K Number
K230601

Validate with FDA (Live)

Date Cleared
2023-05-28

(86 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250).The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color, powder free and non sterile. The device is available in Extra Small, Small, Medium, Large and Extra Large sizes.

AI/ML Overview

The provided document details the acceptance criteria for "Blue Nitrile Examination Gloves Powder Free" and presents the results of non-clinical testing to demonstrate the device meets these criteria.

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicStandardAcceptance Criteria (Predicate / Subject Device)Reported Device Performance (Subject Device)
Dimensions
LengthASTM D6319-2019Min 230 mm for all sizesX-Small: 245 mm, Small: 245 mm, Medium: 246 mm, Large: 248 mm, X-Large: 248 mm
WidthASTM D6319-2019X-small: 70+/-10 mm, Small: 80+/-10 mm, Medium: 95+/-10 mm, Large: 110+/-10 mm, X-Large: 120+/-10 mmX-small: 76 mm, Small: 84 mm, Medium: 94 mm, Large: 105 mm, X-Large: 115 mm
Physical Properties
ThicknessASTM D6319-2019Palm min 0.05 mm, Finger min 0.05 mm for all sizesPalm: 0.09 mm (all sizes), Finger: 0.15 mm (all sizes)
Tensile Strength (Before Aging)ASTM D6319-2019Min 14 Mpa for all sizesX-Small: 22.76 Mpa, Small: 22.77 Mpa, Medium: 24.46 Mpa, Large: 24.51 Mpa, X-Large: 24.59 Mpa
Tensile Strength (After Aging)ASTM D6319-2019Min 14 Mpa for all sizesX-Small: 20.50 Mpa, Small: 20.50 Mpa, Medium: 21.81 Mpa, Large: 21.95 Mpa, X-Large: 22.05 Mpa
Ultimate Elongation (Before Aging)ASTM D6319-2019Min 500% for all sizesX-Small: 884%, Small: 885%, Medium: 888%, Large: 891%, X-Large: 892%
Ultimate Elongation (After Aging)ASTM D6319-2019Min 400% for all sizesX-Small: 867%, Small: 868%, Medium: 870%, Large: 872%, X-Large: 875%
Water Tight (Detection of Holes)ASTM D5151-2019Passes AQL-1.5Gloves Passes AQL 1.5
Powder ResidueASTM D6124-06 (Reapproved 2017)≤2 mg/gloveX-small: 0.16 mg/glove, Small: 0.16 mg/glove, Medium: 0.16 mg/glove, Large: 0.16 mg/glove, X-Large: 0.16 mg/glove
Biocompatibility
Primary Skin IrritationISO 10993-10Under the condition of study not an irritantUnder the condition of study not an irritant
Dermal SensitizationISO 10993-10Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
In vitro CytotoxicityISO 10993-5:2009Under the conditions of study non cytotoxicUnder the conditions of the study cytotoxic (Note: This result is listed as "cytotoxic" while the acceptance criteria is "non cytotoxic", which could be a typo or indicates further consideration is needed).
Acute Systemic ToxicityISO 10993-11:2017Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern
Material Mediated PyrogenicityISO 10993-11:2017/USP 41<151>Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity responseUnder the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response

From the table, the device performance meets or exceeds the acceptance criteria for all tested characteristics, with the exception of the "In vitro Cytotoxicity" test where the reported result is "cytotoxic" while the acceptance criteria is "non cytotoxic". This discrepancy needs clarification. However, in the "SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS", the predicate also states "cytotoxic" for in vitro cytotoxicity, implying it might be an acceptable characteristic for this type of device or a different interpretation of cytotoxicity specific to gloves.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the specific sample sizes used for each test. For example, for "Water Tight (1000 ml)" test, it mentions "Passes AQL-1.5", which implies a sampling plan was used according to the ASTM standard, but the specific number of gloves tested is not provided. The data provenance (country of origin, retrospective/prospective) is also not specified, beyond the fact that the tests were conducted according to international standards (ASTM, ISO).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable. The device under review is a medical glove, and the evaluation relies on adherence to established physical, chemical, and biological performance standards, not expert interpretation of medical images or cases. Ground truth for these tests is based on objective measurements and established test methodologies defined in the ASTM and ISO standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of data (e.g., medical images) where expert consensus is required. The testing for this device involves objective, quantitative measurements against predefined standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is a medical glove. MRMC studies are relevant for AI-powered diagnostic tools where human readers (e.g., radiologists) interpret images with or without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a medical glove, not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for this device's performance is based on the objective measurements and predefined thresholds within recognized international standards (ASTM D6319-2019, ASTM D5151-2019, ASTM D6124-06, ISO 10993-10, ISO 10993-5:2009, ISO 10993-11:2017, USP 41<151>). For example:

  • Dimensions: Measured length and width against minimum/range requirements.
  • Physical Properties: Measured tensile strength and elongation against minimum percentage requirements.
  • Water Tightness: Statistical sampling and inspection against an Acceptable Quality Level (AQL).
  • Biocompatibility: Observation of biological responses (e.g., irritation, sensitization) following standardized protocols, compared against criteria for "not an irritant" or "not a sensitizer".

8. The sample size for the training set

This information is not applicable. The device is a medical glove, not an AI or machine learning model that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable, as there is no training set for this type of medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 28, 2023

Latrile Gloves Private Limited % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE114 Aurora, Illinois 60504

Re: K230601

Trade/Device Name: Blue Nitrile Examination Gloves Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: February 23, 2023 Received: March 3, 2023

Dear Manoi Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230601

Device Name

Blue Nitrile Examination Gloves Powder Free

Indications for Use (Describe)

Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY K230601 As required by: 21CFR§807.92

A. APPLICANTINFORMATION

510(K) Owner's NameLatrile Gloves Private Limited
AddressPlot No : B5 - Sipcot Induatrial Complex, 8th Main road,Pallapatti Village, Nilakottai Taluk, Dindigul Dist - 624201
Phone-----
Fax-----
E-mailinfo@latrile.com
Contact PersonRagu. M
DesignationQMS - Head
Contact Number8754672072
Contact Emailquality@latrile.com
Date Submitted22nd February 2023

B. DEVICE IDENTIFICATION

Name of the deviceBlue Nitrile Examination Gloves Powder Free
Product proprietary or trade nameQ-loves
Common or usual nameExam Gloves
Classification nameNon-powdered patient examination glove
Device ClassificationClass-1
Product CodeLZA
Regulation Number21 CFR 880.6250
Review PanelGeneral Hospital

C. PREDICATE DEVICE

Predicate DeviceJR Engineering & Medical Technologies (M) SDN.BHD.
510( K) NumberK192333
Regulatory Class1
Product codeLZA

D. DESCRIPTION OF THEDEVICE:

Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250).The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color, powder free and non sterile. The device is available in Extra Small, Small, Medium, Large and Extra Large sizes.

E. INTENDED USE OF THE DEVICE:

Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

{4}------------------------------------------------

F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATEDEVICE

CharacteristicsStandardsDevice PerformanceRemarks
PredicateSubject
510(K) NumberK192333K230601----
Name of deviceJR MEDIC Blue NitrileExamination GlovesPowder-freeBlue NitrileExamination GlovesPowder Free----
DimensionsASTM D6319-2019Length Min 230 m WidthMin 95+/-10mm(for medium size)Length Min 230 mm WidthMin 95+/-10mm(for medium size)Same
Physical PropertiesASTM D6319-2019Before AgingTensile Strengthmin 14 MpaUltimate ElongationMin 500%After AgingTensile Strengthmin 14 MpaUltimate ElongationMin 400%Before AgingTensile Strength min14 MpaUltimate ElongationMin 500%After AgingTensile Strengthmin 14 MpaUltimate ElongationMin 400%Same
ThicknessASTM D6319-2019Palm min 0.05 mmFinger min 0.05 mmPalm min 0.05 mmFinger min 0.05 mmSame
Powder ResidueASTM D6319-2019≤2 mg/glove≤2 mg/gloveSimilar
BiocompatibilityPrimary SkinIrritation-ISO 10993-10:2010(E)Under the condition ofstudy not an irritantUnder the conditionof study not anirritantSame
DermalSensitization- ISO10993-10:2010(E)Under the conditions ofthe study not a sensitizerUnder the conditionsof the study not asensitizerSame
In vitro cytotoxicityISO10993-5:2009(E)Under the conditions ofthe study, cytotoxicUnder the conditionsof the study cytotoxicSame
Acute SystemicToxicity Test ISO10993-11:2017(E)Under the conditionsof study the deviceextracts do not pose asystemic toxicity concernUnder the conditionsof study the deviceextracts do not pose asystemic toxicityconcernSame
Material MediatedPyrogenicity ISO10993-11:2017(E) /USP 41<151>Under the conditions ofthe study non pyrogenicUnder the conditionsof the study nonpyrogenicSame
CharacteristicsStandardsDevice PerformanceRemarks
PredicateCurrent
Water Tight (1000 ml)ASTM D5151-2019Passes AQL-1.5Passes AQL-1.5Similar
Intended useJR MEDIC Blue NitrileExamination GlovesPowder-free isdisposable devicesintended for medicalpurpose that are won onthe examiner's hand toprevent contaminationbetween patient andexaminer.Blue NitrileExamination GlovesPowder free isdisposable devicesintended for medicalpurpose that are wornon the examiner'shand to preventcontaminationbetween patient andexaminer.Similar
MaterialASTMD6319-2019NitrileNitrileSame
ColorBlueBlueSame
TextureFinger TextureFinger textureSame
SizeASTMD6319-2019Extra Small, Small,Medium, Large, ExtraLargeExtra Small, Small,Medium, Large, ExtraLargeSame
Single UseMedical GloveGuidance Manual- LabelingSingle UseSingle UseSame
Manufacturer(s)JR Engineering &MedicalTechnologies(M) SDN.BHD.MalaysiaLatrile Gloves PrivateLimited-----

{5}------------------------------------------------

There are no significant differences between the two products and are identical in terms of intended use, materials, design, manufacturing methods. Both devices meet the ASTM standard D6319-2019.

NON-CLINICAL TESTING SUMMARY

PERFORMANCE DATA
Test MethodPurposeAcceptance CriteriaResult
ASTM D6319-2019 StandardSpecification for NitrileExamination Gloves forMedical ApplicationTo determine thelength of the glovesMin 230 mm for all sizesX-Small:- 245 mmSmall:- 245 mmMedium:- 246mmLarge:- 248 mmX-Large:- 248 mm
ASTM D6319-2019 StandardSpecification for NitrileExamination Gloves forMedical ApplicationTo determine thewidth of the glovesX-small:- 70+/-10 mmSmall:- 80+/-10 mmMedium:- 95+/-10mmLarge:- 110+/-10 mmX-Large:- 120+/-10 mmX-small:- 76 mmSmall:- 84 mmMedium:- 94 mmLarge:- 105 mmX-Large:- 115 mm

{6}------------------------------------------------

Test MethodPurposeAcceptanceCriteriaResult
ASTM D6319-2019 StandardSpecification for NitrileExamination Gloves forMedical ApplicationTo determine thethickness of theglovesPalm 0.05 mm minFinger 0.05 mm minfor all sizesSizeX-SmallSmallMediumLargeX-LargePalm0.09mm0.09mm0.09mm0.09mm0.09mmFinger0.15mm0.15mm0.15mm0.15mm0.15mm
ASTM D6319-2019 StandardSpecification for NitrileExamination Gloves forMedical ApplicationTo Determine thephysical properties-Tensile strengthBefore AgeingTensile Strength14Mpa Min for allsizesAfter AgeingTensile Strength14Mpa Min for allsizesSizeX-SmallSmallMediumLargeX-LargeBeforeageing22.76Mpa22.77Mpa24.46Mpa24.51Mpa24.59MpaAfterageing20.50Mpa20.50Mpa21.81Mpa21.95Mpa22.05Mpa
ASTM D6319-2019 StandardSpecification for NitrileExamination Gloves forMedical ApplicationTo Determine thephysical properties-Ultimate ElongationBefore AgeingUltimate Elongation500% Min for allsizesAfter AgeingUltimateElongation 400%Min for all sizesSizeX-SmallSmallMediumLargeX-LargeBeforeageing884%885%888%891%892%Afterageing867%868%870%872%875%
Test MethodPurposeAcceptance CriteriaResult
ASTM D5151-2019 StandardTest Method for Detection ofHoles in Medical GlovesTo determine theholes in the glovesAQL 1.5Gloves Passes AQL 1.5
ASTM D6124-06 (Reapproved2017) Standard Test Methodfor Residual Powder onMedical GlovesTo determine theresidual powder inthe gloves2 Mg/Glove MaxSizeX-small 0.16mg/gloveSmall 0.16mg/gloveMedium 0.16 mg/gloveLarge 0.16 mg/gloveX-Large 0.16 mg/glove

{7}------------------------------------------------

BIO-COMPATIBILITY DATA

Test MethodPurposeAcceptanceCriteriaResult
ISO 10993-10 BiologicalEvaluation of Medical DevicesTest for Irritation and SkinSensitization. Test done forirritation.To determine the potential ofthe material under test toproduce dermal irritation inRabbitsUnder the conditionof study not anirritantUnder the condition ofstudy not an irritant
ISO 10993-10 BiologicalEvaluation of Medical DevicesTest for Irritation and SkinSensitization. Test done Skinsensitization.To determine the skinsensitization potential of thematerial both in terms ofinduction and elicitation inGuinea Pig.Under theconditions of thestudy not asensitizerUnder the conditions ofthe study not asensitizer
ISO 10993-5:2009 biologicalevaluation of medical devices -part 5, tests for in vitrocytotoxicity.To evaluate the in vitrocytotoxic potential of the testitem (both inner and outersurface) Extracts in L-929mouse fibroblasts cells usingelution method.Under theconditions of studynon cytotoxicUnder the conditions ofthe study cytotoxic.
ISO 10993-11:2017 biologicalevaluation of medical devices -part 11, tests for systemictoxicity.To determine the acutesystemic toxicity potential ofthe test item extracts (bothinside and outer surfaces) inSwiss Albino mice.Under theconditions of studythe device extractsdo not pose asystemic toxicityconcernUnder the conditionsof study the deviceextracts do not pose asystemic toxicityconcern
Material MediatedPyrogenicity ISO 10993-11:2017(E) / USP 41<151>To determine the pyrogenicpotential of the test itemextract following intravenousinjection in New Zealandwhite Rabbits.Under theconditions of thestudy, the devicedid notdemonstrate amaterial mediatedpyrogenicityresponse.Under the conditions ofthe study, the device didnot demonstrate amaterial mediatedpyrogenicity response.

G. Clinical Testing Summary

Not applicable - Clinical data is not needed for gloves.

H. CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.