K Number
K210898

Validate with FDA (Live)

Date Cleared
2021-08-26

(153 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Device Description

The proposed device, Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size and color. The color of the proposed device is Blue and Purple-Blue. The proposed device is not provided as sterilized. The proposed device is made of Nitrile.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for Disposable Nitrile Examination Gloves.

Important Note: This document pertains to the clearance of a medical device – specifically disposable nitrile examination gloves. The concepts of "AI models," "human readers," "effect size," "multi-reader multi-case (MRMC) comparative effectiveness studies," "standalone algorithms," and "ground truth establishment for training sets" as typically applied to AI/ML medical devices do not apply to this product. This submission focuses on the physical and performance characteristics of the gloves themselves compared to a predicate device.

Therefore, the answers to several of your requested points will be "Not Applicable" or refer to standard laboratory testing rather than AI/ML model validation.


1. Table of Acceptance Criteria and Reported Device Performance

Device: Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue)

Test MethodPurposeAcceptance CriteriaReported Device Performance (Worst Case/Representative)
ASTM D5151Freedom from HolesAQL 2.5Blue Color: No water leakage inspected from 200 samples (Meets AQL 2.5)Purple-Blue Color: No water leakage inspected from 200 samples (Meets AQL 2.5)
ASTM D6124Residual Powder for Powder-Free Gloves<2.0 mg per gloveBlue Color: Avg. 0.29 mg (XS), 0.32 mg (S), 0.39 mg (M), 0.34 mg (L), 0.38 mg (XL)Purple-Blue Color: Avg. 0.33 mg (XS), 0.32 mg (S), 0.32 mg (M), 0.37 mg (L), 0.38 mg (XL)(All values are significantly less than 2.0 mg/glove)
ASTM D412, D573Physical Property CharacteristicsBefore Aging: Tensile Strength: ≥ 14 MPa, Ultimate Elongation: ≥ 500%After Aging: Tensile Strength: ≥ 14 MPa, Ultimate Elongation: ≥ 400%Blue Color: Before Aging: Tensile Strength: ≥ 19 MPa, Ultimate Elongation: ≥ 500%. After Aging: Tensile Strength: ≥ 17.8 MPa, Ultimate Elongation: ≥ 472%.Purple-Blue Color: Before Aging: Tensile Strength: ≥ 19 MPa, Ultimate Elongation: ≥ 500%. After Aging: Tensile Strength: ≥ 17.1 MPa, Ultimate Elongation: ≥ 463%.(All values exceed the minimum acceptance criteria)
ASTM D412, D3767Physical Dimensions SpecificationLength: ≥ 220 mm (XS, S), ≥ 230 mm (M, L, XL)Width: 70±10 mm (XS), 80±10 mm (S), 95±10 mm (M), 110±10 mm (L), 120±10 mm (XL)Cuff Thickness: ≥0.05 mm; Finger Thickness: ≥0.05 mm; Palm Thickness: ≥0.05 mm. (All meet ASTM D6319 requirements)Blue Color (Representative for XS): Length: ≥ 230 mm; Width: 75±3 mm (75-78 mm); Cuff/Palm/Finger Thickness: ≥0.06 mm/≥0.07 mm/≥0.10 mm respectively.Purple-Blue Color (Representative for XS): Length: ≥ 230 mm; Width: 75±4 mm (75-79 mm); Cuff/Palm/Finger Thickness: ≥0.05 mm/≥0.07 mm/≥0.08 mm respectively.(All reported dimensions across all sizes and colors consistently meet or exceed the specified minimums and ranges)
ISO 10993-11Systemic Toxicity (Biocompatibility)No evidence of systemic toxicity from the extract.The test article showed no evidence of systemic toxicity from the extract.
ISO 10993-10Irritation (Biocompatibility)The response of the test article extract is negligible.The test result showed that the response of the test article extract was categorized as negligible under the test condition.
ISO 10993-10Sensitization (Biocompatibility)No evidence of causing delayed dermal contact sensitization.The test article showed no evidence of causing delayed dermal contact sensitization in the guinea pig.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Physical/Performance Tests:
    • Freedom from Holes (ASTM D5151): 200 samples were inspected for each color (Blue and Purple-Blue). This is per the standard's sampling requirements for AQL testing.
    • Residual Powder (ASTM D6124): Quantities of powder were measured across various sizes (XS, S, M, L, XL) for both colors. The exact number of gloves per size/color tested isn't explicitly stated but would be based on the ASTM D6124 standard.
    • Physical Property Characteristics (Tensile Strength, Elongation): Not explicitly stated, but standard test methods (ASTM D412, D573) specify the number of specimens to be tested per batch.
    • Physical Dimensions: Not explicitly stated, but standard test methods (ASTM D412, D3767) specify the number of specimens to be measured.
    • Biocompatibility Tests (ISO 10993-10, -11): These tests involve a specific number of animal subjects (e.g., guinea pigs for sensitization, mice for systemic toxicity, rabbits for irritation). The exact number isn't in the summary but is defined by the ISO standards.
  • Data Provenance: The tests appear to be internal bench tests conducted by or for the manufacturer. No specific country of origin for the test data is given beyond the company's location (China). These are prospective tests performed specifically for this 510(k) submission, not retrospective data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not Applicable. This is not an AI/ML device where "ground truth" is established by human experts reviewing medical images. The "ground truth" here is the objective measurement against established ASTM and ISO standards for physical properties, chemical residue, and biological safety of gloves. These measurements are performed by trained technicians in a laboratory setting, not by medical experts interpreting data like radiologists.

4. Adjudication Method for the Test Set

Not Applicable. As this is not an AI/ML device; there is no human interpretation or adjudication involved in the test results. The tests are objective measurements against pre-defined quantitative criteria.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

Not Applicable. MRMC studies are used to assess the impact of AI on human reader performance, typically in imaging diagnostics. This device is a physical product (gloves) and does not involve human readers interpreting AI output.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not Applicable. This is not an algorithm or AI model. Its performance is measured directly through physical, chemical, and biological testing.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is established by:

  • Reference Standards: Adherence to established and internationally recognized industry standards (ASTM D6319, D5151, D6124, ISO 10993-10, ISO 10993-11).
  • Laboratory Measurements: Objective measurements of physical properties (e.g., length, width, thickness, tensile strength, elongation), chemical properties (residual powder), and biological responses (irritation, sensitization, systemic toxicity) using calibrated equipment and validated protocols.

8. The Sample Size for the Training Set

Not Applicable. This product is not an AI/ML model, and therefore, there is no "training set." The product's characteristics are inherent to its manufacturing process and materials, and its performance is validated through defined testing, not a learned model.

9. How the Ground Truth for the Training Set was Established

Not Applicable. As there is no training set for this type of device, this question is irrelevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 26, 2021

Tangshan Lanhai Medical Supplies Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm. 912, Building #15, XiYueHui, No. 5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China

Re: K210898

Trade/Device Name: Disposable Nitrile Examination Gloves (Powder Free, Purple-Blue, Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: June 1, 2021 Received: June 4, 2021

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K210898

Device Name

Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue)

Indications for Use (Describe)

The Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@.fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K210898

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number: K210898

    1. Date of Preparation: 08/24/2021
    1. Sponsor

Tangshan Lanhai Medical Supplies Co., Ltd.

South Side of Zhongda Street, East Side of Jinxiu Road, Chengxi Industrial Zone, Luannan County, Tangshan City, Hebei Province, P.R. China 063500 Contact Person: Li Ruidong Position: Manager Tel: +86-15931581811 Fax: +86-315-4155192 Email: 718356353@qq.com

  • Submission Correspondent 3.
    Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China,102401 Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com

  • Proposed Device Identification 4.
    Trade Name: Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue) Common Name: NITRILE Patient Examination Gloves (Powder Free) Regulatory Information: Classification: I Product Code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

Indication For Use Statement:

The Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

  • న్న Predicate Device Identification

{4}------------------------------------------------

510(k) Number: K150340

Product Name: POWDER FREE Nitrile GLOVES (White, Cobalt Blue, Black, Ice Blue) Manufacturer: HEBEI HONGSEN PLASTICS TECHNOLOGY CO., LTD

  • Device Description 6.
    The proposed device, Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size and color. The color of the proposed device is Blue and Purple-Blue.

The proposed device is not provided as sterilized

The proposed device is made of Nitrile.

DesignationSizeTolerance
XSSMLXL
Length, mm220220230230230min
Width, mm708095110120±10
Thickness, mm:
Finger0.05min
Palm0.05min
Cuff0.05min

Table 1 Device Size Specifications

Before AgingAfter AgingPinhole AQL
TensileStrengthUltimateElongationTensileStrengthUltimateElongation
14 MPa, min500 % min14 MPa, min400 % min2.5

Table 2 Performance and Physical Specifications

The above data of size, performance, and physical specifications of proposed gloves meet all the current specifications listed in the ASTM standard D6319.

    1. Comparison of technological characteristics between the subject and predicate devices
ITEMProposed Device (K210898)Predicate Device (K150340)Remark
Disposable Nitrile Examination Gloves(Powder free, Purple-Blue, Blue)POWDER FREE Nitrile GLOVES (White,Cobalt Blue, Black, Ice Blue)
Product CodeLZALZASAME
Regulation No.21 CFR 880.625021 CFR 880.6250SAME

Table 1 General Comparison

{5}------------------------------------------------

ClassIISAME
IntendedUse/IndicationFor UseThe Disposable Nitrile Examination Gloves(Powder free, Purple-Blue, Blue) is adisposable device intended for medicalpurposes that is worn on the examiner'shands to prevent contamination betweenpatient and examiner.The POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, Ice Blue) is adisposable device intended for medicalpurposes that is worn on the examiner'shands to prevent contamination betweenpatient and examiner.SAME
Powdered orPowered freePowdered freePowdered freeSAME
Proposed Device (K210898)SizeTolerance
Disposable Nitrile ExaminationGloves (Powder free,Purple-Blue, Blue)DesignationXSSMLXL
Length, mm220220230230230min
Width, mm708095110120±10
Thickness, mm:
Finger0.05min
Palm0.05min
Cuff0.05min
Predicate Device (K150340)Size
POWDER FREE NitrileGLOVES (White, Cobalt Blue,Black, Ice Blue)DesignationXSSMLXLTolerance
Length, mm230230230230230min
Width, mm708095110120±10
Thickness, mm:
Finger0.10-0.12±0.03
Palm0.08-0.10±0.03
Cuff0.06-0.09±0.03
RemarkSimilar

Table 2 Device Dimensions Comparison

Different Analysis:

The proposed device has different size specification to the predicate device, but all proposed devices are meet the specifications of ASTM D 6319.

ITEMProposed Device (K210898)Disposable Nitrile Examination Gloves(Powder free, Purple-Blue, Blue)Predicate Device (K150340)POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, IceBlue)Remark
ColorantBlue, Purple BlueWhite, Cobalt Blue, Black, Ice BlueDifferent
Table 3 Performance Comparison

{6}------------------------------------------------

Single UseYesYesSAME
PhysicalPropertiesBeforeAgingTensileStrength14 MPa, min15 MPa, minDifferent
UltimateElongation500 % min500 % minSAME
AfterAgingTensileStrength14 MPa, min14 MPa, minSAME
UltimateElongation400 % min400 % minSAME
Freedom from HolesComply with ASTM D6319Comply with ASTM D6319SAME
Be free from holes when tested inaccordance with ASTM D5151, AQL2.5Be free from holes when tested inaccordance with ASTM D5151,AQL 1.5Different
Powder ContentLess than 2 mg per glove when testedin accordance with ASTM D6124Meet the requirements of ASTM6124SAME

Different Analysis:

The proposed device has different color to the predicate device, this different may causes potential biocompatibility risk, for this risk we conducted the biocompatibility test according to the ISO 10993-10 and ISO 10993-11 and the test results showed that the proposed devices did not induce skin irritation and showed no significant evidence of causing skin sensitization and systemic toxicity response.

Different Analysis:

The proposed device has different Tensile Strength before aging specification to the predicate device, but all proposed device meets the specification requirements of ASTM D 6319.

Different Analysis:

The proposed device has different AQL to the predicate device, but all proposed device meets the specification requirements of ASTM D 6319.

ITEMProposed DevicePredicate Device (K150340)Remark
Disposable Nitrile ExaminationGloves (Powder free, Purple-Blue,Blue)POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, IceBlue)
Material usedNitrileNitrileSAME
BiocompatibilityIrritationUnder the conditions of the study,not an irritantUnder the conditions of the study,not an irritantSAME
SensitizationUnder conditions of the study, not asensitizer.Under conditions of the study, not asensitizer.
acutesystemicUnder the conditions of the study,there was no evidence of systemicNot AvailableDifferent
Table 4 Safety Comparisor

{7}------------------------------------------------

toxicitytoxicity from the extract.
Label and LabelingMeet FDA's RequirementsMeet FDA's RequirementsSAME

Different Analysis:

The proposed device has conducted the acute systemic toxicity testing, and the test results showed that there was no evidence of systemic toxicity.

8. Summary of Non-Clinical Testing

Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6124-17, Standard Test Method for Residual Powder on Medical Gloves.

ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

Test MethodPurposeAcceptance CriteriaResults
ASTM D5151Testing for Freedomfrom holesFreedom from holesBlue Color
from holesAQL 2.5No water leakage is inspected form 200samples
Purple-Blue Color
No water leakage is inspected form 200samples
ASTM D6124Determine the powder<2.0 mg per gloveBlue Color
residue for powder freeResidual Powder of Size XS: Average 0.29 mg;
glovesResidual Powder of Size S: Average 0.32 mg;
Residual Powder of Size M: Average 0.39 mg;
Residual Powder of Size L: Average 0.34 mg;
Residual Powder of Size XL: Average 0.38mg;
Purple-Blue Color
Residual Powder of Size XS: Average 0.33mg;
Residual Powder of Size S: Average 0.32 mg;
Residual Powder of Size M: Average 0.32 mg;
Residual Powder of Size L: Average 0.37 mg;
Residual Powder of Size XL: Average 0.38mg;
ASTM D412Testing for Physicalproperty characteristicsBefore AgingBlue Color
ASTM D573property characteristicsTensile Strength: 14 MPa min.Ultimate Elongation: 500% min.Before AgingTensile Strength: ≥ 19 MPa;
ASTM D412ASTM D3767Testing For physicaldimensions specificationBefore AgingTensile Strength: 14 MPa min.Ultimate Elongation: 400% min.Ultimate Elongation: ≥ 500%.After AgingTensile Strength: ≥ 17.8 MPa;Ultimate Elongation: ≥ 472%.
Purple-Blue Color
Before AgingTensile Strength: ≥ 19 MPa;Ultimate Elongation: ≥ 500%.After AgingTensile Strength: ≥ 17.1 MPa;Ultimate Elongation: ≥ 463%.
Length: 220 mm min. for size (XS, S);Length: 230 mm min. for size (M, L, XL);Width: 70±10 mm for XS; 80±10 mmfor S; 95±10 mm for M; 110±10 mm forL; 120±10 mm for XL.Cuff Thickness: ≥0.05 mm;Finger Thickness: ≥0.05 mm;Palm Thickness: ≥0.05 mm;All acceptance criteria above meet therequirements in Table 1 Dimensions andTolerances of ASTM D6319Blue ColorLength of Size XS: ≥ 230 mm;Width of Size XS: 75±3 (75-78) mm;Cuff Thickness of Size XS: ≥0.06 mm;Palm Thickness of Size XS: ≥0.07 mm;Finger Thickness of Size XS: ≥0.10 mm.Length of Size S: ≥ 228mm;Width of Size S: 85±2 (85-87) mm;Cuff Thickness of Size S: ≥0.06 mm;Palm Thickness of Size S: ≥0.07 mm;Finger Thickness of Size S: ≥0.10 mm.Length of Size M: ≥ 240 mm;Width of Size M: 95±2 (95-97) mm;Cuff Thickness of Size M: ≥0.06 mm;Palm Thickness of Size M: ≥0.07 mm;Finger Thickness of Size M: ≥0.10 mm.Length of Size L: ≥ 240 mm;Width of Size L: 105±4 (105-109) mm;Cuff Thickness of Size L: ≥0.06 mm;Palm Thickness of Size L: ≥0.07 mm;Finger Thickness of Size L: ≥0.10 mm.Length of Size XL: ≥ 241 mm;Width of Size XL: 115±2 (115-117) mm;Cuff Thickness of Size XL: ≥0.06 mm;Palm Thickness of Size XL: ≥0.07 mm;Finger Thickness of Size XL: ≥0.10 mm
Purple-Blue Color
Length of Size XS: ≥ 230 mm;
Width of Size XS: 75±4 (75-79) mm;
Cuff Thickness of Size XS: ≥0.05 mm;
Palm Thickness of Size XS: ≥0.07 mm;
Finger Thickness of Size XS: ≥0.08 mm.
Length of Size S: ≥ 235mm;
Width of Size S: 85±2 (85-87) mm;
Cuff Thickness of Size S: ≥0.05 mm;
Palm Thickness of Size S: ≥0.05 mm;
Finger Thickness of Size S: ≥0.08 mm.
Length of Size M: ≥ 236 mm;
Width of Size M: 95±3 (95-98) mm;
Cuff Thickness of Size M: ≥0.05 mm;
Palm Thickness of Size M: ≥0.05 mm;
Finger Thickness of Size M: ≥0.07 mm.
Length of Size L: ≥ 240 mm;
Width of Size L: 105±3 (105-108) mm;
Cuff Thickness of Size L: ≥0.05 mm;
Palm Thickness of Size L: ≥0.06 mm;
Finger Thickness of Size L: ≥0.08 mm.
Length of Size XL: ≥ 240 mm;
Width of Size XL: 115±3 (116-118) mm;
Cuff Thickness of Size XL: ≥0.05 mm;
Palm Thickness of Size XL: ≥0.05 mm;
Finger Thickness of Size XL: ≥0.08 mm.
ISO 10993-11Evaluate the endpoint ofSystemic Cytotoxicityfor biocompatibilityThe test article showed no evidence ofsystemic toxicity from the extract.The test article showed no evidence ofsystemic toxicity from the extract.
ISO 10993-10Evaluate the endpoint ofirritant forbiocompatibilityThe response of the test article extract isnegligible.The test result showed that the response of thetest article extract was categorized asnegligible under the test condition.
Evaluate the endpoint ofsensitization forbiocompatibilityThe test article showed no evidence ofcausing delayed dermal contactsensitizationThe test article showed no evidence of causingdelayed dermal contact sensitization in theguinea pig.

Table 5 Performance Test Results Summary

{8}------------------------------------------------

{9}------------------------------------------------

  • Summary of Clinical Testing 9. Clinical Testing is not applicable.

{10}------------------------------------------------

10. Conclusion

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, POWDER FREE Nitrile GLOVES (White, Cobalt Blue, Black, Ice Blue) cleared under K150340.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.