K Number
K223314

Validate with FDA (Live)

Date Cleared
2023-01-25

(89 days)

Product Code
Regulation Number
878.4850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The product is used to obtain capillary blood samples from fingertip in a home. The lancet is assembled with lancing device, once the lancing device is launched, the needle of lancet can prick the skin.

Device Description

The Disposable Blood Lancets consists of a needle, a body and a cap. The models of the Disposable Blood Lancets are 21G; 23G; 26G; 30G. Disposable Blood Lancets are used to obtain blood samples for testing purposes. It is sterilized by Irradiation and is a single-use product. The device is compatible with the reusable lancing device which met the specification and size requirements of the following figure in the market.

AI/ML Overview

The document provided is a 510(k) summary for a medical device called "Disposable Blood Lancets". It outlines the device's technical characteristics and compares them to a predicate device to establish substantial equivalence.

Based on the provided text, the device is a medical lancet and not an AI/ML powered device. Therefore, a study proving the device meets acceptance criteria related to AI/ML performance (such as MRMC studies, standalone algorithm performance, ground truth establishment methods for AI, or training set sizes) would not be applicable and is not present in this document.

However, I can extract information regarding the acceptance criteria and the studies performed to demonstrate the device's physical performance, biocompatibility, and sterilization for this specific medical device.

1. Table of Acceptance Criteria and Reported Device Performance

NoTesting ItemSpecification (Acceptance Criteria)Result
01AppearanceLancet should have the same color, no bubble, no flash, no slip.Pass
02Launch LengthThe length of the needles in the Disposable Blood Lancets is different in different gauge. The general depth is 3.2mmPass
03Sharpness/Penetration testingPenetration force ≤1.00N.Pass
04Exterior (Needle connection)The connection between needle and needle body should be firm.Pass
05Exterior (Cap twist)Cap twist should be smooth.Pass
06Initial bioburdenInitial bioburden of the device shall be less than 100CFU/g.Pass
07SterileThe sterile blood lancet shall be sterile.Pass
08Cap removal forceThe moment for breaking the safe mode should range from 30 Ncm to 35 Ncm.Pass
09Needle removal forceThe bond between the lancet body and needle should be greater than or equal to 10N/15s.Pass
10Drop testingThe carton box should have no puncture after the drop test.Pass

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes used for each individual performance test (e.g., how many lancets were tested for sharpness, bioburden, etc.). It generally states that "The following performance data were provided in support of the substantial equivalence determination."

The data provenance is not specified in terms of country of origin of the data or whether it was retrospective or prospective. It is implied to be internal testing conducted by the manufacturer, Huaian Hening Medical Instruments Co., Ltd.

3. Number of Experts Used to Establish Ground Truth and Their Qualifications

Not applicable for this type of medical device's performance testing. The "ground truth" for these tests is based on objective, measurable physical and biological parameters defined by established standards and internal specifications, not expert consensus on interpretations.

4. Adjudication Method

Not applicable. The tests are objective measurements against predefined specifications.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is not an AI/ML powered device, so MRMC studies are not relevant.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

Not applicable. This is a physical medical device, not an algorithm.

7. Type of Ground Truth Used

The "ground truth" for the performance tests outlined in the table consists of objective physical and biological measurements against predefined specifications and established international standards (e.g., ISO for biocompatibility and sterilization). This is not expert consensus, pathology, or outcomes data in the context of an AI/ML diagnostic device.

8. Sample Size for the Training Set

Not applicable. This is a physical device, not an AI/ML model.

9. How the Ground Truth for the Training Set Was Established

Not applicable. This is a physical device, not an AI/ML model.

Additional Information on Studies Conducted:

The document also details other non-clinical tests performed:

  • Biocompatibility Testing: Evaluated in accordance with ISO 10993-1:2018 for the body contact category. Specific tests performed include:
    • Cytotoxic test (ISO 10993-5:2009)
    • Skin sensitization test (ISO 10993-23:2021)
    • Intracutaneous test (ISO 10993-10:2021)
    • Acute systemic toxicity test (ISO 10993-11:2017)
    • Hemolysis test (ISO 10993-4:2017)
    • Pyrogen Test (ISO 10993-11:2017)
  • Sterilization and Shelf Life Testing:
    • Irradiation sterilization validation per ISO 11137-1 and ISO 11137-3.
    • Pyrogen testing per ISO 10993-11:2017.
    • Transportation test per ISTA 2A:2011.
    • Shelf life of 5 years determined based on stability study which includes aging test.

These studies support the substantial equivalence claim by demonstrating that the device meets safety and effectiveness requirements through validated physical, chemical, and biological testing, rather than complex AI/ML performance evaluations.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 25, 2023

Huaian Hening Medical Instruments Co., Ltd Zhengcan Da General Manager No. 6 West Hongdou Road Economic Development Zone Huaian, Jiangsu China

Re: K223314

Trade/Device Name: Disposable Blood Lancets Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: QRK, QRL Dated: December 29, 2022 Received: December 29, 2022

Dear Zhengcan Da:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Carr -S

for Long Chen, PhD Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223314

Device Name Disposable Blood Lancets

Indications for Use (Describe)

Disposable Blood Lancets

The product is used to obtain capillary blood samples from fingertip in a home. The lancet is assembled with lancing device, once the lancing device is launched, the needle of lancet can prick the skin.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K223314 510(k) summary

l Submitter

Device submitter: HUAIAN HENING MEDICAL INSTRUMENTS CO., LTD.

NO.6 West Hongdou Road Economic Development Zone Huaian,

Jiangsu China

Contact person: Zhengcan Da General Manager Phone: +86-13952318668 Fax: +86-517-83800501 Email: hn-lancets@vip.163.com

Date: 10/12/2022

II Device

Trade Name of Device: Disposable Blood Lancets Common Name: Blood Lancets Regulation Number: 21 CFR 878.4850 Regulation Name: Single Use Only Blood Lancet without an Integral Sharps Injury Prevention Feature Regulatory Class: II Product code: QRK; QRL Review Panel: General & Plastic Surgery

III Predicate Devices

Trade name:Promisemed Blood Lancet
VeriFine Safety Lancet
VeriFine Mini-Safety Lancet
Common name:Blood Lancet
Classification:Class I, Lancet, Blood, 21CFR 878.4800
Product Code:FMK
Premarket Notification:K192666
Manufacturer:Promisemed Hangzhou Meditech Co., Ltd.

IV Device description

The Disposable Blood Lancets consists of a needle, a body and a cap. The models of the Disposable Blood Lancets are 21G; 23G; 26G; 30G. Disposable Blood Lancets

{4}------------------------------------------------

are used to obtain blood samples for testing purposes. It is sterilized by Irradiation and is a single-use product. The device is compatible with the reusable lancing device which met the specification and size requirements of the following figure in the market.

Image /page/4/Picture/1 description: The image shows three diagrams of a cylindrical object with measurements. The first diagram shows the object with a ring on one end and a measurement of 32.5mm ± 0.5. The second diagram shows the object with a different end and a measurement of 22.7mm ± 0.5. The third diagram shows the object with a measurement of 3.2mm ± 0.6.

V Indications for use

The product is used to obtain capillary blood samples from fingertip in a hospital or at home. The lancet is assembled with lancing device, once the lancing device is launched, the needle of lancet can prick the skin.

VI Comparison of technological characteristics with the predicate devices

The Disposable Blood Lancets have the same intended use, technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the Disposable Blood Lancets and predicate devices do not alter suitability of the proposed device for its intended use.

Device featureSubject Device(Disposable BloodLancets)Predicate DeviceK192666(PromisemedBloodLancet)Comment
Indications for useThe product is used toobtain capillary bloodsamples from fingertip in ahospital or at home. Thelancet is assembled withlancing device, once thelancing device is launched,the needle of lancet canprick the skin.It is intended forcapillary bloodsampling.SimilarComment 1
Product codeQRK, QRLFMKDifferentComment 1
Safety protectionfeaturesNoYesDifferent
Device featureSubject Device(Disposable BloodLancets)Predicate DeviceK192666(PromisemedBloodLancet)Comment
Reuse durabilitySingle useSingle useEquivalent
SterilizationIrradiationNot availableDifferentComment 2
Model21G; 23G; 26G; 28G; 30GBL-30 (30G)BL-28 (28G)(Informationgathered fromPromisemedHangzhouMeditech Co.,Ltd. officialwebsite)DifferentComment 3
Launch length3.2mm±0.63mm(Informationgathered fromPromisemedHangzhouMeditech Co.,Ltd. officialwebsite)
Materials of partsin contact withhuman bodyLancet needle: stainlesssteel;Body and cap: PELancet needle:stainless steel;Body and cap: PEEquivalent
Principle ofOperationThe Disposable BloodLancets comprises astainless steel needleencapsulated with a plasticbody and cap, the cap istwisted off to expose theneedle for use.The PromisemedBlood Lancetcomprises astainless steelneedleencapsulated witha plastic body andcap, the cap isEquivalent
Device featureSubject Device(Disposable BloodLancets)Predicate DeviceK192666(PromisemedBloodLancet)Comment
twisted off toexpose theneedle for use.
ManufacturingStainless steel needle isfed into an injectionmolding machine to over-mold plastic material(polyethylene) forming abody and cap,encapsulating the stainlesssteel needles. Terminalsterilization process isperformed to ensuresterility of an entireproduct.For thePromisemedBlood Lancet,stainless steelneedle is fed intoan injectionmolding machineto over-moldplastic material(polyethylene)forming a bodyand cap,encapsulating thestainless steelneedles. Terminalsterilizationprocess isperformed toensure sterility ofan entire product.Equivalent

{5}------------------------------------------------

{6}------------------------------------------------

Discussion:

Comment 1

The subject device and the predicate device have the same intended use, to puncture the skin to obtain drops of blood for testing purposes. While the Disposable Blood Lancets have no sharp's prevention features and without the intended use of protecting the user from a needlestick injury. This difference does not affect the clinical safety of the subject device.

Comment 2

{7}------------------------------------------------

The sterilization method of predicate device is not available. However, the subject device was ensured sterility by sterilization validation. Therefore, the differences on sterilization do not raise new questions about safety and effectiveness.

Comment 3

The models and Launch length of subject device are different from the predicate device. The model was more than as the predicated products, while the puncture depths are same. Different models are only different in the outer diameter of the needle, which allowed to choose to meet blood volume needs. Different needle specification will be selected by physician per patient's condition and this different were addressed by performance tests. Testing of performance shows no impact of launch length on the sharpness /penetration force or bond between the lancet body and needle because the puncture depth is tuned by the lancing device, not the lancet itself, this difference does not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.

VII Summary of non-clinical testing

The following performance data were provided in support of the substantial equivalence determination.

NoTesting itemSpecificationResult
01AppearanceLancet should have the same color, nobubble, no flash, no slip.Pass
02Launch LengthThe length of the needles in the DisposableBlood Lancets is different in differentgauge. The general depth is 3.2mmPass
03Sharpness/PenetrationtestingPenetration force ≤1.00N.Pass
04ExteriorThe connection between needle andneedle body should be firm.Pass
05ExteriorCap twist should be smooth.Pass
06Initial bioburdenInitial bioburden of the device shall be lessthan 100CFU/gPass
07SterileThe sterile blood lancet shall be sterilePass
08Cap removal forceThe moment for breaking the safe modeshould range from 30 Ncm to 35 Ncm.Pass
09Needle removal forceThe bond between the lancet body andneedle should be greater than or equal to10N/15s.Pass

{8}------------------------------------------------

10Drop testingThe carton box should have no punctureafter the drop test.Pass
----------------------------------------------------------------------------------------

Biocompatibility testing

Biocompatibility of the Disposable Blood Lancets and Disposable Safety Lancets were evaluated in accordance with ISO 10993-1:2018 for the body contact category. The following tests were performed, as recommended:

Cytotoxic testISO 10993-5:2009
Skin sensitization testISO 10993-23:2021
Intracutaneous testISO 10993-10:2021
Acute systemic toxicity testISO 10993-11:2017
Hemolysis testISO 10993-4:2017
Pyrogen TestISO 10993-11:2017

Sterilization and shelf life testing

  • A Irradiation sterilization validation per ISO 11137-1 and ISO 11137-3.
  • Pyrogen testing per ISO 10993-11:2017 A
  • Transportation test per ISTA 2A:2011

  • A The 5 years shelf life of the device is determined based on stability study which includes ageing test.

VIII Conclusion

The Disposable Blood Lancets is substantially equivalent to its predicate device (Promisemed Blood Lancet). The differences between the predicate and subject device do not raise any new or different questions of safety or effectiveness. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.

§ 878.4850 Blood lancets.

(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.