K Number
K211714

Validate with FDA (Live)

Date Cleared
2021-08-04

(62 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extralarge.

AI/ML Overview

The provided text describes the acceptance criteria and performance data for "Nitrile Exam Gloves, Powder Free, Blue (Tested for Use with Chemotherapy Drugs)" in the context of an FDA 510(k) premarket notification.

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D6319Physical Dimensions TestLength (mm): S: ≥220 mm; M/L/XL: ≥230 mm. Width (mm): S: 80±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mm. Thickness (mm): Finger: ≥0.05mm; Palm: ≥0.05mm.Length: >230 mm (Pass) Width: S: 81-86, M: 93-99, L: 104-109, XL: 113-116 (Pass) Thickness: Finger: 0.11-0.15mm, Palm: 0.08-0.15mm (Pass)
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125 leaks (Pass)
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.08 (mg/glove) (Pass)
ASTM D412Physical properties (Tensile Strength)Before Aging: ≥14MPa. After Aging: ≥14MPa.Before Aging: 14.0-15.4 (Pass). After Aging: 14.0-15.0 (Pass).
ASTM D412Physical properties (Ultimate Elongation)Before Aging: ≥500%. After Aging: ≥400%.Before Aging: 506-664 (Pass). After Aging: 420-511 (Pass).
ISO 10993-5CytotoxicityNon-cytotoxicDid not show potential toxicity to L-929 cells (Pass)
ISO 10993-10IrritationNon-irritatingNot an irritant (Pass)
ISO 10993-10SensitizationNon-sensitizingNot a sensitizer (Pass)
ASTM D6978-05Permeation by Chemotherapy Drugs(Varying breakthrough detection times for different drugs. Implied acceptance is to demonstrate performance and clearly state findings, especially for drugs with low permeation.)Carmustine (BCNU): 23.6 Minutes. Cisplatin: > 240 Minutes. Cyclophosphamide: > 240 Minutes. Dacarbazine: > 240 Minutes. Doxorubicin HCl: > 240 Minutes. Etoposide: > 240 Minutes. Fluorouracil: > 240 Minutes. Methotrexate: > 240 Minutes. Mitomycin C: > 240 Minutes. Paclitaxel: > 240 Minutes. Thio Tepa: 57.4 Minutes. Vincristine Sulfate: > 240 Minutes.

2. Sample Size Used for the Test Set and Data Provenance

  • Physical and Biocompatibility Tests: The specific sample sizes for each physical property test (e.g., length, width, thickness, tensile strength, elongation, powder content, holes) and biocompatibility tests (cytotoxicity, irritation, sensitization) are not explicitly stated. However, the results (e.g., "0/125 leaks" for watertightness) indicate that a sample was tested.
  • Chemotherapy Drug Permeation Test: The sample size for the chemotherapy drug permeation testing is not explicitly provided.
  • Data Provenance: The document does not specify the country of origin of the data for these tests. The tests were presumably conducted by or for Hengchang (Dongying) Medical Technology Co., Ltd. in China. The study is a prospective evaluation of the manufactured gloves against specific standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This is not applicable as the studies described are bench tests and biocompatibility tests for a physical medical device (gloves), not studies involving human interpretation of clinical data or images. Ground truth is established by standardized testing protocols and measurements, not expert consensus.

4. Adjudication Method for the Test Set

Not applicable, as expert adjudication is not relevant for physical and chemical standardized tests. Test results are objective measurements against defined criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This device is a pair of examination gloves, not an AI-powered diagnostic or assistive technology for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device does not involve an algorithm or AI.

7. The Type of Ground Truth Used

  • For physical properties and safety (holes, powder content, dimensions, tensile strength, elongation): Ground truth is established by adherence to recognized industry standards and their specified test methodologies (e.g., ASTM D6319, ASTM D5151, ASTM D6124, ASTM D412). The "ground truth" is the quantitative measurement of these properties.
  • For biocompatibility (cytotoxicity, irritation, sensitization): Ground truth is established by adherence to recognized international standards for biological evaluation of medical devices (ISO 10993-5, ISO 10993-10). The "ground truth" is the observed biological response in the specified test models.
  • For chemotherapy drug permeation: Ground truth is established by adherence to recognized industry standards for assessing resistance to chemotherapy drugs (ASTM D6978-05). The "ground truth" is the measured breakthrough detection time.

8. The Sample Size for the Training Set

Not applicable. This is a physical device being tested against standards, not an AI model requiring a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set is relevant to this type of device and testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 4, 2021

Hengchang (Dongying) Medical Technology Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161 Lujiazui East Rd., Pudong Shanghai. 200120 China

Re: K211714

Trade/Device Name: Nitrile Exam Gloves, Powder Free,Blue (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: May 25, 2021 Received: June 3, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211714

Device Name

Nitrile Exam Gloves, Powder Free,Blue (Tested for Use with Chemotherapy Drugs)

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes
Carmustine (BCNU)3.3 mg/ml (3,300 ppm)23.6 Minutes
Cisplatin1.0 mg/ml(1,000 ppm)> 240 Minutes
Cyclophosphamide (Cytoxan)20.0 mg/ml(20,000 ppm)> 240 Minutes
Dacarbazine (DTIC)10.0 mg/ml(10,000 ppm)> 240 Minutes
Doxorubicin HCl2.0 mg/ml(2,000 ppm)> 240 Minutes
Etoposide20.0 mg/ml(20,000 ppm)> 240 Minutes
Fluorouracil50.0 mg/ml(50,000 ppm)> 240 Minutes
Methotrexate25 mg/ml(25,000 ppm)> 240 Minutes
Mitomycin C0.5 mg/ml(500 ppm)> 240 Minutes
Paclitaxel6.0 mg/ml(6,000 ppm)> 240 Minutes
Thio Tepa10.0 mg/ml(10,000 ppm)57.4 Minutes
Vincristine Sulfate1.0 mg/ml(1,000 ppm)> 240 Minutes

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 23.6 Minutes Thio-Tepa 10.0 mg/ml 57.4 Minutes

Caution: Testing showed an average breakthrough time of 57.4 minutes with Thio-Tepa WARNING: Do not use with Carmustine

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary (K211714)

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: Hengchang (Dongying) Medical Technology Co.,Ltd. Address: Room 101,Building 2,No.26 Xinghe Road,Niuzhuang Town,Dongying District,Dongying City,China. Phone Number: +86-18653343268 Contact: Li Jing Date of Preparation: Aug.4,2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Nitrile Exam Gloves, Powder Free,Blue (Tested for Use with Chemotherapy Drugs) Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL

3.0_Classification

Production code: LZA,LZC Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

  • Manufacturer: Medline Industries, Inc. Device: Medline Powder-Free Light Blue Nitrile Exam Glove (Tested for Use with Chemotherapy Drugs)

{5}------------------------------------------------

510(k) number: K201390

5.0 Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extralarge.

6.0 Indication for Use_

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs, per ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugConcentrationBreakthrough DetectionTime in Minutes
Carmustine (BCNU)3.3 mg/ml (3,300 ppm)23.6
Cisplatin1.0 mg/ml(1,000 ppm)> 240
Cyclophosphamide(Cytoxan)20.0 mg/ml(20,000 ppm)> 240
Dacarbazine (DTIC)10.0 mg/ml(10,000 ppm)> 240
Doxorubicin HCI2.0 mg/ml(2,000 ppm)> 240
Etoposide20.0 mg/ml(20,000 ppm)> 240
Fluorouracil50.0 mg/ml(50,000 ppm)> 240
Methotrexate25 mg/ml(25,000 ppm)> 240
Mitomycin C0.5 mg/ml(500 ppm)> 240
Paclitaxel6.0 mg/ml(6,000 ppm)> 240
Thio Tepa10.0 mg/ml(10,000 ppm)57.4
Vincristine Sulfate1.0 mg/ml(1,000 ppm)> 240

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 23.6 Minutes; Thio Tepa 10.0 mg/ml 57.4Minutes.

Caution: Testing showed an average breakthrough time of 57.4 minutes with Thio-Tepa WARNING: Do not use with Carmustine

7.0 Technological Characteristic Comparison Table

{6}------------------------------------------------

ItemSubject DevicePredicate DeviceRemark
Product CodeLZA,LZCLZA,LZCSame
510(k) ReferenceK211714K201390Same
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseA patient examination gloveis a disposable deviceintended for medicalpurposes that is worn on theexaminer's hands to preventcontamination betweenpatient and examiner.These gloves were tested foruse with chemotherapydrugs, per ASTM D6978-05Standard Practice forAssessment of Resistanceof Medical Gloves toPermeation byChemotherapy Drugs.A patient examination gloveis a disposable deviceintended for medicalpurposes that is wornon the examiner's hand toprevent contaminationbetween patient andexaminer.These gloves were tested forusewith chemotherapy drugsas per ASTM D6978-05(2019) Standard Practicefor Assessment ofResistance of MedicalGloves to Permeation byChemotherapy Drugs.Same
Powdered orPowered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
Sterile vs Non-SterileNon-SterileNon-SterileSame
ColorBlueLight BlueDifferent
Labeling InformationSingle-use indication,powder free, device color,device name, glove size andquantity,Non-Sterile, astatement of standardASTM D6978-05compliance and asummary of the testingresults.Single-use indication, powderfree, device color, devicename, glove size andquantity, Non-Sterile, astatement of standardASTM D6978-05compliance and a summaryof the testing results.Similar
Dimensions - LengthComplies with ASTMD6319-19:S: ≥220 mm;M/L/XL: ≥230 mm.Complies with ASTM D6319-19: ≥240 mm.Similar
Dimensions - WidthComplies with ASTMD6319-19:S:80±10mm;M:95±10mm;L: 110±10mm;Complies with ASTM D6319-19:S:85±10mm;M:95±10mm;L: 105±10mm;Similar

{7}------------------------------------------------

XL: 120±10mm;XL: 115±10mm;
Dimensions -ThicknessComplies with ASTM D6319-19Palm:≥0.05mmFinger: ≥0.05mmComplies with ASTM D6319-19Palm:≥0.16mmFinger: ≥0.14mmSimilar
Physical Properties -Tensile StrengthComplies with ASTM D6319-19:Before Aging: ≥14MPaAfter Aging: ≥500%Complies with ASTM D6319-19:Before Aging: ≥17MPaAfter Aging: ≥500%Similar
Physical Properties -ElongationComplies with ASTM D6319-19:Before Aging: ≥14MPaAfter Aging: ≥400%Complies with ASTM D6319-19:Before Aging: ≥14MPaAfter Aging: ≥400%Same
Freedom from HolesComplies with ASTM D6319-19and ASTM D5151-19G-1, AQL 2.5Complies with ASTM D6319-19and ASTM D5151-19G-1, AQL 2.5Same
Powder ContentComplies with ASTM D6319-19,< 2 mg per gloveComplies with ASTM D6319-19,< 2 mg per gloveSame
BiocompatibilityComplies with ISO10993-5 (2009)* Under the conditions of thestudy, the device is notcytotoxic.Complies with ISO10993-10 (2010)* Under the conditions of thestudy, the device is a non-irritant and a non-sensitizer.Comply with ISO10993-10(2010) and ISO10993-5 (2009)Same
Chemotherapy drugs testedBreakthrough Detection Time in Minutes/
Subject DevicePredicate Device
Carmustine (BCNU), 3.3 mg/ml23.625.3Different
Cisplatin 1 mg/ml> 240> 240Same
Cyclophosphamide 20 mg/ml> 240> 240Same
Dacarbazine (DTIC), 10.0 mg/ml> 240> 240Same
Doxorubicin Hydrochloride, 2.0mg/ml> 240> 240Same
Etoposide (Toposar), 20.0 mg/ml> 240> 240Same
Fluorouracil, 50.0 mg/ml> 240> 240Same
Methotrexate 25 mg/ml> 240> 240Same
Mitomycin C 0.5 mg/ml> 240> 240Same
Paclitaxel, 6.0 mg/ml> 240> 240Same
Thiotepa, 10.0 mg/ml57.443.7Different
Vincristine Sulfate 1.0 mg/ml> 240> 240Same

{8}------------------------------------------------

Analysis:

  1. The physical properties are a little different with that of the predicate, but they all meet the requirements of ASTM D6319-19.

  2. The color of the subject device is different with that of the predicate. The subject device was evaluated according to ISO 10993-1 standards, and there were no risks identified.

3: Breakthrough detection times of Carmustine (BCNU) and Thio Tepa are different. The IFU Statement has clearly defined on the labeling.

Test MethodPurposeAcceptance CriteriaResults
ASTM D6319PhysicalDimensions TestLength(mm):S: ≥220 mm;M/L/XL: ≥230 mm.Length:>230
Width(mm):S:80±10mm;M:95±10mm;L: 110±10mm;XL: 120±10mm.Width:S: 81-86M: 93-99L: 104-109XL: 113-116
Thickness (mm):Finger: ≥0.05Palm: ≥0.05
Finger: 0.11-0.15Palm: 0.08-0.15
Pass
ASTM D5151WatertightnessTest forDetection ofHolesMeet the requirements ofASTM D5151 AQL 2.50/125 leaksPass
ASTM D6124Powder ContentMeet the requirements ofASTM D6124 < 2.0mg0.08 (mg/glove)Pass
ASTM D412PhysicalpropertiesBeforeAging
Tensile Strength ≥14MPa 14.0-15.4Pass
UltimateElongation≥500% 506-664Pass
AfterAgingTensile Strength ≥14MPa 14.0-15.0Pass
Ultimate Elongation ≥400% 420-511Pass
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions ofthe study, did notshow potential toxicityto L-929 cells.Pass
ISO 10993-10IrritationNon-irritatingUnder the conditionsof the study, not anirritant.Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditions ofthe study, not asensitizer.Pass

8.0 Summary of Non-Clinical Testing

{9}------------------------------------------------

The biocompatibility evaluation for Nitrile Exam Gloves, Powder Free,Blue (Tested for Use with Chemotherapy) was conducted in accordance with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

Performance Testing (Bench)

Physical performance qualities of the proposed device were evaluated per ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs.

In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • -ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
  • -ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
  • -ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

ASTM D 6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

9.0 Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Nitrile Exam Gloves, Powder Free,Blue (Tested for Use with Chemotherapy Drugs) is as safe, as effective, and performs as well as or better than the legally marketed predicated device under K201390.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.