K Number
K202892

Validate with FDA (Live)

Date Cleared
2021-01-28

(122 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended to be used for measuring, storing, and to generate prompts for, multiple physiological parameters of adults and pediatrics. The device is intended for use by trained healthcare professionals in hospital environments.

Parameters include: NIBP, SpO2, PR (pulse rate), TEMP.

The F3000 Quick TEMP module is not intended for neonates.

The device is not intended for MRI environments.

Device Description

The iM3s series vital signs monitors including iM3s\iM3As\iM3Bs\ iHM3s are intended to be used for measuring, storing, reviewing of, and generating prompts for multiple physiological parameters of adults, pediatrics and neonates.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study data for the Vital Signs Monitor (iM3s, iM3As, iM3Bs, iHM3s).

Please note: The provided document is an FDA 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. It primarily presents performance specifications and shows that the new device meets those, often by referencing existing standards or previous clearances. It does not typically include detailed experimental setups for clinical or performance studies in the way a research paper would. Therefore, some information requested (like sample size for training sets, specific adjudication methods, MRMC studies, or detailed ground truth establishment for training data) is not present in this type of regulatory document.


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by the predicate device's specifications and compliance with relevant ISO/IEC standards. The "Reported Device Performance" for the subject device is stated as meeting these same specifications.

ParameterAcceptance Criteria (from Predicate/Standards)Reported Device Performance (Subject Device)
SpO2 AccuracyAdult/Pediatric:Adult/Pediatric:
70 to 100%: ±2 %±2% (70% to 100% SpO2)
0-69% unspecifiedUndefined (0% to 69% SpO2)
Neonate:Neonate:
70 to 100%: ±3%±3% (70% to 100% SpO2)
0-69% unspecifiedUndefined (0% to 69% SpO2)
PR from SpO2 Accuracy±2 bpm±2 bpm
TEMP Measurement Range30°C~43°C30°C~43°C
TEMP Prediction Measurement Range35°C~43°C35°C~43°C
NIBP PR Accuracy(Referenced to NIBP module of X series cleared by K192514, which would presumably have its own stated accuracy. The document states NIBP PR accuracy as)±3 bpm or 3.5%, whichever is greater
Electrical SafetyCompliance with ANSI AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009(R) 2012 and A2:2010/(R) 2012Complies (assessed for conformity)
EMCCompliance with IEC 60601-1-2:2014 (Fourth Edition)Complies (assessed for conformity)
NIBP PerformanceCompliance with IEC 80601-2-30:2009+A1:2013Meets accuracy specification and relevant consensus standards
TEMP PerformanceCompliance with ISO 80601-2-56: 2017+A 1:2018Meets accuracy specification and relevant consensus standards
Pulse Oximeter PerformanceCompliance with ISO 80601-2-61: 2017Meets accuracy specification and relevant consensus standards
Software Verification & ValidationAdherence to FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"Documentation provided and testing conducted

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not specified in the document. The performance is validated through "functional and system level testing" and compliance with standards, not a clinical trial with a specified patient sample size for device comparison.
  • Data Provenance: The studies are described as "non-clinical data" and "bench testing." No country of origin is mentioned for patient data, as no clinical studies with human subjects are detailed. The nature of the testing implies it's retrospective, based on existing standards and validation procedures, rather than prospective clinical data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • This information is not provided in the document. The "ground truth" for these performance tests would typically be established by highly calibrated reference instruments or simulated physiological signals according to the requirements of the cited ISO/IEC standards, rather than expert human interpretation in a clinical context.

4. Adjudication Method for the Test Set

  • This information is not applicable and not provided. As no clinical studies with human subjects or interpretation tasks are described, there is no need for an adjudication method. The assessment is against technical specifications and standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC comparative effectiveness study was not done. The document explicitly states "Clinical data: Not applicable." The focus is on the device's standalone performance against engineering and international standards.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, the performance reported is essentially standalone. The "bench testing" and "software verification and validation testing" assess the device's accuracy and functionality in isolation, without an explicit human-in-the-loop component for interpreting outputs in the context of the device's core physiological measurements (NIBP, SpO2, PR, TEMP).

7. The Type of Ground Truth Used

  • The ground truth for the performance testing is based on:
    • Reference measurements from calibrated equipment: For parameters like SpO2, NIBP, and TEMP, the "ground truth" would be simulated physiological signals or measurements from highly accurate, traceable reference devices as mandated by the respective ISO/IEC standards.
    • Compliance with international standards: The device is tested against the specific requirements and accuracy tolerances defined in standards such as IEC 80601-2-30 (NIBP), ISO 80601-2-56 (thermometers), and ISO 80601-2-61 (pulse oximeters).

8. The Sample Size for the Training Set

  • This information is not provided and is not applicable in the context of this device and document. This device is a vital signs monitor, not typically an AI/machine learning diagnostic device that relies on "training sets" in the conventional sense of machine learning algorithms. Its parameters are measured directly through hardware and firmware, validated against physical standards.

9. How the Ground Truth for the Training Set Was Established

  • This information is not provided and is not applicable as there is no "training set" for an AI algorithm described for this device. The device's fundamental operational principles are based on established physiological measurement techniques, not on learning from a dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 28, 2021

Edan Instruments, Inc. Alice Yang Regulatory Engineer #15 Jinhui Road, jinsha Community, Kengzi Sub-District Pingshan District Shenzhen, Guangdong 518122 China

Re: K202892

Trade/Device Name: Vital Signs Monitor:iM3s, iM3As, iHM3s Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, FLL, DQA, DXN Dated: December 25, 2020 Received: December 28, 2020

Dear Alice Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202892

Device Name

Vital Signs Monitor ( iM3s\iM3As\iM3Bs\ iHM3s)

Indications for Use (Describe)

The device is intended to be used for measuring, storing, and to generate prompts for, multiple physiological parameters of adults and pediatrics. The device is intended for use by trained healthcare professionals in hospital environments.

Parameters include: NIBP, SpO2, PR (pulse rate), TEMP.

The F3000 Quick TEMP module is not intended for neonates.

The device is not intended for MRI environments.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K202892

510(K) Summary

Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92

1. Submitter:Edan Instruments, Inc.
#15 Jinhui Road, Jinsha Community,
Kengzi Sub-District, Pingshan District,
Shenzhen, 518122 P.R.China.
Tel: +86(0755) 26858736
Fax: +86(0755) 26882223
Contact person:Alice Yang
Preparing date:September 25, 2020
2. Device name andTrade Name: Vital Signs Monitor, Model: iM3s, iM3As, iM3Bs, iHM3s
classification:Common/Usual Name: Vital Signs Monitor
Classification NameProduct code
21 CFR 870.2300MWI
Monitor, Physiological, Patient (Without ArrhythmiaDetection Or Alarms)
Subsequent Product Code
21 CFR 870.1130DXN
System, Measurement, Blood-Pressure, Non-Invasive
21 CFR 880.2910FLL
Thermometer, Electronic, Clinical
21 CFR 870.2700DQA
Oximeter

3. Predicate Device(s):

    1. Mindray North America, Accutorr 3 Vital Signs Monitor, K132037 (Primary)
    1. Edan Instruments, Inc, Vital signs monitor Model iM3, K180380 (Reference)
    1. Edan Instruments, Inc, Patient Monitor Model X8,X10,X12, K192514 (Reference)
    1. HeTaiDa Technology Co., Ltd, HTD8808C Non-Contact Infrared Body

{4}------------------------------------------------

K202892

Thermometers , K171888 (Reference)

    1. Capsule Technologie, SAS, SmartLinx Vitals Plus Patient Monitoring System, K171751(Reference)
  • The iM3s series vital signs monitors including iM3s\iM3As\iM3Bs\ iHM3s are 4. Device Description: intended to be used for measuring, storing, reviewing of, and generating prompts for multiple physiological parameters of adults, pediatrics and neonates.
    1. Indication for Use The device is intended to be used for measuring, storing, and reviewing of, and to generate prompts for, multiple physiological parameters of adults and pediatrics. The device is intended for use by trained healthcare professionals in hospital environments.

Parameters include: NIBP, SpO2, PR (pulse rate), TEMP.

The F3000 Quick TEMP module is not intended for neonates.

The device is not intended for MRI environments.

6. Predicate Device Comparison

The intended use of the subject devices iM3s series is the same as the primary predicate Mindray Accutor 3 Vital Signs Monitor which is intended for spot-checking monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP) and Temperature (TEMP), on adult, pediatric, and neonatal patients in healthcare facilities by physicians or appropriate medical staff under the direction of physicians.

The table below compares the indication for use and parameter modules of the subject devices to the reference predicate devices.

ItemEDAN Instrument Inc.iM3 Vital Signs MonitorEDAN Instrument Inc.iM3s series Vital Signs MonitorComparisonResult
Manufacturer/K#K180380Current Submission--
DescriptionThe monitor is intended to be used formonitoring, storing, recording, andreviewing of, and to generate alarmsfor, multiple physiological parametersof adults, pediatrics and neonates. Themonitor is intended for use by trainedhealthcare professionals in hospitalThe device is intended to be usedfor measuring, storing, andreviewing of, and to generateprompts for, multiple physiologicalparameters of adults and pediatrics.The device is intended for use bytrained healthcare professionals inSimilar
environments.Monitored parameters include: NIBP,SpO2, PR (pulse rate), QuickTEMP/Infrared TEMP.The monitor is not intended for MRIenvironments.hospital environments.Parameters include: NIBP, SpO2,PR (pulse rate), TEMP.The F3000 Quick TEMP module isnot intended for neonates.The device is not intended for MRIenvironments.
MeasurementParametersSpO2, PR, NIBP, TEMPSpO2, PR, NIBP, TEMPSame
Operation modeMonitoring, Spot-checking, WardRoundSpot-checking, Ward roundDifferent
SpO2 (EDAN Module)
Measurement Range0% to 100%0% to 100%Same
AccuracyAdult/pediatric, non-motion conditions70 to 100%: ±2 %0-69% unspecifiedNeonate70 to 100%: ±3%0-69% unspecifiedAdult /Pediatric:±2% (70% to100% SpO2)Undefined (0% to 69% SpO2)Neonate:±3% (70% to100% SpO2)Undefined (0% to 69% SpO2)Same
PR from SpO2
Measurement range25 to 300 bpm25 to 300 bpmSame
Accuracy±2 bpm±2 bpmSame
TEMP (Covidien F3000 Quick Temp Module)
Measuring range30°C~43°C30°C~43°CSame
Predictionmeasurement range35°C~43°C35°C~43°CSame
Sensor typeOral /axillary /rectalOral /axillary /rectalSame
Measuring ModeDirect Mode /Adjusted ModeDirect Mode /Adjusted ModeSame
Wi-Fi
IEEE802.11b/g/n802.11b/g/nSame
Frequency Band2.4GHz ISM band5GHz & 2.4GHzDifferent
Max. TransmitPower(±2 dBm)17 dBm for 802.11b DSSS17 dBm for 802.11b CCK15 dBm for 802.11g/n OFDM2.4GHz:17dBm (802.11b DSSS)17dBm (802.11b CCK)17dBm (802.11g/n OFDM)5GHz:10dBm (802.11g OFDM)9dBm (802.11n)Different
e-LinkBluetooth Low Energy 4.0Bluetooth Low Energy 4.0Same
NIBP-Edan
PrincipleOscillometry
MeasuringparameterSYS, DIA, MAP, PR
Measurement RangeAdult Mode:SYS: 25 mmHg to 290 mmHgDIA: 10 mmHg to 250 mmHgMAP: 15 mmHg to 260 mmHgPediatric Mode:SYS: 25 mmHg to 240 mmHgDIA: 10 mmHg to 200 mmHgMAP: 15 mmHg to 215 mmHgNeonatal Mode:SYS: 25 mmHg to 140 mmHgDIA: 10 mmHg to 115 mmHgMAP: 15 mmHg to 125 mmHgThe NIBPmodule issimilar toNIBP moduleof X seriescleared byK192514
NIBP PRMeasurement range40 bpm to 240 bpm
NIBP PR Accuracy±3 bpm or 3.5%, whichever isgreater
HTD8808CNon-contact InfraredThermometeriM3s series is compatible with HTD8808C Non-contact Infrared Thermometer cleared byK171888
ExergenTemporalScannerThermometerTAT5000SiM3s series is compatible with Exergen TemporalScanner Thermometer TAT5000S cleared byK011291

{5}------------------------------------------------

{6}------------------------------------------------

As seen in the comparison tables, the subject and predicate devices have similar design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness.

7. Performance Data:

Non-clinical data:

Electrical safety and electromagnetic compatibility (EMC)

iM3s Series Vital Signs Monitor were assessed for conformity with the relevant requirements of the following standards and found to comply:

  • ANSI AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009(R) 2012 and A2:2010/(R) 2012 Medical

{7}------------------------------------------------

K202892

electrical equipment - Part 1: General requirements for basic safety and essential performance.

  • IEC 60601-1-2:2014 (Fourth Edition) Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: electromagnetic disturbances - Requirements and tests.

Performance testing-Bench

Edan has conducted functional and system level testing to validate the performance of the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.

  • IEC 80601-2-30:2009+A1:2013 Medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
  • ISO 80601-2-56: 2017+A 1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
  • ISO 80601-2-61: 2017 Medical electrical equipment - part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

Clinical data: Not applicable.

Summary

The non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate device.

8. Conclusion

The bench testing data and software verification and validation demonstrate that iM3s Series Vital Signs Monitor are substantially equivalent to the predicate devices.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).