K Number
K223872

Validate with FDA (Live)

Date Cleared
2023-01-19

(27 days)

Product Code
Regulation Number
870.1340
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Merit Low Profile Companion Sheath is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.

Device Description

The Low Profile Companion Sheath is a sterile, disposable device consisting of (a) a coil reinforced shaft with and the distal end; (b) a hemostasis valve with a side port and (c) a tapered tip dilator with snap-fit hub at the proximal end.

(a) Shaft. The coil reinforced, multi-layer polymer shaft contains a tapered tip at the distal end. A continuous inner PTFE tube forms the core of the shaft and provides a circular working lument through which devices can be passed. A hydrophilic coating is applied to the entire outer surface of the shaft. A radiopaque marker made of platinum iridium is embedded 5mm from the dista At the proximal end of the shaft, a female, winged luer hub is over-molded onto the shaft to support handling and to provide for the connection of the hemostasis valve.

(b) Hemostasis valve. A removable hemostasis valve is thread onto the proximal end of the shaft. Inside the valve housing, a lubricated, silicone slit disc provides a seal around devices passed through the sheath, thereby preventing blood leakage through the valve. Just distal of the valve housing is connected to a side port leading to a three-way stopcock valve. The sideport is used for flushing the introducer sheath.

(c) Dilator. The dilator made of a polypropylene blend contains a full-length round lumen to allow placement over a guidewire. The distal end of the dilator is configured as a tapered tip that extends about 2 cm beyond the end of the dilator is fully inserted through the sheath.

The Low Profile Companion Sheath is a prescription medical device that is used only in healthcare facilities and hospitals. The device is placed in patients for up to 24 hours.

AI/ML Overview

This is a 510(k) summary for a medical device called the "Low Profile Companion Sheath." This document describes the device and claims substantial equivalence to a previously cleared device (predicate device). For such submissions, the acceptance criteria and study information provided generally focus on demonstrating that the new device performs comparably to the predicate or meets established performance specifications. The details for acceptance criteria and studies are typically more concise than for novel device approvals.

Here's an analysis of your requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

Based on the text, the acceptance criteria are implicitly tied to meeting "predetermined specifications" for the "changed dimensions of the device." The specific numerical acceptance criteria are not explicitly provided in this summary, but the general categories of tests and their successful outcome are stated.

Acceptance Criteria CategoryReported Device Performance
Hydrophilic Coated LengthMet predetermined specifications for changed dimensions.
Sheath Tip to Dilator Taper LengthMet predetermined specifications for changed dimensions.
Sheath Effective LengthMet predetermined specifications for changed dimensions.
Dimensions (General)Met acceptance criteria applicable to changed dimensions.
Other performance tests (implicitly similar to predicate)No new questions of safety and effectiveness; performs comparably to predicate.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample size used for the design verification tests. It mentions "design verification tests" were performed. The "data provenance" (country of origin, retrospective/prospective) is also not specified, as these are typically bench-top engineering tests rather than clinical studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is a technical device submission, not an AI/diagnostic software submission where expert adjudication is common for ground truth. Therefore, this information is not applicable and not provided in the document. The "ground truth" here is defined by engineering specifications and physical measurements.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

As this is a technical device submission involving engineering measurements and performance testing, an "adjudication method" in the context of expert review for diagnostic accuracy is not applicable and not mentioned. The tests would likely be performed by qualified engineers/technicians according to established protocols.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is relevant for diagnostic software (often AI-powered) where human interpretation is involved. This submission is for a physical medical device (catheter introducer sheath). Therefore, such a study was not performed and is not applicable to this device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This refers to AI algorithm performance. Since this is a physical medical device, there is no AI algorithm involved, and thus no standalone performance was evaluated in this context.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the design verification tests, the "ground truth" would be the predetermined engineering specifications and physical measurements for the device's dimensions and characteristics. This is not "expert consensus," "pathology," or "outcomes data" in the clinical sense.

8. The sample size for the training set

This question refers to the training of an AI algorithm. Since this is a physical medical device and does not involve an AI algorithm, there is no "training set."

9. How the ground truth for the training set was established

Again, this question is relevant for AI algorithm development. As there is no AI algorithm or training set for this physical medical device, this information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 19, 2023

Merit Medical Systems, Inc. Kirk Mcintosh Senior Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, Utah 84095

Re: K223872

Trade/Device Name: Low Profile Companion Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB, DRE Dated: December 23, 2022 Received: December 23, 2022

Dear Kirk Mcintosh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Finn E. Finn E. Donaldson -S Date: 2023.01.19 10:26:47 Donaldson -S -05'00'

For

Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223872

Device Name Low Profile Companion Sheath

Indications for Use (Describe)

The Merit Low Profile Companion Sheath is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Construction (ID #1-077-001-0001 or Plan)
On-Site Septic System (ID #1-077)

|× | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Prepared on: 2023-01-19

Contact Details21 CFR 807.92(a)(1)
Applicant NameMerit Medical Systems, Inc.
Applicant Address1600 West Merit Parkway South Jordan UT 84095 United States
Applicant Contact Telephone(801) 316-3695
Applicant ContactMr. Kirk McIntosh
Applicant Contact Emailkirk.mcintosh@merit.com
Device Name21 CFR 807.92(a)(2)
Device Trade NameLow Profile Companion Sheath
Common NameCatheter introducer
Classification NameIntroducer, Catheter
Regulation Number870.1340
Product CodeDYB
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K211405Prelude® Guide SheathDYB
Device Description Summary21 CFR 807.92(a)(4)
-------------------------------------------------

General Description:

The Low Profile Companion Sheath is a sterile, disposable device consisting of (a) a coil reinforced shaft with and the distal end; (b) a hemostasis valve with a side port and (c) a tapered tip dilator with snap-fit hub at the proximal end.

(a) Shaft. The coil reinforced, multi-layer polymer shaft contains a tapered tip at the distal end. A continuous inner PTFE tube forms the core of the shaft and provides a circular working lument through which devices can be passed. A hydrophilic coating is applied to the entire outer surface of the shaft. A radiopaque marker made of platinum iridium is embedded 5mm from the dista At the proximal end of the shaft, a female, winged luer hub is over-molded onto the shaft to support handling and to provide for the connection of the hemostasis valve.

(b) Hemostasis valve. A removable hemostasis valve is thread onto the proximal end of the shaft. Inside the valve housing, a lubricated, silicone slit disc provides a seal around devices passed through the sheath, thereby preventing blood leakage through the valve. Just distal of the valve housing is connected to a side port leading to a three-way stopcock valve. The sideport is used for flushing the introducer sheath.

(c) Dilator. The dilator made of a polypropylene blend contains a full-length round lumen to allow placement over a guidewire. The distal end of the dilator is configured as a tapered tip that extends about 2 cm beyond the end of the dilator is fully inserted through the sheath.

The Low Profile Companion Sheath is a prescription medical device that is used only in healthcare facilities and hospitals. The device is placed in patients for up to 24 hours.

Principle of Operation:

The device is operated manually and is to be handled aseptically by qualified medical personnel familiar with its application. After removal from the sterile pouch packaging using aseptic techniques, the device is placed into the vasculature. Prior to use, the

{4}------------------------------------------------

introducer sheath and dilator are flushed with hepanized solution. The dilator is then inserted completely into the introducer sheath and locked into place through the snap fit connection at the housing of the diator-sheath-combination is then passed as one unit over a quide wire with a maximum size of up to 0.038 inches. Once the introducer sheath is fully placed in the patient, the quide wire and dilator are removed and compatible catheters and inserted through the introducer sheath. For removal of the sheath, a guide wire is inserted past the dilator is passed over the wire into the sheath. Sheath and dilator are then removed as one unit.

Device Models:

The lone configuration of the Low Profile Companion Sheath is a 7-French diameter, 30 cm in length and a straight shape.

Comparison of Technological Characteristics with the Predicate Devices:

The Low Profile Companion Sheath is manually operated, sterile, single use sheath system made predominantly of thermoplastic polymers. The sheath is reinforced with a stainless-steel coll which provides kink resistance when paths. With regard to the design, device features, method of sterilization, and mode of operation, the Low Profile Companion Sheath does not differ from the predicate devices. Materials used to manufacture the Low Profile Companion Sheath are identical to those used in the legally marketed predicate device.

Other than the shorter length there are no differences in the technological characteristics between the predicate device. The Low Profile Companion Sheath complies to the same standards as the predicate device.

Intended Use/Indications for Use

The Merit Low Profile Companion Sheath is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.

Indications for Use Comparison

There is no change in the Indications for Use statement from the predicate to the subject device.

Technological Comparison

The subject device is substantially equivalent to the based on an identical Indications for Use statement, the same basic performance and safety profile, principle of operation, funciples, and manufacturing technology.

The primary reason for submitting this Special 510(k) is to document an extension to the product line. The subject device is shorter in length than the shortest cleared configuration under K211405.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The following design verification tests were performed to demonstrate that the Low Profile Guide Sheath met predetermined specifications and to assure that there were no unacceptable risks associated with the modifications:

Hydrophilic Coated Length Sheath Tip to Dilator Taper Length Sheath Effective Length

The results of the testing demonstrated that the subject Low Profile Companion Sheath met the acceptance criteria applicable to the changed dimensions of the device.

Based on the indications for use, design, safety and performance testing, the subject Low Profile Companion Sheath raises no new questions of safety and effectiveness compared to the predicate device and is substantially equivalent to the predicate device manufactured by Merit Medical Systems, Inc.

21 CFR 807.92(a)(6)

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(5)

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).