K Number
K203359
Device Name
BD Flu+ Syringe
Date Cleared
2021-07-28

(254 days)

Product Code
Regulation Number
880.5860
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD TM Flu+ Syringe is intended for aspiration and injection of fluids.

Device Description

The BD Flu+ Syringe is a sterile two-piece single use, sterile syringe with an integral needle. It allows for a variable dose up to 1 ml to be aspirated and injected. It is intended for general-purpose aspiration and injection of fluids from a vial or ampoule. The BD Flu+ syringe has been designed for low dead space to reduce medication waste. The BD Flu+ Syringe is a 1.0mL maximum dosage with 0.5mL and 1.0mL barrel marking and 0.25mL incremental markings. The Flu+ syringe is assembled with a pre attached needle in the following gauges and sizes: 23G (0.6mm) x 1 inch (25mm), 25G (0.5mm) x 5/8 inch (16mm.), 25G (0.5mm) x 1 inch (25mm).

AI/ML Overview

The device in question is the BD™ Flu+ Syringe. This submission is a 510(k) for an updated indication for use, stating it is "intended for aspiration and injection of fluids." The acceptance criteria and the studies that prove the device meets these criteria are outlined below:

1. Table of Acceptance Criteria and Reported Device Performance

Test CategoryAcceptance Criteria (Standard / Requirement)Reported Device Performance
Bench Performance
LeakageISO 7886-1:2017 (Syringes for medical use - Part 1: Syringes for manual use)Met acceptance criteria. (Implied to be same as predicate device, which also met ISO 7886-1)
Dose AccuracyISO 7886-1:2017Met acceptance criteria. (Implied to be same as predicate device, which also met ISO 7886-1)
Dead SpaceISO 7886-1:2017Met acceptance criteria. (Implied to be same as predicate device, which also met ISO 7886-1)
Activation ForcesISO 7886-1:2017Met acceptance criteria. (Implied to be same as predicate device, which also met ISO 7886-1)
Cannula Pull ForceISO 7864:2016 (Sterile hypodermic needles for single use)Met acceptance criteria. (Implied to be same as predicate device, which also met ISO 7864)
Shield Pull ForceBD Internal RequirementMet acceptance criteria. (Implied to be same as predicate device, which also met BD Internal Requirement)
Scale Mark PermanencyBD Internal RequirementMet acceptance criteria. (Implied to be same as predicate device, which also met BD Internal Requirement)
Package IntegrityISO 11607-1:2010 (Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems)Met acceptance criteria during testing after environmental conditioning and simulated transportation.
Particulate MatterUSP <788> Particulate Matter in InjectionsMet USP acceptance criteria.
Biocompatibility
CytotoxicityISO 10993-5 (Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity) & USP <87> (Biological Reactivity Tests, In Vitro)Non-cytotoxic. (Implied to be same as predicate device, which also met these standards).
SensitizationISO 10993-10 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization) & ASTM F2148 (Method for Assessment of Local Lymph Node Assay for Allergic Reaction) / Maximization: ISO 10993-11 (Biological evaluation of medical devices - Part 11: Tests for systemic toxicity)Non-sensitizer. (Implied to be same as predicate device, which also met these standards).
Intracutaneous Reactivity/IrritationISO 10993-10 & USP <88> (Biological Reactivity Tests, In Vivo - Intracutaneous Reactivity)Non-irritant. (Implied to be same as predicate device, which also met these standards).
Primary Dermal IrritationISO 10993-10Met acceptance criteria. (Implied to be same as predicate device, which also met this standard).
Acute Systemic ToxicityISO 10993-11 & USP <88>Non-toxic. (Implied to be same as predicate device, which also met these standards).
Material-Mediated PyrogenicityISO 10993-11:2017 & USP <151> (Pyrogen Test)Non-pyrogenic. (Implied to be same as predicate device, which also met these standards).
Extractables/LeachablesISO 10993-18:2005 (Biological evaluation of medical devices - Part 18: Chemical characterization of materials)Acceptable. (Implied to be same as predicate device, which also met this standard).
HemolysisISO 10993-4 (Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood)Non-hemolytic. (Implied to be same as predicate device, which also met this standard).
Sterility
Sterilization MethodEthylene Oxide (ETO) Sterilization per Standard EN ISO 11135-1 (Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices). Product meets EN 556 (Sterilization of Medical Devices - Requirements for Medical Devices to be Labelled Sterile) requirements.Validation conducted in accordance with EN ISO 11135-1. The sterilized product meets EN 556.
Sterilization ResidualsMaximum EO mg/device limit: 4mg/device. Maximum ECH mg/device Limit: 9mg/device. Acceptable per ISO 10993-7:2008 (Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals).Residuals are within acceptable limits per ISO 10993-7:2008.
Pyrogenicity / Endotoxin TestingUSP 40-NF 35 monographs 85 (Bacterial Endotoxins Test) and 161 (Medical Devices Bacterial Endotoxin and Pyrogen Test)Test methods per USP 40-NF 35 monographs 85 and 161 were conducted, implying results met acceptance criteria.
Shelf-LifeASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices for a 5-year shelf life.5-year shelf life validated using ASTM F1980-16. Device continues to meet all predetermined acceptance criteria for performance tests over this period.

Study Details:

The primary study type reported is bench testing and biocompatibility testing. The submission emphasizes that modifications of the subject device's intended use does not affect the results of this analysis, suggesting that the underlying performance characteristics tested are largely identical to the predicate device.

2. Sample Size for the Test Set and Data Provenance

  • Test Set Sample Size: The document does not explicitly state the specific sample sizes used for each individual bench or biocompatibility test. However, it indicates these tests were performed by BD on the BD™ Flu+ Syringe.
  • Data Provenance: The data is generated from bench testing and biocompatibility testing conducted by the manufacturer, Becton, Dickinson and Company (BD). It does not involve patient data. The context implies these tests were performed internally by BD, likely in the US, given the submission to the FDA. The nature of these tests (bench and biocompatibility) means they are prospective in the sense that they are conducted specifically for this device and submission.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: This information is not applicable. The device's performance is evaluated against recognized international standards (ISO, ASTM) and USP monographs, as well as internal BD requirements, not against expert consensus on a test set.
  • Qualifications of Experts: Not applicable for establishing ground truth in this context. The "ground truth" is defined by the objective performance criteria outlined in the referenced standards and internal requirements.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. The tests performed are objective, measurable performance characteristics evaluated against predetermined pass/fail criteria from established standards. There is no subjective interpretation requiring an adjudication method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No, an MRMC comparative effectiveness study was not performed. This type of study is relevant for diagnostic devices that involve human interpretation of images or other data, where the AI assists human readers. The BD™ Flu+ Syringe is a physical medical device (syringe) and does not involve human readers in its operational use that could be augmented by AI.

6. Standalone Performance Study (Algorithm Only)

  • Standalone Performance: Not applicable. This device is a physical product and does not incorporate an algorithm or AI component. Therefore, no standalone algorithm-only performance study was conducted.

7. Type of Ground Truth Used

  • Ground Truth Type:
    • Objective Performance Standards: The ground truth for the device's performance is established by objective, quantitative performance standards defined by international organizations (ISO, ASTM) and national pharmacopeias (USP), as well as internal engineering specifications (BD Internal Requirement). Examples include specific limits for leakage, dose accuracy, particulate matter concentration, and biological responses in biocompatibility tests.

8. Sample Size for the Training Set

  • Training Set Sample Size: Not applicable. This device does not use machine learning or AI, and therefore no "training set" is required or used in its development or validation.

9. How the Ground Truth for the Training Set was Established

  • Ground Truth for Training Set: Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym in a blue square and the full name "U.S. Food & Drug Administration" in blue text.

July 28, 2021 Becton, Dickinson and Company Nikita Kumar Senior Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417

Re: K203359

Trade/Device Name: BD Flu+ Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: June 26, 2021 Received: June 30, 2021

Dear Nikita Kumar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203359

Device Name BD™ Flu+ Syringe

Indications for Use (Describe)
The BD TM Flu+ Syringe is intended for aspiration and injection of fluids.
Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for BD, a global medical technology company. The logo consists of two parts: a circular symbol on the left and the letters "BD" on the right. The circular symbol is orange and features a stylized image of a person with outstretched arms inside of a circle. The letters "BD" are in blue and are written in a bold, sans-serif font.

K203359 510(k) Summary

BD™ Flu+ Syringe

SubmitterInformationSubmitter Name:Becton, Dickinson and Company
Submitter Address:1 Becton DriveFranklin Lakes, NJ 07417
Contact Person:Nikita Abirami Mahendra KumarSenior Regulatory Affairs Specialist
Email Address:Nikita.Abirami.Mahendra.Kumar@bd.com
Phone Number:(201) 847-5641
Date of Preparation:July 28, 2021
Subject DeviceTrade Name:BD™ Flu+ Syringe
Common Name:Piston Syringe
Regulation Number:21 CFG 880.5860
Regulation Name:Piston Syringe
Regulatory Class:Class II device
Product Code:FMF
Classification Panel:General Hospital
Predicate DeviceTrade Name:BD™ FLU+ Syringe
510(k) Reference:K091377
Common Name:Piston Syringe
Regulation Number:21 CFR 880.580
Regulation Name:Piston Syringe
Regulatory Class:Class II Device
Product Code:FMF
Classification Panel:General Hospital
Reason forSubmissionThe basis of this submission is to modify the indications for useof the BD™ Flu+ Syringe.
Device DescriptionThe BD Flu+ Syringe is a sterile two-piece single use, sterilesyringe with an integral needle. It allows for a variable dose upto 1 ml to be aspirated and injected. It is intended for general-purpose aspiration and injection of fluids from a vial or ampoule.The BD Flu+ syringe has been designed for low dead space toreduce medication waste.
The BD Flu+ Syringe is a 1.0mL maximum dosage with 0.5mLand 1.0mL barrel marking and 0.25mL incremental markings.The Flu+ syringe is assembled with a pre attached needle in thefollowing gauges and sizes:
23G (0.6mm) x 1 inch (25mm)
25G (0.5mm) x 5/8 inch (16mm.)
Indications for UseThe BD™ Flu+ Syringe is intended for aspiration and injection of fluids.
TechnologicalCharacteristicsThe subject devices are equivalent to the predicate devices in intended use, materials and performance characteristics.

{4}------------------------------------------------

25G (0.5mm) x 1 inch (25mm)

Element ofComparisonSubject Device:BD™ Flu+ SyringePredicate DeviceBD™ Flu+ SyringeComparison
510(k) NumberK203359K091377N/A
Intended UseIntended foraspiration andinjection of fluidsIntended foraspiration andinjection of theinfluenza vaccineDifferent; Subjectdevice is intendedfor general use
Syringe Type2 Piece (barrel andplunger)2 Piece (barrel andplunger)Same
Dose SavingFeatureLow DeadSpace/VolumeLow DeadSpace/VolumeSame
IntegratedNeedleYesYesSame
DoseSetting/VolumesVariable dose scalemarkings with1.0mL max deliveryVariable dose scalemarkings with1.0mL max deliverySame
Needle Size(Gauge x Length)23G x 1"25G x 1"25G x 5/8"23G x 1"25G x 1"25G x 5/8"Same
Barrel MaterialPlasticPlasticSame
Plunger MaterialPlastic+Colorant(Blue, Orange)Plastic+Colorant(Blue, Orange)Same
IntegratedCannulaStainless SteelStainless SteelSame
ShieldPlasticPlasticSame
AdhesiveEpoxyEpoxySame
CannulaLubricantSiliconeSiliconeSame
SterilizationEthylene Oxide (SAL10-6)Ethylene Oxide (SAL10-6)Same
Shelf Life5 years5 yearsSame
Functional Testing:
LeakageISO 7886-1ISO 7886-1Same
Dose AccuracyISO 7886-1ISO 7886-1Same
Dead SpaceISO 7886-1ISO 7886-1Same
Activation ForcesISO 7886-1ISO 7886-1Same
Cannula Pull ForceISO 7864ISO 7864Same
Shield Pull ForceBD Internal RequirementBD Internal RequirementSame
Scale Mark PermanencyBD Internal RequirementBD Internal RequirementSame
Biocompatibility Testing:
Testing per ISO 10993:2018
CytotoxicityISO 10993-5 & USP <87>, Non-cytotoxic• LLNA: ISO 10993-10 & ASTM F2148ISO 10993-5 & USP <87>, Non-cytotoxic• LLNA: ISO 10993-10 & ASTM F2148Same
Sensitization• Maximization: ISO 10993-11Non-sensitizer• Maximization: ISO 10993-11Non-sensitizerSame
Intracutaneous ReactivityPer ISO 10993-10 & USP<88>, Non-irritantPer ISO 10993-10 & USP<88>, Non-irritantSame
Primary Dermal IrritationISO 10993-10ISO 10993-10Same
Acute Systemic ToxicityISO 10993-11 & USP<88>, Non-toxicISO 10993-11 & USP<88>, Non-toxicSame
PyrogenicityPer ISO 10993-11:2017 & USP<151>, Non-pyrogenicPer ISO 10993-11:2017 & USP<151>, Non-pyrogenicSame
Extractables/LeachablesPer ISO 10993-18:2005, AcceptablePer ISO 10993-18:2005, AcceptableSame
HemolysisPer ISO 10993-4, Non-hemolyticPer ISO 10993-4, Non-hemolyticSame

{5}------------------------------------------------

Performance Tests

BD has performed the following bench and biocompatibility testing on BD™ Flu+ Syringe. Modifications of the subject device's intended use does not affect the results of this analysis.

Bench Performance:

  • Leakage (ISO 7886-1:2017) .
  • Dose Accuracy (ISO 7886-1:2017) .

Confidential & Proprietary Page 3 of 5

{6}------------------------------------------------

  • Dead Space (ISO 7886-1:2017) ●
  • Activation Forces (ISO 7886-1:2017)
  • Cannula Pull Force (ISO 7864:2016) ●
  • Shield Pull Force (BD Internal Requirement)
  • Scale Mark Permanency (BD Internal Requirement) ●
  • Package Integrity (ISO 11607-1:2010). ●

Material Biocompatibility Performance:

In accordance with ISO 10993-1, the Flu+ Syringe is classified as: Externally Communicating Device, Blood Path Indirect, Limited Contact (<24 hours). The following testing was conducted:

  • Cytotoxicity ●
  • Sensitization
  • Irritation
  • Acute Systemic Toxicity
  • Material-Mediated Pyrogenicity
  • . Hemocompatibility

Particulate matter testing was conducted in accordance with USP<788> Particulate Matter in Injections and met the USP acceptance criteria.

Sterility, Shipping and Shelf-Life

SterilizationMethodEthylene Oxide (ETO)
SterilizationResidualsMaximum EO mg/device limit:4mg/deviceMaximum ECH mg/device Limit:9mg/deviceThe above limits are acceptable perISO 10993-7:2008
ValidationMethodThe sterilization process is validatedin accordance with Standard EN ISO11135-1."Sterilization of health careproducts-Ethylene oxide- Part 1:Requirements for development,validation and routine control of asterilization process for medicaldevices".The sterilized product meets the

{7}------------------------------------------------

requirements of EN 556 "Sterilizationof Medical Devices - Requirements forMedical Devices to be Labelled Sterile".
Pyrogenicity/EndotoxinTestingTest methods are per USP 40-NF 35monographs 85 (Bacterial EndotoxinsTest) and 161 (Medical DevicesBacterial Endotoxin and PyrogensTest).
Packaging integrity testing, after environmental conditioning aresimulated transportation in accordance with ISO 11607-1:2010and 11135:2014, was conducted on the final packaged, andsterile device. All packaging deemed acceptable for protection ofproduct and sterility maintenance.
Shelf Life of 5 years is validated using the FDA recognizedstandards ASTM F1980-16 Standard Guide for Accelerated Agingof Sterile Barrier Systems for Medical Devices.
The subject device continue to meet all the predeterminedacceptance criteria for the above-listed performance tests,demonstrating substantial equivalence to the predicate device.
Clinical TestingClinical testing was not required for this submission
Summary ofSubstantialEquivalenceThe BD™ Flu+ Syringe is substantially equivalent to thepredicate device in principles of operation, technology, design,

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).