K Number
K222148

Validate with FDA (Live)

Date Cleared
2023-08-16

(392 days)

Product Code
Regulation Number
864.5220
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, EO%, EO#, BASO%, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.

Device Description

HemoScreen is a point of care (POC), automated hematology analyzer that provides 20 common CBC parameters, including a 5-part leukocyte (WBC) differential, in capillary and venous whole blood samples. The HemoScreen analyzer (reader) is a tabletop device that is designed to use with a disposable reagent Cartridge. In addition to the Cartridge, the system includes a disposable Sampler with two glass capillaries which is used to collect the blood sample and then transfer it to the Cartridge.

Once the Cartridge is inserted into the reader, there are no further procedural steps; blood is expelled from the capillaries (Sampler) into the reagent compartments (Cartridge). The reader then mixes the blood sample with the reagents by alternately pressing compressible portions of the Cartridge, eventually causing the suspension of cells to flow into the microfluidic chamber. Cells flowing in the microfluidic chamber focus into a single-cell plane due to a patented physical phenomenon known as viscoelastic focusing.

The reader then captures images of the focused cells and analyzes them in real time using machine vision algorithms. When analysis is complete, the results are displayed to the user on the reader's touch screen and may be printed to an adjacent printer or exported to a USB flash drive. The Cartridge is ejected by the analyzer after analysis, and can then be safely disposed of, as the reagents and blood sample remain within the Cartridge.

The basic staining and microscopic image analysis performed by HemoScreen closely resembles the traditional blood smear and the hemocytometer counting chamber. Leukocytes are classified based on their staining properties and morphology, whereas absolute counts are obtained by counting the cells contained in a chamber of predetermined volume. Test results are obtained within six (6) minutes and the results are saved.

Quality Control: Commercial 3-level liquid quality controls, PIX-CBC Hematology Controls, are recommended for use with the HemoScreen. These controls cover all the tested parameters and are sampled the same way whole blood is sampled.

Software: The HemoScreen software displays an intuitive, simple-to-use user interface that is operated via the touch screen. The software is responsible for operating the device, performing the measurements, and recording the results.

AI/ML Overview

The provided text describes the HemoScreen Hematology Analyzer's performance and the study conducted to demonstrate its substantial equivalence. The key change evaluated was the introduction of direct fingerstick sampling, compared to the previously cleared indirect fingerstick sampling method.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a numerical or categorical format for each parameter. Instead, it relies on correlation and regression analysis to show agreement between the new direct sampling method and the previously cleared indirect sampling method. The implicit acceptance criteria are likely based on acceptable Passing-Bablok regression parameters (slope and intercept ideally close to 1 and 0, respectively) and high Pearson Correlation coefficients.

However, the reported performance is provided as follows:

Performance of HemoScreen (Direct Fingerstick) vs. HemoScreen (Indirect Fingerstick) for 20 Parameters

ParameterNResult Range (Indirect)Intercept [95% CI]Slope [95% CI]Pearson Correlation
WBC (10³/μL)424.54-12.52-0.216 [-0.893, 0.412]1.028 [0.955, 1.108]0.978
RBC (10⁶/μL)424.14-5.780 [-0.401, 0.498]0.994 [0.881, 1.078]0.97
HGB (g/dL)4211.84-17.12-0.749 [-2.348, 1.097]1.048 [0.913, 1.156]0.969
HCT (%)4235.76-50.11-1.77 [-7.048, 3.527]1.032 [0.901, 1.157]0.967
MCV (fL)4276.6-93.841.013 [-2.325, 3.887]0.988 [0.954, 1.025]0.994
MCH (pg)4224.67-32.98-0.191 [-0.907, 0.605]1.007 [0.981, 1.032]0.998
MCHC (g/dL)4232.2-35.83-1.519 [-4.767, 1.859]1.047 [0.949, 1.143]0.961
RDW (%)4211.64-14.16-0.25 [-0.876, 0.266]1.019 [0.977, 1.068]0.991
PLT (10³/μL)42142.4-399.63.646 [-16.602, 21.01]0.963 [0.894, 1.051]0.979
MPV (fL)429.18-13.460.572 [-1.036, 1.837]0.949 [0.832, 1.113]0.92
NEUT (10³/μL)422.19-7.95-0.221 [-0.531, 0.062]1.049 [0.995, 1.109]0.984
LYMP (10³/μL)421.51-4.17-0.153 [-0.562, 0.251]1.075 [0.925, 1.273]0.953
MONO (10³/μL)420.27-0.86-0.087 [-0.256, 0.023]0.996 [0.796, 1.358]0.825
EO (10³/μL)420.04-0.53-0.021 [-0.032, -0.005]1.017 [0.945, 1.081]0.988
BASO (10³/μL)420.01-0.070.003 [-0.01, 0.012]1.225 [0.842, 1.659]0.598
NEUT (%)4243.75-68.8-2.615 [-11.774, 4.007]1.05 [0.941, 1.194]0.957
LYMP (%)4219.95-45.05-1.418 [-4.219, 1.404]1.073 [0.983, 1.162]0.969
MONO (%)423.95-11.25-2.108 [-3.977, -0.95]1.147 [0.985, 1.459]0.823
EO (%)420.7-7.3-0.27 [-0.429, -0.028]1.029 [0.968, 1.097]0.987
BASO (%)420.1-0.7-0.075 [-0.3, 0.15]1.5 [1, 2]0.614

2. Sample size and Data Provenance

  • Test Set Sample Size: 42 subjects.
  • Data Provenance: The study described is a "prospective clinical study" involving "matched fingerstick collections." The document does not explicitly state the country of origin, but the company (PixCell Medical Technologies, Ltd.) is based in Israel. Given the FDA submission, it's likely the study was conducted to meet US regulatory requirements, but the specific location isn't provided.

3. Number of Experts and Qualifications for Ground Truth

The document does not mention the use of experts to establish a "ground truth" in the traditional sense of consensus reading for images. The study compares two different methods of sampling with the same device (HemoScreen) – direct vs. indirect fingerstick. The "ground truth" for the comparison is essentially the measurement obtained by the indirect sampling method, which was previously cleared. There are no mentions of radiologists, pathologists, or similar experts reviewing cases to establish a ground truth for a diagnostic AI.

4. Adjudication Method for the Test Set

Not applicable. This was a method comparison study for quantitative measurements, not a diagnostic AI study requiring adjudication of expert interpretations.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This study focused on comparing two sampling methods for a highly automated hematology analyzer, not on human readers' interpretive performance with or without AI assistance.

6. Standalone (Algorithm Only) Performance

The HemoScreen device itself is described as using "machine vision algorithms" to analyze images of cells. Therefore, the reported performance metrics (Pearson correlation, Passing-Bablok regression) represent the standalone performance of the device's algorithms in quantifying blood parameters. The study specifically compared the input method (direct vs. indirect fingerstick) to the device, not an AI assisting a human.

7. Type of Ground Truth Used

The ground truth used for performance comparison was the quantitative measurements obtained from the HemoScreen device itself, using the indirect fingerstick sampling method. This method was previously cleared (K180020) and served as the reference for evaluating the new direct sampling method. It's a "device-to-device" or "method-to-method" comparison, rather than comparison to a clinical expert consensus, pathology, or outcomes data.

8. Sample Size for the Training Set

The document does not explicitly describe a "training set" for the "machine vision algorithms." The algorithms are inherent to the device's operation. It's possible the algorithms were developed and validated internally during the device's initial design and prior K180020 clearance, but this specific submission focuses on validating the direct fingerstick sampling method rather than training new algorithms or updating existing ones. Thus, the sample size for a training set is not provided in this document.

9. How Ground Truth for Training Set Was Established

As mentioned above, information regarding a specific "training set" and its ground truth establishment is not provided in this document. The existing HemoScreen device's core technology and algorithms were already cleared under K180020. This submission primarily validates a new sample collection modality for the same device and its existing algorithms.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

August 16, 2023

PixCell Medical Technologies, Ltd. % Erika Ammirati President Ammirati Regulatory Consulting 575 Shirlynn Court Los Altos, California 94022

Re: K222148

Trade/Device Name: HemoScreen Hematology Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Dated: March 29, 2023 Received: March 30, 2023

Dear Erika Ammirati:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Min Wu-S

Min Wu, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222148

Device Name HemoScreen Hematology Analyzer

Indications for Use (Describe)

The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, EO%, EO#, BASO%, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K222148.

807.92 (a)(1): Name:PixCell Medical Technologies, Ltd.
Address:6 Hayezira St.
Yoknaem Ilit, Israel 2069202
Phone:+972-4-9593516
Email:yaara@pixcell-medical.com
Contact:Yaara Ben-Yosef, PhD

807.92 (a)(2): Device name- trade name and common name, and classification

Trade name: HemoScreen Hematology Analyzer

Common Name: Automated differential cell counter

Classification: 21 CFR 864.5220

807.92 (a)(3): Identification of the legally marketed predicate devices

HemoScreen Hematology Analyzer (PixCell Medical Technologies, Ltd), cleared under K180020.

807.92 (a)(4): Device Description

HemoScreen is a point of care (POC), automated hematology analyzer that provides 20 common CBC parameters, including a 5-part leukocyte (WBC) differential, in capillary and venous whole blood samples. The HemoScreen analyzer (reader) is a tabletop device that is designed to use with a disposable reagent Cartridge. In addition to the Cartridge, the system includes a disposable Sampler with two glass capillaries which is used to collect the blood sample and then transfer it to the Cartridge.

Once the Cartridge is inserted into the reader, there are no further procedural steps; blood is expelled from the capillaries (Sampler) into the reagent compartments (Cartridge). The reader then mixes the blood sample with the reagents by alternately pressing compressible portions of the Cartridge, eventually causing the suspension of cells to flow into the microfluidic chamber. Cells flowing in the microfluidic chamber focus into a single-cell plane due to a patented physical phenomenon known as viscoelastic focusing.

The reader then captures images of the focused cells and analyzes them in real time using machine vision algorithms. When analysis is complete, the results are displayed to the user on the reader's touch screen and may be printed to an adjacent printer or exported to a USB flash drive. The Cartridge is ejected by the analyzer after analysis, and can then be safely disposed of, as the reagents and blood sample remain within the Cartridge.

{4}------------------------------------------------

The basic staining and microscopic image analysis performed by HemoScreen closely resembles the traditional blood smear and the hemocytometer counting chamber. Leukocytes are classified based on their staining properties and morphology, whereas absolute counts are obtained by counting the cells contained in a chamber of predetermined volume. Test results are obtained within six (6) minutes and the results are saved.

Quality Control: Commercial 3-level liquid quality controls, PIX-CBC Hematology Controls, are recommended for use with the HemoScreen. These controls cover all the tested parameters and are sampled the same way whole blood is sampled.

Software: The HemoScreen software displays an intuitive, simple-to-use user interface that is operated via the touch screen. The software is responsible for operating the device, performing the measurements, and recording the results.

807.92 (a)(5): Intended Use

The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, EO%, EO#, BASO%, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.

807.92 (a)(6): Technological Similarities and Differences to the Predicate

The following chart describes similarities and differences between HemoScreen and the predicate.

ComparisonON-TESTHemoScreen with direct fingersticksamplingPREDICATEHemoScreen with indirect fingersticksampling (K180020)
Intended UseAutomated hematology analyzerSame
ParametersMeasured• Red Blood Cells (RBC),• White Blood Cells (WBC),• Platelets (PLT),• Hemoglobin (HGB),• Hematocrit (HCT),• Mean Corpuscular (erythrocyte)Volume (MCV),• Mean Cell (erythrocyte) Hemoglobin(MCH),• Mean Cell (erythrocyte) HemoglobinConcentration (MCHC),• Red Blood Cell Distribution Width(RDW)-CV• Mean Platelets Volume (MPV),• Neutrophils (NEUT; #/%),Same
ComparisonON-TESTHemoScreen with direct fingersticksamplingPREDICATEHemoScreen with indirect fingersticksampling (K180020)
Lymphocytes (LYMP; #/%), Eosinophils (EO; #/%) and Basophiles (BASO; #/%)
ClassClass IISame
RegulationNumber21 CFR 864.5220Same
Product CodeGKZSame
FDA BranchHematologySame
Throughput10 samples/hourSame
Test PrincipleThe HemoScreen uses a novel focusingmethod called viscoelastic focusing whichcauses the cells to perfectly align into aplane. High resolution microscopic imagesare taken of the flowing cells. Each imageis analyzed using machine visionalgorithms and the different cell types aredifferentiated and counted. WBCs arestained prior to analysis so as to enabledifferentiation between their subtypes andabnormal cells.HGB is calculated based on the opticaldensity measured on intact individual cells.Same
CalibrationFactory calibratedSame
Sample Type-venousK2EDTA anticoagulated whole bloodSame
Sample Type-fingerstickDirect and indirectDirect: Capillary blood from fingertipdrawn directly into Sampler.Indirect: Capillary blood from fingertipdelivered to microtube, then transferredinto SamplerIndirect, only
Sample Volume40μLSame

{5}------------------------------------------------

{6}------------------------------------------------

807.92 (b)(1): Brief Description of Nonclinical Data- no changes from K180020

Limit of Blank

Please refer to the 510(k) Summary for K180020.

Limits of Detection and Quantitation

Please refer to the 510(k) Summary for K180020.

Linearity

Please refer to the 510(k) Summary for K180020.

Repeatability of Venous Samples (internal and external operators)

Please refer to the 510(k) Summary for K180020.

Reproducibility of OC Materials Please refer to the 510(k) Summary for K180020.

Interference

Please refer to the 510(k) Summary for K180020.

Reference Intervals- Adult Males and Females (CLSI EP28-A3C)

Please refer to the 510(k) Summary for K180020.

807.92 (b)(2): Brief Description of Clinical Data-

No changes from K180020 for venous sampling and indirect capillary sampling into a K2EDTA microtube. The data in this section reflect the comparisons between direct capillary sampling (ontest), and indirect capillary sampling (predicate).

Comparisons between Direct Capillary Sampling and Indirect Capillary Sampling.

A comparison HemoScreen to HemoScreen study was performed where results from direct sampling of fingerstick blood into the Sampler were compared to results from indirect sampling, where fingerstick blood was first collected into a K2EDTA microtube, and then transferred to the Sampler. This was a prospective clinical study that included 42 subjects and matched fingerstick collections. The data were evaluated by Passing-Bablok regression and Pearson's correlation for all 20 parameters, and the summarized data are provided below.

{7}------------------------------------------------

ParameterNResult Range(Indirect)Intercept [95% CI]Slope [95% CI]PearsonCorrelation
WBC (10³/μL)424.54-12.52-0.216 [-0.893, 0.412]1.028 [0.955, 1.108]0.978
RBC (10⁶/μL)424.14-5.780 [-0.401, 0.498]0.994 [0.881, 1.078]0.97
HGB (g/dL)4211.84-17.12-0.749 [-2.348, 1.097]1.048 [0.913, 1.156]0.969
HCT (%)4235.76-50.11-1.77 [-7.048, 3.527]1.032 [0.901, 1.157]0.967
MCV (fL)4276.6-93.841.013 [-2.325, 3.887]0.988 [0.954, 1.025]0.994
MCH (pg)4224.67-32.98-0.191 [-0.907, 0.605]1.007 [0.981, 1.032]0.998
MCHC (g/dL)4232.2-35.83-1.519 [-4.767, 1.859]1.047 [0.949, 1.143]0.961
RDW (%)4211.64-14.16-0.25 [-0.876, 0.266]1.019 [0.977, 1.068]0.991
PLT (10³/μL)42142.4-399.63.646 [-16.602, 21.01]0.963 [0.894, 1.051]0.979
MPV (fL)429.18-13.460.572 [-1.036, 1.837]0.949 [0.832, 1.113]0.92
NEUT (10³/μL)422.19-7.95-0.221 [-0.531, 0.062]1.049 [0.995, 1.109]0.984
LYMP (10³/μL)421.51-4.17-0.153 [-0.562, 0.251]1.075 [0.925, 1.273]0.953
MONO (10³/μL)420.27-0.86-0.087 [-0.256, 0.023]0.996 [0.796, 1.358]0.825
EO (10³/μL)420.04-0.53-0.021 [-0.032, -0.005]1.017 [0.945, 1.081]0.988
BASO (10³/μL)420.01-0.070.003 [-0.01, 0.012]1.225 [0.842, 1.659]0.598
NEUT (%)4243.75-68.8-2.615 [-11.774, 4.007]1.05 [0.941, 1.194]0.957
LYMP (%)4219.95-45.05-1.418 [-4.219, 1.404]1.073 [0.983, 1.162]0.969
MONO (%)423.95-11.25-2.108 [-3.977, -0.95]1.147 [0.985, 1.459]0.823
EO (%)420.7-7.3-0.27 [-0.429, -0.028]1.029 [0.968, 1.097]0.987
BASO (%)420.1-0.7-0.075 [-0.3, 0.15]1.5 [1, 2]0.614

Passing-Bablok Regression Data and Pearson Correlation

CI = confidence interval

Precision of Direct Capillary Sampling

Twenty-three (23) subjects were enrolled for the direct capillary precision study. Two capillary punctures were performed on each subject, one puncture by each of two operators. The fingersticks were done on the same hand, with one operator collecting blood using the HemoScreen Sampler from one finger and completing the test, and the other operator collecting blood using the HemoScreen Sampler from another finger and completing the test. The operator order was randomized to counteract order effects, and the two punctures were done within 10 minutes of each other. One HemoScreen analyzer and multiple Cartridge lots were used for testing. Neither operator was aware of the test results from the other operator. Precision in capillary blood was shown to be adequate.

807.92 (b)(3): Conclusions from Nonclinical and Clinical Data

The conclusions drawn from the updated clinical data demonstrate that the device is safe and effective for its intended use.

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”