Intended Use

EONS Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

EONS Nitrile Examination Gloves Powder Free is a Class I device bearing the product code LZA (21CFR 880.6250). They meet all the current specifications listed under the ASTM Specification D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application. They are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color having Finger Texture, Ambidextrous and are powder free. The product is non-sterile.

AI/ML Overview

The document provided describes the acceptance criteria and performance of the EONS Nitrile Examination Gloves Powder Free, a medical device. This is a 510(k) premarket notification for a Class I, reserved medical device (non-powdered patient examination glove, product code LZA).

Here's an analysis based on the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Feature ComparedAcceptance CriteriaReported Device Performance (EONS Nitrile Examination Gloves Powder Free)Result
Dimensions
Length (ASTM D6319-19)Min 230 mmS: 243.69 mm; M: 243.30 mm; L: 242.84 mm; XL: 242.15 mmPass
Width (ASTM D6319-19)S: 80±10 mm; M: 95±10 mm; L: 110±10 mm; XL: 120±10 mmS: 83.84 mm; M: 94.15 mm; L: 109.07 mm; XL: 114.53 mmPass
Thickness (ASTM D3767-03)Palm min 0.05 mm; Finger min 0.05 mmS: Palm 0.08 mm, Finger 0.12 mm; M: Palm 0.12 mm, Finger 0.08 mm; L: Palm 0.08 mm, Finger 0.10 mm; XL: Palm 0.09 mm, Finger 0.16 mmPass
Physical Properties
Tensile Strength (Before Aging) (ASTM D6319-19, ASTM D412-16)Min 14 MPaS: 31.26 MPa; M: 23.61 MPa; L: 20.55 MPa; XL: 19.9 MPaPass
Tensile Strength (After Aging) (ASTM D6319-19, ASTM D412-16)Min 14 MPaS: 23.51 MPa; M: 24.33 MPa; L: 21.6 MPa; XL: 19.9 MPaPass
Ultimate Elongation (Before Aging) (ASTM D6319-19, ASTM D412-16)Min 500%S: 799.76%; M: 885.53%; L: 849.69%; XL: 555.30%Pass
Ultimate Elongation (After Aging) (ASTM D6319-19, ASTM D412-16)Min 400%S: 815.61%; M: 738.76%; L: 708%; XL: 444.15%Pass
Detection of Holes (ASTM D6319-19, ASTM D5151-06)AQL 2.5AQL 2.5 (for all sizes S, M, L, XL)Pass
Powder Free Residue (ASTM D6124-06)≤2 mg/gloveS: 0.6 mg/glove; M: 0.7 mg/glove; L: 0.4 mg/glove; XL: 0.7 mg/glovePass
Biocompatibility
In vitro CytotoxicityAcceptable (after additional testing for systemic toxicity concern)Under the conditions of the study, the device is cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concern (Result: Pass)Pass
Skin SensitizationNot a sensitizerUnder the conditions of the study not a sensitizerPass
Skin IrritationNot an irritantUnder the condition of study not an irritantPass
Acute Systemic ToxicityNo systemic toxicity concernUnder the condition of study, the device extracts do not pose a systemic toxicity concernPass
Material-Mediated PyrogenicityNon-pyrogenic responseUnder the conditions of the study, the device demonstrate a non-pyrogenic response.Pass

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample sizes for each specific test (e.g., how many gloves were tested for dimensions, tensile strength, barrier properties, or biocompatibility). However, it mentions that the tests were performed "according to the requirements of Guidance for Industry and FDA Staff - Medical Glove Guidance Manual" and specific ASTM and ISO standards. These standards typically specify the sampling plans and test methodologies.

The provenance of the data is not explicitly stated in terms of country of origin for the testing itself, but the manufacturer is EONS GLOVES (Thailand) Co., Ltd. The study type is non-clinical data, focusing on performance and biocompatibility rather than clinical trials.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This question is not applicable to this type of device and study. The "ground truth" for the performance and biocompatibility tests of medical gloves is established by the specified ASTM and ISO standards and their defined methodologies, not by expert consensus or interpretation of images/data by human experts. The results are quantitative measurements against predefined criteria in the standards.

4. Adjudication Method for the Test Set:

This question is not applicable. The performance testing involves objective measurements and adherence to specific standards. There's no human interpretation or adjudication process in the sense of consensus among experts for diagnostic output.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This question is not applicable. This is a physical medical device (gloves), not an AI-powered diagnostic or interpretive tool. Therefore, MRMC studies involving human readers and AI assistance are not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This question is not applicable. This is a physical medical device. It does not involve an algorithm or AI.

7. The Type of Ground Truth Used:

The "ground truth" for evaluating the EONS Nitrile Examination Gloves Powder Free is based on:

  • Established ASTM Standards: For physical properties and barrier properties (e.g., ASTM D6319-19, ASTM D3767-03, ASTM D412-16, ASTM D5151-06, ASTM D6124-06).
  • Established ISO Standards: For biocompatibility tests (e.g., ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2010, ISO 10993-11:2017).
  • Guidance from Regulatory Bodies: Specifically, the "Guidance for Industry and FDA Staff - Medical Glove Guidance Manual."

These standards define the acceptable ranges and methodologies for testing, which serve as the objective performance criteria.

8. The Sample Size for the Training Set:

This question is not applicable. This device does not involve machine learning or AI, and therefore, there is no "training set." The tests are designed to evaluate the physical and biological characteristics of the manufactured product against established quality standards.

9. How the Ground Truth for the Training Set was Established:

This question is not applicable for the reason stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 19, 2022

Eons Gloves (Thailand) Co., Ltd % Manoj Zacharias US Agent Liberty Management Group Limited 75 Executive Drive Suite 114 Aurora, Illinois 60504

Re: K213286

Trade/Device Name: Eons Nitrile Examination Gloves Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: January 24, 2022 Received: January 31, 2022

Dear Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213286

Device Name

EONS Nitrile Examination Gloves Powder Free

Indications for Use (Describe)

EONS Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for EONS GLOVES. The logo is in the shape of a silver shield with a blue ribbon across the middle. The words "EONS GLOVES" are written in white on the blue ribbon. There is a plus sign in the center of the shield above the ribbon.

EONS GLOVES (THAILAND) CO., LTD

486/14 Moo 9 Nai Klong bang Pla Kot, Phra Samut Chedi District, Samut Prakan Province 10290 Thailand Email: eonsglovesthailand@gmail.com Tel: 02-464-0842 HP: +6681-934-3579 www.eonsgloves.com

510(k) SUMMARY (K213286)

[AS REQUIRED BY 21CFR807.92]

SUBMITTER I.

510(k) Owner's Name:EONS GLOVES (THAILAND) CO., LTD
Address:486/14 Moo 9 Nai Klong Bang Pla Kot, Phra, Samut ChediSamutPrakan,10290Thailand
Telephone:+66 958638585
Contact person:Ms. Gliaophan Supharkarn
Designation:Vice President
Contact Number:+66 958638585
Contact Email:toutah.13@gmail.com
Date of Summary Prepared:24.01.2022

II. DEVICE

Device Name:EONS Nitrile Examination Gloves Powder Free
Device Common Name:Nitrile Examination Gloves Powder Free
Device Classification name:Non-powdered patient examination glove
Regulation Number:21 CFR 880.6250
Class:I
Product Code:LZA

III. PREDICATE DEVICE

Predicate Device Name: JR MEDIC Blue Nitrile Examination Gloves Powder Free
510(k) Number: K192333
Regulation Number: 21 CFR 880.6250
Class: I
Product Code: LZA

{4}------------------------------------------------

IV. DEVICE DESCRIPTION

EONS Nitrile Examination Gloves Powder Free is a Class I device bearing the product code LZA (21CFR 880.6250). They meet all the current specifications listed under the ASTM Specification D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application. They are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color having Finger Texture, Ambidextrous and are powder free. The product is non-sterile.

V. INTENDED USE

EONS Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

SI.NoFeatures comparedProposed DevicePredicate DeviceResult
General Information
1.510(k) NumberK213286K192333-
2.ManufacturerEONS GLOVES (Thailand) CO.,LTDJR Engineering & MedicalTechnologies (M) SDN.BHD-
3.ClassificationIISame
4.Regulation number21 CFR 880.625021 CFR 880.6250Same
5.Product CodeLZALZASame
6.Indication For UseEONS Nitrile Examination GlovesPowder Free is a disposable deviceintended for medical purpose thatis worn on the examiner's hand toprevent contamination betweenpatient and examiner.JR MEDIC Blue NitrileExamination Gloves Powder Freeis disposable devices intended formedical purpose that are worn onthe examiner's hand to preventcontamination between patientand examiner.Same
7.MaterialNitrileNitrileSame
8.ColorBlueBlueSame
9.TextureFinger TextureFinger textureSame
10.AmbidextrousYesData Not available-
11.SizeS, M, L, XLXS, S, M, L, XLSimilar1
12.OTC UseYesYesSame
13.ReusabilitySingle useSingle useSame
14.SterilityNon- sterileNon- sterileSame
15.DimensionsLength Min 230 mWidth Min 95±10Mm (for medium size)Length Min 230 mWidth Min 95±10Mm (for medium size)Same
16.ThicknessPalm min 0.05 mmFinger min 0.05 mmPalm min 0.05 mmFinger min 0.05 mmSame

Table 1: General Comparison

{5}------------------------------------------------

Sl.NoFeatures comparedProposed DevicePredicate DeviceResult
17.Physical PropertiesBefore AgingTensile Strength min 14 MPaUltimate Elongation Min 500%After AgingTensile Strength min 14 MPaUltimate Elongation Min 400%Before AgingTensile Strength min 14 MPaUltimate Elongation Min 500%After AgingTensile Strength min 14 MPaUltimate Elongation Min 400%Same
18.Detection of HolesPasses AQL 2.5Passes AQL 1.5Similar²
19.Powder Free Residue≤2 mg/glove≤2 mg/gloveSame
20.Biocompatibility StudyIn vitro CytotoxicityUnder the conditions of the study, cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concernUnder the conditions of the study, cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concernSame
Skin SensitizationUnder the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizerSame
Skin IrritationUnder the condition of study not an irritantUnder the condition of study not an irritantSame
Acute systemic toxicityUnder the condition of study, the device extracts do not pose a systemic toxicity.Under the condition of study, the device extracts do not pose a systemic toxicity.Same
Material mediated pyrogenicityUnder the conditions of the study, the device demonstrate a non-pyrogenic response.Under the conditions of the study, the device did not demonstrate a material mediated Pyrogenicity response.Same

There are no significant differences between the two products and are identical in terms of intended use, materials, design, manufacturing methods. Both devices met the performance standards.

VII. PERFORMANCE DATA

A. Non- Clinical Data

Performance Tests

EONS Nitrile Examination Gloves Powder Free is subjected to the following performance tests according to the requirements of Guidance for Industry and FDA Staff - Medical Glove Guidance Manual and found to be safe and efficient with respect to its intended use:

  • · Dimension
  • Physical property
  • · Barrier property tests
    • Detection of Holes in Medical Gloves

  • Powder Free Residue

{6}------------------------------------------------

TestsProposed DeviceActual DataAcceptance CriteriaResult
SizeLengthWidthSizeLengthWidth
1.DimensionLength, Width and ThicknessASTM D6319-19ASTM D3767-03S243.69 mm83.84 mmS80±10 mmPass
M243.30 mm94.15 mmM230 mm95±10 mm
L242.84 mm109.07 mmLmin110±10 mm
XL242.15 mm114.53 mmXL120±10 mm
ASTM D3767-03(Reapproved 2014)Standard Specification for NitrileExamination Gloves for MedicalApplicationThickness
SizePalmFingerSizePalmFinger
S0.08 mm0.12 mmS
M0.12 mm0.08 mmM0.05 mm0.05 mm
L0.08 mm0.10 mmLminmin
XL0.09 mm0.16 mmXL
Tensile Strength
SizeBeforeAgingAfterAgingSizeBeforeAgingAfterAging
Physical propertyTensile strength and UltimateElongationASTM D6319-19ASTM D412-16Standard Specification for NitrileExamination Gloves for MedicalApplicationS31.26 MPa23.51 MPaSPass
2.M23.61 MPa24.33 MPaM14 MPa14 MPa
L20.55 MPa21.6 MPaLminmin
XL19.9 MPa19.9 MPaXL
Ultimate Elongation
SizeBeforeAgingAfterAgingSizeBeforeAgingAfterAging
S799.76 %815.61 %S
M885.53 %738.76 %M500%400%
L849.69 %708 %Lminmin
XL555.30 %444.15 %XL
3.Barrier property testsDetection of Holes in MedicalGlovesASTM D6319-19ASTM D5151-06(Reapproved 2015)Standard Test Method forDetection of Holes in MedicalGlovesSAQL 2.5SAQL 1.5Pass
MM
LL
XLXL
4.Powder Free ResidueASTM D6124-06(Reapproved 2017)Standard Test Method forResidual Powder on MedicalGlovesSizeResidual Powder ContentPass
S0.6 mg/gloveS≤2 mg/glove
M0.7 mg/gloveM
L0.4 mg/gloveL
XL0.7 mg/gloveXL

Table 2: Performance Testing Summary

{7}------------------------------------------------

B. Biocompatibility

The materials used in the EONS Nitrile Examination Gloves Powder Free are biocompatible based on the biocompatibility tests mentioned in the Guidance for Industry and FDA Staff - Medical Glove Guidance Manual:

  • · In Vitro Cytotoxicity
  • . Skin Sensitization
  • . Skin Irritation
  • · Acute Systemic Toxicity
  • · Material-Mediated Pyrogenicity

These tests are performed according to ISO 10993-1:2018, Biological Evaluation of Medical Devices -Part 1, Evaluation and Testing within a Risk Management Process.

SI.NoTest PerformedStandardProposed DeviceResult
1.In vitro CytotoxicityISO 10993-5:2009Under the conditions of the study, thedevice is cytotoxic. Additional testingwas performed to determine if this wasa systemic toxicity concernPass
2.Skin SensitizationISO 10993-10:2010Under the conditions of the study not asensitizerPass
3.Skin IrritationISO 10993-10:2010Under the condition of study not anirritantPass
4.Acute Systemic ToxicityISO 10993-11:2017Under the condition of study, the deviceextracts do not pose a systemic toxicityconcernPass
5.Material-MediatedPyrogenicityISO 10993-11:2017(E)Under the conditions of the study, thedevice demonstrate a non-pyrogenicresponse.Pass

Table 3: Biocompatibility Test Summary

C. Clinical Test Data

Clinical study was not conducted, as clinical data is not needed for EONS Nitrile Examination Gloves Powder Free.

VIII. CONCLUSION

The conclusion drawn from the non-clinical tests demonstrate that the subject device, EONS Nitrile Examination Gloves Powder Free are as safe, as effective and perform as well as or better than legally marketed predicated device in K192333.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.