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate device.

8. Conclusion

The performance data demonstrates that iX series Patient Monitor are substantially equivalent to the predicate device.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.