Search Results
Found 1 results
510(k) Data Aggregation
(207 days)
TalWire
The TalWire is intended for percutaneous entry of peripheral vessels using the Seldinger Technique. The TalWire is not intended for use in the coronary or cerebral vasculature.
The TalWire is a family of vascular access mandrel guidewires. The TalWire is available in various sizes: between 40 to 80 cm in length and in the 0.018" diameter range. The TalWire is available with Nitinol (NiTi) cores and Stainless Steel coils. The TalWire consist of a solid core shaft with a ground tapered section at the distal end of the guidewire. A microcoil is wound with a lumen that is then placed over the tapered distal end of the coil is secured to the shaft via a weld and the proximal end is bonded to the shaft using adhesive. The TalWire includes an articulatable tip at the distal end of the guidewire. The finished guidewire is placed in a protective polymer dispenser hoop in a labeled Tyvek pouch. The TalWire is provided as a sterile, single-use device, and is sterilized using a validated ethylene oxide process.
The provided text describes a 510(k) premarket notification for a medical device called TalWire, a catheter guidewire. It compares the TalWire to a predicate device and includes performance data to support substantial equivalence.
Here's an analysis of the acceptance criteria and study information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The document states that "All tests passed and met the predefined acceptance criteria." However, it does not explicitly list the quantitative acceptance criteria for each performance test. It only lists the types of performance tests conducted.
Test | Acceptance Criteria (Not Explicitly Stated, Assumed to be "Pass") | Reported Device Performance |
---|---|---|
Dimensional Attributes | (Assumed: Meets design specifications) | Passed |
Visual Inspection | (Assumed: No defects) | Passed |
Simulated Use | (Assumed: Performs as intended in simulated environment) | Passed |
Tensile Strength | (Assumed: Meets strength requirements) | Passed |
Torque Strength | (Assumed: Meets torque requirements) | Passed |
Torqueability | (Assumed: Adequate torque response) | Passed |
Corrosion Resistance | (Assumed: No significant corrosion) | Passed |
Kink Resistance | (Assumed: Resists kinking) | Passed |
Tip Flexibility | (Assumed: Adequate flexibility) | Passed |
Radiopacity | (Assumed: Visible under fluoroscopy) | Passed |
Flex Resistance | (Assumed: Resists fatigue/damage from flexing) | Passed |
Fracture Resistance | (Assumed: Resists fracture) | Passed |
Prolapse Force | (Assumed: Meets prolapse force requirements) | Passed |
Biocompatibility:
- Evaluation in compliance with ISO 10993-1.
- Specific tests conducted: Cytotoxicity, Guinea Pig Maximization Sensitization, Intracutaneous Irritation, Material Mediated Pyrogenicity, Acute Systemic Toxicity, ASTM Hemolysis, SC5b-9 Complement Activation Assay, Standard Thrombogenicity in Canine.
- Reported device performance: "The device was found biocompatible."
Sterilization, Packaging and Shelf Life Testing:
- EtO Sterilization validation (SAL 10^-6).
- Shelf life and packaging testing after environmental conditioning and transportation simulation.
- Reported device performance: "All tests were successfully completed."
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample sizes for any of the performance tests.
The data provenance is not explicitly stated beyond being "performance data." This type of testing is typically done in a lab setting rather than using patient data, so concepts like "country of origin" or "retrospective/prospective" studies are not applicable in the usual sense for this kind of device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This information is not applicable as the tests described are performance tests of the physical properties and functionality of the guidewire, not assessments of diagnostic images or clinical outcomes that would require expert human interpretation as "ground truth." These tests rely on standardized methods and measurement equipment.
4. Adjudication Method for the Test Set
This information is not applicable for the same reasons as point 3. Performance tests are typically objectively measured against predefined criteria, not adjudicated by experts.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating the substantial equivalence of the device through performance testing against established standards and comparison to a predicate device, not on assessing human reader performance with or without AI assistance. This device is a physical medical device (guidewire), not an AI-driven diagnostic tool.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
No, a standalone algorithm performance study was not done. This device is a physical guidewire, not an algorithm.
7. The Type of Ground Truth Used
For the performance and biocompatibility testing, the "ground truth" is established by:
- Standardized test methods: Based on FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling Guidance for Industry and Food and Drug Administration Staff" and ISO 10993-1 for biocompatibility.
- Predefined acceptance criteria: While not explicitly listed, these criteria would typically be quantitative thresholds or qualitative observations (e.g., "no visible corrosion," "no fracture").
8. The Sample Size for the Training Set
This information is not applicable. This is a physical medical device, not an AI model that requires a "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reasons as point 8.
Ask a specific question about this device
Page 1 of 1