Intended Use

Intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

USA Gloves Nitrile Gloves and aforementioned predicate device are Non-powdered patient examination glove, as defined in 21 CFR 880.6250. Former Release Powder or Chemical: No release powder or chemical is used. USA Gloves Nitrile Gloves is not intended to be sterilized. The device is delivered in packaging which may be used as a protective carrying case.

AI/ML Overview

The document provided is a 510(k) Premarket Notification from the FDA for USA Gloves Nitrile Powder Free Exam Gloves (K211624). This type of document is for medical devices and focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel efficacy through clinical studies with "AI" components or human readers improving with AI assistance. As such, many of the requested categories (2-9) are not applicable or cannot be extracted from this specific type of regulatory submission.

Here's the information that can be extracted and a clear indication where the requested information is not applicable:

1. A table of acceptance criteria and the reported device performance

Test MethodologyAcceptance CriteriaReported Device Performance
ASTMD 6319-10 (Reapproved 2015)Physical Properties: Before Aging Tensile Strength min 14 Mpa, Ultimate Elongation Min 500%; After Aging Tensile Strength min 14 Mpa, Ultimate Elongation Min 400%Pass
ASTMD 6319-10 (Reapproved 2015)Dimensions: Length Min 230 mm, Width Min 95+/-10mm (for medium size)Pass
ASTMD 6319-10 (Reapproved 2015)Thickness: Palm min 0.05 mm, Finger min 0.05 mmPass
ASTMD 6319-10 (Reapproved 2015)Powder Residue: < 2 mg/glovePass
ISO 10993-5Biocompatibility - Cytotoxicity: No CytotoxicityPass
ISO 10993-10Skin Irritation: No Skin IrritationPass
ISO 10993-11Acute Systemic Toxicity: No Systemic ToxicityPass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample sizes used for each test. The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective, beyond the fact that these are nonclinical tests conducted to standard methodologies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is a medical glove, and "ground truth" in the context of expert review for image analysis or diagnostic performance is not relevant. The testing involves standardized laboratory methods for physical and biocompatibility properties.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are typically used in clinical studies or expert reviews. The tests performed are objective, standardized laboratory measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a medical glove, not an AI-powered diagnostic tool. There were no MRMC studies or AI involvement mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a medical glove, not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the nonclinical tests is established by the specified international standards (ASTM and ISO). For example, physical properties are compared against defined thresholds (e.g., tensile strength min 14 Mpa), and biocompatibility tests look for the absence of specific biological responses (e.g., no cytotoxicity, no skin irritation). These are objective, measurable criteria defined by the standards.

8. The sample size for the training set

Not applicable. This document describes the testing of a manufactured physical device (medical glove), not the development of an algorithm or AI model that requires a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 18, 2021

USA Gloves Zishan Momin Chief Operating Office 12505 Reed Rd Ste 110 Sugar Land, Texas 77478

Re: K211624

Trade/Device Name: USA Gloves Nitrile Powder Free Exam Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: November 16, 2021 Received: November 16, 2021

Dear Zishan Momin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211624

Device Name

USA Gloves Nitrile Powder Free Exam Gloves

Indications for Use (Describe)

Intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for USA Gloves. The logo features a blue hand with a white star on the palm, and red stripes on the fingers, resembling the American flag. Below the hand is the text "USA GLOVES" in blue, and below that is the text "AMERICAN MADE NITRILE GLOVES" in red.

May 16, 2021

Zishan Momin Tel: (281) 827-3691 Email: sharez@usagloves.com

510(K) Summary - K211624

Submitter Information

FDA 510(K) K211624
Product/Trade NameUSA Gloves Nitrile Powder Free Exam Gloves
510(K)K211624
Applicant Name:Glove Ventures LLCdba USA Gloves
Applicant Address (city, state, zip):Stafford, TX 77477
Applicant Contact Person:Zishan Momin
Correspondent Name:Zishan Momin
Correspondent Address (full address):12505 Reed Rd Ste 110Sugar Land, TX 77478USA
Correspondent Contact Person:Zishan Momin
Correspondent Contact Phone#:
Correspondent Contact Email:zishan@usagloves.com

Predicate Devices

The USA Gloves Nitrile Powder Free Exam Gloves is identical to the Hi-Care Thai Gloves Co. Ltd. Blue Nitrile Examination Gloves Powder free which is the subject of a clearance letter from the Agency under 510(k) number K202384, which itself, was predicated on the subject of K192333. The predicate Hi-Care Thai Gloves Co. Ltd. Blue Nitrile Examination Gloves Powder free and USA Gloves Global Nitrile Gloves are the same in design and use.

Indications for Use:

USA Gloves Nitrile Powder Free Exam Gloves is intended as a powder-free patient examination glove and is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The USA Gloves Nitrile Powder Free Exam Gloves is a single use, disposable device, provided non-sterile.

{4}------------------------------------------------

Device Description:

USA Gloves Nitrile Gloves and aforementioned predicate device are Non-powdered patient examination glove, as defined in 21 CFR 880.6250.

Former Release Powder or Chemical: No release powder or chemical is used.

USA Gloves Nitrile Gloves is not intended to be sterilized. The device is delivered in packaging which may be used as a protective carrying case.

CHARACTERSTICSSTANDARDSDEVICE PERFORMANCEComparison
SubjectPredict
510(K) NumberK211624K202384----
Name of deviceUSA Gloves NitrilePowder Free ExamGloveBlue NitrileExamination GlovesPowder free----
Product CodeLZALZA----
DimensionsASTMD 6319-10(Reapproved 2015)Length Min 230 mWidth Min 95+/-10mm(for medium size)Length Min 230 mmWidth Min 95+/mm(for medium size)Same
Physical PropertiesASTMD 6319-10(Reapproved 2015)Before AgingTensile Strength min14 MpaUltimate ElongationMin 500%After AgingTensile Strengthmin 14 MpaUltimateElongationMin 400%Before AgingTensile Strength min14 MpaUltimate ElongationMin 500%After AgingTensile Strength min14 MpaUltimate ElongationMin 400%Same
ThicknessASTMD6319-10(Reapproved 2015)Palm min 0.05 mm Fingermin 0.05 mmPalm min 0.05 mmFinger min 0.05 mmSame
Powder ResidueASTMD 6319-10<2 mg/glove<2 mg/gloveSame
Primary SkinIrritation-ISO10993-10:2010(E)Under the condition ofstudy not an irritantUnder the conditionof study not anirritantSame

Technological Characteristics Comparison Table

Table1: USA Gloves Nitrile Gloves / the Hi-Care Thai Gloves Co. Ltd. Blue Nitrile Examination Gloves Powder Free Comparison

{5}------------------------------------------------

Summary of Nonclinical Testing

Provided below is a summary of the standards and test methodology that was used to evaluate and demonstrate that the subject device met the performance specification and acceptance criteria for a nitrile glove.

Test MethodologyPurposeAcceptance CriteriaResults
ASTMD 6319-10(Reapproved 2015)Physical PropertiesBefore Aging Tensile Strength min14 Mpa Ultimate Elongation Min500% After Aging Tensile Strengthmin 14 Mpa Ultimate ElongationMin 400%Pass
ASTMD 6319-10(Reapproved 2015)DimensionsLength Min 230 m Width Min95+/-10mm (for medium size)Pass
ASTMD 6319-10(Reapproved 2015)ThicknessPalm min 0.05 mm Finger min 0.05mmPass
ASTMD 6319-10(Reapproved 2015)Powder Residue$<$ 2 mg/glovePass
ISO 10993-5Biocompatibility-CytotoxicityNo CytotoxicityPass
ISO 10993-10Skin IrritationNo Skin IrritationPass
ISO 10993-11Acute SystemicToxicityNo Systemic ToxicityPass

Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.