QRL · Multiple Use Blood Lancet For Single Patient Use Only

General, Plastic Surgery · 21 CFR 878.4850 · Class 2

Overview

Product CodeQRL
Device NameMultiple Use Blood Lancet For Single Patient Use Only
Regulation21 CFR 878.4850
Device ClassClass 2
Review PanelGeneral, Plastic Surgery

Identification

The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.

Classification Rationale

Class II (special controls) for paragraphs (a), (b), and (c). Class III (premarket approval) for paragraph (d).

Special Controls

*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (b) *Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (c) *Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that: (A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and (B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected. (vi) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use. (C) Handwashing instructions for the user before and after use of the device. (D) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device. (F) Instructions for the safe disposal of the device. (G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades. (H) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.” (B) “Used lancet blades must be safely discarded after a single use.” (C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.” (d) *Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class III (premarket approval).(3) *Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Recent Cleared Devices (20 of 24)

Showing 20 most recent of 24 cleared devices.

RecordDevice NameApplicantDecision DateDecision
K253706Lancing device (HH-XV-T)Tianjin Huahong Technology Co., Ltd.Dec 9, 2025SESE
K250813MICROLET®NEXT 2 Lancing DeviceAscensia Diabetes Care U.S., Inc.May 14, 2025SESE
K243306Lancet (IA, IB, IC, ID, IE, IK, IL, IM, IIA, IIB, III, V, VI, VII, VIII, IX); Lancing device (HH-X-T, HH-XV-T, HH-XVI-T, HH-XVII-T, HH-XVIII-T, HH-XIX, HH-XXI-T, HH-XXII-T, HH-XXIII-T, HH-XXIV-T, HH-XXV-T, HH-XXVI-T, HH-XXVIII-T, HH-XXIX-T, HH-XXX-T, HH-XIII-T, HH-XXVII-T)Tianjin Huahong Technology Co., Ltd.Nov 15, 2024SESE
K232912Facet Aurora Reusable Lancet BaseFacet Technologies, LLCJan 25, 2024SESE
K231797Medifun Lancing Device, Model No. LD-E1Medifun CorporationNov 22, 2023SESE
K223643Verifine® Ease Lancing Device, Verifine® Lancing DevicePromisemed Hangzhou Meditech Co., Ltd.Feb 21, 2023SESE
K221419TD-5010 Lancing Device and TD-5084 Sterile LancetsGostar Co., Ltd.Jan 20, 2023SESE
K222472Lancing System, Sterile Lancet for Single Use, Lancing DeviceShandong Lianfa Medical Plastic Products Co. , Ltd.Nov 30, 2022SESE
K223099Facet Manatee Reusable Lancing BaseFacet Technologies, LLCNov 28, 2022SESE
K222656LDE4 Lancing DeviceI-Sens, Inc.Nov 23, 2022SESE
K222246Disposable Sterile LancetTianjin Rilifine Medical Device Co., Ltd.Nov 21, 2022SESE
K222539Facet Blood LancetsFacet Technologies, LLCNov 18, 2022SESE
K221072TRUEdraw Lancing Device, Mini Lancing DeviceTrividia HealthOct 18, 2022SESE
K221175Multi-Lancet Device 2, ReliOn Premier Lancing DeviceArkray, Inc.Oct 17, 2022SESE
K221613Freestyle Lancing Device II, Autolet, Autolet Lite, Unilet LancetsOwen Mumford, Ltd.Oct 3, 2022SESE
K221062RIGHTEST Lancing Device GD500, GE Lancing Device, iGlucose Lancing DeviceBionime CorporationSep 26, 2022SESE
K222034RightLance Blood Lancing SystemBeijing Ruicheng Medical Supplies Co., Ltd.Sep 7, 2022SESE
K221970Lancing deviceSteriLance Medical (Suzhou), Inc.Aug 30, 2022SESE
K221546OneTouch Delica Plus Lancing SystemAsahi Polyslider Co., Ltd.Aug 23, 2022SESE
K220387Disposable Blood LancetsSuzhou Kyuan Medical Apparatus Co., Ltd.Aug 17, 2022SESE

Top Applicants

Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...