K Number
K221108

Validate with FDA (Live)

Date Cleared
2022-07-12

(88 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nitrile Medical Examination Gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

Nitrile Medical Examination Gloves is a patient examination glove and it is made from nitrile compound. It is a single-use, powder-free and non-sterile product. The glove is worn on the examiner's hand to prevent contamination between patient and examiner, and this device meets all the requirement specifications in the ASTM D6319-19 Standard.

AI/ML Overview

The provided text describes the acceptance criteria and the study conducted for "Nitrile Medical Examination Gloves." This device is a Class I medical device, and the evaluation focuses on its substantial equivalence to a predicate device (K192333).

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Test ItemAcceptance Criteria (from ASTM D6319-19 unless otherwise specified)Reported Device Performance
Freedom from HolesAny glove showing droplet, stream, or other water leakage considered failed. AQL 2.5Did not show droplet, stream, or other type of water leakage (0 failures out of 200 samples tested).
Physical Dimension TestLength: S: Min 220mm, M: Min 230mm, L: Min 230mm, XL: Min 230mmWidth: S: 80±10mm, M: 95±10mm, L: 110±10mm, XL: 120±10mmThickness: Finger Min 0.05mm, Palm Min 0.05mmLength: S: 241-246mm, M: 243-252mm, L: 287-298mm, XL: 288-292mmWidth: S: 85-88mm, M: 92-94mm, L: 109-110mm, XL: 114-117mmThickness: Finger 0.08-0.125mm, Palm 0.063-0.082mm(0 failures out of 13 samples tested/size).
Physical Requirement Test (Tensile Strength)Before Aging: 14 Mpa, min.After Aging: 14 Mpa, min.Before Aging: 19.6-26.7 MpaAfter Aging: 20.9-26.1 Mpa(0 failures out of 13 samples tested/L size).
Physical Requirement Test (Elongation)Before Aging: 500% min.After Aging: 400% min.Before Aging: 650-700%After Aging: 560-620%(0 failures out of 13 samples tested/L size).
Powder Free GlovesMax. 2.0mg/gloveAverage: 1.9 mg (5 samples tested).
In Vitro CytotoxicityUnder conditions of the study, the test article must not show potential toxicity (ISO 10993-5:2009)Fail – Under the condition of the test, the test article was found to be cytotoxic. (Note 2 states that additional testing for systemic toxicity was performed due to this result, and it passed).
Skin SensitizationUnder the conditions of the study, the test article must be found to be non-sensitizing (ISO 10993-10:2010)Pass – Under the condition of the test, the test article was found to be non-sensitizing.
Skin Irritation TestUnder the conditions of the test, the test article must be found to be non-irritating (ISO 10993-10:2010)Pass – Under the conditions of the test, the test article was found to be non-irritating.
Acute Systemic ToxicityUnder the conditions of the test, the test article must be found to be non-systemic toxicity (ISO 10993-11:2017) (Performed in response to cytotoxicity failure, as per Note 2 and comparison to predicate)Pass – Under the condition of the test, the test article was found to be non-systemic toxicity. (This was done to address the initial cytotoxicity finding, and the device met the requirements, similar to the predicate device which also had initial cytotoxicity but acceptable systemic toxicity.)

2. Sample Size Used for the Test Set and Data Provenance

  • Freedom from Holes: 200 samples
  • Physical Dimension Test: 13 samples per size (S, M, L, XL - so a total of 52 samples for dimensions)
  • Physical Requirement Test (Tensile Strength & Elongation): 13 samples for L size (implies this size was representative or the primary one tested for these characteristics).
  • Powder Free Gloves: 5 samples.
  • Biocompatibility Tests (Cytotoxicity, Skin Sensitization, Skin Irritation, Acute Systemic Toxicity): The exact number of samples for each biocompatibility test is not explicitly stated in this document (e.g., number of animals or cell cultures). However, these tests are generally conducted according to the specified ISO standards, which define their own sample size requirements.

Data Provenance: The document does not explicitly state the country of origin for the data generation. However, the submitter is based in China, suggesting the testing was likely conducted or overseen by facilities associated with the manufacturer, potentially in China or by internationally recognized labs. The studies are non-clinical bench testing, not clinical studies with human patients. Thus, the distinction between retrospective and prospective is less relevant here.


3. Number of Experts Used to Establish Ground Truth and Qualifications

This device did not involve human experts establishing ground truth for a diagnostic AI. The evaluation is based on bench testing against established ASTM and ISO standards. The "ground truth" is defined by the objective pass/fail criteria within those standards.


4. Adjudication Method for the Test Set

Not applicable. This was a non-clinical bench testing study, not a medical image or diagnostic test requiring expert adjudication. The results are objective measurements against defined standards.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This device is a medical examination glove, not a diagnostic AI system or a device intended to assist human readers in interpretation. Therefore, an MRMC comparative effectiveness study is not applicable.


6. Standalone Performance (Algorithm only without human-in-the-loop performance)

No. This is a physical medical device (glove), not an algorithm or AI system.


7. Type of Ground Truth Used

The ground truth for the performance evaluation is based on established industry standards and objective measurements:

  • ASTM D6319-19: Standard Specification for Nitrile Examination Gloves for Medical Application.
  • ASTM D5151: Test Method for Detection of Holes in Medical Gloves.
  • ASTM D412: Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—Tension.
  • ASTM D3767: Practice for Rubber—Measurement of Dimensions.
  • ASTM D573: Test Method for Rubber—Deterioration in an Air Oven.
  • ASTM D6124: Test Method for Residual Powder on Medical Gloves.
  • ISO 10993-5:2009: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
  • ISO 10993-10:2010: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
  • ISO 10993-11:2017: Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.

8. Sample Size for the Training Set

Not applicable. This is a physical medical device, not a machine learning model that requires a training set.


9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 12, 2022

Anhui Fulewei Electronic Technology Co.,Ltd Beina Gong Manager South Of Binheroad, West Of Gangkou Road, Economic Development Zone, Fuyang, Anhui 236000 China

Re: K221108

Trade/Device Name: Nitrile Medical Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: April 1, 2022 Received: April 15, 2022

Dear Beina Gong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221108

Device Name Nitrile Medical Examination Gloves

Indications for Use (Describe)

Nitrile Medical Examination Gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary K221108

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: July 12, 2022

1. Submitter's Information

Name:ANHUI FULEWEI ELECTRONIC TECHNOLOGY CO., LTD
Address:SOUTH OF BINHEROAD, WEST OF GANGKOU ROAD,ECONOMIC DEVELOPMENT ZONE, FUYANG CITY, ANHUI,CHINA 236000
Contact person:Beina Gong
Title:Manager
E-mail:19158151@qq.com
Tel:86-15055553699

2. Device Identification

Trade/Device Name:Nitrile Medical Examination Gloves
Size of gloveS, M, L, XL
Common name:Polymer Patient Examination Glove
Regulation Number:880.6250
Regulation Name:Non-powdered patient examination glove
Regulation Class:Class I
Panel:General Hospital
Product Code:LZA

3. Predicate Device

510(K) number:K192333
Device Name:Blue Nitrile Examination Gloves Powder Free
Manufacturer:JR Engineering & Medical Technologies (M) SDN.BHD
Common namePolymer Patient Examination Glove
Regulation Number:880.6250
Regulation Name:Non-powdered patient examination glove
Regulation Class:Class I
Panel:General Hospital
Product Code:LZA

4. Device Description

Nitrile Medical Examination Gloves is a patient examination glove and it is made from nitrile compound. It is a single-use, powder-free and non-sterile product. The glove is worn on the 510(k) Summary Page 1 of Page 6

{4}------------------------------------------------

examiner's hand to prevent contamination between patient and examiner, and this device meets all the requirement specifications in the ASTM D6319-19 Standard.

5. Indication for use

Nitrile Medical Examination Gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

6. Summary of the technological characteristics of the device compared to the Predicate Device

Compared to the predicate device, the subject device has the same intended use, similar product design, and similar performance as the predicate device, the summarized comparison information is listed in the following table:

SE ComparisonsSubject DeviceK221108Predicate DeviceK192333Remarks
NameNitrile MedicalExamination GlovesBlue Nitrile ExaminationGloves Powder Free/
SizeSmall,Medium, Large, ExtraLargeExtra Small, Small,Medium, Large, ExtraLargeSee Note1
ClassificationClass IClass ISame
Intended use / indications foruseNitrile MedicalExamination Gloves isdisposable deviceintended for medicalpurpose that is worn onthe examiner's hand toprevent contaminationbetween patient andexaminer.JR MEDIC Blue NitrileExamination GlovesPowder Free isdisposable devicesintended for medicalpurpose that are worn onthe examiner's hand toprevent contaminationbetween patient andexaminer.Same
DimensionsASTMD6319- 19Length Min: 220mmWidth: 80±10mm(for S size)Length Min: 230mmWidth: 95±10mm(for M size)Length Min: 230mmWidth: 110±10mm(for L size)Length Min: 230mmWidth: 120±10mm(for XL size)Length Min 230 mmWidth Min 95±10mm(for medium size)Note1
PhysicalPropertiesASTMD6319- 19Before AgingTensile Strength Min 14MpaBefore AgingTensile Strength Min 14MpaSame
Ultimate Elongation Min500%After AgingTensile Strength Min 14MpaUltimate Elongation Min400%Ultimate Elongation Min500%After AgingTensile Strength Min 14MpaUltimate Elongation Min400%
Water leakASTM D6319-19AQL 2.5AQL 2.5Same
ThicknessASTM D6319-19Palm Min 0.05 mmFinger Min 0.05 mmPalm Min 0.05 mmFinger Min 0.05 mmSame
PowderFreeASTM D6319-19$≤$ 2 mg/glove$≤$ 2 mg/gloveSame
BiocompatibilityISO 10993-10:2010Under the condition ofstudy not anirritantUnder the condition ofstudy not anirritantSame
ISO 10993-10:2010Under the conditions ofthe study not a sensitizerUnder the conditions ofthe study not a sensitizerSame
ISO 10993-5:2009Under the conditions ofthe study,cytotoxic.Additional testing wasperformed to determine ifthis was a systemictoxicity concernUnder the conditions ofthe study,cytotoxic.Additional testing wasperformed to determine ifthis was a systemictoxicity concernSameNote 2
ISO 10993-11:2017Under the condition ofstudy the device extractsdo not pose a systemictoxicity concernUnder the condition ofstudy the device extractsdo not pose a systemictoxicity concernSameNote 2
ISO 10993-11:2017/Under the conditions ofthe study, the device didnot demonstrate amaterial mediatedpyrogenicity response.Note 3
Single UseYesYesSame
Non-SterileYesYesSame
ColorBlueBlueSame
Intended use populationAdultAdultSame
WearambidextrousambidextrousSame
EXPIRE DATENot claimIncludedSee Note4

{5}------------------------------------------------

{6}------------------------------------------------

Note 1:

In this submission, the sizes "S, M, L and XL" of Nitrile Medical Examination Gloves have been performed the testing according to the ASTM D6319-19, the dimension and performance meet the requirements of ASTM D6319-19, the difference in sizes does not raise different questions of safety and effectiveness.

Note 2:

We evaluated In vitro cytotoxicity in accordance with ISO 10993-5:2009, the result demonstrated toxicity to L929 cells. Additionally, we evaluated acute systemic toxicity in accordance with ISO 10993-11:2017. Under the conditions of the study, there was no evidence of systemic toxicity from the test article extract. Similar to the predicate device, the proposed device met the requirements of the study.

Note 3:

We do not claim "non-pyrogenic" and we evaluated the biocompatibility per ISO 10993-1, the material mediated pyrogenicity testing is unnecessary.

Note 4: According to FDA guidance, claiming shelf life is non-mandatory.

7. Performance Data

Clinical test:

Clinical testing is not required.

Non-clinical data

The following bench testing was conducted.

ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical 1. Application.

Freedom from Holes - Testing for freedom from holes was conducted in accordance with 2. Test Method ASTM D5151.

  1. Physical Dimension test - Determine the dimension as directed in Table 2 of ASTMD6319-19.

{7}------------------------------------------------

  1. Physical Requirement Test - Before and accelerated aging, the physical requirement specified in Table 3 of ASTM D6319-19, tests were conducted in accordance with test method ASTM D412 and accelerated aging test were conducted in accordance with Test Method ASTMD573.

  2. Powder Free Gloves - Determine the powder residue in accordance with Test Method

  • ASTM D6124.
StandTest itemTestCriteriaResultConclusion
ardmethod
ASTMD6319-19freedomfrom holesASTM D5151Any glove that shows adroplet, stream, orother type of waterleakage shall beconsidered to havefailed the test.AQL 2.5Did not show adroplet, stream,or other type ofwater leakage200samplestested0 failuresPass
PhysicalDimensiontestASTM D412ASTMD3767Length Min:220mmWidth Min:80±10mm(for S size)Length Min:230mmWidth Min:95±10mm(for M size)Length Min:230mmWidth Min:110±10mm(for L size)Length Min:230mmWidth Min:120±10mm(for XL size)Thickness: Finger0.05minPalm 0.05minLength:241246mmWidth: 8588mm(for S size)Length:243252mmWidth: 9294mm(for M size)Length:287298mmWidth:109110mm(for L size)Length:288292mmWidth:114117mm(for XL size)Thickness:Finger0.080.125mmPalm0.0630.082mm13 samplestested/size0 failurePass
PhysicalRequirementTestASTMD412ASTMD573Tensile Strength:Before Aging 14 Mpa,min.After Aging 14 Mpa,min.Elongation:Before Aging 500%min.After Aging 400% min.Tensile Strength:Before Aging19.626.7 Mpa,After Aging20.926.1 Mpa,Elongation:Before Aging650700%After Aging560620%13 samplestested/Lsize0 failurePass
PowderFree GlovesASTM D6124Max. 2.0mgsample quantity:5pcsAverage: 1.9 mgPass

{8}------------------------------------------------

Biocompatibility:

  1. ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic ನ toxicity.

  1. ISO 10993-5:2009 Biological evaluation of medical devices - Part 5 Tests for In vitro cytotoxicity
Test itemTest StandardCriteriaResults
In Vitro CytotoxicityISO 10993-5: 2009Under conditions of thestudy, the test articlemust not show potentialtoxicity.Fail—Under the condition ofthe test, the test articlewas found to becytotoxic
Skin SensitizationISO 10993-10: 2010Under the conditionsof the study, the testarticle must be foundto be non-sensitizing.Pass—Under the condition ofthe test, the test articlewas found to be non-sensitizing
Skin Irritation testISO 10993-10: 2010Under the conditionsof the test, the testarticle must be foundto be non- irritatingPass—Under the conditions ofthe test, the test articlewas found to be non-irritating
Acute systemictoxicityISO 10993-11:2017Under the conditionsof the test, the testarticle must be foundto be non- systemictoxicityPass—Under the condition ofthe test, the test articlewas found to be non-systemic toxicity

8. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the predicated device(K192333).

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.