K Number
K223791

Validate with FDA (Live)

Device Name
TalWire
Date Cleared
2023-07-14

(207 days)

Product Code
Regulation Number
870.1330
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TalWire is intended for percutaneous entry of peripheral vessels using the Seldinger Technique. The TalWire is not intended for use in the coronary or cerebral vasculature.

Device Description

The TalWire is a family of vascular access mandrel guidewires. The TalWire is available in various sizes: between 40 to 80 cm in length and in the 0.018" diameter range. The TalWire is available with Nitinol (NiTi) cores and Stainless Steel coils. The TalWire consist of a solid core shaft with a ground tapered section at the distal end of the guidewire. A microcoil is wound with a lumen that is then placed over the tapered distal end of the coil is secured to the shaft via a weld and the proximal end is bonded to the shaft using adhesive. The TalWire includes an articulatable tip at the distal end of the guidewire. The finished guidewire is placed in a protective polymer dispenser hoop in a labeled Tyvek pouch. The TalWire is provided as a sterile, single-use device, and is sterilized using a validated ethylene oxide process.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called TalWire, a catheter guidewire. It compares the TalWire to a predicate device and includes performance data to support substantial equivalence.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The document states that "All tests passed and met the predefined acceptance criteria." However, it does not explicitly list the quantitative acceptance criteria for each performance test. It only lists the types of performance tests conducted.

TestAcceptance Criteria (Not Explicitly Stated, Assumed to be "Pass")Reported Device Performance
Dimensional Attributes(Assumed: Meets design specifications)Passed
Visual Inspection(Assumed: No defects)Passed
Simulated Use(Assumed: Performs as intended in simulated environment)Passed
Tensile Strength(Assumed: Meets strength requirements)Passed
Torque Strength(Assumed: Meets torque requirements)Passed
Torqueability(Assumed: Adequate torque response)Passed
Corrosion Resistance(Assumed: No significant corrosion)Passed
Kink Resistance(Assumed: Resists kinking)Passed
Tip Flexibility(Assumed: Adequate flexibility)Passed
Radiopacity(Assumed: Visible under fluoroscopy)Passed
Flex Resistance(Assumed: Resists fatigue/damage from flexing)Passed
Fracture Resistance(Assumed: Resists fracture)Passed
Prolapse Force(Assumed: Meets prolapse force requirements)Passed

Biocompatibility:

  • Evaluation in compliance with ISO 10993-1.
  • Specific tests conducted: Cytotoxicity, Guinea Pig Maximization Sensitization, Intracutaneous Irritation, Material Mediated Pyrogenicity, Acute Systemic Toxicity, ASTM Hemolysis, SC5b-9 Complement Activation Assay, Standard Thrombogenicity in Canine.
  • Reported device performance: "The device was found biocompatible."

Sterilization, Packaging and Shelf Life Testing:

  • EtO Sterilization validation (SAL 10^-6).
  • Shelf life and packaging testing after environmental conditioning and transportation simulation.
  • Reported device performance: "All tests were successfully completed."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample sizes for any of the performance tests.
The data provenance is not explicitly stated beyond being "performance data." This type of testing is typically done in a lab setting rather than using patient data, so concepts like "country of origin" or "retrospective/prospective" studies are not applicable in the usual sense for this kind of device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not applicable as the tests described are performance tests of the physical properties and functionality of the guidewire, not assessments of diagnostic images or clinical outcomes that would require expert human interpretation as "ground truth." These tests rely on standardized methods and measurement equipment.

4. Adjudication Method for the Test Set

This information is not applicable for the same reasons as point 3. Performance tests are typically objectively measured against predefined criteria, not adjudicated by experts.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating the substantial equivalence of the device through performance testing against established standards and comparison to a predicate device, not on assessing human reader performance with or without AI assistance. This device is a physical medical device (guidewire), not an AI-driven diagnostic tool.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

No, a standalone algorithm performance study was not done. This device is a physical guidewire, not an algorithm.

7. The Type of Ground Truth Used

For the performance and biocompatibility testing, the "ground truth" is established by:

  • Standardized test methods: Based on FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling Guidance for Industry and Food and Drug Administration Staff" and ISO 10993-1 for biocompatibility.
  • Predefined acceptance criteria: While not explicitly listed, these criteria would typically be quantitative thresholds or qualitative observations (e.g., "no visible corrosion," "no fracture").

8. The Sample Size for the Training Set

This information is not applicable. This is a physical medical device, not an AI model that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 14, 2023

Embrace Medical Ltd. % Orly Maor Regulatory Consultant 25 Sirkin Street Kfar Saba, 4442157 Israel

Re: K223791

Trade/Device Name: TalWire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: June 12, 2023 Received: June 14, 2023

Dear Orly Maor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Digitally signed by Lydia S. Lydia S. Glaw Glaw -S Date: 2023.07.14 12:19:11 -04'00'

Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use Statement

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K223791

Device Name

TalWire

Indications for Use (Describe)

The TalWire is intended for percutaneous entry of peripheral vessels using the Seldinger Technique. The TalWire is not intended for use in the coronary or cerebral vasculature.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.The burden time for this collection of information is estimated to average 79 hours per response, includingthe time to review instructions, search existing data sources, gather and maintain the data needed andcomplete and review the collection of information. Send comments regarding this burden estimate or anyother aspect of this information collection, including suggestions for reducing this burden, to:Department of Health and HumanServices Food and Drug AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA)Staff PRAStaff@fda.hhs.gov"An agency may not conduct or sponsor, and a person is not required to respond to, a collection ofinformation unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1 FDA

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

K223791

Traditional Premarket Notification Submission - 510(k) Embrace Medical Ltd.- TalWire

Date Prepared: December 14, 2022

SUBMITTER I.

Embrace Medical Ltd. 21(B) Habarzel Street. Tel-Aviv, 6971029, Israel Tel: +972-775055645 E-mail: anat@accessmv.com

Contact Person

Orly Maor 25 Sirkin Street Kfar Saba, 4442157 Israel Tel: +972-9-7453607 oram.ma@gmail.com

II. DEVICE

Name of Device: TalWire Common or Usual Name: TalWire Classification Name: 21CFR §870.1330- Wire, Guide, Catheter Regulatory Class: II Product Code: DQX

PREDICATE DEVICE II.

Embrace Medical Ltd. believes that the Tal Wire is substantially equivalent to the following predicate device:

  • Vascular Solutions, Inc., VSI Guidewire, cleared under K112631. .

IV. DEVICE DESCRIPTION

The TalWire is a family of vascular access mandrel guidewires. The TalWire is available in various sizes: between 40 to 80 cm in length and in the 0.018" diameter range. The TalWire is available with Nitinol (NiTi) cores and Stainless Steel coils.

The TalWire consist of a solid core shaft with a ground tapered section at the distal end of the guidewire. A microcoil is wound with a lumen that is then placed over the tapered distal end of the coil is secured to the shaft via a weld and the proximal end is bonded to the shaft using adhesive. The TalWire includes an articulatable tip at the distal end of the guidewire.

The finished guidewire is placed in a protective polymer dispenser hoop in a labeled Tyvek pouch. The TalWire is provided as a sterile, single-use device, and is sterilized using a validated ethylene oxide process.

INDICATIONS FOR USE V.

The TalWire is intended for percutaneous entry of peripheral vessels using the Seldinger Technique. The TalWire is not intended for use in the coronary or cerebral vasculature.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VI.

The TalWire has the same intended use as the predicate device. Its indications for use are identical to that of the predicate device.

The TalWire has similar technological characteristics as the predicate device as demonstrated in the table below:

{4}------------------------------------------------

SpecificationEmbrace Medical Ltd.TalWireVascular Solutions, Inc.,VSI GuidewireSE Justification
510K NumberK 223791K112631
Regulation Number21CFR §870.133021CFR §870.1330Same
Product CodeDQXDQXSame
ClassificationClass IIClass IISame
Indications for UseThe TalWire is intended forpercutaneous entry of peripheralvessels using the SeldingerTechnique. The TalWire is notintended for use in the coronaryor cerebral vasculature.The VSI Guidewire is indicatedfor percutaneous entry ofperipheral vessels using theSeldinger Technique. The deviceis not indicated for use in thecoronary or cerebral vasculature.Same
Anatomical LocationPeripheral vesselsPeripheral vesselsSame
Wire Diameter0.018"-0.021"0.018"SEWithin the range ofcleared guidewiresSuccessfulverification testingconfirmed theequivalence
Device Length40-80cm40cm - 130cmSEWithin the predicaterange
Tip Type and ShapeStraightStraight or AngledSEWithin the predicaterange
Wire Mandrel (core)MaterialNitinolStainless Steel or NitinolSEWithin the predicaterange
Coating Material,Length and LocationNoneNone or PTFE coatedSEWithin the predicaterange
Coil MaterialStainless SteelStainless Steel, TungstenSEWithin the predicaterange
AccessoriesNoneNone or may be supplied in amicro-introducer access kitSE
PackagingconfigurationThe wire is placed in a protectivepolymer dispenser hoop in alabeled TyveK pouch.5 pack configurationThe wire is placed in aprotective polymer dispenserhoop in a labeled TyveK pouch.5 or 10 pack configurationSEWithin the predicaterange
Single Use orReusableSingle UseSingle UseSame
Sterilization MethodSterilized by EtO, SAL 10-6Sterilized by EtO, SAL 10-6Same
SpecificationEmbrace Medical Ltd.TalWireVascular Solutions, Inc.,VSI GuidewireSE Justification
BiocompatibilityISO 10993-1ISO 10993-1Same

{5}------------------------------------------------

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination: The tests were performed based on Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling Guidance for Industry and Food and Drug Administration Staff. Dated: October 10, 2019

Biocompatibility -

Biocompatibility evaluation in compliance with ISO 10993-1 was performed. The following tests were conducted: Cytotoxicity Study Using the ISO Elution Method ISO Guinea Pig Maximization Sensitization Test ISO Intracutaneous -irritation Study in Rabbits Material Mediated Pyrogenicity in Rabbits Acute Systemic Toxicity ASTM Hemolysis Study SC5b-9 Complement Activation Assay Standard Thrombogenicity in Canine In addition, chemical characterization and Biological Risk Assessment were performed. The device was found biocompatible.

Sterilization, Packaging and Shelf Life Testing

EtO Sterilization validation was performed. In addition, shelf life and packaging testing were performed to support the labeled shelf life. All tests, including packaging integrity and device performance, were performed after environmental conditioning and transportation simulation. All tests were successfully completed.

Performance Testing

Performance testing per FDA guidance Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling Guidance for Industry and Food and Drug Administration Staff. Dated: October 10, 2019

included the following:

Tests
• Dimensional Attributes• Kink resistance
• Visual Inspection• Tip flexibility
• Simulated use• Radiopacity
• Tensile strength• Flex resistance
• Torque strength• Fracture resistance
• Torqueability• Prolapse force
• Corrosion resistance

All tests passed and met the predefined acceptance criteria.

{6}------------------------------------------------

VIII. CONCLUSIONS

The results of the comparison of design, intended use and technological characteristics demonstrate that the device is as safe and effective as the legally marketed predicate device. Therefore, Embrace Medical Ltd. concludes that the proposed TalWire is substantially equivalent to the identified predicate device.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.