K Number
K193555

Validate with FDA (Live)

Date Cleared
2020-06-04

(167 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs and fentanyl citrate in accordance with ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical gloves to Permeation by Chemotherapy Drugs.

Device Description

"Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate" is a patient examination glove made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of this medical device is to provide single use barrier protection for the wearer and the device meets the specifications for Barrier Protection and tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

AI/ML Overview

This document describes the premarket notification (510(k)) for a medical device: "Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate" (K193555). The core of the submission, as it relates to proving the device meets acceptance criteria, focuses on the glove's resistance to permeation by chemotherapy drugs and fentanyl citrate.

Here's an analysis of the provided information to answer your questions:

1. A table of acceptance criteria and the reported device performance

The primary acceptance criteria for the chemotherapy drug and fentanyl citrate resistance are the "Minimum Breakthrough Detection Time (Min.)" as per ASTM D6978-05. The device aims to achieve a breakthrough time of "> 240 minutes" for most listed drugs.

Test Chemotherapy Drug / Fentanyl CitrateConcentration (mg/ml or mcg/2mL)Acceptance Criteria (Minimum Breakthrough Detection Time)Reported Device Performance (Breakthrough Detection Time - Short Cuff)Reported Device Performance (Breakthrough Detection Time - Long Cuff)
Arsenic Trioxide1.0> 240 min> 240 minN/A
Azacitidine (Vidaza)25.0> 240 min> 240 minN/A
Bendamustine HCl5.0> 240 min> 240 minN/A
Bleomycin Sulfate15.0> 240 min> 240 minN/A
Bortezomib (Velcade)1.0> 240 min> 240 minN/A
Busulfan6.0> 240 min> 240 minN/A
Carboplatin10.0> 240 min> 240 minN/A
Carfilzomib2.0> 240 min> 240 minN/A
Carmustine (BCNU)3.3Not specified for new device directly, but predicate showed 6.2 min. New device aims to show improved or similar performance.21.5 min37.5 min
Cetuximab (Erbitux)2.0> 240 min> 240 minN/A
Chloroquine50.0> 240 min> 240 minN/A
Cisplatin1.0> 240 min> 240 minN/A
Cladribine1.0> 240 min> 240 minN/A
Cyclophosphamide20.0> 240 min> 240 minN/A
Cyclosporine A100.0> 240 min> 240 minN/A
Cytarabine100.0> 240 min> 240 minN/A
Cytovene (Ganciclovir)10.0> 240 min> 240 min (Found in Indications for Use table, not explicitly in the detailed comparison table)N/A
Dacarbazine10.0> 240 min> 240 minN/A
Daunorubicin5.0> 240 min> 240 minN/A
Decitabine5.0> 240 min> 240 minN/A
Docetaxel10.0> 240 min> 240 minN/A
Doxorubicin Hydrochloride2.0> 240 min> 240 minN/A
Epirubicin (Ellence)2.0> 240 min> 240 minN/A
Etoposide20.0> 240 min> 240 minN/A
Fludarabine25.0> 240 min> 240 minN/A
Fluorouracil50.0> 240 min> 240 minN/A
Fulvestrant50.0> 240 min> 240 minN/A
Gemcitabine38.0> 240 min> 240 minN/A
Idarubicin1.0> 240 min> 240 minN/A
Ifosfamide50.0> 240 min> 240 minN/A
Irinotecan20.0> 240 min> 240 minN/A
Mechlorethamine HCl1.0> 240 min> 240 minN/A
Melphalan5.0> 240 min> 240 minN/A
Methotrexate25.0> 240 min> 240 minN/A
Mesna100.0> 240 min> 240 minN/A
Mitomycin C0.5> 240 min> 240 minN/A
Mitoxantrone2.0> 240 min> 240 minN/A
Oxaliplatin5.0> 240 min> 240 minN/A
Paclitaxel6.0> 240 min> 240 minN/A
Paraplatin10.0> 240 min> 240 minN/A
Pemetrexed25.0> 240 min> 240 minN/A
Pertuzumab30.0> 240 min> 240 minN/A
Raltitrexed0.5> 240 min> 240 minN/A
Retrovir10.0> 240 min> 240 minN/A
Rituximab10.0> 240 min> 240 minN/A
Temsirolimus25.0> 240 min> 240 minN/A
Thiotepa10.0Not specified for new device directly, but predicate showed 38.8 min. New device aims to show improved or similar performance.23.1 min13.6 min
Topotecan HCl1.0> 240 min> 240 minN/A
Trastuzumab21.0> 240 min> 240 minN/A
Triclosan2.0> 240 min> 240 minN/A
Trisenox1.0> 240 min> 240 minN/A
Vinblastine1.0> 240 min> 240 minN/A
Vincristine Sulfate1.0> 240 min> 240 minN/A
Vinorelbine10.0> 240 min> 240 minN/A
Zoledronic Acid0.8> 240 min> 240 minN/A
Fentanyl Citrate Injection100 mcg/2mL> 240 min> 240 minN/A

Note on Acceptance Criteria: For most drugs, the stated acceptance criterion is explicitly given as "> 240 minutes" in the "Indications for Use" section and confirmed by the reported performance. For Carmustine and Thiotepa, the document highlights their "extremely low permeation time" and provides their specific breakthrough times. The comparison table (Table 2) then compares these to the predicate, showing the new device's performance. The "Warning: do not use with Carmustine and Thiotepa" statement effectively sets the acceptance criteria for these specific drugs as less than the ideal >240 minutes, acknowledging their known rapid permeation through gloves. The aim for these is to show performance comparable or better than the predicate, which the subject device achieves for Carmustine, but is slightly worse for Thiotepa (Thiotepa: predicate 38.8 min, subject device 23.1 min (short cuff) and 13.6 min (long cuff)). However, since a warning is issued, this outcome appears to be accepted.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: The document does not explicitly state the specific number of gloves/samples tested for each drug permeation. It refers to testing "in accordance with ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs," which would define the sample size requirements for the test. However, the exact sample numbers (e.g., how many gloves were tested for each drug) are not provided in this summary.
  • Data Provenance: The tests were conducted by Ever Global (Vietnam) Enterprise Corp (Vietnam). The type of study for these tests would be prospective bench testing, as new gloves were manufactured and then tested for their characteristics.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is a device performance study based on laboratory bench testing, not an AI or diagnostic study involving expert interpretation of medical images. Therefore:

  • Number of experts: Not applicable. Ground truth for material properties (like drug permeation) is established through standardized laboratory procedures and measurements.
  • Qualifications of experts: Not applicable in the context of human expert review of "ground truth" for clinical data. The "experts" here would be the laboratory personnel qualified to perform ASTM D6978-05 and other relevant material testing standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not a study involving human reader interpretation or adjudication. Device performance is determined by direct physical and chemical testing according to established ASTM standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a 510(k) submission for a physical medical device (gloves), not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical device, not an algorithm. The "standalone performance" is the intrinsic material property of the glove against chemical permeation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the device's performance regarding chemotherapy drug and fentanyl permeation is established through direct laboratory measurement using the methodology outlined in ASTM D6978-05. This standard defines how resistance to permeation is assessed, with breakthrough time being the key metric. For other properties like tensile strength, freedom from holes, powder residual, and biocompatibility, the ground truth is also established by specific ASTM standards and biological evaluation guidelines (e.g., ISO 10993).

8. The sample size for the training set

Not applicable. This is a traditional device submission, not an AI/machine learning product that requires a "training set." The gloves are manufactured, and then samples from the manufacturing batches are subjected to testing (the "test set" in this context).

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

June 4, 2020

Ever Global (Vietnam) Enterprise Corp % Elizabeth Deng U.S. Representative Elizabeth Deng 5748 Eaglewood Place Ranch Cucamonga, California 91730

Re: K193555

Trade/Device Name: Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZC, LZA, QDO Dated: April 29, 2020 Received: May 5, 2020

Dear Elizabeth Deng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Ever Global (Vietnam) Enterprise Corporation

Long Thanh Industrial Zone, Taman Village, Long Thanh District, Dong Nai Province, Vietnam

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationForm Approved: OMB No. 0910-0120Expiration Date: 06/30/2020See PRA Statement below.
Indications for Use
510(k) Number (if known)K193555
Device NameDisposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate
Indications for Use (Describe)A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand orfingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use withchemotherapy drugs and fentanyl citrate in accordance with ASTM D6978-05 Standard Practice for Assessment ofResistance of Medical gloves to Permeation by Chemotherapy Drugs.
Test Chemotherapy DrugConcentration (mg/ml)Minimum Breakthrough Detection Time (Min.)
01 Arsenic Trioxide1.0> 240
02 Azacitidine (Vidaza)25.0> 240
03 Bendamustine HCl5.0> 240
04 Bleomycin Sulfate15.0> 240
05 Bortezomib (Velcade)1.0> 240
06 Busulfan6.0> 240
07 Carboplatin10.0> 240
08 Carfilzomib2.0> 240
09 Carmustine (BCNU)3.36.2
10 Cetuximab (Erbitux)2.0> 240
11 Chloroquine50.0> 240
12 Cisplatin1.0> 240
13 Cladribine1.0> 240
14 Cyclophosphamide20.0> 240
15 Cyclosporine A100.0> 240
16 Cytarabine mg/ml)100.0> 240
17 Cytovene (Ganciclovir)10.0> 240
18 Dacarbazine10.0> 240
19 Daunorubicin5.0> 240
20 Decitabine5.0> 240
21 Docetaxel10.0> 240
22 Doxorubicin Hydrochloride2.0> 240
23 Epirubicin (Ellence)2.0> 240
24 Etoposide20.0> 240
25 Fludarabine25.0> 240
26 Fluorouracil50.0> 240
27 Fulvestrant50.0> 240
28 Gemcitabine38.0> 240
29 Idarubicin1.0> 240
30 Ifosfamide50.0> 240
31 Irinotecan20.0> 240
32 Mechlorethamine HCl1.0> 240
33 Melphalan5.0> 240
34 Methotrexate25.0> 240
35 Mesna100.0> 240
36 Mitomycin C0.5> 240
37 Mitoxantrone2.0> 240
FORM FDA 3881 (7/17)Page 1 of 2PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

Ever Global (Vietnam) Enterprise Corporation

Long Thanh Industrial Zone, Taman Village, Long Thanh District, Dong Nai Province, Vietnam

38 Oxaliplatin5.0> 240
39 Paclitaxel6.0> 240
40 Paraplatin10.0> 240
41 Pemetrexed25.0> 240
42 Pertuzumab30.0> 240
43 Raltitrexed0.5> 240
44 Retrovir10.0> 240
45 Rituximab10.0> 240
46 Temsirolimus25.0> 240
47 Thiotepa10.013.6
48 Topotecan HCl1.0> 240
49 Trastuzumab21.0> 240
50 Triclosan2.0> 240
51 Trisenox1.0> 240
52 Vinblastine1.0> 240
53 Vincristine Sulfate1.0> 240
54 Vinorelbine10.0> 240
55 Zoledronic Acid0.8> 240
Warning: do not use with Carmustine and Thiotepa.The maximum testing time is 240 minutes. Please note that the following drug has an extremely low permeation time:
Carmustine (BCNU)3.3 mg/ml6.2 minutes
Thiotepa10.0 mg/ml13.6 minutes
Fentanyl Permeation Resistance Claim - Under the testing conditions of ASTM D6978-05, Fentanyl CitrateInjection (100mcg/2mL) was found to have no breakthrough detected up to 240 minutes.
Type of Use (Select one or both, as applicable)
Image: CheckboxOver-The-Counter Use (21 CFR 801 Subpart C)
Image: Unchecked checkboxPrescription Use (Part 21 CFR 801 Subpart D)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY

K193555

1.0 Submitter:

Submitter's name :Ever Global (Vietnam) Enterprise Corp.
Submitter's address :Long Thanh Industrial ZoneTaman Village Dong Nai Province, VN 810000
Phone number :84-61-3514022
Fax number :84 -61-3514023
Name of contact person:Jerry Lin
Summary Preparation Date:May 27th,2020
2.0 US Agent:
US representative name:Elizabeth Deng
Company address:5748 Eaglewood PlaceRancho Cuamonga, CaliforniaRancho Cucamonga, CA 91739
Telephone number:909 4659188
Contact email:Baxianunited48@Yahoo.Com
3.0 Name of the Device
Proprietary/Trade name:Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested ForUse With Chemotherapy Drugs and Fentanyl Citrate
Common Name:Nitrile Examination Gloves
Classification Name:Patient Examination Glove
Device Classification:Class I
Regulation Number:21 CFR 880.6250
Product Code:LZA, LZC, QDO
4.0 Predicate device
Device Name:Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested ForUse With Chemotherapy Drugs
Company name:Ever Global (Vietnam) Enterprise Corp.
510(K) Number:K190403

5.0 Device Description:

"Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate" is a patient examination glove made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of this medical device is to provide single use barrier protection for the wearer and the device meets the specifications for Barrier Protection and tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

6.0 Indication for use:

Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs.

{5}------------------------------------------------

No.Test Chemotherapy Drug andConcentrationMinimum BreakthroughDetection Time (Min.)
1Arsenic Trioxide (1.0 mg/ml)> 240
2Azacitidine (Vidaza) (25.0 mg/ml)> 240
3Bendamustine HCl (5.0 mg/ml)> 240
4Bleomycin Sulfate (15.0 mg/ml)> 240
5Bortezomib (Velcade) (1.0 mg/ml)> 240
6Busulfan (6.0 mg/ml)> 240
7Carboplatin (10.0 mg/ml)> 240
8Carfilzomib (2.0 mg/ml)> 240
9Carmustine (BCNU), (3.3 mg/ml)6.2
10Cetuximab (Erbitux) (2.0 mg/ml)> 240
11Chloroquine (50.0 mg/ml)> 240
12Cisplatin (1.0 mg/ml)> 240
13Cladribine (1.0 mg/ml)> 240
14Cyclophosphamide (20.0 mg/ml)> 240
15Cyclosporine A (100.0 mg/ml)> 240
16Cytarabine (100.0 mg/ml)> 240
17Cytovene (Ganciclovir) (10.0 mg/ml)> 240
18Dacarbazine (10.0 mg/ml)> 240
19Daunorubicin (5.0 mg/ml)> 240
20Decitabine (5.0 mg/ml)> 240
21Docetaxel (10.0 mg/ml)> 240
22Doxorubicin Hydrochloride (2.0 mg/ml)> 240
23Epirubicin (Ellence) (2.0 mg/ml)> 240
24Etoposide (20.0 mg/ml)> 240
25Fludarabine (25.0 mg/ml)> 240
26Fluorouracil (50.0 mg/ml)> 240
27Fulvestrant (50.0 mg/ml)> 240
28Gemcitabine (38.0 mg/ml)> 240
29Idarubicin (1.0 mg/ml)> 240
30Ifosfamide (50.0 mg/ml)> 240
31Irinotecan (20.0 mg/ml)> 240
32Mechlorethamine HCl (1.0 mg/ml)> 240
33Melphalan (5.0 mg/ml)> 240
34Methotrexate (25 mg/ml)> 240
35Mesna (100 mg/ml)> 240
36Mitomycin C (0.5 mg/ml)> 240
37Mitoxantrone (2.0 mg/ml)> 240
38Oxaliplatin (5.0 mg/ml)> 240
39Paclitaxel (6.0mg/ml)> 240
40Paraplatin (10.0 mg/ml)> 240
41Pemetrexed (25.0 mg/ml)> 240
42Pertuzumab (30.0 mg/ml)> 240
43Raltitrexed (0.5 mg/ml)> 240
44Retrovir (10.0 mg/ml)> 240
45Rituximab (10.0 mg/ml)> 240
46Temsirolimus (25.0 mg/ml)> 240
47Thiotepa (10.0 mg/ml)13.6
48Topotecan HCl (1.0 mg/ml)> 240
49Trastuzumab (21.0 mg/ml)> 240
50Triclosan (2.0 mg/ml)> 240
51Trisenox (1.0 mg/ml)> 240
52Vinblastine (1.0 mg/ml)> 240
53Vincristine Sulfate (1.0 mg/ml)> 240
54Vinorelbine (10.0 mg/ml)> 240
55Zoledronic Acid (0.8 mg/ml)> 240

Table 1 Tested for use with 55 chemotherapy drugs.

{6}------------------------------------------------

Warning: do not use with Carmustine and Thiotepa.

The maximum testing time is 240 minutes. Please note that the following drug has an extremely low permeation time:

Carmustine (BCNU), 3.3 mg/ml6.2 minutes
Thiotepa, 10.0 mg/ml13.6 minutes

Fentanyl Permeation Resistance Claim - Under the testing conditions of ASTM D6978-05, Fentanyl Citrate Injection (100mcg/2mL) was found to have no breakthrough detected for up to 240 minutes.

{7}------------------------------------------------

7.0 Comparison of Technological characteristics between the predicate and subject devices:

Device CharacteristicPredicate DeviceSubject DeviceComparison
Product nameDisposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy DrugsDisposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citratesimilar
510(K) No.K190403K193555N/A
Product OwnerEver Global (Vietnam) Enterprise CorporationEver Global (Vietnam) Enterprise Corporationsame
Product CodeLZA, LZCLZA, LZC, QDOsimilar
Regulation21 CFR 880.625021 CFR 880.6250same
ClassIIsame
Intended UseThe Nitrile Powder Free patient examination glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.The Nitrile Powder Free patient examination glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs And Fentanyl Citrate in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.similar
Powder freeYesYessame
SizeSmall/ Medium/Large/X LargeX Small/ Small/Medium/Large/X LargeX Small is additional.
Single UseYESYESsame
Non-SterileNoNosame
Dimensions-LengthComplies with ASTM D6319-10 230 mm min.Short cuff ≥230mmLong cuff ≥300mmLong cuff is additional.
Dimensions - Palm WidthComplies with ASTM D6319-10Complies with ASTM D6319-10Similar.X Small is additional.
SmallMediumLargeX Small 70±10Small 80±10Medium 95 ±10Large 110±10

Table 2 Comparison of Technological Characteristics

{8}------------------------------------------------

Dimensions - ThicknessComplies with ASTM D6319-10Palm 0.05mm min.Finger 0.05 mm min.Cuff 0.05 mm minComplies with ASTM D6319-10Palm 0.05mm min.Finger 0.05 mm min.Cuff 0.05 mm minsame
PhysicalPropertiesTensile Strength:Before Aging 14 MPa, min.After Aging 14 MPa, min.Elongation:Before Aging 500% min.After Aging 400% min.Tensile Strength:Before Aging 14 MPa, min.After Aging 14 MPa, min.Elongation:Before Aging 500% min.After Aging 400% min.same
ResidualpowderComplies with ASTM D6319-10< 2mg per gloveComplies with ASTM D6319-10< 2mg per glovesame
Freedom fromHolesIn accordance with ASTM D6319-10 and ASTM D5151- 06(reapproved 2015), G-1, AQL 2.5In accordance with ASTM D6319-10 and ASTM D5151-06(reapproved 2015), G-1, AQL 2.5same
BiocompatibilityAAMI/ANSI/ISO 10993-10PassesNot a skin irritant & Not a skinsensitizerAAMI/ANSI/ISO 10993-10 &AAMI/ANSI/ISO 10993-5PassesNot a skin irritant, Not a skinsensitizer & No cytotoxicityreactionsame
Chemotherapy drugs testedBreakthrough Detection Time in MinutesComparison
PredicatedeviceSubject Device
Device CharacteristicK190403Short CuffLong Cuff
Arsenic Trioxide (1.0 mg/ml)-> 240Different
Azacitidine (Vidaza) (25.0 mg/ml)-> 240Different
Bendamustine HCI (5.0 mg/ml)-> 240Different
Bleomycin Sulfate (15.0 mg/ml)-> 240Different
Bortezomib (Velcade) (1.0 mg/ml)-> 240Different
Busulfan (6.0 mg/ml)-> 240Different
Carboplatin (10.0 mg/ml)-> 240Different
Carfilzomib (2.0 mg/ml)-> 240Different
Carmustine (BCNU), (3.3 mg/ml)6.221.537.5Similar
Cetuximab (Erbitux) (2.0 mg/ml)-> 240Different
Chloroquine (50.0 mg/ml)-> 240Different
Cisplatin (1.0 mg/ml)> 240> 240Same
Cladribine (1.0 mg/ml)-> 240Different
Cyclophosphamide (20.0 mg/ml)> 240> 240Same
Cyclosporin A (100.0 mg/ml)-> 240Different
Cytarabine (100.0 mg/ml)-> 240Different
Dacarbazine (DTIC), (10.0 mg/ml)> 240> 240Same
Daunorubicin (5.0 mg/ml)-> 240Different
Decitabine (5.0 mg/ml)-> 240Different
Docetaxel (10.0 mg/ml)-> 240Different
Doxorubicin Hydrochloride (2.0mg/ml)> 240> 240Same
Epirubicin (Ellence) (2.0 mg/ml)-> 240Different
Etoposide, (20.0 mg/ml)> 240> 240Same
Fludarabine (25.0 mg/ml)-> 240Different
Fluorouracil, (50.0 mg/ml)> 240> 240Same
Fulvestrant (50.0 mg/ml)-> 240Different
Gemcitabine (38.0 mg/ml)-> 240Different
Idaribicin (1.0 mg/ml)-> 240Different
Ifosfamide (50.0 mg/ml)-> 240Different
Irinotecan (20.0 mg/ml)-> 240Different
Mechlorethamine HCl (1.0 mg/ml)-> 240Different
Melphalan (5.0 mg/ml)-> 240Different
Methotrexate (25 mg/ml)-> 240Different
Mesna (100 mg/ml)-> 240Different
Mitomycin C (0.5 mg/ml)-> 240Different
Mitoxantrone (2.0 mg/ml)-> 240Different
Oxaliplatin (5.0 mg/ml)-> 240Different
Paclitaxel (Taxol), (6.0 mg/ml)> 240> 240Same
Paraplatin (10.0 mg/ml)-> 240Different
Pemetrexed (25.0 mg/ml)-> 240Different
Pertuzumab (30.0 mg/ml)-> 240Different
Raltitrexed (0.5 mg/ml)-> 240Different
Retrovir (10.0 mg/ml)-> 240Different
Rituximab (10.0 mg/ml)-> 240Different
Temsirolimus (25.0 mg/ml)-> 240Different
Thiotepa (10.0 mg/ml)38.823.113.6Similar
Topotecan HCl (1.0 mg/ml)-> 240Different
Trastuzumab (21.0 mg/ml)-> 240Different
Triclosan (2.0 mg/ml)-> 240Different
Trisenox (1.0 mg/ml)-> 240Different
Vinblastine (1.0 mg/ml)-> 240Different

{9}------------------------------------------------

{10}------------------------------------------------

Vincristine Sulfate (1.0 mg/ml)-> 240Different
Vinorelbine (10.0 mg/ml)-> 240Different
Zoledronic Acid (0.8 mg/ml)-> 240Different
Fentanyl Permeation Resistance
Fentanyl Citrate Injection, (100mcg/2ml)-> 240Different

8.0 Summary of Non-clinical testing results:

Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate is summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandard
DimensionASTM standard D 6319-10(Reapproved 2015)
LengthShort cuff≥230mmLong cuff≥300mm
WidthX Small$70 \pm 10 mm$Small$80 \pm 10 mm$Medium$95 \pm 10 mm$Large$110 \pm 10 mm$X large$120 \pm 10 mm$
ThicknessFinger tip≥0.05mmPalm≥0.05mmCuff≥0.05mm
PhysicalPropertiesASTM standard D 6319-10(Reapproved 2015)Tensile strength (Before aging)≥14MPaTensile strength (After aging)≥14MPaElongated rate (Before aging)≥500%Elongated rate (After aging)≥400%
Freedom frompinholes21 CFR 800.20ASTM standard D 6319-10(Reapproved 2015)Test method in accordance with ASTM D5151-06(Reapproved 2015)Passed Standard Acceptance Criteria
Powder ResidualASTM standard D6319-10 (Reapproved 2015)Test method in accordance with D6124-06(Reaffirmation 2011)< 2 mg/glove
BiocompatibilityPrimary Skin Irritation in rabbitsPassUnder the conditions of the study, the subject device is not a primary skin irritant.Dermal sensitization in the guinea pigPassUnder the conditions of the study, the subject device is not a primary skin sensitizer.In vitro cytotoxicityPassUnder the conditions of the study, the subject device is not shown cytotoxicity.

Table 3 Summary of the Technological Characteristics

The following bench testing was conducted for the Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate made by Ever Global (Vietnam) Enterprise Corp:

  • Dimension per ASTM D6319-10 (Reapproved 2015) Standard Specification for Nitrile Examination ● Gloves for Medical Application

{11}------------------------------------------------

  • Tensile strength(Before aging/After aging) and Elongation(Before aging/After aging) per ASTM D6319-10(Reapproved 2015) Standard Specification for Nitrile Examination Gloves for Medical Application
  • Water leak test on pinhole per ASTM D6319-10(Reapproved 2015) Standard Specification for ● Nitrile Examination Gloves for Medical Application and per 21 CFR 800.20 Patient examination gloves and surgeons' gloves; sample plans and test method for leakage defects;adulteration.
  • Powder Residual tests per ASTM D6319-10(Reapproved 2015) Standard Specification for Nitrile Examination Gloves for Medical Application.
  • . Biocompatibility test per ISO 10993-5 Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity & ISO 10993-10: Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitization.
  • . Assessment Of Resistance To Permeation By Chemotherapy Drugs And Fentanyl Citrate per ASTM D6978-05(R 2013) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs

9.0 Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device the Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K190403.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.