K Number
K232336

Validate with FDA (Live)

Device Name
ES Series V2
Manufacturer
Date Cleared
2023-09-26

(53 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound tissue brightness parameters such as ultrasound beam attenuation and backscattering coefficient, and estimates of speed of sound, in internal structures of the body. The device is intended to be used by trained healthcare professionals, in a healthcare environment.

The device is indicated for imaging of anatomical structures in the abdomen and measurements of physical properties in the liver and the spleen.

In particular, the device is intended to provide:

  • Linear distance measurements of anatomical structures,

  • Measurement of shear wave speed at selected shear wave frequencies, and estimates of tissue stiffness in the liver and the spleen.

  • Estimates of ultrasound tissue brightness parameters in the liver at selected ultrasound frequencies,

  • Measurement of brightness ratio between structures and in particular between the liver and the kidnev.

  • Estimates of speed of sound in the liver.

The shear wave speed measurements, ultrasound tissue brightness parameters, elastic properties estimates, brightness ratio may be used as an aid to the diagnosis, monitoring and clinical management of adult and pediatric patients with liver disease.

Device Description

E-Scopics' ES Series V2 is an ultraportable ultrasound imaging system used to perform non-invasive diagnostic general purpose ultrasound imaging and quantitative imaging studies. ES Series V2 consists of a Software App (running on a consumer off-the-shelf Selected Host) and an accessory curved array probe. The system produces images and quantifications, which are displayed on the monitor of the Selected Host. ES Series V2 is operated from the Selected Host multi-touch screen. ES Series V2 also allows the user to perform measurements, to capture images, and to generate printable reports.

ES Series V2 is designed to perform non-invasive measurements of liver/spleen shear wave speed and to estimate tissue stiffness. The e.C5-1 probe is equipped with a mechanical vibrator to produce shear waves in tissue. When used on the spleen applications, shear waves travel through the skin and intercostal space into the abdominal organ. 2D ultrasound is used to track the shear wave and measure its speed. The system then provides an estimate of tissue stiffness. In addition, ES Series V2 is designed to measure several quantitative parameters from B Mode imaging: ultrasound tissue brightness parameters such as ultrasound attenuation and backscattering coefficient, speed of sound, and compute ultrasound brightness ratio between 2 regions in the image. The results of ES Series V2 quantitative imaging modalities are displayed on the selected host monitor.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the ES Series V2 device, based on the provided text:

Acceptance Criteria and Device Performance

The device's performance was evaluated against precision (within subject coefficient of variation, wCV) and bias (difference from ground truth) on calibrated phantoms. The stated acceptance criteria are reflected in the reported performance.

ParameterAcceptance Criteria (Predicate Device)Reported Device Performance (ES Series V2) - wCV (Precision)Reported Device Performance (ES Series V2) - Bias
B-Mode ImagingAxial Resolution: < 1 mmLateral Resolution: < 2mm
Contrast Resolution: distinguishes targets with +/- 3 dB contrast ratio
Speed of Sound (SOS)Bias: (-0.7%) - (0.7%)wCV: [0.1%, 0.4%] (CI 95% [0.1%, 2.5%])Bias: [-3.2%, -1.2%] (CI 95% [-4.0%, -0.2%])
Precision: (0.05%) - (0.3%)
Shear Wave Speed (SWS) / Liver Stiffness (2D TE)Bias (kPa): (-12.5%) - (28.5%)wCV (SWS): [0.8%, 3.2%] (CI 95% [0.4%, 20%])Bias (SWS): [-15%, 2.2%] (CI 95% [-20%, 9%])
Bias (m/s): (-6.4%) - (13.2%)wCV (Young's modulus/stiffness): [1.2%, 6.3%] (CI 95% [0.6%, 40%])Bias (Young's modulus/stiffness): [-27%, 5.6%] (CI 95% [-35%, 18%])
Precision (kPa): (0.8%) - (2.7%)
Precision (m/s): (0.4%) - (1.4%)
Elasticity ImagingwCV (SWS): [0.1 %, 0.9 %] (CI 95% [0.1%, 5.3%])Bias (SWS): [3.8 %, 11.2 %] (CI 95% [3.0%, 13.6%])
wCV (Young's moduli): [0.3 %, 1.7 %] (CI 95% [0.1%, 10.6%])Bias (Young's moduli): [7.5 %, 23.7 %] (CI 95% [5.9%, 28.9%])
Average axial resolution: 4.3 mmAverage lateral resolution: 4.0 mm
Ultrasound Attenuation Coefficient (ATT)Bias (dB/m @ 3.5 MHz): (-11.6%) - (1.4%)wCV: [1.2%, 2.0%] (CI 95% [0.6%, 12.6%])Bias: [-26%, 6.1%] (CI 95% [-28%, 16%])
Precision (dB/m @ 3.5 MHz): (1.5%) - (4.5%)
Ultrasound Backscatter Coefficient (BSC)(Predicate Device used Hepatpreanal Index (HRI), not direct BSC)wCV: [0.8%, 1.7%] (CI 95% [0.3%, 10.8%])Bias: [-13%, -1.1%] (CI 95% [-18%, 0.8%])

Study Details

  1. Sample size used for the test set and the data provenance:

    • The test set consisted of a "sample of calibrated phantoms." The exact number of phantoms is not specified.
    • The data provenance is in-vitro performance assessment. The document does not specify the country of origin for the phantoms or if the data was retrospective or prospective, as it's an in-vitro study.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for this study. The ground truth for the phantoms was established by their inherent calibrated properties, not by expert consensus.
  3. Adjudication method for the test set:

    • Not applicable for this study, as the ground truth was based on calibrated phantoms.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical testing is necessary to demonstrate substantial equivalence between ES Series V2 and its Predicate Device." This study focused on in-vitro technical performance and comparison to a predicate device.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance testing described is a standalone (algorithm only) assessment. The device's quantitative modes (Speed of Sound, Shear Wave Speed/Stiffness, Ultrasound Attenuation, Ultrasound Backscatter Coefficient) were evaluated algorithmically against known ground truth values from calibrated phantoms.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth used was calibrated phantoms with known reference values.
  7. The sample size for the training set:

    • The document does not explicitly state the sample size for the training set. However, given that this is a 510(k) submission primarily focused on demonstrating substantial equivalence to a predicate device through performance testing, the details of the training set used for the underlying algorithms are not typically provided in this summary. The focus is on the performance of the final, developed algorithms.
  8. How the ground truth for the training set was established:

    • The document does not explicitly describe how the ground truth for the training set was established. This information is typically proprietary to the developer's internal development process and not required for a 510(k) summary demonstrating equivalence via performance testing.

{0}------------------------------------------------

September 26, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

E-Scopics % Carine Kessali Quality Assurance and Regulatory Affairs Manager 931 Chemin de la Bosque d'Antonelle Aix en Provence, 13090 FRANCE

Re: K232336

Trade/Device Name: ES Series V2 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: July 31, 2023 Received: September 18, 2023

Dear Carine Kessali:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Marjan Nabili -S for

Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K232336

Device Name

ES Series V2

Indications for Use (Describe)

The device is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound tissue brightness parameters such as ultrasound beam attenuation and backscattering coefficient, and estimates of speed of sound, in internal structures of the body. The device is intended to be used by trained healthcare professionals, in a healthcare environment.

The device is indicated for imaging of anatomical structures in the abdomen and measurements of physical properties in the liver and the spleen.

In particular, the device is intended to provide:

  • Linear distance measurements of anatomical structures,

  • Measurement of shear wave speed at selected shear wave frequencies, and estimates of tissue stiffness in the liver and the spleen.

  • Estimates of ultrasound tissue brightness parameters in the liver at selected ultrasound frequencies,

  • Measurement of brightness ratio between structures and in particular between the liver and the kidnev.

  • Estimates of speed of sound in the liver.

The shear wave speed measurements, ultrasound tissue brightness parameters, elastic properties estimates, brightness ratio may be used as an aid to the diagnosis, monitoring and clinical management of adult and pediatric patients with liver disease.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for "escopics" with the tagline "ULTRASOUND AGILITY" underneath. The logo features the letters "e.sco" in yellow and blue, with the "o" made up of small blue and yellow dots. Below the logo is the text "DOC-1895 ver2".

K232336

Premarket Notification [510(k)] Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements detailed in 21 CFR 807.92

The assigned 510(k) number is: K232336

  • Company: E-Scopics 931 chemin de la Bosque d'Antonelle 13090 Aix-en-Provence France
    Telephone:

Contact Person: Claude COHEN- BACRIE

Date Prepared: 20th July 2023

Device Names:

Trade/Proprietary Name:ES Series V2
Common or Usual Name:Diagnostic Ultrasound System and Accessories
Device ClassClass II
Classifications:§892.1550 : Ultrasonic pulsed doppler imaging system (IYN)§892.1560 : Ultrasonic Pulsed Echo Imaging System (IYO)§892.1570 : Diagnostic Ultrasound Transducer (ITX)

Substantially Equivalent/Predicate Devices:

The data and information supplied in this submission demonstrates substantial equivalence to the predicate device: Predicate Device: ES1 System (K213102), cleared on 11th of January 2022.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for "e.scopics ULTRASOUND AGILITY". The "e" is yellow, and the rest of the word "scopics" is dark blue. The "o" in scopics is made up of small blue squares that fade into yellow squares on the right side. Below the word "e.scopics" is the phrase "ULTRASOUND AGILITY".

Device description:

E-Scopics' ES Series V2 is an ultraportable ultrasound imaging system used to perform non-invasive diagnostic general purpose ultrasound imaging and quantitative imaging studies. ES Series V2 consists of a Software App (running on a consumer off-the-shelf Selected Host) and an accessory curved array probe. The system produces images and quantifications, which are displayed on the monitor of the Selected Host. ES Series V2 is operated from the Selected Host multi-touch screen. ES Series V2 also allows the user to perform measurements, to capture images, and to generate printable reports.

ES Series V2 is designed to perform non-invasive measurements of liver/spleen shear wave speed and to estimate tissue stiffness. The e.C5-1 probe is equipped with a mechanical vibrator to produce shear waves in tissue. When used on the spleen applications, shear waves travel through the skin and intercostal space into the abdominal organ. 2D ultrasound is used to track the shear wave and measure its speed. The system then provides an estimate of tissue stiffness. In addition, ES Series V2 is designed to measure several quantitative parameters from B Mode imaging: ultrasound tissue brightness parameters such as ultrasound attenuation and backscattering coefficient, speed of sound, and compute ultrasound brightness ratio between 2 regions in the image. The results of ES Series V2 quantitative imaging modalities are displayed on the selected host monitor.

Intended Use & Indications for Use:

The device is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound tissue brightness parameters such as ultrasound beam attenuation and backscattering coefficient, and estimates of speed of sound, in internal structures of the body. The device is intended to be used by trained healthcare professionals, in a healthcare environment.

The device is indicated for imaging of anatomical structures in the abdomen and measurements of physical properties in the liver and the spleen.

In particular, the device is intended to provide:

  • Linear distance measurements of anatomical structures,

  • Measurement of shear wave speed at selected shear wave frequencies, and estimates of tissue stiffness in the liver and the spleen,

  • Estimates of ultrasound tissue brightness parameters in the liver at selected ultrasound frequencies.

  • Measurement of brightness ratio between structures and in particular between the liver and the kidney.

  • Estimates of speed of sound in the liver.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Escopics. The logo features the word "e.scopics" in a stylized font, with the "e" in yellow and the rest of the letters in blue. The "o" in "scopics" is made up of small blue squares, with some of the squares fading away to the right. Below the word "e.scopics" is the phrase "ULTRASOUND AGILITY" in a smaller font.

The shear wave speed measurements, ultrasound tissue brightness parameters, elastic properties estimates, brightness ratio may be used as an aid to the diagnosis, monitoring and clinical management of adult and pediatric patients with liver disease.

Substantial Equivalence discussion

ES Series V2 operating modalities, principles, and features (B-Mode, 2D-Transient Elastography, estimation of shear wave speed, ultrasound attenuation, backscattering coefficient & speed of sound, and associated combinations as listed in the Diagnostic Ultrasound Indications for Use Table) have been demonstrated to be substantially equivalent to operating modalities, principles, and features of the Predicate Device.

Shear wave speed measurement and tissue stiffness estimation as available on ES Series V2 are equivalent to those available on the Predicate Device, as demonstrated through performance testing , and they serve the same intended use, without raising any safety & effectiveness issues.

ES Series V2 provides estimates of ultrasound tissue brightness, such as ultrasound attenuation and backscattering coefficient in tissue and is substantially equivalent to the Predicate Device as demonstrated by performance testing (verification and validation testing, in vitro performance assessment). The differences in ultrasound sequences and data processing algorithms between methods implemented on ES Series V2 and the Predicate Device do not raise issues of safety and effectiveness.

ES Series V2 provides estimates of speed of sound in tissue, and is substantially equivalent to the Predicate Device as demonstrated by performance testing. The differences in ultrasound sequences and data processing algorithms between methods implemented on ES Series V2 and the Predicate Device do not raise issues of safety and effectiveness.

ES Series V2 operates the same curvilinear array as the Predicate Device to perform all ultrasound modes, including conventional B-Mode. Although ES Series V2 B-Mode imaging has a different post-processing filtering algorithm, ultrasound transmit/receive sequences are identical to the ones implemented in the Predicate Device, without raising issues of safety and effectiveness.

Application of Recognized Standards

ES Series V2 operates the same curvilinear array as the Predicate Device (e.C5-1 Probe), which was designed & developed to ensure compliance with the requirements of standards covering biocompatibility, acoustic output, cleaning & disinfection, thermal, mechanical, electrical & electromagnetic safety performance.

Changes in ES Series V2 as compared to the Predicate Device only reside in ultrasonic sequences updates under track 1, ultrasonic data processing algorithms and

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image is the logo for Escopics. The logo has the word "escopics" in blue, with the "e" in yellow. Underneath the word is the phrase "ULTRASOUND AGILITY" in a smaller font. The "o" in escopics is made up of small blue squares that fade out to the right.

software parts. These limited changes have been designed & developed to ensure compliance with the requirements of standards covering acoustic output and software safety performance:

  • IEC 60601-2-37 Ed2.1 2015 : Medical electrical equipment Part 2-37: Particular ● requirements for the safety of ultrasonic medical diagnostic and monitoring equipment
  • IEC 62304: Medical Device Software - Software Life Cycle Processes; First Edition 2006- 05, Equivalent to IEC 62304: 2006/A1:2016.
  • ISO 14971 Second: Medical Devices Application of Risk Management To . Medical Devices; Edition 2007-03-01.
  • AAMI TIR57:2016 Principles for medical device security Risk management. ●

Discussion of Performance Data:

The accuracy and precision of the device ultrasound modalities were tested on calibrated phantoms with known reference values.

For B-Mode imaging, the in-vitro performances were good. Contrast resolution enables to distinguish targets having a contrast ratio of +/- 3 dB as compared to the phantom background. The axial resolution of the system is below 1 mm, and the lateral resolution of the system is below 2mm.

The accuracy of quantitative ultrasound modes of ES Series V2 was assessed in-vitro by looking at the precision and the bias obtained on a sample of calibrated phantoms. Precision (in %), also known as the within subject coefficient of variation (wCV), was defined as the standard deviation of the independent measurements normalized by the mean of the independent measurements. Bias (in %) was defined as the difference between the mean value of the independent measurements and the ground truth value of the phantom, normalized by the ground truth value of the phantom.

In addition, the accuracy of shear wave speed/stiffness, speed of sound measurements, ultrasound attenuation measurements and ultrasound basckscatter coefficient measurements of ES Series V2 and the counterpart in its Predicate Device were compared looking at the precision and the bias obtained with both systems on the same sample of calibrated speed of sound, shear wave speed, ultrasound attenuation and ultrasound backscatter coefficient phantoms.

For the estimation of the Speed of Sound, the in-vitro performance tests revealed a strong inter-system repeatability and demonstrated that the algorithm is accurate and robust on a wide range of speed of sound values. Indeed, the wCV lies in the range [0.1%, 0.4%] (CI 95% [0.1%, 2.5%]) while bias values range in [-3.2%, -1.2%] (CI 95% [-4.0%, -0.2%]). When comparing measurements performed with ES Series V2 and

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for e.scopics, a company specializing in ultrasound agility. The logo features the company name in a stylized font, with the "e" in yellow and the rest of the letters in blue. Below the logo is the text "DOC-1895 ver2".

its Predicate Device, a perfect coefficient of determination of 1.000 was obtained between the 2 systems and methods.

For shear wave speed measurements and liver stiffness estimation by the 2D TE method, in-vitro experiments reveal that the stiffness estimation is robust on a wide range of stiffness values and satisfies QIBA technical performance claims. Indeed, the wCV lies in the range [0.8%, 3.2%] (CI 95% [0.4%, 20%]) for the shear wave speed and in the range [1.2%. 6.3%] (CI 95% [0.6%, 40%]) for the corresponding Young's modulus. Bias values lie in the range [-15%, 2.2%] (CI 95% [-20%, 9%]) for the SWS and in the range [-27%, 5.6%] (CI 95% [-35%. 18%]) for Young's modulus. It also reveals a strong inter-system repeatability.

When comparing measurements performed with ES Series V2 and its Predicate Device, a high coefficient of determination of 0.99 was obtained between the 2 systems and methods.

For elasticity imaging, the in-vitro performances were good. Average axial and lateral resolution have been measured at 4.3 and 4.0 mm. In addition, the in-vitro performance tests demonstrated that the measurements of the shear wave speed and stiffness within inclusions are accurate and robust on a wide range of values. Indeed, the wCV lies in the range [0.1 %, 0.9 %] (CI 95% [0.1%, 5.3%]) for the shear wave speed and in the range [0.3 %. 1.7 %] (CI 95% [0.1%. 10.6%]) for the corresponding Young moduli while bias values range in [3.8 %, 11.2 %] (CI 95% [3.0%, 13.6%]) for the shear wave speed and in the range [7.5 %, 23.7 %] (CI 95% [5.9%, 28.9%]) for the corresponding Young moduli. When comparing measurements performed with ES Series V2 and its Predicate Device, a high coefficient of determination of 0.99 was obtained between the 2 systems and methods.

For measurement of the ultrasound attenuation coefficient, the in-vitro performance tests revealed a strong inter-system repeatability and demonstrated that the algorithm is accurate and robust on a wide range of attenuation values. Indeed the wCV lies in the range [1.2%, 2.0%] (CI 95% [0.6%, 12.6%]), while bias values range in [-26%, 6.1%] (CI 95% [-28%. 16%]). On the same set of calibrated phantoms, ultrasound attenuation estimations with ES Series V2 were highly correlated with estimations performed with its Predicate Device, with a coefficient of determination correlation of 0.987.

For measurement of the ultrasound backscatter coefficient, the in-vitro performance tests revealed a strong inter-system repeatability and demonstrated that the algorithm is accurate and robust on a wide range of backscatter coefficient values. Indeed the wCV lies in the range [0.8%, 1.7%] (CI 95% [0.3%, 10.8%]), while bias values range in [-13%, -1.1%] (CI 95% [-18%, 0.8%]). On the same set of calibrated phantoms, ultrasound backscatter coefficient estimations with ES Series V2 were highly correlated

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image is a logo for Escopics. The logo features the word "escopics" in blue letters, with the "e" in yellow. Below the word "escopics" is the phrase "ULTRASOUND AGILITY" in smaller blue letters. The "o" in "escopics" is a circle made up of small blue squares that fade out on the right side.

with hepatorenal index estimations performed with its Predicate Device, with a coefficient of determination correlation of 0.993.

Acoustic Pressure and Intensities of changed ultrasound modalities available to the user on ES Series V2 were measured: ATT and SOS. 2 transmit patterns (TP) were measured, each of them being used respectively in the three modified composed imaging modes available on the system. These test results show that the upper-limit values for MI, TI and ISPTA3 are always below the limits recommended in the FDA Guidance for applications other than for ophthalmic use, for ultrasound systems, therefore legitimating the Track 1 approach.

Conclusions for Performance Testing:

The performance testing data conclude that the safety and effectiveness of ES Series V2 is not compromised and meet all acceptance criteria, demonstrating that the system is substantially equivalent to its Predicate Device.

Conclusions for Clinical Testing:

No clinical testing is necessary to demonstrate substantial equivalence between ES Series V2 and its Predicate Device.

Conclusion regarding Substantial Equivalence:

In conclusion, ES Series V2 was demonstrated to be substantially equivalent to its Predicate Device K213102, as they have the same intended use, equivalent indications for use, and the technological differences do not raise issues of safety and effectiveness.

This has been demonstrated via safety & performance testing between ES Series V2 and its Predicate Device.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Escopics. The logo features the word "escopics" in a stylized font, with the "e" in yellow, the "s", "c", "p", "i", and "s" in dark blue, and the "o" made up of small blue squares. Below the word "escopics" is the phrase "ULTRASOUND AGILITY" in a smaller font.

Key pointsPredicate Device (K213102)New Device
Device NameES1 SystemES Series V2
510(K) ownerE-Scopics S.A.S., FranceE-Scopics S.A.S., France
Manufactured byE-Scopics S.A.S., FranceE-Scopics S.A.S., France
Classification NameUltrasonic pulsed doppler imaging systemUltrasonic pulsed doppler imaging system
ClassIIII
Product Code§892.1550 : Ultrasonic pulsed doppler imaging system (IYN)§892.1560 : Ultrasonic Pulsed Echo Imaging System (IYO)§892.1570 : Diagnostic Ultrasound Transducer (ITX)§892.1550 : Ultrasonic pulsed doppler imaging system (IYN)§892.1560 : Ultrasonic Pulsed Echo Imaging System (IYO)§892.1570 : Diagnostic Ultrasound Transducer (ITX)
Common or usualnameDiagnostic Ultrasound System and AccessoriesDiagnostic Ultrasound System and Accessories
Intended useDiagnostic ultrasound imaging, soft tissue elasticity imaging, fluidflow analysis of the human body.Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluidflow analysis of the human body.
Indications for useThe ES1 Ultrasound Diagnostic System is intended for generalpurpose pulse echo ultrasound imaging, soft tissue elasticityimaging of the human body and provides measurements of shearwave speed and tissue stiffness, ultrasound beam attenuation andestimates of speed of sound, in internal structures of the body.The ES1 Ultrasound Diagnostic System is indicated for imagingand measurements of anatomical structures in the abdomen.In particular, the ES1 ultrasound diagnostic system is intended toprovide:The device is intended for general purpose pulse echo ultrasoundimaging, soft tissue elasticity imaging of the human body andprovides measurements of shear wave speed and tissue stiffness,ultrasound tissue brightness parameters such as ultrasound beamattenuation and backscattering coefficient, and estimates of speedof sound, in internal structures of the body. The device is intendedto be used by trained healthcare professionals, in a healthcareenvironment.The device is indicated for imaging of anatomical structures in theabdomen and measurements of physical properties in the liver andthe spleen.In particular, the device is intended to provide:
Key pointsPredicate Device (K213102)New Device
- Linear distance measurements of anatomical structures,- Measurement of shear wave speed at selected shear wavefrequencies, and estimates of tissue stiffness in the liverand the spleen,- Estimates of ultrasound attenuation coefficient in the liverat 3.5 MHz,- Measurement of brightness ratio between structures andin particular between the liver and the kidney,- Estimates of speed of sound in the liverThe shear wave speed measurements, ultrasound beam attenuation,elastic properties estimates, brightness ratio may be used as an aidto the diagnosis, monitoring and clinical management of adult andpediatric patients with liver disease.- Linear distance measurements of anatomical structures,- Measurement of shear wave speed at selected shear wavefrequencies, and estimates of tissue stiffness in the liverand the spleen,- Estimates of ultrasound tissue brightness parameters in theliver at selected ultrasound frequencies,- Measurement of brightness ratio between structures and inparticular between the liver and the kidney,- Estimates of speed of sound in the liver.The shear wave speed measurements, ultrasound tissue brightnessparameters, elastic properties estimates, brightness ratio may beused as an aid to the diagnosis, monitoring and clinicalmanagement of adult and pediatric patients with liver disease.
Intended UsersTrained healthcare professionalsTrained healthcare professionals
Imaging &quantificationmodesB-modeTransient Elastography2D Transient Elastography and Shear Wave ImagingAttenuation measurementSpeed of sound measurementB-modeTransient Elastography2D Transient Elastography and Shear Wave ImagingAttenuation measurementBackscattering coefficient measurementSpeed of sound measurement
B-ModeB-ModeB-Mode(using equivalent technological characteristics as the PredicateDevice as presented in DOC-1797; Theory of Operations - V2 - B-Mode)
UltrasoundattenuationName: ATTProbe: e.C5-1 probeName: ATTProbe: e.C5-1 probe
Key pointsPredicate Device (K213102)New Device
Technological characteristics: Diverging wave ultrasoundtransmission and receptionWide bandwidth attenuation slope estimation AND Estimation ofattenuation at 3,5MhzRange:100 to 350 dB/m0.2 dB/cm/MHz to 1 dB/cm/MHzTechnological characteristics: Diverging wave ultrasoundtransmission and receptionEstimation of attenuation at 3.5 MHzRange:100 to 350 dB/m @ 3.5 MHz
Display:Median, interquartile range (IQR) and IQR/Median ratio (%) (Seedetails in the User Manual)Display:Median, interquartile range (IQR) and IQR/Median ratio (%) (Seedetails in the User Manual)
Bias:(dB/m @ 3.5 MHz): (-11.6%) - (1.4%)(dB/cm/MHz): (-11.6%) - (1.4%)Bias:(dB/m @ 3.5 MHz): (-26%) - (6.1%)(95% CI): (-28%) - (16%)
Precision:(dB/m @ 3.5 MHz): (1.5%) - (4.5%)(dB/cm/MHz): (1.5%) - (4.5%)Precision:(dB/m @ 3.5 MHz): (1.2%) - (2.0%)(95% CI): (0.6%) - (12.6%)
Speed of soundName: SOSName: SOS
Probe: e.C5-1 probeProbe: e.C5-1 probe
Technological Characteristics: Diverging wave ultrasoundtransmission and receptionWide bandwidth frequency estimationTechnological Characteristics: Plane wave ultrasoundtransmission and receptionSpeed of sound estimation
Range:1400 m/s to 1650 m/sRange:1400 m/s to 1650 m/s
Display:Median, interquartile range (IQR) and IQR/Median ratio (%) (Seedetails in the User Manual)Display:Median, interquartile range (IQR) and IQR/Median ratio (%) (Seedetails in the User Manual)
Key pointsPredicate Device (K213102)New Device
Bias:(m/s): (-0.7%) - (0.7%)Bias:(m/s): (-3.2%) - (-1.2%)(95% CI): (-4.0%) - (-0.2%)
Precision:(m/s): (0.05%) - (0.3%)Precision:(m/s): (0.1%) - (0.4%)(95% CI): (0.1%) - (2.5%)
Other ModesElastography ModeName: Transient Elastography (2D-TE)Name: Transient Elastography (2D-TE)
Type: 2D, quantification with imaging capabilitiesType: 2D, quantification with imaging capabilities
Shear wave source: External electromechanical vibrator creating anon-punctual source of vibration, spread along the probe surfaceShear wave source: External electromechanical vibrator creating anon-punctual source of vibration, spread along the probe surface
Ultrasound tracking: Plane wave transmission and receptionUltrasound tracking: Diverging wave transmission and reception
Range:Liver: shear wave speed (0.0-5.8); Stiffness values (0.0-100 kPa)Elastography range displayed on the screen (next to color bar) anddepends on vibration frequencyRange:Liver: shear wave speed (0.0-5.8); Stiffness values (0.0-100 kPa)Elastography range displayed on the screen (next to color bar) anddepends on vibration frequency
Display:Median value; Interquartile range (IQR); IQR/Median ratio (Seedetails in the User Manual)Display:Median value; Interquartile range (IQR); IQR/Median ratio (Seedetails in the User Manual)
Bias:(kPa): (-12.5%) - (28.5%)(m/s): (-6.4%) - (13.2%)Bias:(kPa): (-27%) - (5.6%)(95% CI): (-35%) - (18%)(m/s): (-15%) - (2.2%)(95% CI): (-20%) - (9.0%)
Key pointsPredicate Device (K213102)New Device
Precision:(kPa): (0.8%) - (2.7%)(m/s): (0.4%) - (1.4%)Precision:(kPa): (1.2%) - (6.3%)(95% CI): (0.6%) - (40%)(m/s): (0.8%) - (3.2%)(95% CI): (0.4%) - (20%)
BackscatteringcoefficientThe predicate device ES1 System (K213102) does not providequantitative BSC measurements but semi-quantitativemeasurements by means of the hepatorenal index (HRI) (see DOC-1793 Theory of Operations - V2 - Appendices).The backscatter coefficient estimation algorithm as available in theNew Device provides a quantitative way to estimate correctedbrightness which is similar to HRI from the Predicate Device.Technological characteristics of the two algorithms implementedin the two devices have a high degree of similarity as:- Both devices rely on the same energy-source: pulse-echoUS imaging;- They rely on the same electronic boards and the sameacoustic head to transmit / receive US data;- The transmit strategy (i.e. 12 Archimedean spiral waves),the transmit waveform are highly similar;Both techniques estimate average brightness ratios between aregion of interest i.e the region labeled "L" in the HRI and theregion of interest surrounded by yellow borders in the BSCestimation algorithm, and a reference region i.e. the region labeled"K" in the HRI and the region surrounded by yellow borderswithin the reference phantom for the BSC estimation algorithm.Name: BSCProbe: e.C5-1 probeTechnological Characteristics: Diverging wave ultrasoundtransmission and receptionEstimation of backscatter coefficient in dB/cm-sr at 3 MHzRange:-40 dB/cm-sr @ 3 MHz to -20 dB/cm-sr @ 3MHzDisplay:Median, interquartile range (IQR) and IQR/Median ratio (%)Bias:(dB/cm-sr @ 3 MHz): (-13%) - (-1.1%)(95% CI): (-18%) - (0.8%)Precision:(dB/cm-sr @ 3 MHz): (0.8%) - (1.7%)(95% CI): (0.3 %) - (10.8 %)
Key pointsPredicate Device (K213102)New Device
Combined ModesCombined modes include:- B + 2D-TE or B + SWS- B + ATT- B + SOS- B + Q.US (ATT + SOS)Combined modes include:- B + 2D-TE or B + SWS- B + TB (where TB is ATT + BSC)- B + SOS- B + Q.US (ATT + BSC + SOS)
System Components- Software- USB digital hardwired probe- Consumer off-the-shelf (COTS) Selected Host computer (notprovided)- Software- USB digital hardwired probe- Consumer off-the-shelf (COTS) Selected Host computer (notprovided)
Form factor (Cartbased / portablemobile )Portable product with COTS selected hostPortable product with COTS selected host
Acoustic OutputLevels (Track 1 & 3)Track 1Track 1
ControlsImage DepthImage GainMode SelectFreezeImage DepthImage GainMode SelectFreeze
Power SourceProbe powered via COTS selected hostProbe powered via COTS selected host
ScreenCOTS selected host display screenCOTS selected host display screen
Software operatingsystemCOTS selected host equipped with Windows OSCOTS selected host equipped with Windows OS
TransducersUltrasound sourcePiezoelectric ultrasound sourcePiezoelectric ultrasound source
TransducersProbe typeMulti-elements ultrasound transducer:Convex array: e.C5-1Multi-elements ultrasound transducer:Convex array: e.C5-1
Key pointsPredicate Device (K213102)New Device
Cleaning /Disinfectingmethods &materialsMethods:Soft cloth wipingSoaking prohibitedMethods:Soft cloth wipingSoaking prohibited
Materials:Isopropyl alcohol / AlkylamineQuaternary ammoniumChlorine dioxideMaterials:Isopropyl alcohol / AlkylamineQuaternary ammoniumChlorine dioxide
ReportYes, available in pdfYes, available in pdf
Target PatientPopulation (ifapplicable)General purpose imaging systemGeneral purpose imaging system
Patient Populationexclusion (ifapplicable)- Patients with active implants such as pacemakers, defibrillators,pumps, etc.- Wounds- Patients with active implants such as pacemakers, defibrillators,pumps, etc.- Wounds

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image is a logo for Escopics Ultrasound Agility. The logo features the word "escopics" in a stylized font, with the "e" in yellow and the rest of the letters in blue. The "o" in "escopics" is made up of small blue squares that fade out towards the right. Below the word "escopics" is the phrase "ULTRASOUND AGILITY" in a smaller font.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image is a logo for a company called "escopics". The "e" is yellow, and the rest of the letters are dark blue. The "o" in "escopics" is made up of small blue squares. Below the company name is the phrase "ULTRASOUND AGILITY" in a smaller font.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for eScopics. The logo features the word "eScopics" in a stylized font, with the "e" in yellow and the rest of the letters in blue. The "o" in "Scopics" is made up of a grid of blue squares that fade out to the right. Below the word "eScopics" is the phrase "ULTRASOUND AGILITY" in a smaller font.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for "escopics". The "e" is yellow, and the rest of the letters are blue. The "o" is made up of small blue squares that fade out to the right. Underneath the logo is the text "ULTRASOUND AGILITY".

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image is a logo for eScopics. The logo features the word "eScopics" in a stylized font, with the "e" in yellow and the rest of the letters in blue. The "o" in "Scopics" is made up of small blue squares that fade out to the right. Below the word "eScopics" is the phrase "ULTRASOUND AGILITY" in a smaller font.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for eScopics. The logo features the word "eScopics" in a stylized font, with the "e" in yellow and the rest of the letters in blue. The "o" in "Scopics" is made up of a grid of blue squares that fade out to the right. Below the word "eScopics" is the phrase "ULTRASOUND AGILITY" in a smaller font.

N/A