K Number
K232946

Validate with FDA (Live)

Device Name
Yumizen H2500
Manufacturer
Date Cleared
2024-05-21

(244 days)

Product Code
Regulation Number
864.5220
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Yumizen H2500 is a quantitative multiparameter fully automated hematology analyzer intended for in-vitro diagnostic use in clinical laboratories by qualified healthcare professionals for the screening of patient populations.

The Yumizen H2500 is intended to perform tests on the following specimens:

  • Anticoagulated whole blood specimens ●
  • Body fluids (synovial fluids, serous fluids and cerebrospinal fluids). .

The Yumizen H2500 classifies and enumerates the following parameters:

  • A complete blood count (CBC) consisting of TNC, WBC, RBC, HGB, calculated . HCT, MCV, calculated MCH, calculated MCHC, RDW-SD, RDW-CV, PLT, PLT-Ox, LPF, MPV.
  • A leukocyte differential count consisting of LYM (%#), MON (%#), NEU (%/#), ● EOS (%/#), BAS (%/#), IMG (%/#)
  • A nucleated red blood cell count consisting of NRBC (%/#). ●
  • A reticulocyte analysis consisting of RET (%/#), calculated CRC, IRF, RHCC. ●
  • Quantitative determination of blood cells in synovial fluids, serous fluids and . cerebrospinal fluids consisting of BFWBC, BFRBC, BFPN (%/#), BFMN (%/#).

Note: Venous and capillary whole blood should be collected in K2EDTA anticoagulant. Serous and synovial fluids should be collected without anticoagulant or in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with cerebrospinal fluid specimens is neither required nor recommended. Alternatively, Sodium Heparin or Lithium Heparin may be used for synovial fluid.

Device Description

The HORIBA Medical analyzer modules Yumizen H2500 are multi-parameter hematology analyzers intended to perform tests on whole blood samples collected in K2EDTA and body fluids (synovial and serous) collected in K2EDTA anticoaqulant. The analyzers can also perform tests on cerebrospinal fluids which should not be collected in any anticoaqulant.

The Analyzer Units (Yumizen H2500) aspirate, dilute, mix, and analyze blood and body fluid samples.

The Yumizen H2500 model provides Complete Blood Count (CBC), Differential (DIFF), Reticulocyte counts (RET) and Optical Platelet counts as well as Body Fluid counts (BF).

The analyzer models may function with:

  • · a Data Management Unit (Yumizen P8000) which is the interface with the laboratory connections (LIS) and the Analyzer Unit(s). Through its interface, the Yumizen P8000 enables the user to monitor the workflow of patient data, centralize result data, perform reflex testing, customize rules, centralize the validation operations, run quality control, manage quality assurance on results.
AI/ML Overview

1. A table of acceptance criteria and the reported device performance

Since specific acceptance criteria values were not explicitly stated for all performance aspects, I will infer them as generally "met predefined acceptance criteria" or "demonstrated comparable performance" where stated in the document.

Test CategoryParameterAcceptance Criteria (Inferred)Reported Device Performance
Analytical Performance
Repeatability (Whole Blood)All parametersAll components of variation met predefined acceptance criteria.Max %CV and Max SD values were reported for various parameters across normal, low, and high target ranges (Tables 4 and 5) and were found to meet the predefined acceptance criteria. For example, normal WBC has Max %CV of 2% and Max SD of 0.150, while high WBC (10-30 10^9/L) has Max %CV of 1.9% and Max SD of 0.330.
Repeatability (Body Fluid)All parametersAll components of variation met predefined acceptance criteria.Max %CV and Max SD values were reported for various parameters (BFWBC, BFRBC, BFPN#, BFPN%, BFMN#, BFMN%) across different levels and fluid types (serous, synovial, CSF) (Tables 6, 7, 8) and were found to meet the predefined acceptance criteria. For example, BFWBC (Level 1) for serous fluids has Max %CV of 10.4% and Max SD of 8.5.
Reproducibility (Whole Blood)All parametersMet acceptance criteria per CLSI EP05-A3.Detailed SD and CV% reported for within-run, between-run, between-day, and between-site variations for whole blood control materials (ABX Difftrol - Table 9, ABX Minotrol Retic - implicit in text, though table only provided for Difftrol). All results met the acceptance criteria. For example, WBC (Low) total CV% was 3.62%; HGB (Normal) total CV% was 0.89%.
Reproducibility (Body Fluids)All parametersMet acceptance criteria per CLSI EP05-A3.Detailed SD and CV% reported for within-run, between-run, between-day, and between-site variations for body fluid control material (BFTROL - Table 11). All results met the acceptance criteria. For example, BFWBC (Level 2) total CV% was 6.34%; BFRBC (Level 3) total CV% was 3.55%.
Linearity (Whole Blood)WBC, TNC, RBC, HGB, HCT, PLT, PLT-Ox, RET#, NRBC#All results met predefined acceptance criteria and were acceptable.Linearity ranges were established for each parameter (Table 12). For example, WBC: 0.06 – 344.50 10^9/L; HGB: 0.5 – 25.8 g/dL.
Linearity (Body Fluids)BFWBC, BFRBCAll results met predefined acceptance criteria and were acceptable.Linearity ranges were established for BFWBC (3 – 11 345 10^6/L) and BFRBC (1079 – 5 394 633 10^6/L) (Table 13).
Interferences (Whole Blood)All parametersNo interference detected or no significant effect. Conjugated bilirubin may have a significant effect on low HGB levels.The device was found not susceptible to interference from Hemoglobin, Lipemia, Bilirubin (except conjugated bilirubin on low HGB), Glucose, and Yeast for various parameters. Intrinsic interferences from elevated WBC, RBC, and PLT measurands showed no interference for specific parameters. Some potential impacts (e.g., PLT-Ox and LYM# from macrothrombocytosis; RDW-SD from dual RBC population) were noted, but overall deemed acceptable within context of the study (Table 15).
Interferences (Body Fluids)All parametersNo interference detected or no significant effect. Interference from yeast was detected.No significant effect was detected on BFWBC, BFRBC from Hemoglobin, Lipemia, Bilirubin, and Total Protein across various fluid types. Interference from yeast was detected. Known interferences from crystals and liposomal particles were acknowledged as per literature (Table 17).
Stability (Whole Blood)All parametersAcceptance criteria for each parameter met for defined time intervals.Whole blood samples are stable for 24h at room temperature for CBC/LMNE/NRBC/RET parameters, and 48h (CBC/LMNE/NRBC) or 72h (RET) at refrigerated temperature (Table 18).
Stability (Body Fluid)All parametersAcceptance criteria for sample stability (max bias) met.Serous and synovial fluids are stable for 24h at room temperature for BFWBC/BFRBC/BFPN/BFMN parameters. CSF is stable for 4h at room temperature for BFWBC/BFRBC/BFPN/BFMN parameters (Table 19).
Detection Limits (Whole Blood)WBC, TNC, RBC, HGB, HCT, PLT, PLT-Ox, RET#All results met predefined acceptance criteria and were acceptable.LoB, LoD, and LoQ values were determined for various parameters (Table 20). For example, WBC: LoB 0.05, LoD 0.07, LoQ 0.10 (10^9/L).
Detection Limits (Body Fluid)BFWBC, BFRBCAll results met predefined acceptance criteria and were acceptable.LoB, LoD, and LoQ values were determined for BFWBC (LoB 2, LoD 4, LoQ 5 (10^6/L)) and BFRBC (LoB 500, LoD 1000, LoQ 1500 (10^6/L)) (Table 21).
Carry-over (Whole Blood)All parametersAll carry-over results are within specifications.Not applicable - Carry-over results are within specifications.
Carry-over (Body Fluids)All parametersAll carry-over results are within specifications.Not applicable - Carry-over results are within specifications.
Comparison Studies
Method Comparison (Whole Blood)All parametersAll results were within the predefined acceptance criteria and acceptable. Yumizen H2500 demonstrated comparable performance to predicate device.Passing-Bablok regression analysis (r, slope, intercept with 95% CI) was performed (Table 22). Correlations ranged from 0.184 (MCHC, potentially an outlier or specific context needed) to 0.998 (HGB). Most parameters showed high correlations and slopes close to 1, with intercepts close to 0, indicating strong agreement with the predicate device. For example, WBC: r=0.996, slope=1.012, intercept=0.045.
Method Comparison (Body Fluids)All parametersAll results were within the predefined acceptance criteria and acceptable. Yumizen H2500 demonstrated comparable performance to predicate device.Passing-Bablok regression analysis (r, slope, intercept with 95% CI) was performed for synovial, serous, and CSF (Tables 23, 24, 25). Correlations varied, with BFRBC showing very high correlation (0.999-1.000) across all fluid types. Other parameters showed good correlations (e.g., BFWBC r=0.923-0.980, BFMN% r=0.816-0.967) indicating comparable performance. For example, CSF BFWBC: r=0.980, slope=0.99, intercept=5.14.
Comparability (Sampling types)All parametersAcceptance criteria met for all parameters at all levels.Bias estimated at low, mid, and high points for each parameter showed comparability between capillary and venous whole blood samples.
Comparability (Anticoagulants)All parametersNo difference linked to anticoagulant or significant effect linked to matrix observed.Visual examination of Bland-Altman difference plots showed no difference linked to K2EDTA, Lithium Heparin, or Sodium Heparin anticoagulants for synovial fluid. No difference linked to anticoagulant for K2EDTA in serous fluids.
Comparability (Analytical Modes)All parametersAcceptance criteria met for all parameters at all levels.Bias estimated at low, mid, and high points for each parameter demonstrated comparable performance characteristics for all Yumizen H2500 modes (DIF, DIR, RBC_PLTO, DIF_LV).
Comparability (Manual vs Auto)All parametersAcceptance criteria met for all parameters at all levels.Bias estimated at low, mid, and high points for each parameter demonstrated comparable performance characteristics between automatic rack mode and manual (STAT) mode.
Clinical SensitivityMorphological Flags, Distributional Abnormality, Combined FlagsMet predefined acceptance criteria for both sensitivity and specificity.Sensitivity: 80.5% (Morphological), 91.9% (Distributional), 90.0% (Combined). Specificity: 83.6% (Morphological), 92.7% (Distributional), 90.4% (Combined). Efficiency: 82.2% (Morphological), 92.1% (Distributional), 90.1% (Combined) (Table 27).
Expected Values/Reference RangeAll parametersEstablishment of reference intervals.Reference intervals were established for adult (male/female) and pediatric (neonate, infant, child, adolescent) whole blood samples, and for synovial, serous, and CSF body fluids (Tables 28, 29, 30). This demonstrates the ability to define expected values for various populations.

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Repeatability:
    • Whole Blood: 116 residual K2EDTA whole blood samples (mixed normal and contrived for extremes) used for within-run repeatability.
    • Body Fluid: 87 residual body fluid samples (Synovial, Serous and Cerebrospinal Fluids) (mixed normal and contrived for extremes) used for within-run repeatability.
    • Provenance: Retrospective, samples around medical decision levels or contrived to cover analytical measuring range. Collected from multiple sites (4 test sites for whole blood, 4 test sites for body fluid). Country of origin is not explicitly stated beyond "4 test sites" which include "2 US sites and 2 European sites" for some studies, implying data from both regions.
  • Reproducibility:
    • Whole Blood: Three levels of control material (ABX Difftrol and ABX Minotrol Retic) run in duplicate twice a day for a minimum of 20 days (total 320 runs per level for each parameter if 4 sites and 2 runs/day * 20 days * 2 replicates * 2 instruments = 320 runs).
    • Body Fluids: Two levels of control material (BFTROL) run in duplicate twice a day for a minimum of 20 days (total 320 runs per level for each parameter).
    • Provenance: Control materials, conducted at 4 test sites (4 instruments, one per site).
  • Linearity:
    • Whole Blood: Minimum of seven concentration levels (commercial or prepared from dilutions) for each parameter. Each level tested in a minimum of 4 replicates on 2 instruments using 2 reagent lots.
    • Body Fluids: Mimimum of seven concentration levels (prepared from dilutions) for each parameter. Each level tested in a minimum of 4 replicates on 2 instruments using 2 reagent lots.
    • Provenance: One test site (for both whole blood and body fluids).
  • Interferences:
    • Whole Blood: Two concentration levels of interferent (Hemoglobin, Lipemia, Bilirubin, Glucose, Yeast) for direct interference. A subset of samples from the method comparison study (unique, native whole blood specimens identified with potential interference analysis, minimum of 17 specimens per interferent) for intrinsic interferences.
    • Body Fluids: Two concentration levels of interferent (Hemoglobin, Lipemia, Bilirubin, Total Protein, Yeast) for direct interference. A subset of samples from the method comparison study (unique, native body fluid specimens identified with potential interferents) for intrinsic interferences.
    • Provenance: 4 test sites (for both whole blood and body fluids).
  • Stability:
    • Whole Blood: 14 whole venous blood specimens.
    • Body Fluid: 28 body fluid specimens (13 serous, 7 synovial, and 8 cerebrospinal fluids).
    • Provenance: 3 test sites for whole blood, 2 test sites for body fluid.
  • Detection Limits (LoB, LoD, LoQ):
    • Six blank samples, six low concentration samples (for LoD), at least four low concentration samples (for LoQ). Each sample run 10 repeated times.
    • Provenance: Not specified beyond "different analyzers" and "two different reagent lots".
  • Carry-over:
    • High and low target value samples run consecutively (3 high, 3 low) for each parameter. Three sets of carry-over sequences.
    • Provenance: One test site, three analyzers.
  • Method Comparison:
    • Whole Blood: 969 venous and/or capillary specimens (pediatric (≤21 years) and adult). 143 with known medical conditions. Maximum 10% contrived.
    • Body Fluids: 427 residual body fluid specimens (pediatric (≤21 years) and adult) - 174 synovial, 138 serous, 115 CSF.
    • Provenance: 4 clinical sites (2 US sites and 2 European sites for whole blood); 3 clinical sites (2 US sites and 1 European site for body fluids). Mostly retrospective (leftover specimens).
  • Matrix Comparison (Comparability between sampling types):
    • 84 normal and pathological paired capillary and venous whole blood specimens.
    • Provenance: One clinical site. Prospective collection.
  • Matrix Comparison (Comparability between body fluid anticoagulants):
    • Synovial fluid: 9 without anticoagulant, 39 with K2EDTA, 92 with Lithium Heparin, 34 with Sodium Heparin.
    • Serous fluid: 82 without anticoagulant, 56 with K2EDTA.
    • Provenance: 3 clinical sites.
  • Matrix Comparison (Comparability between analytical modes):
    • DIR vs DIF: 166 normal and pathological residual whole blood specimens.
    • RBC_PLTO vs DIF: 172 normal and pathological residual whole blood specimens.
    • DIF_LV vs DIF: 187 normal and pathological residual whole blood specimens.
    • Provenance: One clinical site for each comparison.
  • Matrix Comparison (Comparability mode to mode):
    • 83 normal and pathological residual whole blood samples. (Automatic rack mode vs manual STAT mode).
    • Provenance: One clinical site.
  • Clinical Sensitivity:
    • 456 residual normal and abnormal whole blood samples (from method comparison study).
    • Provenance: 4 clinical sites.
  • Expected Values/Reference Range:
    • Adult Whole Blood: 240 apparently healthy adults (120 male, 120 female).
    • Pediatric Whole Blood: At least 80 apparently healthy neonates, infants, children, and adolescents (<22 years) (approx. 40 male, 40 female).
    • Body Fluids: At least 100 apparently healthy men and women (Synovial, serous and cerebrospinal fluids).
    • Provenance: 2 US clinical sites for adult and pediatric whole blood; 2 US clinical sites for body fluids.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The document does not specify the number of experts or their qualifications for establishing ground truth for the test set.

For the clinical sensitivity (flagging analysis), the "reference (predicate or manual count results, depending on the parameter)" was used as ground truth. If manual counts were performed, the number and qualifications of the personnel (e.g., medical technologists, pathologists) performing these counts are not mentioned.

For the method comparison studies, the predicate device (Sysmex XN-series) served as the reference method. This is not "expert ground truth" in the typical sense for imaging or pathology, but rather comparison against an already cleared and accepted method.

For expected values/reference ranges, healthy adult and pediatric populations were studied, but the diagnosis of "healthy" is usually based on clinical assessment and laboratory results, not a separate expert panel for ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe an adjudication method (like 2+1 or 3+1 consensus) for the test set. For flagging analysis, ground truth was established by either the predicate device or manual counts. For method comparison, it was against the predicate device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an automated hematology analyzer, not an AI software intended to assist human readers in interpretation. The studies focus on the analytical performance of the instrument itself compared to a predicate device or established laboratory methods.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the studies are essentially standalone performance evaluations of the Yumizen H2500 automated hematology analyzer. The device generates results automatically, which are then compared to reference methods (predicate device, manual counts, or defined analytical limits). There is no "human-in-the-loop" component in the reported performance metrics for the device itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The types of "ground truth" used are:

  • Predicate Device Measurements: For the method comparison studies, the Sysmex XN-Series (an already cleared automated hematology analyzer) served as the reference method.
  • Manual Count Results: For clinical sensitivity (flagging analysis), manual count results were used as ground truth for certain parameters, likely for differential counts where visual microscopy is the gold standard.
  • Defined Analytical Limits/Scientific Principles: For analytical performance studies (precision, linearity, detection limits, carry-over, stability, interference), adherence to established scientific and statistical methods (e.g., CLSI guidelines) defines the "ground truth" for acceptable performance.
  • Clinically Defined Healthy Populations: For establishing expected values/reference ranges, samples from apparently healthy adult and pediatric populations were used.

8. The sample size for the training set

The document does not specify a training set size. This device is an automated hematology analyzer, implying it likely uses established algorithms and hardware for measurement, rather than a machine learning model that requires a distinct "training set" in the context of AI. The performance studies focus on validation and verification rather than development and training.

9. How the ground truth for the training set was established

Since there is no mention of a traditional "training set" for an AI model, the question of how ground truth was established for it is not applicable in the context of this device and documentation. The device's operation is based on established impedance, optical cytometry, and spectrophotometry principles, not on a machine learning model trained on a specific dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 21, 2024

Horiba ABX Sas Caroline Ferrer Regulatory Affairs Manager Parc Euromedecine, Rue du Caducee BP7290 Montpellier Cedex 4, 34184 France

Re: K232946

Trade/Device Name: Yumizen H2500 Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Dated: September 14, 2023 Received: September 20, 2023

Dear Caroline Ferrer:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

K232946 - Caroline Ferrer

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Min Wu -

Min Wu, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Device Name Yumizen H2500

Indications for Use (Describe)

The Yumizen H2500 is a quantitative multiparameter fully automated hematology analyzer intended for in-vitro diagnostic use in clinical laboratories by qualified healthcare professionals for the screening of patient populations.

The Yumizen H2500 is intended to perform tests on the following specimens:

  • · Anticoagulated whole blood specimens
  • · Body fluids (synovial fluids, serous fluids and cerebrospinal fluids).

The Yumizen H2500 classifies and enumerates the following parameters:

· A complete blood count (CBC) consisting of TNC, WBC, RBC, calculated HCT, MCV, calculated MCH, calculated MCHC, RDW-SD, RDW-CV, PLT, PLT-Ox, LPF, MPV.

  • · A leukocyte differential count consisting of LYM (%), NEU (%), EOS (%(#), BAS (%/#), IMG (%/#), IMG (%/#),
  • · A nucleated red blood cell count consisting of NRBC (%/#).
  • · A reticulocyte analysis consisting of RET (%/#), calculated CRC, IRF, RHCC.

· Quantitative determination of blood cells in synovial fluids, serous fluids and cerebrospinal fluids consisting of BFWBC, BFRBC, BFPN (%/#), BFMN (%/#).

Note: Venous and capillary whole blood be collected in K2EDTA anticoagulant. Serous and synovial fluids should be collected without anticoagulant or in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with cerebrospinal fluid specimens is neither required nor recommended. Alternatively, Sodium Heparin or Lithium Heparin may be used for synovial fluid.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

510(k) Safety and Effectiveness Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements 21 CFR 807.92 (b)(3).

Submitted by:

Company: HORIBA ABX SAS Address: Parc Euromédecine Rue du Caducée – BP 7290 34184 Montpellier Cedex 4 FRANCE

Contact Person: Caroline FERRER Telephone (France): +33 467 141843 Email: caroline.ferrer@horiba.com

Date submitted: May 20th, 2024

{4}------------------------------------------------

VOL 004 : 510(k) Safety and Effectiveness Summary

Yumizen H2500

Traditional 510(k): Supplement to Original Submission K232946

I.BACKGROUND INFORMATION................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................
g.Accuracy (Instrument):32
h.Carry-over:32
B.Comparison Studies34
a.Method Comparison with Predicate Device:34
b.Matrix Comparison:39
c.Clinical Sensitivity:42
C.Clinical Studies43
D.Clinical Cut-Off43
E.Expected Values/Reference Range:44
F.Other Supportive Instrument Performance Characteristics Data46
VIII.PROPOSED LABELING46
IX.CONCLUSION46

{5}------------------------------------------------

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

{6}------------------------------------------------

l. Background Information

A. 510(k) number

K232946

B. Applicant

HORIBA ABX SAS (Brand HORIBA Medical)

C. Proprietary and Established Names

Yumizen H2500

D. Regulatory Information

Product CodeClassRegulation sectionRegulation DescriptionPanel
GKZII864.5220Automated Differential CellCounterHematology

{7}------------------------------------------------

Submission / Device Overview ll.

A. Purpose for Submission

Clearance for a new device: Yumizen H2500

B. Parameters

TNC, WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, LYM (%/#), MON (%/#), NEU (%/#), EOS (%/#), BAS (%/#), NRBC (%/#), BFWBC, BFRBC, BFPN (%/#), BFMN (%/#), PLT-Ox, LPF, RET (%/#), CRC, IRF, RHCC

C. Type of Test

A complete blood count (CBC)

A leukocyte differential count

A nucleated red blood cell count

A reticulocyte analysis

A body fluid count

{8}------------------------------------------------

���. Intended Use/Indications for Use

A. Intended Use(s)

See Indications for Use below.

B. Indications for Use

The Yumizen H2500 is a quantitative multiparameter fully automated hematology analyzer intended for in-vitro diagnostic use in clinical laboratories by qualified healthcare professionals for the screening of patient populations.

The Yumizen H2500 is intended to perform tests on the following specimens:

  • Anticoagulated whole blood specimens ●
  • Body fluids (synovial fluids, serous fluids and cerebrospinal fluids). .

The Yumizen H2500 classifies and enumerates the following parameters:

  • A complete blood count (CBC) consisting of TNC, WBC, RBC, HGB, calculated . HCT, MCV, calculated MCH, calculated MCHC, RDW-SD, RDW-CV, PLT, PLT-Ox, LPF, MPV.
  • A leukocyte differential count consisting of LYM (%#), MON (%#), NEU (%/#), ● EOS (%/#), BAS (%/#), IMG (%/#)
  • A nucleated red blood cell count consisting of NRBC (%/#). ●
  • A reticulocyte analysis consisting of RET (%/#), calculated CRC, IRF, RHCC. ●
  • Quantitative determination of blood cells in synovial fluids, serous fluids and . cerebrospinal fluids consisting of BFWBC, BFRBC, BFPN (%/#), BFMN (%/#).

Note: Venous and capillary whole blood should be collected in K2EDTA anticoagulant. Serous and synovial fluids should be collected without anticoaqulant or in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with cerebrospinal fluid specimens is neither required nor recommended. Alternatively, Sodium Heparin or Lithium Heparin may be used for synovial fluid.

C. Special Conditions for Use Statement

Rx - For Prescription Use Only

{9}------------------------------------------------

IV. Device/System Characteristics

A. Device Description

The HORIBA Medical analyzer modules Yumizen H2500 are multi-parameter hematology analyzers intended to perform tests on whole blood samples collected in K2EDTA and body fluids (synovial and serous) collected in K2EDTA anticoaqulant. The analyzers can also perform tests on cerebrospinal fluids which should not be collected in any anticoaqulant.

The Analyzer Units (Yumizen H2500) aspirate, dilute, mix, and analyze blood and body fluid samples.

The Yumizen H2500 model provides Complete Blood Count (CBC), Differential (DIFF), Reticulocyte counts (RET) and Optical Platelet counts as well as Body Fluid counts (BF).

The analyzer models may function with:

  • · a Data Management Unit (Yumizen P8000) which is the interface with the laboratory connections (LIS) and the Analyzer Unit(s). Through its interface, the Yumizen P8000 enables the user to monitor the workflow of patient data, centralize result data, perform reflex testing, customize rules, centralize the validation operations, run quality control, manage quality assurance on results.

B. Principles of Operation

The HORIBA Yumizen H2500 uses impedance variation, optical and fluorescence flow cytometry, hydrodynamic focusing, and absorption spectrophotometry technologies to measure, count, and calculate hematological parameters in samples.

  • . Impedance variation consists in measuring the proportional response to the volume of the particle generated by its passage through a calibrated micro aperture concurrent with an electric current.
  • Optical and fluorescence flow cytometry is a process used to count and measure . the properties of cells or particles as they are carried by fluid through a sensing zone.

The cells are counted with impedance variation technology and physical and chemical characteristics of cells or particles are measured via light absorbency response from a light beam, or fluorescence response from a laser.

The Yumizen H2500 uses double hydrodynamic sequential system (DHSS) flow cytometry to focus and align cells into a single-file micro aperture through the sensing zone.

Two cell-free liguid sheaths surround the diluted sample and move with it in a laminar flow. The laminar flow prevents any mixing between the two liquid sheaths and the diluted sample.

  • Absorption spectrophotometry is based on the linear relationship between the ● amount of light that a well-mixed, nonflowing sample absorbs at a particular

{10}------------------------------------------------

absorption band and the concentration of an absorbing entity in the sample (Beer-Lambert Law).

To perform absorption spectrophotometry, the system uses the hemoglobin dilution as the sample and a specific reagent to release, oxidize and stabilize all heme iron from erythrocytes. The resulting complexes are quantified by spectrophotometry at a wavelength of 555nm.

Impedance variation technology through a calibrated micro aperture is used to analyse whole blood samples for RBC, PLT, TNC and body fluids samples (BFRBC). Flow cytometry technology is used to:

  • Separate NRBC from TNC to get the WBC and make the differential white blood ● cell formula with absorbency measurement.
  • Analyse whole blood samples for RET with fluorescence measurement.
  • . Analyse whole blood samples for PLT with absorbency measurement (PLT-Ox)
  • Analyse body fluids samples (BFWBC) ●

Absorption spectrophotometry is used to measure the HGB concentration.

All the other hematological parameters are calculated or derived from these different measurements.

Based on the available technologies and types of tests, the following analytical modes are available on the analyzers:

Analytical ModesCorresponding tests
CBCComplete blood count
DIFComplete blood count
Leukocyte differential count
DIF_LVComplete blood count
Leukocyte differential count for low values
RBC_PLTOComplete blood count
PLT-Ox
RETReticulocyte analysis
CBR (CBC + RET)Complete blood count and
Reticulocyte analysis
DIR (DIF + RET)Complete blood count
Leukocyte differential count
Reticulocyte analysis
CBFBody fluid count

Table 1 - Analvtical modes

{11}------------------------------------------------

C. Instrument Description

  • a. Instrument Name:

Yumizen H2500

b. Specimen Identification:

Tube sample ID can be identified manually using an external keyboard or the virtual keyboard or automatically using the integrated barcode reader or an external barcode reader.

c. Specimen Sampling and Handling:

The Yumizen H2500 operates in two sampling modes:

  • Automatic sampling (auto mixer and auto loader) .
  • Manual sampling (STAT mode) ●

d. Calibration:

Calibration is a procedure that is performed during specific situations such as installation, maintenance or service intervention. It ensures that the precision and accuracy of the analyzer are acceptable, so that accurate measurements are performed by the analyzer. Calibration is performed using materials with assigned values that are traceable to standard reference methods. It is recommended that the laboratory calibrate using ABX Minocal, which is a commercial whole blood calibrator.

e. Quality Control:

Quality control allows the user to monitor a set of analyses based on known sample values and ranges over a period of several months. Statistical computation performed on these populations allows the extraction of qualitative information related to the stability of the instrument.

The manufacturer's instructions are to be followed for material and frequency of quality control analysis.

Commercial controls:

ABX Difftrol, for CBC and Differential parameters ABX Minotrol Retic, for Retic parameters

BFTROL, for Body Fluid parameters

{12}------------------------------------------------

V. Substantial Equivalence Information

A. Predicate Device Name(s)

Sysmex XN-series

B. Predicate 510(k) Number(s)

K112605

C. Comparison with Predicate

Table 2 - Comparison with predicate

Device & PredicateDevice(s)Device Trade NameK112605
Yumizen H2500Sysmex® XN-Series (XN-10, XN-20)Automated Hematology Analyzers (K112605)
General Device Characteristics Similarities
Regulation21 CFR 864.522021 CFR 864.5220
Product codeGKZGKZ
IntendedUse/Indications foruseThe Yumizen H2500 is a quantitativemultiparameter fully automated hematologyanalyzer intended for in-vitro diagnostic usein clinical laboratories by qualifiedhealthcare professionals for the screeningof patient populations.The Yumizen H2500 is intended to performtests on the following specimens:• Anticoagulated whole blood specimens• Body fluids (synovial fluids, serousfluids and cerebrospinal fluids).The Yumizen H2500 classifies andenumerates the following parameters:• A complete blood count (CBC)consisting of TNC, WBC, RBC, HGB,calculated HCT, MCV, calculated MCH,calculated MCHC, RDW-SD, RDW-CV,PLT, PLT-Ox, LPF, MPV.• A leukocyte differential count consistingof LYM (%/#), MON (%/#), NEU (%/#),EOS (%/#), BAS (%/#), IMG (%/#)• A nucleated red blood cell countconsisting of NRBC (%/#).• A reticulocyte analysis consisting ofRET (%/#), calculated CRC, IRF,RHCC.• Quantitative determination of blood cellsin synovial fluids, serous fluids andcerebrospinal fluids consisting ofBFWBC, BFRBC, BFPN (%/#), BFMNThe Sysmex® XN-10 and XN-20 modules arequantitative multi-parameter automatedhematology analyzers intended for in vitrodiagnostic use in screening patient populationsfound in clinical laboratories. The XN-Seriesmodules classify and enumerate the followingparameters for whole blood: WBC, RBC, HGB,HCT, MCV, MCH, MCHC, PLT (PLT-I, PLT-F),NEUT%/#, LYMPH%/#, MONO%/#, EO%/#,BASO%/#, NRBC%/#, IG%/#, RDW-CV, RDW-SD, MPV, RET%/#, IRF, IPF, RET-He and has aBody Fluid mode for body fluids. The Body Fluidmode enumerates the WBC-BF, RBC-BF,MN%/#, PMN%/# and TC-BF# parameters incerebrospinal fluids (CSF), serous fluids(peritoneal, pleural) and synovial fluids. Wholeblood should be collected in K2 or K3EDTAanticoagulant and, Serous and Synovial fluids inK2EDTA anticoagulant to prevent clotting offluid. The use of anticoagulants with CSFspecimens is neither required norrecommended.
Device & PredicateDevice(s)K112605
Device Trade NameYumizen H2500Sysmex® XN-Series (XN-10, XN-20)Automated Hematology Analyzers (K112605)
(%/#).
Note: Venous and capillary whole bloodshould be collected in K2EDTAanticoagulant. Serous and synovial fluidsshould be collected without anticoagulant orin K2EDTA anticoagulant to prevent clottingof fluid. The use of anticoagulants withcerebrospinal fluid specimens is neitherrequired nor recommended. Alternatively,Sodium Heparin or Lithium Heparin may beused for synovial fluid.
Test Principle(same)Performs hematology analyses according tothe Double Hydrodynamic SequentialSystem, flow cytometry method (using asemiconductor laser and fluorescent dyes)Performs hematology analyses according to theHydro Dynamic Focusing (DC Detection), flowcytometry method (using a semiconductor laserand fluorescent dyes)
Test parameters*(same)WBC, RBC, HGB, HCT, MCV, MCH,MCHC, PLT, NEU%/#, LYM%/#, MON%/#,EOS%/#, BAS%/#, NRBC%/#, RDW-CV,RDW-SD, MPV, RET%/#, IRF, IMG%/#,BFWBC, BFRBC, BFMN%/#, BFPMN%/#WBC, RBC, HGB, HCT, MCV, MCH, MCHC,PLT (PLT-I), NEUT%/#, LYMPH%/#,MONO%/#, EO%/#, BASO%/#, NRBC%/#,RDW-CV, RDW-SD, MPV, RET%/#, IRF, IG%/#,WBC-BF, RBC-BF, MN%/#, PMN%/#
Specimen typesWhole bloodBody fluidWhole bloodBody fluid
Use ofControls/CalibratorsYesYes
Modes of OperationAutomatic Rack modeBody Fluid Analysis ModeSampler Analysis ModeBody Fluid Analysis Mode
Manual mode (both close and open cap)(Sample placed in tube holder position)Manual Closed Analysis ModeManual Open Cap Analysis Mode (Sampleplaced in tube holder position)
Low Value mode (DIF_LV)Low WBC Mode (LWBC)
Type of tubesVenous and capillary tubesVenous and capillary tubes
Patient populationAdult and PediatricGeneral Patient population
Instrument operatingtemperature15 to 30°C15 to 30°C
Relative humidityUp to 85%30 to 85%
TransportationconditionFrom -20°C (-4°F) to +60°C (+140°F).-10 to 60°C
* Different names/formats of equivalent parameters are used between the Yumizen H2500 device andSysmex® XN-series; therefore, equivalent parameters are listed in the same order.

{13}------------------------------------------------

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

{14}------------------------------------------------

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

Device & PredicateDevice(s)K112605
Device Trade NameYumizen H2500Sysmex® XN-Series (XN-10, XN-20)Automated Hematology Analyzers (K112605)
General Device Characteristics - Differences
Test Principle(different)HGB: Spectrophotometry using cyanide-free lysisHGB: Spectrophotometry using SLS-hemoglobinmethod
Test parameters(different)Whole blood parameters:PLT-Ox, LPF, RHCC, CRCWhole blood parameters:PLT-F, IPF, RETHe#
Body Fluid parameters:/Body Fluid parameters:TC-BF
ReagentsABX DILUENT 10L (Diluent)ABX DILUENT 20L (Diluent)ABX LYSEBIO 1L (Lyse)ABX BASOLYSE 5L (Lyse)NUCEDIFF 1L (Lyse)ABX FLUOCYTE 0.5L (Stain)ABX CLEANER 1L (Cleaner)ABX MINOCLAIR 0.5L (Cleaner)CELLPACKTM DCL (Diluent)CELLPACKTM DFL (Diluent)LYSERCELL WNR (Lyse)LYSERCELL WDF (Lyse)LYSERCELL WPC (Lyse)SULFOLYSER (Lyse)FLUOROCELL WNR (Stain)FLUOROCELL WDF (Stain)FLUOROCELL RET (Stain)FLUOROCELL PLT (Stain)FLUOROCELL WPC (Stain)
Controls / CalibratorsWhole Blood:ABX DIFFTROL (3 Levels)ABX MINOTROL RETIC (3 Levels)ABX MINOCALBody Fluid:BFTROL (2 Levels)Whole Blood:XN-Check - 3 LevelsXN CAL (XN-10/X-20 Calibrator)XN CAL PF (Platelet F Calibrator)Body Fluid:XN Check BF - 2 Levels
Modes of OperationNo-predilute analysis modeNo dilution of samplePre-dilute Analysis ModeDilute sample 1 :7
ThroughputWhole Blood:120 samples/hour maximum depending onmode used.Body Fluid20 samples/hour maximumWhole Blood100 samples/hour maximum depending onmode used.Body Fluid40 samples/hour maximum
Sample AspirationVolumeRack analysis: 110 µLManual analysis: 110 µLSampler Mode – 88 µLManual (Closed Cap) Mode – 88 µLManual (Open Cap) Mode - 88 μLDilution Mode – 70 µLBody Fluid Mode – 88 µL
Dimension870(W) x 730(H) x 670(D) mm645(W) x 855(H) x 755(D) mm
Weightabout 99 kgabout 78 kg (including the sampler)

{15}------------------------------------------------

VI. Standards/Guidance Documents Referenced

Table 3 - List of standards

Standard referenceStandard Title [FDA recognition number]
CLSI EP05-A3Evaluation of Precision of Quantitative Measurement Procedures; ApprovedGuideline - Third Edition [7-251]
CLSI EP06-A2Evaluation of the Linearity of Quantitative Measurement Procedures [7-306]
CLSI EP07-A3Interference Testing in Clinical Chemistry [7-275]
CLSI EP09-A3cMeasurement Procedure Comparison and Bias Estimation Using PatientSamples [7-296]
CLSI EP17-A2Evaluation of Detection Capability for Clinical Laboratory MeasurementProcedures; Approved Guideline - Second Edition [7-233]
CLSI EP28-A3cDefining, Establishing, and Verifying Reference Intervals in the ClinicalLaboratory; Approved Guideline - Third Edition [7-224]
CLSI EP37-ASupplemental Tables for Interference Testing in Clinical Chemistry [7-284]
CLSI H20-A2Reference Leukocyte (WBC) Differential Count (Proportional) and Evaluation ofInstrumental Methods; Approved Standard - Second Edition [7-165]
CLSI H26-A2Validation, Verification, and Quality Assurance of Automated HematologyAnalyzers; Proposed Standard - Second Edition [7-210]
CLSI H44-A2Methods for Reticulocyte counting (Automated blood cell counters, Flowcytometry, and Supravital dyes); Approved Guideline – Second edition
CLSI H56-ABody Fluid Analysis for Cellular Composition; Approved Guideline. [7-163]
IEC 61010-1:2010+A1Safety requirements for electrical equipment for measurement, control, andlaboratory use - Part 1: General requirements [19-34]
UL 61010-1:2012R7.19Standard for Safety for Electrical Equipment For Measurement, Control andLaboratory Use; Part 1: General Requirements [19-41]
IEC 60601-1-2: 2014Medical electrical equipment - Part 1-2: General requirements for basic safetyand essential performance - Collateral Standard: Electromagnetic disturbances- Requirements and tests [19-8]
IEC 60825-1: 2014Safety of laser products - Part 1: Equipment classification, and requirements
IEC 62304: 2015Medical device software - Software life cycle processes [13-79]
ISO 14971: 2019Medical devices - Application of risk management to medical devices [5-125]
ISO 15223-1: 2021Medical devices - Symbols to be used with information to be supplied by themanufacturer - Part 1: General requirements [5-134]
Standard referenceStandard Title [FDA recognition number]
ISO 18113-3: 2022In vitro diagnostic medical devices - Information supplied by the manufacturer(labelling) - Part 3: In vitro diagnostic instruments for professional use

{16}------------------------------------------------

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

{17}------------------------------------------------

Performance Characteristics (if/when applicable) VII.

  • A. Analytical Performance
  • a. Precision/Reproducibility:

Repeatability - Whole Blood

Repeatability studies were performed according to CLSI H26-A2 guide recommendations at four (4) test sites with seven (7) instruments using DIF, DIF, DIF LV, or RBC PLTO modes.

Within-run repeatability studies were performed using a total of 116 residual K2EDTA whole blood samples around medical decision levels and within the laboratory reference range. Contrived samples were utilized to cover the extremes of the analytical measuring range of some parameters.

Each sample was measured 12 consecutive times.

For each parameter, the mean and coefficient of variation (CV) were computed. All components of variation that were calculated met the pre-defined acceptance criteria. The results are summarized below.

TargetRangeRecommended TargetValuesParameterNResult RangeMax %CVMax SD
Normal4-10 109/LWBC374.27-9.8420.150
Normal4-10 109/LTNC364.27-9.841.80.150
Normal3.6-6.2 1012/LRBC583.66-5.921.50.057
Normal12.0-18.0 g/dLHGB4412.0-17.60.60.096
Normal36-54 %HCT4636.4-53.61.20.5
Normal80-100 fLMCV8780.4-99.40.50.530
Normal27-32 pgMCH7927.1-31.81.20.344
Normal320-360 g/LMCHC84320-3551.23.954
Normal5-17 %RDWcv7112.6-16.92.70.350
Normal10-49 fLRDWsd5331.4-48.93.41.453
Normal150-500 109/LPLT64156-479511.595
Normal8-12 fLMPV838.2-1230.242
Normal150-500 109/LPLT-Ox16155-4694.712.317
Normal2-12 %LPF152.1-11.220.70.834
Normal25-50 %LYM%2625.5-43.23.81.101
Normal2-10 %MON%413.8-9.79.10.845
Normal45-80 %NEU%3646.8-79.21.91.093
Normal0.5-5 %EOS%380.5-4.918.20.442
Normal0-2.5 %BAS%750-2.239.30.370

Table 4 - Repeatability - Whole Blood normal samples

{18}------------------------------------------------

TargetRangeRecommended TargetValuesParameterNResult RangeMax %CVMax SD
Normal0.5-3 %IMG%460.5-2.819.40.355
Normal0.8-5 109/LLYM#480.84-4.257.20.129
Normal0.1-1 109/LMON#420.33-0.989.10.058
Normal2 - 8 109/LNEU#362.18-7.322.60.113
Normal0-0.5 109/LEOS#440.02-0.48220.027
Normal0-0.25 109/LBAS#730.00-0.2138.40.053
Normal0-0.2 109/LIMG#480.01-0.1911.60.039
Normal0-2.5 %NRBC%590.00-2.1818.80.481
Normal0-0.15 109/LNRBC#540.00-0.020.00.017
Normal0.5-3 %RET%330.52-2.9411.60.209
Normal20-150 109/LRET#3528-14717.112.525
Normal0.75-3 %CRC320.83-2.9312.20.249
Normal0.15-0.3IRF170.15-0.2914.40.033
Normal27-35 pgRHCc2128.41-34.863.51.077

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

Table 5 - Repeatability – Whole Blood abnormal samples

TargetRangeRecommended TargetValuesParameterNResult RangeMax %CVMax SD
Low0.1-2 109/LWBC10.813.90.032
Low2-4 109/LWBC42.81-3.971.80.065
Low0.5-4 109/LTNC50.81-3.973.90.065
Low0.3-3.0 1012/LRBC251.17-2.991.70.038
Low3.0-3.6 1012/LRBC233.02-3.560.90.028
Low4.0 - 8.0 g/dLHGB184.1-7.91.00.098
Low8.0-12.0 g/dLHGB438.0-11.80.90.121
Low4-36 %HCT599.8-35.61.60.4
Low40-80 fLMCV1459.6-80.00.40.322
Low30 -150 109/LPLT956-13087.236
Low30 -150 109/LPLT-Ox2106-1314.64.830
Low10 - 25%LYM%2010.9-24.460.889
Low20 - 45%NEU%626.7-43.53.70.981
Low0.00 - 0.15IRF180.03-0.1525.60.031
High10-30 109/LWBC3110.03-19.91.90.330
High30-300 109/LWBC534.4-150.7310.966
High10-300 109/LTNC4110.01-153.243.65.540
High6.2-8 1012/LRBC106.43-7.830.70.045

{19}------------------------------------------------

TargetRangeRecommended TargetValuesParameterNResult RangeMax %CVMax SD
High18.0-24.0 g/dLHGB919.0-22.00.40.134
High54-67 %HCT1054.1-65.70.80.5
High100-200 fLMCV13101.7-137.70.50.531
High17-50 %RDWcv1317.1-28.62.90.644
High49-100 fLRDWsd3149.4-82.82.82.317
High500-2500 109/LPLT18515-959213.968
High500-2500 109/LPLT-Ox10678-10472.323.649
High12-100 %LPF215.2-15.45.30.811
High50-100LYM%254.6-58.81.30.735
High10-100MON%1510.1-20.55.60.903
High80-100NEU%184.5-84.50.50.447
High5-100EOS%56.4-22.74.90.770
High2.5-100BAS%23.4-611.40.458
High3-100IMG%73.6-26.816.11.272
High5-300 109/LLYM#25.2-6.852.60.136
High1-300 109/LMON#131.04-6.1560.297
High8-300 109/LNEU#58.02-13.941.90.266
High0.5-300 109/LEOS#50.51-2.35.60.064
High0.25-300 109/LBAS#20.33-2.2280.091
High0.2-300 109/LIMG#180.2-22.9219.11.145
High2.5-100NRBC%104.0-45.98.12.827
High0.15-20 109/LNRBC#140.16-6.03170.414
High3 - 35 %RET%73.3-11.17.40.286
High150-1200 109/LRET#5160-3014.80.008
High3 - 35 %CRC53.2-6.74.70.235
High0.3-1IRF40.31-0.398.30.026
High35-100 ngRHCc835.2-45.22.91.055

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

Repeatability – Body Fluid

Repeatability studies were performed according to CLSI H56-A guide recommendations at four (4) test sites with seven (7) instruments using CBF mode.

Within-run repeatability studies were performed using a total of 87 residual body fluid samples (Synovial, Serous and Cerebrospinal Fluids) around medical decision levels and

{20}------------------------------------------------

within the laboratory reference range. Contrived samples were utilized to cover the extremes of the analytical measuring range of some parameters.

Each sample was measured 4 to 10 consecutive times.

For each parameter, the mean and coefficient of variation (CV) were computed. The data supports the validity of the repeatability performance claims for the instrument models tested, as summarized in Table below.

Table 6 - Repeatability – Body Fluid serous samples
TargetRangeRecommended TargetValuesParameterNResult RangeMax %CVMax SD
Level 15-100 106/LBFWBC80-2910.48.5
Level 11500-5000 106/LBFRBC50-759717500.8
Level 15-100 106/LBFPN#210-9736.64.4
Level 15-25BFPN%210-739.73.6
Level 15-300v106/LBFMN#210-549.25.9
Level 15-80BFMN%140-411.73.6
Level 2100-600 106/LBFWBC18101-49213.628.8
Level 25000-5000000 106/LBFRBC218445-7758289.47596.8
Level 2100-10000 106/LBFPN#11110-33168.196.7
Level 225-100BFPN%825-8515.46.7
Level 2300-10000 106/LBFMN#17322-25547.853.8
Level 280-100BFMN%2481-994.43.7
Level 3600-10000 106/LBFWBC12699-38874.4111.2

Table 6 - Repeatability – Body Fluid serous samples

Table 7 - Repeatability – Body Fluid synovial samples

TargetRangeRecommendedTarget ValuesParameterNResult RangeMax %CVMax SD
Level 15-100 106/LBFWBC719-8211.89.7
Level 11500-5000 106/LBFRBC21864-344018.1482.2
Level 15-100 106/LBFPN#86-6337.66.3
Level 15-25BFPN%68-2420.33.4
Level 15-300 106/LBFMN#119-23620.67.7
Level 15-80BFMN%1713-7917.63.6
Level 2100-600 106/LBFWBC7110-52010.523.2
Level 25000-5000000 106/LBFRBC165084-37987318.730369.5
Level 2100-10000 106/LBFPN#12103-61798.8213.2
Level 225-100BFPN%1534-879.85.4
Level 2300-10000 106/LBFMN#11484-670014.4604.4
Level 280-100BFMN%583-936.65.6
Level 3600-10000 106/LBFWBC131144-95404.8222.0

{21}------------------------------------------------

Table 8 - Repeatability – Body Fluid CSF samples
TargetRangeRecommendedTarget ValuesParameterNResult RangeMax %CVMax SD
Level 15-100 106/LBFWBC917-8919.28.6
Level 11500-5000 106/LBFRBC32080-367816.0357.2
Level 15-100 106/LBFPN#11816.22.9
Level 15-25BFPN%26-80.00.0
Level 15-300 106/LBFMN#911-20927.55.6
Level 15-80BFMN%45-4518.23.5
Level 2100-600 106/LBFWBC3260-54610.054.8
Level 25000-5000000 106/LBFRBC78555-28879457.011989.8
Level 2100-10000 106/LBFPN#6144-292914.765.8
Level 225-100BFPN%655-9917.710.8
Level 2300-10000 106/LBFMN#14387-43871.566.1
Level 280-100BFMN%484-979.37.8
Level 3600-10000 106/LBFWBC3640-46572.369.2

Table 8 - Re eatability Body Fluid CSF .

{22}------------------------------------------------

Reproducibility - Whole Blood

Reproducibility studies were performed to evaluate the within-run, between-run, between-day, between-instrument, and total precision of the Yumizen H2500 analyzer. Testing was performed using three (3) levels of each whole blood control material (ABX Difftrol and ABX Minotrol Retic). Each level was run in duplicate twice each day for a minimum of 20 days using a single calibration and reagent lot at each of the four (4) test sites. The results were analyzed in accordance with the CLSI EP05-A3 approved guideline and met the acceptance criteria. Results are summarized below.

ParameterLevelNMeanWithin runBetween runBetween dayBetween site(a)Total
SDCV%SDCV%SDCV%SDCV%SDCV%
WBCL3202.280.041.600.010.350.000.000.073.220.083.62
10%LN3207.280.060.830.060.800.000.000.192.660.212.90
H32018.220.130.690.080.450.110.600.492.700.532.88
L3202.410.020.780.010.340.000.000.020.640.031.07
RBC1012/LN3204.820.030.660.000.000.010.120.010.180.030.69
H3205.540.040.760.000.000.010.200.040.630.061.01
L3206.1410.0250.400.180.290.000.000.550.890.631.02
HGBg/dLN32013.6790.0410.300.0000.000.0220.160.1130.830.1220.89
H32017.0860.0490.290.0300.180.0200.120.1310.7701440.84
L32018.610.150.800.100.520.050.270.291.570.351.86
НСТN32041.020.310.740.080.190.080.200.561.350.921.57
%H32051.440.420.820.000.000.160.300.801.560.901.78
L32077.080.230.300.210.270.270.350.801.040.901.17
MCVN32085.180.240.280.160.190.220.261.061.241.121.32
tH32092.820.280.300.160.170.280.301.151.241.221.32
L32025.420.210.840.100.410.000.000.381.480.441.75
MCHN32028.400.190.680.030.090.000.000.260.910.321.14
baH32030.830.230.750.000.000.080.250.351.120.421.37
L320329.872.740.832.050.620.850.268.232.508.962.72
MCHCN320333.502.540.760.760.230.570.177.172.157.662.30
g/LH320332.252.600.780.000.001.280.397.612.298.142.45
L32015.930.251.570.050.330.040.230.291.830.392.45
RDW-CV%N32013.590.201.500.020.120.130.970.302.200.392.84
H32012.340.201.580.100.820.090.730.272.160.362.90
RDW-SDL32042.150.962.290.230.540.360.851.072.541.503.57
frN32039.790.781.960.330.830.611.540.932.341.403.51
H32040.400.791.970.441.090.621.540.922.281.443.56
PLTL32063.073.365.320.000.001.001.580.921.463.625.74
N320247.036.452.610.000.001.610.657.272.959.863.99
10%LH320483.158.151.692.530.520.000.0014.943.0917.213.56
L32065.025.638.661.722.652.684.123.465.327.3411.28
PLT-OxN320244.786.792.772.491.023.561.456.302.5810.234.18
10%LH320478.7010.772.250.000.003.770.798.281.7314.092.94
L32011.170.171.530.060.540.040.390.070.590.201.77
MPVN32010.660.090.810.030.320.030.270.100.960.141.32
ELH32010.440.060.600.050.470.000.000.111.020.131.27
LPFL3202.200.6830.910.000.000.3717.050.3314.930.8438.33
Within runBetween runBetween dayBetween site(a)Total
ParameterLevelNMeanSDCV%SDCV%SDCV%SDCV%SDCV%
%N3201.540.3119.910.085.440.2516.310.3119.910.5132.99
H3201.360.2115.380.096.450.1712.510.2518.090.3727.61
LYM#109/LL3200.400.024.390.011.530.000.000.013.340.025.72
N3201.330.032.160.010.830.000.000.043.100.053.87
H3201.600.042.480.010.510.021.300.063.470.074.49
LYM%L32017.520.854.830.000.000.160.890.201.120.885.04
N32018.300.382.090.130.720.000.000.110.600.422.29
H3208.800.232.570.070.760.020.260.070.780.252.81
MON#109/LL3200.150.019.160.016.080.000.000.013.890.0211.67
N3200.490.024.940.012.310.000.000.023.810.036.65
H3200.820.045.280.033.390.000.000.044.870.077.95
MON%L3206.670.609.050.416.130.000.000.182.760.7511.28
N3206.770.334.880.202.890.000.000.121.740.405.93
H3204.490.235.160.163.630.000.000.173.810.337.37
NEU#109/LL3201.430.042.650.000.000.010.690.053.320.064.30
N3204.670.061.210.051.120.000.000.122.520.143.02
H32013.490.120.870.060.470.070.540.362.640.392.87
NEU%L32062.910.991.570.490.770.230.370.000.001.131.79
N32064.140.490.770.250.390.000.000.200.310.590.92
H32074.060.360.480.090.120.000.000.190.250.410.56
EOS#109/LL3200.120.019.620.003.540.015.800.003.330.0112.24
N3200.230.028.040.012.330.000.000.001.280.028.47
H3200.720.045.610.011.940.011.980.011.630.056.47
EOS%L3205.280.499.360.183.340.295.580.183.420.6311.90
N3203.100.258.000.062.030.000.000.072.340.278.57
H3203.930.235.730.040.980.051.390.051.190.246.10
BAS#109/LL3200.100.003.370.001.370.001.030.004.860.016.16
N3200.320.001.460.001.020.000.770.014.000.014.45
H3200.840.011.090.000.000.000.590.033.230.033.46
BAS%L3204.370.061.440.030.700.051.190.102.190.132.97
N3204.370.051.050.020.390.040.840.081.840.102.32
H3204.580.040.820.000.000.010.250.030.760.051.15
NRBC#109/LL3200.170.014.630.001.750.000.000.013.320.015.96
N3200.570.012.220.000.750.000.000.023.080.023.87
H3200.690.022.480.000.700.011.280.023.470.034.51
NRBC%L3207.500.364.840.000.000.070.990.081.120.385.06
N3207.840.162.070.060.710.000.000.050.600.182.27
H3203.770.102.750.000.000.000.000.030.700.112.84
IMG#109/LL3200.070.005.690.001.420.000.000.003.880.017.03
N3200.240.001.650.001.220.000.090.012.480.013.22
H3200.750.011.070.000.410.000.420.022.650.022.92
IMG%L3203.250.082.510.041.150.000.000.010.410.092.79
N3203.320.041.320.020.750.000.000.020.500.051.59
H3204.140.041.080.000.030.000.000.020.540.051.21
TNC109/LL3202.450.041.490.010.490.000.000.083.200.093.57
N3207.850.070.830.060.780.000.000.212.690.232.92
H32018.910.130.680.080.440.120.630.522.730.552.91
Within runBetween runBetween dayBetween site(a)Total
ParameterLevelNMeanSDCV%SDCV%SDCV%SDCV%SDCV%
RET#109/L132065.314.406.731.632.500.300.452.203.365.197.94
RET#109/L2320180.316.263.472.701.501.340.740.850.477.003.88
RET#109/L3320355.788.312.343.591.012.920.820.750.219.542.68
RET%13201.860.126.690.052.450.000.000.073.600.157.99
RET%23205.140.173.400.071.400.040.850.040.690.203.84
RET%332010.040.232.310.100.950.070.680.040.390.262.62
CRC[%]13201.180.086.720.032.540.010.450.053.900.108.19
CRC[%]23203.210.113.470.051.490.030.940.051.550.134.19
CRC[%]33206.230.152.340.060.980.050.850.101.550.193.09
IRF13200.310.027.700.014.090.000.000.012.150.038.98
IRF23200.480.013.120.012.110.000.000.035.610.036.75
IRF33200.530.011.740.012.260.000.000.035.700.036.37
RBC1012/L13203.510.030.750.010.250.000.000.010.280.030.84
RBC1012/L23203.510.020.630.010.420.000.000.010.390.030.85
RBC1012/L33203.540.020.610.010.210.010.150.010.360.030.76
RHCCpg132023.660.582.440.291.220.241.030.241.030.733.09
RHCCpg232017.010.362.110.251.470.120.720.211.210.502.93
RHCCpg332012.520.302.440.262.060.141.130.211.710.473.79

Table 9 - Reproducibility ABX Difftrol - combined sites analysis

HORIBA ABX SAS, FRANCE

VOL 004-20

{23}------------------------------------------------

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

Control materials (ABX Diffrol, ABX Minotrol Retic and BFTROL) were analyzed at four sites using four instruments (one (a) instrument per site). Between site analysis corresponds to between instrument analysis.

{24}------------------------------------------------

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

Table 10 - Reproducibility ABX Minotrol Retic - combined sites analysis

Control materials (ABX Diffrol, ABX Minotrol Retic and BFTROL) were analyzed at four sites using four instruments (one (a) instrument per site). Between site analysis corresponds to between instrument analysis.

Reproducibility - Body Fluids

Reproducibility studies were performed to evaluate the within-run, between-run, between-day, between-instrument, and total precision of the Yumizen H2500 analyzer. Testing was performed using two (2) levels of the body fluid control material (BFTROL). Each level was run in duplicate twice each day for a minimum of 20 days using a single calibration and reagent lot at each of the four (4) test sites. The results were analyzed in accordance with the CLSI EP05-A3 approved quideline and met the acceptance criteria. Results are summarized below.

ParameterLevelNMeanWithin runBetween runBetween dayBetween site(a)Total
SDCV%SDCV%SDCV%SDCV%SDCV%
BFRBC2320777833258.204.190.000.000.000.002297.832.953986.965.13
BFRBC332052211910520.512.01816.660.161142.110.2215093.992.8918452.133.55
BFWBC23202277.933.504.872.150.000.0010.974.8414.396.34
BFWBC332091716.621.815.570.617.780.8532.663.5737.874.14
BFMN#2320675.257.830.620.931.392.072.694.016.099.09
BFMN#33201557.764.990.000.001.761.134.362.819.075.84
BFMN%2320302.097.050.000.000.471.580.180.602.157.25
BFMN%3320170.865.060.000.000.120.720.181.070.895.18

Table 11 - Reproducibility BFTROL - combined sites analysis

HORIBA ABX SAS, FRANCE

{25}------------------------------------------------

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

ParameterLevelNMeanWithin runBetween runBetween dayBetween site(a)Total
SDCV%SDCV%SDCV%SDCV%SDCV%
BFPN#23201607.354.603.262.040.000.008.245.1611.517.21
BFPN#332076116.142.127.480.985.690.7528.453.7534.044.48
BFPN%2320702.092.960.000.000.470.660.180.252.153.05
BFPN%3320830.861.030.000.000.120.150.180.220.891.06

Control materials (ABX Difftrol, ABX Minotrol Retic and BFTROL) were analyzed at four sites using four instruments (a) (one instrument per site). Between site analysis corresponds to between instrument analysis.

{26}------------------------------------------------

b. Linearity:

Linearity - Whole Blood

The study was performed according to CLSI EP06-A2 quide recommendations at one test site, using two (2) instruments:

  • For CBC parameters: two (2) Yumizen H2500 with one different reagent lot on . each instrument.
  • For RET and PLT-Ox parameters, two (2) Yumizen H2500 with one different ● reagent lot on each instrument

For each parameter, the linearity range (commercial or prepared from dilutions of samples) using a minimum of seven (7) concentration levels was analyzed:

  • . One (1) sample underneath the lower limit of the linearity interval
  • At least five (5) samples with concentrations over the entire linearity range .
  • One (1) sample over the upper limit of the linearity interval .

Each level is tested in a minimum of four (4) replicates on two (2) instruments using two (2) reagent lots.

Results are presented in the table below. All results met the predefined acceptance criteria and were determined to be acceptable.

ParameterUnitLinearity range
WBC$10^9/L$( $10^3/\mu L$ )0.06 – 344.50
TNC$10^9/L$( $10^3/\mu L$ )0.06 – 344.50
RBC$10^{12}/L$( $10^6/\mu L$ )0.17-8.63
HGBg/dL0.5 – 25.8
HCT%1.5 – 72.7
PLT$10^9/L$( $10^3/\mu L$ )5 – 2706
PLT-Ox$10^9/L$( $10^3/\mu L$ )7 – 2570
RET#$10^9/L$8 – 1276
NRBC#$10^9/L$0.00 – 21.21

Table 12 - Whole Blood linearity summary

{27}------------------------------------------------

Linearity - Body Fluids

The study was performed according to CLSI EP06-A2 guide recommendations at one test site, using two (2) instruments, with one different reagent lot on each instrument.

For each parameter, the linearity range (prepared from dilutions of samples) using a minimum of seven (7) concentration levels was analyzed:

  • One (1) sample underneath the lower limit of the linearity interval .
  • At least five (5) samples with concentrations over the entire linearity range ●
  • One (1) sample over the upper limit of the linearity interval ●

Each level is tested in a minimum of four (4) replicates on two (2) instruments using two (2) reagent lots.

Results are presented in the table below. All results met the predefined acceptance criteria and were determined to be acceptable.

ParametersUnitLinearity range
BFWBC106/L3 – 11 345
BFRBC106/L1079 – 5 394 633

Table 13 - Body Fluid linearity summary

c. Analytical Specificity/Interference:

Interferences - Whole Blood

The susceptibility of the Yumizen H2500 device to interference in the presence of hemoglobin, triglycerides, bilirubin, cholesterol, elevated RBCs, elevated RBCs, elevated PLTs, microcytic RBCs and other potential interferents was tested in whole blood samples collected in K2EDTA tubes, at four (4) test sites.

Hemoglobin, Lipemia, Bilirubin, Glucose and Yeast

For the Hemoglobin, Lipemia, Bilirubin, Glucose and Yeast studies, interferent levels were tested with samples at two concentration levels, according to CLSI EP7-A3 quidelines.

The potential interfering substance is added to a sample and the bias relative to a control portion of the sample is evaluated ("paired-difference testing"). This bias was compared to the acceptance criteria.

The following parameters were evaluated for each potential interferent:

{28}------------------------------------------------

Table 14 - Whole Blood Parameters - Potential Interferences
ParametersPotential interferent
WBC, HGB, LYM#, MON#, NEU#, EOS#,RET#Hemoglobin
WBC, LYM#, MON#, NEU#, EOS#, HGB,PLT, PLT-OxLipemia (Triglycerides or Cholesterol)
HGB, PLT, PLT-Ox, MPVBilirubin (Total and Direct)
MCVGlucose
PLT, PLT-Ox, MPVYeast

Whole Blood Parameters Potential Interferents

The Yumizen H2500 device were considered not susceptible to interference from Hemoglobin, Lipemia. Bilirubin, Glucose and Yeast, if the absolute difference or % difference criteria were met for each measurand.

Whole Blood Intrinsic Interferences

A study was conducted to evaluate the susceptibility of the Yumizen H2500 device to interference from Intrinsic Interferences in native whole blood specimens, according to CLSI H26-A2 guidelines.

This study utilized a subset of samples tested in the method comparison study, in four (4) test sites. where unique, native whole blood specimens collected in K2EDTA tubes were identified with potential interference analysis. A minimum of 17 specimens were evaluated per interferent.

The Yumizen H2500 device results for a parameter were considered not susceptible to interference from an interferent condition if the visual inspection of the difference plot does not differentiate the two groups (with and without interferent). The visual inspection was used for this analysis because the study compares the Yumizen H2500 device and another method (Sysmex XN-10) in specimens with the interfering condition. Overall, the Yumizen H2500 device demonstrated no interference in measuring WBC, RBC, PLT, MCV, and HGB from elevated WBC, RBC, and PLT measurands in native whole blood specimens.

InterferentLevelConclusion
Hemoglobin1.4 g/dLNo interference was detected on WBC,HGB, Differential and RET parameters.
LipemiaCholesterol 453.6 mg/dLand/orTriglycerides 2621.7 mg/dLNo significant effect was detected on WBCand Differential, HGB, MCH, MCHC, PLT,PLT-Ox and MPV parameters.
Total Bilirubin60 mg/dLNo interference was detected with totalbilirubin on HGB, MCH, MCHC, PLT, PLT-Ox, MPV parameters.
Direct Bilirubin(Conjugated)19 mg/dLConjugated bilirubin has a significanteffect on low HGB level (4 g/dL).No interference was detected withconjugated bilirubin on MCH, MCHC, PLT,
InterferentLevelConclusion
Glucose960 mg/dLPLT-Ox, MPV parameters.No interference detected on MCV, RDW-CV and RDW-SD parameters.
Yeast(Saccharomyces boulardii)50 mg/LNo interference was detected on PLT, PLT-Ox, MPV parameters.
Leukocytosis(Elevated WBCs)>32 109/L WBC>59 109/L WBC300 109/LSignificant effect on low MCVSignificant effect on high MCVNo interference detected on RBC, HGB, MCH, MCHC, RDW-CV, RDW-SD, PLT, PLT-Ox, MPV and NRBC parameters.
Fragile WBC-No interference was detected.
Thrombocytosis(Elevated Platelets)PLT>600 109/LNo interference detected on WBC, WBC Differential, RBC, HGB, HCT, MCV, RDW-CV, RDW-SD, MPV parameters.
Macrothrombocytosis-PLT-Ox and LYM# potentially increased.No interference detected on WBC/WBC differential, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, NRBC, RET parameters.
Platelet Aggregates-PLT, PLT-Ox and LYM# potentially impacted.No interference was detected on WBC/WBC differential, MPV, NRBC and RET.
Dual RBC population-RDW-SD potentially impacted.No interference detected on RBC, HCT, MCV, MCH, MCHC, RDW-CV parameters.
RBC abnormal morphology-MCV, RDW-SD, PLT-Ox and MPV potentially impacted.No interference detected on RBC, HGB, HCT, RDW-CV and RET parameters.
RBC Fragments(schiztocytes)-PLT-Ox and RET potentially impacted.No interference detected on PLT, MPV and RET parameters.
RBC inclusions-PLT-Ox potentially impacted.No interference detected on RBC, HCT, MCV, MCH, MCHC, RDW-CV parameters.
RBC MicrocytosisMCV<75 fLPLT, MPV and RET potentially impacted.No interference was detected on RBC and PLT-Ox parameters.
ErythroblastsNRBC%>10%WBC and Differential potentially impacted.No interference detected on RET parameter.

Table 15 - Potential interferent study summary

{29}------------------------------------------------

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

{30}------------------------------------------------

Interferences - Body Fluids

The susceptibility of the Yumizen H2500 device to interference in the presence of hemoglobin, triglycerides, bilirubin, total protein, mesothelial cells, and other interferents was tested in body fluid samples collected in K2EDTA tubes, at four (4) test sites.

Hemoglobin, Lipemia, Bilirubin, Total Protein and Yeast

For the Hemoglobin, Lipemia, Bilirubin, Total Protein and Yeast studies, interferent levels were tested with samples at two concentration levels, according to CLSI EPT-A3 auidelines.

The potential interfering substance is added to a sample and the bias relative to a control portion of the sample is evaluated ("paired-difference testing"). This bias was compared to the acceptance criteria.

The following parameters were evaluated for each potential interferent:

ParametersPotential interferent
BFWBC, BFRBCHemoglobin
BFWBC, BFRBCLipemia (Triglycerides)
BFWBC, BFRBCBilirubin (Total and Direct)
BFWBC, BFRBCTotal protein
BFWBCYeast

Table 16 - Whole Blood Parameters - Potential Interferents

The Yumizen H2500 device were considered not susceptible to interference from Hemoglobin, Lipemia, Bilirubin and Total Protein, if the absolute difference or % difference criteria were met for each measurand.

Interference from yeast was detected.

Body Fluid Intrinsic Interferences

A study was conducted to evaluate the susceptibility of the Yumizen H2500 device to interference from Intrinsic Interferences in native body fluid specimens, according to CLSI H26-A2 guidelines.

This study utilized a subset of samples tested in the method comparison study, in four (4) test sites, where unique, native body fluid specimens were identified with potential interferents for interference analysis.

The Yumizen H2500 device results for a parameter were considered not susceptible to interference from an interferent condition if the visual inspection of the difference plot does not differentiate the two groups (with and without interferent). The visual inspection was used for this analysis because the study compares the Yumizen H2500 device and another method (Sysmex XN-10) in specimens with the interfering condition. Overall, the Yumizen H2500 device demonstrated no interference in measuring BFWBC and BFRBC from presence of mesothelial cells or malignant cells in native body fluid specimens.

{31}------------------------------------------------

Interferences from the presence of crystals or liposomal particles on BFWBC is well described in literature.

InterferentLevelConclusion
Hemoglobin2 g/dLNo significant effect was detected onBFWBC, BFRBC on three (3) types ofbody fluids.
LipemiaTriglycerides 1593 mg/dLNo significant effect was detected onBFWBC, BFRBC on serous fluids.
Total Bilirubin25.9 mg/dLNo significant effect was detected onBFWBC, BFRBC on serous fluids.
Direct Bilirubin(Conjugated)39.3 mg/dLNo significant effect was detected onBFWBC, BFRBC on serous fluids.
Total Protein50 mg/LNo interference detected on BFWBC,BFRBC parameters on all three (3) typesof body fluids.
Yeast(Saccharomyces boulardii)50 mg/dLSignificant effect was detected on BFWBCon all three (3) types of body fluids.
Crystals-This interference is known anddocumented in bibliography: crystals mayinterfere with BFWBC and generate apollution on the scattergram.
Mesothelial cells-No interference detected with mesothelialcells on native serous fluids.
Malignant cells-No interference detected with mesothelialcells on native serous and CSF fluids.
Liposomal particles-This interference is known anddocumented in bibliography: liposomalparticles may interfere with BFWBC andgenerate a pollution on the scattergram.

d. Traceability, Stability, Expected Values:

Stability - Whole Blood samples:

A total of 14 whole venous blood specimens (collected in K2EDTA) were analyzed on the Yumizen H2500 at three (3) test sites. Following the collection (T0), each specimen was divided in half, with one sample stored at ambient temperature (20-24°C) and the other stored under refrigerated conditions (2-8℃). Testing for stability was performed at regular intervals after (TO) When sufficient volume is available, each time point is tested in duplicate.

Acceptance criteria for each parameter was established for defined time intervals at each storage condition as the difference between the recovery of the parameter at defined time points and time point zero (TO). The acceptance criteria were met for each claimed storage conditions:

{32}------------------------------------------------

Table 18 - Whole Blood - Sample stability Study results
Room temperatureRefrigerated temperature
WHOLE BLOOD24h for CBC/LMNE/NRBC/RET parameters48h for CBC/LMNE/NRBC parameters72h for RET parameters

Sample stability study results Whole Blood -

Stability - Body Fluid samples:

28 body fluid specimens (13 serous, 7 synovial, and 8 cerebrospinal fluids) were analyzed on the analyzer at two (2) test sites. Following the collection (TO), each specimen was stored at ambient temperature (20-24°C). For serous and synovial fluids, testing for stability was performed at 6, 24 and 27 hours after (TO), and for cerebrospinal fluids, testing for stability was performed at three (3), four (4) and 4.5 hours after (TO). The acceptance criteria for sample stability is given as an acceptable maximum bias of the value at T with the value at (T0).

Table 19 – Body Fluid - Sample stability study results
------------------------------------------------------------
Room temperature
Serous fluids24h for BFWBC/BFRBC/BFPN/BFMN parameters
Synovial fluids24h for BFWBC/BFRBC/BFPN/BFMN parameters
CSF4h for BFWBC/BFRBC/BFPN/BFMN parameters

Per established literature, body fluid samples should be stored at room temperature and analyzed within 1 hour of collection.

e. Detection Limits:

Limit of Blank - Whole Blood

The Limit of Blank (LoB) for WBC, TNC, RBC, HGB, HCT, PLT, PLT-Ox and RET# was determined according to CLSI EP17-A2 guide recommendations. Testing was conducted over a minimum of five (5) days using six (6) blank samples, with two (2) different reagent lots on different analyzers, each sample being run 10 repeated times. The results obtained from all samples were used to calculate the LoB for each parameter.

{33}------------------------------------------------

Limit of Detection - Whole Blood

The Limit of Detection (LoD) for WBC, TNC, RBC, HGB, HCT, PLT, PLT-Ox and RET# was determined according to CLSI EP17-A2 quide recommendations. Testing was conducted over a minimum of three (3) days using six (6) low concentration whole blood samples, with two (2) different reagent lots on different analyzers, each sample being run 10 repeated times. The results obtained from all samples were used to calculate the LoD for each parameter.

Limit of Quantitation - Whole Blood

The Limit of Quantitation (LoQ) for WBC, TNC, RBC, HGB, HCT, PLT, PLT-Ox, RET# and RET% was determined according to CLSI EP17-A2 quide recommendations. Testing was conducted over a minimum of three (3) days using at least four (4) low concentration whole blood samples, with two (2) different reagent lots on different analyzers, each sample being run 10 repeated times. The results obtained from all samples were used to calculate the Mean, Total error, and LoQ for each parameter. All results met the predefined acceptance criteria and were determined to be acceptable.

ParameterUnitLoBLoDLoQ
WBC109/L(103/μL)0.050.070.10
TNC109/L(103/μL)0.050.070.10
RBC1012/L(106/µL)0.010.050.20
HGBg/dL0.10.30.8
HCT%0.10.32
PLT109/L(103/μL)247
PLT-Ox109/L(103/μL)2610
RET#109/L(103/μL)2710

Table 20 - Whole Blood - Analytical Sensitivity Summary

Not applicable, ratio result

Limit of Blank - Body Fluid

The Limit of Blank (LoB) for BFWBC and BFRBC was determined according to CLSI EP17-A2 quide recommendations. Testing was conducted over a minimum of five (5) days using six (6) blank samples, with two (2) different reagent lots on different analyzers, each sample being run 10 repeated times. The results obtained from all samples were used to calculate the LoB for each parameter.

{34}------------------------------------------------

Limit of Detection - Body Fluid

The Limit of Detection (LoD) for BFWBC and BFRBC was determined according to CLSI EP17-A2 guide recommendations. Testing was conducted over a minimum of three (3) days using six (6) low concentration body fluid samples, with two (2) different reagent lots on different analyzers, each sample being run 10 repeated times. The results obtained from all samples were used to calculate the LoD for each parameter.

Limit of Quantitation - Body Fluid

The Limit of Quantitation (LoQ) for BFWBC and BFRBC was determined according to CLSI EP17-A2 guide recommendations. Testing was conducted over a minimum of three (3) days using at least four (4) low concentration body fluid samples, with two (2) different reagent lots on different analyzers, each sample being run 10 repeated times. The results obtained from all samples were used to calculate the Mean, Total error, and LoQ for each parameter. All results met the predefined acceptance criteria and were determined to be acceptable.

ParameterUnitLoBLoDLoQ
BFWBC106/L245
BFRBC106/L50010001500

Table 21 - Body Fluids - Analytical Sensitivity Summary

f. Assay Cut-Off:

Not applicable

q. Accuracy (Instrument):

Not applicable

h. Carry-over:

Carry-over - Whole Blood

Carryover was evaluated by assaying whole blood K2EDTA samples with high WBC, RBC, HGB, PLT, PLT-Ox, RET% and RET# counts three consecutive times (high 1, high 2, high 3) followed immediately by testing samples with low target values consecutively three times (low 4, low 5, low 6) in accordance with CLSI H26-A2 guide. Three sets of carry-over sequences were run for each parameter on three analyzers at one (1) test site.

The percentage of carryover is calculated using the formula below:

{35}------------------------------------------------

low 4 – low 6 Ct % = × 100 high 3 – low 6

Where:

is the first low sample analyzed after the high sample runs low 4:

is the last low sample analyzed after the high sample runs low 6:

high 3: is the last high sample analyzed

All carry-over results are within specifications for the Yumizen H2500 systems.

Carry-over - Body Fluids

Carryover was evaluated by assaving body fluid samples (serous and synovials) with high BFWBC and BFRBC counts three consecutive times (high 1, high 2, high 3) followed immediately by testing samples with low target values consecutively three times (low 4, low 5, low 6) in accordance with CLSI H26-A2 quide. Three sets of carry-over sequences were run for each parameter on three analyzers at one (1) test site.

The percentage of carryover is calculated using the formula below:

low 4 – low 6 Ct % = high 3 – low 6

Where:

low 4: is the first low sample analyzed after the high sample runs

  • is the last low sample analyzed after the high sample runs low 6:
    high 3: is the last high sample analyzed

All carry-over results are within specifications for the Yumizen H2500 systems.

{36}------------------------------------------------

B. Comparison Studies

a. Method Comparison with Predicate Device:

Method comparison studies were conducted to assess the performance of the Yumizen H2500 device, compared to the predicate device (Sysmex XN-Series, XN-10, K112605), according to EP9-A3 quidelines.

Whole Blood

A method comparison study was conducted to assess the performance of the Yumizen H2500 device when compared to the predicate device, Sysmex XN-Series (XN-10, XN-20) (K112605). A total of 969 venous and/or capillary specimens collected in K2EDTA from pediatric (≤21 years) and adult subjects including a wide variety of disease states (clinical conditions) were tested across four (4) clinical sites (two (2) US sites and two (2) European sites).

Venous and/or capillary whole blood leftover specimens were collected in K2EDTA tubes from a wide range of demographics (age and sex) and disease states (clinical conditions). In total, there were 143 specimens collected from subjects with one or more known medical conditions. Study sites aimed to cover the target assay reportable range for the parameters. A maximum of 10% samples were permitted to be contrived to cover the entire target assay reportable range.

Each specimen was tested within six (6) hours from the time of collection in two (2) replicates on the Yumizen H2500 and one (1) replicate on the Sysmex XN-10 System, when sufficient volume was available. Specimens were tested on the Yumizen H2500 and the Sysmex XN-Series (XN-10, XN-20) within three (3) hours of each other.

A Passing-Bablok regression analysis was performed. Bias at medical decision points were also evaluated for all sites combined. All results were within the predefined acceptance criteria and found to be acceptable. Overall, the Yumizen H2500 demonstrated comparable performance to the predicate device, Sysmex XN-Series (XN-10, XN-20) (K112605) in an intended use population in a clinical laboratory setting.

{37}------------------------------------------------

A summary of the results is presented in the table below:

Parameter(unit)NSample Ranger(95% CI)Slope(95% CI)Intercept(95% CI)
WBC (109/L)(103/μL)9360.19 - 2320.996[0.996;0.997]1.012[1.007 ; 1.017]0.045[0.014 ; 0.076]
TNC (109/L)(103/μL)9370.19 - 2320.996[0.996;0.997]1.006[1.00 ; 1.01]0.080[0.05 ; 0.11]
RBC (1012/L)(106/μL)9410.54 - 7.310.997[0.997;0.998]1.006[1.00 ; 1.01]-0.017[-0.03 ; 0.01]
HGB (g/dL)9412 - 22.10.998[0.998;0.998]1.000[1 ; 1]0.000[0 ; 0]
HCT (%)9415.8 - 65.30.985[0.983;0.987]1.036[1.024 ; 1.048]-1.4[-1.9 ; -1.0]
MCV (fL)94060.6 - 1400.893[0.88;0.906]1.015[0.983 ; 1.052]-2.119[-5.381 ; 0.804]
MCH (pg)94018.6 - 46.10.976[0.973;0.979]1.000[1.000 ; 1.022]0.000[-0.662 ; 0.000]
MCHC (g/L)940278 - 3790.184[0.106;0.238]0.545[0.5 ; 0.6]149.500[132.0 ; 164.5]
RDW-CV (%)92410.7 - 30.40.872[0.857;0.888]0.868[0.833 ; 0.901]2.316[1.860 ; 2.843]
PLT (109/L)(103/µL)9399 - 17800.995[0.995;0.996]0.977[0.970 ; 0.984]2.841[1.447 ; 3.996]
PLT-Ox(109/L)(103/µL)80110 - 17800.988[0.986;0.99]0.962[0.949 ; 0.974]3.570[1.503 ; 5.583]
LPF (%)7910.4 - 40.90.872[0.856;0.889]0.833[0.786 ; 0.888]0.067[-0.100 ; 0.222]
MPV (fL)8777.2 - 13.30.901[0.888;0.913]1.000[0.976 ; 1.000]-1.000[-1.000 ; -0.763]
NRBC% (%)9300 - 66.70.927[0.918;0.936]0.609[0.455 ; 0.73]0.000[0.000 ; 0.000]
NRBC#(109/L)(103/μL)9300.00 - 9.080.951[0.945;0.957]0.636[0.466 ; 0.736]0.000[0.000 ; 0.000]
Parameter(unit)NSample Ranger(95% CI)Slope(95% CI)Intercept(95% CI)
LYM% (%)8210.7 - 99.20.985[0.983;0.987]0.990[0.984 ; 0.996]0.685[0.501 ; 0.875]
LYM# (109/L)(103/µL)8210.05 - 204.000.993[0.993;0.994]1.005[0.995 ; 1.015]0.064[0.047 ; 0.080]
MON% (%)8430.1 - 87.90.938[0.93;0.946]1.008[1.000 ; 1.042]0.402[0.122 ; 0.600]
MON#(109/L)(103/µL)8430.00 - 98.900.955[0.949;0.961]1.128[1.102 ; 1.152]-0.028[-0.041 ; -0.015]
NEU% (%)8340.2 - 95.60.984[0.982;0.986]0.988[0.982 ; 0.995]0.373[-0.051 ; 0.769]
NEU# (109/L)(103/µL)8340.0 - 75.30.968[0.964;0.972]1.014[1.008 ; 1.021]0.013[-0.006 ; 0.031]
EOS% (%)9280.0 - 78.10.984[0.982;0.986]0.918[0.902 ; 0.933]0.345[0.310 ; 0.385]
EOS# (109/L)(103/μL)9280.00 - 23.300.976[0.973;0.979]1.000[0.984 ; 1.000]0.020[0.020 ; 0.021]
BAS# (109/L)(103/µL)8970.00 - 12.300.902[0.89;0.914]1.000[0.818 ; 1.000]-0.010[-0.010 ; -0.006]
BAS% (%)8970.0 - 13.10.712[0.679;0.744]1.000[0.931 ; 1.0]-0.200[-0.200 ; -0.1]
IMG% (%)8340 - 370.867[0.85;0.884]0.733[0.674 ; 0.778]0.080[0.050 ; 0.129]
IMG# (109/L)(103/µL)8340.00 - 72.400.982[0.979;0.984]0.854[0.802 ; 0.901]0.001[-0.002 ; 0.002]
RET% (%)9190.31 - 20.200.901[0.888;0.913]1.020[0.997 ; 1.045]-0.140[-0.182 ; -0.100]
RET# (109/L)(103/µL)91410 - 5650.925[0.916;0.934]0.997[0.968 ; 1.026]-3.355[-5.626 ; -1.907]
CRC (%)9190.1 - 12.70.939[0.931;0.947]0.986[0.961 ; 1.012]-0.067[-0.099 ; -0.034]
IRF9140.00 - 0.540.778[0.752;0.803]0.833[0.807 ; 0.868]0.023[0.017 ; 0.028]
RHCC (pg)91312.7 - 49.10.654[0.616;0.691]1.226[1.159 ; 1.294]-9.781[-12.089 ; -7.528]
ParameterNSample Ranger(95% CI)Slope(95% CI)Intercept(95% CI)
BFWBC(106/L)15922 - 89400.923[0.9;0.946]0.89[0.86 ; 0.91]26.37[18.62 ; 39.67]
BFRBC(106/L)751500 - 35200000.999[0.999;0.999]1.04[1.0 ; 1.07]-370.53[-580.7 ; 61.48]
BFMN% (%)1439.0 - 98.00.967[0.957;0.978]0.90[0.88 ; 0.91]9.75[8.13 ; 11.05]
BFMN#(106/L)14310 - 31400.943[0.925;0.961]1.09[1.01 ; 1.25]23.05[9.94 ; 35.66]
BFPN% (%)1432.0 - 91.00.967[0.957;0.978]0.90[0.88 ; 0.92]0.64[-0.29 ; 2.25]
BFPN#(106/L)1431 - 74300.911[0.883;0.938]0.79[0.77 ; 0.82]5.04[1.68 ; 9.17]

Table 22 - Whole Blood method comparison data summary

HORIBA ABX SAS, FRANCE

VOL 004-35

{38}------------------------------------------------

VOL 004 : 510(k) Safety and Effectiveness Summary Yumizen H2500 Traditional 510(k): Supplement to Original Submission K232946

{39}------------------------------------------------

Body Fluids

A method comparison study was conducted to assess the performance of the Yumizen H2500 device when compared to the predicate device. Sysmex XN-Series (XN-10. XN-20) (K112605). A total of 427 residual body fluid specimens from pediatric (≤21 years) and adult subjects including a wide variety of disease states (clinical conditions) were tested across three (3) clinical sites (two (2) US sites and one (1) European site).

The study was conducted using 174 synovial fluids, 138 serous fluids and 115 cerebrospinal fluids from a wide range of demographics (age and sex) and disease states (clinical conditions). Study sites aimed to cover the target assay reportable range for the parameters.

Each specimen was analysed as soon as possible from the time of collection in two (2) replicates on the Yumizen H2500 and one (1) replicate on the Sysmex XN-10 System, when sufficient volume was available. Specimens were tested on the Yumizen H2500 and the Sysmex XN-Series (XN-10, XN-20) within two (2) hours of each other.

A Passing-Bablok regression analysis was performed. Bias at medical decision points were also evaluated for all sites combined. All results were within the predefined acceptance criteria and found to be acceptable. Overall, the Yumizen H2500 demonstrated comparable performance to the predicate device, Sysmex XN-Series (XN-10, XN-20) (K112605) in an intended use population in a clinical laboratory setting.

A summary of the results is presented in the tables below:

Table 23 - Synovial Body Fluid method comparison data summary

{40}------------------------------------------------

Table 24 – Serous Body Fluid method comparison data Summary
ParameterNSample Ranger(95% CI)Slope(95% CI)Intercept(95% CI)
BFWBC(106/L)13514 - 93700.965[0.954;0.977]0.91[0.89 ; 0.94]14.89[6.58 ; 23.34]
BFRBC(106/L)871610 - 45900000.999[0.999;1]1.04[1.02 ; 1.08]31.75[-231.53 ; 262.80]
BFMN% (%)13210.0 - 99.00.965[0.954;0.977]0.90[0.85 ; 0.94]11.66[7.78 ; 15.94]
BFMN#(106/L)13223 - 45900.954[0.939;0.969]1.00[0.96 ; 1.04]13.31[7.09 ; 21.67]
BFPN% (%)1321.0 - 90.00.965[0.954;0.977]0.90[0.85 ; 0.95]-1.58[-2.69 ; -0.67]
BFPN#(106/L)1321 - 75600.972[0.963;0.981]0.78[0.74 ; 0.82]-1.36[-2.82 ; 0.04]

Table 24 – Serous Body Fluid method comparison data summary

Table 25 – CSF Body Fluid method comparison data summary

ParameterNSample Ranger(95% CI)Slope(95% CI)Intercept(95% CI)
BFWBC(106/L)1095 - 84600.980[0.973;0.988]0.99[0.95 ; 1.02]5.14[3.87 ; 7.25]
BFRBC(106/L)711650 - 46900001.000[1;1]1.07[1.04 ; 1.08]15.80[-533.69 : 487.19]
BFMN% (%)771 - 1000.816[0.738;0.887]0.94[0.88 ; 1.04]10.87[6.31 ; 14.89]
BFMN#(106/L)776 - 29800.693[0.566;0.8]1.11[1.00 ; 1.18]6.36[2.36 ; 11.25]
BFPN% (%)770 - 990.817[0.739;0.888]0.94[0.88 ; 1.05]-5.36[-13.33 ; -0.86]
BFPN#(106/L)770 - 59400.863[0.805;0.918]0.87[0.81 ; 0.99]-0.49[-2.46 ; 0.58]

{41}------------------------------------------------

b. Matrix Comparison:

Comparability between sampling types (capillary / venous)

The study was performed to demonstrate comparability between capillary whole blood samples and venous whole blood samples on the Yumizen H2500 device.

A total of 84 normal and pathological paired capillary and venous whole blood specimens were prospectively collected. Paired specimens collected from the same individuals were assayed in duplicate on the Yumizen H2500 device at one (1) clinical site. The clinical site aimed to enroll subject samples that covered all relevant medical decision points and were representative of the proposed analytical measurement ranges to the extent possible.

Bias was estimated at three points for each parameter: the low end of the distribution of observations, the mid-point, and the high end of the distribution, based on EP9-A3 guidelines. Acceptance criteria were met for all parameters at all levels. Therefore, the study results support the claim of using the two specimen types for measurement on the Yumizen H2500 device.

Comparability between body fluid anticoaqulants

The study was performed to demonstrate comparability between body fluid anticoagulants on the Yumizen H2500 device.

A total of 9 synovial body fluids without anticoagulant, 39 synovial body fluids with K2EDTA anticoagulant, 92 synovial body fluids with Lithium Heparin anticoagulant and 34 synovial body fluids with Sodium Heparin anticoagulant were run on the Yumizen H2500 device at three (3) clinical sites. In the Bland-Altman difference plots, a visual examination of each anticoagulant subgroup of samples was performed. No difference linked to anticoaqulant was observed on these graphs. No significant effect linked to the matrix was observed on these data.

A total of 82 serous body fluids without anticoagulant and 56 serous body fluids with K2EDTA were run on the Yumizen H2500 device at three (3) clinical sites.

In the Bland-Altman difference plots, a visual examination of each anticoagulant subgroup of samples was performed. No difference linked to anticoagulant was observed on these graphs. No significant effect linked to the matrix was observed on these data.

{42}------------------------------------------------

Comparability between analytical modes

The study was performed to demonstrate equivalency between Yumizen H2500 available analytical modes:

Available Analytical ModesCorresponding Analytical Cycle
DIFDIF
DIRDIR
RBC_PLTORBC_PLTO
DIF_LVDIF_LV
CBCDIF
RETDIR
CBRDIR

Table 26 - List of Yumizen H2500 Modes and cycles

There are seven (7) analytical modes based on four (4) different cycles only. The equivalency between those four (4) cycles: DIF, RBC PLTO and DIF LV was verified.

A total of 166 normal and pathological residual whole blood specimens were assayed in duplicate in the DIR and DIF modes on one (1) Yumizen H2500 at one (1) clinical site to compare performance between analytical modes, and the results were analyzed for all applicable parameters.

A total of 172 normal and pathological residual whole blood specimens were assayed in duplicate in the RBC_PLTO and DIF modes one (1) Yumizen H2500 at one (1) clinical site to compare performance between analytical modes, and the results were analyzed for all applicable parameters.

A total of 187 normal and pathological residual whole blood specimens were assayed in duplicate in the DIF_LV and DIF modes on one (1) Yumizen H2500 at one (1) clinical site to compare performance between analytical modes, and the results were analyzed for all applicable parameters.

Study site aimed to cover the target assay reportable range for the parameters.

Bias was estimated at three points for each parameter: the low end of the distribution of observations, the mid-point, and the high end of the distribution, based on EP9-A3 guidelines. Acceptance criteria were met for all parameters at all levels.

The results show comparable performance characteristics for all Yumizen H2500 modes.

{43}------------------------------------------------

Comparability mode to mode

The study was performed to demonstrate comparability between automatic rack mode versus manual (STAT) mode with whole blood samples on the Yumizen H2500 device. A total of 83 normal and pathological residual whole blood samples were run in automatic rack mode version manual (STAT) mode on the Yumizen H2500 device at one (1) clinical site. The clinical site aimed to enroll subject samples that covered all relevant medical decision points and were representative of the proposed analytical measurement ranges to the extent possible.

Bias was estimated at three points for each parameter: the low end of the distribution of observations, the mid-point, and the high end of the distribution, based on EP9-A3 guidelines. Acceptance criteria were met for all parameters at all levels.

The results show comparable performance characteristics for all Yumizen H2500 modes.

{44}------------------------------------------------

c. Clinical Sensitivity:

Flagging analysis was conducted to evaluate the flagging capabilities of the Yumizen H2500 using the flagging results obtained from the samples used in the method comparison study, from four (4) clinical sites, based on CLSI H20-A2 quidelines. A total of 456 residual normal (no flags, marked as negative) and abnormal (contained flags, marked as positive) whole blood samples were tested and compared to the reference (predicate or manual count results, depending on the parameter).

The results were classified according to the following categories:

  • TN (True negative): Yumizen H2500 result and reference result are both negative (normal sample).
  • Yumizen H2500 result and reference result are both positive TP (True positive): (abnormal sample).
  • FN (False Negative): Yumizen H2500 gives a neqative result (normal sample) whereas reference gives a positive result (abnormal sample).
  • FP (False Positive): Yumizen H2500 gives a positive result (abnormal sample) whereas reference gives a negative result (normal sample).

The estimated sensitivity, estimated specificity and estimated efficiency were calculated for each pathology and abnormality, using the following calculations:

Estimated sensitivity (%): TP / (TP+FN) x 100 Estimated specificity (%): TN / (TN+FP) x 100 Estimated efficiency (%): (TP+TN) / (TP+FN+FP+TN) x 100

The overall flagging capabilities of the Yumizen H2500 device met the predefined acceptance criteria for both sensitivity and specificity.

A summary of the results is presented in the table below:

Category ofAbnormalitiesNTPFPENTNSensitivity(95%CI)Specificity(95%CI)Efficiency(95%CI)
Any MorphologicalFlag455165414020980.5%(74.5% to 85.3%)83.6%(78.5% to 87.7%)82.2%(78.4% to 85.6%)
Any DistributionalAbnormality45630692711491.9%(88.5% to 94.4%)92.7%(86.7% to 96.1%)92.1%(89.2% to 94.4%)
Any MorphologicalFlag and/orDistributionalAbnormality455307113410390.0%(86.4% to 92.8%)90.4%(83.5% to 94.5%)90.1%(87.0% to 92.7%)

Table 27 - Clinical Sensitivity Study acceptance criteria

{45}------------------------------------------------

C. Clinical Studies

Not applicable

D. Clinical Cut-Off

Not applicable.

{46}------------------------------------------------

E. Expected Values/Reference Range:

Whole Blood - Adult samples

The adult reference intervals study was conducted at two (2) US clinical sites to establish adult reference intervals for the Yumizen H2500 device. The study followed CLSI EP28-A3c quidelines and was performed using whole blood samples collected from 240 (120 male and 120 female) apparently healthy adults (>21 years). Each of the samples was analyzed on one Yumizen H2500 device for all parameters. The reference intervals were calculated for each parameter. The lower and upper limits of the 95% reference intervals were determined based on the 2.5th and 97.5th percentiles of all valid measurements for each sex group, respectively.

ParameterTable 28 - Whole blood - Adult reference intervalsFemaleMale
WBC (109/L) (103/μL)4.71-12.034.00-10.74
RBC (1012/L) (106/μL)3.93-5.354.08-6.06
HGB (g/dL)10.8-15.011.8-17.2
HCT (%)33.8-46.235.9-50.6
MCV (fL)77.8-95.869.5-96.5
MCH (pg)25.2-32.022.3-32.2
MCHC (g/L)313-344313-347
RDW-SD (fL)30.9-47.230.9-48.0
RDW-CV (%)12.0-16.012.7-19.1
PLT (109/L) (103/μL)183-402144-379
MPV (fL)8.0-11.27.3-11.6
TNC (109/L) (103/μL)4.71-12.034.00-10.74
NRBC# (109/L) (103/μL)0.00-0.000.00-0.00
NRBC% (%)0.0-0.00.0-0.0
LYM# (109/L) (103/μL)1.11-3.781.21-3.88
LYM% (%)18.6-49.816.5-46.5
MON# (109/L) (103/μL)0.27-0.970.28-1.04
MON% (%)4.6-11.75.1-12.3
NEU# (109/L) (103/μL)1.96-8.391.88-6.62
NEU% (%)38.0-72.341.4-75.7
EOS# (109/L) (103/μL)0.04-0.470.03-0.43
EOS% (%)0.6-6.00.5-5.2
BAS# (109/L) (103/μL)0.00-0.070.00-0.08
BAS% (%)0.0-1.20.0-1.1
IMG# (109/L) (103/μL)0.01-0.150.00-0.19
IMG% (%)0.1-2.00.1-2.5
PLT-Ox (109/L) (103/μL)175-420134-370
LPF (%)1.3-8.01.0-11.5
RET# (109/L) (103/μL)26-11827-137
RET% (%)0.55-2.620.54-2.39
CRC (%)0.50-2.350.53-2.49
IRF0.00-0.160.01-0.20
RHCC (pg)24.7-35.225.1-36.8

Table 28 - Whole blood – Adult reference intervals

{47}------------------------------------------------

Whole Blood - Pediatric samples

The pediatric reference intervals study was conducted at two (2) US clinical sites to establish pediatric reference intervals for the Yumizen H2500 device. The study followed CLSI EP28-A3c guidelines and was performed using pediatric venous or capillary whole blood samples collected from at least 80 (approx. 40 male and 40 female) apparently healthy neonates, infants, children and adolescents (<22 years). Each of the samples was analyzed on one Yumizen H2500 device for all parameters. The Robust method is used to determine the most appropriate weighed mean and calculate the desired reference interval of all valid measurements for each age group, respectively.

ParameterNeonateInfantChildAdolescent
WBC (109/L) (103/μL)4.28-18.625.11-13.991.87-12.543.40-10.40
RBC (1012/L) (106/µL)2.21-6.433.59-5.603.82-5.214.14-5.54
HGB (g/dL)6.1-20.89.4-14.510.7-14.212.2-16.2
HCT (%)19.9-65.029.2-44.931.7-44.037.6-48.6
MCV (fL)78.8-119.970.9-91.475.3-93.680.5-97.9
MCH (pg)24.9-38.722.5-30.224.5-31.126.4-32.4
MCHC (g/L)306-334312-335307-353312-343
RDW-SD (fL)43.8-83.926.5-51.231.6-45.733.1-46.5
RDW-CV (%)12.0-22.311.3-18.712.1-16.812.1-15.7
PLT (109/L) (103/μL)55-564123-610175-452147-398
MPV (fL)7.7-12.46.6-11.57.6-10.67.4-11.5
TNC (109/L) (103/μL)4.10-19.115.01-13.891.80-12.592.17-11.5
NRBC# (109/L) (103/μL)0.00-0.000.00-0.000.00-0.002.17-11.39
NRBC% (%)0.0-0.00.0-0.00.0-0.00.00-0.00
LYM# (109/L) (103/μL)0.17-6.561.70-10.260-5.630.0-0.0
LYM% (%)7.1-58.326.2-100.022.9-66.50.0-4.49
MON# (109/L) (103/μL)0-3.290-1.910.21-0.960.14-0.93
MON% (%)0-24.30-17.32.98-13.41.49-14.1
NEU# (109/L) (103/μL)0-11.320-5.540-6.030.64-6.93
NEU% (%)17.6-75.90-53.319.6-63.425.8-90.1
EOS# (109/L) (103/μL)0-0.640-0.560-0.960-0.23
EOS% (%)0-6.50-5.60-10.90-3.2
BAS# (109/L) (103/μL)0-0.090-0.090-0.080-0.44
BAS% (%)0-0.80-0.80-1.10-7.0
IMG# (109/L) (103/μL)0-0.300-0.110-0.060-0.06
IMG% (%)0.06-2.220-0.960.10-0.590-0.78
PLT-Ox (109/L) (103/μL)0-61096-593177-433142-382
LPF (%)0-11.50-6.30-4.70-9.4
RET# (109/L) (103/μL)0-38816-9118-7612-90
RET% (%)0-9.000.18-2.180.39-1.680.22-1.92
CRC (%)0-9.080.24-1.690.33-1.410.24-1.77
IRF0-0.420-0.160-0.140-0.55
RHCc (pg)24.2-45.920.4-32.123.6-33.416.0-47.5

Table 29 - Whole blood - Pediatric reference interval

{48}------------------------------------------------

Body Fluids

The body fluid reference intervals study was conducted at two (2) clinical US sites to establish body fluid reference intervals for the Yumizen H2500 device. The study followed CLSI EP28-A3c guidelines and was performed using body fluid samples (Synovial, serous and cerebrospinal fluids) collected from at least 100 apparently healthy men and women. Each of the samples was analyzed on one Yumizen H2500 device for all parameters. The reference intervals were calculated for each parameter. The Robust method is used to determine the most appropriate weighed mean and calculate the desired reference interval of all valid measurements for each body fluid type, respectively.

ParameterSynovialSerousCSF
BFWBC (106/L)0-19020-9510-301
BFRBC (106/L)0-262750-138130-17679
BFPN# (106/L)0-13580-3840-188
BFPN% (%)0-960-530-95
BFMN# (106/L)0-8640-6720-217
BFMN% (%)0-1000-1000-100
Table 30 - Body fluid - Reference interval

F. Other Supportive Instrument Performance Characteristics Data

Not applicable.

VIII. Proposed Labeling

The labeling is written as per the recommendations given in standard ISO 15223-1 and EN 18113-2. It takes into account the requirements of 21 CFR Part 809.10.

IX. Conclusion

As per 21CFR Part 807.92(b)(3), the nonclinical tests demonstrate that the Yumizen H2500 devices are as safe, as effective, and perform as well as the predicate device.

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”