K Number
K221507

Validate with FDA (Live)

Date Cleared
2022-08-31

(100 days)

Product Code
Regulation Number
878.4850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Disposable Blood Lancet is used for capillary blood collection.

Device Description

Disposable Blood Lancet is a single use, sterile medical device designed to be used for capillary blood collection. The device comprises a stainless needle encapsulated with a plastic needle body and protective cap, the protective cap is twisted off to expose the needle for use.

The device was sterilized by Radiation. The needle body and protective cap form a sterile barrier to maintain the needle sterile.

AI/ML Overview

The provided text describes the regulatory clearance of a medical device, the "Disposable Blood Lancet," and details its non-clinical testing for substantial equivalence to a predicate device. It does not involve AI or algorithms, and therefore, the requested information regarding AI-specific criteria, such as MRMC studies, training/test set details, and expert adjudication, is not applicable.

Here's an analysis based on the information provided regarding the non-AI device:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are implicitly derived from the non-clinical tests performed to demonstrate substantial equivalence to the predicate device. The "reported device performance" is the result of these tests.

Acceptance Criteria (Implicit from Tests)Reported Device Performance (Test Result)
Appearance and CleanlinessConforms to specifications
Needle tip appearanceConforms to specifications
Needle dimensionConforms to specifications
Exposed length of productConforms to specifications
Cap siteConforms to specifications
Needle tip sharpnessConforms to specifications
Binding StrengthConforms to specifications
Double needlesNone detected
Empty needleNone detected
Reverse needleNone detected
Compatibility test with Lancing DeviceCompatible
In Vitro Cytotoxicity (ISO 10993-5: 2009)No Cytotoxicity
Skin Sensitization (ISO 10993-10: 2010)No Skin sensitization
Intracutaneous reactivity (ISO 10993-10: 2010)No irritation
Acute Systemic Toxicity (ISO 10993-11: 2017)No Acute Systemic Toxicity
Material-Mediated Pyrogenicity (ISO 10993-11: 2017)Absence of pyrogen
Shelf life (5 years)Performance and sterile barrier integrity remain stable (verified per ASTM F1980)

2. Sample sized used for the test set and the data provenance

The document does not specify the exact sample sizes used for each non-clinical performance test. It only states that "bench testing performed verifies that the performance of the proposed device is substantially equivalent." For biocompatibility, it states "biocompatibility evaluations were conducted." The provenance of the data is from non-clinical lab testing conducted by the manufacturer, SteriLance Medical (Suzhou) Inc., likely in China. The nature of these tests is retrospective in the context of the 510(k) submission, as they were completed prior to the review.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable as the device is a physical medical device, not an AI/algorithm-driven one. Ground truth is established through standardized laboratory testing and measurement procedures defined by the relevant ISO and ASTM standards. Experts in laboratory testing and quality assurance would perform and evaluate these tests, but their qualifications are not detailed beyond their adherence to the standards.

4. Adjudication method for the test set

This question is not applicable for a non-AI physical device. The testing described involves objective measurements and standardized biological evaluations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable as the device is a physical "Disposable Blood Lancet" and does not involve AI assistance or human readers in an interpretive capacity.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable as the device is a physical "Disposable Blood Lancet" and does not contain an algorithm.

7. The type of ground truth used

For the performance tests (e.g., appearance, dimensions, sharpness, binding strength, etc.), the "ground truth" is established by engineering specifications, design requirements, and applicable international standards (e.g., ISO 9626 for needle size). For biocompatibility, the ground truth is established by standardized biological evaluation methods and criteria outlined in ISO 10993 series.

8. The sample size for the training set

This question is not applicable as the device is a physical "Disposable Blood Lancet" and does not involve an AI training set.

9. How the ground truth for the training set was established

This question is not applicable as the device is a physical "Disposable Blood Lancet" and does not involve an AI training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

August 31, 2022

SteriLance Medical (Suzhou) Inc. Ms. Susan Sun Manager No.168 PuTuoShan Road, New District Suzhou, Jiangsu 215153 China

Re: K221507

Trade/Device Name: Disposable Blood Lancet Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: QRK,QRL Dated: July 28, 2022 Received: August 4, 2022

Dear Ms. Sun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221507

Device Name Disposable Blood Lancet

Indications for Use (Describe) Disposable Blood Lancet is used for capillary blood collection.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date of Summary prepare: August 31, 2022

1. Submission Sponsor

Applicant NameSteriLance Medical (Suzhou) Inc.
AddressNo.168 PuTuoShan Road, New District, 215153 Suzhou,Jiangsu, P.R.China
Contact personSusan Sun
Phone+86-512-65799308

2. Submission correspondent

NameSteriLance Medical (Suzhou) Inc.
AddressNo.168 PuTuoShan Road, New District, 215153Suzhou, Jiangsu, P.R.China
Post Code215153
Phone No.+86-512-65799308
Contact PersonSusan Sun
Emailregistration2@sterilance.com

3. Device Identification

Type of 510(k) submission:Traditional
Trade Name:Disposable Blood Lancet
Model:Soft
Classification name:Lancet, Blood
Review Panel:General & Plastic Surgery
Product Code:QRK, QKL
Device Class:II
Regulation Number:21 CFR 878.4850

4. Legally Marketed Predicate Device

Trade NamePromisemed Blood Lancet
Regulation number21 CFR 878.4800
Regulation classI
Regulation nameManual Surgical Instrument for General Use
510(k) NumberK192666
Product CodeFMK
ManufacturerPromisemed Hangzhou Meditech Co., Ltd.

{4}------------------------------------------------

5. Device Description

Disposable Blood Lancet is a single use, sterile medical device designed to be used for capillary blood collection. The device comprises a stainless needle encapsulated with a plastic needle body and protective cap, the protective cap is twisted off to expose the needle for use.

The device was sterilized by Radiation. The needle body and protective cap form a sterile barrier to maintain the needle sterile.

6. Intended Use/ Indications for Use

Disposable Blood Lancet is used for capillary blood collection.

ComparisonitemSubject Device:Disposable Blood Lancet(K221507)Predicate Device:Promisemed Blood Lancet(K192666)Comments
Product CodeQRK, QKLFMKDifferent 1
RegulationNumber21 CFR § 878.485021 CFR § 878.4800Different1
ClassificationClass IIClass IDifferent 1
Type of useOTCOTCSame
Intendeduse &Indications for UseDisposable Blood Lancet isused for capillary bloodcollection.It is intended for capillaryblood sampling.Same
ApplicableuserAdult and pediatricAdult and pediatricSame
ReusedurabilitySingle useSingle useSame
Sterilizationmethod andSALSterilized by RadiationSAL=10-6Sterilized by RadiationSAL=10-6Same
Shelf life5 Years3 YearsDifferent 2
ComponentNeedle, Needle body,andProtective capNeedle, Needle body, andCapSame

7. Technological characteristics comparison

{5}------------------------------------------------

ComparisonitemSubject Device:Disposable Blood Lancet(K221507)Predicate Device:Promisemed Blood Lancet(K192666)Comments
Specification(needlediameter)21G, 23G, 26G, 28G, 30G,32G, 33GContain a variety of gauge(unknown the detail)Similar
MaterialsNeedle: stainless steelNeedle body and cap:PolyethyleneNeedle: stainless steelNeedle body and cap:PolyethyleneSame

Different 1 : On November 22, 2021, FDA reclassified blood lancet, therefore, the product code, regulation number and classification are different.

Different 2 : The shelf life of the subject device is different from the predicate device. The shelf life

of the subject device has been verified in accordance with ASTM F1980, demonstrating that the performance of the device and the integrity of the sterile barrier remain stable over the claimed 5year shelf life. Therefore, this difference does not affect substantially equivalence on safety and effectiveness.

Similar: In ISO 9626, the needle size is designated by nominal diameter, corresponding gauge size. Gauge size ranges from 10G-34G. The difference in gauge size does not affect intended use. Therefore, this difference does not affect substantially equivalence on safety and effectiveness.

8. Summary of non-clinical testing

*Performance Testing

The bench testing performed verifies that the performance of the proposed device is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests are as follow:

  • Appearance and Cleanliness and Needle tip appearance
  • . Needle dimension, Exposed length of product
  • Cap site, Needle tip sharpness and Binding Strength ●
  • Double needles. Empty needle and Reverse needle
  • Compatibility test between Disposable blood lancet and Lancing Device ●

*Biocompatibility

The biocompatibility evaluations were conducted in accordance with the 2020 FDA Guidance document Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk

{6}------------------------------------------------

ItemTest methodTest result
In Vitro CytotoxicityISO 10993-5: 2009No Cytotoxicity
Skin SensitizationISO 10993-10: 2010No Skin sensitization
Intracutaneous reactivityISO 10993-10: 2010No irritation
Acute Systemic ToxicityISO 10993-11: 2017No Acute Systemic Toxicity
Material-MediatedPyrogenicityISO 10993-11: 2017Absence of pyrogen

Management Process". The tests include the following tests:

9. Brief discussion of clinical tests

No clinical tests were performed.

10. Conclusions

The conclusion drawn from the nonclinical tests demonstrates that the subject device, the Disposable Blood Lancets are as safe and effective, and performs as well as or better than the legally marketed predicate device cleared under K192666.

§ 878.4850 Blood lancets.

(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.