K Number
K201390

Validate with FDA (Live)

Date Cleared
2020-09-18

(114 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

The Medline Powder-Free Light Blue Nitrile Exam Gloves (Tested for Use with Chemotherapy Drugs) are single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, light blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large.

AI/ML Overview

This document is a 510(k) summary for the Medline Powder-Free Light Blue Nitrile Exam Gloves, which have been tested for use with chemotherapy drugs. It is a report from Medline Industries, Inc. to the U.S. Food & Drug Administration (FDA) dated September 18, 2020. The purpose of this summary is to demonstrate substantial equivalence to a legally marketed predicate device (K172525).

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document provides performance data for resistance to permeation by chemotherapy drugs, evaluated per ASTM D6978-05 (Reapproved 2019). The "acceptance criteria" can be inferred from the "≥ 240 Minutes" value for most drugs, indicating that a barrier of at least 240 minutes is considered acceptable for practical use, though specific quantitative acceptance criteria are not explicitly stated for all tests. For Carmustine and Thio-Tepa, specific (lower) breakthrough times are reported, along with cautions/warnings.

CharacteristicAcceptance Criteria (Inferred from Predicate/Standard)Reported Device Performance (Medline Powder-Free Light Blue Nitrile Exam Gloves)
Chemotherapy Drug Permeation (ASTM D6978-05)
Carmustine (BCNU) 3.3 mg/mlPredicate: 12.4 Minutes25.3 Minutes
Cisplatin 1.0 mg/mlPredicate: ≥ 240 Minutes≥ 240 Minutes
Cyclophosphamide (Cytoxan) 20.0 mg/mlPredicate: ≥ 240 Minutes≥ 240 Minutes
Dacarbazine (DTIC) 10.0 mg/mlPredicate: ≥ 240 Minutes≥ 240 Minutes
Doxorubicin Hydrochloride 2.0 mg/mlPredicate: ≥ 240 Minutes≥ 240 Minutes
Etoposide (Toposar) 20.0 mg/mlPredicate: ≥ 240 Minutes≥ 240 Minutes
Fluorouracil 50.0 mg/mlPredicate: ≥ 240 Minutes≥ 240 Minutes
Methotrexate 25 mg/mlNot Tested by Predicate≥ 240 Minutes
Mitomycin C 0.5 mg/mlNot Tested by Predicate≥ 240 Minutes
Mixotantrone (2.0 mg/ml)Predicate: ≥ 240 MinutesNot Tested
Paclitaxel (Taxol) 6.0 mg/mlPredicate: ≥ 240 Minutes≥ 240 Minutes
Thio-Tepa 10.0 mg/mlPredicate: 24.4 Minutes43.7 Minutes
Vincristine Sulfate (Oncovin) 1.0 mg/mlPredicate: ≥ 240 Minutes≥ 240 Minutes
Physical Properties (ASTM D6319-10)
Tensile Strength (Before Aging)Predicate: ≥15 MPa, min≥17 MPa, min
Tensile Strength (After Aging)Predicate: ≥14 MPa, min≥14 MPa, min
Elongation (Before Aging)Predicate: 500% Min500% Min
Elongation (After Aging)Predicate: 400% Min400% Min
Freedom from Holes (ASTM D5151-06)G-1, AQL 2.5Complies with G-1, AQL 2.5
Residual PowderMax. 0.52mg per gloveMax. 0.52mg per glove
Biocompatibility (AAMI/ANSI/ISO 10993-10 & -05)
Skin IrritationNot a skin irritantNot a skin irritant
Skin SensitizationNot a skin sensitizerNot a skin sensitizer
CytotoxicityNot cytotoxicCytotoxic (Reported as "Non-Toxic" in the proposed device section of the biocompatibility, possibly a typo or referring to a specific outcome like not inducing significant cytotoxicity)

Notes on Biocompatibility: The "Cytotoxic / Non-Toxic" entry under the proposed device's biocompatibility in the table (page 5) is contradictory. Under the "Summary of Non-Clinical Testing" (page 8), it explicitly states that ISO 10993-05: Cytotoxicity was performed. However, the result in the comparison table is listed as "Cytotoxic / Non-Toxic." Assuming the test was conducted and the device is intended for medical use, it should generally be "Non-Toxic." For this response, I've noted the apparent contradiction.

2. Sample size used for the test set and the data provenance

The document does not specify the sample sizes for the various tests (e.g., number of gloves tested for chemotherapy permeation, physical properties, or biocompatibility). It only states that the tests were performed "per ASTM D6978-05 (Reapproved 2019)" and other standards.

The provenance (country of origin, retrospective/prospective) of the data is not mentioned. It is implied to be prospective testing conducted for the purpose of this 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to this type of device and study. The "ground truth" here is established through standardized laboratory testing (ASTM standards) of physical and chemical properties, not through expert human interpretation of medical images or clinical observations. Therefore, there are no medical experts establishing ground truth in the context of this submission. The "experts" would be the accredited laboratories and technicians conducting the ASTM tests.

4. Adjudication method for the test set

This information is not applicable. Adjudication methods like "2+1" or "3+1" are used in studies where there is subjective interpretation of data (e.g., medical images) by multiple human readers, and a consensus needs to be reached. For standardized laboratory tests of physical properties, the results are quantitative and objective, and thus adjudication is not necessary.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This submission is for medical gloves, which are a physical device, and does not involve AI or human readers for diagnostic or interpretive tasks.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable. This device is a physical product, not an algorithm.

7. The type of ground truth used

The ground truth for the device's performance is established by standardized laboratory test methods and their defined parameters. Specifically:

  • Chemotherapy Permeation: ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. The "ground truth" is the breakthrough time measured according to this standard.
  • Physical Properties: ASTM D6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application. The "ground truth" is the measured tensile strength, elongation, dimensions, etc., according to this standard.
  • Freedom from Holes: ASTM D5151-06 (Reapproved 2011) Standard Test Method for Detection of Holes in Medical Gloves.
  • Residual Powder: ASTM D6124-06 (Reapproved 2011) Standard Test Method for Residual Powder on Medical Gloves.
  • Biocompatibility: ISO 10993-10 (Primary Skin Irritation, Dermal Sensitization), and ISO 10993-05 (Cytotoxicity). The "ground truth" is the biological response observed according to these standards.

8. The sample size for the training set

This information is not applicable. There is no "training set" in the context of laboratory testing of a physical medical device. This term typically refers to machine learning algorithms.

9. How the ground truth for the training set was established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 18, 2020

Medline Industires, Inc. Adam Ostrower Regulatory Affairs Specialist Three Lakes Drive Northfield, Illinois 60093

Re: K201390

Trade/Device Name: Medline Powder-Free Light Blue Nitrile Exam Gloves (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: June 24, 2020 Received: June 25, 2020

Dear Adam Ostrower:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201390

Device Name

Medline Powder-Free Light Blue Nitrile Exam Gloves (Tested for Use with Chemotherapy)

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Carmustine (BCNU) 3.3 mg/ml 25.3 Minutes Cisplatin 1.0 mg/ml ≥ 240 Minutes Cyclophosphamide (Cytoxan) 20.0 mg/ml ≥ 240 Minutes Dacarbazine (DTIC) 10.0 mg/ml ≥ 240 Minutes Doxorubicin Hydrochloride 2.0 mg/ml ≥ 240 Minutes Etoposide (Toposar) 20.0 mg/ml ≥ 240 Minutes Fluorouracil 50.0 mg/ml 0.5 mg/ml ≥ 240 Minutes Methotrexate 25 mg/ml ≥ 240 Minutes Mitomycin C 0.5 mg/ml ≥ 240 Minutes Paclitaxel (Taxol) 6.0 mg/ml ≥ 240 Minutes Thio-Tepa 10.0 mg/ml 43.7 Minutes Vincristine Sulfate (Oncovin) 1.0 mg/ml ≥ 240 Minutes

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 25.3 Minutes Thio-Tepa 10.0 mg/ml 43.7 Minutes

Caution: Testing showed an average breakthrough time of 43.7 minutes with Thio-Tepa WARNING: Do not use with Carmustine

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, sans-serif font, stacked on top of a white, stylized starburst shape. The background is a solid dark blue color. The logo is simple and modern, and the colors are clean and professional.

Medline Industries, Inc.
Three Lakes Drive
Northfield, IL 60093

510(k) SUMMARY (K201390) [AS REQUIRED BY 21CFR807.92(c)]

Submitter / 510(k) Sponsor

Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093

Registration Number: 1417592

Contact Person

Adam Ostrower Regulatory Affairs Sr. Specialist Phone: 224-931-1176 Email: aostrower@medline.com

Summary Preparation Date September 18, 2020

Type of 510(k) Submission Traditional

Device Name / Classification

Trade Name: Medline Powder-Free Light Blue Nitrile Exam Glove (Tested for Use with Chemotherapy Drugs) Common Name: Patient Examination Glove, Specialty Classification Name: Patient Examination Glove, Specialty Product Code: LZA, LZC Classification Panel: General Hospital Regulatory Class: I Regulation Number: 21 CFR 880.6250

Predicate Device

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs – K172525

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image features the logo for Medline, a healthcare company. The logo is set against a blue square background. A white, stylized starburst or cross shape is prominently displayed, with the word "MEDLINE" in bold, white letters positioned horizontally across the center of the shape. A gray bar is visible on the right side of the image.

Medline Industries, Inc.
Three Lakes Drive
Northfield, IL 60093

Device Description

The Medline Powder-Free Light Blue Nitrile Exam Gloves (Tested for Use with Chemotherapy Drugs) are single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, light blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large.

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs, per ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

CHEMOTHERAPY DRUG TESTEDBREAKTHROUGH TIME (IN MINUTES)
Carmustine (BCNU)3.3 mg/ml (3,300 ppm)25.3
Cisplatin1.0 mg/ml (1,000 ppm)≥ 240
Cyclophosphamide (Cytoxan)20.0 mg/ml (20,000 ppm)≥ 240
Dacarbazine (DTIC)10.0 mg/ml (10,000 ppm)≥ 240
Doxorubicin Hydrochloride2.0 mg/ml (2,000 ppm)≥ 240
Etoposide (Toposar)20.0 mg/ml (20,000 ppm)≥ 240
Fluorouracil50.0 mg/ml (50,000 ppm)≥ 240
Methotrexate25 mg/ml (25,000 ppm)≥ 240
Mitomycin C0.5 mg/ml (500 ppm)≥ 240
Paclitaxel (Taxol)6.0 mg/ml (6,000 ppm)≥ 240
Thio-Tepa10.0 mg/ml (10,000 ppm)43.7
Vincristine Sulfate (Oncovin)1.0 mg/ml (1,000 ppm)≥ 240

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 25.3 Minutes Thio-Tepa 10.0 mg/ml 43.7 Minutes

CAUTION: Testing showed an average breakthrough time of 43.7 minutes with Thio-Tepa WARNING: Do not use with Carmustine

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in white, with a stylized star shape behind it. The star shape has four points, with the top and bottom points being longer than the side points.

Summary of Technological Characteristics

TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES

DEVICECHARACTERISTICSPROPOSED DEVICEPREDICATE DEVICECOMPARISONANALYSIS
Product NamePowder-Free Light Blue NitrileExam Gloves (Tested for Usewith Chemotherapy Drugs)Blue Non-Sterile Powder FreeNitrile Examination GlovesTested for Use withChemotherapy DrugsSimilar
510(k) ReferenceK201390K172525
Product OwnerMedline Industries, Inc.Central MedicareDifferent
Product CodeLZA, LZCLZA, LZCSame
Intended UseA patient examination glove is adisposable device intended formedical purposes that is wornon the examiner's hand toprevent contaminationbetween patient and examiner.These gloves were tested foruse with chemotherapy drugsas per ASTM D6978-05(2019)Standard Practice forAssessment of Resistance ofMedical Gloves to Permeationby Chemotherapy Drugs.Blue Non-Sterile Powder FreeNitrile Examination GlovesTested for Use withChemotherapy Drugs is adisposable device intendedfor medical purposes that isworn on the examiner's handto prevent contaminationbetween patient andexaminer.Same
Regulation Number21 CFR 880.625021 CFR 880.6250Same
MaterialsNitrileNitrileSame
ColorLight BlueBlueSimilar
SizesSmall, Medium, Large, ExtraLargeExtra-Small, Small, Medium,Large, Extra LargeSimilar
Dimensions - LengthComplies with ASTM D6319-10240mm min.Complies with ASTM D6319-10230mm min.Different
Dimensions - WidthComplies with ASTM D6319-10XS: N/AS: 85±10M: 95±10L: 105±10XL: 115±10Complies with ASTM D6319-10XS: 70±10S: 80±10M: 95±10L: 110±10XL: 120±10Similar
Dimensions -ThicknessComplies with ASTM D6319-10Palm - 0.14Finger - 0.16Cuff - 0.12Complies with ASTM D6319-10Palm - 0.06±10Finger - 0.08±10Cuff - 0.10±10Different
Physical PropertiesComplies with ASTM D6319-10minimum:Tensile Strength:Before Aging ≥17 MPa, minAfter Aging ≥14 MPa, minElongation:Before Aging 500% MinAfter Aging 400% MinTensile Strength:Before Aging ≥15 MPa, minAfter Aging ≥14 MPa, minElongation:Before Aging 500% MinAfter Aging 400% MinSimilar
Disposable vs. Non-DisposableDisposableDisposableSame
Freedom from HolesComplies with ASTM D6319-10and ASTM D5151-06 G-1, AQL2.5In accordance with ASTMD5151-06, following ASTMD6319 AQL 2.5/InspectionLevel G-1Same
Powder or Powder-FreePowder-FreePowder-FreeSame
Residual PowderMax. 0.52mg per gloveMax. 0.52mg per gloveSame
Contact DurationLimited 24 hoursLimited 24 hoursSame
BiocompatibilityAAMI/ANSI/ISO 10993-10:Not a skin irritantNot a skin sensitizerCytotoxicNon-ToxicNot a skin irritantNot a skin sensitizer
SterilityNon-SterileNon-SterileSame
Carmustine (BCNU) 3.3 mg/ml:25.3 MinutesCarmustine (3.3 mg/ml): 12.4MinutesDifferent
Cisplatin 1.0 mg/ml: ≥240MinutesCisplatin (1.0 mg/ml): ≥240MinutesSame
Cyclophosphamide (Cytoxan)20.0 mg/ml: ≥240 MinutesCyclophosphamide (20mg/ml) : ≥240 MinutesSame
Dacarbazine (DTIC) 10.0 mg/ml:≥240 MinutesDacarbazine (10.0 mg/ml):≥240 MinutesSame
Chemotherapy DrugsTested with MinimumBreakthroughDetection Time asTested per ASTM D6978Doxorubicin Hydrochloride 2.0mg/ml: ≥240 MinutesDoxorubicin HCI (2.0 mg/ml):≥240 MinutesSame
Etoposide (Toposar) 20.0mg/ml: ≥240 MinutesEtoposide (20.0 mg/ml): ≥240MinutesSame
Fluorouracil 50.0 mg/ml: ≥240MinutesFluorouracil (50.0 mh/ml):≥240 MinutesSame
Methotrexate 25 mg/ml: ≥240MinutesNot TestedDifferent
Mitomycin C 0.5 mg/ml: ≥ 240MinutesNot TestedDifferent
Not TestedMixotantrone (2.0 mg/ml):≥240 MinutesDifferent
Paclitaxel (Taxol) 6.0 mg/ml:≥240 MinutesPaclitaxel (6.0 mg/ml): ≥240MinutesSame
Thio-Tepa 10.0 mg/ml: 43.7MinutesThiotepa (10.0 mg/ml): 24.4MinutesDifferent
Vincristine Sulfate (Oncovin) 1.0mg/ml: ≥240 MinutesVincristine Sulfate (1.0mg/ml): ≥240 MinutesSame

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image features the Medline logo, which is set against a blue background. The logo consists of the word "MEDLINE" in a bold, sans-serif font, positioned to the left of a white, stylized starburst or cross-like symbol. The symbol is composed of four pointed shapes converging at a central point. A small portion of a gray rectangle is visible on the right edge of the image.


{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, sans-serif font, stacked on top of a white, stylized starburst shape. The background is a solid dark blue color. The logo is simple and modern, and the colors are clean and professional.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in a bold, sans-serif font, placed on the left side of the image. To the right of the word is a white, stylized starburst shape. The background is a solid dark blue color.

Summary of Non-Clinical Testing

The biocompatibility evaluation for Medline Powder-Free Light Blue (Tested for Use with Chemotherapy Drugs) was conducted in accordance with ANSI/AAMI/ISO 10993-1:2009 Biological Evaluation of Medical Devices – Part 1: Evaluation within a Risk Management Process, as recognized by FDA.

The following tests were performed to evaluate the biocompatibility of the Medline Powder-Free Light Blue Nitrile Exam Glove (Tested for Use with Chemotherapy Drugs):

  • ISO 10993-10: Primary Skin Irritation
  • ISO 10993-10: Dermal Sensitization
  • ISO 10993-05: Cytotoxicity

Performance Testing (Bench)

Physical performance qualities of the proposed device were evaluated per ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs.

In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • ASTM D 6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application
  • ASTM D 6124-06 (Reapproved 2011) Standard Test Method for Residual Powder on Medical Gloves
  • ASTM D 5151-06 (Reapproved 2011) Standard Test Method for Detection of Holes in Medical Gloves
  • ASTM D 6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Summary of Clinical Testing

No clinical study is included in this submission.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image features the logo of Medline, a healthcare company. The logo is set against a blue background and consists of the word "MEDLINE" in bold, sans-serif font, positioned to the left of a stylized white symbol that resembles a compass or a star. The symbol has four points, with the top and bottom points being longer than the side points. A small gray rectangle is visible in the upper right corner of the image.

Conclusion

In accordance with 21 CFR Part 807, and based on the nonclinical information provided in this premarket notification Medline Industries, Inc. concludes that Medline Powder-Free Light Blue Nitrile Exam Glove (Tested for Use with Chemotherapy Drugs) is as safe, as effective, and performs as well as or better for its intended use as the legally marketed predicate device, Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs, K172525.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.