K Number
K211405

Validate with FDA (Live)

Date Cleared
2021-10-04

(151 days)

Product Code
Regulation Number
870.1340
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Prelude Guide Sheath Introducer is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.

Device Description

The Prelude Guide Sheath Introducer is a sterile, disposable device consisting of a (a) a coil reinforced shaft with an atraumatic tip and the distal end: (b) a hemostasis valve with a side port and color-coded stopcock; and (c) a tapered tip dilator with snap-fit hub at the proximal end.

AI/ML Overview

This is a 510(k) summary for a medical device (Prelude Guide Sheath). The document focuses on demonstrating substantial equivalence to a predicate device based on non-clinical performance data and material comparisons, not on evaluating an AI/ML algorithm. Therefore, many of the requested criteria related to AI/ML device studies (such as sample size for test/training sets, expert ground truth, MRMC studies, standalone performance, etc.) are not applicable or readily extractable from this specific document.

However, I can provide the acceptance criteria and performance as described for the non-clinical tests.

Acceptance Criteria and Reported Device Performance

The document states that "The results of the testing demonstrated that the subject Prelude Guide Sheath met the predetermined acceptance criteria applicable to the performance of the device." The specific numerical acceptance criteria for each test are not provided in this summary, but the general categories of tests performed are listed. The "Reported Device Performance" for each test is simply that the device met those unstated criteria.

Here's a table based on the provided "Performance Tests" section:

No.Verification/Validation TestApplicable Standard(s)Acceptance Criteria (Implied)Reported Device Performance
1Insertion Force/Tip Insertion, Peel BackInternal RequirementMet internal requirementsMet predetermined criteria
2Peak Tensile Force Shaft/Hub & TipISO 10555-1:2013, ISO 10555-3:2013, ISO 11070:2014Met standards' requirementsMet predetermined criteria
3Visual (Extraneous/Surface Defects)ISO 11070:2014Met standards' requirementsMet predetermined criteria
4Sheath StiffnessInternal RequirementMet internal requirementsMet predetermined criteria
5Sheath KinkTested through ValidationMet validation requirementsMet predetermined criteria
6Radiopacity marker band & tipISO 11070:2014, ASTM F640-12, ISO 10555-1:2013Met standards' requirementsMet predetermined criteria
7Hydrophilic coating durability & coverage/coating lengthInternal RequirementMet internal requirementsMet predetermined criteria
8Hydrophilic Coating LubricityTested through ValidationMet validation requirementsMet predetermined criteria
9Coating integrity test (particulate evaluation)ASTM F1877-16, Class II Special Controls Guidance for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) CathetersMet standards' requirementsMet predetermined criteria
10Corrosion resistanceISO 11070:2014Met standards' requirementsMet predetermined criteria
11Sheath assembly leakISO 11070:2014Met standards' requirementsMet predetermined criteria
12Hydrogen Peroxide ResistanceISO 10555-1:2013, ISO 10555-3:2013Met standards' requirementsMet predetermined criteria
13Radiopacity DilatorISO 11070:2014, ISO 10555-1:2013Met standards' requirementsMet predetermined criteria
14Dilator tip fractureInternal RequirementMet internal requirementsMet predetermined criteria
15Dilator-to-Hub TensileISO 10555-1:2013Met standards' requirementsMet predetermined criteria
16Sheath tip to dilator taper lengthInternal RequirementMet internal requirementsMet predetermined criteria
17Dilator Stiffness & Hub Snap ForceNo Standard - Tested through ValidationMet validation requirementsMet predetermined criteria
18Dilator Insertion/Removal from SheathNo Standard - Testing through ValidationMet validation requirementsMet predetermined criteria
19Dimensional tests for OD & ID of sheath & dilatorISO 11070:2014Met standards' requirementsMet predetermined criteria
20Curve shape & visual for sheath & dilatorInternal RequirementMet internal requirementsMet predetermined criteria
21Curve RetentionNo standard - Testing through ValidationMet validation requirementsMet predetermined criteria
22Sheath Effective LengthISO 10555-1:2013Met standards' requirementsMet predetermined criteria
23Luer Design TestsISO 11070:2014, ISO 594-1: 1986, ISO 594-2:1998Met standards' requirementsMet predetermined criteria
24Hemostasis Valve LeakageISO 11070:2014Met standards' requirementsMet predetermined criteria
25EO Sterilization ValidationISO 11135:2014, ISO 10993-7:2008, AAMI TIR28:2016Met standards' requirementsMet predetermined criteria
26Package Integrity TestingISO 11607-1:2019, ISO 11607-2:2019, ASTM D4169-16Met standards' requirementsMet predetermined criteria
27Biocompatibility Testing (Cytotoxicity, Sensitization, etc.)ISO 10993-1:2009, ISO 10993-4: 2017, ISO 10993-5:2009, ISO 10993-6:2009, ISO 10993-7:2008, ISO 10993-10:2010, ISO 10993-11:2017, ISO 10993-12:2012Met standards' requirementsMet predetermined criteria
28Chemical Characterization – Particulate AnalysisASTM F1877-16Met standards' requirementsMet predetermined criteria

The remaining requested information is largely irrelevant to this 510(k) submission, as it pertains to AI/ML algorithm validation for which this device is not. This submission focuses on the physical and material characteristics of a medical device (a guide sheath introducer) and its substantial equivalence to existing predicate devices, rather than the performance of an AI/ML system.

Therefore, the following information is not applicable or extractable from the provided document:

  1. Sample sized used for the test set and the data provenance: Not an AI/ML study, so "test sets" in that context don't apply. The non-clinical tests would have their own sample sizes for a variety of physical or chemical tests, but these are not detailed as a single "test set" for an algorithm.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not an AI/ML study; no "ground truth" in the diagnostic/interpretive sense is established by experts here.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable.
  7. The sample size for the training set: Not applicable.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side, there is a symbol representing the Department of Health & Human Services - USA. To the right, there is a blue square with the letters 'FDA' in white. Next to the blue square, the words 'U.S. FOOD & DRUG' are written in blue, with 'ADMINISTRATION' written in a smaller font size below.

October 4, 2021

Merit Medical Systems, Inc. David Thomas Principal Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, Utah 84095

Re: K211405

Trade/Device Name: Prelude Guide Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB, DRE Dated: August 31, 2021 Received: September 1, 2021

Dear David Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Finn Donaldson Assistant Director (Acting) DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211405

Device Name Prelude® Guide Sheath

Indications for Use (Describe)

The Prelude Guide Sheath Introducer is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K211405

Prelude Guide Sheath

GeneralProvisionsSubmitter Name:Address:Telephone Number:Fax Number:Contact Person:Date of Preparation:Registration Number:Merit Medical Systems, Inc.1600 West Merit ParkwaySouth Jordan, UT 84095(801) 316-4956(801) 208-3365David ThomasAugust 31, 20211721504
SubjectDeviceTrade Name:Common/Usual Name:Classification Name:Prelude® Guide SheathSheath IntroducerCatheter Introducer
PredicateDeviceTrade Name:Classification Name:Premarket Notification:Manufacturer:Manufacturer:DuraSheath Introducer Sheath SystemCatheter IntroducerK181463 (Predicate)Contract Medical InternationalK091329 (Reference Device) – PinnacleDestination Peripheral Guiding SheathTerumo Medical Corp
ClassificationClass II21 CFR § 870.1340FDA Product Code: DYB, DREReview Panel: Cardiovascular

{4}------------------------------------------------

Indications for Use

The Prelude Guide Sheath Introducer is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.

A. Device Description

The Prelude Guide Sheath Introducer is a sterile, disposable device consisting of a (a) a coil reinforced shaft with an atraumatic tip and the distal end: (b) a hemostasis valve with a side port and color-coded stopcock; and (c) a tapered tip dilator with snap-fit hub at the proximal end.

(a) Shaft. The coil reinforced, multi-layer polymer shaft contains a tapered tip at the distal end. A continuous inner PTFE tube forms the core of the shaft and provides a circular working lumen through which devices can be passed. A single, full length Pebax tube creates the outer cover of the shaft. A stainless-steel flat wire coil is fused between the two polymer tubes along the entire length of the shaft. Hydrophilic coating is applied on the distal 35cm of the shaft. A radiopaque marker made of platinum iridium is embedded 5mm from the distal end of the shaft. At the proximal end of the shaft, a female, winged luer hub is over-molded onto the shaft to support handling and to provide for the connection of the hemostasis valve. This hub is color coded to match the French size of the device.

(b) Hemostasis valve. A removable hemostasis valve is thread onto the luer hub at the proximal end of the shaft. Inside the valve housing, a lubricated, silicone slit disc provides a seal around devices passed through the sheath, thereby preventing blood leakage through the valve. Just distal of the valve housing is connected to a side port leading to a threeway stopcock valve. The sideport is used for flushing the introducer sheath. At the proximal end of the valve housing, the color-coded housing cap provides a snap fit connection the hub of the dilator.

(c) Dilator. The dilator made of a polypropylene blend with Pebax and Barium sulphate (4F) or polypropylene blend with Bismuth Oxychloride (5-8F) contains a full-length round lumen to allow placement over guidewire. The distal end of the dilator is configured as a tapered tip that extends about 2 cm beyond the end of the sheath when the dilator is fully inserted through the sheath. An integral, color coded luer hub that is over-molded onto the proximal end the dilator supports handling of the dilator and provides a snap fit connection to the valve housing cap at the proximal end of the sheath introducer.

{5}------------------------------------------------

B. Operation and Compatibility

After removal from the sterile pouch packaging using aseptic techniques, the device is placed into the vasculature. Prior to use, the introducer sheath and dilator are flushed with heparinized solution. The dilator is then inserted completely into the introducer sheath and locked into place through the snap fit connection at the housing of the hemostasis valve. The dilator-sheath-combination is then passed as one unit over a guide wire with a maximum size of up to 0.035 inches for 4 French and up to 0.038 inches for 5 French through 8 French models. Guide wire is not part of the device. Once the introducer sheath is fully placed in the patient, the guide wire and dilator are removed and compatible catheters and instruments can be inserted through the introducer sheath.

C. Device Models

The Prelude Guide Sheath consists of twenty-nine (29) models of different sizes (4 French, 5 French, 6 French, 7 French and 8 French) and effective lengths (45cm, 65cm and 90 cm).

D. Comparison of Technological Characteristics with the Predicate Devices

With regard to the design, device features, method of sterilization, and mode of operation, the Prelude Guide Sheath does not differ from the predicate devices. Materials used to manufacture the Prelude Guide Sheath are identical to that used in the legally marketed predicate device, or very similar to those contained in the legally marketed reference device.

Technological characteristics of the subject device do not differ from the predicate device and differ only with respect to materials for selected components and the choice of lubricious coating on the sheath from a hydrophobic coating to a hydrophilic coating which is similar to the reference device. Both the Pebax used for the outer layer and the hydrophilic coating applied to the distal end of the Prelude Guide Sheath are commonly used in medical devices, including introducer sheaths and catheters. There are no differences in the material and technological characteristics between the subject device and the predicate device.

E. Performance Tests

Nonclinical performance data demonstrates that the Prelude Guide Sheath is substantially equivalent to the predicate. The following performance data from the non-clinical tests were provided to support the substantial equivalence determination:

  • Manual testing, including tests required under relevant international standards, ● coating adhesion and particulate testing, transportation integrity testing, performed to verify the validate the design.
  • Biocompatibility Risk Assessment (BRA) and biocompatibility device testing to ● demonstrate compatibility.
  • Sterilization information to confirm sterility of the device upon exposure to the . selected sterilization cycle.
  • Accelerated aging testing to confirm product performance at end of shelf life. ●

{6}------------------------------------------------

The list of tests performed to support determination of substantial equivalence is provided in the table below.

Testing Performed
No,Verification/Validation TestApplicable Standard(s)
1Insertion Force/Tip InsertionPeel BackInternal Requirement
2Peak Tensile Force Shaft/Hub &TipISO 10555-1:2013ISO 10555-3:2013ISO 11070:2014
3Visual (Extraneous/SurfaceDefects)ISO 11070:2014
4Sheath StiffnessInternal Requirement
5Sheath KinkTested through Validation
6Radiopacity marker band & tipISO 11070:2014, ASTM F640-12ISO 10555-1:2013
7Hydrophilic coating durability &coverage/coating lengthInternal Requirement
8Hydrophilic Coating LubricityTested through Validation
9Coating integrity test (particulateevaluation)ASTM F1877-16Class II Special Controls Guidance forCertain Percutaneous TransluminalCoronary Angioplasty (PTCA) Catheters
10Corrosion resistanceISO 11070:2014
11Sheath assembly leakISO 11070:2014
12Hydrogen Peroxide ResistanceISO 10555-1:2013ISO 10555-3:2013
13Radiopacity DilatorISO 11070:2014ISO 10555-1:2013
14Dilator tip fractureInternal Requirement
15Dilator-to-Hub TensileISO 10555-1:2013
16Sheath tip to dilator taper lengthInternal Requirement
17Dilator Stiffness & Hub SnapForceNo Standard-Tested through Validation
18Dilator Insertion/Removal from SheathNo Standard - Testing through Validation
19Dimensional tests for OD & ID of sheath & dilatorISO 11070:2014
20Curve shape & visual for sheath & dilatorInternal Requirement
21Curve RetentionNo standard - Testing through Validation
22Sheath Effective LengthISO 10555-1:2013
23Luer Design TestsISO 11070:2014ISO 594-1: 1986ISO 594-2:1998
24Hemostasis Valve LeakageISO 11070:2014
25EO Sterilization ValidationISO 11135:2014ISO 10993-7:2008AAMI TIR28:2016
26Package Integrity TestingISO 11607-1:2019ISO 11607-2:2019ASTM D4169-16
27Biocompatibility TestingCytotoxicity TestingSensitizationIrritationAcute Systemic ToxicityPyrogenicityHemocompatibility• ASTM Hemolysis Study – Direct Contact and Extract Method: Phosphate Buffered Saline ExtractionISO 10993-1:2009ISO 10993-4: 2017ISO 10993-5:2009ISO 10993-6:2009ISO 10993-7:2008ISO 10993-10:2010ISO 10993-11:2017ISO 10993-12:2012
οComplement Activation(SC5b-9 only) Assay
Thrombogenicity Study – In-vitroBlood Loop Assay, PartialThromboplastin Time (PTT)
28Chemical Characterization –Particulate AnalysisASTM F1877-16

{7}------------------------------------------------

{8}------------------------------------------------

F. Conclusions

The results of the testing demonstrated that the subject Prelude Guide Sheath met the predetermined acceptance criteria applicable to the performance of the device.

Based on the indications for use, design, safety and performance testing, the subject Prelude Guide Sheath raises no new questions of safety and effectiveness compared to the predicate device and is substantially equivalent to the predicate device, DuraSheath Introducer Sheath System K181463 manufactured by Contract Medical International.

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).