K Number
K230662

Validate with FDA (Live)

Date Cleared
2023-10-26

(230 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

In addition, these gloves were tested for use with chemotherapy drugs, Fentanyl Citrate, and select other drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permention by Chemotherapy Drugs.

Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, blue colored, nitrile, and tested for use with chemotherapy drugs And Fentanyl Citrate. The design of the subject device is addressing the standards as ASTM D6124, ASTM D5151, ASTM D6319, ASTM D6978-05. The gloves have a minimum length of 270mm. The Subject device is nonsterile.

AI/ML Overview

Below are the acceptance criteria and the study proving the device meets them, based on the provided text:

1. A table of acceptance criteria and the reported device performance:

TestAcceptance CriteriaReported Device PerformanceResult
Biocompatibility
Skin Sensitization PotentialUnder the conditions of the testing, not a sensitizer.Primary irritation indexes of polar and non-polar test group were both 0%.Pass
Irritation PotentialUnder the conditions of the testing, not an irritant.Skin sensitization rates of polar and non-polar extract group were both 0%.Pass
In Vitro Cytotoxicity PotentialUnder the conditions of the testing, non-cytotoxic.Cell viability of the 100% test article extract was 86.2%.Pass
Physical Performance
Residual Powder on Medical GlovesPowder residue limit of 2.0 mg/glove (ASTM D6124-06)1.9 mg/glovePass
Detection of HolesAQL: 2.5 (ISO 2859-1) (ASTM D5151-19)0 gloves of failurePass
Physical Dimensions (ASTM D6319-19)
Length≥270 mm for all sizes (XS, S, M, L, XL, XXL)Range: 287-301 mm (across sizes)Pass
WidthXS: 70±10mm; S: 80±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mm; XXL: 130±10mmXS: 78-80mm; S: 84-86mm; M: 96-97mm; L: 105-107mm; XL: 113-115mm; XXL: 122-124mmPass
Thickness (Finger)≥0.05 mm0.082-0.185 mmPass
Thickness (Palm)≥0.05 mm0.065-0.123 mmPass
Physical Properties: Before Aging (ASTM D6319-19 & ASTM D412-16)
Tensile Strength≥14 MPa19.9-28.9 MPaPass
Ultimate Elongation≥500%540-580%Pass
Physical Properties: After Accelerated Aging (ASTM D6319-19 & ASTM D412-16 & ASTM D573-04)
Tensile Strength≥14 MPa20.2-29.1 MPaPass
Ultimate Elongation≥400%508-528%Pass
Permeation Characteristics (ASTM D6978-05)
Carmustine (3.3 mg/ml)Not explicitly stated as acceptance criteria, but listed as having a minimum breakthrough time in the indications for use.21.1 (24.2, 21.1, 25.3) min. (Note: Warning - Do not use)N/A
Thiotepa (10 mg/ml)Not explicitly stated as acceptance criteria, but listed as having a minimum breakthrough time in the indications for use.74.9 (74.9, 96.7, 79.0) min. (Note: Warning - Do not use)N/A
Other Chemotherapy Drugs and Fentanyl CitrateGenerally, Breakthrough Detection Time > 240 minutes for most tested drugs.> 240 minutes for most listed drugs (see full list in original document)Pass

2. Sample size used for the test set and the data provenance:

  • Sample size: The document does not explicitly state the sample size used for each specific test (e.g., number of gloves tested for permeation, holes, etc.). However, it refers to standard test methods like ASTM D6319-19 which would define typical sample sizes for such evaluations.
  • Data provenance: The data is generated from non-clinical (bench) testing performed by the manufacturer or a contracted lab. The country of origin of the data is not specified beyond the manufacturer's address in Hebei, P.R. China. The data is prospective in nature, as it was generated to demonstrate device performance for regulatory submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as the document describes non-clinical (bench) testing of physical properties and chemical permeation, not a study involving human expert interpretation for establishing ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not applicable as the document describes non-clinical (bench) testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable as the document describes non-clinical (bench) testing of a medical glove, not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

This information is not applicable as the document describes non-clinical (bench) testing of a medical glove, not an algorithm.

7. The type of ground truth used:

The "ground truth" for the non-clinical tests is established by industry standards and defined acceptance criteria. For example:

  • Physical dimensions: Measured values compared against specified ranges (e.g., length ≥270mm).
  • Physical properties (tensile strength, elongation): Measured values compared against minimum performance requirements defined in ASTM standards.
  • Freedom from holes: Tested according to ASTM D5151-19, with acceptance based on AQL (Acceptable Quality Limit) of 2.5.
  • Residual powder: Measured amount compared against a limit of < 2.0 mg/glove.
  • Biocompatibility: Results of standardized biological evaluations (ISO 10993-10, ISO 10993-5) indicating non-sensitization, non-irritation, and non-cytotoxicity.
  • Chemotherapy drug permeation: Breakthrough detection time measured according to ASTM D6978-05, compared against a maximum testing time of 240 minutes.

8. The sample size for the training set:

This information is not applicable as the document describes non-clinical (bench) testing, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established:

This information is not applicable as the document describes non-clinical (bench) testing, not a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 26, 2023

Zhonghong Pulin Medical Products Co.,Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608,No.738,Shangcheng Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K230662

Trade/Device Name: Nitrile Exam Glove, Extended cuff, Tested For Use With Chemotherapy Drugs, Fentanyl Citrate, And Select Other Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QDO Dated: September 26, 2023 Received: September 26, 2023

Dear Boyle Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Allan Guan -S

For Bifeng Qian, M.D., Ph.D. Assistant Director

{2}------------------------------------------------

DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K230662

Device Name

Nitrile Exam Glove, Extended cuff, Tested For Use With Chemotherapy Drugs, Fentanyl Citrate, And Select Other Drugs

Indications for Use (Describe)

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

In addition, these gloves were tested for use with chemotherapy drugs, Fentanyl Citrate, and select other drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permention by Chemotherapy Drugs.

Chemotherapy DrugConcentrationBreakthrough Detection Time (minutes)
Bendamustine HCl5 mg/ml (5,000 ppm)> 240
Bleomycin Sulfate15 mg/ml (15,000 ppm)> 240
Busulfan6 mg/ml(6,000 ppm)> 240
Carboplatin10.0 mg/ml (10,000 ppm)> 240
Carfilzomib2 mg/ml (2,000 ppm)> 240
Carmustine3.3 mg/ml (3,300 ppm)21.1 (24.2, 21.1, 25.3)
Cetuximab2 mg/ml (2,000 ppm)> 240
Cisplatin1 mg/ml (1000 ppm)> 240
Cladribine1 mg/ml (1000 ppm)> 240
Cyclophosphamide20 mg/ml (20.000 ppm)> 240
Cytarabine(Cytosine)10 mg/ml(10,000 ppm)> 240
Cytovene(Ganciclovir)10 mg/ml(10,000 ppm)> 240
Dacarbazine10 mg/ml(10,000 ppm)> 240
Daunorubicin HCl5 mg/ml (5,000 ppm)> 240
Decitabine5 mg/ml(5,000 ppm)> 240
Docetaxel10 mg/ml (10.000 ppm)> 240
Doxorubicin HCl2 mg/ml (2,000 ppm)> 240
Epirubicin HCI (Ellence)20 mg/ml (20,000 ppm)> 240
Etoposide20 mg/ml (20,000 ppm)> 240
Fludarabine25.0 mg/ml (25,000 ppm)> 240
Fluorouracil50 mg/ml,(50,000 ppm)> 240
Fulvestrant50 mg/ml(50,000 ppm)> 240
Gemcitabine38 mg/ml (38,000 ppm)> 240
Idarubicin HCL1.0 mg/ml (1,000 ppm)> 240
lfosfamide50 mg/ml (50,000 ppm)> 240
Irinotecan HCI20 mg/ml (20,000 ppm)> 240
Mechlorethamine HCl1 mg/mI(1,000 ppm)> 240
Melphalan5 mg/ml(5000 ppm)> 240
Methotrexate25 mg/ml (25,000 ppm)> 240
Mitomycin C0.5 mg/ml (500 ppm)> 240
Mitoxantrone HCL2 mg/ml (2,000 ppm)> 240
Oxaliplatin5 mg/ml (5,000 ppm)> 240
Paclitaxel6 mg/ml (6,000 ppm)> 240
Paraplatin10 mg/ml (10,000 ppm)> 240
Pemetrexed25 mg/ml (25,000 ppm)> 240
Raltitrexed0.5 mg/ml (500 ppm)> 240
Rituximab10 mg/ml (10,000 ppm)> 240
Temsirolimus25 mg/ml (25,000 ppm)> 240
Thiotepa10 mg/ml(10,000 ppm)74.9 (74.9, 96.7, 79.0)
Topotecan1 mg/ml (1,000 ppm)> 240
Trisenox (Arsenic Trioxide)1 mg/ml (1,000 ppm)> 240
Vinblastine,1 mg/ml (1,000 ppm)> 240
Vincristine sulfate (Oncovin)1 g/ml (1,000 ppm)> 240
Vinorelbine10 mg/ml(10,000 ppm)> 240
Fentanyl Citrate and other DrugsConcentrationBreakthrough Detection Time (minutes)
Chloroquine50 mg/ml (50,000 ppm)> 240
Cyclosporin A100.0 mg/ml (100,000 ppm)> 240
MESNA100 mg/ml(100,000 ppm)> 240
Retrovir10 mg/ml(10,000 ppm)> 240
Triclosan3 mg/ml(3,000 ppm)> 240
Zoledronic Acid1 mg/25ml (40 ppm)> 240
Fentanyl Citrate Injection100mcg/2mL> 240

{4}------------------------------------------------

Note: The maximum testing time is 240 minutes. And testing showed minimum breakthrough time of 21.1 minutes with Carmustine (3.3 mg/ml), 74.9 minutes with Thio Tepa (10 mg/ml). Warning: Do not use with Carmustine and Thiotepa.

Type of Use (Select one or both, as applicable)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains a logo for a company called "ZHONGHONG PULIN". The logo consists of two parts: a red square with the letters "Z" and "/" repeated inside, and the company name written in black text to the right of the square. The company name is written in two lines, with "ZHONGHONG" on the top line and "PULIN" on the bottom line.

510(k) Summary K230662

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: Zhonghong Pulin Medical Products Co.,Ltd. Address: West Industrial Park, BDA, Luannan County, Tangshan City, 063500 Hebei, P.R. China. Phone Number: +86-18731635203 Contact: Li Yunjiao Date of Preparation: 2023.10.23

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Nitrile Exam Glove, Extended cuff, Tested For Use With Chemotherapy Drugs, Fentanyl Citrate, And Select Other Drugs

Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove

3.0 Classification

Production code: LZA, LZC, OPJ, QDO Requlation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Name: Ever Global (Vietnam) Enterprise Corp. Device Trade Name:

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains a logo for a company called "ZHONGHONG PULIN". The logo consists of a red square with the letters "Z" and "Z" separated by a vertical line inside. To the right of the square, the company name "ZHONGHONG" is written in a larger, bolder font, with "PULIN" written below it in a smaller font.

Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate 510(k) number: K193555

5.0 Indication for Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

In addition, these gloves were tested for use with chemotherapy drugs, Fentanyl Citrate, and select other drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugConcentrationBreakthrough Detection Time (minutes)
Bendamustine HCI5 mg/ml (5,000 ppm)> 240
Bleomycin Sulfate15 mg/ml (15,000 ppm)> 240
Busulfan6 mg/ml(6,000 ppm)> 240
Carboplatin10.0 mg/ml (10,000 ppm)> 240
Carfilzomib2 mg/ml (2,000 ppm)> 240
Carmustine3.3 mg/ml (3,300 ppm)21.1 (24.2, 21.1, 25.3)
Cetuximab2 mg/ml (2,000 ppm)> 240
Cisplatin1 mg/ml (1000 ppm)> 240
Cladribine1 mg/ml (1000 ppm)> 240
Cyclophosphamide20 mg/ml (20.000 ppm)> 240
Cytarabine(Cytosine)10 mg/ml(10,000 ppm)> 240
Cytovene(Ganciclovir)100.0 mg/ml (100,000 ppm)> 240
Dacarbazine10 mg/ml(10,000 ppm)> 240
Daunorubicin HCI5 mg/ml (5,000 ppm)> 240
Decitabine5 mg/ml(5,000 ppm)> 240
Docetaxel10 mg/ml (10.000 ppm)> 240
Doxorubicin HCI2 mg/ml (2,000 ppm)> 240
Epirubicin HCI (Ellence)20 mg/ml (20,000 ppm)> 240
Etoposide20 mg/ml (20,000 ppm)> 240
Fludarabine25.0 mg/ml (25,000 ppm)> 240
Fluorouracil50 mg/ml (50,000 ppm)> 240
Fulvestrant50 mg/ml(50,000 ppm)> 240
Gemcitabine38 mg/ml (38,000 ppm)> 240
Idarubicin HCI1.0 mg/ml (1,000 ppm)> 240
Ifosfamide50 mg/ml (50,000 ppm)> 240
Irinotecan HCI20 mg/ml (20,000 ppm)> 240
Mechlorethamine HCI1 mg/ml(1,000 ppm)> 240
Melphalan5 mg/ml(5000 ppm)> 240
Methotrexate25 mg/ml (25,000 ppm)> 240
Mitomycin C0.5 mg/ml (500 ppm)> 240
Mitoxantrone HCI2 mg/ml (2,000 ppm)> 240
Oxaliplatin5 mg/ml (5,000 ppm)> 240
Paclitaxel6 mg/ml (6,000 ppm)> 240
Paraplatin10 mg/ml (10,000 ppm)> 240
Pemetrexed25 mg/ml (25,000 ppm)> 240
Raltitrexed0.5 mg/ml (500 ppm)> 240
Rituximab10 mg/ml (10,000 ppm)> 240
Temsirolimus25 mg/ml (25,000 ppm)> 240
Thiotepa10 mg/ml(10,000 ppm)74.9 (74.9, 96.7, 79.0)
Topotecan1 mg/ml (1,000 ppm)> 240

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for ZHONGHONG PULIN. The logo consists of a red square with the letters "Z" and "Z" separated by a vertical line inside. To the right of the square, the words "ZHONGHONG" are stacked on top of the word "PULIN".

ConcentrationBreakthrough Detection Time (minutes)
Trisenox (Arsenic Trioxide)1 mg/ml (1,000 ppm)> 240
Vinblastine,1 mg/ml (1,000 ppm)> 240
Vincristine sulfate (Oncovin)1 g/ml (1,000 ppm)> 240
Vinorelbine10 mg/ml(10,000 ppm)> 240
Fentanyl Citrate and other Drugs
Chloroquine50 mg/ml (50,000 ppm)> 240
Cyclosporin A100.0 mg/ml (100,000 ppm)> 240
MESNA100 mg/ml(100,000 ppm)> 240
Retrovir10 mg/ml(10,000 ppm)> 240
Triclosan3 mg/ml(3,000 ppm)> 240
Zoledronic Acid1 mg/25ml (40 ppm)> 240
Fentanyl Citrate Injection100mcg/2mL> 240

Note: The maximum testing time is 240 minutes. And testing showed minimum breakthrough time of 21.1 minutes with Carmustine (3.3 mg/ml), 74.9 minutes with Thio Tepa (10 mg/ml).

Warning: Do not use with Carmustine and Thiotepa.

6.0 Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, blue colored, nitrile, and tested for use with chemotherapy drugs And Fentanyl Citrate. The design of the subject device is addressing the standards as ASTM D6124, ASTM D5151, ASTM D6319, ASTM D6978-05. The gloves have a minimum length of 270mm. The Subject device is nonsterile.

7.0 Comparison of Technical Characteristics

ItemSubject devicePredicate deviceComparison
510(k)numberNitrile Exam Glove, Extended cuff,Tested For Use WithChemotherapy Drugs, FentanylCitrate, And Select Other DrugsDisposable Powder Free NitrileExamination Glove, Blue Color,Tested For Use With ChemotherapyDrugs And Fentanyl Citrate/
ProductCodeLZA, LZC, OPJ, QDOLZA, LZC, QDOSimilar
RegulationNo.21CFR880.625021CFR880.6250Same
ClassIISame
IndicationsFor UseA patient examination gloves is adisposable device intended for medicalpurpose that is worn on the examiner'shand or fingers to preventA patient examination gloves is adisposable device intended formedical purpose that is worn on theexaminer's hand or fingers toSimilar
contamination between patient and examiner.In addition, these gloves were tested for use with chemotherapy drugs, Fentanyl Citrate, and select other drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.Chemotherapy Drug Concentration Breakthrough Detection Time (minutes)prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs.Test Chemotherapy Drug Concentration (mg/ml) Minimum Breakthrough Detection Time (Min.)
Bendamustine HCl 5 mg/ml (5,000 ppm) > 2401 Arsenic Trioxide (1.0 mg/ml) > 240
Bleomycin Sulfate 15 mg/ml (15,000 ppm) > 2402 Azacitidine (Vidaza) (25.0 mg/ml) > 240
Busulfan 6 mg/ml (6,000 ppm) > 2403 Bendamustine HCl (5.0 mg/ml) > 240
Carboplatin 10.0 mg/ml (10,000 ppm) > 2404 Bleomycin Sulfate (15.0 mg/ml) > 240
Carfilzomib 2 mg/ml (2,000 ppm) > 2405 Bortezomib (Velcade) (1.0 mg/ml) > 240
Carmustine 3.3 mg/ml (3,300 ppm) 21.1 (24.2, 21.1, 25.3)6 Busulfan (6.0 mg/ml) > 240
Cetuximab 2 mg/ml (2,000 ppm) > 2407 Carboplatin (10.0 mg/ml) > 240
Cisplatin 1 mg/ml (1000 ppm) > 2408 Carfilzomib (2.0 mg/ml) > 240
Cladribine 1 mg/ml (1000 ppm) > 2409 Carmustine (BCNU), (3.3 mg/ml) 6.2
Cyclophosphamide 20 mg/ml (20.000 ppm) > 24010 Cetuximab (Erbitux) (2.0 mg/ml) > 240
Cytarabine(Cytosine) 10 mg/ml(10,000 ppm) > 24011 Chloroquine (50.0 mg/ml) > 240
Cytovene (Ganciclovir) 10 mg/ml(10,000 ppm) > 24012 Cisplatin (1.0 mg/ml) > 240
Dacarbazine 10 mg/ml (10,000 ppm) > 24013 Cladribine (1.0 mg/ml) > 240
Daunorubicin HCl 5 mg/ml (5,000 ppm) > 24014 Cyclophosphamide (20.0 mg/ml) > 240
Decitabine 5 mg/ml (5,000 ppm) > 24015 Cyclosporine A (100.0 mg/ml) > 240
Docetaxel 10 mg/ml (10.000 ppm) > 24016 Cytarabine (100.0 mg/ml) > 240
Doxorubicin HCl 2 mg/ml (2,000 ppm) > 24017 Cytovene (Ganciclovir) (10.0 mg/ml) > 240
Epirubicin HCl (Ellence) 20 mg/ml (20,000 ppm) > 24018 Dacarbazine (10.0 mg/ml) > 240
Etoposide 20 mg/ml (20,000 ppm) > 24019 Daunorubicin (5.0 mg/ml) > 240
Fludarabine 25.0 mg/ml (25,000 ppm) > 24020 Decitabine (5.0 mg/ml) > 240
Fluorouracil 50 mg/ml, (50,000 ppm) > 24021 Docetaxel (10.0 mg/ml) > 240
Fulvestrant 50 mg/ml(50,000 ppm) > 24022 Doxorubicin Hydrochloride (2.0 mg/ml) > 240
23 Epirubicin (Ellence) (2.0 mg/ml) > 240
24 Etoposide (20.0 mg/ml) > 240
25 Fludarabine (25.0 mg/ml) > 240
Gemcitabine 38 mg/ml (38,000 ppm)> 24026 Fluorouracil (50.0 mg/ml) > 24027 Fulvestrant (50.0 mg/ml) > 240
Idarubicin HCL 1.0 mg/ml (1,000 ppm)> 24028 Gemcitabine (38.0 mg/ml) > 240
Ifosfamide 50 mg/ml (50,000 ppm)29 Idarubicin (1.0 mg/ml) > 240

General Comparison Table

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image contains a logo with the letters "Z|Z" in white on a red square. To the right of the square, the words "ZHONGHONG" and "PULIN" are written in black, with "ZHONGHONG" on top of "PULIN". The logo appears to represent a company or organization named Zhonghong Pulin.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains a logo for Zhonghong Pulin. The logo consists of a red square with the letters "Z" inside, separated by a vertical line. To the right of the square are the words "ZHONGHONG" on the top line and "PULIN" on the bottom line, both in a sans-serif font.

Irinotecan HCl 20 mg/ml (20,000 ppm) > 24030 Ifosfamide (50.0 mg/ml) > 240
Mechlorethamine HCl 1 mg/ml(1,000 ppm) > 24031 Irinotecan (20.0 mg/ml) > 240
Melphalan 5 mg/ml (5000 ppm) > 24032 Mechlorethamine HCl (1.0 mg/ml) > 240
Methotrexate 25 mg/ml (25,000 ppm) > 24033 Melphalan (5.0 mg/ml) > 240
Mitomycin C 0.5 mg/ml (500 ppm) > 24034 Methotrexate (25 mg/ml) > 240
Mitoxantrone HCL 2 mg/ml (2,000 ppm) > 24035 Mesna (100 mg/ml) > 240
Oxaliplatin 5 mg/ml (5,000 ppm) > 24036 Mitomycin C (0.5 mg/ml) > 240
Paclitaxel 6 mg/ml (6,000 ppm) > 24037 Mitoxantrone (2.0 mg/ml) > 240
Paraplatin 10 mg/ml (10,000 ppm) > 24038 Oxaliplatin (5.0 mg/ml) > 240
Pemetrexed 25 mg/ml (25,000 ppm) > 24039 Paclitaxel (6.0mg/ml) > 240
Raltitrexed 0.5 mg/ml (500 ppm) > 24040 Paraplatin (10.0 mg/ml) > 240
Rituximab 10 mg/ml (10,000 ppm) > 24041 Pemetrexed (25.0 mg/ml) > 240
Temsirolimus 25 mg/ml (25,000 ppm) > 24042 Pertuzumab (30.0 mg/ml) > 240
Thiotepa 10 mg/ml(10,000 ppm) 74.9 (74.9, 96.7, 79.0)43 Raltitrexed (0.5 mg/ml) > 240
Topotecan 1 mg/ml (1,000 ppm) > 24044 Retrovir (10.0 mg/ml) > 240
Trisenox (Arsenic Trioxide) 1 mg/ml (1,000 ppm) > 24045 Rituximab (10.0 mg/ml) > 240
Vinblastine, 1 mg/ml (1,000 ppm) > 24046 Temsirolimus (25.0 mg/ml) > 240
Vincristine sulfate (Oncovin) 1 g/ml (1,000 ppm) > 24047 Thiotepa (10.0 mg/ml) 13.6
Vinorelbine 10 mg/ml (10,000 ppm) > 24048 Topotecan HCI (1.0 mg/ml) > 240
Fentanyl Citrate and other DrugsConcentration BreakthroughDetection Time (minutes)49 Trastuzumab (21.0 mg/ml) > 240
Chloroquine 50 mg/m (50,000 ppm) > 24050 Triclosan (2.0 mg/ml) > 240
Cyclosporin A 100.0 mg/ml (100,000 ppm) > 24051 Trisenox (1.0 mg/ml) > 240
MESNA 100 mg/ml(100,000 ppm) > 24052 Vinblastine (1.0 mg/ml) > 240
Retrovir 10 mg/ml(10,000 ppm) > 24053 Vincristine Sulfate (1.0 mg/ml) > 240
Triclosan 3 mg/ml(3,000 ppm) > 24054 Vinorelbine (10.0 mg/ml) > 240
Zoledronic Acid 1 mg/25ml (40 ppm) > 24055 Zoledronic Acid (0.8 mg/ml) > 240
Fentanyl Citrate Injection 100mcg/2mL > 240Warning: do not use withCarmustine and Thiotepa.The maximum testing time is 240 minutes. Please note that the following drug has an extremely low permeation time:
Note: The maximum testing time is 240Carmustine (BCNU), 3.3 mg/ml 6.2 minutesThiotepa, 10.0 mg/ml 13.6 minutes
Fentanyl Permeation ResistanceClaim - Under the testing conditionsof ASTM D6978-05, Fentanyl CitrateInjection (100mcg/2mL) was foundto have no breakthrough detectedup to 240 minutes.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image contains a logo for a company called "ZHONGHONG PULIN". The logo consists of two parts: a red square with the letters "Z" and "Z" separated by a vertical line, and the company name written in black text to the right of the square. The company name is written in two lines, with "ZHONGHONG" on the top line and "PULIN" on the bottom line.

breakthrough time of 21.1 minutes withCarmustine (3.3 mg/ml), 74.9 minuteswith Thio Tepa (10 mg/ml).Warning: Do not use with Carmustineand Thiotepa.
Powdered orPower freePowder freePowder freeSame
Design FeatureAmbidextrousAmbidextrousSame
SterilityNon-SterileNon-SterileSame
Main MaterialPowder-Free NitrileNitrileSimilar
ColorsBlueBlueSame

Device Dimensions Comparison Table

SubjectdeviceK230662DesignationSize (mm)
XSSMLXLXXL
Length, mm287-293288-297287-300290-297283-301292-300
Criteria≥270≥270≥270≥270≥270≥270
Complywith ASTMD6319ResultsPassPassPassPassPassPass
Width, mm78-8084-8696-97105-107113-115122-124
Criteria70±1080±1095±10110±10120±10130±10
ResultsPassPassPassPassPassPass
Thickness, mm:
Finger(0.082-0.185)
Criteria> 0.05
Resultspass
Palm(0.065-0.123)
Criteria> 0.05
Resultspass
DesignationSize (mm)
PredicateDevice(K193555)XSSMLXL
Length, mm≥300≥300≥300≥300≥300
Criteria≥220≥220≥230≥230≥230
ResultsPassPassPassPassPass
Width, mm70±1080±1095±10110±10120±10
Criteria70±1080±1095±10110±10120±10
ResultsPassPassPassPassPass
Complywith ASTMD6319Thickness, mm:
Finger0.05mm min.
Criteria> 0.05
Resultspass
Palm0.05mm min.
Criteria> 0.05
Resultspass
RemarksSimilar. The subject gloves have a minimum length of 270mm.

Performance Comparison Table

ItemSubject deviceK230662Predicate device(K193555)Remark
------------------------------------------------------------------------

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image contains a logo for "ZHONGHONG PULIN". The logo consists of a red square with the letters "Z" and "Z" separated by a vertical line inside. To the right of the square, the words "ZHONGHONG" are stacked above the word "PULIN".

PhysicalPropertiesComply with ASTMD6319Comply with ASTMD6319Same
BeforeAgingTensileStrength(19.9-28.9)MPa14MPa, minSame
Criteria14MPa, min14MPa, min
ResultsPassPass
UltimateElongation(540-580)%500% minSame
Criteria500% min500% min
ResultsPassPass
AfterAgingTensile Strength(20.2-29.1)MPa14MPa, minSame
Criteria14MPa, min14MPa, min
ResultsPassPass
Ultimate Elongation(508-528)%400%minSame
Criteria400%min400%min
ResultsPassPass
Freedom from Holesaccordance withASTMD5151accordance withASTMD5151Same
CriteriaAQL: 2.5 (ISO 2859)G-1, AQL 2.5
Results0 gloves of failurePassPass
Powder ContentMeet the requirements ofASTM D6319Meet the requirements ofASTM D6319Same
Criteria< 2.0mg per glove< 2.0mg per glove
Results1.9 mg/glovePassPass
Chemotherapy Drugs andFentanyl Citrate InjectionTested with MinimumBreakthrough DetectionTime as Tested per ASTMD6978Bendamustine HCl5 mg/ml (5,000 ppm)> 240 min.Bendamustine HCl(5.0 mg/ml)> 240 min.Same
Bleomycin Sulfate15 mg/ml (15,000 ppm)> 240 min.Bleomycin Sulfate15 mg/ml> 240 min.Same
Busulfan6 mg/ml(6,000 ppm)> 240 min.Busulfan6 mg/ml> 240 minSame
Carboplatin10.0 mg/ml (10,000 ppm)> 240 min.Carboplatin10 mg/ml> 240 min.Same
Carfilzomib2 mg/ml (2,000 ppm)> 240 min.Carfilzomib(2.0 mg/ml)> 240 min.Same
Carmustine3.3 mg/ml (3,300 ppm)21.1 (24.2, 21.1, 25.3)min.Carmustine (BCNU)3.3 mg/ml37.5 min.Similar
Cetuximab2 mg/ml (2,000 ppm)> 240 min.Cetuximab (Erbitux)(2.0 mg/ml)> 240 min.Same
Chloroquine50 mg/ml (50,000 ppm)> 240 min.Chloroquine(50.0 mg/ml)> 240 min.Same
Cisplatin1 mg/ml(1,000 ppm)> 240 min.Cisplatin1.0 mg/ml> 240 min.Same
Cladribine1 mg/ml (1000 ppm)> 240 min.Cladribine(1.0 mg/ml)> 240 min.Same
> 240 min.Cyclophosphamide20 mg/ml (20.000 ppm)> 240 min.> 240 min.Cyclophosphamide(20.0 mg/ml)> 240 minSame
Cyclosporin ACyclosporin ASame

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image contains a logo for a company called "ZHONGHONG PULIN". The logo consists of a red square with the letters "Z" and "Z" separated by a vertical line. To the right of the square, the company name is written in black, with "ZHONGHONG" on the top line and "PULIN" on the bottom line.

100 mg/ml (100,000 ppm)> 240 min.(100.0 mg/ml)> 240 min.
Cytarabine(Cytosine)100.0 mg/ml (100,000 ppm)> 240 min.Cytarabine(100.0 mg/ml)> 240 minSame
Cytovene(Ganciclovir)10 mg/ml(10,000 ppm)> 240 min.Cytovene (Ganciclovir)(10.0 mg/ml)> 240 min.Same
Dacarbazine10 mg/ml(10,000 ppm)> 240 min.Dacarbazine (DTIC)10.0 mg/ml> 240 min.Same
Daunorubicin HCl5 mg/ml (5,000 ppm)> 240 min.Daunorubicin5 mg/mL> 240 min.Same
Decitabine5 mg/ml(5,000 ppm)> 240 min.Decitabine(5.0 mg/ml)> 240 min.Same
Docetaxel10 mg/ml (10.000 ppm)> 240 min.Docetaxel10.0 mg/ml> 240 minSame
Doxorubicin HCl2 mg/ml (2,000 ppm)> 240 min.Doxorubicin HCl2.0 mg/ml:> 240 minSame
Epirubicin HCl (Ellence)2 mg/ml (2,000 ppm)> 240 min.Epirubicin (Ellence)2.0 mg/ml> 240 min.Same
Etoposide20 mg/ml (20,000 ppm)> 240 min.Etoposide,(20.0 mg/ml)> 240 min.Same
Fludarabine25.0 mg/ml (25,000 ppm)> 240 min.Fludarabine25.0 mg/ml> 240 min.Same
Fluorouracil50 mg/ml,(50,000 ppm)> 240 min.Fluorouracil50.0 mg/ml:> 240 minSame
Fulvestrant50 mg/ml(50,000 ppm)> 240 min.Fulvestrant(50.0 mg/ml)> 240 minSame
Gemcitabine38 mg/ml (38,000 ppm)> 240 min.Gemcitabine38 mg/ml> 240 min.Same
Idarubicin HCl1.0 mg/ml (1,000 ppm)> 240 min.Idarubicin1 mg/ml> 240 min.Same
Ifosfamide50 mg/ml (50,000 ppm)> 240 min.Ifosfamide50.0 mg/ml> 240 min.Same
Irinotecan HCl20 mg/ml (20,000 ppm)> 240 min.Irinotecan20.0 mg/ml> 240 min.Same
Mechlorethamine HCl1 mg/ml(1,000 ppm)> 240 min.Mechlorethamine HCI1.0 g/ml> 240 min.Same
Melphalan5 mg/ml(5000 ppm)> 240 min.Melphalan5 mg/ml> 240 min.Same
Methotrexate25 mg/ml (25,000 ppm)> 240 min.Methotrexate25 mg/ml> 240 min.Same
MESNA100 mg/ml(100,000 ppm)Mesna(100 mg/ml)Same

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image shows the logo for ZHONGHONG PULIN. The logo consists of a red square with the letters "Z|Z" in white, followed by the text "ZHONGHONG" in black, and then "PULIN" in black below it. The logo is simple and modern, with a clear and easy-to-read font.

> 240 min.> 240 min.
Mitomycin CMitromycin C.Same
0.5 mg/ml (500 ppm)0.5 mg/ml
> 240 min.> 240
Mitoxantrone HCLMitoxantroneSame
2 mg/ml (2,000 ppm)2.0 mg/ml
> 240 min.> 240 min.
OxaliplatinOxaliplatinSame
5 mg/ml (5,000 ppm)5 mg/ml
> 240 min.> 240 min.
PaclitaxelPaclitaxel (Taxol)Same
6 mg/ml (6,000 ppm)6.0 mg/ml
> 240 min.> 240 min.
ParaplatinParaplatinSame
10 mg/ml (10,000 ppm)10 mg/ml
> 240 min.> 240 min.
PemetrexedPemetrexedSame
25 mg/ml (25,000 ppm)(25.0 mg/ml)
> 240 min.> 240 min.
RaltitrexedRaltitrexedSame
0.5 mg/ml (500 ppm)(0.5 mg/ml)
> 240 min.> 240 min.
RetrovirRetrovirSame
10 mg/ml(10,000 ppm)10 mg/ml
> 240 min.> 240 min.
RituximabRituximabSame
10 mg/ml (10,000 ppm)10.0 mg/ml
> 240 min.> 240 min.
TemsirolimusTemsirolimusSame
25 mg/ml (25,000 ppm)(25.0 mg/ml)
> 240 min.> 240 min.
ThiotepaThiotepaSimilar
10 mg/ml(10,000 ppm)10.0 mg/ml
74.9 (74.9, 96.7, 79.0)min.13.6min.
TopotecanTopotecan HClSame
1 mg/ml (1,000 ppm)(1.0 mg/ml)
> 240 min.> 240 min.
TriclosanTriclosanSimilar
3 mg/ml(3,000 ppm)(2.0 mg/ml)
> 240 min.> 240 min.
Trisenox (Arsenic Trioxide)TrisenoxSame
1 mg/ml (1,000 ppm)1.0 mg/ml
> 240 min.> 240 min.
Vinblastine,Vinblastine,Same
1 mg/ml (1,000 ppm)1 mg/ml (1,000 ppm)
> 240 min.> 240 min.
Vincristine sulfate (Oncovin)1 g/ml (1,000 ppm)Vincristine SulfateSame
> 240 min.1.0 mg/ml
> 240 min.
VinorelbineVinorelbineSame
10 mg/ml(10,000 ppm)10 mg/ml
> 240 min.> 240 min.
Zoledronic AcidZoledronic AcidSimilar
1 mg/25ml (40 ppm)(0.8 mg/ml)
> 240 min.> 240 min.
Fentanyl Citrate InjectionFentanyl Citrate Injection,Same
100mcg/2mL(100 mcg/2ml)
> 240 min.> 240 min

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image contains a logo with the text "ZHONGHONG PULIN". To the left of the text is a red square with two white Z's separated by a vertical line. The text is in a simple, sans-serif font and is aligned to the right of the square. The overall design is clean and professional.

ItemsPredicate deviceSubject deviceRemark
(K193555)K230662
MaterialNitrileNitrile compoundsSimilar
Bio-compatibilityIrritationComply with ISO10993-10Under the conditions of thestudy, not an irritantComply with ISO10993-10Under the conditions of thestudy, not an irritantSimilar
SensitizationComply with ISO10993-10Under conditions of the study,not a sensitizer.Comply with ISO10993-10Under conditions of the study,not a sensitizer.
CytotoxicityComply with ISO10993-5Comply with ISO10993-5Similar
Extraction conditions:37°C and 72 hoursTest conditions:No descriptions in summary
Under the conditions of thestudy, the device is not showncytotoxicityUnder conditions of the study,not cytotoxic

bi

8.0 Summary of Non-Clinical Testing

Biocompatibility Testing

The biocompatibility evaluation for Nitrile Exam Glove, Extended cuff, Tested For Use With Chemotherapy Drugs, Fentanyl Citrate, and Select Other Drugs was conducted in accordance with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

Performance Testing (Bench)

Physical performance qualities of the Subject device were evaluated per ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

Permeation testing was conducted according to ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs to support the addition of the labeling claim: Tested for use with chemotherapy drugs, fentanyl citrate, and select other drugs.

In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • ASTM D6124-06 (Reapproved 2017), Standard Test Method for -

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image contains a logo for Zhonghong Pulin. The logo consists of a red square with the letters "Z" and a vertical line inside, followed by the text "ZHONGHONG" on one line and "PULIN" on the line below. The text is in a simple, sans-serif font and is black.

Residual Powder on Medical Gloves

  • ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
  • Assessment Of Resistance To Permeation By Chemotherapy Drugs - And Fentanyl Citrate per ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
No.Purpose of theTestStandardAcceptance CriteriaResultsSummary
Methodology
1Test for skinsensitizationpotentialISO10993-10:2010 BiologicalUnder the conditions of the testing,not a sensitizerThe primary irritation indexes of the polar andnon-polar test group were both calculated to be 0.Pass
2Test for IrritationpotentialEvaluation OfMedical Devices- Part 10Under the conditions of the testing,not an irritant.The skin sensitization rates of polar and non-polar extract group were both determined with 0%.Pass
3Evaluate In VitroCytotoxicitypotentialISO10993-5:2009Under the conditions of the testing,non-cytotoxic.The cell viability of the100% test article extractwas 86.2%Pass
4Measure residualpowder onmedical glovesASTM D6124-06(Reapproved2017)powder residue limit of 2.0mg/glove1.9 mg/glovePass
5Detection ofholes in medicalglovesASTM D5151-19Freedom free holes AQL: 2.5 (ISO2859-1)0 gloves of failurePass
6Evaluate physicaldimensionsASTM D6319-19, (7.4 &ASTM D3767-03(2 020))XS: width 70±10mm Length ≥270mmLength: 287-293 mmWidth: 78-80mmPass
S: width 80±10mm Length ≥270mmLength: 288-297 mmWidth: 84-86mmPass
M: width 95±10mm Length ≥270mmLength: 287-300mmWidth: 96-97mmPass
L: width 110±10mm Length ≥270mmLength: 290-297 mmWidth: 105-107mmPass
XL: width 120±10mm Length ≥270mmLength: 283-301 mmWidth: 113-115mmPass
XXL: width 130±10mm Length≥270 mmLength: 292-300 mmWidth: 122-124mmPass
ThicknessFinger ≥0.05 mm Palm ≥0.05 mmFinger :0.082-0.185mmPalm: 0.065-0.123mmPass
Evaluate physical properties: Before aging
Tensile strengthASTM D6319-≥14MPa(19.9-28.9)MPaPass
Ultimate19,≥500%(540-580)%Pass
Elongation7.5& ASTMD412-16 (2021)
After Accelerated Aging
Tensile strengthASTM D6319-≥14MPa(20.2-29.1)MPaPass
Ultimate19, 7.5& ASTM≥400%(508-528)%Pass
ElongationD412-16 (2021)&ASTM D573-04 (2019)
7EvaluatepermeationcharacteristicsASTM D6978Bendamustine HCI5 mg/ml (5,000 ppm)> 240 min.> 240 min.
meBleomycin Sulfate15 mg/ml (15,000 ppm)> 240 min.> 240 min.
Busulfan6 mg/ml(6,000 ppm)> 240 min.> 240 min.
Carboplatin10.0 mg/ml (10,000 ppm)> 240 min.> 240 min.
Carfilzomib2 mg/ml (2,000 ppm)> 240 min.> 240 min.
Carmustine3.3 mg/ml (3,300 ppm)21.1 (24.2, 21.1, 25.3)min.21.1 min
Cetuximab2 mg/ml (2,000 ppm)> 240 min.> 240 min.
Chloroquine50 mg/m (50,000 ppm)> 240 min.> 240 min.
Cisplatin1 mg/ml(1,000 ppm)> 240 min.> 240 min.
Cladribine1 mg/ml (1000 ppm)> 240 min.> 240 min.
Cyclophosphamide20 mg/ml (20.000 ppm)> 240 min.> 240 min.
Cyclosporin A100 mg/ml (100,000 ppm)> 240 min.> 240 min.
Cytarabine(Cytosine)100.0 mg/ml (100,000 ppm)> 240 min.> 240 min.
Cytovene(Ganciclovir)10 mg/ml(10,000 ppm)> 240 min.> 240 min.
Dacarbazine10 mg/ml(10,000 ppm)> 240 min.> 240 min.
Daunorubicin HCI5 mg/ml (5,000 ppm)> 240 min.Decitabine5 mg/ml(5,000 ppm)> 240 min.> 240 min.> 240 min.
Docetaxel10 mg/ml (10.000 ppm)> 240 min.> 240 min.
Doxorubicin HCI2 mg/ml (2,000 ppm)> 240 min.> 240 min.
Epirubicin HCI (Ellence)2 mg/ml (2,000 ppm)> 240 min.> 240 min.
Etoposide20 mg/ml (20,000 ppm)> 240 min.> 240 min.
Fludarabine25.0 mg/ml (25,000 ppm)> 240 min.> 240 min.
Fluorouracil> 240 min.
50 mg/ml,(50,000 ppm)
> 240 min.
Fulvestrant> 240 min.
50 mg/ml(50,000 ppm)
> 240 min.
Gemcitabine> 240 min.
38 mg/ml (38,000 ppm)
> 240 min.
Idarubicin HCL> 240 min.
1.0 mg/ml (1,000 ppm)
> 240 min.
Ifosfamide> 240 min.
50 mg/ml (50,000 ppm)
> 240 min.
Irinotecan HCI> 240 min.
20 mg/ml (20,000 ppm)
> 240 min.
Mechlorethamine HCI> 240 min.
1 mg/ml(1,000 ppm)
> 240 min.
Melphalan> 240 min.
5 mg/ml(5000 ppm)
> 240 min.
Methotrexate> 240 min.
25 mg/ml (25,000 ppm)
> 240 min.
MESNA> 240 min.
100 mg/ml(100,000 ppm)
> 240 min.
Mitomycin C> 240 min.
0.5 mg/ml (500 ppm)
> 240 min.
Mitoxantrone HCL> 240 min.
2 mg/ml (2,000 ppm)
> 240 min.
Oxaliplatin> 240 min.
5 mg/ml (5,000 ppm)
> 240 min.
Paclitaxel> 240 min.
6 mg/ml (6,000 ppm)
> 240 min.
Paraplatin> 240 min.
10 mg/ml (10,000 ppm)
> 240 min.
Pemetrexed> 240 min.
25 mg/ml (25,000 ppm)
> 240 min.
Raltitrexed> 240 min.
0.5 mg/ml (500 ppm)
> 240 min.
Retrovir> 240 min.
10 mg/ml(10,000 ppm)
> 240 min.
Rituximab> 240 min.
10 mg/ml (10,000 ppm)
> 240 min.
Temsirolimus> 240 min.
25 mg/ml (25,000 ppm)
> 240 min.
Thiotepa10 mg/ml(10,000 ppm)74.9 min.
74.9 (74.9, 96.7, 79.0)min.
Topotecan> 240 min.
1 mg/ml (1,000 ppm)

Table 1 Summary of Non-Clinical Testing

{16}------------------------------------------------

Image /page/16/Picture/1 description: The image contains a logo for "ZHONGHONG PULIN". The logo consists of a red square with the letters "Z" and "/" repeated inside. To the right of the square, the words "ZHONGHONG" are stacked on top of the word "PULIN". The text is in a simple, sans-serif font.

{17}------------------------------------------------

Image /page/17/Picture/1 description: The image contains a logo for a company called "ZHONGHONG PULIN". The logo consists of two parts: a red square with the letters "Z" and "Z" separated by a vertical line, and the company name written in black text to the right of the square. The company name is written in two lines, with "ZHONGHONG" on the top line and "PULIN" on the bottom line.

{18}------------------------------------------------

Image /page/18/Picture/1 description: The image contains a logo for Zhonghong Pulin. The logo consists of a red square with the letters "Z" and a vertical line separating them. To the right of the square are the words "ZHONGHONG" in a larger font and "PULIN" in a smaller font below it. The text is in a sans-serif font and is black.

> 240 min.> 240 min.
Triclosan3 mg/ml(3,000 ppm)> 240 min.> 240 min.
Trisenox (Arsenic Trioxide)1 mg/ml (1,000 ppm)> 240 min.> 240 min.
Vinblastine,1 mg/ml (1,000 ppm)> 240 min.> 240 min.
Vincristine sulfate (Oncovin)1 g/ml (1,000 ppm)> 240 min.> 240 min.
Vinorelbine10 mg/ml(10,000 ppm)> 240 min.> 240 min.
Zoledronic Acid1 mg/25ml (40 ppm)> 240 min.> 240 min.
Fentanyl Citrate Injection100mcg/2mL> 240 min.> 240 min.

9.0 Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Nitrile Exam Glove, Extended cuff, Tested For Use With Chemotherapy Drugs, Fentanyl Citrate, And Select Other Drugs is as safe, as effective, and performs as well as or better than the legally marketed predicate device K193555.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.