K Number
K222246

Validate with FDA (Live)

Date Cleared
2022-11-21

(118 days)

Product Code
Regulation Number
878.4850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Disposable Sterile Lancet: The lancet is intended for capillary blood sampling.

Lancing Device:

The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is single patient use only and should not be shared with anyone else, even a family member.

Device Description

Disposable Sterile Lancet is Class II Single Use Only Blood Lancet Without An Integral Sharps Injury Prevention Feature bearing the product code ORK (21CFR 878.4850), and The Disposable Sterile Lancet and lancing device is Class II Multiple Use Blood Lancet For Single Patient Use Only bearing the product code QRL (21CFR878.4850).

The lancet is for single use, disposable sterile devices, which is designed to collect capillary blood samples. And the lancing device is intended for capillary blood sampling with Disposable Sterile Lancet. The lancet can be used independently, or used together with lancing device.

The intended users include healthcare personnel, patients and lay users.

The lancet is provided sterile. The sterile barrier is the needle sleeve and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The shelf life of the lancet is 5 years.

AI/ML Overview

This document describes a 510(k) submission for a Disposable Sterile Lancet and Lancing Device. The submission aims to demonstrate substantial equivalence to a predicate device (K220475).

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

ItemsAcceptance CriteriaReported Device Performance
Appearance
Product AppearanceNo obvious foreign body, fracture, unformed defects etc.Meet the requirements
Needle-tipNo obvious burrs, curved hooks, oil stains etc. Lancet should be well assembled, no obvious dislocation at the fitting of the shellMeet the requirements
Dimensions
Diameter of needle coreProduct dimensions shall be consistent to the drawingsMeet the requirements
Length of exposed needle-tipThe length of exposed needle-tip shall be consistent to the drawings.Meet the requirements
Performance
Puncture Force17-20G≤4N 21-25G<1.5N 26-38G≤1NMeet the requirements
FirmnessNeedle should connect firmly with plastic handleMeet the requirements
SterilityThe sterility for the device shall be at SAL of 10-6Meet the requirements
Biocompatibility
CytotoxicityUnder the conditions of the study, the subject device extract was determined to be non-cytotoxic.Pass
IrritationUnder the conditions of the study, the subject device non-polar and polar extracts were determined to be non-irritating.Pass
SensitizationUnder the conditions of the study, the subject device non-polar and polar extracts were determined to be non-sensitizing.Pass
Acute Systemic ToxicityUnder the conditions of the study, the subject device is no acute systemic toxicityPass
PyrogenUnder the conditions of the study, the subject device is no pyrogenicity reactionPass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample size used for the non-clinical performance and biocompatibility tests. It also does not mention the country of origin of the data or whether the studies were retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document. The studies conducted are non-clinical (performance and biocompatibility), which do not typically involve human expert ground truth establishment in the same way clinical studies or AI diagnostic device studies would.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and not provided. The studies are non-clinical laboratory tests, not clinical evaluations requiring adjudication of expert opinions.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a physical medical device (lancet and lancing device), not an AI-powered diagnostic tool. Therefore, the concept of human readers improving with AI assistance is not relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical medical device. There is no algorithm or AI component involved.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the non-clinical tests was established by objective measurements and standardized testing procedures against defined acceptance criteria (e.g., measuring puncture force, examining appearance for defects, confirming sterility via standard methods, and conducting biocompatibility tests according to ISO standards). It's essentially objective laboratory measurements and adherence to international standards.

8. The sample size for the training set

This is not applicable as there is no mention of a training set. This is a traditional medical device submission, not an AI/machine learning device that requires training data.

9. How the ground truth for the training set was established

This is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

11/21/2022

Tianjin Rilifine Medical Device Co., Ltd. Mr. Qiusheng Jiang Manager No 32, Jingguan Road, Yixingbu, Beichen District Tianjin, 300402 China

Re: K222246

Trade/Device Name: Disposable Sterile Lancet Regulation Number: 21 CFR 878.4850 Regulation Name: Blood lancets Regulatory Class: Class II Product Code: QRL, QRK Dated: September 26, 2022 Received: September 27, 2022

Dear Mr. Jiang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Carr -S

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K22246

Device Name

Disposable Sterile Lancet; lancing device

Indications for Use (Describe) Disposable Sterile Lancet: The lancet is intended for capillary blood sampling.

Lancing Device:

The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is single patient use only and should not be shared with anyone else, even a family member.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

(In accordance with 21 CFR 807.92)

1.0 Submitter's Information

Name:Tianjin Rilifine Medical Device Co., Ltd.
Address:No 32, Jingguan Road, Yixingbu, Beichen District
Tianjin, 300402, CHINA
Phone Number:86-22-23233999
Primary Submitter:Mr. Qiusheng Jiang
TitleCo-founder and Manager
Email:henry@rilifine.com
Secondary Submitter:Ms. Belinda Wang
Title:Regulatory manager
Email:belindaw@126.com
Date of Preparation:Nov 17, 2022

2.0 Device Information

Device Name:Disposable Sterile Lancet; Lancing device
Common Name:Rilifine Blood Lancet, lancing device
Classification Name:Single Use Only Blood Lancet Without An Integral SharpsInjury Prevention FeatureMultiple Use Blood Lancet For Single Patient Use Only

3.0 Classification

Product Code:QRK, QRL
Regulation Number:21 CFR 878.4850
Classification:II
Review Panel:General & Plastic Surgery

4.0 Predicate Device Information

Manufacturer:Tianjin Huahong Technology Co., Ltd.
Device:Lancet, lancing device
510(k) Number:K220475
ClassificationII
Product CodeQRL, QRK

5.0 Intended Use

{4}------------------------------------------------

Disposable Sterile Lancet: The lancet is intended for capillary blood sampling -.

Lancing Device:

The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is single patient use only and should not be shared with anyone else, even a family member.

6.0 Device Description

Disposable Sterile Lancet is Class II Single Use Only Blood Lancet Without An Integral Sharps Injury Prevention Feature bearing the product code ORK (21CFR 878.4850), and The Disposable Sterile Lancet and lancing device is Class II Multiple Use Blood Lancet For Single Patient Use Only bearing the product code QRL (21CFR878.4850).

The lancet is for single use, disposable sterile devices, which is designed to collect capillary blood samples. And the lancing device is intended for capillary blood sampling with Disposable Sterile Lancet. The lancet can be used independently, or used together with lancing device.

The intended users include healthcare personnel, patients and lay users.

The lancet is provided sterile. The sterile barrier is the needle sleeve and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The shelf life of the lancet is 5 years.

7.0 Comparison of Technological Characteristics with Predicate Device

following table is the summary of the technological characteristics, The biocompatibility of the proposed subject device and predicate device.

ItemSubject DevicePredicate Device(K220475)ComparisonResult
ManufacturerTianjin Rilifine MedicalDevice Co., Ltd.Tianjin HuahongTechnology Co., Ltd.--
510K number--K220475--
Product nameDisposable Sterile LancetLancing deviceLancet,Lancing Device--
ClassificationIIIISame
Product CodeQRK, QRLQRK, QRLSame
Regulation Number21 CFR 878.485021 CFR 878.4850Same
ItemSubject DevicePredicate Device(K220475)ComparisonResult
Intended UseDisposable Sterile Lancet:The lancet is intended forcapillary blood sampling.Lancing Device:The Lancing Device isused with lancets to drawcapillary blood from thefingertip, for testingutilizing small amounts ofblood. The LancingDevice is single patientuse only and should not beshared with anyone else,even a family member.Lancet:Lancet is intended forcapillary bloodsampling.Lancing device:The lancing device isused with lancet to drawcapillary blood fromfingertip, for testingutilizing small amountsof blood. The lancingdevice is intended to beused by a single patientand should not beshared.Same
Single UseThe disposable sterilelancet is single Use, andthe lancing device isreusable, but single patientuse onlyThe disposable sterilelancet is single Use, and thelancing device is reusable,single patient use onlySame
technicalspecifica-tionsNeedlegauge21G (0.82±0.01mm)23G (0.64±0.01mm)26G (0.46±0.01mm)28G (0.36±0.01mm)30G (0.31±0.01mm)31G (0.26±0.01mm)32G (0.24±0.01mm)33G (0.21±0.01mm)16G (1.50±0.02mm)18G (0.20±0.01mm)19G (1.07±0.01mm)20G (0.91±0.01mm)21G (0.82±0.01mm)22G (0.72±0.01mm)23G (0.64±0.01mm)24G (0.57±0.01mm)25G (0.51±0.01mm)26G (0.46±0.01mm)27G (0.41±0.01mm)28G (0.36±0.01mm)29G (0.34±0.01mm)30G (0.31±0.01mm)31G (0.26±0.01mm)32G (0.24±0.01mm)33G (0.21±0.01mm)34G (0.19±0.01mm)Similar, Seenote 1
ExposedneedleNormally the length ofexposed needle tip is3.2±0.3mm (Model: IA, IB,IC, ID, IE, IK, IL, IM, IIA,
ItemSubject DevicePredicate Device(K220475)ComparisonResult
Which can also vary percustomer requirementsfrom 0.7mm to 3.5mm(±0.3mm).
PunctureForce17-20G≤4N21-25G≤1.5N26-38G<1NThe needle-tip of the needleshould have good punctureability.
Principles ofoperationPressure activated ormanualPressure activated ormanualSame
Sterilization methodand SALThe disposable sterilelancet is sterilized byradiationSAL=10-6Sterilized by radiationSAL=10-6Same
Shelf lifeThe disposable sterilelancet is 5 years5 yearsSame
Materials of parts incontact with humanbodyNeedle: stainless steelOther parts: plasticmaterials, such as PE etc.The lancet has a needle thatis made of stainless steeland silicone oil. The bodyand cap are made ofpolyethylene (PE) andEthylene Vinyl Acetate(EVA) and calcium powderSimilar, Seenote 2
BiocompatibilityConforms to therequirements of ISO10993 series standards.Conforms to therequirements of ISO 10993series standards.Same

{5}------------------------------------------------

{6}------------------------------------------------

Note 1: The needle gauge, and Puncture Force of the subject device is not complete same with predicate device, but the needle gauge for the subject device is within the scope of the subject device and have the same specification with the predicate device. And performance testing showed that the subject devices comply with the specification and is similar to that of the predicate device, and no concerns caused related to the device safety and effectiveness, and the product can be guaranteed to be safe and effective.

Note 2: The component and material of proposed devices is different from the predicate devices. However, all the materials are known biocompatible materials that have been used in lancets or other similar medical devices, and the biocompatibility testing showed that the subject device is biocompatible.

{7}------------------------------------------------

There are no other significant differences between the two products in terms of design and technological characteristics.

8.0 Non-Clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

  • . ISO 11137-1 First edition 2006-04-15, Sterilization of health care products -Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2018)]
  • ISO 11137-2 Third edition 2013-06-01, Sterilization of health care products -. Radiation - Part 2: Establishing the sterilization dose
  • . ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
  • ISO 10993-5:2009 Biological Evaluation of Medical Devices Part 5: Tests . For In Vitro Cytotoxicity
  • ISO 10993-23 First edition 2021-01,Biological evaluation of medical devices -Part 23: Tests for irritation

Performance testing verified that the proposed device is as effective and performs as well as the legally marketed predicate device in terms of critical performance characteristics as follows:

ItemsAcceptance CriteriaResults
AppearanceProduct AppearanceNo obvious foreign body, fracture, unformed defects etc.Meet the requirements
Needle-tipNo obvious burrs, curved hooks, oil stains etc. Lancet should be well assembled, no obvious dislocation at the fitting of the shellMeet the requirements
DimensionsDiameter of needle coreProduct dimensions shall be consistent to the drawingsMeet the requirements
Length of exposed needle-tipThe length of exposed needle-tip shall be consistent to the drawings.Meet the requirements
PerformancePuncture Force$17-20G≤4N$$21-25G<1.5N$Meet the requirements

{8}------------------------------------------------

26-38G≤1N
FirmnessNeedle should connect firmly with plastic handleMeet the requirements
SterilityThe sterility for the device shall be at SAL of 10-6Meet the requirements

Biocompatibility testing as per ISO 10993 standards:

ItemSubject deviceResult
CytotoxicityUnder the conditions of the study, the subjectdevice extract was determined to benon-cytotoxic.Pass
IrritationUnder the conditions of the study, the subjectdevice non-polar and polar extracts weredetermined to be non-irritating.Pass
SensitizationUnder the conditions of the study, the subjectdevice non-polar and polar extracts weredetermined to be non-sensitizing.Pass
Acute SystemicToxicityUnder the conditions of the study, the subjectdevice is no acute systemic toxicityPass
PyrogenUnder the conditions of the study, the subjectdevice is no pytogenicity reactionPass

9.0 Clinical Test

No clinical study is included in this submission.

10. Comparison to the Predicate Device and Conclusion

The conclusion drawn from the nonclinical tests demonstrate that the subject device Disposable Sterile Lancet has the same indication for use and has similar design features and technological characteristic as the predicate device, and the proposed device is as safe, as effective and performs as well as the legally marketed predicate device K220475.

§ 878.4850 Blood lancets.

(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.