K Number
K213954

Validate with FDA (Live)

Date Cleared
2023-02-03

(413 days)

Product Code
Regulation Number
866.3981
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BIOFIRE® SPOTFIRE® Respiratory (R) Panel (SPOTFIRE R Panel) is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE® System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19.

The following organism types are identified and differentiated using the SPOTFIRE R Panel:

Viruses: Adenovirus Coronavirus (seasonal) Coronavirus SARS-CoV-2 Human metapneumovirus Human rhinovirus/enterovirus Influenza A virus Influenza A virus A/H1-2009 Influenza A virus A/H3 Influenza B virus Parainfluenza virus Respiratory syncytial virus

Bacteria: Bordetella parapertussis Bordetella pertussis Chlamydia pneumoniae Mycoplasma pneumoniae

Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS specimens during the acute phase of infection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of respiratory infection are indicative of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by an NPS specimen. Positive results do not rule out coinfection with other organisms. The agent(s) detected by the SPOTFIRE R Panel may not be the definite cause of disease.

Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

Device Description

The BIOFIRE® SPOTFIRE® Respiratory (R) Panel (SPOTFIRE R Panel) simultaneously identifies 15 different respiratory viral and bacterial pathogens in nasopharyngeal swabs (NPS) from individuals with signs and symptoms of respiratory tract infection (see Table 1) . The SPOTFIRE R Panel is compatible with the BIOFIRE® System, a polymerase chain reaction (PCR)-based in vitro diagnostic system for infectious disease testing. The SPOTFIRE System Software executes the SPOTFIRE R Panel test and reports the test results. The SPOTFIRE R Panel was designed to be used in CLIA-waived environments.

A test is initiated by loading Hydration into one port of the SPOTFIRE R Panel pouch and a NPS specimen mixed with the provided Sample Buffer into the other port of the SPOTFIRE R Panel pouch and placing it in the SPOTFIRE System. The pouch contains all of the reagents required for specimen testing and analysis in a freezedried format; the addition of Hydration and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the SPOTFIRE System Software guides the user through the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

The SPOTFIRE System contains coordinated systems of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the appropriate times. Two Pettier devices control heating and cooling of the PCR reactions and the melt curve analysis.

Nucleic acid extraction occurs within the SPOTFIRE R Panel pouch using mechanical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the SPOTFIRE system performs a nested multiplex PCR that is executed in two stages. During the first stage, the SpotFire system s a single, large volume, highly multiplexed reverse transcription PCR (tt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double-stranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in singleplex fashion in each well of the conclusion of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

The SPOTFIRE System Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

AI/ML Overview

Here's a breakdown of the requested information, based on the provided FDA 510(k) summary for the BIOFIRE® SPOTFIRE® Respiratory (R) Panel:

1. Table of Acceptance Criteria (Implicit) and Reported Device Performance

While explicit acceptance criteria (e.g., "PPA must be >95%") are not directly stated as pass/fail thresholds in this summary but rather derived from the observed performance of the predicate device, we can infer the expected performance based on the clinical evaluation results. The reported performance is shown in the "Clinical Performance" section.

Acceptance Criteria (Inferred from Clinical Performance)Reported Device Performance (Positive Percent Agreement, PPA)Reported Device Performance (Negative Percent Agreement, NPA)
Viruses
Adenovirus97.0% (95% CI: 84.7-99.5%) - Prospective97.8% (95% CI: 96.7-98.5%) - Prospective
100% (95% CI: 89.0-100%) - Archived96.9% (95% CI: 94.9-98.2%) - Archived
Coronavirus SARS-CoV-297.3% (95% CI: 90.5-99.2%) - Prospective99.4% (95% CI: 98.7-99.7%) - Prospective
Coronavirus (seasonal)99.0% (95% CI: 94.7-99.8%) - Prospective98.7% (95% CI: 97.8-99.2%) - Prospective
99.0% (95% CI: 94.3-99.8%) - Archived98.2% (95% CI: 96.3-99.1%) - Archived
Human metapneumovirus100% (95% CI: -) - Prospective100% (95% CI: 99.7-100%) - Prospective
97.0% (95% CI: 84.7-99.5%) - Archived100% (95% CI: 99.2-100%) - Archived
Human rhinovirus/enterovirus99.1% (95% CI: 97.5-99.7%) - Prospective90.6% (95% CI: 88.3-92.5%) - Prospective
96.7% (95% CI: 83.3-99.4%) - Archived96.7% (95% CI: 94.6-98.0%) - Archived
Influenza A virus0/0 (not detected in prospective study)100% (95% CI: 99.7-100%) - Prospective
98.3% (95% CI: 91.0-99.7%) - Archived100% (95% CI: 99.1-100%) - Archived
Influenza A virus A/H1-20090/0 (not detected in prospective study)100% (95% CI: 99.7-100%) - Prospective
96.9% (95% CI: 84.3-99.4%) - Archived100% (95% CI: 99.2-100%) - Archived
Influenza A virus A/H30/0 (not detected in prospective study)100% (95% CI: 99.7-100%) - Prospective
100% (95% CI: 87.5-100%) - Archived100% (95% CI: 99.2-100%) - Archived
Influenza B virus0/0 (not detected in prospective study)100% (95% CI: 99.7-100%) - Prospective
100% (95% CI: 88.6-100%) - Archived100% (95% CI: 87.9-100%) - Archived
Parainfluenza virus98.0% (95% CI: 92.9-99.4%) - Prospective98.9% (95% CI: 98.1-99.4%) - Prospective
98.3% (95% CI: 94.0-99.5%) - Archived98.1% (95% CI: 96.1-99.1%) - Archived
Respiratory syncytial virus96.3% (95% CI: 81.7-99.3%) - Prospective99.8% (95% CI: 99.3-99.9%) - Prospective
100% (95% CI: 90.6-100%) - Archived98.4% (95% CI: 96.8-99.2%) - Archived
Bacteria
Bordetella parapertussis0/0 (not detected in prospective study)100% (95% CI: 99.7-100%) - Prospective
96.0% (95% CI: 80.5-99.3%) - Archived100% (95% CI: 89.6-100%) - Archived
Bordetella pertussis0/0 (not detected in prospective study)100% (95% CI: 99.7-100%) - Prospective
96.4% (95% CI: 82.3-99.4%) - Archived99.1% (95% CI: 97.8-99.7%) - Archived
Chlamydia pneumoniae0/0 (not detected in prospective study)100% (95% CI: 99.7-100%) - Prospective
100% (95% CI: 88.6-100%) - Archived99.6% (95% CI: 98.4-99.9%) - Archived
Mycoplasma pneumoniae0/0 (not detected in prospective study)100% (95% CI: 99.7-100%) - Prospective
100% (95% CI: 89.6-100%) - Archived98.9% (95% CI: 97.4-99.5%) - Archived
Overall ReproducibilityPositive Agreement: 99.1% (95% CI: 98.7-99.4%)Negative Agreement: (Implicitly 100% for negative samples, as success rates are high)

2. Sample Size Used for the Test Set and Data Provenance

  • Prospective Clinical Evaluation:

    • Sample Size: 1120 valid Nasopharyngeal Swab (NPS) specimens (out of 1215 enrolled).
    • Data Provenance: Five geographically distinct urgent care or emergency department study sites. Four in the US and one in the UK. Data was collected prospectively from December 2020 to June 2021. Both fresh (861 specimens) and immediately frozen (259 specimens) samples were included, with no performance difference observed.
  • Testing of Preselected Archived Specimens:

    • Sample Size: 542 valid archived NPS specimens (out of 562 obtained).
    • Data Provenance: Frozen archived NPS specimens obtained from 15 external laboratories worldwide. Retrospectively tested at four US clinical sites.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document states, "The performance of the SPOTFIRE R Panel was evaluated by comparing the test results with those from FDA-cleared multiplexed respiratory pathogen panels." These panels serve as the comparator (ground truth), but the document does not specify:

  • The number of experts
  • Their specific qualifications (e.g., radiologist with 10 years of experience)
  • The process by which those experts, if any, established the ground truth for the comparator methods. It relies on the prior FDA clearance of the comparator devices for their accuracy. Discrepant results were investigated using "additional molecular method" or "standard of care," but the specific expertise applied to these investigations is not detailed.

4. Adjudication Method for the Test Set

The document mentions "investigations of discrepant results are summarized in the footnotes." Discrepant results (FN and FP) were retested with the SPOTFIRE R Panel, or further investigated using an "additional molecular method" or "standard of care." However, a specific "adjudication method" involving multiple expert readers/reviewers (e.g., 2+1, 3+1) is not described. The ground truth relies on the comparator molecular methods.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable as the BIOFIRE® SPOTFIRE® Respiratory (R) Panel is a nucleic acid test (PCR-based in vitro diagnostic system), not an AI-powered diagnostic imaging device that assists human readers. Therefore, an MRMC comparative effectiveness study involving human readers with/without AI assistance was not performed.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

This is an algorithm-only (device-only) performance study. The device is an in vitro diagnostic test that provides a qualitative result (positive/negative) for the presence of specific nucleic acids. The performance data presented (PPA, NPA) directly reflect the device's ability to detect these targets without human interpretation of the raw data. The "SPOTFIRE System Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel."

7. The Type of Ground Truth Used

The ground truth for the clinical studies was established by comparison to FDA-cleared multiplexed respiratory pathogen panels (molecular methods). For discrepant results, "additional molecular method" or "standard of care" was used as further reference. For analytical performance (Limit of Detection, Inclusivity, Specificity), the ground truth was based on known concentrations of characterized isolates (e.g., TCID50/mL, CFU/mL, copies/mL).

8. The Sample Size for the Training Set

The document describes pre-market validation studies for a device, which typically involves analytical and clinical performance testing. It does not specify a "training set" in the context of machine learning model development. The data described (prospective and archived clinical specimens, analytical samples) are primarily for validation/testing of the device's performance based on its defined reagent and instrument protocols. The device's underlying "algorithm" (PCR and melt curve analysis interpretation) is inherent to its design and not typically "trained" in the same way an AI model is.

9. How the Ground Truth for the Training Set Was Established

Since there is no "training set" described in the context of an AI/ML model for this device, this question is not applicable. The device's mechanisms are based on established molecular biology principles (PCR and melt curve analysis) rather than iterative learning from a labeled dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 3, 2023

BioFire Diagnostics Kevin Bourzac Vice President, Regulatory & Clinical Affairs 515 Colorow Drive Salt Lake City, Utah 84108

Re: K213954

Trade/Device Name: BIOFIRE SPOTFIRE Respiratory (R) Panel Regulation Number: 21 CFR 866.3981 Regulation Name: Device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test Regulatory Class: Class II Product Code: QOF, OEM, OOU, OTG, OZE, OZX, OZY, OZZ, OCC, NSU Dated: December 16, 2021 Received: December 17, 2021

Dear Kevin Bourzac:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/5 description: The image shows the text "Joseph Briggs-S" in a large, sans-serif font. The name "Joseph" is on the top line, followed by "Briggs-S" on the second line. The text is black, and the background is a light blue color.

Digitally signed by Joseph Briggs -S Date: 2023.02.03 13:34:47 -05'00'

Joseph Briggs Deputy Branch Chief Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213954

Device Name BIOFIRE® SPOTFIRE® Respiratory (R) Panel

Indications for Use (Describe)

The BIOFIRE® SPOTFIRE® Respiratory (R) Panel (SPOTFIRE R Panel) is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE® System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19.

The following organism types are identified and differentiated using the SPOTFIRE R Panel:

Viruses: Adenovirus Coronavirus (seasonal) Coronavirus SARS-CoV-2 Human metapneumovirus Human rhinovirus/enterovirus Influenza A virus Influenza A virus A/H1-2009 Influenza A virus A/H3 Influenza B virus Parainfluenza virus Respiratory syncytial virus

Bacteria: Bordetella parapertussis Bordetella pertussis Chlamydia pneumoniae Mycoplasma pneumoniae

Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS specimens during the acute phase of infection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of respiratory infection are indicative of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by an NPS specimen. Positive results do not rule out coinfection with other organisms. The agent(s) detected by the SPOTFIRE R Panel may not be the definite cause of disease.

Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with a blue top half and a yellow-green bottom half. The word "BIOMÉRIEUX" is written in white letters in the blue portion of the circle.

BIOFIRE® SPOTFIRE® Respiratory (R) Panel 510(k) Summary BioFire Diagnostics, LLC

Introduction:

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitted by:

BioFire Diagnostics, LLC 515 Colorow Drive Salt Lake City, UT 84108

Telephone: 801-736-6354

Facsimile: 801-588-0507

Contact: Kevin Bourzac, Ph.D ext. 1358

Date Submitted: December 16, 2021

Device Name and Classification:

Trade Name: BIOFIRE® SPOTFIRE® Respiratory (R) Panel

Requlation Number: 21 CFR 866.3981

Classification Name: Multi-target Respiratory Specimen Nucleic Acid Test Including SARS-CoV-2 and Other Microbial Agents

Predicate Device:

DEN200031 - BIOFIRE® Respiratory Panel 2.1 (RP2.1)

Intended Use:

The BIOFIRE® SPOTFIRE® Respiratory (R) Panel (SPOTFIRE R Panel) is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE® System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19.

The following organism types and subtypes are identified and differentiated using the SPOTFIRE R Panel:

{5}------------------------------------------------

VirusesBacteria
AdenovirusBordetella parapertussis
Coronavirus (seasonal)Bordetella pertussis
Coronavirus SARS-CoV-2Chlamydia pneumoniae
Human metapneumovirusMycoplasma pneumoniae
Human rhinovirus/enterovirus
Influenza A virus
Influenza A virus A/H1-2009
Influenza A virus A/H3
Influenza B virus
Parainfluenza virus
Respiratory syncytial virus

Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS specimens during the acute phase of infection. The detection of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of respiratory infection of the presence of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by an NPS specimen. Positive results do not rule out coinfection with other organisms. The agent(s) detected by the SPOTFIRE R Panel may not be the definite cause of disease.

Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

Device Description:

The BIOFIRE® SPOTFIRE® Respiratory (R) Panel (SPOTFIRE R Panel) simultaneously identifies 15 different respiratory viral and bacterial pathogens in nasopharyngeal swabs (NPS) from individuals with signs and symptoms of respiratory tract infection (see Table 1) . The SPOTFIRE R Panel is compatible with the BIOFIRE® System, a polymerase chain reaction (PCR)-based in vitro diagnostic system for infectious disease testing. The SPOTFIRE System Software executes the SPOTFIRE R Panel test and reports the test results. The SPOTFIRE R Panel was designed to be used in CLIA-waived environments.

VirusesBacteria
AdenovirusBordetella parapertussis
Coronavirus (seasonal)Bordetella pertussis
Coronavirus SARS-CoV-2Chlamydia pneumoniae
Human metapneumovirusMycoplasma pneumoniae
Human rhinovirus/enterovirus
Influenza A virus
Influenza A virus A/H1-2009
Influenza A virus A/H3
Influenza B virus
Parainfluenza virus
Respiratory syncytial virus
Table 1. Analytes Detected by the SPOTFIRE (R) Panel

{6}------------------------------------------------

A test is initiated by loading Hydration into one port of the SPOTFIRE R Panel pouch and a NPS specimen mixed with the provided Sample Buffer into the other port of the SPOTFIRE R Panel pouch and placing it in the SPOTFIRE System. The pouch contains all of the reagents required for specimen testing and analysis in a freezedried format; the addition of Hydration and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the SPOTFIRE System Software guides the user through the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

The SPOTFIRE System contains coordinated systems of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the appropriate times. Two Pettier devices control heating and cooling of the PCR reactions and the melt curve analysis.

Nucleic acid extraction occurs within the SPOTFIRE R Panel pouch using mechanical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the SPOTFIRE system performs a nested multiplex PCR that is executed in two stages. During the first stage, the SpotFire system s a single, large volume, highly multiplexed reverse transcription PCR (tt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double-stranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in singleplex fashion in each well of the conclusion of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

The SPOTFIRE System Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

Substantial Equivalence:

The SPOTFIRE R Panel is substantially equivalent to the BioFire Respiratory Panel 2.1 (BioFire RP2.1) (DEN200031), which was cleared on March 17, 2021 and determined to be a Class II device under the classification code 21 CFR 866.3981.

ElementPredicate: BioFire RP2.1 (DEN200031)New Device: SPOTFIRE R Panel
Intended UseThe BioFire Respiratory Panel 2.1 (RP2.1)is a PCR-based multiplexed nucleic acidtest intended for use with the BioFire®FilmArray® 2.0 or BioFire® FilmArray®Torch Systems for the simultaneousqualitative detection and identification ofmultiple respiratory viral and bacterialnucleic acids in nasopharyngeal swabs(NPS) obtained from individuals suspectedof respiratory tract infections, includingCOVID-19The SPOTFIRE R Panel is a PCR-basedmultiplexed nucleic acid test intended foruse with the SPOTFIRE System for thesimultaneous, qualitative detection andidentification of multiple respiratory viral andbacterial nucleic acids in nasopharyngealswab (NPS) specimens obtained fromindividuals with signs and symptoms ofrespiratory tract infection, including COVID-19.
Specimen TypesNasopharyngeal swab in transport media orsalineNasopharyngeal swab in transport media
Organisms detectedVirusesAdenovirusCoronavirus 229ECoronavirus HKU1Coronavirus NL63Coronavirus OC43Same, except:No assay for Influenza A subtype H1Coronavirus 229E/HKU1/NL63/OC43combined call (coronavirus seasonal)

A table comparing the SPOTFIRE R Panel to the BioFire RP2.1 is provided in Table 2.

Table 2. Similarities and differences between the BioFire RP2.1 and the SPOTFIRE R Panel

{7}------------------------------------------------

Severe Acute Respiratory SyndromeCoronavirus 2 (SARS-CoV-2)Human MetapneumovirusHuman Rhinovirus/EnterovirusInfluenza ASubtypes: H1, H3 and H1-2009Influenza BParainfluenza Virus 1Parainfluenza Virus 2Parainfluenza Virus 3Parainfluenza Virus 4Respiratory Syncytial VirusBacteriaBordetella parapertussisBordetella pertussisChlamydia pneumoniaeMycoplasma pneumoniaeCombined Parainfluenza call for serotypes1-4
AnalytesDNA/RNASame
Technological PrinciplesHighly multiplexed nested nucleic acidamplification test with melt analysisSame
InstrumentationBioFire FilmArray 2.0 or BioFire FilmArrayTorch SystemsSPOTFIRE System (this submission)
Time to resultAbout 1 hourAbout 15 minutes
Reagent StorageRoom TemperatureSame
Test InterpretationAutomated test interpretation and reporting.User cannot access raw data.Same
ControlsTwo controls are included in each reagentpouch to control for sample processing andboth stages of PCR and melt analysis.Same
User complexityModerateLow (CLIA-waived)
Defines panel-specific parameters,instrument protocols and reportrequirements.Same
Panel Software FunctionsAnalyzes processed image data(fluorescence and temperature data) andprovides test results.Same

{8}------------------------------------------------

Summary of Performance Data

Clinical Performance

Prospective Clincal Evaluation

The clinical performance (encompassing both accuracy and ease of use) of the SPOTFIRE R Panel was established during a prospective multi-center study that was further supplemented with archived specimens. Five geographically distinct urgent care or emergency department study sites representative of the intended use setting (four in the US and one in the UK) participated in these studies from December 2020 to June 2021. All SPOTFIRE R Panel testing was performed according to the manufacturer's instructions by minimally trained operators. No hands-on training was provided to the SPOTFIRE R Panel test operators; rather, training was limited to written materials (i.e. quick reference guides) that were intended to be included with the BioFire SPOTFIRE System.

A total of 1215 NPS specimens were enrolled from consented volunteers or obtained as residual leftover specimens from subjects of all ages for the prospective clinical study; 95 of these NPS specimens were excluded. The most common reason for specimen exclusion was that a valid SPOTFIRE R Panel test was not obtained due to an invalid SPOTFIRE R Panel test. The final data set consisted of 1120 NPS specimens. Across the five study sites, 259 NPS specimens were initially collected and immediately frozen for later testing at the source study site. The remaining 861 NPS specimens were collected and tested fresh (without freezing). No difference in performance was observed when fresh and frozen specimen results were compared. Therefore, the data collected from 259 valid frozen specimens are combined with data from the valid 861 fresh specimens for all analyses.

Table 3 provides a summary of demographic information for the specimens included in the study.

CategoryProspective NPS Specimens
SexMale587 (52.4%)
Female533 (47.6%)
Age≤5 years457 (40.8%)
6-18 years258 (23.0%)
19-40 years160 (14.3%)
41-60 years147 (13.1%)
61+ years98 (8.8%)
Total1120

Table 3. Demographic Summary for Prospective SPOTFIRE R Panel Clinical Evaluation

The performance of the SPOTFIRE R Panel was evaluated by comparing the test results with those from FDAcleared multiplexed respiratory pathogen panels. The prospective study is summarized in Table 4. Positive percent agreement (PPA) for each analyte was calculated as 100% x (TP / (TP + FN)). True positive (TP) indicates that both the SPOTFIRE R Panel and the comparator method had a positive result for the specific analyte, and false negative (FN) indicates that the SPOTFIRE R Panel was negative while the comparator result was positive. Negative percent agreement (NPA) was calculated as 100% x (TN / TN + FP)). True negative (TN) indicates that both the SPOTFIRE R Panel and the comparator method had negative results, and false positive (FP) indicates that the SPOTFIRE R Panel was positive while the comparator method was negative. The exact binomial two-sided 95% confidence interval (95%Cl) was calculated. Investigations of discrepant results are summarized in the footnotes.

{9}------------------------------------------------

Table 4. SPOTFIRE R Panel Clinical Performance Summary for NPS Specimens
SPOTFIRE R Panel AnalytePositive Percent AgreementNegative Percent Agreement
TP/(TP + FN)%95%CITN/(TN + FP)%95%CI
Viruses
Adenovirusa32/3397.084.7-99.5%1058/108297.896.7-98.5%
Coronavirus SARS-CoV-2b71/7397.390.5-99.2%1031/103799.498.7-99.7%
Coronavirus (seasonal)c101/10299.094.7-99.8%1000/101398.797.8-99.2%
Human metapneumovirus1/1100-1114/111410099.7-100%
Human rhinovirus/enterovirusd345/34899.197.5-99.7%695/76790.688.3-92.5%
Influenza A virus0/0--1115/111510099.7-100%
Influenza A virus A/H1-20090/0--1115/111510099.7-100%
Influenza A virus A/H30/0--1115/111510099.7-100%
Influenza B virus0/0--1110/111010099.7-100%
Parainfluenza viruse96/9898.092.9-99.4%1006/101798.998.1-99.4%
Respiratory syncytial virusf26/2796.381.7-99.3%1086/108899.899.3-99.9%
Bacteria
Bordetella parapertussis0/0--1110/111010099.7-100%
Bordetella pertussis0/0--1115/111510099.7-100%
Chlamydia pneumoniae0/0--1115/111510099.7-100%
Mycoplasma pneumoniae0/0--1115/111510099.7-100%

Adenovirus was not detected in the single FN specimen upon SPOTFIRE R Panel retest. Adenovirus was detected in 21/24 FP specimens using an additional molecular method.

SARS-Col-2 was detected in 1/2 FN specimens upon SPOTFIRE R Panel retest. SARS-CoV-2 was detected in 2/6 FP specimens using an additional molecular method.

Seasonal coronavirus (229E/HKU1/NL63/OC43) was detected in the single FN specimen upon SPOTFIRE R Panel retest. Seasonal coronavirus (229E/HKU1/NL63/OC43) was detected in 8/13 FP specimens using an additional mothod.

4 Human rhinovirus/enterovirus was detected in 1/3 FN specimens upon SPOTFIRE R Panel retest. Human rhinovirus was detected in 48/72 FP specimens using an additional molecular method.

Parainfluenza virus was detected in both FN specimens during discrepancy investigation upon SPOTFIRE R Parainfluenza virus was detected in 9/11 FP specimens using an additional molecular method.

f Respiratory syncytial virus was detected in the single FN specimen upon SPOTFIRE R Panel retest. Respiratory syncytial virus was detected in 1/2 FP specimens using an additional molecular method.

The overall success rate for initial specimen tests was 96.7% (1158/1198). Eight tests (8/1198; 0.7%) did not complete on the initial test attempt, resulting in an instrument success rate of 99.3% (1190/1198) for initial specimen tests. Retests were not possible due to insufficient specimen volume. Of the 1190 tests that successfully produced a completed run on the initial test, 1158 had valid internal process controls. This represents a 97.3% (1158/1190) success rate for internal process controls in completed runs in the initial specimen tests.

Testing of Preselected Archived Specimens

A number of analytes on the SPOTFIRE R Panel were of low prevalence during the prospective study and were not encountered in large enough numbers to adequately demonstrate system performance. To supplement the results of the prospective clinical study, an evaluation of preselected archived retrospective NPS specimens was performed.

A total of 562 frozen archived NPS specimens were obtained from 15 external laboratories world-wide and retrospectively tested at the four US clinical sites. Of these, 542 NPS specimens had valid results that were included in performance analysis. The analyte composition of the archived specimens was confirmed using the same comparator methods as the prospective study (described above) for the analyte result to be included in the performance analysis.

The specimens were randomized such that the users performing both the confirmation and the SPOTFIRE R Panel testing were blinded to the expected test result. A summary of the available demographic information of the tested specimens is provided in Table 5, and the results of the SPOTFIRE R Panel performance for these archived specimens is shown in Table 6.

{10}------------------------------------------------

Table 5. Demographic Summary for Valid Archived NPS Specimens
CategoryArchived NPS Specimens
SexMale 254 (46.9%)
Female 185 (34.1%)
Unknown 103 (19.0%)
Age≤5 years 234 (43.2%)
6-18 years 98 (18.1%)
19-40 years 36 (6.6%)
41-60 years 35 (6.5%)
61+ years 39 (7.2%)
Unknown 100 (18.5%)
Total542

Table 6. SPOTFIRE R Panel Archived Performance Summary for NPS Specimens

Positive Percent AgreementNegative Percent Agreement
SPOTFIRE R Panel AnalyteTP/(TP + FN)%95% CITN/(TN + FP)%95% CI
Viruses
Adenovirusa31/3110089.0-100%439/45396.994.9-98.2%
Coronavirus SARS-CoV-20/0-0/0-
Coronavirus (seasonal)b95/9699.094.3-99.8%381/38898.296.3-99.1%
Human metapneumovirusc32/3397.084.7-99.5%451/45110099.2-100%
Human rhinovirus/enterovirusd29/3096.783.3-99.4%439/45496.794.6-98.0%
Influenza A viruse58/5998.391.0-99.7%423/42310099.1-100%
Influenza A virus A/H1-2009e31/3296.984.3-99.4%450/45010099.2-100%
Influenza A virus A/H327/2710087.5-100%455/45510099.2-100%
Influenza B virus30/3010088.6-100%28/2810087.9-100%
Parainfluenza virusf116/11898.394.0-99.5%359/36698.196.1-99.1%
Respiratory syncytial virusg37/3710090.6-100%440/44798.496.8-99.2%
Bacteria
Bordetella parapertussish24/2596.080.5-99.3%33/3310089.6-100%
Bordetella pertussisi27/2896.482.3-99.4%452/45699.197.8-99.7%
Chlamydia pneumoniaej30/3010088.6-100%452/45499.698.4-99.9%
Mycoplasma pneumoniaek33/3310089.6-100%446/45198.997.4-99.5%

ª Adenovirus was detected in 6/12 FP specimens during discrepancy investigation: one was detected by standard of care and five were detected using an additional mothod; two additional FP specimens were unable to be investigated.

b The single FN specimen was unable to be investigated. Seasonal coronavirus (229E/HKU1/NL63/OC43) was detected in 3/6 FP specimens during discrepancy investigation using an additional molecular method; one additional FP specimen was unable to be investigated.

° Human metapneumovirus was detected in the single FN specimen by standard of care.

d The single FN specimen was unable to be investigated. Human rhinovirus/enterovirus was detected in 4/14 FP specimens during discrepancy investigation using an additional mothod; one additional FP specimen was unable to be investigated.

· Influenza A virus A/H1-2009 was detected in the single FN specimen by standard of care.

' Parainfluenza virus was detected in the single FN specimen by standard of care; one additional FN specimen was unable to be investigated. Parainfluenza virus was detected in all seven FP specimens during discrepancy

investigation: six were detected by standard of care and one was detected using an additional mothod. 9 Respiratory syncytial virus was detected in 4/6 FP specimens during discrepancy investigation using an additional

molecular method; one additional FP specimen was unable to be investigated.

b Bordetella parapertussis was detected in the single FN specimen by standard of care.

i Bordetella pertussis was detected in the single FN specimen by standard of care. Bordetella pertussis was detected in 3/4 FP specimens by standard of care.

i Chlamydia pneumoniae was detected in both FP specimens by standard of care.

  • Mycoplasma pneumoniae was detected in all five FP specimens during discrepancy investigation: four were detected by standard of care and one was detected using an additional molecular method.

{11}------------------------------------------------

Analytical Performance Characteristics

Limit of Detection

The limit of detection (LoD) for SPOTFIRE R Panel analytes was estimated by testing dilutions of contrived samples containing known concentrations of organism in artificial nasopharyngeal swab matrix in VTM. The LoD concentrations were confirmed by testing at least 20 replicates at the estimated LoD. Confirmation of LoD required detection in at least 95% of replicates tested. The confirmed LoD concentrations for the SPOTFIRE R Panel are listed in Table 7.

AnalyteIsolateSource IDVirusesLoD Concentrationa
AdenovirusSpecies A Serotype 31ZeptoMetrix 0810073CF4.1E-03 TCID50/mL(1.0E+02 copies/mL)
Species B Serotype 3ZeptoMetrix 0810062CF8.0E-01 TCID50/mL(8.4E+02 copies/mL)
Species C Serotype 2WHO I.S. NIBSC 16-3248.2E+02 IU/mLb(8.2E+02 copies/mLb)
Species D Serotype 37ZeptoMetrix 0810119CF1.1E-02 TCID50/mL(4.5E+02 copies/mL)
Species E Serotype 4ZeptoMetrix 0810070CF1.8E-02 TCID50/mL(1.0E+04 copies/mL)
Species F Serotype 41ZeptoMetrix 0810085CF1.4E-02 TCID50/mL(1.0E+02 copies/mL)
229EATCC VR-7406.5E-01 TCID50/mL(1.1E+01 copies/mL)
HKU1Clinical Specimenc1.8E+04 copies/mL
Coronavirus (seasonal)OC43ZeptoMetrix 0810024CF1.6E-02 TCID50/mL(6.3E+01 copies/mL)
NL63ZeptoMetrix 0810228CF2.5E-03 TCID50/mL(4.7E+00 copies/mL)
USA-WA1/2020(heat inactivated)ATCC VR-1986HK1.1E-01 TCID50/mL(2.5E+02 copies/mL)
Coronavirus SARS-CoV-2A1-16 Iowa 10/2003ZeptoMetrix 0810161CF3.2E+00 TCID50/mL(2.4E+02 copies/mL)
Human metapneumovirusB1-3 Peru2-2002ZeptoMetrix 0810156CF2.5E-01 TCID50/mL(5.4E+02 copies/mL)
A2-27 Iowa A/2004ZeptoMetrix 0810164CF5.8E-01 TCID50/mL(1.8E+03 copies/mL)
B2-18 IA18-2003ZeptoMetrix 0810162CF2.0E+00 TCID50/mL(7.7E+02 copies/mL)
Human Rhinovirus 1AZeptoMetrix 0810012CFN2.1E-01 TCID50/mL(1.1E+00 copies/mL)
Human rhinovirus/enterovirusEnterovirus D68 US/MO/14-18947ATCC VR-18231.1E+01 TCID50/mL(5.4E+01 copies/mL)
Influenza A virusSubtype H1-2009Influenza A H1N1 pdmA/Michigan/45/15ZeptoMetrix 0810538CF8.2E-01 TCID50/mL(3.4E+02 copies/mL
Influenza A virusSubtype H3Influenza A H3N2A/Hong Kong/4801/14ZeptoMetrix 0810526CF8.6E-01 TCID50/mL(3.4E+02 copies/mL)
Influenza B virusB/Florida/02/06 (Victoria Lineage)ZeptoMetrix 0810037CF3.3E-02 TCID50/mL(1.6E+02 copies/mL)
B/Nevada/03/2011 (Victoria Lineage)BEI NR-440231.6E+00 CEID50/mL(4.3E+00 copies/mL)
B/Florida/04/06 (Yamagata Lineage)ZeptoMetrix 0810255CF4.0E-01 TCID50/mL(3.2E+01 copies/mL)
Parainfluenza virusSerotype 1ZeptoMetrix 0810014CF4.6E+00 TCID50/mL(1.4E+03 copies/mL)
Serotype 2ZeptoMetrix 0810015CF1.4E+01 TCID50/mL(1.6E+02 copies/mL)
Serotype 3ZeptoMetrix 0810016CF2.6E+01 TCID50/mL(6.1E+01 copies/mL)
AnalyteIsolateSource IDLoD Concentrationa
Respiratory syncytial virusSerotype 4ZeptoMetrix 0810060CF2.0E+02 TCID50/mL(1.1E+03 copies/mL)
Respiratory syncytial virusType A 2006ZeptoMetrix 0810040ACF6.2E-02 TCID50/mL(2.2E+01 copies/mL)
Respiratory syncytial virusType B 3/2015 Isolate #1ZeptoMetrix 0810479CF2.8E-02 TCID50/mL(2.4E+01 copies/mL)
Bacteria
Chlamydia pneumoniaeAR-39ATCC 535922.0E+01 IFU/mL(1.4E+02 copies/mL)
Mycoplasma pneumoniaeM129ZeptoMetrix 08015791.0E+01 CCU/mL(2.1E+03 copies/mL)
Bordetella parapertussisE595ZeptoMetrix 08014624.0E+01 CFU/mL
Bordetella pertussisA639ZeptoMetrix 08014593.3E+02 CFU/mL(3.8E+02 copies/mL)

Table 7. Limit of Detection (LoD) for the SPOTFIRE R Panel Analytes

{12}------------------------------------------------

' LoD concentration may vary from what is listed based on the accuracy and precision of the quantification method.

  • U = International Units: BioFire Diagnostics quantitative real-time PCR to demonstrate that 8,2E+02 U/ml=8,2E+02 conies/ml

· Testing for Coronavirus HKU utilized a clinicality of a cultured isolate. Viral concentration was determined in RNA copies/mL by quantitative real-time RT-PCR.

NOTE: LoD concentrations in copies/mL in Error! Reference source not found. above are based on extraction of nucleic acids from isolate cultures followed by quantitative real-time PCR (qPCR) or digital PCR (dPCR). The accuracy of concentrations may be affected by extraction efficiency, standard curve accuracy (qPCR only), assay conditions, inhibitors, and/or sequence variance. The quantification has not been compared to a reference material or other quantification methods.

NOTE: LoD concentrations of cultured viruses provided in units of TCIDso (50% Tissue Culture Infectious Dose) or CEIDso (50% Chicken Embryo Infectious Dose) are not a direct count of viral particles or nucleic acid, but an indirect measure of viral concentration based on infectivity and cytotoxicity. TCIDs/mL and CEIDso/mL will therefore vary depending on technique and methodology (including cell type, culture media and conditions, cytotoxicity of the virus, etc.). It is not appropriate to make determinations on relative sensitivity of detection for different cultures and/or different molecular assays based on LoD values measured in TCID50/mL or CEID50/mL.

Analytical Reactivity (Inclusivity)

The analytical reactivity (inclusivity) of the SPOTFIRE R Panel assessed by testing viral and bacterial isolates that represented clinical and genetic diversity and included the available phylogenetic, geographic, and temporal diversity of each SPOTFIRE R Panel analyte. Isolates were tested in triplicate at concentrations near the LoD.

A summary of analytical reactivity is included in Table 8 to Table 19.

NOTE: Influenza A assays will react variably with non-human influenza A viruses and rarely encountered human influenza A viruses that are not H1-2009 or H3; generally producing Uncertain: Influenza A virus or Influenza A virus (No Subtype Identified) results.

NOTE: The SPOTFIRE R Panel assays may react with vaccines that contain specific segments of the pathogen genome or full genome or vaccines containing attenuated/inactivated pathogen, including vaccines for SARS-CoV-2, influenza A (various subtypes), influenza B, poliovirus (human rhinovirus/enterovirus), and Bordetella pertussis. Care should be taken to minimize contamination of samples with vaccines, and clinical history of vaccine administration should be considered in the interpretation of results, particularly for vaccines administered by nasal spray.

{13}------------------------------------------------

Table 8. Summary of Reactivity to Adenovirus Isolates
SpeciesTypeSource/Isolate ID[Strain/Location/Year]Result
A12ATCC VR-863[Huie/Massachusetts]
A18ATCC VR-19[Washington DC/1954]
A31ZeptoMetrix 0810073CF-
A3ZeptoMetrix 0810062CF-
A7ATCCVR-7[Gomen/California/1954]
A7aZeptoMetrix 0810021CF-
A7d/d2UIRF[Iowa/2001]
B11ATCC VR-12[Slobitski/Massachusetts]
B14ATCC VR-15[De Wit/Netherlands/1955]
B16ATCC VR-17[CH.79/Saudi Arabia/1955]
B21ATCC VR-1833[128/Saudi Arabia/1956]
B34ATCC VR-716[Compton/1972]
B35ATCC VR-718[Holden]
B50ATCC VR-1602[Wan, RIVM no. 88-1773]Adenovirus Positive
C1ZeptoMetrix 0810050CF-
C2W.H.O NIBSC 16/324-
C5ZeptoMetrix 0810020CF-
C6ATCC VR-6[Tonsil 99]
D8ZeptoMetrix 0810069CF-
D20ZeptoMetrix 0810115CF[KB]
D37ZeptoMetrix 0810119CF-
E4ZeptoMetrix 0810070CF-
E4ATCC VR-1572[RI-67/Missouri/1952-1953]
E4aUIRF[S.Carolina/2004]
F40NCPV 0101141v-
F40ZeptoMetrix 0810084CF-
F41ATCC VR-930[Tak (73-3544)]
F41ZeptoMetrix 0810085CF[Tak]

Table 9. Summary of Reactivity to Bordetella parapertussis Isolates

Source/Isolate IDStrain/Location/YearResult
ZeptoMetrix 0801462[E595]
ATCC 9305[517]
ATCC 53892[PT28G]
ATCC 53893[PT 26/28G]
ATCC 15237[NCTC 10853]
ATCC 15311[NCTC 5952]
ATCC BAA-587[12822/Germany/1993]
ZeptoMetrix 0801461[A747]
ZeptoMetrix 0801643[C510]
ZeptoMetrix 0801644[E838]
Bordetella parapertussis Positive

Table 10. Summary of Reactivity to Bordetella pertussis Isolates

Source/Isolate IDStrain/Location/YearResult
ZeptoMetrix 0801459[A639]
ATCC 10380[10-536]
ATCC 51445[CNCTC Hp 12/63 [623]]
ATCC 8467[F]
ATCC 9340[5 [17921]Bordetella pertussis Positive
ATCC 9797[18323 [NCTC 10739]
ATCC BAA-1335[MN2531]
ATCC BAA-589[Tohama]
ZeptoMetrix 0801460[E431]

Table 11. Summary of Reactivity to Chlamydia pneumoniae Isolates

Source/Isolate IDStrain/Location/YearResult
ATCC 53592[AR-39]
ATCC VR-1310[CWL-029]
ATCC VR-1356[TWAR strain 2023]
ATCC VR-1360[CM-1]Chlamydia pneumoniaePositive
ATCC VR-1435[J-21]
ATCC VR-1452[A03]
ATCC VR-2282[TWAR strain, TW-183/Taiwan/1965]

Table 12. Summary of Reactivity to Coronavirus Isolates

TypeSource/Isolate IDStrain/Location/YearResult
NL63ZeptoMetrix 0810228CF-Coronavirus (seasonal)

{14}------------------------------------------------

TypeSource/Isolate IDStrain/Location/YearResult
229EBEI NR-470[Amsterdam/2003]Positive
ATCC VR-740-
ZeptoMetrix 0810229CF-
HKU1Clinical Specimen[Columbus OH, 2016]
Clinical Specimen[South Carolina/2010]
Clinical Specimen[France/2016]
Clinical Specimen[France/2016]
OC43ZeptoMetrix 0810024CF-
ATCC VR-759-
ATCC VR-1986HK[USA-WA1/2020]
SARS-CoV-2 aATCC VR-1991D[Hong Kong/VM20001061/2020]
ATCC VR-1992D[2019-nCoV/Italy-INMI1]
ATCC VR-1994D[Germany/BavPat1/2020]
ATCC VR-3326D[USA/CA_CDC_5574/2020]
BEI NR-52499 b[England/02/2020]
BEI NR-52501 c[Singapore/2/2020]Coronavirus SARS-CoV-2Positive
BEI NR-52503 d[USA-IL1/2020]
BEI NR-52505 e[USA-AZ1/2020]
BEI NR-52507 f[USA-CA3/2020]
BEI NR-52510 g[Chile/Santiago_op4d1/2020]
BEI NR-53518 h[New York-PV08410/2020]
LGC SeraCare AccuPlex™0505-0298 i[Omicron B.1.1.529 Variant]

a See Error! Reference source not found. for additional SARS-CoV-2 reactivity predictions based on in silico analysis.

9 The following reagent was deposited by Professor Maria Zambon and obtained through BEI Resources, NAID, NH: Genomic RNA from SARS-Related Coronavirus 2, Isolate England/02/2020, NR-52499.

· The following reagent was contributed by Duke-National University of Singapore, Programme in Emerging Infectious Diseases for distribution through BEI Resources, NIAID, NIH: Genomic RNA from SARS-Related Coronavirus 2, Isolate Singapore/2/2020, NR-52501.

9 The following reagent was deposited by the Centrol and Prevention and obtained through BEI Resources, NAID, NH: Genomic RNA from SARS-Related Coronavirus 2, Isolate USA-IL1/2020, NR-52503.

® The following reagent was deposited by the Centrol and Prevention and obtained through BEI Resources, NIAID, NH: Genomic RNA from SARS-Related Coronavirus 2, Isolate USA-AZ1/2020, NR-52505.

' The following reagent was deposited by the Centrol and Prevention and Prevention and obtained through BEI Resources, NAID, NH: Genomic RNA from SARS-Related Coronavirus 2, Isolate USA-CA3/2020, NR-52507.

9 The following reagent was obtained through BEI Resources, NAD, NH: Genomic RNA from SARS-Related Coronavirus 2, Isolate Chile/Santiago_qp4d1/2020, NR-52510.

h The following reagent was obtained through BEI Resornic RNA from SARS-Related Coronavirus 2, Isolate New York-PV08410/2020, NR-53518

' Recombinant alphavirus that contains the full SARS-Col-2 genome with mutations identified in the SARS-CoV-2 Omicron variant B.1.1.529.

SpeciesSerotypeSource/Isolate ID[Strain/Location/Year]Result
Human Rhinovirus
1ZeptoMetrix0810012CFN[1A]Human rhinovirus/enterovirusPositive
77ATCC VR-1187[130-63]
85ATCC VR-1195[50-525-CV54]
A34ATCC VR-1365[137-3]
57ATCC VR-1600[Ch47]
7ATCC VR-1601[68-CV11]
16ATCC VR-283[11757]
2ATCC VR-482[HGP]
17ATCC VR-1663[33342]
14ATCC VR-284[1059]
42ATCC VR-1950[56822]
B3ATCC VR-483[FEB]
27ATCC VR-1137[5870]
83ATCC VR-1193[Baylor 7]
Enterovirus
AEnterovirus 71ATCC VR-1432[71 H]Human rhinovirus/enterovirusPositive
Coxsackievirus 10ATCC VR-168[NY/1950]
Coxsackievirus 9ZeptoMetrix 0810017CF-
Echovirus 11ZeptoMetrix 0810023CF-
BCoxsackievirus B3ZeptoMetrix 0810074CF-
Coxsackievirus B4ZeptoMetrix 0810075CF-
Echovirus 6ZeptoMetrix 0810076CF-
Echovirus 9ZeptoMetrix 0810077CF-
CCoxsackievirus A24ATCC VR-583[DN-19/TX/1963]
Coxsackievirus A21ATCC VR-850[Kuykendall/CA/1952]
DEnterovirus D68ATCC VR-1823[US/MO/14-18947]

Table 13. Summary of Reactivity to Human Rhinovirus and Enterovirus Isolates

{15}------------------------------------------------

Table 14. Summary of Reactivity to Human Metapneumovirus Isolates
GenotypeSerotypeSource/Isolate ID[Location/Year]Result
A19ZeptoMetrix 0810160CF[lowa 3/2002]Human metapneumovirusPositive
A116ZeptoMetrix 0810161CF[lowa 10/2003]
A220ZeptoMetrix 0810163CF[lowa 14/2003]
A227ZeptoMetrix 0810164CF[lowa 27/2004]
B13ZeptoMetrix 0810156CF[Peru2-2002]
B15ZeptoMetrix 0810158CF[Peru 3/2003]
B24ZeptoMetrix 0810157CF[Peru 1/2002]
B28ZeptoMetrix 0810159CF[Peru 6/2003]
B218ZeptoMetrix 0810162CF[IA18-2003]
B2UnknownBEI NR-22232[TN/91-316]

Table 15. Summary of Reactivity to Influenza A Isolates

TypeHostSource/Isolate IDStrain/Location/YearResult
H1N1pdm09HumanZeptoMetrix 0810538CF[Michigan/45/15]Influenza A virus(Subtype H1-2009) Positive
BEI NR-19823[Netherlands/2629/2009]
BEI NR-42938[Georgia/F32551/2012]
BEI NR-44345[Hong Kong/H090-761-V1(0)/2009]
H3N2HumanZeptoMetrix 0810109CFJ[Canada/6294/2009]Influenza A virus(Subtype H3) Positive
ZeptoMetrix 0810165CF[California/07/2009]
ZeptoMetrix 0810166CF[Mexico/4108/2009]
ZeptoMetrix 0810249CF[SwineNY/03/2009]
ATCC VR-544[Hong Kong/8/1968]
H3N2HumanATCC VR-547[Aichi/2/1968]Influenza A virusPositive (No SubtypeIdentified)
ATCC VR-776[Alice]
ATCC VR-810[Port Chalmers/1/1973]
H3N2HumanATCC VR-822[Victoria/3/1975]
H3N2HumanZeptoMetrix 0810138CF[Brisbane/10/2007]
H3N2HumanZeptoMetrix 0810238CF[Texas/50/2012]
H3N2HumanZeptoMetrix 0810252CF[Wisconsin/67/2005]
H3N2HumanZeptoMetrix 0810526CF[Hong Kong/4801/14]
H1N1HumanZeptoMetrix 0810036CF[New Caledonia/20/1999]
H1N1HumanZeptoMetrix 0810036CFN[Solomon Islands/3/2006]
H1N1HumanZeptoMetrix 0810244CF[Brisbane/59/2007]
H1N1SwineATCC VR-333[Swine/Iowa/15/1930]
H1N1SwineATCC VR-897[Swine/A/New Jersey/8/76]
H1N1SwineATCC VR-99[Swine/1976/1931]
H2N2HumanBEI NR-2775 a[A/Japan/305/1957]
H5N3AvianBEI NR-9682 b[A/Duck/Singapore/645/97]Uncertain: Influenza A virus
H1N2RecombinantBEI NR-3478 c[Kilbourne F63 A/NWS/1934(HA) x A/RockefellerInstitute/5/1957 (NA)]
H10N7AvianBEI NR-2765 d

® The following reagent was obtained through BEI Resources, NIAID, NH: Genomic RNA from Influenza A Virus, A Japan/305/1957 (H2N2), NR-2775. ® The following reagent was obtained through BEI Resources, NJAD, NH: Genomic RNA from Kilbourne F181: Adusk Singapore/645/1997 (H5N3), Wild Type, NR-9682.

® The following reagent was obtained through BEI Resources, NAID, NH: Kilbourne F63: ANWS/1934 (HA) x Althute/5/1957 (NA) (HN2),

Reassortant NWS-F, NR-3478. d The following reagent was obtained through BEI Resources, NAID, NH: Genomic RNA from Influenza A Virus, NR-2765.

Table 16. Summary of Reactivity to Influenza B Isolates

LineageSource/Isolate ID[Strain/Location/Year]Result
YamagataZeptoMetrix 0810255CF[Florida/04/06]Influenza B virusPositive
ZeptoMetrix 0810239CF[2/Massachusetts/2012]
ZeptoMetrix 0810241CF[1/Wisconsin/2010]
ZeptoMetrix 0810256CF[07/Florida/2004]
ZeptoMetrix 0810037CF[B/Florida/02/06]
VictoriaBEI NR-44023[B/Nevada/03/2011]
ATCC VR-823[5/Hong Kong/1972]
CDC 2005743348[1/Ohio/2005]
ZeptoMetrix 0810258CF[2506/Malaysia/2004]
UnknownATCC VR-101[Lee/1940]
ATCC VR-102[Allen/1945]
ATCC VR-103[GL/1739/1954]
ATCC VR-295[2/Taiwan/1962]
ATCC VR-296[1/Maryland/1959]
ATCC VR-786[Brigit/Russia/1969]

Table 17. Summary of Reactivity to Mycoplasma pneumoniae Isolates

Isolate IDStrain/Location/Year

Source ZeptoMetrix 0801579 [M129] Result

{16}------------------------------------------------

Source/Isolate IDStrain/Location/YearResult
ATCC 29085[PI 1428]
ATCC 29342[M129-B7]
ATCC 15492[Mac]
ATCC 15531-TTR[FH strain of Eaton Agent [NCTC 10119]]Mycoplasma pneumoniae
ATCC 15293[M52]Positive
ATCC 15377[Bru]
ATCC 39505[Mutant 22]
ATCC 49894[UTMP-10P]

Table 18. Summary of Reactivity to Parainfluenza Virus Isolates

SerotypeSubtypeSource/Isolate IDStrain/Location/YearResult
1ZeptoMetrix 0810014CF-Parainfluenza virusPositive
1ATCC VR-94[C-35/1957]
1BEI NR-48680[FRA/29221106/2009]
2ZeptoMetrix 0810015CF-
2ATCC VR-92[Greer/1955]
3ZeptoMetrix 0810016CF-
3ATCC VR-93[C-243/1957]
3BEI NR-3233[NIH 47885 Wash/47885/57]
4AZeptoMetrix 0810060CF-
AATCC VR-1378[M-25/1958]
4BZeptoMetrix 0810060CF-
4BATCC VR-1377[CH-19503/1962]
4BZeptoMetrix 0810060BCF-

Table 19. Summary of Reactivity to Respiratory Syncytial Virus Isolates

TypeSource/Isolate IDStrain/Location/YearResult
AZeptoMetrix 0810040ACF[2006]Respiratory syncytial virusPositive
ATCC VR-26[Long/Maryland/1956]
ATCC VR-1540[A2/Melbourne/1961]
ZeptoMetrix 0810474CF[2/2015 Isolate #2]
ZeptoMetrix 0810452CF[12/2014 Isolate #2]
BZeptoMetrix 0810479CF[3/2015 Isolate #1]
ZeptoMetrix 0810040CF[Ch-93 (18)-18]
ATCC VR-1400[WV/14617/1985]
ATCC VR-955[9320/Massachusetts/1977]
ATCC VR-1580[18537/WashingtonDC/1962]
ZeptoMetrix 0810451CF[11/2014 Isolate #2]

In Silico Reactivity Predictions for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Assays

Evaluation of analytical reactivity for the SPOTFIRE R Panel SARS-CoV-2 assays (SARS-CoV2-1 and SARS-CoV2-2) was based on in silico analysis of all available sequences in the GISAID database as of December 21, 2022. In total, 11,989,970 sequences from around the globe were aligned to the assay primers.

This analysis determined that the >99.99% of 11,079,180 sequences will be detected by one or both SPOTFIRE R Panel SARS-CoV-2 assays based on homology and mismatch location with one or both sets of primers. A limitation on detection (both assays impaired) is predicted for less than 0.004% of the sequences evaluated (427 / 11,989,970) (Table 20).

The sequences evaluated include lineages and variants of concern (VOC) or variants under investigation (VUI) that may have important epidemiological, or pathogenic properties from a public health perspective, such as Delta and Omicron variants. Variants evaluated are listed in the BioFire® Respiratory Panels SARS-CoV-2 Reactivity Tech Note technical note at www.biofiredx.com/support/documents.

All lineages and variants of public health interest identified as of December 2022 are predicted to be detected; new sequences and variants will continue to be monitored for impacts on detection by the SPOTFIRE R Panel assays.

{17}------------------------------------------------

Table 20. In silico Prediction of SARS-CoV-2 Detection by SPOTFIRE R Panel Assays

+/+ indicates detected by both assays with no impairment, +/- indicates detection by one assay with no impairment and potential for impaired detection by the other assay, -1- indicates potential for impaired detection by both assays

In silico prediction
Predicted Assay ResultSARS-CoV2-1Total Sequences
Number of Sequences+-
SARS-CoV2-2+11,759,674185,74511,989,543 / 11,989,970
-44,12442799.9964%

Analytical Specificity (Cross-Reactivity & Exclusivity)

The potential for non-specific amplification and detection by the SPOTFIRE R Panel assays was evaluated by in silico analysis of available sequences and also by testing of high concentrations of on-panel and off-panel organisms. The organisms evaluated included relevant bacteria, fungi, and viruses that are either phylogenetically related to organisms detected by the SPOTFIRE R Panel or pathogenic/commensal organisms that may be present in NPS specimens. Each organism was tested in triplicate at the highest possible concentration (generally ≥1.0E+07 units/mL for bacteria and ≥1.0E+05 units/mL for viruses).

In silico analysis and testing identified a risk of SARS-CoV-2 assay cross-reactivity with a few sequences of SARSlike viruses isolated from bats and pangolin as well as intra-panel cross-reactivity with Bordetella species and Influenza A subtypes of swine origin. A summary of potential cross-reactivity is provided in Table 21. The on-panel and off-panel isolates and concentrations tested are listed in Table 22 and Table 23 respectively.

Cross-Reactive Organism/SequenceSPOTFIRE R Panel Analyte ResultDescription
Bat coronavirusPangolin coronavirusBat SARS-like coronavirusCoronavirus SARS-CoV-2aThe SARS-CoV-2 assays can amplify selectsequences from closely related sarbecovirusesisolated from bats and pangolin.
Bordetella bronchiseptica(isolates containing IS 1001 gene)Bordetella parapertussisSome strains of B. bronchiseptica containIS 1001 insertion sequences identical to thosepresent in B. parapertussis. These sequenceswill be amplified by the IS1001 assay andreported as Bordetella parapertussis.
Bordetella bronchisepticabBordetella parapertussisbBordetella pertussisHuman rhinovirus/enteroviruscThe HRV/EV assay designed to target HumanRhinovirus/Enterovirus may amplify theoxidoreductase gene from Bordetella species ( B. pertussis , B. parapertussis , and B. bronchiseptica ) when organisms are present at ahigh concentration. Cross-reactivity with B. pertussis was observed at a concentration of$≥1.3E+10$ CFU/mL.
Bovine picornavirusCanine picornavirusHuman rhinovirus/enterovirusaBovine and canine picornaviruses may bedetected and reported as Humanrhinovirus/enterovirus when present at highconcentration.
Influenza A H1N1(swine origin)Influenza A Subtype H1-2009aThe FluA-H1-2009 assay may react with the H1hemagglutinin gene sequences from viruses ofswine origin.The SPOTFIRE R Panel will report eitherInfluenza A Virus (Subtype H1-2009) or InfluenzaA Virus (No Subtype Identified), depending onthe strain and concentration in the sample.
Chlamydia gallinaceaChlamydia pneumoniaeaChlamydia gallinacea may be detected andreported as Chlamydia pneumoniae whenpresent at high concentration.

Table 21. Predicted and Observed Cross-Reactivity of the SPOTFIRE R Panel

ª Indicated cross-reactivity is predicted based on in silico analysis.

· Cross-reactivity between the HRV/EV assay and B. parapertussis is predicted based on in silico analysis but was not observed when testing organisms at the highest possible concentrations (8.3E+09 cells/mL for B. bronchiseptica and 4.6E+09 CFU/mL for B. parapertussis). · Cross-reactivity between the HRVEV assay and B. perfussis will have positive results reported for Human Rhinovirus and B. paraperussis or B. pertussis, respectivity between the HRV/EV assay and B. bronchiseptica isolates that do not cary the IS 1001 sequence will only have positive results reported for Human rhines entern the HRV/EV assay and B. bronchiseptica isolates that contain the IS 1001 sequence will have positive results reported for Human Rhinovirus and B. parapertussis.

{18}------------------------------------------------

Table 22. Summary of Results for On-Panel Organisms Tested During Evaluation of Analytical Specificity of the SPOTFIRE R Panel

OrganismIsolate IDConcentration TestedObserved Cross-Reactivity
Bacteria
Bordetella parapertussisZeptoMetrix 08014624.6E+09 CFU/mLNone
Bordetella pertussisZeptoMetrix 08014591.3E+10 CFU/mL1.3E+09 CFU/mLHuman rhinovirus/enterovirusa
Chlamydia pneumoniaeATCC 535922.9E+07 IFU/mLNone
Mycoplasma pneumoniaeZeptoMetrix 08015792.5E+07 CCU/mLNone
Viruses
Adenovirus AZeptoMetrix 0810073CF1.4E+05 TCID50/mLNone
Adenovirus BZeptoMetrix 0810062CF1.2E+07 TCID50/mLNone
Adenovirus CZeptoMetrix 0810110CF2.2E+06 TCID50/mLNone
Adenovirus DZeptoMetrix 0810119CF1.7E+05 TCID50/mLNone
Adenovirus EZeptoMetrix 0810070CF1.4E+05 TCID50/mLNone
Adenovirus FZeptoMetrix 0810085CF1.1E+06 TCID50/mLNone
Coronavirus 229EATCC VR-7408.9E+06 TCID50/mLNone
Coronavirus HKU1Clinical Specimens4.5E+07 copies/mLNone
Coronavirus NL63ZeptoMetrix 0810228CF5.0E+05 TCID50/mLNone
Coronavirus OC43ZeptoMetrix 0810024CF3.6E+05 TCID50/mLNone
Coronavirus SARS-CoV-2 (heat-inactivated)ATCC VR-1986HK7.6E+07 copies/mLNone
Enterovirus D68ATCC VR-18231.6E+07 TCID50/mLNone
Human Metapneumovirus A1ZeptoMetrix 0810161CF2.5E+05 TCID50/mLNone
Human Metapneumovirus A2ZeptoMetrix 0810164CF3.6E+05 TCID50/mLNone
Human Metapneumovirus B1ZeptoMetrix 0810156CF1.6E+04 TCID50/mLNone
Human Metapneumovirus B2ZeptoMetrix 0810162CF1.3E+06 TCID50/mLNone
Human rhinovirus A1ZeptoMetrix 0810012CFN1.3E+06 TCID50/mLNone
Influenza A H1N1pdm09ZeptoMetrix 0810538CF1.4E+05 TCID50/mLNone
Influenza A H3N2ZeptoMetrix 0810526CF7.2E+05 TCID50/mLNone
Influenza B (Victoria Lineage)BEI NR-440232.8E+08 CEID50/mLNone
Influenza B (Yamagata Lineage)ZeptoMetrix 0810256CF2.1E+04 TCID50/mLNone
Parainfluenza virus 1ZeptoMetrix 0810014CF4.2E+05 TCID50/mLNone
Parainfluenza virus 2ZeptoMetrix 0810015CF1.2E+07 TCID50/mLNone
Parainfluenza virus 3ZeptoMetrix 0810016CF3.4E+07 TCID50/mLNone
Parainfluenza virus 4ZeptoMetrix 0810060CF3.4E+07 TCID50/mLNone
Respiratory syncytial virus AZeptoMetrix 0810040ACF4.2E+05 TCID50/mLNone
Respiratory syncytial virus BZeptoMetrix 0810479CF4.2E+05 TCID50/mLNone

ª The HRV/EV assay may ampility off-target sequences (B. peries (B. periussis, B. perpertussis, and B. bronchiseptica) when present at a concentration ≥1.3E+10 CFU/mL.

Table 23. Summary of Results for Off-Panel Organisms Tested During Evaluation of Analytical Specificity of the SPOTFIRE R Panel

OrganismIsolate IDConcentration TestedObserved Cross-Reactivity
Bacteria
Arcanobacterium bernardiaeATCC BAA-4411.6E+09 cells/mLNone
Arcanobacterium haemolyticumATCC 93451.5E+08 cells/mLNone
Arcanobacterium pyogenesATCC 496986.7E+09 cells/mLNone
Bacillus cereusATCC 70648.3E+09 cells/mLNone
ATCC 105808.3E+09 cells/mLNone
ATCC 46177.9E+09 cells/mLNone
ATCC 193957.9E+09 cells/mLNone
Bordetella bronchisepticaNRRL B-599147.1E+09 cells/mLNone
NRRL B-59909a2.8E+01 cells/mLbBordetella parapertussisa
2.8E+00 cells/mLbNone
Bordetella holmesiiATCC 7000528.3E+09 cells/mLNone
Burkholderia cepaciaATCC 516717.9E+09 cells/mLNone
Campylobacter rectusATCC 332387.6E+07 cells/mLNone
Chlamydia trachomatisZeptoMetrix 08017751.3E+08 IFU/mLNone
Corynebacterium diptheriaeATCC 270108.0E+09 cells/mLNone
Corynebacterium pseudodiphtheriticumATCC 107008.7E+09 cells/mLNone
Enterococcus casseliflavusATCC 496058.0E+09 cells/mLNone
Enterococcus faecalisZeptoMetrix 08016378.0E+09 CFU/mLNone
Escherichia coliATCC BAA-21967.2E+09 cells/mLNone
Fusobacterium necrophorum ssp. funduliformeATCC 513574.4E+08 cells/mLNone
Fusobacterium nucleatumATCC 255864.9E+08 cells/mLNone
Fusobacterium variumATCC 277251.6E+08 cells/mLNone
Gemella haemolysansATCC 103794.0E+09 cells/mLNone
Gemella morbillorumATCC 278241.0E+08 cells/mLNone
Granulicatella adiacensATCC 491751.3E+09 cells/mLNone
Haemophilus influenzaeATCC 102118.3E+09 cells/mLNone
Haemophilus parahaemolyticusATCC 497008.7E+09 cells/mLNone
Klebsiella pneumoniaeCDC AR#01157.3E+09 CFU/mLNone

{19}------------------------------------------------

OrganismIsolate IDConcentration TestedObserved Cross-Reactivity
Lactobacillus rhamnosusATCC 74697.9E+09 cells/mLNone
Lactococcus lactisATCC 291466.2E+09 cells/mLNone
Legionella pneumophilaATCC 332157.0E+09 cells/mLNone
Leptotrichia buccalisATCC 142014.4E+08 cells/mLNone
Moraxella catarrhalisATCC 436277.2E+09 cells/mLNone
Mycobacterium tuberculosisZeptoMetrix 08016606.1E+06 CFU/mLNone
Mycoplasma buccaleMycoplasma ExperienceNC101361.4E+07 CFU/mLNone
Mycoplasma fauciumMycoplasma ExperienceNC101741.4E+06 CFU/mLNone
Mycoplasma fermentansMycoplasma ExperienceNC101172.8E+07 CFU/mLNone
Mycoplasma genitaliumMycoplasma ExperienceNC101951.8E+06 CFU/mLNone
Mycoplasma hominisMycoplasma ExperienceNC101111.2E+07 CFU/mlNone
Mycoplasma lipophilumMycoplasma ExperienceNC101731.5E+06 CFU/mLNone
Mycoplasma oraleMycoplasma ExperienceNC101122.2E+07 CFU/mLNone
Mycoplasma salivariumMycoplasma ExperienceNC101134.4E+06 CFU/mLNone
Neisseria elongataATCC 252958.5E+09 cells/mLNone
Neisseria gonorrhoeaeZeptoMetrix 08014824.9E+07 CFU/mLNone
Neisseria lactamicaATCC 239712.7E+09 cells/mLNone
Neisseria meningitidisATCC 131137.4E+09 cells/mLNone
Neisseria siccaATCC 99137.2E+09 cells/mLNone
Neisseria subflavaATCC 492758.0E+09 cells/mLNone
Parvimonas micraATCC 332706.0E+07 cells/mLNone
Pneumocystis cariniiATCC PRA-1591.0E+07 nuclei/mLNone
Porphyromonas endodontalisATCC 354061.6E+07 cells/mLNone
Porphyromonas gingivalisATCC BAA-3085.0E+08 cells/mLNone
Prevotella histicolaBEI HM-4719.0E+08 cell/mLNone
Prevotella melaninogenicaATCC 258456.9E+08 cells/mLNone
Prevotella oralisATCC 333226.2E+08 cells/mLNone
Pseudomonas aeruginosaCDC AR#00928.3E+09 cells/mLNone
Rhodococcus equiATCC 337067.3E+09 cells/mLNone
Serratia marcescensATCC 271378.9E+09 cells/mLNone
Staphylococcus aureusATCC BAA-17007.4E+09 cells/mLNone
Staphylococcus epidermidisATCC 122288.0E+09 cells/mLNone
Staphylococcus haemolyticusATCC 299688.0E+09 cells/mLNone
Staphylococcus intermediusATCC 296638.2E+09 cells/mLNone
Staphylococcus saprophyticusATCC 153058.1E+09 cells/mLNone
Streptococcus agalactiaeATCC 138136.0E+09 cells/mLNone
Streptococcus anginosusATCC 7002317.1E+09 cells/mLNone
Streptococcus constellatus ssp. pharyngisNCTC 131225.6E+08 cells/mLNone
Streptococcus dysgalactiae ssp. dysgalactiaeATCC 430786.7E+09 cells/mLNone
NCTC 46697.4E+09 cells/mLNone
NCTC 43358.4E+09 cells/mLNone
NCTC 46706.6E+09 cells/mLNone
CCUG 276657.4E+09 cells/mLNone
CCUG 281126.7E+09 cells/mLNone
CCUG 281147.5E+09 cells/mLNone
Streptococcus dysgalactiae ssp. equisimilis(isolated from human)ZeptoMetrix 08015167.8E+08 CFU/mLNone
Streptococcus dysgalactiae ssp. equisimilis(isolated from pig)CCUG 281177.1E+09 cells/mLNone
Streptococcus dysgalactiae ssp. equisimilis(isolated from horse)ATCC 100096.9E+09 cells/mLNone
Streptococcus gallolyticusATCC 431432.8E+09 cells/mLNone
Streptococcus gordoniiATCC 105584.5E+09 cells/mLNone
Streptococcus intermediusATCC 273352.9E+09 cells/mLNone
Streptococcus mitisATCC 159143.2E+09 cells/mLNone
Streptococcus mutansATCC 251752.3E+09 cells/mLNone
Streptococcus oralisATCC 105571.1E+09 cells/mLNone
Streptococcus parasanguinisATCC 159127.8E+09 cells/mLNone
Streptococcus pneumoniaeATCC 496192.5E+08 cells/mLNone
Streptococcus pyogenesStreptococcus pyogenesATCC 12344ATCC 123443.4E+08 cells/mL3.4E+08 cells/mLNoneNone
Streptococcus salivariusATCC 134196.6E+09 cells/mLNone
Streptococcus sanguinisATCC 105561.1E+09 cells/mLNone
Tannerella forsythiaATCC BAA-27172.6E+08 cells/mLNone

{20}------------------------------------------------

OrganismIsolate IDConcentration TestedObserved Cross-Reactivity
Treponema denticolaATCC 335202.2E+08 cells/mLNone
Ureaplasma urealyticumATCC 276185.7E+07 cells/mLNone
Veillonella parvulaATCC 107904.7E+08 cells/mLNone
Fungi
Candida albicansATCC MYA-28762.8E+08 cells/mLNone
Saccharomyces cerevisiaeATCC 188241.9E+08 cells/mLNone
Viruses
CytomegalovirusZeptoMetrix 0810003CF1.9E+05 TCID50/mLNone
Epstein-Barr virusZeptoMetrix 0810008CF5.9E+06 copies/mLNone
Human herpes simplex virus 1ATCC VR-2608.9E+06 TCID50/mLNone
Measles virusZeptoMetrix 0810025CF2.5E+05 TCID50/mLNone
Middle East respiratory syndrome coronavirus(heat-inactivated)ZeptoMetrix 0810575CFHI1.2E+05 TCID50/mLNone
Mumps virusZeptoMetrix 0810079CF2.0E+06 TCID50/mLNone
Severe acute respiratory syndrome coronavirus(purified genomic RNA)BEI NR-523465.3E+05 genomes/mLNone

Bordelella bronchiseptica strain NRRL B-5999 contains genomic insertions of the IS 1001 gene and is reactive with the SPOTFIRE R Panel Bordelela parapertussis IS1001 assay .

Positive results were observed for Bordeles in all regicates tested at concentrations ≥2.8E+01 cells/ml, and one out of three regionates tested at 2 8F+00 cells/ml

· Parvimonas micra was formerly classified as Micromonas micros and Peptostreptococcus micros.

Near-LoD/Reproducibility Evaluation

A near-LoD/reproducibility evaluation was performed to demonstrate that the SPOTFIRE R Panel could reproducibly provide accurate results for weak-positive samples when used by minimally trained operators. Contrived samples were tested at three of the prospective clinical study sites and additionally on three unique SPOTFIRE Systems at BioFire Diagnostics (BioFire) by trained BioFire personnel. The contrived samples contained combinations of SPOTFIRE R Panel analytes prepared in artificial matrix at or near (1x to 3x) the LoD. For testing performed at clinical sites, samples were testing events (non-consecutive days) by two operators during the course of their normal workday routine. Each site was equipped with a single SPOTFIRE System. Testing at all three sites was performed with a single reagent lot. For each testing event, each operator ran two replicate pouches for a total of 20 replicates per site and 60 total replicates across all three sites. For testing performed at BioFire, samples were tested over five consecutive days, by two operators per system, using three different reagent lots. Each day of testing, the two operators each tested three replicates on each system for a total of 30 replicates per system and 90 total replicates across all systems. When combined, each analyte was tested in a total of 150 replicates by at least 12 different operators across six different SPOTFIRE Systems.

A summary of results (percent (%) agreement with the expected positive result) for each analyte (by site and system) is provided in Table 24. The SPOTFIRE R Panel reported the expected positive results for panel analytes in 94% -100% of samples and the expected negative results for all analytes in 100% of samples. Comparison of the positive percent agreement between user groups (98.8% for trained operators at BioFire versus 98.9% for minimally trained operators) demonstrates that the accuracy of the SPOTFIRE R Panel is not dependent upon the specific expertise of the user.

{21}------------------------------------------------

SpotFire System testingAll Sites/Systems
AnalyteIsolate(Source ID)ConcentrationTested(test level)ExpectedResultBioFire DxClinical[95%ConfidenceInterval]
System ASystem BSystem CTotalSite 1Site 2Site 3Total
AdenovirusSpecies B(ZeptoMetrix 0810062CF)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
2.4E+00 TCID50/mL(3× LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)150/150100%[97.6-100%]
Bordetella parapertussis(ZeptoMetrix 0801462)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
1.2E+02 CFU/mL(3× LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)150/150100%[97.6-100%]
Bordetella pertussis(ZeptoMetrix 0801459)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
9.9E+02 CFU/mL(3× LoD)Positive29/30(96.7%)29/30(96.7%)30/30(100%)88/90(97.8%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)148/15098.7%[95.3-99.8%]
Chlamydia pneumoniae(ATCC 53592)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
2.0E+01 IFU/mL(1× LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)150/150100%[97.6-100%]
No AnalyteNegative60/60(100%)60/60(100%)60/60(100%)180/180(100%)40/40(100%)40/40(100%)40/40(100%)120/120(100%)300/300100%[98.8-100%]
229E(ATCC VR-740)2.0E+00 TCID50/mL(3× LoD)Positive30/30(100%)30/30(100%)29/30(96.7%)89/90(98.9%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)149/15099.3%[96.3-100%]
Coronavirus (seasonal)OC43(ZeptoMetrix0810024CF)1.6E-02 TCID50/mL(1× LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)150/150100%[97.6-100%]
NL63(ZeptoMetrix0810228CF)2.5E-03 TCID50/mL(1× LoD)Positive27/30(90.0%)30/30(100%)30/30(100%)87/90(96.7%)18/20(90.0%)20/20(100%)18/20(90.0%)56/60(93.3%)143/15095.3%[90.6-98.1%]
Coronavirus SARS-CoV-2Severe Acute RespiratoryNo AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
Syndrome Coronavirus 2(ATCC VR-1986HK)2.5E+02 copies/mL(1× LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)150/150100%[97.6-100%]
SpotFire System testingClinicalAll Sites /Systems [95% Confidence Interval]
Analyte Isolate(Source ID)Concentration Tested(test level)Expected ResultSystem ASystem BSystem CTotalSite 1Site 2Site 3Total
Human metapneumovirus3 Type B1(ZeptoMetrix 0810156CF)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
7.5E-01 TCID50/mL(3x LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)19/20(95.0%)20/20(100%)19/20(95.0%)58/60(96.7%)148/15098.7%[95.3-99.8%]
Human rhinovirus/enterovirusEnterovirus D68US/MO/14-18947(ATCC VR-1823)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
1.1E+01 TCID50/mL(1x LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)150/150100%[97.6-100%]
Influenza A virusNo AnalyteNegative90/90(100%)90/90(100%)90/90(100%)270/270(100%)60/60(100%)60/60(100%)60/60(100%)180/180(100%)450/450100%[99.2-100%]
Influenza A H1N1pdm(ZeptoMetrix 0810538CF)2.5E+00 TCID50/mL(3x LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)150/150100%[97.6-100%]
Influenza A H3N2(ZeptoMetrix 0810526CF)2.6E+00 TCID50/mL(3x LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)19/20(95.0%)19/20(95.0%)20/20(100%)58/60(96.7%)148/15098.7%[95.3-99.8%]
Influenza A virus A/H1-2009Influenza A H1N1pdm(ZeptoMetrix 0810538CF)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
2.5E+00 TCID50/mL(3x LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)150/150100%[97.6-100%]
Influenza A virus A/H3Influenza A H3N2(ZeptoMetrix 0810526CF)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
2.6E+00 TCID50/mL(3x LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)19/20(95.0%)19/20(95.0%)20/20(100%)58/60(96.7%)148/15098.7%[95.3-99.8%]
Influenza B virus(ZeptoMetrix 0810037CF)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
9.9E-02 TCID50/mL(3x LoD)Positive29/30(96.7%)30/30(100%)30/30(100%)89/90(98.9%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)149/15099.3%[96.3-100%]
Mycoplasma pneumoniae(ZeptoMetrix 0801579)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
AnalyteIsolate(Source ID)ConcentrationTested(test level)ExpectedResultSpotFire System testingClinicalAll Sites/Systems[95%ConfidenceInterval]
System ASystem BSystem CTotalSite 1Site 2Site 3Total
Parainfluenza virus1.0E+01 CCU/mL(1x LoD)Positive29/30(96.7%)30/30(100%)30/30(100%)89/90(99.3%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)149/15099.3%[96.3-100%]
No AnalyteNegative30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)150/150100%[97.6-100%]
Parainfluenza virus 1(ZeptoMetrix0810014CF)4.6E+00 TCID50/mL(1x LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)150/150100%[97.6-100%]
Parainfluenza virus 2(ZeptoMetrix0810015CF)4.2E+01 TCID50/mL(3x LoD)Positive30/30(100%)29/30(96.7%)28/30(93.3%)87/90(96.7%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)147/15098.0%[94.3-99.6%]
Parainfluenza virus 3(ZeptoMetrix0810016CF)8.8E+00 TCID50/mL(1x LoD)Positive28/30(93.3%)30/30(100%)30/30(100%)88/90(97.8%)19/20(95.0%)18/20(90.0%)20/20(100%)57/60(95.0%)145/15096.7%[92.4-98.9%]
Parainfluenza virus 4(ZeptoMetrix0810060CF)2.0E+02 TCID50/mL(1x LoD)Positive29/30(96.7%)30/30(100%)30/30(100%)89/90(98.9%)20/20(100%)20/20(100%)20/20(100%)60/60(100%)149/15099.3%[96.3-100%]
Respiratory syncytial virus(ZeptoMetrix 0810040ACF)No AnalyteNegative120/120(100%)120/120(100%)120/120(100%)360/360(100%)80/80(100%)80/80(100%)80/80(100%)240/240(100%)600/600100%[99.4-100%]
6.2E-02 TCID50/mL(1x LoD)Positive30/30(100%)30/30(100%)30/30(100%)90/90(100%)20/20(100%)19/20(95.0%)20/20(100%)59/60(98.3%)149/15099.3%[96.3-100%]
Total positive agreement (%)by system/user group621/63098.6%628/63099.7%627/63099.5%1876/189099.3%414/42098.6%415/42098.8%417/42099.3%1246/126098.9%3122/3150
Overall positive agreement (%)[95% Confidence Interval]99.1%[98.7-99.4%]

Table 24. Reproducibility of Results on the SPOTFIRE R Panel and BIOFIRE SPOTFIRE System

{22}------------------------------------------------

{23}------------------------------------------------

{24}------------------------------------------------

Interference

Potentially interfering substances that could be present in NPS specimens or may be introduced into specimens during collection or subsequent handling and testing were evaluated for their effect on SPOTFIRE R Panel performance. The substances tested included endogenous substances that may be found at normal or elevated levels in clinical specimens (e.g. blood, mucus/mucin, human genomic DNA), various commensal or infectious microorganisms, medications, washes or topical applications for the nasal passage, various swabs and transport media for specimen collection, and substances used to clean, decontaminate, or disinfect work areas.

Each substance was added to contrived samples containing representative organisms at concentrations near (3x) the LoD. The concentration of substance added to the samples was equal to or greater than the highest level expected to be in NPS specimens.

Valid and accurate results were obtained for each sample containing substances and microorganisms at the concentrations listed in Table 25.

Substance TestedConcentration Tested
Endogenous Substances
Human Whole Blood (with Na Citrate)10% (v/v)
Human Sputum/Mucus1% (v/v)
Human Genomic DNA20 ng/µL
Exogenous Substancesa
Promethazine hydrochloride1.04 µmol/L (3.34E-04 mg/mL)
Acetaminophen (paracetamol)1.0E+03 µmol/L (1.5E-01 mg/mL)
Acetylsalicylic acid (Aspirin)167 µmol/L (3.0E-02 mg/mL)
Ibuprofen1060 µmol/L (2.2E-01 mg/mL)
Albuterol sulfate (common ingredient in rescue inhalers)0.188 µmol/L (5.4E-05 mg/mL)
Triple antibiotic ointment (neomycin/polymyxin B/bacitracin)2% w/v
Mucinex® Severe Nasal Congestion Relief Clear & Cool Nasal Spray (Oxymetazoline hydrochloride 0.05%)1% w/v
Saline nasal spray (sodium chloride 0.65%, disodium phosphate, phenylcarbinol, monosodium phosphate andbenzalkonium chloride solution)1% w/v
Vicks® VapoRub® Cough Suppressant Topical Analgesic (Camphor 4.8%, eucalyptus oil 1.2%, and menthol 2.6%)1% w/v
Vaseline® Petroleum Jelly (100% white petrolatum)1% w/v
Orajel™ (benzalkonium chloride 0.13%, benzocaine 20%, menthol 0.5%, zinc chloride 0.15%)2% w/v
Chloraseptic® Sore Throat Spray (Phenol 1.4%)1% v/v
Vicks® Formula 44™ DM (dextromethorphan hydrobromide 0.67 mg/mL, guaifenesin 13 mg/mL) (cough syrup)1% v/v
Phenylephrine hydrochloride (common ingredient in nasal decongestants)1% w/v
Nasal spray (fluticasone propionate 50 mcg)1% v/v
Sucrets® Sore Throat (dyclonine hydrochloride 2.0 mg/lozenge)1% w/v
Benadryl® Allergy Liqui-gels® (diphenhydramine hydrochloride 25 mg/capsule)1% v/v
Zicam® Cold Remedy (Galphimia Glauca 4x, Luffa Operculata 4x, Sabadilla 4x)1% v/v
Cold-eeze® (zinc gluconate 2.3%)1% w/v
HALLS lozenge (menthol 5 mg/lozenge)1% w/v
Listerine® Cool Mint® (menthol 0.042%, thymol 0.064%, methyl salicylate 0.06%, eucalyptol 0.092%)6.5% v/v
Copenhagen® Snuff (Tobacco)1% w/v
JUICE HEAD (30% propylene glycol, 70% vegetable glycerin) (e-juice)1% v/v
Technique-Specific Substances
Rayon swab (COPAN Diagnostics Inc.)1 swab
Nylon flocked swab (COPAN Diagnostics Inc.)1 swab
Polyester swab (COPAN Diagnostics Inc.)1 swab
Calcium Alginate swab (Puritan®)1 swab
Cary-Blair90% v/v
Dulbecco's Modified Eagles-Medium (DMEM)90% v/v
Hanks Balanced Salt Solution100% v/v
0.9% Normal Saline100% v/v
BD™ Universal Viral Transport100% v/v
Remel MicroTest™ M4RT Tube w/o beads100% v/v
Remel MicroTest™ M4 Tube w/o beads100% v/v
Viral Preservative Media (VPM)90% v/v
Phosphate Buffered Saline (PBS)90% v/v
PrimeStore® MTM Molecular Transport Media90% v/v
Stuart Transport Medium90% v/v
eNAT™ Molecular Transport Medium90% v/v
Bleach1% v/v, 2% v/vb
Substance TestedConcentration Tested
Ethanol7% v/v
Disinfecting wipes (ammonium chloride)0.25 - 0.5 inch square/sample
DNAZapTM1% v/v
RNaseZapTM1% v/v
Competing Microorganisms
On-Panel
Coronavirus 229E1.5E+07 copies/mL
Enterovirus D687.8 E+07 copies/mL
Parainfluenza virus 38.0E+06 copies/mL
Respiratory syncytial virus A1.5E+07 copies/mL
Adenovirus A311.6E+07 copies/mL
Bordetella pertussis1.6E+09 copies/mL
Off-Panel
Cytomegalovirus (CMV)4.2E+04 TCID50/mL
Herpes simplex virus 19.0E+06 TCID50/mL
Staphylococcus aureus7.4E+08 CFU/mL
Streptococcus pneumoniae2.5E+07 CFU/mL
Streptococcus pyogenes2.2E+08 copies/mL
Haemophilus influenzae8.3E+08 CFU/mL
Candida albicans2.8E+07 CFU/mL

Table 25. Substances Tested on the SPOTFIRE R Panel - No Interference Observed

{25}------------------------------------------------

4 Nasal influenza vacines (e.g. FluMis®) were not evaluated but are predicted to be reactive with the Influenza B assays

b houbation of sample with 1% (vi) bleach for 15 minutes, 4 hours, or ~18.5-hour (overnight) or 2% (vir) bleach for 15 minutes aid not result in interference

NOTE: Avoid contact between samples and bleach prior to testing (bleach can damage nucleic acids and prevent amplification and detection by the panel).

NOTE: Compatibility of the SPOTFIRE R Panel with NPS in PrimeStore® MTM has not been evaluated in the intended use setting. PrimeStore® MTM and BIOFIRE Sample Buffer contain guanidine salts that will react with bleach to form a toxic gas. Use caution if using bleach for disinfection purposes when collecting or testing NPS specimens.

External Control Material

Quality Control Testing

External control material is available directly from bioMérieux and Maine Molecular Quality Controls, Inc. (MMQC) (online at www.mmqci.com). The MMQCl controls listed in Table 26 are positive and negative control solutions for all viruses and bacteria detected by the SPOTFIRE R Panel, along with an insert explaining the expected results.

VendorProduct NamePart Number
Maine Molecular QualityControls, Inc. (MMQCI)SPOTFIRE® RSPNegative ControlM42738
Maine Molecular QualityControls, Inc. (MMQCI)SPOTFIRE® RSPPositive ControlM42638

Table 26. Control Materials Compatible with all Analytes on the SPOTFIRE R Panel

O NOTE: MMQCI control test solutions can be stored at room temperature (18°C-25°C).

It is recommended that when using the SPOTFIRE R Panel that Positive QC testing should be performed when:

  • Training a new operator .
  • Receiving a new shipment or lot of SPOTFIRE R Panel test kits .
  • Receiving a new SpotFire Control Station

A Negative QC test should be run at least monthly for each testing location to monitor for environmental contamination. Additional Negative QC tests may be run if contamination is suspected.

§ 866.3981 Device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test.

(a)
Identification. A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
e.g., BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
e.g., when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
e.g., regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
e.g., saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
i.e., H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.