K Number
K213604

Validate with FDA (Live)

Date Cleared
2022-04-03

(139 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ELAZ Nitrile Powder Free Examination Glove is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

ELAZ Nitrile Powder Free Examination Glove is non-sterile, single use only, disposable, powder free examination glove. The glove is made of Acrylonitrile-butadiene rubber. The glove is designed to meets the specification of ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

AI/ML Overview

This document describes the performance testing for the ELAZ Nitrile Powder Free Examination Glove, a Class I medical device. The information provided is for a traditional 510(k) submission, confirming substantial equivalence to a predicate device.

It's important to note that this is a medical glove, not a software or AI-driven diagnostic device. Therefore, many of the typical acceptance criteria and study designs for AI/software-based medical devices (like MRMC studies, expert ground truth adjudication, training/test set provenance for AI, etc.) are not applicable here. The performance evaluation focuses on physical properties, biocompatibility, and compliance with established ASTM and ISO standards for examination gloves.

Here's a breakdown of the provided information, framed within the context of a medical glove substantial equivalence submission:


1. Table of Acceptance Criteria and Reported Device Performance

The full table from the document (page 6-7) is replicated here, summarizing the acceptance criteria and the test results for the ELAZ Nitrile Powder Free Examination Glove.

Test MethodologyPurposeAcceptance CriteriaResults
ASTM D6319-19: Physical Dimensions Test (mm)
Length (mm)S ≥ 220, M ≥ 230, L ≥ 230, XL ≥ 230S ≥ 220, M ≥ 230, L ≥ 230, XL ≥ 230 (All Pass)
Width (mm)S: 80±10, M: 95±10, L: 110±10, XL: 120±10S: 87-89, M: 99-101, L: 110-113, XL: 112-115 (All Pass)
Finger Thickness (mm)≥ 0.05S ≥ 0.16, M ≥ 0.15, L ≥ 0.15, XL ≥ 0.13 (All Pass)
Palm Thickness (mm)≥ 0.05S ≥ 0.11, M ≥ 0.11, L ≥ 0.11, XL ≥ 0.10 (All Pass)
ASTM D5151-19Testing for Freedom from holesAQL 2.5Gloves pass AQL 2.5
ASTM D6124-17Determine the powder residue for powder free gloves< 2 mg/gloveAverage 0.43 mg/glove
ASTM D412, ASTM D573Testing for Physical property characteristicsBefore Aging: Tensile strength ≥14 MPa, Ultimate Elongation ≥500% After Aging: Tensile strength ≥14 MPa, Ultimate Elongation ≥400%Before Aging: Tensile strength ≥21 MPa, Ultimate Elongation ≥502% After Aging: Tensile strength ≥18 MPa, Ultimate Elongation ≥465% (All Pass)
ISO 10993-5Tests for In vitro cytotoxicityUnder the conditions of the study non-cytotoxic.Under conditions of the study, it was considered "non-cytotoxic" at 25%, 12.5%, and 6.25% and demonstrate cytotoxicity at the 50% and 100% of the test item extract. (Meets non-cytotoxic criteria at relevant concentrations)
ISO 10993-10Evaluate the endpoint of irritant for biocompatibilityUnder the conditions of the study, not an irritant.Under the conditions of the study, not an irritant.
ISO 10993-10Evaluate the endpoint of sensitization for biocompatibilityUnder the conditions of the study, not a sensitizer.Under the conditions of the study, not a sensitizer.
ISO 10993-11Tests for systemic toxicityUnder the conditions of the study, the device extracts do not pose a systemic toxicity concernUnder the conditions of the study, the device extracts do not pose a systemic toxicity concern

Regarding the other questions, which are highly relevant for AI/software medical devices but not for this traditional glove submission:

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • For physical and performance tests, sample sizes are typically determined by the specific ASTM standard (e.g., AQL levels for freedom from holes, number of samples for tensile strength). These are standardized industry practices, not "data provenance" in the sense of patient data.
  • The biocompatibility tests would involve animal models (for in vivo tests like irritation, sensitization, systemic toxicity) or cell cultures (for in vitro cytotoxicity). The document does not specify the exact sample sizes for each test, but they would follow the requirements of the respective ISO standards. This is not "patient data."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. Ground truth for medical gloves is established by objective physical measurements and standardized biological tests according to the protocols defined in ASTM and ISO standards. There are no "experts" establishing ground truth in the sense of clinical image review.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not applicable. This applies to clinical interpretation studies, typically for AI/software devices.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This relates to AI/software performance in diagnostic tasks and human-in-the-loop studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This refers to the performance of an AI algorithm. The device is a physical glove.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for medical gloves is defined by the validated and universally accepted specifications and test methods outlined in the ASTM and ISO standards (e.g., a hole is detected if water leaks, tensile strength is measured by a machine, biological response is quantified by established cytotoxicity assays). It is not clinical ground truth from patient outcomes or expert consensus.

8. The sample size for the training set

  • Not applicable. There is no "training set" in the context of a physical medical device. Manufacturing processes are continually monitored for quality control, but this isn't machine learning training data.

9. How the ground truth for the training set was established

  • Not applicable.

In summary, this document presents a standard premarket notification for a Class I medical glove. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" are entirely based on established engineering and biomedical testing standards (ASTM and ISO), not on clinical or AI performance metrics. The device successfully met all specified acceptance criteria derived from these standards, demonstrating its substantial equivalence to legally marketed predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

April 3, 2022

Protect Gloves Co., Ltd. % Paweena U-Thainual CEO MDR Solutions Co., Ltd. 1435 Kanchanapisek Rd., Bang Khae Nuea Bangkok, Bang Khae 10160 Thailand

Re: K213604

Trade/Device Name: ELAZ Nitrile Powder Free Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: March 21, 2022 Received: March 24, 2022

Dear Paweena U-Thainual:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213604

Device Name

ELAZ Nitrile Powder Free Examination Glove

Indications for Use (Describe)

ELAZ Nitrile Powder Free Examination Glove is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K213604

510(k) Summary

1.General Information
Applicant/Submitter:Protect Gloves Company Limited
Address:60/138 Moo 3 Mabyangporn Pluakdang
Rayong, 21140 THAILAND
Tel: +66-2-384-3049
Email: nathaphon@me.com
Contact Person:Paweena U-Thainual, PhD
Address:MDR Solutions, Co., Ltd.
1435 Kanchanapisek Rd., Bang Khae Nuea
Bang Khae, Bangkok 10160 THAILAND
Email: paweena@mdrsolutions.co.th
Preparation Date:November 8, 2021

2. Device Name and Code

Device Trade Name:ELAZ Nitrile Powder Free Examination Glove
Common Name:Nitrile Patient Examination Glove
Classification Name:Non-Powdered Patient Examination Glove
Product Code:LZA
Regulation Number:21 CFR 882.6250
Classification:I
Review Panel:General Hospital

Predicate Device 3.

Provided below is the legally marketed predicate device.

Table 1 Primary Predicate device

ApplicantDevice Name510(k) Number
Tangshan LanhaiMedical Supplies Co.,Ltd.Disposable NitrileExamination Gloves (Powderfree, Purple-Blue, Blue)K210898

{4}------------------------------------------------

4. Device Description

ELAZ Nitrile Powder Free Examination Glove is non-sterile, single use only, disposable, powder free examination glove. The glove is made of Acrylonitrile-butadiene rubber. The glove is designed to meets the specification of ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

Indications / Intended Use 5.

ELAZ Nitrile Powder Free Examination Glove is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Technological Characteristics Comparison: 6.

ELAZ Nitrile Powder Free Examination Glove is compared to the legally marketed predicate device.

Proposed Subject DevicePredicate DeviceComparison
ManufacturerProtect Gloves Co., Ltd.Tangshan Lanhai Medical Supplies Co., Ltd.N/A
Trade NameELAZ Nitrile Powder Free Examination GloveDisposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue)N/A
510(k) NumberK213604K210898N/A
ClassificationClass IClass IYES
Product CodeLZALZAYES
Regulation Number880.6250880.6250YES
Intended Use/ Indications For useELAZ Nitrile Powder Free Examination Glove is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examinerThe DISPOSABLE NITRILE EXAMINATION GLOVES (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.YES
Powdered FreeYesYesYes

Page 2 of 5

{5}------------------------------------------------

Device Dimensions Comparison
SizeS, M, L, XLXS, S, M, L, XLYES*
ColorBluePurple-Blue, BlueYES
SterilityNon-SterileNon-SterileYES
SMLXLXSSMLXLYES
Length (mm)≥ 220≥ 230≥ 230≥ 230≥ 220≥ 220≥ 230≥ 230≥ 230
Width (mm)SMLXLXSSMLXLYES
80±1095±10110±10120±1070±1080±1095±10110±10120±10
FingerThickness(mm)≥ 0.05≥ 0.05YES
PalmThickness(mm)≥ 0.05≥ 0.05YES
Performance and Physical Comparison
Before AgingAfter AgingBefore AgingAfter Aging
TensileStrength(MPa)≥ 14≥ 14≥ 14≥ 14YES
UltimateElongation(%)Before AgingAfter AgingBefore AgingAfter AgingYES
≥ 500≥ 400≥ 500≥ 400
Single UseYesYesYES
Be free from holes when tested inBe free from holes when tested in
Freedomfrom holeaccordance with ASTM D5151AQL 2.5accordance with ASTM D5151AQL 2.5YES
PowderContentPowder residue ≤ 2.0 mgPowder residue ≤ 2.0 mgYES
Comply withASTM D6319YesYesYES
MaterialUsedNitrileNitrileYES
Shelf life3 yearsN/AN/A
Safety Comparison - Biocompatibility Test
IrritationUnder the conditions of the study,not an irritantUnder the conditions of the study, notan irritantYES
SensitizationUnder conditions of the study, not asensitizerUnder conditions of the study, not asensitizerYES
CytotoxicityUnder conditions of the study, itwas considered as "non-cytotoxic"at 25%, 12.5%, and 6.25% anddemonstrate cytotoxicity at the 50%and 100% of the test item extract.N/AN/A
AcuteSystemicToxicityUnder the conditions of the study,there was no evidence of systemictoxicity from the extractYES

{6}------------------------------------------------

  • When compared to the same size, the dimensions, and physical design are identical.

Performance Data 7.

Non-clinical tests:

Bench tests were conducted according to the FDA-recognized consensus standard, to verify that the proposed device met all design specifications. The proposed device complies with the following standards:

  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical । Application.
  • ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. -
  • -ASTM D6124-17, Standard Test Method for Residual Powder on Medical Gloves.

Biocompatibility tests were conducted according to the FDA-recognized consensus standard, to verify the safety of the device.

  • ISO 10993-5:2009, Biological Evaluation of Medical Devices - Part 5: Tests for In vitro cytotoxicity.
  • ISO 10993-10:2010, Biological Evaluation of Medical Devices Part 10: Tests for irritation and skin sensitization.
  • -ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
TestMethodologyPurposeAcceptance CriteriaResults
ASTM D6319-19PhysicalDimensionsTest (mm)Length (mm)Length (mm)
S ≥ 220S ≥ 220
M ≥ 230M ≥ 230
L ≥ 230L ≥ 230
XL ≥ 230XL ≥ 230
Width (mm)Width (mm)
S: 80±10S: 87-89
M: 95±10M: 99-101
L: 110±10L: 110-113
XL: 120±10XL: 112-115
Finger Thickness (mm)Finger Thickness (mm)
≥0.05S ≥ 0.16
M ≥ 0.15
L ≥ 0.15
XL ≥ 0.13
Palm Thickness (mm)Palm Thickness (mm)
>0.05S > 0.11

{7}------------------------------------------------

M ≥ 0.11L ≥ 0.11XL ≥ 0.10
ASTM D5151-19Testing for Freedomfrom holesAQL 2.5Gloves pass AQL 2.5
ASTM D6124-17Determine thepowder residue forpowder free gloves< 2 mg/gloveAverage 0.43 mg/glove
ASTM D412ASTM D573Testing for PhysicalpropertycharacteristicsBefore AgingTensile strength ≥14 MPaUltimate Elongation≥500%After AgingTensile strength ≥14 MPaUltimate Elongation≥400%Before AgingTensile strength ≥21 MPaUltimate Elongation≥502%After AgingTensile strength ≥18 MPaUltimate Elongation≥465%
ISO 10993-5Tests for In vitrocytotoxicityUnder the conditions ofthe study non-cytotoxic.Under conditions of thestudy, it was considered"non-cytotoxic" at 25%,12.5%, and 6.25% anddemonstrate cytotoxicityat the 50% and 100% ofthe test item extract.
ISO 10993-10Evaluate theendpoint ofirritant forbiocompatibilityUnder the conditions ofthe study, not an irritant.Under the conditions ofthe study, not an irritant.
ISO 10993-10Evaluate theendpoint ofsensitization forbiocompatibilityUnder the conditions ofthe study, not a sensitizer.Under the conditions ofthe study, not a sensitizer.
ISO 10993-11Tests for systemictoxicityUnder the conditions ofthe study, the deviceextracts do not pose asystemic toxicityconcernUnder the conditions ofthe study, the deviceextracts do not pose asystemic toxicityconcern

Conclusions 8.

The conclusions drawn from the non-clinical tests demonstrate that the subject device, ELAZ Nitrile Powder Free Examination Glove, is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.